9
6. Calibration
6.1. The device is pre-calibrated at the factory for operation with the glass tubes 16 mm
in external diameter (see the label on the bottom side of the unit) at temperature
range from +15°C to +25°C and saves calibration data when being switched off.
6.2. Before using the standards. The standard suspensions must be sufficiently
resuspended. To do this, proceed as follows:
- Mix adequately at the highest RPM setting of your vortex mixer device, i.e. Biosan
V-1 plus.
- Allow the tubes to stand for 2 to 5 minutes to allow all air bubbles to rise.
- Invert the tubes carefully before comparison. Avoid bubble formation during
operation.
6.3. Perform calibration from lower to higher calibration value. Use at least 2 points for
calibration. Different calibration points are available:
For DEN-1, calibration points are 0.5, 1.0, 2.0, 3.0, 4.0 and 5.0.
For DEN-1B, calibration points are 0.00, 0.50, 1.00, 2.00, 3.00, 4.00, 5.00 and 6.00.
6.4. Connect the external power supply to electric circuit. Switch on the unit using the
Power switch (fig. 2/1) on the rear panel.
Note. For DEN-1B, connecting external power supply is optional when the
batteries are in use.
Attention! Make sure that the tube socket is empty.
6.5. Press the Select button (fig. 2/3) on the rear of the unit.
Note. Use a thin pin of maximum diameter of 2 mm for pressing the Select
and Install buttons.
6.6. To calibrate DEN-1B model, it is necessary to set values for an empty socket -.-- and
transparent standard 0.00.
6.6.1. Empty socket. Display shows -.-- indication. Press Install button (fig. 2/4) to save
empty socket value. Display shows next required calibration value.
6.6.2. Transparent standard. Display shows 0.00 indication. Insert standard for 0.00 value
into the socket (fig. 3/1) of the unit. If the standard for 0.00 value is not available, fill
the tube (of the kind that is used for operations) with distilled water. Use the tube as
the 0.00 value standard. Press Install button to save empty socket value. Display
shows next required calibration value.
Note! Calibrate the unit using as many points as possible to obtain precise
results. Minimum requirement are 2 points closest to the working
range limits (e.g. 0.00 and 6.00 for operation in 0.00 - 6.00 McF
range).