Delta T® Heated Culture Dish Cover Instructions
Description:
The Bioptechs Delta T® Heated Lid is a device, which will provide a condensate free optical surface
on the top of a Delta T® Dish through which specimens can be trans-illuminated on an inverted
microscope. It also provides a port for the introduction of gas above the specimen.
Instructions:
When preparing to use the heated lid, it should be pre-warmed by either placing it in an incubator
or connecting it to the heated lid output socket in the back of the controller for about 15 minutes.
Connection is made to the Heated Lid through a three pin 0.1 pin header. Both of the outer pins are
earth ground, the center pin carries current. Therefore, orientation of the connector does not matter.
The controller end of the connector wire is a standard 5.5mm DC coaxial plug.
After the Heated Lids is warmed, it can be mounted to the Delta TÆ dish by applying a combination of
downward and slight rotational force to the lid in such a manner to compress the o-ring seal to seat the
lid on the dish. Removal also requires a slight rotating motion. It is powered by a 2.5-volt adjustable
power supply located on the back of the Delta TÆ4 Controller. The voltage should be adjusted to a point
where the temperature of the glass surface of the lid prevents condensation in the dish. The controller
is factory set to 2.5 volts. However, It may be necessary to adjust the voltage depending on ambient
conditions. An adjustment screw for this setting is provided adjacent to the Heated Lid Socket on the
back of the Delta T4 controller. Alternate power supplies can be used including a DC battery or an AC
adapter.
Note: A heat lter set is recommended in the transmitted light path. (Reection lter before absorption
lter).
Cleaning:
The heating lid can be cleaned with mild laboratory soap and water, rinsed and dried. If sterilization is a concern
it can be placed in an autoclave bag and autoclaved on a short cycle of about 15 minutes. Exercise extreme care
in handling.
Note: After autoclaving it must be thoroughly dry on the outside surface before use.
Important!
Do not apply electrical power to the lid if liquids are present on the coated surface of the lid (same
surface as contacts). Make sure the lid is dry before use.