3560 Beck Road • Butler, PA 16002 • Voice: 724-282-7145 • Fax: 724-282-0745 • E-mail: info@bioptechs.com • Web: www.bioptechs.com
Delta T Cooling Ring Instructions
Description:
The Bioptechs cooling ring is an immersion device, which provides a thermally conductive physical barrier between
chilled uid passing through the ring and the uid surrounding the specimen thus absorbing heat from the specimen. This
cooling ring is made of autoclavable 304 stainless steel and provides the microscopist with a convenient and inexpensive
method of reducing the temperature of specimens in Delta T Culture Dishes. The cooling ring is supported on the stage
adapter and translates along with the dish. It is easily ipped up and out of the way to enable easy exchange of dishes in
the stage adapter.
Instructions:
1. Mount an empty Delta T Dish into the stage adapter.
2. Press t the 1/8” end of the hinged cooling ring adapter into the holes provided adjacent to the rectangular
opening on the stage adapter.
3. Using the 5/64” Allen driver provided, adjust the immersion ring to t into the dish and ip up and down as
needed to replace dishes.
4. Adjust the tension of the hinge with the 5/64” wrench.
5. Press 1/16” Tygon tubing onto the 14 gauge tube and attach to a chilled uid source and drain.
Note: A simple Dewar of ice water above the stage, Tygon tubing, and a pinch clamp will reduce the
temperature in the dish to 4 - 10°C. It is simplistic, inexpensive and it works!
6. When you need temperatures above ambient simply stop the chilled water ow and turn on the Delta T
Controller.
3560 Beck Road • Butler, PA 16002 • Voice: 724-282-7145 • Fax: 724-282-0745 • E-mail: info@bioptechs.com • Web: www.bioptechs.com