BIO PROTECH MAXEMS 2000, MAXEMS 1000 Instruction Manual

PRODUCT CONFIGURATION
EMS Unit x 1 Electrodes x 4 Leadwire x 2 9V Battery, type 6F22 x 1 Instruction Manual x 1 Carring Case x 1
CAUTION Federal Law (U.S.A) restricts this device to sale by or on the order of a physician.
INSTRUCTION MANUAL
MAXEMS 2000
Manufactured by Bio Protech Inc.
Donghwa Medical Instrument complex, 151-3, Donghwagongdan-ro, Munmak-eup, Wonju-si, Gangwon-do, 26365 Korea Tel: 82-33-735-7720 , Fax : 82-33-735-7736
Made in China. Rev.005 05/2017
INDEX
1. INTRODUCTION
2. CONTRAINDICATION
3. WARNINGS
4. PRECAUTIONS
5. ADVERSE REACTIONS
6. GENERAL DESCRIPTION
7. CONSTRUCTION
8. TECHNICAL SPECIFICATION
9. ACCESSORIES
10. GRAPHIC SYMBOLS
11. ATTACHMENT OF ELECTRODE LEAD WIRES
12. LEAD WIRE MAINTENANCE
13. ELECTRODE OPTIONS
14. ELECTRODE PLACEMENT
15. TIP FOR SKIN CARE
16. APPLICATION OF RE-USABLE SELF ADHESIVE ELECTRODES
17. ADJUSTING THE CONTROLS
18. INSTRUCTION FOR USE
19. MAINTENANCE, TRANSPORTATION AND STORAGE OF MAXEMS 2000
20. SAFETY-TECHNICAL CONTROLS
21. MALFUNCTIONS
22. WARRANTY
1
2
2
2
3
3
4
4
5
6
6
7
7
7
8
8
9
10
12
14
15
16
16
Chapter 1: INTRODUCTION
EXPLANATION OF EMS
Electrical Muscle Stimulation is an internationally accepted and proven way of treating muscular injuries. It works by sending electronic pulses to the muscle needing treatment: this causes the muscle to exercise passively.
EMS has low frequency and this, in conjunction with the square wave pattern, allows direct work on muscle groupings. This is being widely used in hospitals and sports clinics for the treatment of muscular injuries and for the re-education of paralyzed muscles, to prevent atrophy in affected muscles and improving muscle tone and blood circulation.
INTENDED USE
EMS Unit is applicable to patients diagnosed with wide range of orthopedic and neurologic deficiencies who demonstrate difficulties during voluntary movements. This is a prescription device and should be used under continued medical supervision. And it is intended use as below:
1. Relaxation of muscle spasms.
2. Prevention or retardation of disuse atrophy.
3. Increasing local blood circulation.
4. Muscle re-education.
5. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis.
6. Maintaining or increasing range of motion.
Chapter 2: CONTRAINDICATION
Powered muscle stimulators should not be used on patients with cardiac demand pacemakers.
Chapter 3: WARNINGS
1. The long-term effects of chronic electrical stimulation are unknown.
2. Stimulation should not be applied over the carotid sinus nerves, particularly in patients with a known sensitivity to the carotid sinus reflex.
3. Stimulation should not be applied over the neck or mouth. Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing.
4. Stimulation should not be applied transthoracically in that the introduction of electrical current into the heart may cause cardiac arrhythmias.
5. Stimulation should not be applied transcerebrally.
2
6. Stimulation should not be applied over swollen, infected, or inflamed areas or skin eruptions, e.g., phlebitis, thrombophlebitis, varicose veins, etc.
7. Stimulation should not be applied over, or in proximity to, cancerous lesions.
8. Remove batteries if the unit is not used for a long period of time.
9.Keep the unit away from sources of high magnetic fields such as TV’s, microwave ovens and hi-fi speakers, as these may affect the LCD screen.
10. Do not apply stimulation before attaching electrodes on the skin.
11. During stimulation, do not touch the patient and treatment site at a time
12. Do not apply stimulation in the presence of electronic monitoring equipment (e.g., cardiac monitors, ECG alarms), which may not operate properly when the electrical stimulation device is in use.
Chapter 4: PRECAUTIONS
1. Safety of powered muscle stimulators for use during pregnancy has not been established.
2. Caution should be used for patients with suspected or diagnosed heart problems.
3. Caution should be used for patients with suspected or diagnosed epilepsy.
4. Caution should be used in the presence of the following: a) When there is a tendency to hemorrhage following acute trauma or fracture; b) Following recent surgical procedures when muscle contraction may disrupt the healing process; c) Over the menstruating or pregnant uterus; and d) Over areas of the skin which lack normal sensation.
5. Some patients may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical conductive medium. The irritation can usually be reduced by using an alternate conductive medium or alternate electrode placement.
6. Electrode placement and stimulation settings should be based on the guidance of the prescribing practitioner.
7. Powered muscle stimulators should be kept out of the reach of children.
8. Powered muscle stimulators should be used only with the leads and electrodes recommended for use by the manufacturer.
9. Portable powered muscle stimulators should not be used while driving, operating machinery, or during any activity in which involuntary muscle contractions may put the user at undue risk of injury.
Chapter 5: ADVERSE REACTIONS
Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators.
3
Chapter 6: GENERAL DESCRIPTION
Chapter 8: TECHNICAL SPECIFICATIONS
The MAXEMS 2000 is a battery operated pulse generator that sends electrical impulses through electrodes to the body and reaches the underlying muscle group. The device is provided with two controllable output channels, each independent of each other. An electrode pair can be connected to each output channels. The electronics of the MAXEMS 2000 create electrical impulses whose Amplitude, Pulse Width. Pulse Rate, Contraction, Relaxation and Ramp may be altered with the switches. Press buttons are very easy to use and the slide cover prevents accidental changes in the setting.
Chapter 7: CONSTRUCTION
(5)MENU UP
(6)MENU DOWN
Select Menu
Menu Up
Menu Down
(7)SELECT MENU
(8)
BATTERY STRIP
(9)
BATTERY COMPARTMENT
The technical specification details of MAXEMS 2000 are as follows.
MECHAN
MECHANISM TECHNICAL DESCRIPTION
M TECHNICAL DESCRIPTION
1 Channel Dual
2 Pulse Amplitude Adjustable by intensity control knob.
Adjustable, 0~80mA peak into 500 ohm load
each channel.
3 Output Voltage Adjustable by intensity control knob
Adjustable, 0~40V peak into 500 ohm load
each channel.
4 Wave Form Asymmetrical biphasic pulse
5 Pulse Rate Adjustable, from 1~150Hz, 1Hz/step
6 Pulse Width Adjustable, from 50~250㎲, 10㎲/step
7 Contraction Time Adjustable, from 1~30 seconds, 1sec/step
8 Relaxation Time
Adjustable, from 1~45 seconds, 1sec/step
9 Ramp Time Adjustable, from 1~8 seconds, 1sec/step
10 Timer 15, 30, 45, 60 minutes
11 Power supply 9V Battery, type 6F22
12 Size 105(H) x 70(W) x 24(T) mm
13 Weight 130 grams with battery
14 Operating Condition Temperature: 0~40℃
Relative Humidity: 30%~75%
Atmosphere Pressure: 700Hpa~1060Hpa
15 Remark There maybe up to a ±5% tolerance of
parameters and ±10% tolerance of
Amplitude & Voltage.
4
5
1. Waveform Asymmetrical biphasic pulse
Pulse Rate: 5, 30, 100 Hz (load: 500 ohm) Pulse Width: Fixed at 250us (load: 500 ohm) Contraction Time: Adjustable, 1-30 seconds (load: 500 ohm) Relaxation Time: Adjustable, 1-45 seconds (load: 500 ohm) Power supply: 9V Battery (type: 6F22)
Chapter 9: ACCESSORIES
Each MAXEMS 1000 set is completed with standard accessories and standard
labels as given below.
Chapter 11: ATTACHMENT OF ELECTRODE LEAD WIRES
The wires provided with the system insert into the jack sockets located on top of the device. Holding the insulated portion of the connector, push the plug end of the wire into one of the jacks (see drawing); one or two sets of wires may be used. After connecting the wires to the stimulator, attach each wire to an electrode. Use care when you plug and unplug the wires. Jerking the wire instead of holding the insulated connector body may cause wire breakage.
MAXEMS2000
Do not insert the plug of the patient lead wire into the AC power supply socket.
1. Electrodes 4 pieces
2. Electrodes Leads 2 pieces
3. 9V Battery, type 6F22 1 piece
4. Instruction manual
5. Carrying Case
1 piece
1 piece
Chapter 10: GRAPHIC SYMBOLS
1. Direct current
2. Serial number
3. Type BF applied part
4. Caution, consult accompanying documents
6SN8
Chapter 12: LEAD WIRE MAINTENANCE
Clean the wires by wiping with a damp cloth. Coating them lightly with talcum powder will reduce tangling and prolong life.
Chapter 13: ELECTRODE OPTIONS
Your clinician will decide which type of electrode is best for your condition. Follow
application procedures outlined in electrode packaging to maintain stimulation and
prevent skin irritation. Use of legally marketed adhesive EMS electrodes is recommended.
EMS electrode must satisfy below specification.
1. Carbon film: standard carbon film
2. Gel: biocompatible adhesive hydrogel
3. size: more than 4.9cm² area.
4. AC impedance: less than 2K ohms
7
Chapter 14: ELECTRODE PLACEMENT
The placement of electrodes can be one of the most important parameters in achieving success with EMS therapy. Of utmost importance is the willingness of the clinician to try the various styles of electrode placement to find which method best fits the needs of the individual patient. Every patient responds to electrical stimulation differently and their needs may vary from the conventional settings suggested here. If the initial results are not positive,feel free to experiment.Once an acceptable placement has been achieved, mark down the electrode sites and the settings on the patient’s Reference sheet of this manual so the patient can easily continue treatment at home.
Chapter 15: TIP FOR SKIN CARE
To avoid skin irritation, especially if you have sensitive skin, follow these suggestions:
1. Wash the area of skin where you will be placing the electrodes, using mild soap and water before applying electrodes, and after taking them off. Be sure to rinse soap off thoroughly and dry skin well.
2. Excess hair may be clipped with scissors; do not shave stimulation area.
3. Wipe the area with the skin preparation your clinician has recommended. Let this dry. Apply electrodes as directed.
4. Many skin problems arise from the “pulling stress” from adhesive patches that are excessively stretched across the skin during application. To prevent this, apply electrodes from center outward; avoid stretching over the skin.
5. To minimize “pulling stress”, tape extra lengths of lead wires to the skin in a loop to prevent tugging on electrodes.
6. When removing electrodes, always remove by pulling in the direction of hair growth.
7. It may be helpful to rub skin lotion on electrode placement area when not wearing electrodes.
8. Never apply electrodes over irritated or broken skin.
For the safety of the user, Please turn the adjuster slowly because The EMS unit is not running if you turn the adjuster rapidly.
Chapter 16: APPLICATION OF RE-USABLE SELF
ADHESIVE ELECTRODES
Application
1. Clean and dry skin at the prescribed area thoroughly with soap and water prior to application of electrodes.
2. Insert the lead wire into the pin connector on the pre-wired electrodes.
3. Remove the electrodes from the protective liner and apply the electrodes firmly to the treatment site.
Remove
1. Lift at the edge of electrodes and peel; do not pull on the lead wires because it may damage the electrodes.
2. Place the electrodes on the liner and remove the lead wire by twisting and pulling at the same time.
Care and Storage
1. Between uses, store the electrodes in the resealed bag in a cool dry place.
2. It may be helpful to improve repeated application by spreading a few drops of cold water over the adhesive and turn the surface up to air dry. Over Saturation with water will reduce the adhesive properties.
NOTE
1. Do not apply to broken skin.
2. The electrodes should be discarded when they are no longer adhering.
3. The electrodes are intended for single patient use only.
4. If irritation occurs, discontinue use and consult your clinician.
5. Read the instructions for use of self-adhesive electrode before application.
9
Chapter 17: ADJUSTING THE CONTROLS
Relaxation Ramp
Contraction
Ramp
Contraction
Relaxation
1. Slide Cover A slide-on panel covers the controls for Contraction Time, Relaxation Time, Ramp Time,Pulse Width and Pulse Rate. Your medical professional may wish to set these controls for you and request that you leave the cover in place.
2. Power On/Off Switch and Intensity Controls: If both controls are in the off position, the device is switched off. By turning the controls clockwise, the appropriate channel is switched on and the power indicator (a dot) will appear on the LCD. The current strength of the impulse transmitted to the electrodes increases further when the control is turned clockwise. To reduce the current strength or switch the device off, turn the control counter clockwise to the required setting or off-position, respectively.
Select Menu
Menu Up
Menu Down
MAXEMS2000
3. Lead-wire Connector Connection of the electrodes is made with a two-lead connector. The device must be switched off before connecting the cables. Both intensity controls must be at the off position. Electrodes must be pressed firmly on the skin.
Select Menu
Menu Up
Menu Down
4.Function Selector Expose the controls by sliding front cover down from top of unit. When the button is pressed, it can change menu from contraction time, relaxation time, ramp time, pulse rate, pulse width to timer. The parameter of each function can be adjusted when it is revealed on the menu.
5. Parameter Increase Control This button controls the increase of parameter. When pressing this button, the parameter will increase. Press the button until the desired value is reached.
6. Parameter Decrease Control This button controls the decrease of parameter. When pressing this button, the parameter will decrease. Press the button until the desired value is reached.
7. Function Indicator This drawing on the top of LCD shows the function that is working. Each part of the drawing represents one function of the course.
Contraction
Relaxation
Ramp
Contraction
Relaxation
Ramp
10 11
8. Step to Set Parameter The parameter of a treatment course can be adjusted according to the following steps. Press “Select” button to select the function you wish to set.
a. Turn on the intensity After the electrodes are placed firmly on skin and the lead wires are plugged in the socket of device, turn the on/off control clockwise. The menu will appear on LCD. Notice the indication of power and function on the LCD.
b. Set Contraction Time The contraction time controls the time of stimulation. By pressing parameter controls, the contraction time can be pre-set. The range is adjustable from 1 second to 30 seconds.
c. Set Relaxation Time The relaxation time determines the time of relaxation. The stimulation ceases at setting. Relaxation time of both channels is changed by pressing the parameter controls. The range of it is adjustable from 1 second to 45 seconds.
d. Set Ramp Time This ramp time controls the time of output current that increases from 0 to the setting level and from the setting value to 0. When the ramp time is set, each contraction may be ramped in order that signals come on and come off gradually and smoothly. The ramp time is adjustable from 1 second to 8 seconds by pressing the parameter controls.
Note: Gradual “Ramp” time The stimulator special circuitry is designed so the electrical impulses gradually build to a peak and then level off. This “ramped” pulse produces a gradual muscle contraction emulating natural muscle movement. It can also prevent spastic patients from reacting adversely. The duration and intensity of the electrical pulses are predetermined by the clinician.
e. Set Pulse Rate The pulse rate determines how many electrical impulses are applied through the skin each second. By pressing the parameter controls, the number of current impulses per second (Hz) for both channels can be continually adjusted.
The pulse rate is adjustable from 1 Hz to 150 Hz.
f. Set pulse Width The pulse width determines the length of time. Each electrical signal is applied through the skin, which controls the strength and sensation of the stimulation.The pulse width is adjustable from 50 to 250 ㎲. Press the buttons until the desired value is reached.
12 13
g. Set Timer The treatment time can be adjusted 15 minutes, 30 minutes, 45 minutes, 60 minutes.
9. Check/Replace the Battery: Over time, in order to ensure the functional safety of EMS, changing the battery is necessary.
1) Make sure that both intensity controls are switched to off-position.
2) Slide the battery compartment cover and remove.
3) Remove the battery from the compartment.
4) Insert the battery into the compartment. Note the polarity indicated on the battery and in the compartment.
5) Replace the battery compartment cover and slide
Select Menu
Menu Up
Menu Down
to close.
6) If the battery symbol on the LCD shows “Low” battery, replace battery with new one.
Chapter 18: INSTRUCTION FOR USE
Protocol for the safe application of EMS
1) Check contraindications with patient.
2) Test skin for normal sensation using blunt/sharp test.
3) EMS device should be switched off and electrode leads disconnected.
4) Set electrical characteristics of EMS while device is switched off. They
should be prescribed.
5) Connect electrodes to pins on lead wire and position electrodes on patient’s skin.
6) Ensure EMS device is still switched off and connect the electrode wire to the
EMS device.
7) Switch the EMS device ON.
14
8) Gradually (slowly) increase the intensity until the patient experiences the first ‘tingling’ sensation from the stimulator.
9) Gradually (slowly) increase the intensity further until the patient experiences a ‘strong but comfortable’ tingling sensation.
10) This intensity should not be painful.
Protocol for the safe termination of EMS
1) Gradually (slowly) decrease the intensity until the patient experiences no tingling sensation.
2) Switch the EMS device OFF.
3) Disconnect the electrode wire from the EMS device.
4) Disconnect electrodes from the pins on lead wire.
5) Remove the electrodes from the patient’s skin.
Chapter 19: MAINTENANCE, TRANSPORTATION
AND STORAGE OF MAXEMS 1000
1. Non-flammable cleaning solution is suitable for cleaning the device. Note: Do not smoke or work with open flames (for example, candles, etc.) when working with flammable liquids
2. Stains and spots can be removed with a cleaning agent.
3. Do not submerge the device in liquids or expose it to large amounts of water.
4. Return the device to the carrying box with sponge foam to ensure that the unit is well-protected before transportation.
5. If the device is not to be used for a long period of time, remove the batteries from the battery compartment (acid may leak from used batteries and damage the device). Put the device and accessories in carrying box and keep it in a cool dry place.
6. The packed MAXEMS 1000 should be stored and transported under the
temperature range of atmosphere pressure 700 hPa ~ 1060hPa.
-10℃~40℃(14℉ ~104℉), relative humidity 30% ~ 75%,
Chapter 20: SAFETY-TECHNICAL CONTROLS
For safety reasons, check your MAXEMS 1000 each week based on the following checklist.
1. Check the device for external damage.
- Deformation of the housing.
- Damaged or defective output sockets.
15
2. Check the device for defective operating elements.
- Legibility of inscriptions and labels.
- make sure the inscriptions and labels are not distorted.
3. Check LED
- LED must be illuminated when switched on.
4. Check the usability of accessories.
- Patient cable undamaged.
- Electrodes undamaged.
Chapter 21: MALFUNCTIONS
Should any malfunctions occur while using the MAXEMS 1000, check :
1. Whether the switch/control is set to the appropriate form of therapy. Adjust the control correctly.
2. Whether the cable is correctly connected to the device. The cables should be inserted completely into the sockets.
3. Whether the impulse display LED is illuminated. If necessary, insert a new battery.
4. For possible damage to the cable. Change the cable if any damage is detected. ※ If there is any other problem, please return the device to your distributor. Do not try to repair a defective device.
Chapter 22: WARRANTY
All MAXEMS 2000 models carry a warranty of one year from the date of delivery. The warranty applies to the stimulator only and covers both parts and labor relating thereto. The warranty does not apply to damage resulting from failure to follow the operating instructions, accidents, abuse, alteration or disassembly by unauthorized personnel.
Do not dispose of your used batteries with other household waste. Hand your used batteries over at a designated collection point. If the device is not going to be used for a long period of time, remove the batteries to avoid any damage due to possible battery leakage.
16
Loading...