Instructions for Assembly/Disassembly:
1. To assemble the Lighted Suction Handle, insert the blunt end of the tip into the light source by aligning the pegs
with the grooves inside the light source. Push down and twist FIRMLY clockwise until rotation stops. This will lock
the handle into place and activate the light. NOTE: This product is a spring-loaded device; improper assembly
may cause ejection of the tip.
2. Attach the suction hose (extending from pump) to the barbed hose adapter of the handle by pushing rmly.
3. To attach the magnication lens, snap the lens onto the neck of the blue light source so that the top of the lens
angles toward the lighted tip.
4. To disassemble, push the Lighted Suction Handle into the light source and rotate counter clockwise to remove.
The light will go o when the tip is removed.
The Lighted Suction Handle is intended for single-patient use. The light source and magnication lens
are designed for multi-procedure use. A light source and lens are provided in each box of product.
Instructions for Use:
1. Examine the ear using an Otoscope noting the location and depth of cerumen or foreign body.
2. Under the direct visualization of the ear canal (using the Lighted Suction Handle with or without magnication)
carefully approach the cerumen with your thumb removed from the ne tuning suction port on the handle.
3. Once in contact with the cerumen, adjust handle pressure by gradually covering the opening until cerumen is
removed or grasped against the suction tip. NOTE: Keep a cup of hot water nearby in case the cerumen suction
handle becomes clogged. Drawing hot water into the Lighted Suction Handle may help clear the clog.
The Lighted Suction Handle is also eective for round foreign bodies. Simply slide the provided silicon
suction tip on to the end of the Lighted Suction Handle and use as previously instructed.
4. Once the procedure is complete, remove the Lighted Suction Handle from the light source and tubing then
dispose of handle. Save the magnication lens and light source for your next procedure.