Bionics IFM-500 User manual

Nemko ref. no.:NK-2E-M-055
DESCRIPTION OF EQUIPMENT UNDER TEST:
This equipment is a fetal monitor, Fetal heart rate(FHR), fetal movement(FM), and uterine
For description of operating, please refer to User’s Manual supplied by Man ufacturer.
NAME AND ADDRESS OF PRODUCTION-SITES (FACTORIES):
See front page.
INFORMATION ABOUT THE STANDARDS / DOCUMENTS CONSIDERED:
EN 60601-1 (1990) + A1 (1993) + A2 (1995) +A13 (1996) + Corrigenda (July 1994). EMC standard : EN 60601-1-2 (2001)
TESTED ACCORDING TO NATIONAL REQUIREMENTS FOR THE
FOLLOWING COUNTRIES:
LIST OF APPENDIXES / ENCLOSURES TO THE TEST REPORT:
This Part 1 TRF contains 69 pages, numbered from 1 to 69
This TRF contains attachments. Attachment #1: Photos. Page 1 to 5
Page 2 of 69
Nemko ref. no.:NK-2E-M-055
SUMMARY OF TESTING
Evaluation of Results If not explicitly stated otherwise in the standard, the test is passed if the measurement value is equal to or below (above) the
limit line, regardless of the uncertainty of the measurement. If the measurement value is above (below) the limit line, the test is not passed - ref. IECEE-CTL Guide 001 and Nemko WM L220. The instrumentation accuracy is within limits agreed (ref. IECEE-CTL/251A and Nemko procedure WM L226) :
The AC/DC adapter is certified by TÜV Rheinland according to EN60601-1. (refer to “56. list of Critical Component Parts”) It has double insulation between primary and secondary.
Maximum normal load : Refer to sub-clause 7 (power input) This equipment was tested with these probes:
DOP I
DOP II
UC Probe
Event Mark jack
Clause Remarks Information/Comments
Cord set not checked The equipment shall be provided with a cord set
complying with the national regulation of the countries intended for sale
Page 3 of 69
Nemko ref. no.:NK-2E-M-055
LIST OF ATTACHMENTS PROVIDING FURTHER INFORMATION
ON THE EQUIPMENT TESTED AND THE TEST METHODS
The following attachments are added to this TRF or kept in file at the Testing Station mentioned at the cover page:
Statements on calibrations and measurement uncertainties (where relevant) of the measurement and test equipment as identified throughout this TRF:
Calibration
Measurement uncertainties
Document title/identification: List of worksheets which describe measurement procedures or test methods where
practicable: Document title/identification: List of equipment or units tested and/or accompanying units and accessories (in case not
all individual units can be mentioned on the front-page):
All instruments used in the tests given in this report are calibrated and traceable to international standards. Further information about tractability will be given on request.
Measurements uncertainties are calculated for all instruments and instruments set-ups given in this report. Calculations are based on the principles given in the standard EA-4/02 (Dec. 1999), IECEE­CTL/153/NF (2002-04-19), Nemko WM L226 and actual internal Nemko procedures. Further information about measurements uncertainties will be given on request
Kept in file at Nemko; P31000
Kept in file at Nemko; P31000
ATT. No.: Not attached
ATT. No.: Not attached
Document title/identification: Technical design documents of components having basic, supplementary of reinforced
insulation (e.g. transformer specification mentioning the insulation class of the insulation materials applied - A to H):
Document title/identification: Description of safety systems (specifying for example safety circuits, redunda nt circuits,
double circuits, hardware/software safety measures, temperature controls, etc.): Document title/identification:
See summary of testing in this report.
Kept in file at Nemko
Kept in file at Nemko
ATT. No.: Not attached
ATT. No.: Not attached
ATT. No.: Not attached
Page 4 of 69
Nemko ref. no.:NK-2E-M-055
3 GENERAL REQUIREMENTS
3.1 EQUIPMENT when transported, stored, installed, operated in NORMAL USE and maintained according to the instructions of the manufacturer, causes no SAFETY HAZARD which could reasonably be fore- seen and which is not connected with its intended application in NORMAL CONDITION and in S.F.C.
3.4 An alternative means of construction is used to that detailed in this standard and it can be demonstrated that an equivalent degree of safety is obtained.
The equipment causes no hazards when used according the manufacturers instructions.
No alternative construction.
5 CLASSIFICATION
5.1 Type of protection against electric shock.
5.2 Degree of protection against electric shock.
5.3 Classification of the equipment against ingress of liquids.
Class I equipment
Type BF applied part
IPX0
5.5 Degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE
5.6 Mode of operation.
The equipment is not an AP or APG category equipment.
Continuous operation.
Page 5 of 69
E
Nemko ref. no.:NK-2E-M-055
INSULATION DIAGRAM
Protection against electric shock -
Block diagram of system
Drawing
Enclosure connected to PE
E
L
N
G
A
F
TUV Certified by EN60601-1 AC/DC Adapter
(100-240Vac, 50/60Hz, 1.2A, Output: +12Vdc, 3.5A)
D
B
Refer to table on next page. P
Main board
DOP1 DOP2 UC Probe Event Mark jack
PLASTIC ENCLOSUR
Comments
TABLE TO INSULATION DIAGRAM ON PREVIOUS PAGE
Area Distance
(Test Refer to
20.1 + 20.2)
Insulation type Basic/Supplement./
Double/Reinforced Insulation
A A-f Basic 250 V a.c. 1.6 mm 3.0 mm Not applicable B A-f Basic 12 V d.c. 1.6 mm 3.0 mm Not applicable D B-a Double/Reinforce 250 V a.c 5.0 mm 8.0 mm 4 000 V a.c. E A-a1 Basic 250 V a.c. 2.5 mm 4.0 mm 1 500 V a.c. F A-a1 Basic 250 V a.c. 2.5 mm 4.0 mm 1 500 V a.c. G A-a2 Double/Reinforce. 250 V a.c. 5.0 mm 8.0 mm 4 000 V a.c.
Comments
Maximum circuit
Required distances (mm) Dielectric strength
test voltage
voltage Clearance Creepage Refer to 20.3
Page 6 of 69
Nemko ref. no.:NK-2E-M-055
6 IDENTIFICATION, MARKING AND DOCUMENTS
6.1
Marking on the outside Example of marking label:
6.1 a Markings of Mains operated EQUIPMENT
6.1 b Markings of Internally Powered EQUIPMENT
See below P
Not applicable. No Internally Powered EQUIPMENT N
6.1 c Markings of EQUIPMENT supplied from a specified power supply
6.1 d Minimum requirements if limited space for marking
6.1 e Name and/or trademark of the manufacturer or supplier
6.1 f Model or type reference
6.1 g Rated supply voltage(s) or voltage range(s) Number of phases Type of current
6.1 h Rated frequency or rated frequency range(s) in Hz
6.1 j Rated power input (VA, W or A)
6.1 k Power output of auxiliary mains socket outlets
6.1 l Class II symbol
Symbol for degree of protection with
respect to harmful ingress of water according to EN 60529 (IPX0 not required to be marked)
N
P
Symbol 14
from DI is included in accompanying
documents. Bionics P
IFM-500 P 100-240 V
P
Single phase
50/60 Hz. P
0.25-0.15A P No socket outlet is provided N
Not applicable. N Ordinary equipment. IPX 0. P
Symbol for protection against electric
shock:
Type BF
. Symbol
P
Page 7 of 69
6.1 m Mode of operation (if no marki n g, suitable for continuous operation)
Nemko ref. no.:NK-2E-M-055
Continuous operation. N
6.1 n Types and rating of external accessible fuses
6.1 p Rated output voltage and current or power, output frequency (where applicable)
6.1 q Symbol for physiological effect(s):
6.1 r Anaesthetic-proof symbol:
6.1 s High voltage symbol:
6.1 t Special cooling requirements
6.1 u) Limited mechanical stability
6.1 v Protective packing requirement(s). Marking(s) for unpacking safety hazard(s)
6.1 y Earth terminals:
6.1 z Removable protective means
Durability of markings
“T1AL 250V” marked P
No auxiliary mains socket outlets. N
Not applicable. No need for marking. N Not of category AP or APG. N Not applicable. No parts with a voltage over 1 000 V a.c.
N
or over 1 500 V d.c. or 1 500 V peak value. No special cooling requirements N No limited mechanical stability N No special measures during transport or storage N
Symbol 9 of table DI, POTENTIAL EQUALIZATION TERMINAL
is used.
P
No removable protective means N Tested with a cloth of rack soaked with, in turn, water,
P
methylated spirit and isopropyl alcohol.
6.2
Marking on the inside
6.2 a Marking clearly legible
Nominal supply voltage of
permanently installed equipment marked inside or outside of equipment
6.2 b Maximum power loading of heating elements or lamp holders for heating lamps clearly and indelibly marked near or in the heater
Heating elements or lamp holders for
heating lamps not intended to be changed by operator and which can be changed only with the use of a tool, have at least an identifying marking referring to information stated in the accompanying documents
6.2 c High voltage parts:
6.2 d Type of battery and mode of insertion (if applicable) marked (see Sub-clause
56.7 b)
The sample label withstood the test.
Not applicable. No permanently installed equipment
No heating elements or holders for heating lamps
Not applicable.
P N
N
N
No high voltage part N Not applicable.
N
Batteries not intended be changed by
operator and which can be changed only with use of tool have at least marking referring to information in accompanying documents
6.2 e Fuse type and rating or reference stated
Not applicable.
N
N
Page 8 of 69
6.2 f Protective earth terminal marked
6.2 g Functional earth terminal marked
6.2 h Terminals for supply neutral conductor in permanently installed equipment marked (N)
Nemko ref. no.:NK-2E-M-055
Appliance inlet is used in an according to IEC 60320
No Functional earth terminal.
No permanently installed equipment.
N
N
N
6.2 j Markings required in Sub-clause 6.2 f, h, k and l
6.2 k The supply connections are clearly marked adjacent to the terminals or in accompanying documents (for small equipment)
6.2 l Statement for suitable wiring materials (at temperatures over 75°C) locates at or near the point of the supply connections and is clearly discernible after connection
6.2 n Capacitors and/or circuit parts are marked as required in Sub-clause 15 c
6.3
Marking of controls and instruments
6.3 a Mains switch clearly identified. ”On” and “off” positions marked or otherwise indicated
6.3 b Adequate indications of different positions of controls and other switches
6.3 c Adequate indication of the direction of setting devices if change of setting of a control could cause a safety hazard
Not applicable. No temp. over 75°C.
No such a capacitor
P
N
P
N
“I/O” marked close to the switch on front side P
P
P
6.3 f Functions of operator controls and indicators identified
6.3 g Numeric indications of parameters are in SI units according to ISO 1000. Units outside the International System as specified
All control buttons marked. All functions and marking described in user manual.
All control buttons marked. All functions and marking described in user manual.
P
P
Page 9 of 69
Nemko ref. no.:NK-2E-M-055
6.4
Symbols
6.4 a Marking symbols compliance with appendix D, where applicable
6.4 b Symbols for controls and performance conform to IEC 60 878, where applicable
Durability of marking symbols
6.5
Colours of insulation of conductors
6.5 a Protective earth conductor has green/yellow insulation
6.5 b All insulations of internal protective earth conductors are green/yellow, at least at the terminations of the conductors
6.5 c Only protective earth-, functional earth-, potential equalisation and inside earthing conductors (cf. 6.5 b) are green/yellow
6.5 d Colour of neutral conductor: light blue according to IEC 60 227 or 245
Complied with appendix D, where applicable P
P
Tested in Clause 6.1. P
P
P
Protective earth conductor P
Light Blue P
6.5 e Colours of phase conductors in power supply cord according to IEC 60 227 or 245
6.5 f Additional protective earthing in multi­conductor cords are marked green/yellow at the ends of the additional conductors
6.6
Identification of medical gas cylinders and connections
6.6 a Identification of content in accordance with ISO/R32
6.6 b Identification of connection point that errors are avoided when a replacement is made
Brown, VDE Certified P
No Additional protective earthing N
Not applicable. Not use of medical gas cylinders N
Not applicable. N
Page 10 of 69
Nemko ref. no.:NK-2E-M-055
6.7
6.7 a Colour red only used for warning of
Indicator lights and push-buttons
Not used of colour red.
danger or/and need for urgent action. Dot-matrix/alphanumeric displays not
The unit used dot-matrix / alphanumeric display
considered to be indicator lights
6.7 b Colours of unilluminated push-but t o ns.
No emergency switch N
Colour red only used in case of emergency
Location Meaning of indicator lights Illuminated push-button (see IEC 60
73)
Front of enclosure
Operating / LED Green
(For “Volume”)
Front of enclosure
Operating / LED Green
(For “Record”)
Front of enclosure
Operating / LED Green
(For “Alarm”)
6.8
6.8.1 Equipment is accompanied at least by
- instructions for use
Accompanying documents
User Manual P
P
Colour
- technical description
- an address to which the user can refer
All applicable classifications specified
in Clause 5 are included in instructions for use and technical description
markings in Sub-clause 6.1 included in
accompanying documents if not permanently affixed to equipment
Warning statements and explanations
of warning symbols are provided in accompanying documents
Language of accompanying documents
P In both the above mentioned booklets P In both the above mentioned booklets P
P
Properly described in user manual P
English. Will be in the language required for the market
P
where the equipment is intended to be sold.
Page 11 of 69
6.8.2 Instructions for use
Nemko ref. no.:NK-2E-M-055
6.8.2 a General information
Necessary information to operate the
Included in user manual P
equipment
Explanation of the function of controls,
Included in user manual P
displays and signals
Sequence of operation Connection and disconnection of
Included in user manual P Included in user manual P
detachable parts and accessories
Replacement of material which is
Included in user manual P
consumed during operation
Indications of recognised accessories,
Included in user manual P
detachable parts and materials, if the use of other parts or materials can degrade minimum safety
Cleaning, preventive inspection and
Included in user manual P
maintenance to be performed including the frequency of such maintenance
Information about safe performance of
Included in user manual P
routine maintenance
Information about preventive
Included in user manual P
inspection and maintenance to be performed by other persons
Meanings of figures, symbols, warning
statements and abbreviations on equipment explained in the instructions for use
Instruction for use states the function
and intended application of equipment
Instructions for use provide user with
information regarding potential electromagnetic or interference and advice how to avoid such interference
6.8.2 c Signal output and signal input parts
Signal output or signal input parts
intended only for connection to specified equipment stated in instruction for use
6.8.2 d Cleaning, disinfection and sterilisation of parts in contact with the patient
Details about cleaning or disinfection
or sterilisation methods that may be used for equipment parts which come into contact with the patient during normal use given in instruction for use
Included in user manual P
Included in user manual P
Included in user manual N
No SIP/SOP parts N
Properly described. P
Page 12 of 69
6.8.2 e Mains operated equipment with additional power source
Nemko ref. no.:NK-2E-M-055
A warning statement referring to the
necessity for periodical checking or replacement of an additional power source
If Class I equipment a statement saying
that where the integrity of the external protective conductor in the installation or its arrangement is in doubt, equipment shall be operated from its internal electrical power source
6.8.2 f Removal of primary batteries (i.e. not rechargeable batteries):
Instruction for use contains a warning
to remove batteries if equipment is not likely to be used for some time ,unless no risk of safety hazard
6.8.2 g Rechargeable batteries:
Instructions to ensure safe use and
adequate maintenance
6.8.2 h Equipment with a specified power supply or battery charger:
Instructions for use identify power
supplies or battery chargers necessary to ensure compliance with the requirements of IEC 601-1
This equipment is Alternatively using an internal
P
electrical power source and main supply
Included in user manual P
No primary battery N
Not rechargeable batteries N
Not applicable. N
6.8.2 j Instructions for use identify risks associated with disposal of waste, residues etc. and of equipment/accessories at end of their lives. Further it provides advice on minimising these risks
6.8.3 Technical description
6.8.3 a General
Technical description provides all data
essential for safe operation including data in Sub-clause 6.1 and all characteristics of the equipment
A statement whether particular
measures or particular conditions are to be observed for installing equipment and bringing equipment into use
6.8.3 b Replacement of fuses and other parts :
Required type and rating of fuses
utilised in the mains supply circuit external to permanently installed equipment
Instructions for replacement of
interchangeable and/or detachable parts which are subject to deterioration during normal use
The item is properly described. Reference to the user
P
Manual.
The Technical and Service Manual is well written and
P
contains all necessary information required.
P
No permanently installed equipment N
Properly described. P
Page 13 of 69
6.8.3 c Circuit diagrams, component part lists, etc.:
Nemko ref. no.:NK-2E-M-055
The technical description contains a
statement that the supplier will make available on request circuit diagrams, component part lists, descriptions, calibration instructions, etc. in order to assist user in case of necessary reparations
6.8.3 d Environmental conditions for transport and storage:
The technical description contains a
specification of the permissible environmental conditions for transport and storage
The above said specification also
repeated on the outside of the packaging
Included in user manual P
Included in user manual
P
P
Page 14 of 69
Nemko ref. no.:NK-2E-M-055
7 POWER INPUT(Continued)
7.1 Steady state current or power input does not exceed the marked rating by more than specified
Maximum normal load is followings: Continuous operation
1. Applied parts (DOPI, DOPII, UC and Event Mark jack)
2. Max. Volume
3. Printer
Equipment: 50 Hz supply
No. 64 Power input: 100-240VAC, 0.25-0.15A P
Power input Function U Input current Normal operation (A) Standby (A) Input power Normal operation (W) Input power Standby operation (W) Power factor
60 Hz supply
cos
ϕ
Power input: 100-240VAC, 0.25-015A P
Power input Function U Input current Normal operation (A) Standby (A)
P
N
lower
= 90V
UN = 100V UN = 240V
U
N
upper
= 264V
0.24 0.23 0.14 0.13 — — — —
12.0 12.0 13.0 13.0
23.0 23.0 33.0 34.0
0.52 0.52 0.39 0.38
N
lower
= 90V
UN = 100V UN = 240V
U
=264V
N
upper
0.22 0.21 0.13 0.13 — — — —
Input power Normal operation (W) Input power Standby operation (W) Power factor
cos
ϕ
12.0 12.0 13.0 13.0
20.2 21.0 32.0 32.0
0.57 0.55 0.40 0.39
10 ENVIRONMENTAL CONDITIONS
10.1 Equipment capable, while packed for transport and storage, of being exposed to environmental conditions as stated by manufacturer (see 6.8.3 d)
10.2 Operation of equipment according to specified environment and power supply
Considered OK. P
Considered OK. P
13 GENERAL
Equipment so designed that risk of
electric shock obviated as far as practicable (requirements as given in section three)
Considered OK. P
Page 15 of 69
Nemko ref. no.:NK-2E-M-055
14 REQUIREMENTS RELATED TO CLASSIFICATION
14.1
Class I equipment
14.1 a Parts with Double Insulation
Parts with Reinforced Insulation Parts with SELV-Voltage Parts with Protective Impedance
14.1 b If the mains part of equipment specified for an external d.c. power source is isolated from accessible conductive parts by basic insulation only, a separate protective earth conductor is provided
14.2
Class II equipment
14.2 a Class II equipment is of one of the following types:
14.2 b If equipment is fitted with a device for changing over from class I to class II protection, the following requirements are fulfilled:
The change-over device indicates the
selected class clearly
N CB Certified adapter used N CB Certified adapter used N N Not applicable. N
Not applicable N
Not applicable. N
Not applicable. N
A tool for change-over is necessary The equipment complies with all
requirements for the selected class at any given time
Class II position:
The device interrupts the connection of protective earth conductor to equipment or changes it into functional earth conductor (compliance with 18.2)
14.2 c Class II equipment provided with functional earth connections (cf. Sub­clause 18 k and l)
Not applicable. N Not applicable. N
Not applicable. N
Not applicable. N
Page 16 of 69
14.4
Nemko ref. no.:NK-2E-M-055
Class I and II equipment
14.4 a Equipment is provided with an
P
additional protection according to the requirements of Class I or Class II equipment
14.4 b No safety hazard develop, when a
P
connection with the wrong polarity is made in equipment specified for power supply from an external d.c. power source (e.g. for use in ambulances)
14.5
14.5 b Equipment also having means of
Internally powered equipment
Not applicable. No external d.c. power source N
connection to supply mains complies also with requirements for Class I or II while so connected
14.6
14.6 c Applied parts suitable for direct cardiac
Types B, BF and CF applied parts
Type BF N
application are of type CF
15 LIMITATION OF VOLTAGE AND/OR ENERGY
15 b Equipment with mains plug so
designed that the voltage 1 sec. after disconnection does not exceed 60 V
Certified AC/DC Adaptor N
Interference suppression capacitors Capacitance between each line and
N N
protective earth. Measured value
15 c Accessible live parts of capacitors or
Not applicable. N
related circuits having a residual voltage above 60 V does not have a residual energy above 2 mJ (see 15 c below)
A non-automatic discharging device is
Not applicable N
included and marked
Page 17 of 69
Nemko ref. no.:NK-2E-M-055
16 ENCLOSURES AND PROTECTIVE COVERS
16 a Protection against contact with live
parts.
- and with parts which can become live in single fault conditions
Instruction for use instructs operator
A tool is needed to open enclosure. No accessible live parts.
A tool is needed to open enclosure. No accessible live parts in single fault condition.
Not applicable. N
not to touch such parts and the patient simultaneously
Equipment:
Standard test finger : No. 97 Straight test finger : No. 105 Pull force measuring device : No. 52 Test pin : No. 98 Test hook : No. 95 Weight : approximately 4.2kg
Location of opening Test result (finger, pin, hook)
Top-side openings Pass
Rear-side openings Pass
Bottom-side openings Pass
16 b Check of openings in top covers with
the vertical suspended test rod
16 c Conductive parts of actuating
No holes in enclosure of the unit.
No handles, knobs, levers for removal N
mechanism of electrical controls after the removal of handles, knobs, levers etc. have either resistance to the protective earth terminal max. 0.2 ohms or the separation from live parts complies with 17 g
P
P
P
16 d Protection of contact with internal parts
of the equipment with a circuit voltage exceeding 25 V a.c. or 60 V d.c. which cannot be disconnected from the supply by external mains switch or plug device
16 e Protective enclosures are removable
only with the aid of tool or an automatic device makes these parts not live, when the enclosure is opened or removed
16 f Live parts are inaccessible to the test
rod through openings for adjustment of pre-set controls
No parts exceeding 25V ac, or 60Vdc N
Screw driver is required to remove enclosures P
N
Page 18 of 69
Nemko ref. no.:NK-2E-M-055
17 SEPARATION
17 a Separation method of the applied part
from live parts so that allowable leakage currents are not exceeded
17 a 1 Basic insulation - applied part earthed 17 a 2 By protectively earthed conductive part
(e.g. screen)
17 a 3 By separate earthed intermediate circuit 17 a 4 By double or reinforced insulation
17 a 5 By protective impedances limiting
current to applied part
17 a 6 Other method, e.g. specified in
particular standard
17 c There is no conductive connection
between applied parts and accessible conductive parts, which are not protectively earthed
17 d Supplementary insulation between
hand-held flexible shafts and motor shafts (Class I equipment)
Not applicable. N Not applicable. N
Not applicable. N DOP I, DOP II, UC Probe and Event Mark jack
P
See table 19 P
Not applicable. N
All accessible conductive parts are protectively earthed. P
No such a part N
Adequate isolation of accessible metal
N
parts from motor shaft
Rated motor voltage Test voltage Air clearances/creepage distances
N N N
(mm)/(mm)
17 g Separation method of accessible parts
other than applied parts from live parts so that allowable leakage currents are not exceeded
17 g 1 Basic insulation - accessible part
Protectively earthed and the applied part has such a low internal impedance to earth that leakage currents do not exceed the allowable values in normal condition and single fault condition.
See above 1) and below 4) N
earthed
17 g 2 By protectively earthed conductive part
See above 1) and below 4) N
(e.g. screen)
17 g 3 By separate earthed intermediate circuit 17 g 4 By double or reinforced insulation 17 g 5 By protective impedances limiting
Unearthed part to mains parts(see insulation diagram) P See above 1) and 4) P See above 1) and 4) P
current to accessible part
Leakage currents measurements, when
required
Protectively earthed and the applied part has such a low internal impedance to earth that leakage currents do not exceed the allowable values in normal condition and single fault condition.
P
P
Page 19 of 69
17 h Adequate arrangements used to isolate
defibrillation-proof applied parts from other parts
Nemko ref. no.:NK-2E-M-055
No defibrillation-proof applied parts N
Impulse voltage tests Peak voltage between Y1 and Y2 does
not exceed 1 V
After recovery time equipment
continues to perform its intended function
N N
N
Page 20 of 69
Nemko ref. no.:NK-2E-M-055
18 PROTECTIVE EARTHING, FUNCTIONAL EARTHING AND
POTENTIAL EQUALIZATION
18 a Sufficiently low impedance to the
protective earth terminal(18 f and 18 g)
18 b Suitable connection between protective
earth terminal and protective conductor in the installation (18 f)
18 e Means for connection of potentially
equalisation conductor complies with specified requirements
This connection complies with
following requirements:
- readily accessible
- no possibility to accidental disconnection in normal use
- conductor is detachable without the use of a tool
- power supply cord does not include potential equalisation conductor
- connection is marked with symbol
See 18 f) P
P
P
P
P P
P
P
P
18 f Impedance of protective earthing
system: Impedance between any accessible metal part and:
1) protective earth terminal (PE) (requirement: R max. 0.1 Ohm),
2) protective earth contact in appliance inlet (requirement: R max.
0.1 Ohm),
3) protective earth contact in the mains plug Test equipment: See attached list. (requirement: R max. 0,2 Ohm), are tabled below:
Accessible part and its location R (Ohm) measured against (numbers see above) 1 PE 2 inlet 3 plug
Gnd of Inlet - Accessible metal part Gnd of Inlet –Probe connector Gnd of Inlet – Potential equalization connector
18 g If the impedance of protective earth
connections other than in 18 f exceeds
0.1 Ohm, the allowable value of the enclosure leakage current is not exceeded in single fault condition
P
— — —
Not applicable N
0.02
0.021
0.018
— — —
18 k Functional earth terminals are not used
to provide protective earthing
No functional earth N
Page 21 of 69
18 l For Class II equipment with isolated
internal screens and with a power supply cord:
Nemko ref. no.:NK-2E-M-055
Not applicable. Class I equipment. N
- the third conductor is used only as functional earth of these screens and is coloured green/yellow
- insulation of such screens and all internal wiring connected to them is double insulation or reinforced insulation (see also 20.3)
- marking of functional earth terminal is distinguished from protective earth terminal and is explained in accompanying documents
Test of insulation (see Clause 20)
N
N
N
N
Page 22 of 69
Nemko ref. no.:NK-2E-M-055
19 CONTINUOUS LEAKAGE CURRENTS AND PATIENTS
AUXILIARY CURRENTS
19.1 General requirements (tests see 19.4)
19.1 g Equipment with multiple patient connections: Patient leakage current and patient auxiliary current do not exceed allowable values while one or more patient connections are disconnected from the patient or disconnected from the patient and earthed
19.2 Single fault conditions (tests see 19.4)
19.3 Allowable values (see 19.4)
19.4 Tests before humidity preconditioning treatment
Equipment:
In normal condition In single fault condition
P
See comment below.
Mains voltage : No. 83 AC-voltage : No. 19 DC-voltage : No.19 MD : 03 Equipment for supply connection normal or reverse: 06
Type of leakage current Allowed value
(µA) Earth leakage current 500 Enclosure leakage current 100 Patient leakage current (p.l.c.), a.c. : B, BF: 100
CF: 10 Patient leakage current (p.l.c.), d.c. : B, BF: 10
CF: 10 P.l.c. (110 % of the mains voltage on the
signal input part or signal output part) P.l.c. (110 % of mains voltage on the
applied part)
Patient auxiliary current, a.c. : B, BF: 100
Patient auxiliary current, d.c. : B, BF: 10
Comments
CF: 10
CF: 10
The values listed are the maximum values.
Measured max. value (µA)
246
56.3
58.7
0.01
42.1
0.01
Allowed value (µA)
1000 500 B, BF: 500
CF: 50 B, BF: 50
CF: 50 B: 5000
BF, CF: — B: —
BF: 5000 CF: 50
B, BF: 500 CF: 50
B, BF: 50 CF: 50
Measured max. value (µA)
453
57.3
59.3
0.01
45.9
47.5
0.01
Page 23 of 69
Loading...
+ 51 hidden pages