Biolight Meditech WBP202 User Manual

Electronic Sphygmomanometer
Users Manual
Guangdong Biolight Meditech Co., Ltd.
Product Model: WBP100/WBP201/WBP202
Product Name: Electronic Sphygmomanometer Manufacturer Name: Guangdong Biolight Meditech Co.,Ltd. After Service Contact Information:
Address: No.2 Innovation First Road, Technical Innovation Coast,
Hi-tech Zone, Zhuhai, P.R. China Fax: +86-756-3399919 Post code: 519085 Toll-free consultation hot line: +86-400-8818-233
Service life5 years
Revision History
This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change. Contents of this manual are subject to change without prior notice.
Document No.: J/WBPCE-A-008 Revision number: V1.0 Release time: Sept. 2015
Copyright © 2015 Guangdong Biolight Meditech Co., Ltd. All rights reserved.
CE mark
EC Representative Name:
Shanghai International Holding Corp. GmbH (Europe)
EC Representative Address:
Eiffestrasse 80, 20537 Hamburg Germany
Statement
Manufacturer holds the copyright of this manual, and we are also entitled to deal with this manual as confidential files. This manual is only used for operation, maintenance and service of product, someone else can not publish the manual.
This manual contains exclusive information protected by copyright laws and we reserve its copyright. Without written approval of manufacturer no parts of this manual shall be photocopied, Xeroxed or translated into other languages.
The contents contained in this manual are subject to amendments without
notification.
Manufacturer's Responsibility
Only under the following circumstances will manufacturer be responsible
for the safety, reliability and performance of the instrument: All the installation, expansion, readjustment, renovation or repairs are
conducted by the personnel certified by manufacturer.
The storage condition, operation condition and electrical status of the
instrument conform to the product specification.
The instrument is used in accordance with the users manual.
About this manual
This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation. This manual is based on the maximum configuration and therefore some contents may not apply to your product. If you have any question, please contact us. This manual is an integral part of the product. It should always be kept close to the equipment so that it can be obtained conveniently when needed.
Illustrations are for reference only. They may be not in conformity with
your products.
Signs in this manual
Warning: Indicates a potential hazard or unsafe practice that, if
not avoided, will result in death or serious injury.
Caution: Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.
Note: Provides application tips or other useful information to ensure
that you get the most from your product.
Electronic Sphygmomanometer Users manual
Contents
1 Intended Use ......................................................................................................................... 1
2 Application ............................................................................................................................ 1
3 Safety Information ................................................................................................................ 1
4 Appearance ........................................................................................................................... 4
5 Display .................................................................................................................................. 6
6 Setting ................................................................................................................................... 7
7 Historical data review function ............................................................................................. 8
8 Download And Install The Software ..................................................................................... 8
9 Using Steps ........................................................................................................................... 8
10 Maintenance and Cleaning .............................................................................................. 11
11 Accessory and Specifications .............................................................................................. 12
12 Troubleshooting .............................................................................................................. 14
13 Equipment Symbols ........................................................................................................ 15
14 Blood pressure general knowledge ................................................................................. 15
Appendix A Guidance and Manufacturer’s Declaration of EMC ................................................... 17
I
Electronic Sphygmomanometer Users manual
1 Intended Use
The device is intended to measure blood pressure and pulse rate of adult. The measurement data is available for clinical reference. It can store, manage the measure data, and transmit the measuring result via wireless signal.
It is intended to be used in clinic, general ward, physical examination department and at home.
Caution:
The patient is the operator when the device is used at home or in public
areas;
The device is intended for use only by clinical professionals and in continual
attendance when its used in hospital.
2 Application
The device is reusable and intended for measuring human blood pressure and
pulse rate.
Application population
Adult.
Measurement site
Place a bare arm through the cuff and position the cuff 1-2cm above the elbow joint.
3 Safety Information
Warning
Dont take measurement immediately after smoking, drinking alcohol, coffee
and black tea, or within an hour after a meal;
Magnetic and electrical fields are capable of interfering with the proper
performance of the device. For this reason make sure that all external equipments operated in the vicinity of the device comply with the relevant EMC requirements. Mobile phone, X-ray equipment or MRI devices are a possible source of interference as they may emit higher levels of electromagnetic radiation;
The device is designed for adults and should never be used on infants,
1
Electronic Sphygmomanometer Users manual
pregnant, pre-eclampic patients or people with serious arrhythmia. Consult your physician before use on children;
Don’t use the electronic sphygmomanometer in the environment containing
inflammable anesthetic gas;
Consult your physician for any of the following situations:
a) The application of the cuff over a wound or inflamed area; b) The application of the cuff on any limb with intravascular access or
therapy, or an arterio-venous (A-V) shunt;
c) The application of the cuff on the side of a mastectomy; d) Simultaneous use with other medical monitoring equipment on the same
limb; The device cannot be used together with high-frequency electrotomes and
defibrillators;
Please do not use the device while it is charging; Please do not measure immediately after an intense movement or a bath; This product might not meet its performance specifications if stored or used
outside the specified temperature, humidity and altitude ranges;
Please do not share the cuff with any infectious person to avoid
cross-infection;
Battery replacement should only be performed by personnel who have been
confirmed qualified by the manufacturer. To do otherwise will void your warranty and possible damage your unit;
Any form of modification to this device is forbidden.
Caution
Make sure that the appropriate cuff size is used; Cuff is replaceable and its replacement can only be conducted by
manufacturer.
We recommend the users to use the original manufactured cuff and adapter.
If not, it might cause measurement error or product failure;
If the adapter is abnormal, please change the adapter; Too high cuff pressure or too long charging time may make users feel
uncomfortable, and cause ischemia and neuropathy. Please turn off the device in time and take off the cuff;
The device and adapter should be considered as a medical system. The power
2
Electronic Sphygmomanometer Users manual
adapter must be provided or designated by manufacturer (see optional item list). If use another power adapter, it should be verified, otherwise electric shock may occur and cause patient death;
The effective wireless transmission distance is 5m in a barrier-free
environment. If more than this distance, the product might not work. Please use this product within effective wireless transmission distance;
The charging cable should be put beyond children to avoid the danger of
strangulation;
If children or pets beat, throw or tramp on the device, it might affect the
performance of the device. Please put the device beyond children and pets;
The main unit, cable, battery and cuff must be disposed according to the
local regulations at the end of their usage.
Caution
Changes or modifications not expressly approved by the party responsible
for compliance could void the user’s authority to operate the equipment;
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation;
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
-- Reorient or relocate the receiving antenna.
-- Increase the separation between the equipment and receiver.
-- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
-- Consult the dealer or an experienced radio/TV technician for help.
3
Electronic Sphygmomanometer Users manual
Note
Doctor should make diagnosis on clinical manifestation and symptoms, only
with using the device as subsidiary.
Blood pressure can be affected by the position of the cuff, measuring posture
and your physiologic condition;
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation;
The device complies with the standard of IEC 80601-2-30; If the determined blood pressure (systolic or diastolic) is outside the rated
range specified in part specifications, the device will immediately display an error code on screen. In this case, consult a physician or ensure that proper measurement procedures are followed;
Patients can use all functions of the device safely according to this users
manual;
Any questions about setting, using or maintaining the device, please contact
with manufacturer;
If unexpected operation or events appear in use process, please report to the
manufacturer;
When measuring, the following conditions may lead to the measured results
have differences:
a) Talking ; b)After defecate or urinate; c) After the bath; d) Different measurement site or environment; e) After smoking; f) After drinking alcohol, coffee and black tea; g) After the movement; h) Room temperature changes suddenly
i) Within an hour after a meal; j) Mood be agitated when caused by
tension, uneasy.
4 Appearance
WBP100
4
Electronic Sphygmomanometer Users manual
1 3 4 56 4 3 2 1
WBP201/WBP202
1Cuff
2 Display screen
Display systolic blood pressure, diastolic blood pressure, pulse rate and so on.
3Power buttonStart/Stop button
Long press (press and hold it more than 2s) this button to turn off the device;
When the power is off, short press (press and hold it less than 2s) this
button to turn on the device. Otherwise, short press it to start/stop the NIBP measurement.
4Charging cable interface
Connect to the power supply for charging.
5
Electronic Sphygmomanometer Users manual
Error code
Definition
Possible Reason
E02
Self-test failed
Transducer or other hardware failure.
E06
Loose cuff
a. Cuff is completely unwrapped; b. No cuff attached;
E07
Air leak
Air leak in pneumatics, hose, or cuff.
E08
Air pressure error
Unit cannot maintain stable cuff pressure, e.g. kinked.
E09
Weak signal
a. Very weak patient signal due to a loosely wrapped
cuff;
b. Extremely weak pulse from patient.
10 1 3 4 5
8
7
6 9 2
5 Select users button
In measurement sate: Short press this button to select users; In setting state: Short press this button, the time value would increase one.
6 Historical data review button
Not in setting state: Short press this button to review historical data. Long
press it to enter setting mode. In setting state: Short press this button to switch the setting items. Long press it to save the setting and quit setting mode.
5 Display
1. Systolic blood pressure;
2. Blood pressure unit;
3. Diastolic blood pressure;
4. Pulse rate;
5. Time;
6. Users name;
7. Pulse rate unit;
8. Signal quality;
9. Bluetooth state;
10. Battery state.
When measuring error, such as transducer failure, loosely wrapped cuff, screen displays the wrong code. The definition of the error code and the possible reason are as follows:
6
Electronic Sphygmomanometer Users manual
Error code
Definition
Possible Reason
E10
Range exceeded
Measurement range exceeds module specification of 255 mmHg maximum.
E11
Excessive motion
a. Too many retries due to interference of motion
artifact;
b. Signal is too noisy during step-down to detect
pressure pulse, e.g. subject has severe tremor;
c. Irregular pulse rate, e.g. arrhythmia.
E12
Overpressure sensed
Cuff pressure exceeds software safety limit, 295 mmHg in Adult mode. Could be due to rapid squeezing or bumping of cuff.
E13
Signal saturated
Large motion artifact that saturates the BP amplifiers amplitude handling capability.
E14
Air system leak
Module report Air system leakage Failure while in the Air System Leakage Test mode.
E15
System failure
Abnormal events.
E19
Time out
Measurement took more than 120 seconds in adult.
6 Setting
You can set the time and date of this product. The setting method is as follows:
Press the historical data review button and hold it for 3 seconds to enter the year setting interface. The third number of the year starts to flash. When you press the select users button once, the third number increase 1 (It means the year increases ten years).
After you finish setting the third number of year, press the historical data review button to enter the setting interface of the fourth number. The fourth number of the year starts to flash. When you press the select users button once, the fourth number increase 1 (It means the year increases 1 year).
Set month, day, hour, and minute in the same way.
After you finish setting time and date, press the historical data review button ” to enter the setting interface of blood pressure units (The default unit is mmHg). When a flashing mmHg appears on the screen, it means the blood pressure unit is mmHg. Press the select users button , mmHg would disappear and kPa would appear. Press the button again, kPa would disappear and mmHg would appear. So repeatedly, you can switch between the two units. Press the historical data review button and hold it for 3 seconds to end the setting after you finish setting unit.
7
Electronic Sphygmomanometer Users manual
7 Historical data review function
This device can record the blood pressure, pulse rate and time automatically to facilitate the management of blood pressure. Press the historical data review button ”, the latest measurement result would be displayed on the screen. Press the button again, the second latest measurement result would be displayed. And so on, you can review all the historical data.
8 Download And Install The Software
1Software download methods
Users who use the IOS system device can log in iphone App Store, enter “Biolight”
and search for the software of “BP tracker” in the searching results.
Users who use the Android system device can log in the wearcare official website: http://www.wearcare.cn for download. 2Download the software of “BP tracker”, click installation button for free installation.
9 Using Steps
Preparation before measurement
1Keep your body and mood calm, take off the thicker clothes such as coat, sweater, Etc. When measuring, bare upper arm or wear thinner shirt.
2Long press the power button to turn on the device. (Users, who use APP software, need to open the Bluetooth of the device and terminal, open the APP software, and connect with the device )
Caution When a call comes in during the measurement, the measurement will be
terminated. It is recommended that IOS device be set in Airplane mode during measurement to avoid strong magnetism interference.
Using Steps
(1) Apply the cuff
8
Electronic Sphygmomanometer Users manual
Loosen the calibre of the curly cuff slightly. Place a bare arm through the cuff
directly according to the instruction icon , . Position the cuff 1-2cm above the elbow joint and ensure the mark “Φ” on the cuff shall lie above artery.
Pull one end of the cuff to the lateral and tighten. The cuff should fit
comfortably, yet snugly around your arm. You should be able to insert one finger between your arm and the cuff.
Caution
Make sure that the appropriate cuff size is used. Refer to the cuff
circumference range in the Specification section of this manual.
(2) Body posture
Stay still and calm for five minutes before taking a blood pressure measurement.
Be seated with your feet flat on the floor without crossing your legs. Place your hand palm-side up in front of you on a flat surface such as a desk or table. The cuff should be placed at the same level as your heart.
9
Electronic Sphygmomanometer Users manual
Note:
Incorrect measuring posture showed as follows might cause an incorrect
measurement:
a) Bent overlean forward; b) Sitting cross-legged; c) Bend the body because of the low table or sitting on sofa.
(3) Measuring
Select users: Press the Select users button to select users, you can select user
1 or user 2;
Start measuring: Press the power button, then the cuff pressure automatically
and the screen shows rising pressure values. Stay still and calm during measurement. Do not move your arm, body and the device.
(4) Display the measurement value
The device exhausts automatically and displays the measurement results. About the display interface of measurement results, please refer to chapter 5 Display.
Users, who use the APP software, can read the measurement results on the software interface.
(5) Take off the cuff (6) Turn off the device
Long press the power button to turn off the device after the measurement.
Note:
While measuring blood pressure, the users must stay still and calm without
any talk;
The cuff tied on the limb shall be on the same level as the patients heart so
as to avoid the reading error resulting from the hydrostatics effect of the blood flow between the heart and cuff. If the cuff position is higher than heart level, the BP reading will be lower, the measured result shall be added 0.75mmHg (0.1kPa) for each centimeter higher; in case the cuff position is lower than heart level, the BP reading will be higher, the measured result shall be deducted
0.75mmHg (0.1kPa) for each centimeter lower;
If the signal measurement time is more than two minutes and the device
doesnt exhaust automatically, please take off the cuff and turn off the device.
10
Electronic Sphygmomanometer Users manual
Caution
You can stop the measurement process at any time by pressing Start/Stop
button.
Warning:
Prolonged and frequent NIBP measurements are associated with ischemia
and neuropathy in the limb wearing the cuff. Please examine the extremities of the limb frequently for normal color, warmth and sensitivity. If any abnormality is observed, stop the NIBP measurements.
Too frequent measurements may have influence on the measurement results.
10 Maintenance and Cleaning
1Clean the traces on the host or display with a dry, soft cloth or tissue; 2The most commonly used hospital cleaning agent and non-corrosive detergents
can be used for cleaning the device, but please be careful that many types of detergents must be diluted before use; Please use them according to the directions of the manufacturers of the detergents;
3 The casing of the device should be kept from the contamination of filth and
dirt, and it can be wiped with non-velvet soft cloth. When cleaning, do not spill the liquid onto the instrument. Ensure no liquid is allowed to enter the inside of the device;
4 It is recommended that if the cuff is used, for example, in hospital or a clinic, it
be disinfected twice a week, Wipe the inner side (the side that contacts skin) of the cuff with a soft cloth lightly moistened with Ethyl alcohol (75%-90%). Then air dry the cuff.
5Please do not clean or wet cuff. Please do not soak the device in liquid; 6Please store the host and the cuff in a cool and dry place; 7Do not drop the device or subject it to strong impact.
Caution
Avoid using alcohol-based, amido or acetone-based detergents. Please put the main unit and its accessories beyond infants and young
11
Electronic Sphygmomanometer Users manual
Items
Mode
Power adapter (optional)
LXCP12-05(Output DC5V,1.0A)
Production model
WBP100WBP201WBP202
Size
130 mm(L) x 55 mm(W) x 30 mm(H)
Weight
350g
Measure specifications
Measurement way
Automatic oscillometry
Measurement range Systolic: Diastolic:
30 mmHg255 mmHg4.0kPa34.0kPa); 10 mmHg220 mmHg1.3kPa29.3kPa);
Pressure accuracy
Static: Clinic:
±3 mmHg Average error: ±5 mmHg, standard deviation: ≤8 mmHg
PR range and accuracy
40 bpm240 bpm,±5% or ±3bpm, whichever is the greater
Blood pressure unit
“mmHg” or “kPa”
children to avoid dangerous accident.
Under normal circumstances, it is unnecessary for the device to have special
maintenance, and cautions must be exercised on the following points during the use of the device.
——Please use the device in the environment according to the requirements of the performance criteria;
——Avoid exposure or direct sunlight; ——Avoid excessive radioactive infrared rays or ultraviolet rays; ——Avoid contacts with organic solutions, dusts or corrosive gases.
No device component needs to be maintained by the user. Electrical
schematic diagram and component list are only provided to repair station or personnel which have been confirmed qualified by the manufacturer;
Check the device regularly please. If you doubt the measurement result is
not accurate, please contact the manufacturer.
11 Accessory and Specifications
11.1 Optional item
11.2 Technical Specifications
12
Electronic Sphygmomanometer Users manual
Software overpressure protection
295±5 mmHg
Measurement time
Normally, it is 20s to 45s (depending on HR and moving interference typically) Maximal measurement time: 120s
LED indicating lamp
WBP 100 Bluetooth
indicating lamp
1 (Blue) Not connected: The lamp flashes blue; Connected: The lamp lights blue without flashing; (When the device is power off, the lamp is off.)
Charging indicating lamp
1 (Orange) Light up: When the battery is being charged or when the DC power supply is connected; Off: When the battery is fully charged.
WBP 201 WBP 202 Bluetooth state indicating
Not connected: The indicating flashes; Connected: The indicating lights without flashing;
Battery state indicating
The electric quantity of battery is divided into four ranks to display the current electricity. Being Charged : The indicating flashes; Not being charged or fully charged: The indicating lights without flashing.
Bluetooth
Bluetooth version
Bluetooth 4.0 with low power consumption.
Transmission frequency
2440 MHz
Wireless transmission distance
5m
App software platform
IOS system
iPhone 4s、iPhone5、iPhone5s、iPhone 6、iPad3+、iPad mini
Android (4.3+) system
Mobile phone, Palmtop computer
Battery
Type
Rechargeable lithium ion battery
Rated voltage
3.7 VDC
Capability
600 mAh
Operating time
24 h
Charge time
2 h
Cuff
13
Electronic Sphygmomanometer Users manual
Large cuff
32 cm42cmOptional
Medium cuff
22 cm32cmWith
Small cuff
17 cm22cmOptional
Adapter
Input
100240V, 50/60Hz 0.3Amax
Output
DC 5V, 1A
Environment specifications
Temperature
Operating: 5℃~+40; Transportation and storage: -20℃~+55.
Humidity
Operating: 25%95%non condensing; Transportation and storage: 25%95% (non condensing).
Atmospheric pressure
Operating: 700hPa1060hPa; Transportation and storage: 500hPa1060hPa.
Symptom
Possible Reason
Solution
No response or cannot
turn on the device
1. The battery power might be at
shortage or no electricity;
1 The device itself might be
damaged.
1Please charge the device; 2Contact customer service.
The data cannot be display properly or display reads an abnormal
result
1. The cuff position was not
correct or it was not properly tightened;
2. Speaking, moving arm or body,
being angry, excited or nervous during test.
1. Review the cuff
application instructions and retest;
2. Retest when calm; avoid
speaking or movement during the test.
Cannot make connection between device and receiver
1. Bluetooth does not be open on
receiver;
2. The version of Android in the
receiver device isn’t 4.3 or 4.3+;
3. The version of Bluetooth is not
4.0;
4. Out of connection range.
1. Open Bluetooth;
2. Upgrade the Android system
to version 4.3 or 4.3+;
3. Use the receiver device with
Bluetooth 4.0;
4. Close to receiver.
12 Troubleshooting
14
Electronic Sphygmomanometer Users manual
Symbol
Symbol Note
Type BF applied part, without defibrillation protected. Refer to this user’s manual.
Symbol for the marking of electrical and electronics devices according to Directive 2002/96/EC
LOT
Batch code
SN
Serial number
IP22
Enclosure degree of ingress protection.
Federal Communications Commission
IPX0
Power adapter
Degree of protection against ingress of liquid
Fragile. Show transport package contents fragile, so handling should be handled with care.
Upward. It shows the correct position of the transport package is upright.
Guard against wet. Show packages afraid be wet.
Stacking layer limit. Same packing maximum stacking layers, N represents the number of layers limit. (N is 6).
13 Equipment Symbols
14 Blood pressure general knowledge
14.1 Definition of blood pressure
Blood pressure is the pressure on the vessel wall side. When the heart is contracting, the blood pressure is the highest at this time and is called the systolic pressure. When the heart is relaxing, the blood pressure is the lowest at this time and is called the diastolic blood pressure.
15
Electronic Sphygmomanometer Users manual
Scope
Systolic blood pressure (SYS) kPa/mmHg
Diastolic blood pressure (DIA) kPa/mmHg
Corresponding measure
Hypotension
13.3/100
8.0/60
Ask physician for measuring.
Normal blood pressure
13.3/10018.7/140
8.0/6012.0/90
Self measure.
Mild Hypertension
18.7/14021.3/160
12.0/9013.3/100
Consult physician.
Moderate Hypertension
21.3/16024.0/180
13.3/10014.7/110
Consult physician.
Severe Hypertension
24.0/180
14.7/110
Danger! Please see the doctor as soon as possible.
14.2 Classification of blood pressure
The World Health Organization (WHO) has created the following blood pressure classify table for assessing high blood pressure (without regard age or gender). Please note that other factors (e.g. diabetes, obesity, smoking, etc.) also need to be considered. Consult with your physician for accurate assessment.
NoteThis table is not intended to provide a basis for any type of emergency condition or diagnosis. This table only depicts different classification of blood pressure. Consult your physician for proper interpretation of blood pressure results.
16
Guidance and manufacture’s declaration – electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer of the user of the device should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The device use RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emission CISPR 11
Class B
The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Complies
Appendix A Guidance and Manufacturers Declaration of EMC
Note
This product needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
Do not use a mobile phone or other devices that emit electromagnetic fields, near the
unit. This may result in incorrect operation of the unit.
Caution:
This unit has been thoroughly tested and inspected to assure proper performance and
operation.
This machine should not be used adjacent to or stacked with other equipment and that
if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.
17
Guidance and manufacture’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
- guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4
±2 kV for power supply lines
±2kV for power supply lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
± 1 kV line(s) to line(s)
±1 kV differential mode
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 sec
<5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 sec
Mains power quality should be that of a typical commercial or hospital environment. If the user of the device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.
Power frequency (50Hz/60Hz) magnetic field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
18
Guidance and manufacture’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted
RF
IEC
61000-4-6
3 V
rms
150 kHz to 80
MHz
3 Vrms
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz to 2.5
GHz
3 V/m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device. b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
19
Recommended separation distances between
portable and mobile RF communications equipment and the device .
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 KHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
20