BIOFLEX Therapist, Therapist Pro, Home Unit Operation Manual

Page 1
01
Operation Manual
Therapist, Therapist Pro, Home Unit
Page 2
THERAPIST
Therapist Operation
Manual
Instructions
Meditech International Inc.
Page 3
II
Meditech International Inc.
411 Horner Ave. Unit 1 Toronto, Ontario Canada M8W4W3
Meditech’s EU Authorized Representative:
EMERGO EUROPE Molenstraat 15 2513 BH, The Hague
The Netherlands Phone: +31.70.345.8570 Fax: +31.70.346.7299
Disclaimer: Meditech International Inc. adheres to recommended safety standards and proven treatment outcome methodologies. The company has a comprehensive training program that health care professionals must follow, in order to properly apply Low Intensity Laser Therapy (LILT). Meditech International does not support the incorrect application of LILT.
Copyright © 2014 Meditech International Inc. All rights reserved.
No part of this publication may be reproduced, translated into another language, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise without the prior written consent of Meditech International Inc.
Every precaution has been taken in the preparation of this publication.
Meditech assumes no responsibility for errors or omissions. Neither is
any liability assumed for damages resulting from the use of the information contained herein.
All brand and product names mentioned are used for identification purposes only and are trademarks or registered trademarks of their respective holders.
Published: 1
st
Edition March 2013
2
nd
Edition April 2014
3
rd
Edition October 2014
4
th
Edition March 2015
5
th
Edition June 2015
Therapist Operation Manual DOC 100.652
Email: [email protected] Web: www.biolexlaser.com
(416) 251-1055 (888) 557
-4004
(416) 251-2446
Tel:
Fax:
Page 4
III
Mission Statement
Meditech International Inc. initiated research in
the
development of Low
Intensity
Laser Therapy
Systems in 1988.
The goal of the company has
been to engineer
devices
that are superior
to those in
existence and
scientifica
lly develop
accurate, effective
protocols
to treat a
variety
of medical
conditions in
the
musculo-
skeletal field.
These include sports injuries repetitive motion syndrome and the arthritides.
To date, we have surpassed our
initial goals and expectations but each day advances our horizons as the educational process continues.
Our primary objectives are
to significantly improve/ cure the multiple pathologies that we encounter in the shortest possible period of time. Guidelines to that end consist of the elimination of pharmaceuticals and pain and the restoration of mobility and quality of life.
Page 5
IV
Important Safety Instructions
The following safety guidelines protect your system from potential damage and ensure personal safety:
• The system can be used in any location in the world which provides the following AC power ratings:
1. 115 volts (V)/60 hertz (Hz) in most of North and South America and some Far Eastern countries such as Korea and Taiwan using the MPU30A-1-1 AC adapter, or
2. 100V/50Hz in eastern Japan and 100V/60Hz in Western Japan using the MPU30A-1-1 AC adapter, or
3. 230V/50Hz in most of Europe, the Middle East, and the Far East using the MPU30A-1-1 AC adapter.
• AC adapters are available by contacting Meditech or your local distributor.
• To help prevent electric shock, plug the Controller Unit (CU) into properly grounded power sources. For best grounding, plug into a receptacle marked “Hospital Grade.”
• Using extension cords with the system might affect its performance
• All system cables should be periodically inspected for signs of deterioration. If damaged, contact Meditech’s authorized service centers to repair or replace the cable(s).
• Do not pull, kink or pinch the cables. All cables should be loosely wound in a loop and they should never be tightly wrapped around the Treatment Array.
• Never remove the ground pin of the AC power cord.
Page 6
V
• Caution: Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
• Be sure nothing rests on your system’s cables and that the cables are not located where they can be interfered with.
• Only properly trained personnel should use the system.
• Do not spill food or liquids on or in your system.
• Keep system away from radiators and other heat sources.
• If the system is not in use, you should protect it against unqualified use by activating the key lock in the CU’s Main Menu screen.
• To reduce the risk of electric shock, do not remove covers from any part of the system. There are no serviceable parts inside. Refer servicing to Meditech’s qualified service personnel.
• Wear safety glasses when using the LD-Series Treatment Heads. Meditech recommends wearing safety glasses for all Treatment Heads at all times.
• In case of emergency, press the RED “LASER STOP” button
• On the front of the CU to immediately stop the emission of light radiation and to stop the treatment.
• The LD-Series Treatment Heads should only be used by qualified personnel. The LD-Series Treatment Heads should never be provided for patient use.
• The LD-Series Treatment Heads (lasers) should be protected against unauthorized use by using the ‘System Lock’.
Page 7
VI
• Some Treatment Arrays require cooling after
use. See the section “Treatment Array Cooling” in the “LS-Series Treatment Arrays and SLD-Series Treatment Arrays” section under “System Features, Indicators and Controls” for information about Treatment Array cooling.
• Care should be taken when operating the system
as external surfaces of the treatment heads, controller and power adapter may become hot.
• Before treating, always disconnect the system
from any connection to a computer.
• The system should be installed and operated
according to CAN/ CSA-Z386: 2001 “Laser Safety in Health Care Facilities.”
• Equipment should not be used adjacent to or
stacked with other equipment and that if adjacent or stacked use is necessary, the equipment should be observed to verify normal operation in the configuration in which it will be used.
• Equipment must not be disposed of as unsorted
municipal waste, but must be disposed of and collected separately.
• Although Meditech has taken all reasonable
precautions to ensure system safety, it is also the responsibility of the therapist to ensure their personal safety.
• For Contraindications, please refer to the
Clinical User’s Manual.
Page 8
VII
Rules on Laser Safety (LD-Series)
Lasers of the LD-Series Treatment Heads produce a very intense beam of light that is potentially hazardous if the direct beam or specular reflection is viewed by the unprotected eye. The following precautions should be taken to avoid direct beam viewing and diffraction. Read the provided “Laser Safety Guide” from the Laser Institute of America prior to using the system.
• Make certain that your system is located and operated in a controlled area when using the LD-Series Treatment Heads. Remove all shiny objects from the area in which you will be working. This includes rings, watches, metal bands, tools, and glass. Reflections from the beam can be nearly as intense as the beam itself.
• The entrances to the area should be posted with the provided laser danger sign.
• Never look directly into the laser beam or stare at its reflections. OCULAR DAMAGE CAN OCCUR.
• Never use magnifiers (such as binoculars or telescopes) to look at the beam.
• Never point the laser at the eyes directly.
• If you are applying the laser to a region of the face,
please ensure that the patient is wearing the safety eye shields which protect against possible exposure of the eye to the laser beam.
• Never leave a laser unattended while it is turned on.
• Wear safety glasses! Every person within the enclosed
area must wear safety glasses or eye shields when using the LD-Series Treatment Head.
• Specific labeling is required on all laser products. For safety purposes, do not remove any of the labels.
Page 9
VIII
Important Notice for the LD-Series
Treatment Heads
When applying the Treatment Head, ensure that the entire nosecone tip (also called “body switch”) has complete contact with the skin and is not applied at an angle. Failure to do so will allow potentially harmful laser light to irradiate surrounding areas.
Correct Application Incorrect Application
Figure 1 LD-Series Treatment Head Application
Safety Glasses Use
FAILURE TO FOLLOW THESE INSTRUCTION MAY
RESULT IN SERIOUS INJURY AND/OR BLINDNESS
READ ALL INSTRUCTIONS AND WARNINGS
BEFORE USING THE SYSTEM
• Never use the safety glasses (or eyewear) with other manufacturer’s equipment.
• Never use other manufacturer’s safety glasses with the BioFlex system.
Page 10
IX
• Always avoid direct eye exposure to direct or reflected laser light.
• Ensure that the safety glasses it securely at all times.
• Always check safety glasses before use to assure that
they have not become damaged. Exposure to or contact with chemical vapors or liquids may also cause surface cracks or other damage. Do not use if you observe any type of damage to your eyewear and replace your eyewear immediately.
• Always keep your safety glasses in the Therapist carrying case when not in use.
• Never use the safety glasses for any purpose other than for eye protection while using the system.
Cleaning Instructions
Between patient therapy sessions as indicated
• Wash in mild soap and water.
• Rinse in clean water.
• Air dry or pat dry with a clean, soft tissue.
DO NOT USE AMMONIA, ALKALINE CLEANERS, ABRASIVE CLEANING COMPOUNDS OR
SOLVENTS
WHEN CLEANING SAFETY GLASSES.
CERTAIN SOLVENTS MAY LOWER THE IMPACT RESISTANCE OF THE LENSES.
Page 11
X
Safety Kit
The BioFlex Safety Kit consists of the following:
• Laser Safety Guide
• Laser Danger Sign
• Two pairs of safety glasses (one for the practitioner and one for the patient)
Please read the Laser Safety Guide. Post the danger sign outside the entrance to the room where the system is located. If
there is
more than one entrance into the room where the system is located
, please contact Meditech to obtain additional signs
to
post in these areas.
Page 12
XI
Table of Contents
Important Safety Instructions
Rules on Laser Safety
Safety Glasses Use
Safety Kit
Classification
Safety Standards
Electromagnetic Compatibility (EMC)
Explanation of Warning Symbols and Labels
Controller Unit (CU)
Classification Marks
LD-Series Treatment Head Labels
LS-Series Treatment Head Labels
SLD-Series Treatment Head Labels
DUO
+
Treatment Head Labels
Description of the Beam Delivery Systems
LS-Series Treatment Heads
LD-Series Treatment Heads
iv
vii
viii
x
13
14
15
16
16
16
17
21
25
25
27
27
27
XI
Page 13
XII
System Features, Indicators and Controls
Controller Unit
LS-Series Treatment Arrays
Applying and Removing Treatment Heads
LS-Series
LD-Series
Operation
Environmental Conditions
Calibration
Cleaning
If the System Becomes Damaged
Guidelines for Proper Packaging
Error Messages
Specifications
Controller Unit
Treatment Heads
Glossary
System Parts List
29
29
34
37
38
39
41
54
55
56
57
58
59
63
63
64
66
67
Page 14
13
Classification
The system is classified as a medical device. Furthermore, it is in accordance to the Sub-clause 6 of IEC 60601-1 and Clause 6 of IEC 60601-1:2005:
• CLASS I EQUIPMENT, according to the type of protection against electric shock;
• TYPE B APPLIED PART, according to the degree of protection against electric shock;
• NOT CLASSIFIED, according to the degree of protection against harmful ingress of water;
• CONTINUOUS OPERATION, according to the mode of operation;
• NONE OF THE PARTS OF THE SYSTEM ARE STERILIZED.
Page 15
14
Safety Standards
This product is classified as medical equipment and has been tested to comply for the following standards:
Canadian standards
• CAN/CSA C22.2 No. 60601.1, Third ed. - Medical Electrical Equipment - Part 1: General Requirements for Safety
• CAN/CSA C22.2 No. 601-1.M90 - Medical Electrical Equipment – Part 1: General Requirements for Safety
• Canadian and United States of America standards
• UL 60601-1, First ed. - Medical Electrical Equipment - Part 1: General Requirements for Safety
• 21 CFR 1040 - Performance Standards for Light-Emitting Products
International standards
• IEC/EN 60601-1, Second ed. - Medical Electrical Equipment ­Part 1: General Requirements for Safety
• ANSI/AAMI ES60601-1:2005 - Medical electrical equipment
- Part 1: General requirements for basic safety and essential
performance
• IEC 60601-1-2 - Medical Electrical Equipment - Part 1:
General Requirements for Safety - 2. Collateral Standard:
Electromagnetic Compatibility - Requirements and Tests
• IEC 60601-2-22:2007 - Medical Electrical Equipment - Part
2: Particular Requirements for the Safety of Diagnostic and
Therapeutic Laser Equipment
• IEC 60825-1 - Safety of Laser Products - Part 1: Equipment Classification, Requirements and User’s Guide
Page 16
15
Electromagnetic Compatibility (EMC)
This equipment complies with the European rules for EMC according to the safety standard IEC 60601-1-2. This device complies with EMC rules under test conditions that include use of system cables and connectors between system components.
This Equipment requires special precaution regarding the EMC and must be installed and put into service according to the information provided in this manual.
The use of accessories and cables other than those specified and sold by the manufacturer may result in increased emissions or decreased immunity of the equipment and may cause the system to be non-compliant with the requirements of IEC 60601-1-2
Replace cables and connectors between components only with Meditech approved cables and connectors.
This system may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as reorienting or relocating the equipment or shielding the location.
Page 17
16
Explanation of Warning Symbols and Labels
This section explains the warning symbols and labels on the CU and the Treatment Heads. Each Treatment Head has labels which are located on both sides (underside and topside) of the Treatment Head.
“Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked ‘Hospital Grade’.”
Controller Unit (CU)
The CONSULT OPERATION MANUAL symbol. Parts of the system are marked with this warning symbol when it is necessary for the user to refer to the instructions in the Operation Manual.
The TYPE B APPLIED PART symbol. This product is classified as a Type B Applied Part, according to the degree of protection against electric shock.
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately.
Classification Marks
The UL mark is a registered trademark of the Underwriters Laboratories Inc. with respect to electric shock, ire, mechanical and other specified hazards only in accordance with CAN/ CSA C22.2 No. 601.1, CAN/CSA C22.2 No.
601.2.22 and UL 60601-1.
The CE mark is a registered trademark of the European Community
Page 18
17
LD- Series Treatment Head Labels
Figure 2 LD-I 75 Treatment Head Explanatory Label
Figure 3 LD-I 200 Treatment Head Explanatory Label
Page 19
18
Figure 4 LD-R 100 Treatment Head Explanatory Label
“VISIBLE LASER RADIATION” or “INVISIBLE LASER RADIATION”
Laser (LIGHT AMPLIFICATION BY STIMULATED EMISSION OF RADIATION) is a device which can be made to produce or amplify electromagnetic radiation in the wavelength range from 180 nm to 1 mm primarily by the process of controlled stimulated emission. For the LD-series Treatment Heads, the laser light is radiated from the top of the Treatment Head nosecone tip (see section “Description of the Beam Delivery System”).
“AVOID EXPOSURE TO BEAM”
Never look directly into the laser beam or stare at its bright reflections which are radiated from the aperture of the Treatment Head.
“CLASS 3B LASER PRODUCT”
This “CLASS 3B LASER PRODUCT” emits laser light at a wavelength where direct beam viewing is hazardous. Avoid viewing diffuse reflections.
Page 20
19
For this “CLASS 3B LASER PRODUCT” eye protection is required and the system has to be operated in an enclosed area where the entrance should be posted with the provided laser warning sign.
Figure 5 LD-Series Treatment Head Laser Aperture Label
“LASER APERTURE”
This label shows the location of the aperture (LD-series Treatment Head tip) through which the laser radiation is emitted. Please see
“Description of the Beam Delivery System”.
Figure 6 LD-Series Treatment Head Hazard Symbol
“YELLOW TRIANGLE”
This is a hazard label warning that the system emits hazardous light. See “CLASS 3B LASER PRODUCT.” This declaration is required by the U.S. Food and Drug Administration (FDA) in accordance with the standard 21 CFR 1040 - Performance Standards for Light-Emitting Products.
“This product is in conformity with performance standards for laser products under 21 CFR 1040, except with respect to those characteristics authorized by Variance Number 98V0816/7 effective October 7, 1998.”
Page 21
20
Figure 7 LD-Series Treatment Head Side Label Locations
Figure 8 LD-Series Treatment Head Bottom Label Locations
Date of Manufacture
Refer to the laser treatment head’s serial number label for the date of manufacture. The format of the serial number is shown below:
Figure 8a LD-Series Treatment Head Serial Number Format
Page 22
21
LS-Series Treatment Head Labels
Figure 9 LS-R 250 Treatment Head top label
Figure 10 LS-R 250 Treatment Head bottom label
Page 23
22
Figure 11 LS-I 500 Treatment Head top label
Figure 12 LS-I 500 Treatment Head bottom label
Page 24
23
Figure 13 LS-R 750 Treatment Head top label
Figure 14 LS-R 750 Treatment Head bottom label
Page 25
24
Figure 15 LS-I 1500 Treatment Head top label
Figure 16 LS-I 1500 Treatment Head bottom label
Page 26
25
Figure 17a DUO+LS-B 3000 Treatment Head top label
Figure 17b DUO
+
LS-B 3000 Treatment Head bottom label. Refer to
Figure 6a for serial number format and date of manufacture
Figure 18 SLD Treatment Head Label
Page 27
26
“CAUTION: SURFACE MAY BE HOT”
Although Meditech has taken every reasonable precaution to ensure that the treatment arrays do not become too hot or uncomfortable for the patient, they still may.
It is the responsibility of the administering practitioner to ensure the PATIENT’S SAFETY. Prior to starting each treatment, the patient should be warned that the treatment arrays may become uncomfortable and that they should notify the administering practitioner, if they do.
To prevent the treatment array from becoming too warm or uncomfortable to the patient, it is recommended that the operator should cool them after 20 minutes of use (e.g. one patient treated for 20 minutes or four patients treated for 5 minutes each etc.) by using Meditech’s fan for 2 minutes or 6 minutes at ambient temperature (without a cooling fan). The above mentioned cooling procedure applies to protocols with continuous wave mode of operation at 100% power. With protocols using modulation mode of operation and/or less power, treatment times can be longer between cooling periods.
Use caution if very young children or pets are in the patient environment. It is recommended that the operator should prevent children from unintentionally touching the treatment array surfaces because they may not be aware of potential surface temperatures.
DO NOT cover the treatment arrays during the treatment because the top surface removes excess heat into the surrounding environment.
Date of Manufacture
Refer to the “Date:” field on the bottom label of the LS Series treatment head for the date of manufacture.
Page 28
27
Description of the Beam Delivery Systems
This section explains how the LS or laser light travels from its origin to the environment for each Treatment Head.
LS-Series Treatment Heads
(Models: LS-R 250, LS-I 500, LS-R 750, LS-I 1500, SLD-R500, SLD-I1000 and DUO
+
, DUO)
Light radiates from the LS-Series directly into the environment (shown in figure below).
Figure 19 LS-Series Treatment Head Light Radiation
Figure 19a SLD-Series Treatment Head Light Radiation
Page 29
28
LD-Series Treatment Heads
(Models: LD-R 100, LD-I 75, LD-I 200)
Light radiated from the laser diode located inside the Treatment Head nosecone passes through an air gap. Then, it goes through a nosecone window (laser aperture) into the environment as shown in the figure below.
Figure 20 LD-Series Treatment Head Light Radiation
Page 30
29
System Features, Indicators and Controls
Controller Unit
THERAPIST
EXIT
Figure 21 Front of CU
SELECT
Page 31
30
To help prevent electric shock, plug the Controller Unit (CU) into properly grounded power sources. For best grounding, plug into
a
receptacle marked “Hospital Grade.”
Screen
A back-lit liquid crystal display (LCD) provides the user with information and instructions.
POWER / LASER STOP
• The system is turned on and off by using the POWER / LASER
STOP button on the front of the CU.
• Press once to turn on the system.
• Press POWER / LASER STOP for approximately 1 second to shut off the system.
• Pressing POWER / LASER STOP during treatment will immediately stop the treatment and the emission of light.
SELECT and EXIT
Follow the instructions on the bottom of the CU’s screen for when to use SELECT and EXIT.
UP and DOWN
(Unlabeled Arrow Buttons)
Generally, UP and DOWN allow scrolling through menu items or selecting values on the CU’s screen. Follow the instructions on the bottom of the CU’s screen how to use the UP and DOWN buttons.
Page 32
31
Figure 22 Back of the CU
Date of Manufacture
Refer to the CU’s serial number label for the date of manufacture. The format of the serial number is shown below:
Page 33
32
Figure 22a Serial number format for CU
Stand
The stand can be pulled out to allow the CU to rest at an angle for easy viewing of the screen.
If detached from the CU, the stand can be re-attached as shown in the following figure.
Figure 23 Replacing the Stand
Page 34
33
POWER
INPUT
Figure 24 Top of the CU
Treatment Head Port
(Marked with man symbol)
Any Treatment Head may be plugged into port. Plugs and sockets are polarized; therefore they must be correctly orientated before they can lock. Align the mark on the plug with the mark on the socket.
Power Input
Input for the power adapter. Do not use any other adapter except the one provided by Meditech or your local distributor. Use of any other power sources will damage this unit.
USB
This port is used when it is required that the system be connected with a computer.
DO NOT OPERATE THE UNIT WHILE CONNECTED
TO A COMPUTER.
USB
Page 35
34
LS-Series & SLD-Series Treatment Arrays
Figure 25 LS Series
Figure 26 DUO Series
Figure 27 SLD Series
Page 36
35
Treatment Array Cooling
For Models LS-R 250, LS-I 500, LS-R 750, LS-I 1500, SLD-R 500, SLD-I 1000:
Although Meditech has taken every reasonable
precaution to ensure that the Treatment Heads do not become too hot or uncomfortable for the patient, they still may. It is the responsibility of the administering practitioner to ensure the PATIENT’S SAFETY. Prior to starting each treatment, the patient should be warned that the Treatment Heads may become uncomfortable and that they should
notify the administering
practitioner if they do.
To prevent the treatment arrays from becoming
too warm or uncomfortable to the patient, it is recommended that the operator should cool them after 20 minutes of use (e.g. one patient treated for 20 minutes or four patients treated for 5 minutes each etc.) by using Meditech’s fan for 2 minutes or 6 minutes at ambient temperature (without a cooling fan). The above mentioned cooling procedure applies to protocols with continuous wave mode of
operation at 100% power. With protocols using
modulation mode of operation and/or less power, treatment times can be longer between cooling periods.
Page 37
36
For Models DUO (LS-B 2250) and DUO+ (LS-B
3000):
Safety Features
(For Models LS-R 250, LS-R 750, LS-I 500, LS-I 1500)
The LS-Series Treatment Heads are equipped with an “ambient light sensor” which will prevent the Head from starting if not secured to the body. If the Treatment Head is not secure, an error message (see section “Error Messages”) will be displayed on the LCD. The operator may then proceed with discretion. The ambient light sensor is designed to measure the amount of ambient light present in the room prior to invoking a treatment. If the Treatment Head is securely attached to the patient, the amount of ambient light reaching the sensor will be minimal. If the Treatment Head is not secure, the amount of ambient light reaching the sensor will be significant. Therefore, the amount of ambient light may vary in different locations.
To prevent the treatment arrays from becoming
too warm or uncomfortable to the patient, it is recommended that
the operator should cool
them after 40 minutes o
f use (e.g. one
treatment that delivers 20 minutes of red light followed by 20 minutes of infrared
light) by
using Meditech’s fan for 2 minutes or 6 minutes at ambient temperature (without a cooling fan). The above mentioned cooling procedure applies to protocols with continuous wave mo
de of
operation at 100% power. With protocols using modulation mode of operation
and/or less
power, treatment times can be longer between cooling periods.
If the patient finds that Treatment Head is becoming too warm or uncomfortable, STOP or PAUSE the
treatment, remove the Treatment Head and allow it to cool until it is cool to the touch. Use CAUTION, if very young children or pets are being treated. It is recommended that the operator watch children for unintentional contact of the Treatment Head surfaces because they may not be aware of potential surface
temperatures.
Page 38
37
Applying and Removing Treatment Heads
Method of Disconnecting the Treatment Head:
1. Place your thumb and forefinger around the ribbed connector and exert moderate pressure of the digits against the CU.
2. At the same time, squeeze the connector.
3. Simultaneously, pull away from the CU to extract the cable.
Figure 28 Disconnecting Treatment Array
Page 39
38
LS-Series and SLD-Series
The LS-Series Treatment Arrays may be secured directly to the body using the attached straps.
When applying, place the Treatment Array over the area to be treated and wrap the straps around the body part. Place the Velcro from one strap over the Velcro portion of the other strap.
To remove, unlock the Velcro and remove the Treatment Array.
Figure 29 Applying the Treatment Arrays
Page 40
39
Warnings for Using the Treatment Arrays:
• Do not twist
• Do not hyperextend
• Do not drop
• Do not tamper with diodes
Note: The Treatment Arrays must be secured to allow air circulation around the top of the head. Do not cover the heads while running treatments, however you can use the Meditech approved weight set.
LD-Series
Body Switch and Indicator
The LD-Series Treatment Heads are activated when the nosecone tip contacts the skin. After selecting a treatment or continuing to the next phase, wait until the indicators on the Treatment Head turn green before applying the Head. Once the indicators are green, apply the Treatment Head tip to the treatment area and the treatment will begin (the indicators will turn amber). Removal from the skin will cease treatment and the indicators will return to green.
NOTE: When applying the Treatment Head, it is
important to ensure that the flat end makes full contact with the skin.
Failure to do so might
expose bystanders to potentially dangerous laser
radiation. Refer to the pictures below for the
correct and incorrect application of the Treatment Head.
A Treatment sleeve MUST always be used for
broken or compromised skin.
Page 41
40
Correct Application Incorrect Application
Radiation ceases immediately when:
• Contact with the skin ceases.
• Pressing the “UP”, DOWN” or “SELECT” buttons.
• Pressing “LASER STOP” button.
The system is a stand-alone unit.
NOTE: After selecting a treatment or continuing to the next phase, do NOT touch the tip of the Treatment Head until the indicators on the head turn green; otherwise the Treatment Head will not work. If you do touch the tip before the indicators turn green, you will have to stop and restart the treatment to re-activate the Treatment Head.
DO NOT UNPLUG THE TREATMENT HEAD WHILE IT IS ACTIVE.
DO NOT OPERATE THE UNIT WHILE CONNECTED TO A COMPUTER.
Page 42
41
Operation
The following pages show screen shots and describe how to use the various functions of the BioFlex Therapist System:
Main Screen Menu
From the MAIN MENU you can select to run a treatment, customize a treatment, adjust the settings of the unit and also lock the unit from unauthorized use.
This is the main menu when the system is operating.
Select “Run Treatment” from the
main menu.
Use the
UP/DOWN buttons to scroll to
the Anatomy you are treating and press
SELECT
Page 43
42
Scroll: Select: Exit:
Scroll: Select:
Exit:
Scroll: Select: Exit:
UP/DOWN
SELECT
EXIT
UP/DOWN
SELECT
EXIT
1 of 9
LS-R750
6min
CW N/A N/A
100%
UP/DOWN
SELECT
EXIT
Use the UP/DOWN buttons to scroll to the Pathology you want to treat and press
SELECT
Use the UP/DOWN buttons to scroll to the Intensity you want to use and press
SELECT.
Press the
UP/DOWN buttons to view the
Treatment Information. Press the
UP/DOWN buttons to scroll through
the Stages of the Treatment.
Pathology
C-Spine
Intensity
Stage One
Run Treatment
Page 44
43
LS Arrays
View: Start: Exit:
LS Arrays
LS Arrays
DBL CLICK
EXIT
LASER STOP
To start the treatment press SELECT
During this time the Controller performs a treatment Head Check to test functionality of the System.
The screen now asks you to place the treatment head on the area to be treated. Apply the Treatment Heads as described in “Applying and Removing the Treatment Heads” section. Put the safety glasses provided with your system on and press
SELECT to start the treatment.
UP/DOWN
SELECT
EXIT
Run: Cancel: Stop
Page 45
44
LS Arrays
Pause: Pause: Pause:
LS Arrays
PAUSED
DBL CLICK
EXIT
LASER STOP
LS Arrays
SELECT
EXIT
LASER STOP
This screen informs you of the current phase and time remaining for current phase of the protocol. The
LASER STOP button can be
pressed at any time during the treatment to immediately stop the emission of radiation and cease treatment.
Pressing either of the
UP/DOWN, EXIT or
SELECT buttons will pause the treatment.
To continue the treatment press SELECT. To cancel treatment, press
EXIT and follow the
on-screen menu. To stop the treatment press
LASER STOP.
The exit screen.
UP/DOWN
SELECT
EXIT
Run: Cancel: Stop:
No: Exit: Stop:
Page 46
45
Step 1 of 3
TH
Time WaveForm Frequency
DutyCycle Power Run Treatment
Scroll: Select: Exit:
Step 1 of 3
TH50
Time WaveForm Frequency DutyCycle Power
Run Treatment
Scroll: Select: Exit:
Pressing SELECT while the step number is highlighted allows scrolling through the phases.
To set the Treatment Head (TH), Time, Waveform, Frequency, Duty Cycle and Power, highlight the parameter and press SELECT. The highlighted area will change (see below).
Customize a Treatment
This section instructs you on how to customize a protocol using the Therapist CU.
UP/DOWN
SELECT
EXIT
UP/DOWN
SELECT
EXIT
6 min
Squa
5000
90%
100%
TH LS-R750
LS-R750
6 min
Squa
5000
90%
100%
Page 47
46
Step 1 of 3
LS-R750
Time WaveForm
Frequency DutyCycle Power Run Treatment
scroll: select: exit:
Step 1of 3 TH
Time WaveForm Frequency DutyCycle
Power Run Treatment
Scroll: Select:
Exit:
Customize
Step One
Scroll: Select: Exit:
To set the specific parameter use the UP/ DOWN buttons to scroll through the options. When you have found the option you want to use, press SELECT to initiate change.
To Run the protocol just created plug in the appropriate head, scroll to Run Treatment from any phase and press SELECT.
Treatment will always start from step 1.
During this time the Controller performs a Treatment Head Check to test functionality of the System.
Run Treatment
UP/DOWN
SELECT
EXIT
LS-R750
6min
Squa
5000
90%
100%
UP/DOWN
SELECT
EXIT
UP/DOWN
SELECT
EXIT
6min
Squa
5000
90%
100%
TH
Page 48
47
Customize
Step One
Scroll: Select:
Exit:
Customize
Step One
Scroll: Select:
Exit:
Customize
Step One
PAUSED
Scroll: Select: Exit:
The screen now asks you to place the Treatment Head on the area to be treated. Apply the Treatment Heads as described in “Applying and Removing the Treatment Heads” section. Place the safety glasses provided with your system on and press SELECT to initiate treatment.
This screen informs you of the current phase and time remaining for current phase of the protocol. The LASER STOP button can be pressed at any time during the treatment to immediately stop the emission of radiation and cease treatment.
Pressing either of the UP/DOWN, EXIT or SELECT buttons will pause the treatment. To continue the treatment press SELECT. To cancel treatment, press EXIT and follow the on-screen menu. To stop the treatment press LASER STOP.
UP/DOWN
SELECT
EXIT
UP/DOWN
SELECT
EXIT
UP/DOWN
SELECT
EXIT
Page 49
48
Customize
Step One
SELECT
EXIT
LASER STOP
The exit screen.
No: Exit: Stop:
Page 50
49
Settings
This section describes how to change the various settings. The settings will be discussed in more detail in the following illustrations.
Settings
Default Array
Scroll: Select: Exit:
System Info
This screen provides the Treatment Head serial number, firmware version and the Controller serial number and firmware version. Press EXIT to go back to the Settings menu.
Exit:
UP/DOWN
SELECT
EXIT
EXIT
Settings Main Screen
To select an item from the menu, scroll
through the options until the item is
highlighted and press SELECT.
System Info
Page 51
50
Change: Select: Exit:
Adjust: Select: Exit:
Update: Exit:
UP/DOWN
SELECT
EXIT
UP/DOWN
SELECT
EXIT
SELECT
EXIT
Beep Control
The Key Beep can be turned on or off. Key Beep is the audible beep heard when a button is pressed. This does not turn off the beeps heard during treatment. Use the UP/DOWN buttons to select either Key Beep ON or Key Beep OFF and press SELECT.
LCD Contrast
To adjust the contrast of the screen use the UP/DOWN buttons to adjust the levels of contrast to the desired level. Press SELECT to accept or press EXIT to keep current settings.
Update Firmware
This menu option is used to update the Controller or Treatment Head firmware if a new version has been released.
Beep Control
Key Beep ON
LCD Contrast
Update Firmware
Page 52
51
Scroll: Select: Exit:
UP/DOWN
SELECT
EXIT
Lock Password
This option is used to change the Lock Password. If you change the password write it down and place it in a safe and secure location.
Settings
Lock Password
Page 53
52
System Lock
Settings
Locked
Unlock Unit
-
This screen is displayed when the system is locked. You must enter a password to unlock the unit.
To enter the password, use the UP/DOWN buttons to scroll through the numbers 0-9. When you have entered the correct number, press SELECT. The cursor will move to the next digit. Repeat until you have entered the password and Press SELECT.
The system lock is a software lock that password protects the system from unauthorized use. You should lock the system when it is not in use. The default lock password is 0000. If you change the password, write it down and keep it in a safe place. If you have lost your password contact Meditech or a Meditech authorized service center for further assistance.
Page 54
53
Other Screen Shots
You will see the followings screens when the wrong Treatment Head is plugged into the Controller or if the Treatment Head becomes unplugged during a treatment.
This screen is displayed when the correct Treatment Head in not plugged in.
!
Run Treatment
This screen is displayed when the treatment head is unplugged.
Press SELECT
to continue with the treatment. Press EXIT to cancel.
Page 55
54
Environmental Conditions
The system is capable, while packed for transport or storage, of being exposed for a period of up to 15 weeks within the following ranges:
• An ambient temperature range of -20˚C to +70˚C (-4˚F to
+158˚F);
• A relative humidity range of 10% to 100% (non-condensing);
• An atmospheric pressure range of 500hPa to 1,060hPa.
• When the system is operated in accordance with the
instructions for use, the following conditions should not be
exceeded:
• An ambient temperature range of 0˚C to +40˚C (+32˚F to
+104˚F);
• A relative humidity range of 45% to 85% (non-condensing);
• An atmospheric pressure range of 700hPa to 1,060hPa.
Page 56
55
Calibration
• Calibration of the system and/or its components can only be performed at Meditech or at an authorized service center.
• If any LD-Series Treatment Head indicates repeat errors, it may need calibration.
• To maintain your arrays and to keep them working at optimal performance, they should be calibrated periodically, depending on use.
• The Controller Unit (CU) does not require calibration.
Note: Only Meditech authorized service personnel are able to calibrate the system properly. Unauthorized changes to the system may be subject to termination of the warranty at Meditech’s discretion.
WARNING: No modification of this equipment is
allowed. If this equipment is modified, appropriate
inspection and testing must be conducted by
Meditech’s service personnel to ensure continued safe use of the equipment.
Page 57
56
Cleaning
• The CU may be cleaned using a damped soft cloth or a soft disinfectant cloth.
• The arrays may be cleaned with alcohol wipes, a soft disinfectant cloth, a soft damp cloth and/or alcohol/ cotton swab combinations.
• A soft bristle brush can be used to remove particles from the LS Series Treatment Array.
• The LD Series Treatment heads may be cleaned with alcohol wipes a soft disinfectant cloth, a soft damp cloth and/or alcohol / cotton swab combinations.
• It is recommended that the treatment surface of the Treatment Arrays be cleaned after each treatment.
• To keep the Treatment Arrays clean from contaminates from open wounds, etc., use a clear plastic bag (or plastic wrap) to prevent Treatment Array contact with the disease area. Treatment Array Sleeves (covers) can be purchased from Meditech or a Meditech authorized distributor.
Page 58
57
If the System Becomes Damaged
If part or all of the system is dropped or damaged, check to determine if it still functions properly. To troubleshoot a system, perform the following steps:
1. Check all power connections. Make sure all cables are secure and properly connected.
2. Turn on the CU. If the CU does not work, contact a Meditech authorized service center for technical assistance.
Liquids from spills or excessive humidity can also damage your system. If the system becomes wet, perform the following steps:
1. Immediately unplug all system components from the electrical
outlet.
2. Contact a Meditech authorized service center for assistance.
Page 59
58
Guidelines for Proper Packaging
In order to ship the system, please follow the guidelines below for proper packaging:
• Use the original packaging that your system arrived in.
• OR use a box with sufficient room for adequate
cushioning material for all sides of the container. Never exceed the maximum gross weight for the box (which is usually printed on the bottom lap).
• Properly cushion the contents of your package by wrapping each item separately. To prevent damage to the contents, ensure that they are separated from the corners and sides of the box. Various materials can be used for cushioning and protection: i.e. air-encapsulated plastic (bubble wrap), foam-in-place and/or corrugated dividers. Use enough cushioning material to ensure that the contents won’t move easily if the box is handled improperly.
• Make sure to properly close the box. For best results, use either pressure-sensitive plastic or water activated paper tape with at least two inches (5cm) width. Do not use masking tape, cellophane tape, string or paper to seal the box.
Page 60
59
Error Messages
An error message will be displayed on the CU’s screen whenever there is a problem with the system. All Treatment Heads are automatically tested prior to each use.
Error Messages
An error message will be displayed on the CU’s screen whenever
there is a problem with the system. All Treatment Heads are automatically tested prior to each use.
• Should a fault occur within the Treatment Head, the user will be informed with a “beep” and a message via the LCD.
• In some cases the operator may proceed with discretion.
• If your system requires service, please contact a Meditech
authorized service center.
Page 61
60
59
Trouble Shooting Guide
MESSAGES VIA LCD
POSSIBLE
CAUSE
SOLUTION
NO POWER or DISPLAY IS BLANK
Power supply not
connected.
Power not turned on.
Plug power supply into
hospital grade outlet.
Ensure cable is plugged into the Controller Unit.
Press POWER button.
LOCKED
Press any key to enter
password to unlock.
System is LOCKED to
avoid unauthorized use.
Each stage of a treatment specifies a
specific Treatment Head.
Plug in Treatment Head
Incorrect or no
Treatment Head plugged in.
Identify and plug in specified Treatment
Head.
Page 62
61
MESSAGES VIA
LCD
POSSIBLE CAUSE
SOLUTION
Head warning
Refer to warning code
L07 in manual
Treatment Head may be cold compared to the room temperature or needs calibration or
is malfunctioning.
All Treatment Heads are
tested prior to each use.
You can still continue to treat using the Treatment
Head, but use caution.
If this error persists, stop using the Treatment Head and send in for service.
Head warning
Refer to warning code
L06, LD8 in manual
Treatment Head may be hot or needs calibration or is
malfunctioning.
All treatment heads are
tested prior to each use. You can
still continue to
treat using the Treatment
Head,
but use extreme
caution.
Ensure that the Treatment
Head is cooled before continuing in order to prevent dermal skin
damage. Ask the patient to
inform you if the Treatment Head temperature becomes
uncomfor
table or too
warm.
If this error persists, stop using the Treatment Head and send in for service.
Page 63
62
MESSAGES VIA LCD
POSSIBLE CAUSE
SOLUTION
Head error Refer to error code ### in manual
Error codes: L01,
D01, D02, D03, L02, L03, L04, L09
Treatment Head
may need calibration or is malfunctioning.
Stop treating with the Treatment Head.
Contact a Meditech
authorized service center
for further assistance.
Provide error code to the service center.
Send Treatment Head in
for service.
Head warning
Refer to warning
code ### in manual Warning codes: L05
Treatment Head
may need calibration or is malfunctioning.
You may continue treating with the
Treatment Head
If the problem persists
contact a Meditech authorized service center
for further assistance.
Provide warning code to
the service center.
System failure Refer to error code
### in manual
Controller Unit (CU) or Treatment Head may need calibration or is malfunctioning.
Stop treating with the system.
Contact a Meditech authorized service center for further assistance. Provide error code to the service center.
Send system in for service.
Unit failure Refer to error code ### in manual
Error codes: E01, E02, E03, E10, E11
and E12
Controller Unit (CU) is malfunctioning.
Stop treating with the system.
Contact a Meditech authorized service center for further assistance. Provide error code to the service center. Send CU in for service.
Page 64
63
Specifications
Controller Unit
Function
Specifications
Mode of Operation
Waveform
Duration
Frequency Settings
Duty Cycle
Percent of Total Power
CW, Waveform
Sine, Square and Triangle
1-60 minutes (1 min resolution)
10-100Hz (10Hz resolution)
250, 500, 1000, 2500,
5000 & 10000Hz
10-90% (10% steps)
10-100% (10% steps)
Function
Specifications
Mode of Operation
Waveform
Duration
Frequency Settings
Duty Cycle
Percent of Total Power
CW, Waveform
Sine, Square and Triangle
0-3600 seconds (1 sec resolution)
10-10000Hz (1Hz resolution)
1-100% (1% steps)
1-100% (1% steps)
Table 1 Controller Unit (Customized Protocol)
Table 2 Controller Unit Specifications
Page 65
64
PARAMETER
UNITS
LD-I 75
LD-I 200
LD-R100
LS-R750
LS-I
1500
Effective laser- LED
class
3B
3B
3B 1 1
Total optical power
mW
75
180
100
750
1500
Effective surface area
cm2
0.10
0.10
0.10
75
75
Power density
mW/cm2
750
1800
750
10
20
Number of devices
1 1 1 179
179
Medium
GaAlAs
GaAlAs
AlGaInP
GaAlAs
GaAlAs
Type
LD
LD
LD
SLD
SLD
Coherence
mm 1 1
N/A 0 0
Wavelength
nm
830
830
660
660
840
Spectral Width
nm
3.0
3.0
N/A
25.0
50.0
Divergence
Deg
9,30
9,30
10,17
8,8
25,25
Device power
mW
75.0
180.0
75.0
4.2
8.5
MPE
18.19W/m2
18.19W/m2
6.36 J/m2 -
Optical density
2.18
2.48
2.03 -
NOHD
cm
0.0
34.3
49.3 -
Table 3 Treatment Head Specifications
Treatment Heads
Page 66
65
PARAMETER
UNITS
LS-R
250
LS-I
500
LS-B3000
SLD-
R50 0
SLD-
I-
1000
LS-B-2250
Effective laser-
LED class
1 1 1 1 1 1
Total optical
power
mW
250
500
1000(red)
2000(infrared)
500
1000
750(red)
1500 (infrared)
Effective
surface area
cm2
25
25
100
77
77
75
Power density
mW/cm2
10
20
10(red) 20 (infrared)
6.5
13.0
10(red) 20 (infrared)
Number of
devices
59
59
240
120
120
180
Medium
GaAlAs
GaAlAs
AlGaInP (red)
AlGaInP (infrared)
GaAl
As
GaAlAs
AlGaInP (red)
AlGaInP (infrared)
Type SLD
SLD
SLD
SLD
SLD
SLD
Coherence
mm 0 0 0 0 0 0
Wavelength
nm
660
840
660 (red) 840 (infrared)
660
840
660 (red) 840 (infrared)
Spectral Width
nm
25.0
50.0
36 (red) 80 (infrared)
25.0
50.0
36 (red) 80 (infrared)
Divergence
Deg
8,8
25,25
12,12 (red) 12,12 (infrared)
8,8
25,25
12,12 (red) 12,12 (infrared)
Device power
mW
4.2
8.5
4.2 (red)
8.5 (infrared)
4.2
8.5
4.2 (red)
8.5 (infrared)
MPE -
Optical density
W/cm2
-
NOHD
cm -
Page 67
66
Glossary
LED Light Emitting Diode
SLD Super Luminous Diode
LCD Liquid Crystal Display
LD Laser Diode
CU Controller Unit
EMC Electromagnetic Compatibility
MPE Maximum Permissible Exposure
NOHD Nominal Ocular Hazard Distance
Page 68
67
Material
Model
Controller Unit
CU
250 mW Red Large Surface Array Treatment Head
LS-R 250
500 mW Infrared Large Surface Array Treatment
Head
LS-I 500 750mW Red Large Surface Array Treatment Head
LS-R 750
1500mW Large Infrared Large Surface Array Treatment Head
LS-I 1500
1000mW(red)/2000mW(infrared) Large Surface
Array
DUO+ (LS-B 3000) 750 mW(red)/1500mW(infrared) Large Surface
Array
DUO (LS-B 2250) 500mW Red Super Luminous Diode Array
SLD-R 500
1000mW Infrared Super Luminous Diode Array
SLD-I 1000
BioFlex Single Laser LD-I 200
LD-I 200 BioFlex Single Laser LD-I 75
LD-I 75
BioFlex Single Laser LD-R 100
LD-R 100
Carrying Case
BAGT
BioFlex Pro Operational Manual
DOC 100.683
Therapist Clinical Manual
DOC 100.853
Therapist Power supply adapter
CT 100.121
Therapist Power Cord
CT 100.649
Table Fan
CT 100.334 USB Cable
DW 100.650
Treatment Sleeves
CT 100.336
Safety Kit
CT 100.340
Laser safety Guide
Laser Danger Sign
Two Safety Glasses
System Parts List:
The items below might not all be supplied with the system.
Page 69
68
Page 70
69
Page 71
70
415 Horner Ave. Toronto Ontario, M8W 4W3
Phone:(416)251-1055 Fax:(416)251-2446 Toll Free:(888)557-4004
www.biofl exlaser.com info@biofl exlaser.com
copyright©2014 Meditech International Incorporated. ALL RIGHTS RESERVED
MT-ENG-MN-100652
Loading...