you press the STEP key is changed in that Memory. If that Memory is being used again in the current
run, the original value will appear in each subsequent step using that memory. Only after EXITing a
run and returning to the Memory window will the altered STEPped value be visible.
When the gradient has finished, the display shows:
05-25% S11 1:15/88.7/22
FAST Gradient Done EXIT
Press EXIT once to repeat the run with optional editing, or three times to leave the NEW menu. The
screen asks if you want to save the gradient:
Short Sucr 05-25% wv 1St
SAVE Valid Grad or EXIT
If you press SAVE, the gradient is stored in a special memory that is not replaced when a new LIST is
loaded. It will appear in its proper place in the LIST, as well as in LAST and RCNT. If EXIT is pressed,
the gradient is not added to the LIST.
H. ENTERING GRADIENT RUN PARAMETERS ACQUIRED ELSEWHERE.
The gradient run lists that follow this manual are an expanded version of the gradients stored in LIST.
The extra gradients not in the memory involve special conditions (buffers, temperatures, other
solutes) that could not be represented in the limited display of the device. If you find one you want to
use or acquire it from any other source like a paper or from BioComp, enter in into the rotor memory
using NEW(see above) and it will appear in the LIST for that rotor, and in RCNT and LAST as well.
Follow the procedures in Section G to enter the new gradient.
I. FIXED ANGLE AND VERTICAL ROTORS:
USING THE ADJUSTABLE MARKER BLOCK AND
QUICK-SEAL TUBES
I. Setting the Marker Block:
Since the nipple-tipped Quick Seal tubes can not accept a cap during mixing, we supply the 8-place
fixed angle and vertical rotor tube holders with an adjustable marker block that can be altered to fit
the size of the sample you will load into the tube. The idea is to have the tube full to the very top
during mixing and then to remove the top of the gradient to make room for the sample. Here are the
steps to adjust the marker block to fit your sample volume:
1. Tare an empty tube.
2. Fill it to the top of the nipple with water and reweigh it.
3. Subtract your sample volume (1 ml = 1 g) from the total water weight and divide the remainder by
two. This is the half full volume.
4. Fill a new tube with this half full volume.
5. Loosen the set screw (3/32 allen wrench) on the base of the marker block and insert the half-full
tube.
6. While holding the tube so the bottom of the meniscus is just above the rim of the marker block,
slide the plunger up into the base of the marker block until it just touches the bottom of the tube.
Lock it in place by tightening the set screw. Check the adjustment by reinserting the tube and fine
tune the adjustment as necessary.
II. Forming gradients:
1. Make a half-full mark on the tube using the supplied fine tip permanent marker.
2. Add light solution to just above the line.
3. Layer the heavy solution under the light until the interface lies at the half full line.
4. Top up the light solution until the nipple is full. Use a fine needle-tipped syringe for this.
5. Form the gradient.