BioCision CoolCell LX Instructions For Use Manual

CoolCell
®
LX
Instructions for Use
CoolCell
Controlled-rate cell freezing container for 1ml or 2ml cryovials Controlled-rate cell freezing container for 1ml or 2ml cryovials
Quick Start
• The 12 chambers and cryovials should be dry to avoid tube sticking upon freezing.
• Place sample vials containing 1 ml of cell suspension in each well. Each well should contain a filled vial. If freezing batch is fewer than 12 vials, fill each empty well with a BioCision CoolCell Filler Vial (2ml vial, part number BCS-3105) or other vial that contains equivalent volume of freezing media.
Note: Cell suspensions can be inserted into a room temperature CoolCell and successfully preserved. For optimal results, CoolCell should be at the same temperature as your cell suspensions.
• Check that the tubes slide in and out freely.
• Fully seat the lid on CoolCell LX
• Place CoolCell LX upright into a -80OC freezer or dry ice locker. Ensure that there is at least one inch of free space clearance around CoolCell LX module.
• Freeze for four hours before transferring samples to archive storage.
Transferring frozen samples to archive storage
• Prepare an insulated pan or container with a one inch (2.5cm) layer of pulverized or pellet dry ice.
• Remove CoolCell LX from the freezer and gently remove the lid using a gentle twisting and rocking motion.
• Vial tops will be exposed once lid is removed and vials should be quickly extracted and placed onto the dry ice.
Special Notes:
• Always use dry ice to transfer cryovials containing cells to permanent storage to avoid temperature rise and cell damage. Cryovial contents can rise from -8o
O
C to over -50OC in less than one minute if exposed to
room temperature air.
• It is strongly recommended that all frozen cell cultures be checked for viability before the stock culture is terminated.
Recycling CoolCell LX to room temperature
CoolCell LX is ready to freeze again as soon as the foam body and core (black ring) are at room temperature. To rapidly recycle CoolCell LX to room temperature, remove the center solid core ring. CoolCell LX body and lid will return to room temperature in 10 to 15 minutes. Check that all chambers are dry. Dry the core ring before re-inserting into the central chamber.
Quick Start
• The 12 chambers and cryovials should be dry to avoid tube sticking upon freezing.
• Place sample vials containing 1 ml of cell suspension in each well. Each well should contain a filled vial. If freezing batch is fewer than 12 vials, fill each empty well with a BioCision CoolCell Filler Vial (2ml vial, part number BCS-3105) or other vial that contains equivalent volume of freezing media.
Note: Cell suspensions can be inserted into a room temperature CoolCell and successfully preserved. For optimal results, CoolCell should be at the same temperature as your cell suspensions.
• Check that the tubes slide in and out freely.
• Fully seat the lid on CoolCell LX
• Place CoolCell LX upright into a -80 free space clearance around CoolCell LX module.
• Freeze for four hours before transferring samples to archive storage.
Transferring frozen samples to archive storage
• Prepare an insulated pan or container with a one inch (2.5cm) layer of pulverized or pellet dry ice.
• Remove CoolCell LX from the freezer and gently remove the lid using a gentle twisting and rocking motion.
• Vial tops will be exposed once lid is removed and vials should be quickly extracted and placed onto the dry ice.
Special Notes:
• Always use dry ice to transfer cryovials containing cells to permanent storage to avoid temperature rise and cell damage. Cryovial contents can rise from -8o room temperature air.
• It is strongly recommended that all frozen cell cultures be checked for viability before the stock culture is terminated.
Recycling CoolCell LX to room temperature
CoolCell LX is ready to freeze again as soon as the foam body and core (black ring) are at room temperature. To rapidly recycle CoolCell LX to room temperature, remove the center solid core ring. CoolCell LX body and lid will return to room temperature in 10 to 15 minutes. Check that all chambers are dry. Dry the core ring before re-inserting into the central chamber.
About CoolCell LX
CoolCell LX, in combination with a -80
O
C freezer or dry ice locker, will provide the freezing rate of -1OC per min­ute that is ideal for cryo-preservation of most cultured cell lines. CoolCell LX design uses a combination of insulation foam, radial symmetry, and a heat transfer core to regulate heat loss rather than using a large thermal mass (alcohol-based freezing container). As a result, freezing profiles are extremely consistent from one run to the next. Also, because of this low thermal mass, CoolCell LX will not cause a rise in local freezer temperature and will protect nearby samples already stored in the freezer. Low thermal mass also means CoolCell LX will rapidly return to room temperature for another freezing cycle (see fast recycle instructions above).
CoolCell LX freezing performance
A temperature probe was placed into a 2ml cryovial con­taining 1ml of cryopreservative and the tube was inserted into a CoolCell LX sitting at room temperature. CoolCell LX was then placed directly into a -80°C freezer and the temperature rate and profile were observed over a 3 hour period. This experiment was repeated 5 consecutive times and temperature profiles were recorded. Conclusion: CoolCell LX showed identical cooling profiles and phase transition over five consecutive freeze cycles.
Troubleshooting
Care and cleaning
CoolCell LX is constructed of closed cell cross-linked polyethylene foam and a solid thermo-conductive core. CoolCell LX is compatible with prolonged cryogenic temperature exposure. The foam may be cleaned by water and mild soap. Rinse and dry thoroughly. CoolCell LX is resistant to alcohols and 10% bleach solutions. Do not autoclave. Maximum temperature exposure: 60
O
C. Avoid prolonged exposure to UV light sources.
CoolCell®L X2ThermalProle
0 50 100 150 200
-80
-60
-40
-20
0
20
Time in Minutes
CCLX-5
CCLX-1
CCLX-2
CCLX-3
CCLX-4
Problem Solution
Vials do not freely fit in the chambers
CoolCell LX is designed to fit standard screw-top 1ml and 2ml cryovials up to 13mm in diameter and up to 55mm in height. Check that flag style labels, if used, will not bind and hinder insertion or removal.
Vials are stuck in CoolCell after freezing
It is likely moisture was in the vial chambers or on the sample vial prior to freezing. Remove the core (black ring) and tap the CoolCell LX to dislodge vials.
The lid does not fully seat Ensure that sample tubes are fully seated in chamber. The maximum
height of the tube is 55mm.
BioCision, LLC 775 E. Blithedale, Suite 203, Mill Valley, CA 94941, USA, info@biocision.com, www.biocision.com
CoolCell LX Instructions, October 2012 CoolCell LX © 2012. BioCision, LLC. All rights reserved. Patents pending. BioCision, Standardizing Samples, CoolCell designations are trademarks owned by BioCision, LLC.
PURCHASE AND/OR USE OF THIS PRODUCT SHALL CONSTITUTE ACKNOWLEDGMENT AND ACCEPTANCE OF THE TERMS AND CONDITIONS OF THIS RESEARCH LICENSE AGREEMENT. The purchase of this product (“the Product”) conveys to the buyer
only the non-transferable right to use the product in internal non-commercial research (“Research Use”) conducted by the buyer. Research Use does not include the right to use the Product to manufacture products, including but not limited to cells or other biological materials, for subsequent sale; the use the Products to discover, develop, or test any commercial product; and the re-sale of Products, either alone or in combination with other products. The sale or any other transfer or distribution to third parties of (i) the Product, or (ii) any cells or other materials created using the Product is strictly prohibited under this Research License Agreement. Also, the use of cells or other materials created using the Product to discover, develop, or test any commercial product is strictly prohibited under this Research License Agreement. A commercial license is required for any commercial use of the Product, or any commercial use of cells or other materials discovered, developed, or tested using the Product. To obtain a commercial license to use the Product, please contact BioCision.The buyer’s rights to have and use the Product under the Research License Agreement will terminate immediately if the buyer en­gages in commercial use of the Product. Upon such termination of the buyer’s rights, the buyer agrees to return the Products to BioCision. This Research License Agreement shall not be assigned or otherwise transferred by the buyer.
About CoolCell LX
CoolCell LX, in combination with a -80 ute that is ideal for cryo-preservation of most cultured cell lines. CoolCell LX design uses a combination of insulation foam, radial symmetry, and a heat transfer core to regulate heat loss rather than using a large thermal mass (alcohol-based freezing container). As a result, freezing profiles are extremely consistent from one run to the next. Also, because of this low thermal mass, CoolCell LX will not cause a rise in local freezer temperature and will protect nearby samples already stored in the freezer. Low thermal mass also means CoolCell LX will rapidly return to room temperature for another freezing cycle (see fast recycle instructions above).
Troubleshooting
Care and cleaning
CoolCell LX is constructed of closed cell cross-linked polyethylene foam and a solid thermo-conductive core. CoolCell LX is compatible with prolonged cryogenic temperature exposure. The foam may be cleaned by water and mild soap. Rinse and dry thoroughly. CoolCell LX is resistant to alcohols and 10% bleach solutions. Do not autoclave. Maximum temperature exposure: 60
CoolCell®L X2ThermalProle
0 50 100 150 200
-80
-60
-40
-20
0
20
Time in Minutes
CCLX-5
CCLX-1
CCLX-2
CCLX-3
CCLX-4
Problem Solution
Vials do not freely fit in the chambers
Vials are stuck in CoolCell after freezing
The lid does not fully seat Ensure that sample tubes are fully seated in chamber. The maximum
BioCision, LLC 775 E. Blithedale, Suite 203, Mill Valley, CA 94941, USA, info@biocision.com, www.biocision.com
CoolCell LX Instructions, October 2012 CoolCell LX © 2012. BioCision, LLC. All rights reserved. Patents pending. BioCision, Standardizing Samples, CoolCell designations are trademarks owned by BioCision, LLC.
PURCHASE AND/OR USE OF THIS PRODUCT SHALL CONSTITUTE ACKNOWLEDGMENT AND ACCEPTANCE OF THE TERMS AND CONDITIONS OF THIS RESEARCH LICENSE AGREEMENT. The purchase of this product (“the Product”) conveys to the buyer
only the non-transferable right to use the product in internal non-commercial research (“Research Use”) conducted by the buyer. Research Use does not include the right to use the Product to manufacture products, including but not limited to cells or other biological materials, for subsequent sale; the use the Products to discover, develop, or test any commercial product; and the re-sale of Products, either alone or in combination with other products. The sale or any other transfer or distribution to third parties of (i) the Product, or (ii) any cells or other materials created using the Product is strictly prohibited under this Research License Agreement. Also, the use of cells or other materials created using the Product to discover, develop, or test any commercial product is strictly prohibited under this Research License Agreement. A commercial license is required for any commercial use of the Product, or any commercial use of cells or other materials discovered, developed, or tested using the Product. To obtain a commercial license to use the Product, please contact BioCision.The buyer’s rights to have and use the Product under the Research License Agreement will terminate immediately if the buyer en- gages in commercial use of the Product. Upon such termination of the buyer’s rights, the buyer agrees to return the Products to BioCision. This Research License Agreement shall not be assigned or otherwise transferred by the buyer.