Bien Air CHIROPRO L, CHIROPRO Instructions For Use Manual

CHIROPRO L / CHIROPRO
ENG Instructions for use
0120
REF 2100219-0003/2014.05ENG
Options
1x REF 1600613-001
1x REF 1500984-010
1x REF 1600631-001
1x REF 1303393-001
REF 1600598-001
REF 1600632-001
1x REF 1600755-001
1x REF 1600606-001
1x REF 1303711-010
10x
1x REF 1600631-001
REF 1600755-001
REF 1600825-001
REF 1600631-001
CA 20:1 L
CA 20:1
REF 1600692-001
CA 20:1 L Micro-Series
10x
1x REF 1500984-010
1x REF 1303393-001
10x
10x
Set CHIROPRO REF 1700387-001
1x REF 1600724-001
REF 1301560-010
10x
10x
REF 1500984-010
REF 1303711-010
REF 1600606-001
10x
REF 1600785-001 REF 1600786-001
CA 20:1 L KM CA 20:1 L KM
Micro-Series
REF 1303393-001
10x
REF 1501621-010
REF 1501635-010
1x REF 1600825-001
1x REF 1600606-001
1x REF 1303711-010
10x
CHIROPRO L / CHIROPRO
Summary
Starting display
Starting display
Available values
Available values
Implantology
Implantology
Endodontics
Surgery
Round bur 1
Round bur 1
Round bur 2
Round bur 2
Drill 1
Drill 1
Drill 2
Drill 2
Drill 3
Drill 3
Drill 4
Drill 4
Tapping
Tapping
Tap unscrewing
Tap unscrewing
Implant screwing
Implant screwing
Unscrewing
Unscrewing
100 - 40’000 rpm
100 - 40’000 rpm
with a CA 1 : 1
with a CA 1 : 1
Depends on the CA
Depends on the CA
0.48 - 4.8 Ncm
0.48 - 4.8 Ncm
with a CA 1 : 1
with a CA 1 : 1
Depends on the CA
Depends on the CA
30 ml/min 20%
30 ml/min 20%
60 ml/min 40%
60 ml/min 40%
90 ml/min 60%
90 ml/min 60%
120 ml/min 80%
120 ml/min 80%
150 ml/min 100%
150 ml/min 100%
Implantology
Endodontics
Surgery
Open pulp chamber
endo file 1
endo file 2
endo file 3
endo file 4
endo file 5
endo file 6
endo file 7
endo file 8
endo file 9
128:1
128:1
64:1
64:1
30:1
30:1
27:1
27:1
20:1
20:1
16:1
16:1
10:1
10:1
1:1
1:1
1:2
1:2
1:5
1:5
128:1
64:1
30:1
27:1
20:1
16:1
10:1
1:1
1:2
1:5
128:1
64:1
30:1
27:1
20:1
16:1
10:1
1:1
1:2
1:5
100 - 40’000 rpm
with a CA 1 : 1
Depends on the CA
0.48 - 4.8 Ncm
with a CA 1 : 1
Depends on the CA
30 ml/min 20%
60 ml/min 40%
90 ml/min 60%
120 ml/min 80%
150 ml/min 100%
Implantology
Endodontics
Surgery
Apical resection
Tooth extraction
Sinus lift
Free
100 - 40’000 rpm
with a CA 1 : 1
Depends on the CA
0.48 - 4.8 Ncm
with a CA 1 : 1
Depends on the CA
30 ml/min 20%
60 ml/min 40%
90 ml/min 60%
120 ml/min 80%
150 ml/min 100%
MAIN MENU
IMPLANT SYST.
Steps
Steps
Ratio
Ratio
Speed in rpm
Speed in rpm
Torque in Ncm
Torque in Ncm
Irrigation in ml/min
Irrigation in ml/min
CHIROPRO L
CHIROPRO
Straumann
Nobel Biocare
Zimmer
Dentsply Friadent
Biomet 3i
Astra Tech
Thommen Medical
Système
33
ENG
1
1 Symbols Page
Meaning of symbols 34
2
2 Description
Identification 35 Intended use 35 Environment 35 Environmental protection and information for disposal of the instrument 35
3
3 Set supplied 36
4
4 Options 36
5
5 Technical description
Technical data 37 Electromagnetic compatibility 38-39
6
6 Installation 40-41
7
7 Description of keys and elements 42
8
8 Operation
Description of functions 43 Start-up 44 Pre-settings (SETUP) 44-45 Description of programs 46
- Implantology
- Endodontics
*
- Surgery*
9
9 List of errors / Troubleshooting 47
10
10 Default values
Implantology 48+129 Endodontics* 48+130 Surgery* 48+130
11
11 Maintenance
Servicing 48 Information 48 Cleaning-disinfection 48 Important 48
12
12 Generalities and guarantee
General information 48 Terms of guarantee 48
Contenu
Table of contents
*only CHIROPRO L
34
ENG
CE Marking with number of the notified body.
Main switch ON - The instrument is switched on. OFF - The instrument is switched off.
Fuse Ø 5 x 20 mm.
Alternating current.
Device of type B.
CAUTION ! Dangerous voltage.
Element sensitive to electrostatic discharges.
Warning.
CAUTION! Refer to the accompanying documents.
Danger of pinching. Do not put your fingers in rotating parts.
Machine washable.
Variability in steps.
Symbol for “Water-cooling”.
Recyclable materials.
Recyclable electrical and electronic materials.
Sterilisable in autoclave up to the specified temperature.
Operating mode intermittent.
Manufacturer.
Light.
Expiration date.
Product containing phthalates.
Disposable product.
Sterilise with Ethylene Oxyde
0120
STERILE EO
1
1 Meaning of symbols
35
ENG
*only CHIROPRO L
2
2 Description
Electronically controlled tabletop device for dentistry allowing operation of an MX-i LED/MX-i micro­motor with variable speed control by a pedal. A peristaltic pump conveys the physiological liquid via a disposable irrigation line without being contaminated. The device's LCD display indicates the stage of implant fitting, the instrument's ratio, the bur speed, torque value and irrigation flow setting.
The system is to be used by dentists and surgeons in dental offices and hospitals. The system is des­igned to control a dental micromotor which can drive a dental hand-piece fitted with appropriate tools to cut hard and soft tissues in the mouth and to screw dental implants.
The system is intended for use in dentistry for implantology, dental surgery* and endodontic* work. Any use other than that for which this product is intended is unauthorised and may be dangerous. The medical device meets all the current legal requirements.
The device is not designed for use in an explosive atmosphere (anaesthetic gas).
Working Temperature: +10°C (50°F) to +25°C (77°F)
Relative humidity: 30% to 80%, including condensation Atmospheric pressure: 700 hPa to 1060 hPa
Transport Environmental conditions and storage Temperature: -25°C (-13°F) to +70°C (158°F)
Relative humidity: 10% to 100%, including condensation Atmospheric pressure: 500 hPa to 1060 hPa
The disposal and/or recycling of materials must be performed in accordance with the legislation in force.
This device and its accessories must be recycled. Electrical and electronic equipment may contain dangerous substances which constitute health and environmental hazards. The user must return the device to its dealer or establish direct contact with an approved body for treatment and recovery of this type of equipment (European Directive 2002/96/EC).
Identification
Environment
Environmental protection and infor­mation for disposal of the instrument
Intended use
36
ENG
3
3 Set supplied
4
4 Options
Contra-angle handpiece CA 20:1 L (light) REF 1600598-001 Contra-angle handpiece CA 20:1 L Micro-Series (light) REF 1600692-001 Contra-angle handpiece CA 20:1 L KM (light) REF 1600785-001 Contra-angle handpiece CA 20:1 L KM Micro-Series (light) REF 1600786-001 MX-i LED micromotor REF 1600755-001 MX-i micromotor REF 1600825-001 Cable for MX-i LED micromotor REF 1600606-001 Pedal 3 buttons REF 1600631-001 Pack of 10 disposable sterile lines REF 1500984-010 Kirschner/Meyer type detachable irrigation set for CA 20:1 L KM and CA 20:1 L KM Micro-Series, comprising 10 rings and 10 tubes REF 1501621-010 Kirschner/Meyer pack of 10 disposable sterile lines REF 1501635-010 10 attachments collars for fastening the sterile irrigation line to a cable REF 1303711-010 Bracket for fluid bottle REF 1303393-001 Support REF 1301575-001 Cable system 3P, Switzerland, length 2.00 m REF 1300065-001 Cable system 3P, Europe, length 2.50 m REF 1300066-001 Cable system 3P, US / Asia, length 2.00 m REF 1300067-001 10x Fuse T4.0A L 250 VAC breaking capacity 40A REF 1301560-010
1x CHIROPRO L control REF 1600613-001 1x Micromotor MX-i LED REF 1600755-001 1x Cable for MX-i LED micromotor REF 1600606-001 1x Pack of 10 disposable sterile lines REF 1500984-010 1x 10 attachment collars for fastening the sterile irrigation line to a cable REF 1303711-010 1x Bracket for fluid bottle REF 1303393-001 1x Pedal 3 buttons REF 1600631-001 1x Cable system 3P, Switzerland, length 2.00 m REF 1300065-001 1x Cable system 3P, Europe, length 2.50 m REF 1300066-001 1x Cable system 3P, US/Asia, length 2.00 m REF 1300067-001 1x Instructions REF 2100219
1x CHIROPRO control REF 1600724-001 1x Micromotor MX-i REF 1600825-001 1x Cable for micromotor MX-i REF 1600606-001 1x Pack of 10 disposable sterile lines REF 1500984-010 1x 10 attachment collars for fastening the sterile irrigation line to a cable REF 1303711-010 1x Bracket for fluid bottle REF 1303393-001 1x Pedal 3 buttons REF 1600631-001 1x Cable system 3P, Switzerland, length 2.00 m REF 1300065-001 1x Cable system 3P, Europe, length 2.50 m REF 1300066-001 1x Cable system 3P, US/Asia, length 2.00 m REF 1300067-001 1x Instructions REF 2100219
Set CHIROPRO L REF 1700298-001
Set CHIROPRO REF 1700387-001
37
ENG
*only CHIROPRO L
5
5 Technical Description: Technical data
Voltage
100 – 240 VAC 50 – 60 Hz
Fuses
2 fuses T4.0A L 250 VAC, breaking capacity 40A
Power demand
- 100 V /300 VA
- 240 V /300 VA
Classification
Class IIa in accordance with European Directive 93/42/EEC concerning medical devices.
Electric insulation class
Class I, per IEC 60601-1 (apparatus protected against electric shocks).
Degree of protection
IP 40 (protection against insertion of objects larger than 1 mm).
Dimensions L x W x H
309 x 220 x 123 mm. Height with bracket 506 mm
Weight
Housing 2.7 kg Pedal 830 g Cable 105 g Bracket 115 g
Memory
Implantology mode: Storage in memory of 8 implant fitting
sequences of 10 steps each.
Endodontics
* mode: Storage in memory of an endodontics
sequence of 10 steps.
Surgery
* mode: Storage in memory of 4 separate
programs.
Languages
French, German, English, Italian, Spanish, Portuguese, Japanese and Russian.
List of errors & Troubleshooting
Page 47
Bracket for physiological liquid flask
Stainless steel
Intended for use with: see instructions Micromotor MX-i LED REF 2100245 Micromotor MX-i REF 2100245 Cable for micromotor REF 2100163 Contra-angle CA 20:1, without light REF 2100209 Contra-angle CA 20:1 L, with light REF 2100209 Contra-angle CA 20:1 L Micro-Series, with light REF 2100209 Contra-angle CA 20:1 L KM, light REF 2100209 Contra-angle CA 20:1 L KM Micro-Series, light
REF 2100209
The use of the system with other handpieces, motors
or cables has not been validated/certified
Peristaltic pump
Pump delivery: From 30 to 150 ml/min. (5 levels). Hose for pump: External Ø 5.60 mm,
internal Ø 2.40 mm Wall thickness 1.60 mm.
Pedal
REF 1600631-001 Dimensions (LxWxH) 250 x 205 x 54 mm with handle: 250 x 205 x 144 mm The pedal is waterproof (IP X8 in accordance with CEI 529).
Cables
Length of cables: Pedal cable 2.90 m Motor cable 2.00 m
WARNING
To prevent any risk of electric shock, this device must be connected only to a power supply network provided with protective earth. Modification of the device forbidden. The system is not adapted to be used in the presence of inflammable gases (e.g. anaesthetic gas). Do not attempt to open the apparatus when it is connected to the electric mains. Beware of electric shocks.
Applied parts (per IEC 60601-1)
MX-i LED micromotor REF 1600755-001 MX-i Micromotor REF 1600825-001 Cable for micromotor MX-i LED REF 1600606-001 CA 20:1 L REF 1600598-001 CA 20:1 REF 1600632-001 CA 20:1 L Micro-Series REF 1600692-001 CA 20:1 L KM REF 1600785-001 CA 20:1 L KM Micro-Series REF 1600786-001 Irrigation lines REF 1500984-010 KM Irrigation lines REF 1501635-010 KM detachable Irrigation set REF 1501621-010
Operating mode:
Intermittent ON: 5 min. OFF: 40 min.
38
ENG
5
5 Technical Description: Electromagnetic compatibility
Electro-medical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this document. CHIROPRO L/CHIROPRO complies with the EMC requirements according to IEC 60601-1-2. Radio trans­mitting equipment, cellular phones, etc. shall not be used in close proximity of the device since they could influence the performance of the device. Particular precaution is required when using strong emis­sion sources such as High Frequency surgical equipment and similar equipment so that the HF cables are not routed on or near the device. If in doubt, please contact a qualified technician or Bien-Air Dental. CHIROPRO L/CHIROPRO should not be used adjacent or stacked with other equipment. If adjacent or stacked use is necessary, CHIROPRO L/CHIROPRO should be monitored to verify normal operation in the configuration in which it will be used.
Precautions regar­ding Electromagnetic Compatibility (EMC)
The use of accessories, transducers and ca-bles other than those specified, with the exception of transducers and cables sold by Bien-Air Dental as replacements parts for internal compo­nents, may result in increased emissions or decreased immunity of CHIROPRO L/CHIROPRO. Dental professionals need to be aware of potential electromagnetic interference between elec­tronic dental devices and active implantable medical devices, and should always inquire about any devices implanted in the patient.
WARNING!
CHIROPRO L/CHIROPRO is intended for use in the electromagnetic environment specified below. The customer or the user of CHIROPRO L/CHIROPRO should ensure that it is used in such an envi­ronment.
Guidance and manu­facturer’s declaration
- electromagnetic emissions
Emissions test Compliance Electromagnetic environment - guidance
RF emissions Group 1 CHIROPRO L/CHIROPRO uses RF energy only for its internal function. CISPR 11 Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions Class B CISPR 11 CHIROPRO L/CHIROPRO is suitable for use in all Establishments,
including domestic establishments and those directly connected Harmonic emissions Not applicable to the public low-voltage power supply network that supplies IEC 61000-3-2 buildings used for domestic purposes.
Voltage fluctuations/flicker emissions IEC 61000-3-3 Compliant
Guidance and manufacturer’s declaration - electromagnetic immunity
CHIROPRO L/CHIROPRO is intended for use in the electromagnetic environment specified below. The customer or the user of CHIROPRO L/CHIROPRO should ensure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic ±6 kV contact ±6 kV contact discharge (ESD)
±8 kV air ±8 kV air IEC 61000-4-2 Electrical fast ±2 kV for power ±2 kV for power transient burst supply lines supply lines IEC 61000-4-4 ±1 kV for lines no ±1 kV for lines no
input/output input/output Shock waves ±0.5 kV line to line ±0.5 kV line to line IEC 61000-4-5 ±1 kV line to line ±1 kV line to line
±0.5 kV line to earth ±0.5 kV line to earth
±1 kV line to earth ±1 kV line to earth
±2 kV line to earth ±2 kV line to earth Voltage dips <5% U
T
<5% U
T
and outages (>95% dip in UT) (>95% dip in UT)
for 0.5 cycle for 0.5 cycle
40% U
T
40% U
T
(60% dip in UT) (60% dip in UT)
for 5 cycles for 5 cycles
70% U
T
70% U
T
IEC 61000-4-11 (30% dip in UT) (30% dip in UT)
for 25 cycles for 25 cycles
<5% U
T
<5% U
T
(>95% dip in UT) (>95% dip in UT)
for 5 sec for 5 sec Power frequency (50/60 Hz) 3 A/m 3 A/m magnetic field IEC 61000-4-8
NOTE U
T
is the a.c. mains voltage prior to application of the test level. Essential performance: The essential performance is the maintaining of the visual lighting intensity of the LED and the maintaining of motor speed. Maximum allowed speed deviation is ± 5%.
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commer­cial or hospital environment.
Mains power quality should be that of a typical commer­cial or hospital environment.
Mains power quality should be that of a typical commer­cial or hospital environment. If the user of CHIROPRO L/ CHIROPRO requires continued operation during power mains interruptions, it is recommended that CHIROPRO L/ CHIROPRO be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
39
ENG
5
5 Technical Description: Electromagnetic compatibility
Portable and mobile RF communications equipment should be used no closer to any part of CHIROPRO L/CHIROPRO, including cables, than the recom­mended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separa­tion distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnet­ic site survey,ashould be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
Guidance and manufacturer’s declaration - electromagnetic immunity
CHIROPRO L/CHIROPRO is intended for use in the electromagnetic environment specified below. The customer or the user of CHIROPRO L/CHIROPRO should ensure that it is used in such an environment.
Immunity test IEC 60601 Compliance Electromagnetic environment - guidance
test level level
Conducted RF 3 Vrms 3 V IEC 61000-4-6 150 kHz to 80 MHz
Radiated RF 3 V/m 3 V/m IEC 61000-4-3 80 MHz to 2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
CHIROPRO L/CHIROPRO is used exceeds the applicable RF compliance level above, the CHIROPRO L/CHIROPRO should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
CHIROPRO L/CHIROPRO.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the CHIROPRO L/CHIROPRO
The CHIROPRO L/CHIROPRO is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The custo­mer or the user of the CHIROPRO L/CHIROPRO can help prevent electromagnetic interference by maintaining a minimum distance between por­table and mobile RF communications equipment (transmitters) and the CHIROPRO L/CHIROPRO as recommended below, according to the maxi­mum output power of the communications equipment.
Rated maximum output Separation distance according to frequency of transmitter power of transmitter m
W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
40
ENG
6
6 Installation
Installation
A. CHIROPRO L/CHIROPRO may be positioned on a table, on a trolley or
another surface, but in no circumstances on the floor. Power plug is the device for disconnection in case of problems, and it must be easily accessible at all times.
B. The fuse box may be opened with a screwdriver.
100 - 240 Vac = fuse T-4.0 A L 250 VAC REF 1301560-010
C. The equipment is powered by your line voltage (100/115/230 Vac).
Connect the power cable to the plug fig. 1.
D. Connect the pedal cable to the output provided on the rear panel,
guiding the connector and plug by means of the index pin on the connector fig. 2.
Do not raise the pedal using the connection cable.
E. Connect the micromotor cable to the motor output, guiding the
connector and plug by means of the index pin on the connector fig. 3.
F. Align and attach the bracket to the housing provided on the console’s
rear and suspend the flask or bottle fig. 4.
G. Check the packaging integrity, as well as the expiry date of the irrigation
line. Only lines supplied by Bien-Air Dental ensure trouble-free operation. These lines are sterile and for single use. Re-use may result in microbio­logical contamination of the patient.
Fuse
fig. 1
fig. 2
fig. 3
fig. 4
fig. 5
ON/OFF
Plug
=
>
16
14
7
41
ENG
6
6 Installation
fig. 8
H. Remove the single-use sterile irrigation line from its pouch.
Fitting on the spray tube
I. Connect the flexible hose of the irrigation line to the spray tube of the
handpiece or contra-angle fig. 7.
Installation on the peristaltic pump
J. Install the plastic cassette in the peristaltic pump.
Check that the cassette is clipped correctly. Close the pump lid, fig. 8. If there is resistance to closing, open the lid again and check the correct positioning of the cassette.
Warning!
Do not run the pump while the lid is open.
Danger of pinching!
K. Perforate the cap of the physiological liquid flask with the pointed end
of the irrigation line after removing the protective cap.
L. Attach the irrigation line on the motor cable using the attachment collars
REF 1303711-010 fig. 9.
The device can be safely stopped using the main switch .
fig. 6
fig. 7
fig. 9
REF 1303711-010
16
Stopping procedure
42
ENG
*only CHIROPRO L
7
7 Description of keys and elements
0 DEVICE SCREEN
1 COMMAND TO REVERSE THE ROTATION OF
THE MICROMOTOR
2 “SETUP” MENU CALL-UP KEY
3 PARAMETERS BACKUP KEY
4 RETURN KEY
u COMMANDS DEVICE
Down key Up key Left key (–) Right key (+) Confirmation/selection key In implantology/endodontics
* mode: next stage
6 PERISTALTIC PUMP LID
7 MICROMOTOR CONNECTOR
8 IRRIGATION ON/OFF CONTROL BUTTON
ON PEDAL
9 “PROGRAM” BUTTON ON PEDAL
In implantology/endodontics
* mode:
- Short press: next stage
- Long press: previous stage
: BUTTON TO REVERSE THE ROTATION
OF THE MICROMOTOR ON PEDAL
; VARIABLE SPEED DRIVE ON PEDAL
< PEDAL CONNECTOR
= MAINS CONNECTOR (100/115/230 VAC)
> FUSE HOLDER
? MAIN SWITCH ON THE DEVICE
@ LABEL
A BRACKET SUPPORT
B MICROMOTOR
Selection cursor
Instrument transmission ratio
Torque setting
Speed adjustment
Adjusting irrigation
ROUND BUR
20:1
28.0 Ncm
1’000
Name of program
Reversal of motor rotation symbol
Motor overheating
Irrigation stopped
Only in SETUP mode Activation/deactivation of the stage
OK
cursor
adjusting
RPM
43
ENG
*only CHIROPRO L
8
8 Operation
The "reverse" function can be chosen directly in all the programs except endodontics*. Upon selection, a beep and the "reversal of motor rotation" icon indicate reverse rotation. For endodontics
*, settings must be performed under SETUP (Auto-forward/Auto-reverse).
See next page.
Stores the settings of a program: press the key until a beep is emitted, and the values that are flashing will be stored in the memory directly.
Return function. With "ESC", you can leave the current screen. In “Implantology” and “Endodontics
*
mode, can also be used to return to the previous stage. If the name of the program flashes when exiting, the changes will not be taken into consideration. The changes must always be confirmed with "SAVE", otherwise they will be lost.
BLUE Activation/Deactivation of the peristaltic pump.
GREY Activation/Deactivation of reversal of motor rotation.
ORANGE Program key:
- Short press next stage
- Long press previous stage
ESC
SETUP
REV
SAVE
Description of functions
CHIROPRO L
Motor speed control pedal.
1
2
3
4
1
2
3
4
44
ENG
*only CHIROPRO L
8
8 Operation
SETUP
Pre-settings
Start-up
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
Editor
Editor
Editor
Editor
System info
Restore defaults
English
1.
2.
3. 4.
OK: continue ESC: back
5.
OK: continue
*6.
with OK: go directly to pre-setting with no possibility of deactiva­ting the stages
or with SETUP, possibility of de­activating the stages with
Français
Deutsch
Italiano
Español
Português
Russian
Japanese
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
System info
Restore defaults
Straumann
Nobel Biocare
Zimmer
Dentsply Friadent
Biomet 3i
Astra Tech
Thommen Medical
System
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
System info
Restore defaults
Endo steps
Endo parameters
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
System info
Restore defaults
128:1
64:1
30:1 27:1
20:1
16:1
10:1
1:1
1:2
1:5
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
System loading.......
English
Implantology
Please check the pre-programmed values before inserting implant.
Straumann
LANGUAGE
INFORMATION
MAIN MENU
<Name of selected system>
IMPLANT. SYSTEM
Français
Key functions: move cursor UP
move cursor DOWN
decrem./disable
increm./enable
Endodontics*Nobel Biocare
Deutsch
Surgery*Zimmer
Italiano
Dentsply Friaden
Espanol
Biomet 3i
Português
Astra Tech
Russian
Thommen Medical
Japanese
System
All the pre-programmed settings are indicative and MUST be validated by the user.
This storing in memory takes place only at the first connection of the device and is subsequently maintained. These parameters can then be modified in SETUP.
Select the language wanted
and confirm with OK.
Select the ratio to be changed OK, then change of value with and with then save with SAVE. Continue with OK. ESC: back
Select the system wanted
and confirm with OK.
ESC: change
Select with then OK. If modifications with , the screen flashes. save with SAVE continue with OK. Without modification, continue with OK. ESC: back
Select with Confirmation with OK
Select with Confirmation with OK
Select with Confirmation with OK
45
ENG
*only CHIROPRO L
8
8 Operation
Pre-settings
Editor
Editor
Editor
Editor
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
System info
Restore defaults
Light OFF
Light ON
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
System info
Restore defaults
Editor
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
System info
Restore defaults
Implantology
Endodontics*
Surgery*
Implantology (Ncm)
Endodontics* (mNm)
Surgery* (Ncm)
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
System info
Restore defaults
Implant. system
Language
Endo* system
Ratio
Light
Footpedal
Torque units
Contrast
System info
Restore defaults
OK
OK
OK
OK
OK
OK
OK
ON or OFF with then OK. Display under light ON: Level: adjustment with Time delay: setting with , save with SAVE, continue with OK. ESC: back
ON/OFF or progressive with
, save with SAVE, continue with OK. ESC: back
Adjustable with , save with SAVE, continue with OK. ESC: back
Contrast adjustment with , save with SAVE, continue with OK.
System Info: software version, serial number and electronics of the device.
Restore system: Can be used to reinitialise factory settings.
ROUND BUR 1
ROUND BUR 2
DRILL 1
DRILL 2
DRILL 3
DRILL 4
TAPPING
TAP UNSCREWING
IMPLANT SCREWING
UNSCREWING
Implantology (Ncm)
Surgery* (Ncm)
OK
OK
System names
Tool names
OK
OK
Used to rename or customise the name of the system, tool or treatment. Select with then OK,choose the letters on the keypad by moving the cursor using
then OK, save the new name
with SAVE, ESC: back
Apical resection
Tooth extraction
Sinus Lift
Free
OK
Straumann
Nobel Biocare
Zimmer
Dentsply Friadent
Biomet 3i
Astra Tech
Thommen Medical
System
OK
Straumann
Nobel Biocare
Zimmer
Dentsply Friadent
Biomet 3i
Astra Tech
Thommen Medical
System
OK
46
ENG
*only CHIROPRO L
8
8 Operation
Description of functions
Pre-programmed values, see pages 129-130
Implantology
Endodontics*
Surgery*
ROUND BUR 1
ROUND BUR 2
DRILL 1
DRILL 2
DRILL 3
DRILL 4
TAPPING
TAP UNSCREWING
IMPLANT SCREWING
UNSCREWING
100 - 40’000 rpm
with a CA 1:1
Depends on the
CA selected
0.48 - 4.8 Ncm
with a CA 1:1
Depends on the
CA selected
30 ml/min 20%
60 ml/min 40%
90 ml/min 60%
120 ml/min 80%
150 ml/min 100%
Implantology
Endodontics*
Surgery*
OPEN PULP CHAMBER
ENDO FILE 1
ENDO FILE 2
ENDO FILE 3
ENDO FILE 4
ENDO FILE 5
ENDO FILE 6
ENDO FILE 7
ENDO FILE 8
ENDO FILE 9
Implantology
Endodontics*
Surgery*
Apical resection
Tooth extraction
Sinus Lift
Free
MAIN MENU
Selection wanted
and confirm
with
OK
Steps
Each of these stages can be activated or deactivated from the SETUP menu. See also info on the last page.
OK: next step ESC: previous step
Trans­mission ratio
Adjustable:
then SAVE
Speed in rpm
Adjustable:
then SAVE
Torque in Ncm
Adjustable:
then SAVE
Irrigation in ml/min
Adjustable:
then SAVE
Select with the cursor
128:1
64:1
30:1
27:1
20:1
16:1
10:1
1:1
1:2
1:5
Selection cursor
Instrument transmission ratio
Torque setting
Speed adjustment
Adjusting irrigation
ROUND BUR
20:1
28.0 Ncm
1’000
Name of program
Reversal of motor rotation symbol
Motor overheating
Irrigation stopped
Only in SETUP mode Activation/deactivation of the stage
RPM
47
ENG
9
9 List of errors & Troubleshooting
Message
Release the pedal
The pedal is pressed when starting the device.
The motor is blocked for more than 2 sec. The motor control card limits the power supplied to the motor to prevent motor overheating.
Cause of error
Safety
Safety
Action
Release the pedal and press again.
Avoid extended use.
Equipment initialisation error
The following error may occur at start-up of CHIROPRO L/CHIROPRO
1. Check on the integrity of the CHIROPRO L/CHIROPRO memory
INIT ERROR 1 The memory is corrupt!
Please contact Bien-Air Dental SA. ESC: restore
The memory data check failed. Press the ESC key to try to
restore the memory. Contact Bien-Air Dental SA.
Device operating error
The following errors may occur during operation of the device
1. Loss of pedal connection
ERROR 1 The pedal is not connected!
Please check the connection. ESC: exit
The pedal is not connected correctly. Check pedal connection.
Contact Bien-Air Dental SA.
2. Peristaltic pump overheating
ERROR 2 Irrigation pump overheating!
Please wait for it to cool.
ESC: exit
Peristaltic pump motor overheating Wait until the system cools.
Contact Bien-Air Dental SA.
3. Peristaltic pump general error
ERROR 3 Irrigation pump fault!
Please contact Bien-Air Dental SA. ESC: exit
Peristaltic pump electrical fault. Contact Bien-Air Dental SA.
4. Loss of motor connection
ERROR 4 The motor is not connected!
Please check the connection. ESC: exit
Loss of motor phase fault. The motor is not connected correctly.
Check motor connection. Contact Bien-Air Dental SA.
5. Motor cable fault
ERROR 5 Motor cable fault!
Please change cable. ESC: exit
Motor power fault. The motor cable may be defective.
Check motor cable. Contact Bien-Air Dental SA.
6. Motor control overheating
ERROR 6 System overheating!
Please wait for it to cool. ESC: exit
Overheating of motor control card (electrical control of motor).
Wait until the system cools. Contact Bien-Air Dental SA.
7. System electrical fault
GEN ERROR [Error code]
System electrical fault! Please contact Bien-Air Dental SA. ESC: exit
Communication fault with motor control card: [EC100]
Motor control card power supply undervoltage: [EC101]
Motor control card power supply overvoltage: [EC102]
Other motor control card faults: [EC120]
Contact Bien-Air Dental SA.
48
ENG
*only CHIROPRO L
Implantology: Default values page 129
The table shows the default operating values for the 8 implan­tology systems pre-programmed in the system, namely:
• Straumann • Nobel Biocare • Zimmer • Dentsply Friadent
• Biomet 3i • Astra Tech • Thommen Medical • System, available to the user (default settings identical to the Straumann system).
Endodontics*: Default values page 130
The table shows the default operating values for the endodon­tics sequence.
Surgery*: Default values page 130
The table shows the default operating values for 4 types of surgical operations proposed by the system, namely: • Root resection • Extraction of wisdom teeth • Sinus raising • Free program, left available to the user.
10
10 Default values
Only use original Bien-Air Dental maintenance products and parts or those recommended by Bien-Air Dental. Using other products or parts may cause operational failure and/or void the guarantee.
Servicing
Never disassemble the device. For any modification and repair, we recommend that you contact your regular supplier or Bien­Air Dental directly. Bien-Air Dental asks the user to have its dynamic instruments checked or inspected at least once a year.
Information
The technical specifications, illustrations and dimensions con­tained in these instructions are given only as a guide. They may not be the subject of any claim. The manufacturer reserves the right to make technical improvements to its equip­ment, without amending these instructions. For all additional information, please contact Bien-Air Dental SA at the address indicated on the back cover.
Cleaning-disinfection
• Disinfect the surfaces of the console and pedal with a clean cloth soaked in a suitable product. • Do not exert any pres­sure on the screen. • Do not immerse in disinfectant solution
• Not designed for an ultrasonic bath. • Use a new sterile irri­gation line for each patient.
Important
For maintenance: see instructions Micromotor MX-i LED REF 2100245 Micromotor MX-i REF 2100245 Cable for micromotor REF 2100163 Contra-angle CA 20:1, without light REF 2100209 Contra-angle CA 20:1 L, with light REF 2100209 Contra-angle CA 20:1 L Micro-Series, with light REF 2100209 Contra-angle CA 20:1 L KM, light REF 2100209 Contra-angle CA 20:1 L KM Micro-Series, light
REF 2100209
General information
The device must be used by a qualified professional in compli­ance with the current legal provisions concerning workplace safety, health and accident prevention measures, and these working instructions. In accordance with such requirements, the operators:
• must only use devices that are in perfect working order; in the event of irregular functioning, excessive vibration, abnor­mal heating or other signs that may indicate malfunction of the device, the work must be stopped immediately; in this case, contact a repair centre that is approved by Bien-Air Dental;
• must ensure that the device is used only for the purpose for which it is intended, must protect themselves, their patients and third parties from any danger, and must avoid contamina­tion through the use of the product.
Terms of guarantee
Bien-Air Dental grants the user a guarantee covering all func­tional defects, material or production faults. The device is cov­ered by this guarantee for 24 months from the date of invoic­ing. In case of justified claim, Bien-Air Dental or its authorised
representative will fulfil the company’s obligations under this guarantee by repairing or replacing the product free of charge. Any other claims, of whatever nature, in particular in the form of a claim for damages and interest, are excluded.
Bien-Air Dental shall not be held responsible for damage or injury and the consequences thereof, resulting from:
•excessive wear and tear • improper use • non-observance of the instructions for installation, operation and maintenance
•unusual chemical, electrical or electrolytic influences • poor connections, whether of the air, water or electricity supply.
The guarantee does not cover flexible “fibre optic” type light conductors, or any parts made of synthetic materials.
The guarantee shall become null and void if the damage and its consequences are due to improper manipulation of the product, or modifications to the product carried out by persons not authorised by Bien-Air Dental. Claims under the terms of the guarantee will be considered only on presentation, togeth­er with the product, of the invoice or the consignment note, on which the date of purchase, the product reference and the Serial No. should be clearly indicated.
11
11 Maintenance
12
12 Generalities and guarantee
129
Default values
Implantology: default values
All the pre-programmed settings are indicative and MUST
be validated by the user before starting the treatment/
operation. Always follow the implant manufacturer recommendations.
The table shows the default operating values for the 8 implantology systems pre-programmed in the system, namely:
• Straumann • Nobel Biocare • Zimmer • Dentsply Friadent • Biomet 3i • Astra Tech • Thommen Medical
• System, available to the user (default settings identical to the Straumann system).
STRAUMANN /
SYSTEM
NOBEL BIOCARE ZIMMER
DENTSPLY FRIADENT
BIOMET 3i ASTRA TECH
THOMMEN
MEDICAL
ROUND BUR 1
20:1
28.1 Ncm
1’000 RPM
ROUND BUR 2
20:1
28.1 Ncm
1’000 RPM
DRILL 1
20:1
35.3 Ncm 800 RPM
DRILL 2
20:1
35.3 Ncm 600 RPM
DRILL 3
20:1
35.3 Ncm 500 RPM
DRILL 4
20:1
35.3 Ncm 400 RPM
TAPPING
20:1
35.3 Ncm
15 RPM
TAP UNSCREWING
20:1
42.5 Ncm
15 RPM REV
IMPLANT SCREWING
20:1
35.3 Ncm
15 RPM
UNSCREWING
20:1
54.7 Ncm
15 RPM REV
ROUND BUR
20:1
28.1 Ncm
2’000 RPM
PILOT DRILL
20:1
35.3 Ncm 800 RPM
DRILL 1
20:1
35.3 Ncm 800 RPM
DRILL 2
20:1
35.3 Ncm 800 RPM
DRILL 3
20:1
35.3 Ncm 800 RPM
DRILL 4
20:1
35.3 Ncm 800 RPM
TAPPING
20:1
35.3 Ncm
15 RPM
TAP UNSCREWING
20:1
42.5 Ncm
15 RPM REV
IMPLANT SCREWING
20:1
35.3 Ncm
15 RPM
UNSCREWING
20:1
54.7 Ncm
15 RPM REV
ROUND BUR
20:1
28.1 Ncm
1’000 RPM
DRILL 1
20:1
35.3 Ncm 800 RPM
DRILL 2
20:1
35.3 Ncm 800 RPM
DRILL 3
20:1
35.3 Ncm 800 RPM
DRILL 4
20:1
35.3 Ncm 800 RPM
DRILL 5
20:1
35.3 Ncm 800 RPM
TAPPING
20:1
35.3 Ncm
15 RPM
TAP UNSCREWING
20:1
42.5 Ncm
15 RPM REV
IMPLANT SCREWING
20:1
35.3 Ncm
15 RPM
UNSCREWING
20:1
54.7 Ncm
15 RPM REV
ROUND BUR
20:1
28.1 Ncm
1’200 RPM
DRILL 1
20:1
35.3 Ncm 800 RPM
DRILL 2
20:1
35.3 Ncm 800 RPM
DRILL 3
20:1
35.3 Ncm 800 RPM
DRILL 4
20:1
35.3 Ncm 800 RPM
DRILL 5
20:1
35.3 Ncm 800 RPM
TAPPING
20:1
35.3 Ncm
15 RPM
TAP UNSCREWING
20:1
42.5 Ncm
15 RPM REV
IMPLANT SCREWING
20:1
35.3 Ncm
15 RPM
UNSCREWING
20:1
54.7 Ncm
15 RPM REV
ROUND BUR
20:1
28.1 Ncm
1’500 RPM
DRILL 1
20:1
35.3 Ncm 500 RPM
DRILL 2
20:1
35.3 Ncm 500 RPM
DRILL 3
20:1
35.3 Ncm 500 RPM
DRILL 4
20:1
35.3 Ncm 500 RPM
DRILL 5
20:1
35.3 Ncm 500 RPM
TAPPING
20:1
35.3 Ncm
15 RPM
TAP UNSCREWING
20:1
42.5 Ncm
15 RPM REV
IMPLANT SCREWING
20:1
35.3 Ncm
15 RPM
UNSCREWING
20:1
54.7 Ncm
15 RPM REV
ROUND BUR
20:1
35.3 Ncm
1’500 RPM
DRILL 1
20:1
35.3 Ncm
1’500 RPM
DRILL 2
20:1
35.3 Ncm
1’500 RPM
DRILL 3
20:1
35.3 Ncm
1’500 RPM
DRILL 4
20:1
35.3 Ncm
1’500 RPM
DRILL 5
20:1
35.3 Ncm
1’500 RPM
TAPPING
20:1
35.3 Ncm 20 RPM
TAP UNSCREWING
20:1
42.5 Ncm
15 RPM REV
IMPLANT SCREWING
20:1
35.3 Ncm 15 RPM
UNSCREWING
20:1
54.7 Ncm
15 RPM REV
PILOT DRILL 1
20:1
28.1 Ncm 800 RPM
PILOT DRILL 2
20:1
28.1 Ncm 800 RPM
DRILL 1
20:1
35.3 Ncm 600 RPM
DRILL 2
20:1
35.3 Ncm 500 RPM
DRILL 3
20:1
35.3 Ncm 400 RPM
SHAPING DRILL
20:1
35.3 Ncm 250 RPM
TAPPING
20:1
35.3 Ncm
20 RPM
TAP UNSCREWING
20:1
42.5 Ncm
20 RPM REV
IMPLANT SCREWING
20:1
35.3 Ncm
15 RPM
UNSCREWING
20:1
54.7 Ncm
15 RPM REV
130
*only CHIROPRO L
Default values
Endo* : default values
The table shows the default operating values for the endodontics sequence.
Surgery* : default values
The table shows the default operating values for 4 types of surgical operations proposed by the system, namely:
• Root resection
• Extraction of wisdom teeth
• Sinus raising
• Free program, available to the user.
ENDODONTICS
OPEN PULP CHAMBER
1:5
7.20 mNm
100’000 RPM
ENDO FILE 1
1:1
30.2 mNm 250 RPM
ENDO FILE 2
1:1
10.1 mNm 250 RPM
ENDO FILE 3
1:1
14.9 mNm 250 RPM
ENDO FILE 4
1:1
20.2 mNm 250 RPM
ENDO FILE 5
1:1
30.2 mNm 250 RPM
ENDO FILE 6
1:1
20.2 mNm 250 RPM
ENDO FILE 7
1:1
14.9 mNm 250 RPM
ENDO FILE 8
1:1
14.9 mNm 250 RPM
ENDO FILE 9
1:1
10.1 mNm 250 RPM
SURGERY
APICAL RESECTION
1:5
0.72 Ncm
100’000 RPM
TOOTH EXTRACTION
1:2
2.40 Ncm
80’000 RPM
SINUS LIFT
1:5
0.72 Ncm
50’000 RPM
FREE
1:5
0.72 Ncm
100’000 RPM
REF 2100219-0003/2014.05/ENG
Bien-Air Dental SA
Länggasse 60 Case postale 2500 Bienne 6, Switzerland Tel. +41 (0)32 344 64 64 Fax +41 (0)32 344 64 91 office@bienair.com
Bien-Air Deutschland GmbH
Jechtinger Strasse 11 79111 Freiburg, Deutschland Tel. +49 (0)761 45 57 40 Fax +49 (0)761 47 47 28 ba-d@bienair.com
Bien-Air España, SA
Entença, 169 Bajos 08029 Barcelona, España Tel. +34 934 25 30 40 Fax +34 934 23 98 60 ba-e@bienair.com
Bien-Air USA, Inc.
Medical Technologies 5 Corporate Park Suite 160 Irvine, CA 92606 USA Phone 1-800-433-BIEN Phone 949-477-6050 Fax 949-477-6051 ba-usa@bienair.com
Bien-Air France Sàrl
55-57, avenue Jean Lolive 93508 Pantin Cedex, France Tel. +33 (0)1 41 83 60 70 Fax +33 (0)1 48 96 07 40 ba-f@bienair.com
Bien-Air Italia s.r.l.
Via Vaina 3 20122 Milano, Italia Tel. +39 (02) 58 32 12 51/52/54 Fax +39 (02) 58 32 12 53 ba-i@bienair.com
Bien-Air UK Ltd
Arundel House Unit 1 - Ground Floor Amberley Court, Whitworth Road Crawley,West Sussex, RH11 7XL, England Telephone +44 (0)1293 550200 Fax: +44 (0)1293 520481 ba-uk@bienair.com
Bien-Air Asia Ltd.
Nishi-Ikebukuro Daiichi-Seimei Bldg. 10F 2-40-12 Ikebukuro, Toshimaku Tokyo, 171-0014, Japan
Tel. +81 (3) 5954-7661 Fax +81 (3) 5954-7660 ba-asia@bienair.com