Bien Air iOptima, 1700544-001, 1700704-001, iOptima INT, 1700547-001 Instructions For Use Manual

...
REF 2100279-0004/2018.06
iOptima
INT
iOptima

ENG INSTRUCTIONS FOR USE.

other languages available on www.bienair.com/ifu
Set iOptima REF 1700544-001
Set iOptima REF 1700547-001
Options
REF 2100279-0004/2018.06 © Bien-Air Dental
REF 2100279-0004/2018.06 © Bien-Air Dental
Set iOptima
INT
REF 1700704-001
Set iOptima
1x REF 1600677-001 1xREF 1600809-001 1xREF 1601074-001 1xREF 1500666-0011xREF 1502568-0011xREF 1300067-001
INT
REF 1700705-001
2x REF 1600677-001 2xREF 1600809-001 1xREF 1601075-001 1xREF 1500666-0011xREF 1502568-0011xREF 1300067-001
REF 2100279-0004/2018.06 © Bien-Air Dental
Set iOptima
INT
REF 1700730-001
Set iOptima
INT
REF 1700731-001
Options
REF 2100279-0004/2018.06 © Bien-Air Dental

Table of contents

iOptima description ...........................4
1 Symbols............................................................. 5
1.1 Description of symbols for iOptima unit .......................................5
1.2 Description of symbols for iOptima accessories........................ 6
2 Identification, Intended use and Notation ... 7
2.1 Identification..........................................................................................7
2.2 Intended use..........................................................................................7
2.3 Notation...................................................................................................8
3 Warnings & Precautions of Use .................... 9
4 Description...................................................... 10
4.1 iOptima system overview ............................................................... 10
4.2 iOptima system table....................................................................... 11
4.3 Set supplied........................................................................................ 11
4.4 Options ................................................................................................. 12
4.5 Technical data.................................................................................... 12
4.6 Environmental protection and information for disposal....... 13
4.7 Limitation of liability ........................................................................ 13
4.8 Electromagnetic compatibility (technical description) for
iOptima................................................................................................. 13
5 Installation ...................................................... 17
5.1 Install the iOptima app .................................................................... 17
5.2 Install the iOptima system ............................................................. 17
6 Interface overview......................................... 19
6.1 iOptima application........................................................................... 19
6.2 Sound alerts ....................................................................................... 19
6.3 iPod and iOptima connection / disconnection conditions..... 20
iOptima
7 Symbols........................................................... 23
7.1 Description of symbols for iOptima
7.2 Description of symbols for iOptima
INT
description ..................... 22
INT
....................................... 23
INT
accessories............... 24
8 Identification, Intended use and Notation. 25
8.1 Identification....................................................................................... 25
8.2 Intended use....................................................................................... 25
8.3 Notation................................................................................................ 26
9 Warnings & Precautions of Use .................. 27
10 Description ..................................................... 28
10.1 iOptima
10.2 iOptima
10.3 Set supplied........................................................................................ 29
10.4 Options ................................................................................................. 29
10.5 Technical data .................................................................................... 30
10.6 Environmental protection and information for disposal....... 31
10.7 Limitation of liability ........................................................................ 31
10.8 Electromagnetic compatibility (technical description) for
iOptima
INT
system overview ......................................................... 28
INT
system table ................................................................. 28
INT
........................................................................................... 31
11 Installation...................................................... 34
11.1 Install the iOptima app .................................................................... 34
11.2 Install the iOptima
INT
system ....................................................... 34
12 Interface overview......................................... 35
12.1 iOptima application........................................................................... 35
12.2 Sound alerts ....................................................................................... 35
12.3 iPod/iPad and iOptima
conditions............................................................................................ 36
INT
connection / disconnection
1
iOptima/iOptima
INT
operation .........38
13 Getting started ............................................... 39
13.1 Back up data....................................................................................... 39
13.2 Use of the Touch screen.................................................................. 39
13.3 Launch iOptima app ......................................................................... 40
13.4 Welcome screen and disclaimer.................................................. 41
13.5 Information ......................................................................................... 42
13.6 Settings................................................................................................ 43
13.7 iOptima app update .......................................................................... 43
14 Operation - Restorative mode ..................... 45
14.1 Enter Restorative mode .................................................................. 45
14.2 Operative screen description......................................................... 45
14.3 Standard Use - Operating mode (Start) ..................................... 47
14.4 Customize default operations ....................................................... 47
14.5 Create new operations (user-defined)........................................ 49
15 Operation - Endodontics Mode.................... 51
15.1 Enter Endodontics mode................................................................. 51
15.2 Operative screen description......................................................... 51
15.3 Standard Use - Operating mode (Start) ..................................... 53
15.4 Customize settings ........................................................................... 53
15.5 Add, sort or remove files ................................................................ 55
15.6 Restore (default) and delete (user-defined) systems............. 56
15.7 Create new Brands and Systems................................................. 57
15.8 CA ENDO / Reciproq option............................................................. 58
16 List of errors & Troubleshooting ................59
16.1 Safety warning (operating)............................................................. 59
16.2 Device operating error..................................................................... 60
16.3 iOS compatibility errors (disclaimer screen),........................... 61
17 Maintenance ................................................... 63
17.1 Servicing.............................................................................................. 63
17.2 Cleaning-disinfection....................................................................... 63
17.3 Important............................................................................................. 63
17.4 Replace 4VL seal............................................................................... 64
17.5 iOptima
INT
........................................................................................... 64
18 General information and guarantee...........65
18.1 General information ......................................................................... 65
18.2 Terms of guarantee .......................................................................... 65
2
Rx Only
iOptima description
ENG
SN
Rx Only

1Symbols

1.1 Description of symbols for iOptima unit

Symbol Description Symbol Description
REF
CE Marking with number of the notified body.
Manufacturer. Alternating current.
Reference number. Main switch - Power OFF.
Serial number. Main switch - Power ON.
CAUTION! In accordance with federal law (USA), this device is only available for sale upon recom­mendation by an accredited practitioner.
Separate collection of electric and electronic equipment.
Refer to the accompanying documents for the correct use of the product.(www.bienair.com/ifu.)
Recyclable materials. 4-hole connection (4way).
CAUTION! Consult accompanying documents. Provides an instruction that should be observed for safety reasons.
RF emitting device (Interference may occur in the vicinity of equipment marked with this symbol).
Sound alerts
Move fully to the stop, in the direction indicated.
4-hole connection and 4 electric (4VLM).
5
ENG

1.2 Description of symbols for iOptima accessories

SN
REF
Symbol Description Symbol Description
CE Marking with number of the notified body.
Manufacturer. Serial number.
Reference number.
Separate collection of electric and electronic equip­ment.
Electrical safety. Applied part type B.
6
FIG. 1

2 Identification, Intended use and Notation

2.1 Identification

iOptima
Electronically controlled unit for dentistry allowing operation of a single MX2 micromotor with variable speed using the dental unit pedal. It is essential to connect a compatible iPod Touch® using the lightning connector.
The unit provides power to the iPod device as soon as it is connected. Under normal use (power supply = 32Vdc), the unit does not use any battery energy of the iPod. The iOptima unit charges the iPod battery when connected. The restorative, endodontics parameters are defined before operation using the iOptima application installed on the iPod. The iOptima application indicates the operation steps, the ratio of the handpiece, tool speed, torque value, light intensity, footpedal mode and the micromotor rotation direction. iOptima application is compatible with Apple iOS operating system.
CAUTION
In endodontic mode, the predefined torque and speed values are only intended as a guide. The file values used must be adapted according to the manufacturer instructions of the files.

2.2 Intended use

Product intended for professional use only. The iOptima equipments are only intended for use in general dentistry, restorative dentistry and endodontics procedure by dentists and dental professionals in a dental office.
FIG. 1
iOptima system connected to an iPod that controls a MX2 motor.
Any use other than that for which this product is intended is unauthorized and can be harmful.
Note 1
7
ENG

2.3 Notation

A, B, C, etc. Text preceded by a letter indicates a procedure to be carried out step-by-step.
• Indicates a procedure result.
(1), (2), (3), etc. Text preceded by a number indicates text used in conjunction with an illustration.
OK, Save, Settings, etc. Text in bold italic font style indicates, on-screen elements such as buttons, menus, menu items, screen areas, values, fields when they are named and screen names. Tap Settings to open the Settings screen, change parameters and tap Done.
• Interface device «iPod Touch®» is referred to as «iPod».
NOTES
1 The technical specifications, illustrations and dimensions contained in these instructions are given only as a guide. They may not be the subject of any claim. The manufacturer reserves the right to make technical improvements to its equipment, with­out amending these instructions. For all additional information, please contact Bien-Air Dental SA at the address indicated on the back cover.
8

3 Warnings & Precautions of Use

CAUTION
The iOptima unit is not designed for use in an explosive atmosphere (anesthetic gas).
CAUTION
To prevent any risk of electric shock, the iOptima unit must be connected only to a power supply network provided with a protective earth.
CAUTION
The power plug of the iOptima is the device used for disconnection in case of problems, it must be easily accessible at all times.
WARNING
Never connect a handpiece on a running micromotor.
CAUTION
Ensure that the micromotor hose is not bent.
WARNING
The iPod must never be disconnected from the iOptima dental unit during operation!
NOTES
1 The predefined parameters may be subject to modification without notice.
WARNING
The parameters contained in the dental procedures are indicative only. Bien-Air Dental SA cannot be held liable for them.
Note 1
9
ENG

4 Description

5 6 7 8
A B C
A
B
C
1
2
3
4
1011
12
13
14
9

4.1 iOptima system overview

The iOptima system consists of:
• iOptima unit (3) with docking station for iPod (4);
• A MX2 micromotor (12) and MX2 hose (10, A);
• A pneumatic 4-way connection input (6, B);
• A power supply (2) and plug cord (C).
(1) Power supply cord (2) Power Supply (C) (3) iOptima unit (4) iPod [not provided in set] (5) MX2 micromotor hose output (A) (6) 4-way connector input, air and water inlet from your
(7) Power input connector (with lock) (C)
pneumatic unit (B)
FIG. 1
(8) Main power switch (9) Label (under the unit) (10) MX2 micromotor hose* (A) (11) iOptima status light (green power on) (12) MX2 micromotor* (13) Handpiece* [not provided in set] (14) Dental unit with pedal (B) [not provided in set]
*Applied parts (per IEC 60601-1)
10

4.2 iOptima system table

Summary of the iOptima functions and technological features. .
Product Device Technology Motor Restorative Endodontics
iOptima iPod DMX2 MX2 X X - -

4.3 Set supplied

4.3.1 iOptima system
iOptima set REF 1700544-001
Designation REF number
iOptima unit (1x) 1600926-001
MX2 Micromotor (1x) 1600677-001
MX2 Micromotor hose (1x) 1600762-001
Power supply (1x) 1501938-001
3P cable system, Switzerland, length
2.00 m (1x)
3P cable system, Europe, length 2.50 m (1x) 1300066-001
3P cable system, US/Asia, length 2.00 m (1x) 1300067-001
1300065-001

4.4 Options

Designation REF number
Spraynet, cleaning spray 500 ml, box of 6 cans
CA ENDO
option
1600036-006
Surgery
iOptima set REF 1700547-001
Designation REF number
iOptima unit (1x) 1600926-001
MX2 Micromotor hose (1x) 1600762-001
Power supply (1x) 1501938-001
3P cable system, Switzerland, length
2.00 m (1x)
3P cable system, Europe, length 2.50 m (1x) 1300066-001
3P cable system, US/Asia, length 2.00 m (1x) 1300067-001
11
1300065-001
ENG

4.5 Technical data

Dimensions L x W x H
iOptima unit................................................125 x 145 x 75 mm
iOptima unit (with iPod).......................... 125 x 145 x 160 mm
MX2 hose ....................................................L 1.7 m
MX2 micromotor.......................................Ø 21 x L 73.5 mm
Power Supply ............................................130 x 75 x 45 mm
Weight
iOptima unit................................................ 0.4 kg
Power Supply ............................................ 650 g
MX2 micromotor....................................... 94 g
Electrical and pressure data
Voltage.........................................................100-240 VAC
Frequency .................................................. 47-63 Hz
Nominal power ......................................... 90 W
Max. input power......................................160 W
Max. input pneumatic pressure .......... 5 bar / 72.5 psi
Min. input pneumatic pressure ...........3 bar / 43.5 psi
Electric insulation class
Class I per IEC 60601-1 (apparatus protected against electric shocks).
Degree of protection
IP 40 (protection against insertion of objects larger than 1 mm)
Memory
Restorative Storage of 20 user-defined Operations
Endodontics Storage of 10 user-defined Brands Storage of 10 user-defined Systems Storage of 8 Files per System including adjustment of speed and torque values for each file
List of errors & Troubleshooting
See chapter “16 List of errors & Troubleshooting” on page 59.
Important: Consult the Instructions for Use of the following devices:
Environmental conditions
Environmental conditions
Temperat ure
Relative humidity
Atmospheric pressure
Altitude
Operating
+10°C (50°F) to +25°C (77°F)
30% to 80% 30% to 80%
700 hPa to 1060 hPa
0 to 3'048 m (0 to 10'000 ft)
Transport and storage (max. 15 weeks)
-25°C (-13°F) to +70°C (158°F)
500 hPa to 1060 hPa
-
CAUTION
Do not use iOptima outside the range of operating temperature.
Classification
Class IIa in accordance with European Directive 93/42/EEC concerning medical devices.
Product Micromotor IFU Hose IFU
iOptima MX2 LED 2100199 MX2 2100223
12

4.6 Environmental protection and information for disposal

The disposal and/or recycling of materials must be performed in accordance with the legislation in force.
This unit and its accessories must be recycled. Electrical and electronic equipment may contain dangerous substances which constitute health and environmental hazards. The user must return the unit to its dealer or establish direct contact with an approved body for treatment and recovery of this type of equipment (European Directive 2002/96/EC).

4.7 Limitation of liability

Bien-Air Dental SA shall not be held liable for any non-compliant use of the iPod. The conditions for and restrictions on use set by Apple must be respected (jailbreak, hardware modification, etc.). To ensure optimum operation of iOptima the application, iOS updates must be according to version used during validation of the application.

4.8 Electromagnetic compatibility (technical description) for iOptima

Precautions regarding Electromagnetic Compatibility (EMC)
Electro-medical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this document.
CAUTION
Dental professionals need to be aware of potential electromagnetic interference between electronic dental devices and active implantable medical devices, and should always inquire about any devices implanted in the patient.
CAUTION
iOptima complies with the EMC requirements according to IEC 60601-1-2. Radio transmitting equipment, cellular phones, etc. shall not be used in close proximity to the unit since they could influence the performance of the unit. Special precautions must be taken when using strong emission sources such as High Frequency surgical equipment and similar equipment so that the HF cables are not routed on or near the unit. If in doubt, please contact a qualified technician or Bien-Air Dental SA. iOptima should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, iOptima should be monitored to verify normal operation in the configuration in which it will be used.
CAUTION
The use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by Bien-Air Dental SA as replacements parts for internal components, may result in increased emissions or decreased immunity of iOptima.
13
ENG
Guidance and manufacturer’s declaration - electromagnetic emissions
iOptima is intended for use in the electromagnetic environment specified below. The customer or the user of iOptima should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emis­sions IEC 61000-3-3
Group 1
Class B
Compliant
Not applicable
iOptima uses RF energy only for its internal function. CISPR 11 Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
iOptima is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient burst IEC 61000-4-4
Shock waves IEC 61000-4-5
Voltage dips and outages
IEC 61000-4-11
±6 kV contact
±8 kV air
±2 kV for power supply lines ±1 kV for lines no input/output
±0.5 kV line to line ±1 kV line to line ±0.5 kV line to earth ±1 kV line to earth ±2 kV line to earth
<5% U
T
(>95% dip in UT) for 0.5 cycle 40% U
T
(60% dip in UT) for 5 cycles
70% U
T
(30% dip in U for 25 cycles
<5% U (>95% dip in UT) for 5 sec
)
T
T
±6 kV contact
±8 kV air
±2 kV for power supply lines ±1 kV for lines no input/output
±0.5 kV line to line ±1 kV line to line ±0.5 kV line to earth ±1 kV line to earth ±2 kV line to earth
<5% U
T
(>95% dip in UT) for 0.5 cycle 40% U
T
(60% dip in UT) for 5 cycles
70% U
T
(30% dip in U for 25 cycles
<5% U (>95% dip in UT) for 5 sec
)
T
T
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of iOptima requires continued operation during power mains interruptions, it is recommended that iOptima be powered from an uninterruptible power supply or a battery.
14
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
Note : UT is the a.c. mains voltage prior to application of the test level. Essential performance: The essential performance is the maintaining of the visual lighting intensity of the LED and the maintaining of motor speed.
Maximum allowed speed deviation is ± 5%.
3 A/m 3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typi­cal commercial or hospital environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
3 Vrms
Portable and mobile RF communications equipment should be used no closer to any part of iOptima, including cables, than the recommended separation distance calculated from the equa­tion applicable to the frequency of the transmitter. Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Radiated RF IEC 61000-4-3
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an elec­tromagnetic site survey should be considered. If the measured field strength in the location in which the iOptima is used exceeds the applicable RF compliance level above, the iOptima should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the iOptima.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
3 V/m 80 MHz to 2,5 GHz
3 V/m
is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
pliance level in each frequency range Interference may occur in the vicinity of equipment marked
with the following symbol:
where P is the maximum output power rating of the transmitter in watts (W) accor­ding to the transmitter manufacturer and d
a
should be less than the com-
b
.
Note 1 - 2
15
ENG
Recommended separation distances between portable and mobile RF communications equipment and the iOptima
The iOptima is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the iOptima can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the iOptima as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter [m] Rated maximum output power of transmitter [W]
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 - 2
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
NOTES
1 At 80 MHz and 800 MHz, the higher frequency range applies.
2 These guidelines may not apply in all situations. Electro-
magnetic propagation is affected by absorption and reflection from structures, objects and people.
16
FIG. 1
2
1
1 2
3
1
2

5 Installation

FIG. 2 FIG. 3 FIG. 4

5.1 Install the iOptima app

A. Open the "App Store". B.
Search the iOptima application.
C.
Install the iOptima application designed by Bien-Air Dental.
Note 1

5.2 Install the iOptima system

CAUTION
Before installing, please read carefully this product instruction.
Note 2
FIG. 1
A. Place the iOptima on a flat surface capable of bearing its
weight.
CAUTION
It may be positioned on a table, on a trolley or any another surface but in no circumstances on the floor. It is not designed to be placed on wet surfaces or to come in contact with liquids.
FIG. 2
Connect the power cord (2) to the power supply (1) and plug to
B.
the mains.
Note 3
CAUTION
The power plug is the device used for disconnection in case of problems, it must be easily accessible at all times.
FIG. 3
Connect the power supply cable (1) to the input connector (2) and
C.
turn right to lock.
CAUTION
Ensure that the power switch (3) is off «O».
CAUTION
Before connecting the MX2 cable ensure that all o-rings are correctly fitted and connectors are clean of dust.
FIG. 4
Connect the MX2 cable plug (2) to the output connector (1):
D.
• First sleeve up and carefully and firmly insert the cable plug, by guiding the connector and plug with the index pin on the connector;
• Sleeve down and tighten (CW).
17
ENG
FIG. 5 FIG. 6 FIG. 7
1
2
1
2
1
3
2
4
FIG. 5
Connect the MX2 cable (2) to the MX2 micromotor (1), by guiding
E.
the connector and plug with the index pin on the connector and tighten (CW).
NOTES
CAUTION
Never connect a handpiece on a running micromotor.
FIG. 6
Connect the 4-way hose (1) to the 4-way connector of the
F.
iOptima unit (2):
• First guide the sleeve and carefully and firmly insert the hose sleeve, by matching the connector and the coupling;
•Tighten (CW).
FIG. 7
Switch on the water and power supply of the dental unit (refer to
G.
your dental unit instructions).
H.
Switch on the iOptima (1) («I» = ON).
The led (4) turns green (power on).
Once it has been switched on, connect the interface device
I.
«iPod» (2) to the iOptima carefully sliding it along the lightning adapter (3).
The iOptima is ready for use.
See chapter “6 Interface overview” on page 19.
1 The iPod must be correctly connected to a Wi-Fi network be­fore the App Store is opened; refer to Apple's user guide for ap­propriate use of the iPod.
2 In order to conform to the IEC 60601-1-2 standards, take into account the different routes of the wires through the system (bend, fold, section etc) (see chapter “4.1 iOptima system over­view” on page 10 and only use the power supply provided with the iOptima. In order to maintain warranty, this unit must be installed with the greatest care. Follow all the necessary instructions. Pro­tect the unit from direct sunlight and dust. Keep the original pac­kaging for storage and shipment.
3 The equipment is powered by your mains power supply (100­240 VAC).
18
FIG. 1 (iPod Touch®)

6 Interface overview

6.1 iOptima application

The iOptima application allows to carry out endodontics and restorative dentistry operations.
6.1.1 Compatibility
The iOptima application is compatible with iPod from 6th generation.
6.1.2 Notation
In order to simplify the notation, in this manual:
• iOptima application is referred to as «iOptima app»;
• Restorative and Endodontics modes are referred to as «RESTO» and «ENDO».

6.2 Sound alerts

Sound alert Description
Entering in endodontics or restorative
One long beep
Alternate short beeps
Alternate medium beeps
Alternate long beeps
mode
ENDO mode when selecting the first File
Warning notifications
Micromotor REVERSE running indicator
ENDO mode when the micromotor is running in reverse
System failure notification
19
ENG

6.3 iPod and iOptima connection / disconnection conditions

6.3.1 Disconnection
If the iPod is disconnected, the iOptima system is in standby mode (micromotor stopped).
WARNING
The iPod must never be disconnected from the iOptima dental unit during operation! If the iPod is disconnected during an operation, the micromotor stops immediately.
6.3.2 Connection
The micromotor can start only when the iPod is connected and the application is in operative mode.
CAUTION
If the footpedal is pressed before entering in operative mode, a warning message «Please release the pedal…» will be displayed. The micromotor will not start to run until the footpedal is released and pressed again.
6.3.3 iOptima unit switched ON
If the iOptima app is running and the iPod plugged to the iOptima (switch ON):
• The iPod will charge up (even if the app is not running);
• The sleep mode is deactivated;
• The micromotor can be operated with the pedal (iOptima app in operative mode).
20
iOptima
INT
description

7Symbols

SN
ENG
7.1 Description of symbols for iOptima
Symbol Description Symbol Description
CE Marking with number of the notified body.
Manufacturer. Alternating current.
REF
Reference number. Main switch - Power OFF.
Serial number. Main switch - Power ON.
CAUTION! Consult accompanying documents. Provides an instruction that should be observed for safety reasons.
Separate collection of electric and electronic equipment.
Refer to the accompanying documents for the correct use of the product.(www.bienair.com/ifu.)
INT
RF emitting device (Interference may occur in the vicinity of equipment marked with this symbol).
Sound alerts
Move fully to the stop, in the direction indicated.
4-hole connection and 4 electric (4VLM).
Recyclable materials. 4-hole connection (4way).
23
ENG
7.2 Description of symbols for iOptima
SN
REF
Symbol Description Symbol Description
INT
accessories
CE Marking with number of the notified body.
Manufacturer. Serial number.
Reference number.
Separate collection of electric and electronic equip­ment.
Electrical safety. Applied part type B.
24
FIG. 1

8 Identification, Intended use and Notation

8.1 Identification

iOptima
Electronically controlled integrated unit for dentistry allowing operation of multiple micromotors (MX2) with variable speed using the dental unit pedal. It is essential to connect a compatible iPod Touch® or iPad Mini® using the lighting connector with the provided docking station.
The unit provides power to the iPod/iPad device as soon as it is connected. Under normal use (power supply = 32Vdc), the unit does not use any battery energy of the iPod/iPad.
The iOptima/iOptima connected. The restorative and endodontics parameters are defined before operation using the iOptima application installed on the iPod/ iPad. The iOptima application indicates the operation steps, the ratio of the handpiece, tool speed, torque value, light intensity, footpedal mode and the micromotor rotation direction. iOptima application is compatible with Apple iOS operating system.
CAUTION
The predefined torque and speed values are only intended as a guide. The file values used must be adapted according to the manufacturer instructions of the files.
INT
INT
unit charges the iPod/iPad battery when

8.2 Intended use

Product intended for professional use only.
The iOptima dentistry, restorative dentistry and endodontics procedure by dentists and dental professionals in a dental office.
FIG. 1
iOptima micromotor in holder 1 (endodontics + restorative). Any use other than that for which this product is intended is unauthorized and can be harmful.
Note 1
INT
systems are only intended for use in general
INT
system connected to an iPod/iPad that controls a MX2
25
ENG

8.3 Notation

A, B, C, etc. Text preceded by a letter indicates a procedure to be carried out step-by-step.
• Indicates a procedure result.
(1), (2), (3), etc. Text preceded by a number indicates text used in conjunction with an illustration.
OK, Save, Settings, etc. Text in bold italic font style indicates, on-screen elements such as buttons, menus, menu items, screen areas, values, fields when they are named and screen names. Tap Settings to open the Settings screen, change parameters and tap Done.
• Interface device «iPod Touch®/iPad Mini®» is referred to as «iPod/iPad».
CAUTION
All illustrations and screenshots shown in this manual refer to the iPod and may differ from what is displayed on the iPad.
NOTES
1 The technical specifications, illustrations and dimensions contained in these instructions are given only as a guide. They may not be the subject of any claim. The manufacturer reserves the right to make technical improvements to its equipment, with­out amending these instructions. For all additional information, please contact Bien-Air Dental SA at the address indicated on the back cover.
26

9 Warnings & Precautions of Use

CAUTION
The iOptima atmosphere (anesthetic gas).
CAUTION
To prevent any risk of electric shock, the iOptima connected only to a power supply network provided with a protective earth.
WARNING
Never connect a handpiece on a running micromotor.
CAUTION
Ensure that the micromotor hose is not bent.
WARNING
The iPod/iPad must never be disconnected from the iOptima dental unit during operation!
WARNING
Never touch simultaneously the iPod/iPad and the patient.
INT
unit is not designed for use in an explosive
INT
unit must be
NOTES
1 The predefined parameters may be subject to modification without notice.
INT
WARNING
The parameters contained in the dental procedures are indicative only. Bien-Air Dental SA cannot be held liable for them.
Note 1
27
ENG

10 Description

2
1
10.1 iOptima
The iOptima
• A unit as a docking station (2) for iPod/iPad device.
• The iPod/iPad provide the user interface to the iOptima
• The electronic board (1), integrated in the dental unit system, is in charge to drive the micromotor (MX2) and communicate with the interface through the Lightning connector.
• A dental micromotor and the related hose (MX2).
• A power supply 100-240 VAC.
INT
system overview
INT
system mainly consists of:
INT
system, except speed reference input.
FIG. 1
10.2 iOptima
Summary of the iOptima
Product Device Te chn ology Motor Restorative Endodontics
INT
iOptima
INT
system table
INT
functions and technological features.
iPod/iPadDMX3MX2XXX-
CA ENDO
option
Surgery
28

10.3 Set supplied

iOptima
INT
set REF 1700730-001
10.3.1 iOptima
iOptima
iOptima
INT
set REF 1700704-001
Designation REF number
INT
iOptima
MX2 Micromotor (1x) 1600677-001
MX2 Micromotor hose (1x) 1600809-001
Power supply reset switchs 1502568-001
Power supply PMP90 (1x) 1500666-001
3P cable system, US/Asia, length 2.00 m (1x) 1300067-001
1 Motor Control Boards 1601074-001
Designation REF number
iOptima
MX2 Micromotor (2x) 1600677-001
MX2 Micromotor hose (2x) 1600809-001
Power supply reset switch 1502568-001
Power supply PMP90 (1x) 1500666-001
3P cable system, US/Asia, length 2.00 m (1x) 1300067-001
Dual Motor Control Boards 1601075-001
iDevice fixation (1x)
INT
set REF 1700705-001
INT
iDevice fixation (1x)
INT
system
1502475-001
1502475-001
Designation REF number
MX2 Micromotor hose (1x) 1600809-001
Power supply reset switch 1502568-001
Power supply PMP90 (1x) 1500666-001
3P cable system, US/Asia, length 2.00 m (1x) 1300067-001
1 Motor Control Boards 1601074-001
iOptima
INT
set REF 1700731-001
Designation REF number
MX2 Micromotor hose (2x) 1600809-001
Power supply reset switch 1502568-001
Power supply PMP90 (1x) 1500666-001
3P cable system, US/Asia, length 2.00 m (1x) 1300067-001
Dual Motor Control Boards 1601075-001

10.4 Options

Designation REF number
Mechanical interface iPod right mount
Mechanical interface iPod left mount
Mechanical interface iPad right mount
Mechanical interface iPad left mount
Mounting tools 1502567-001
1502620-001
1502621-001
1502622-001
1502623-001
29
ENG

10.5 Technical data

Dimensions L x W x H
MX2 hose ....................................................L 1.7 m
MX2 micromotor.......................................Ø 21 x L 73.5 mm
Weight
MX2 micromotor....................................... 94 g
Memory
Restorative Storage of 20 user-defined Operations
Endodontics Storage of 10 user-defined Brands Storage of 10 user-defined Systems Storage of 8 Files per System including adjustment of speed and torque values for each file
Electrical and pressure data
Voltage.........................................................100-240 VAC
Frequency .................................................. 47-63 Hz
Nominal power ......................................... 90 W
Max. input power......................................160 W
Max. input pneumatic pressure .......... 5 bar / 72.5 psi
Min. input pneumatic pressure ...........3 bar / 43.5 psi
Environmental conditions
Environmental conditions
Temperat ure
Relative humidity
Atmospheric pressure
Altitude
Operating
+10°C (50°F) to +35°C (77°F)
30% to 80% 30% to 80%
700 hPa to 1060 hPa
0 to 3'000 m (0 to 9842 ft)
Transport and storage (max. 15 weeks)
-25°C (-13°F) to +70°C (158°F)
500 hPa to 1060 hPa
-
CAUTION
Do not use iOptima
INT
outside the range of operating temperature.
List of errors & Troubleshooting
See chapter “16 List of errors & Troubleshooting” on page 59.
Important: Consult the Instructions for Use of the following devices:
Product Micromotor IFU Hose IFU
INT
iOptima
MX2 LED 2100199 MX2 2100223
Classification
Class IIa in accordance with European Directive 93/42/EEC concerning medical devices.
Electric insulation class
Class I per IEC 60601-1 (apparatus protected against electric shocks).
30

10.6 Environmental protection and information for disposal

The disposal and/or recycling of materials must be performed in accordance with the legislation in force.
This unit and its accessories must be recycled. Electrical and electronic equipment may contain dangerous substances which constitute health and environmental hazards. The user must return the unit to its dealer or establish direct contact with an approved body for treatment and recovery of this type of equipment (European Directive 2002/96/EC).

10.7 Limitation of liability

Bien-Air Dental SA shall not be held liable for any non-compliant use of the iPod/iPad. The conditions for and restrictions on use set by Apple must be respected (jailbreak, hardware modification, etc.).
To ensure optimum operation of iOptima
INT
the application, iOS updates must be according to version used during validation of the application.
10.8 Electromagnetic compatibility (technical description) for iOptima
INT
10.8.1 Electromagnetic compatibility warnings
The intended EM environment (per IEC 60601-1-2 ed. 4.0) is Professional healthcare facility environment.
CAUTION
The iOptima used in the immediate vicinity of the device, since this could affect its operation. The device is not suitable for being used close to high-frequency surgical equipment, magnetic resonance imaging (MRI) and other similar devices where the intensity of electromagnetic disturbances is high. In any case, ensure that no high frequency cables are routed above or near the device. If in doubt, contact a qualified technician or Bien-Air Dental SA. Special precautions should be taken when using strong emission sources such as high-frequency surgical equipment and other similar devices, to ensure that HF cables are not routed above or near the device. If in doubt, please contact a qualified technician or Bien-Air. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the iOptima performance of this equipment could result.
CAUTION
The use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by Bien-Air as spare parts for internal components, may result in increased emissions or decreased immunity.
CAUTION
Since this device is intended to be used adjacent to or stacked with other equipment, the responsibility of verifying normal operation in the configuration in which it will be used falls onto the dental unit manufacturer.
INT
complies with the EMC requirements according to IEC 60601-1-2. Radio transmitting equipment, cellular phones, etc., should not be
INT
, including cables specified by the manufacturer. Otherwise, degradation of the
31
ENG
10.8.2 Electromagnetic compatibility – emissions & immunity
Guidance and manufacturer’s declaration - electromagnetic emissions
INT
iOptima
The customer or the user of iOptima
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Emissions due to voltage fluctuations IEC 61000-3-3
Guidance and manufacturer’s declaration – Electromagnetic immunity
The iOptima INT is intended for use in the electromagnetic environment specified below. The customer or the user of the iOptima INT must ensure that it is actually used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient burst IEC 61000-4-4
Surge IEC 61000-4-5
is intended for use in the electromagnetic environment specified below.
INT
should ensure that it is used in such an environment.
The iOptima INT uses RF energy for its internal operation only. Therefore, its RF
Group 1
Class B
Class A
Conforming
±8 kV contact ±2 kV air ±4 kV air ±8 kV air ±15 kV air
±2 kV for power supply lines ±1 kV for other lines
±0.5 kV line to line ±1 kV line to line ±0.5 kV line to earth ±1 kV line to earth ±2 kV line to earth
emissions are very low and are not likely to cause any interference in nearby electro­nic equipment.
The iOptima INT is suitable for use in any building, including residential buildings and those directly connected to the public low-voltage power supply network that sup­plies buildings used for residential purposes.
±8 kV contact ±2 kV air ±4 kV air ±8 kV air ±15 kV air
±2 kV for power supply lines ±1 kV for lines no input/output
±0.5 kV line to line ±1 kV line to line ±0.5 kV line to earth ±1 kV line to earth ±2 kV line to earth
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic mate­rial, the relative humidity should be at least 30%.
Mains power quality should be that of a com­mercial or hospital environment.
Mains power quality should be that of a com­mercial or hospital environment.
32
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Voltage dips, short inter­ruptions and voltage variations on power sup­ply input lines IEC 61000-4-11
0% UT for 0.5 cycle, at 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
for 1 cycle and
0% U
T
70% UT for 25/30 cycles at 0°
0% UT for 0.5 cycle, at 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
for 1 cycle and
0% U
T
70% UT for 25/30 cycles at 0°
Mains power quality should be that of a com­mercial or hospital environment. If the user of the iOptima INT requires continued operation during mains power interruptions, it is recom­mended that the iOptima INT be powered from an uninterruptible power supply or a battery.
Magnetic field due to mains frequency (50/60 Hz) IEC 61000-4-8
Conducted disturbances induced by RF fields IEC 61000-4-6
Radiated RF EM fields IEC 61000-4-3
Proximity fields from RF wireless communications equipment IEC 61000-4-3
30 A/m 30 A/m
3 V
3 V
RMS
0,15 MHz – 80 MHz
in ISM bands
6 V
RMS
0,15 MHz – 80 MHz 80% AM at 1 kHz
3 V/m 80 MHz – 2,7 GHz 80 % AM at 1 kHz
Test freq . [ MHz] Max. power [W]
RMS
0,15 MHz – 80 MHz
in ISM and ama-
6 V
RMS
teur bands 0,15 MHz – 80 MHz 80% AM at 1 kHz
3 V/m 80 MHz – 2,7 GHz 80 % AM at 1 kHz
Immunity test level [V/m]
385 1.8 27
450 2 28
710, 745, 780 0.2 9
810, 870, 930 2 28
1720, 1845, 1970
228
2450 2 28
Magnetic fields generated by the mains fre­quency should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
a
should be less than the compliance
survey level in each frequency range. Interference may occur in the vicinity of equipment marked with the following sym­bol:
Distance: 0.3 m
5240, 5500, 5785
0.2 9
NOTE: UT is the AC mains voltage prior to application of the test level. Essential performance per IEC 60601-1: The essential performance is to maintain the visual luminous intensity of the LED and the
motor speed. The maximum speed deviation is ±5%.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and mobile field radios, amateur radios, AM and FM radio
broadcasts and TV broadcasts cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an elec­tromagnetic site survey should be considered. If the measured field strength in the location in which the iOptima INT is used exceeds the RF compliance level mentioned above, the iOptima INT should be observed to verify that it is operating normally. If abnormal operation is observed, additional measures may be nec­essary, such as reorienting or relocating the iOptima INT.
33
ENG

11 Installation

11.1 Install the iOptima app

A. Open the "App Store".
Search the iOptima application.
B. C.
Install the iOptima application designed by Bien-Air Dental.
Note 1
11.2 Install the iOptima
Refer to the iOptima
Note 2
INT
installation manual.
INT
system
NOTES
1 Th e iPo d/iPad mus t be correc tly c onnected t o a Wi-Fi ne twor k before the App Store is opened; refer to Apple's user guide for ap­propriate use of the iPod/iPad.
2 The iOptima neer certified by Bien-Air Dental SA.
INT
system must be installed by a service engi-
34
FIG. 1 (iPod Touch®)

12 Interface overview

12.1 iOptima application

The iOptima application allows to carry out endodontics and restorative dentistry operations.
12.1.1 Compatibility
The iOptima application is compatible with iPod from 6th generation and iPad from 4th generation.
12.1.2 Notation
In order to simplify the notation, in this manual:
• iOptima application is referred to as «iOptima app»;
• Restorative and Endodontics modes are referred to as «RESTO» and «ENDO».

12.2 Sound alerts

Sound alert Description
Entering in endodontics or restorative
One long beep
Alternate short beeps
Alternate medium beeps
Alternate long beeps
mode
ENDO mode when selecting the first File
Warning notifications
Micromotor REVERSE running indicator
ENDO mode when the micromotor is running in reverse
System failure notification
35
ENG
12.3 iPod/iPad and iOptima
INT
connection /
disconnection conditions
12.3.1 Disconnection
If the iPod/iPad is disconnected, the iOptima standby mode (micromotor stopped).
WARNING
The iPod/iPad must never be disconnected from the iOptima dental unit during operation! If the iPod/iPad is disconnected during an operation, the micromotor stops immediately.
INT
system is in
INT
12.3.2 Connection
The micromotor can start only when the iPod/iPad is connected and the application is in operative mode.
CAUTION
If the footpedal is pressed before entering in operative mode, a warning message «Please release the pedal…» will be displayed. The micromotor will not start to run until the footpedal is released and pressed again.
12.3.3 iOptima
If the iOptima app is running and the iPod/iPad plugged to the
INT
iOptima
(switch ON):
• The iPod/iPad will charge up (even if the app is not running);
• The sleep mode is deactivated;
• The micromotor can be operated with the pedal (iOptima app in operative mode).
12.3.4 iOptima
For iOptima the active motor. Memory of each operation set at last usage for each motor is kept (e.g. a user who has two motors can leave a blue CA on «motor 1» and a red CA on «motor 2» and, depending on which motor is activated, the corresponding modes are automatically activated).
INT
INT
unit switched ON
INT
last set operation memory
, the operation displayed by default is associated to
36
iOptima/iOptima operation
INT
FIG. 1 FIG. 2 FIG. 3 FIG. 4 FIG. 5
1
2

13 Getting started

13.1 Back up data

Use iTunes or iCloud to automatically back up your iOptima application user-defined settings. Refer to Apple's user guide for appropriate use of the iPod/iPad, iTunes or iCloud.
13.2.1 Display onscreen keyboard
FIG. 5
• Tap a text field to display the onscreen keyboard;
• Type characters with the onscreen keyboard (tap to delete all characters).

13.2 Use of the Touch screen

The iOptima application is operated with the iPod Touch®/iPad Mini® screen with a few simple gestures:
FIG. 1
Tap to click.
FIG. 2
Slide to scroll and adjust parameters.
FIG. 3
Swipe to restore or remove parameters.
Note 1
FIG. 4
Long tap to access the drag and drop function;
Drag and drop to sort files/operations.
39
ENG
FIG. 6
5
1
3
4
6
2

13.3 Launch iOptima app

FIG. 6
Make sure that the sound volume on your iPod/iPad is ON and set
A.
to an appropriate level according to the ambient noise of the room:
• Volume down (1) and up (2). Close all other applications in order not to disrupt the operation:
B.
• Double-click the home button (6) to show the multitasking display and drag the app up to close.
• Click the home button again to return to the main screen.
Deactivate all alerts. Refer to Apple's user guide for appropriate
C.
use of the iPod/iPad.
D.
Ensure that the latest application updates have been installed
(see chapter “13.7 iOptima app update” on page 43).
E.
Tap the app icon (5) to launch the iOptima application.
Note 2
NOTES
1 The swipe gesture is only active in the delimited area (1).
2 If not using your iPod/iPad, press the Sleep/Wake button (3)
to lock it.
40
FIG. 7 FIG. 8 FIG. 9
2
1
3
4
78
6
5

13.4 Welcome screen and disclaimer

A. To ensure the appropriate working of the iOptima Dental unit, make sure to only use your iOptima application with the iOS version validated by Bien-Air Dental.
FIG. 7
Refer to the current version of your device (1).
See also chapter “16.3 iOS compatibility errors (disclaimer screen),” on page 61.
For more information go to the App Store the selected application.
B.
Tap OK (2) to acknowledge disclaimer message and access the
DISCLAIMER
message second screen (3).
FIG. 8
Make sure to have set all steps of disclaimer (3) and tap OK (4),
C.
see chapter “13.3 Launch iOptima app” on page 40 for details.
The Welcome screen is displayed.
© under «what’s new» of
FIG. 9
The Welcome screen differs whether using the iOptima or
INT
iOptima
system and depends on which type of micromotor is
connected.
See chapter “4.2 iOptima system table” on page 11 for more information.
D.
Tap RO (5) to enter the restorative mode.
See chapter “14.1 Enter Restorative mode” on page 45.
E.
Tap EN (6) to enter the endodontics mode.
See chapter “15.1 Enter Endodontics mode” on page 51 for details.
Tap i (7) to open the
F.
Information
screen.
See chapter “13.5 Information” on page 42 for details.
Tap
Settings
G.
(8) to open the
Settings
screen.
See chapter “13.5 Information” on page 42 for details.
41
ENG
FIG. 10 FIG. 11 FIG. 12 FIG. 13
1
6
2 3 4 5
7
8
9

13.5 Information

FIG. 10
A. Tap (1) to access Bien-Air (catalogue, new products, etc.)
and iOptima (User Manual and About) information screen.
FIG. 11
The information screen allows to navigate to the following
pages:
Latest news (2) (opens Bien-Air website latest news page);
User manual (3) (application detects the language of the iPod/iPad and downloads the User Manual in the same language);
Note 3
Catalogue (4) (opens online catalogue (pdf file));
Web TV (5) (opens Bien-Air web TV);
About (6) (displays About screen (FIG. 12, 8) with iOptima system and Bien-Air factory information);
Contact us (opens email application with predefined email address iOptima@bienair.com).
Note 4
FIG. 13
Chapter links and the table of contents By clicking on a chapter section in the table of contents or in the manual, the user can directly access it. The dedicated button (9) links the user directly to the table.
NOTES
3 If the User Manual is not available in the correct language, the English manual is downloaded.
4 Latest News, Catalogue, Web TV and Contact us functions are available only if the iPod/iPad is connected to a Wi-Fi network.
42
FIG. 14 FIG. 15 FIG. 16 FIG. 17 FIG. 18
1
2
3 4

13.6 Settings

FIG. 14
INT
A. Tap (1) to access iOptima/iOptima screen.
system Settings
The Settings screen allows to set the torque unit for each
available mode.
FIG. 15
B.
Tap
Tor qu e u ni t
FIG. 16
Tap ENDO (3) to access the endodontics screen.
C.
FIG. 17
Tap your preferred torque unit.
D. E.
Select between;
to access its screen.
Ncm; mNm; gcm
and % (RESTO mode only).
13.6.1 Service engineer settings
With service engineer access (password protected), the settings screen also allows access to the micromotor holder configuration
(iOptima
INT
) and lock settings.

13.7 iOptima app update

FIG. 18
When the application is launched, it will check whether a new version is available on the App Store. If a new update is available, a message box will be displayed on Disclaimer screen to notify the user should download the application.
Note 6
Download: user can automatically access App Store application page.
Remind Me Later: user can delay the update and continue to work with currently installed app version (the message appears once a day).
Ignore: User can ignore the update and continue to work with currently installed app version (the message reappears only if a new version is available).
Important: if Lock settings is set, it is not possible to save any user-defined data!
Note 5
43
ENG
NOTES
5 To modify service engineer settings, please contact a service engineer certified by Bien-Air Dental SA.
6 This function is active only when the iPod/iPad is connected to a Wi-Fi network.
44
FIG. 1 FIG. 2 FIG. 3 FIG. 4 FIG. 5
1
2
3
4
7
8
9
10
56

14 Operation - Restorative mode

14.1 Enter Restorative mode

FIG. 1
Tap to enter the restorative mode. Predefined restorative operations are available by default, it is possible to modify and restore them. It is possible to create and delete User-defined operations.
Note 1 - 2

14.2 Operative screen description

FIG. 2
The operative screen displays the selected operation type and settings:
(1) Settings (access to operation types and settings) (2) Selected operation type (3) Micromotor rotation mode: FORWARD and REVERSE (4) Save (customize settings) (5) Footpedal mode: PROGRESSIVE or ON/OFF mode (6) Light intensity (7) Handpiece ratio (8) Maximum micromotor torque in % (or according to set-
(9) Maximum micromotor speed in rpm (10) Back to the previous screen
tings, see chapter “13.6 Settings” on page 43)
Note 3
See chapter “14.3 Standard Use - Operating mode (Start)” on page 47.
14.2.1 Micromotor Speed & Torque
FIG. 3
Slide to adjust the maximum speed and torque values.
Or tap the icons to activate the wheel scroller.
FIG. 4
Slide to precisely adjust the motor speed and torque values and tap Cancel or Done to validate.
FIG. 5
When the power demand of the micromotor is excessive, the
operative screen displays the overheating symbol . In this case the iOptima unit automatically lowers the torque in order to avoid overheating of the micromotor. To restore 100% torque, allow the motor to idle or stop for a few seconds.
14.2.2 Handpiece ratio
FIG. 6
Tap and select the handpiece ratio.
45
ENG
FIG. 6 FIG. 7 FIG. 8 FIG. 9
1
2
CAUTION
Verify that the handpiece corresponds to your selection.
Note 4 - 5
NOTES
14.2.3 Light intensity
FIG. 7
Slide or tap the appropriate dot to set up the light intensity of the micromotor. 11 levels of adjustment are possible:
• light OFF, 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, 90% and 100% of the max light intensity value.
Note 6 - 7
14.2.4 Footpedal mode
FIG. 8
Tap and select the footpedal mode:
ON/OFF (1): The speed consign is equal to the maximum speed once the footpedal is pressed; whatever the pressure on it.
Progressive (2): The speed consign value is (linear) function to the pressure on it.
Note 8
14.2.5 Micromotor rotation direction
FIG. 9
Tap and select the rotation mode of the micromotor:
Forward (clockwise)
Reverse (counterclockwise).
Note 9 - 10
1 In restorative mode, the air spray is always switched ON by
the help of the electro-valve integrated in the iOptima unit.
2 Save and create (
chapter “13.6 Settings” on page 43).
3 All parameters can be changed and saved directly in the op-
erative screen (except if a lock is set).
4 The handpiece ratio is red-colored for multiplication gears,
blue-colored for direct-drive, and green-colored for reduction gears.
5 The operative screen always displays the selected handpiece
ratio.
6 The light intensity value is set for each operation. The default
value of the light intensity is 100%.
7 The operative screen always displays the selected light inten-
sity value.
8 The operative screen always displays the selected footpedal
mode.
9 In reverse mode (CCW), the symbol flashes and there is a
sound alert (alternate medium beeps).
10 The operative screen always displays the selected rotation di-
rection.
) buttons
are disabled if a lock is set (see
46
FIG. 10 FIG. 11 FIG. 12 FIG. 13 FIG. 14
5
3
4

14.3 Standard Use - Operating mode (Start)

FIG. 10
A.
Tap to enter the RESTO mode, the operative screen opens.
Note 11
FIG. 11
Tap
Settings
B.
Note 12
FIG. 12
Tap to select an operation type.
C.
The app navigates back to the operative screen.
D.
Press the dental unit pedal to switch on the micromotor.
to open the
Operations
screen.
Tap the operation type above (3) or below (4) the selected
E.
operation type to navigate to the operation respectively preceding or following the current one in the predefined operation list.
See chapter “14.4.3 Sort operations” on page 48 to sort operations in the list.

14.4 Customize default operations

FIG. 14
From the operative screen, tap
A.
screen.
FIG. 15
Tap to customize an operation type.
B.
The Operation screen opens.
Settings
to open the
Operations
CAUTION
If the footpedal is pressed before entering in operative mode, a warning message «Please release the pedal…» will be displayed. The micromotor will not start to run until the footpedal is released and pressed again.
FIG. 13
Note 13 - 14
• Slide to adjust the micromotor speed and torque values of
• Adjust the handpiece ratio, light intensity, footpedal mode or
•Tap Save (5) to validate the new settings.
Note 15
47
the selected operation.
rotation direction of the selected operation.
FIG. 16
C.
Customize the operation settings.
D.
Tap
Cancel
(2) or
Save
(3) to validate custom settings and tap
back (1).
If saved, the symbol (factory predefined settings) changes
to (custom settings).
E.
Select an operation.
The app navigates back to operative screen.
ENG
FIG. 15 FIG. 16 FIG. 17 FIG. 18 FIG. 19
2 31
1
2
14.4.1 Restore (default) and delete (user­defined) operations
FIG. 17
Swipe left to restore or delete.
Note 16 - 17
14.4.2 Default Operations
FIG. 18
CAUTION
The parameters contained in the dental procedures are indicative only. Bien-Air Dental cannot be held liable for them.
Note 18
14.4.3 Sort operations
FIG. 19
NOTES
11 If using iOptima
holder, the operative mode previously used with that motor is dis­played by default (only from Home page).
12 Each operation in the list is tagged with a colored dot to indi-
cate which type of gear ratio it is associated to: red for multiplication gears, blue for direct-drive, green for reduc­tion gears.
13 When the micromotor is switched ON, the speed and torque
values displayed in the operative screen FIG. 13, toggle from pre­defined values to real time values. Once the motor stops running, the speed and torque values displayed toggle back to maximum predefined values.
14 All parameters can be directly modified in the operative
screen (only when the footpedal is released).
INT
system, each time you pull a motor from a
Do a long tap on an operation to activate the drag and drop
A.
function and display the icons (2).
B.
Tap the icon to drag and drop the operations.
C.
Tap
Done
(1) to validate.
15 Modifications are discarded when leaving the operative
screen without saving.
16 Predefined RESTO operations can not be deleted. But it is pos-
sible to restore the default values if modified. Only user-defined operations can be deleted.
17 Custom and user-defined operations are represented by .
Default operations are represented by .
18 The predefined parameters may be subject to modification
without notice.
48
FIG. 20 FIG. 21 FIG. 22 FIG. 23
5
1 2
4

14.5 Create new operations (user-defined)

FIG. 20
A.
From the operative screen, tap
screen.
FIG. 21
B.
Tap to create a new operation type.
The Operations screen opens.
Note 19
Settings
to open the
Operations
3
FIG. 22
Tap in t he
C.
Operation name
text field (1) to display the onscreen
keyboard.
FIG. 23
Type the ne w
D. E.
Set handpiece ratio and operation settings.
F.
Tap
Cancel
Operation name
or
Save
(2) to validate user-defined operation and tap
(3) and tap
Done
(4).
back, FIG. 22.
If saved, the new operation is displayed in the Operations
screen (5) with the symbol (user-defined) beside it, FIG. 21.
G.
Select an operation.
The app navigates back to operative screen.
49
ENG
NOTES
19 Maximum storage of 20 user-defined operations, if reached
the icon disappears.
50
FIG. 1 FIG. 2 FIG. 3 FIG. 4 FIG. 5
1
2
3
4
7
8
9
12
11
10
2
13
56

15 Operation - Endodontics Mode

15.1 Enter Endodontics mode

FIG. 1
Tap to enter the endodontics mode. Predefined endodontics systems are available by default, it is possible to modify and restore them. It is possible to create and delete User-defined endodontics brands, systems and files.
Note 1 - 2
(9) Maximum micromotor speed in rpm (10) Figure of the file (11) File system name (12) File brand (13) Back to the previous screen
Note 3
15.2.1 Micromotor Speed & Torque
FIG. 3
Slide to adjust the maximum speed and torque values.

15.2 Operative screen description

Or tap the icons to activate the wheel scroller.
FIG. 2
The operative screen displays the selected operation brand, system type and file settings:
(1) Access to brands, systems and file settings (2) File name and color designation of file according to ISO
3630-1 standard
(3) Micromotor rotation mode: FORWARD, AUTO-REVERSE,
AUTO-FORWARD
51
(4) Save (to customize settings) (5) Footpedal mode: PROGRESSIVE or ON/OFF mode (6) Light intensity (7) Handpiece ratio (8) Maximum micromotor torque in Ncm (or according to set-
tings, see chapter “13.6 Settings” on page 43)
FIG. 4
Slide to precisely adjust the motor speed and torque values and tap Cancel or Done to validate.
FIG. 5
When the power demand of the micromotor is excessive, the
operative screen displays the overheating symbol . In this case the iOptima unit lowers the torque automatically in order to avoid overheating of the micromotor. To restore 100% torque, allow the motor to idle or stop for a few seconds.
ENG
FIG. 6 FIG. 7 FIG. 8
1
2
15.2.2 Handpiece ratio
FIG. 6
Tap and select to modify the handpiece ratio.
CAUTION
Verify that the handpiece corresponds to your selection.
Note 4 - 5 - 6
15.2.3 Light intensity
FIG. 7
Slide or tap the appropriate dot to set up the light intensity of the micromotor. 11 levels of adjustment are possible:
• light OFF, 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, 90% and 100% of the max light intensity value.
Note 7 - 8
15.2.4 Footpedal mode
FIG. 8
Tap and select the footpedal mode:
ON/OFF (1): the speed consign is equal to the maximum speed once the footpedal is pressed; whatever the pressure on it.
Progressive (2): The speed consign value is (linear) function to the pressure on it.
Note 9
NOTES
1 In endodontics mode, the air spray is always switched OFF by
the help of the electro-valve integrated in the iOptima/iOptima unit.
)
2 Save and create (
chapter “13.6 Settings” on page 43).
3 All parameters can be changed and saved directly in the op-
erative screen (except if a lock is set).
The operative screen always displays the selected handpiece ratio.
4
5
If a different ratio than 8:1 Bien-Air or 1:1 Bien-Air is chosen, a dis­claimer is displayed on the handpiece ratio popup: «IMPORTANT WARNING: If instruments other than the 8:1 Bien-Air or the 1:1 Bien-Air Dental are used with the iOptima, please note that the displayed torque values cannot be guaranteed any longer (they repre­sent a purely indicative information). In such case, Bien-Air Dental de­clines all responsabilities in case of any injury or any damage resulting of the inappropriate use of the equipment.».
6
The handpiece ratio is red-colored for multiplication gears, blue­colored for direct-drive, and green-colored for reduction gears.
7
The light intensity value is set for each operation. The default value of the light intensity is 100%.
8
The operative screen always displays the selected light intensity value.
9
The operative screen always displays the selected footpedal mode.
buttons are disabled if a lock is set (see
INT
52
FIG. 9 FIG. 10 FIG. 11 FIG. 12 FIG. 13
3
1
2
4
1
15.2.5 Micromotor rotation direction
FIG. 9
Tap to select the rotation mode of the micromotor:
Forward (clockwise) (1)
Auto-reverse (2): the direction of rotation reverses automatically once the maximum torque set is reached (clockwise to counterclockwise).
Auto-forward (3): the direction of rotation reverses automatically once the maximum torque set is reached (clockwise to counterclockwise). In Auto-forward mode (reverse), the motor stops and reverses automatically after the time set (4): 0.5 to 3.0 seconds counterclockwise to clockwise).
Note 10 - 11

15.3 Standard Use - Operating mode (Start)

FIG. 10
A.
Tap to enter the ENDO mode, the operative screen opens.
Note 12
FIG. 11
Tap
Settings
B.
FIG. 12
C.
Tap to choose a brand, the
FIG. 13
Tap to select a file system.
D.
Note 13
to open the
Brands
Systems
screen,
screen opens.
The app navigates back to the operative screen, FIG. 14.
E.
Press the dental unit pedal to switch on the micromotor.
CAUTION
If the footpedal is pressed before entering in operative mode, a warning message «Please release the pedal…» will be displayed. The micromotor will not start to run until the footpedal is released and pressed again.
FIG. 14
Tap the file figure area to jump to the next file, parameters of the
F.
file system previously selected are automatically loaded (or directly tap the required file (3)).
Note 14 - 15

15.4 Customize settings

15.4.1 Customize default settings (operative screen)
FIG. 15
Note 15
A.
Slide to adjust the micromotor speed and torque values of the
selected file.
B.
Adjust the handpiece ratio, light intensity, footpedal mode or
rotation direction to customize settings for the selected file system.
C.
Tap
Save
to validate the new settings.
Note 16
53
ENG
FIG. 14 FIG. 15 FIG. 16 FIG. 17 FIG. 18
3
1
2
4
5
6
3
15.4.2 Customize system settings
FIG. 16
A.
Tap
Settings
to open the
Brands
screen.
FIG. 17
B.
Tap a
Brand
to open the
C.
Tap to customize a file system, the
Systems
screen.
Settings
screen opens.
FIG. 18
D.
Customize the file system settings:
• tap any file (4) to open the File screen and modify motor speed and torque values (tap (3) to go back);
• tap the icons (5) to modify the handpiece ratio, etc...;
•tap in the Add new file field (6) to create a new file, make a long tap to sort the files, or swipe left to remove existing files. See chapter “15.5 Add, sort or remove files” on page 55, for details.
Tap
Cancel
(1) or
Save
E.
(2) to validate custom settings.
If saved, the symbol (factory predefined settings) changes
to (custom settings).
Note 17
See chapter “15.7 Create new Brands and Systems” on page 57, for details
NOTES
10 The operative screen always displays the selected rotation
direction.
11 In reverse mode (CCW), the symbol flashes and there is a
sound alert (alternate medium beeps).
12 If using iOptima
holder, the operative mode previously used with that motor is dis­played by default (only from Home page).
13 If an ENDO file systems foresee
ment is selected, speed, torque, motor direction, light, pedal mode disappear and the gear ratio is fixed.
See chapter “15.8 CA ENDO
14 When the micromotor is switched ON, the speed and torque
values displayed toggle from predefined values to real time val­ues. Once the motor stops running, the speed and torque values displayed toggle back to maximum predefined values.
15 All parameters can be directly modified in the operative
screen (only when the footpedal is released). See chapter “15.4.1 Customize default settings (operative screen)” on page 53.
16 Modifications are discarded when leaving the operative
screen without saving.
17 From the Systems screen it is also possible to chose a differ-
ent brand and create a new system or a new brand with the
symbol.
INT
system, each time you pull a motor from a
ing the reciprocating move-
/ Recipro
q option” on page 58.
54
1
3
5
2
1
2
7
FIG. 19 FIG. 20 FIG. 21 FIG. 22 FIG. 23

15.5 Add, sort or remove files

Note 18
15.5.1 Add new files
FIG. 19
A.
Tap in t he
FIG. 20 - FIG. 21
Tap in t he
B.
type the new
Add new file
File name
File name
(5) and tap
field to open the
File
field (3) to display the onscreen keyboard,
Done
(onscreen keyboard).
4
screen.
6
FIG. 22
C.
Tap the file figure (4) to display the
Color designation
select a color for the new file and tap back (6) (color designat file is according to ISO 3630-1 standard).
D.
Adjust the micromotor speed and torque values (7), FIG. 20.
See “15.2 Operative screen description” on page 51, for details.
E.
Tap back (2) and
Save
(1), FIG. 19.
15.5.2 Sort files
FIG. 23
A.
Do a long tap on a file to activate the drag and drop function and
display the icons (2).
B.
Tap the icon to drag and drop the files.
C.
Tap
55
Done
(1) to validate.
screen,
ion of
ENG
FIG. 24 FIG. 25 FIG. 26 FIG. 27 FIG. 28
15.5.3 Remove files
FIG. 24
A.
Swipe left to remove existing files and tap
Save
or
Cancel
.
15.6 Restore (default) and delete (user­defined) systems
FIG. 25
Swipe left to restore or delete.
Note 19 - 20
15.6.1 Default Systems
FIG. 26
Dentsply Sirona
Note 21
FIG. 27
FKG swiss endo
FIG. 28
VDW
CAUTION
The parameters contained in the dental procedures are indicative only. Bien-Air Dental cannot be held liable for them.
Note 22
NOTES
18 New files can be
user-defined files.
19 Predefined ENDO systems can not be deleted. But it is possi-
ble to restore the default values if modified. Only user-defined systems can be deleted.
20 Custom and user-defined systems are represented with the
symbol .
Default manufacturer files are represented by the symbol .
21 Predefined endo file systems foreseeing the reciprocating
movement need activation on the iOptima movement wasn’t previosuly activated, taping it leads to the acti-
vation page.
See chapter “15
22 The predefined parameters may be subject to modification
without notice.
added, sorted or removed in default and
INT
system. In case the
.8
CA ENDO
/ Recipr
oq option” on page 58.
56
1 2 43
5
FIG. 29 FIG. 30 FIG. 31 FIG. 32 FIG. 33
6

15.7 Create new Brands and Systems

FIG. 29
A.
Tap to add a new brand or a new system, the
opens.
Note 23 - 24
FIG. 30
Tap in t he
B.
FIG. 31
C.
Type the new
drop-down list box and tap
FIG. 30
Tap in
D.
(onscreen keyboard).
E.
Tap
page 55.
F.
Tap the icons (6) to set the iOptima unit parameters:
• Handpiece ratio
• Light intensity
• Footpedal mode: PROGRESSIVE or ON/OFF mode
• Micromotor rotation mode: FORWARD, AUTO-REVERSE,
See chapter “15.2 Operative screen description” on page 51, for details.
Tap
G.
Brand
text field (1) to display the onscreen keyboard.
Brand name
System
field (2), type the
Add new file
AUTO-FORWARD
Save
(3).
(4), refer to chapter “15.5.1 Add new files” on
(5) or select an existing brand from the
Done
(onscreen keyboard).
System name
Settings
and tap
screen
Done
15.7.1 Restore (custom) and delete (user­defined) brand
Note 25 - 26
Restore default brands (factory settings)
FIG. 32
A.
Swipe left on a default brand.
A message box opens: Restore factory brand.
B.
Tap Cancel or Restore to acknowledge.
Delete user-defined brands
FIG. 33
A.
Swipe left on a user-defined brand.
A message box opens: Delete user brand.
B.
Tap
Cancel
or
Delete
to acknowledge.
57
ENG
FIG. 34 FIG. 35 FIG. 36
1
2
3 4

15.8 CA ENDO / Reciproq option

Predefined endo file systems foreseeing the reciprocating
INT
movement need activation on the iOptima
system. In case the movement wasn’t previously activated, tapping it leads to the activation page.
15.8.1 Activation page
FIG. 34
A.
Tap
More information
FIG. 35
Input the code obtained and tap
B.
A message box displays either:
License not Activated. License Activated.
FIG. 36
C.
If activated the app navigates back to the operative screen.
Note 27 - 29
D.
Press the dental unit pedal to switch on the micromotor.
and follow the instructions displayed.
Activate
(2)
NOTES
23 New systems and brands can also be created directly from
the system screen.
24 Maximum storage of 10 user-defined brands and 10 systems
per brand, if reached the icon disappears.
25 Restore ALL default system settings from the Brands screen.
26 It is only possible to delete user-defined brands
27 If an ENDO file systems foreseeing the reciprocating move-
ment is selected, speed, torque, motor direction, light, pedal mode disappear and the gear ratio is fixed.
28
cessive, the operative screen displays the overheating symbol (3). In this case the iOptima/iOptima matically in order to avoid overheating of the MX2 micromotor. To restore 100% torque, allow the motor to idle or stop for a few sec­onds.
29
displays
When the power demand of the MX2 micromotor is ex-
INT
unit lowers the torque auto-
If stress is excessive on the file, the file stress warning
(4).
CAUTION
• Valid only with Bien-Air CA ENDO.
• CA ENDO must be used only with specific tools.
58

16 List of errors & Troubleshooting

16.1 Safety warning (operating)

Warning description Message Cause of warning Action
Pedal is pressed when accessing operative page. Pedal remains pressed when
Footpedal to be released Please release pedal …
Motor torque limitation active
Please switch ON sound vol­ume on your iPod/iPad and
iPod/iPad speaker OFF
set it to an appropriate level according to the ambient noise level in your practice room.
acknowledging any system notification. Motor is jammed for more than 2 seconds.
Motor drive limits delivered torque to prevent motor over­heating.
iPod
/iPad
speaker was switched OFF, or sound volume was dis­abled by user. It is necessary to have iPod
/iPad
speaker turned ON so that system sound notifi­cations and alarms may be heard distinctly.
Release footpedal and press it again.
Avoid extended use. Let system cool down.
Turn iPod/iPad speaker ON and set sound volume to an appropriate level.
Non-verified iOS version in use
User Manual has been updated
Motor torque threshold limit reached, file stress condition encountered
59
User Manual has been updated, and may be con­sulted in the Information page.
iOS version installed on the iPod
/iPad
has not been verified according to Bien-Air validation protocols. Therefore it is NOT recommended to use the sys­tem with this configuration.
As an update of the User Man­ual was available on Bien-Air website, it has been automati­cally downloaded and updated on the iPod
System detects that motor torque threshold limit is reached.
/iPad
.
Do NOT update iPod/iPad with new iOS versions unless Bien-Air recommends to do so. If the iPod/iPad has been updated with a new iOS version despite everything, do not use the system until Bien-Air has verified and approved the new iOS version.
It is highly recommended to consult the updated User Manual before using the system.
Release file pressure. Slow down treatment.
ENG

16.2 Device operating error

Error description Message Cause of error Action
ERROR 4
Motor connection missing
ERROR 5
Motor cable failure
ERROR 6
Motor drive over temperature
GEN ERROR [FailCode]
System electrical failure.
Motor is not connected ! Please check motor connection.
Motor cable fault ! Please replace motor cable.
Overall system overheating ! Please wait until cool.
Electrical system fault ! Please contact Bien-Air Dental SA.
Motor phase missing failure. Motor is not properly connected.
Motor drive power protection failure. Motor cable may be defect.
Motor drive over temperature failure.
[FailCode] = EC100 : Motor drive communication failure [FailCode] = EC101 : Motor drive under voltage failure [FailCode] = EC102 : Motor drive over voltage failure [FailCode] = EC120 : Motor drive other failure
1. Check motor connection.
2. If problem persists, contact Bien-Air Dental SA.
1. Replace motor cable.
2. If problem persists, contact Bien-Air Dental SA.
1. Wait for system cooling.
2. If problem persists, contact Bien-Air Dental SA.
1. Switch OFF unit
2. Disconnect iPod/iPad device from unit
3. Close iOptima application
4. Switch unit back ON
5. Reconnect iPod/iPad device to unit
6. Restart iOptima application
7. If problem persists, contact Bien-Air Dental SA.
60

16.3 iOS compatibility errors (disclaimer screen),

Color iOS version (identifier digit change) Text version Restriction (recommendations)
Green
Orange
Red
Note 1
•None.
• Currently installed iOS is identical to iOS verified during app validation.
• Last digit is different: iOS x.y.Z.
• Currently installed iOS was updated for a minor change compared to iOS verified during app validation.
• First and/or second digit are/is differ­ent: iOS X.y.z, or iOS X.Y.z.
• Currently installed iOS was updated for major or at least significant changes compared to iOS verified during app validation
Currently installed iOS x.y.z text sec­tion is displayed in green color.
Currently installed iOS x.y.z text sec­tion is displayed in orange color. User must be aware that the configu­ration he is using is slightly different from the recommended configura­tion.
Currently installed iOS x.y.z text sec­tion is displayed in red color.
No restrictions
• Configuration in use is slightly different from recommended configuration.
•No restrictions
• Configuration in use is significantly differ­ent from recommended configuration.
CAUTION
The current configuration should NOT be used!
61
ENG
NOTES
1 It is imperative to read and understand this IFU before carry-
ing out any troubleshooting on the iOptima. In case of unclear information or for any error message that is not listed in the following tables contact Bien-Air Dental SA.
62
FIG. 1
1
2

17 Maintenance

CAUTION
Only use original Bien-Air Dental maintenance products and parts or those recommended by Bien-Air Dental. Using other products or parts may cause operational failure and/or void the guarantee.

17.1 Servicing

Never disassemble the device. For any modification and repair, we recommend to contact your regular supplier or
directly at the address indicated on the back cover.
SA
Note 1
Bien-Air Dental

17.3 Important

For maintenance of micromotors and hoses: See IFU
Product Micromotor IFU Hose IFU
iOptima MX2 LED 2100199 MX2 2100223
INT
iOptima
MX2 LED 2100199 MX2 2100223

17.2 Cleaning-disinfection

• Disinfect the surfaces of the iOptima unit / iOptima in and footpedal by gently rubbing for about 15 seconds with a clean cloth soaked in a suitable product (i.e. Bien-Air Dental Spraynet or isopropyl alcohol) or with an appropriate disinfectant wipe.
• Do not immerse in disinfectant solution.
• Do not immerse in an ultrasonic bath.
INT
dock-
63
ENG

17.4 Replace 4VL seal

FIG. 1
CAUTION
Immediately replace any damaged or leaking O-rings and seals. Never use sharp tools!
A. Switch off the water and the dental unit power supply.
Switch OFF the iOptima unit «O».
B. C.
Unscrew and unplug the 4VL hose (1).
D.
Remove the damaged 4VL seal (2).
E.
Replace with a new 4VL seal (REF 1302403-010).
Refit hose, switch ON units and water.
See chapter “5.1 Install the iOptima app” on page 17 for details.
17.5 iOptima
Refer to the iOptima
INT
INT
installation manual
CAUTION
Only a qualified Bien-Air Dental service engineer is permitted to
open the dental unit and repair the iOptima
INT
system.
NOTES
1 Bien-Air Dental SA recommends the user to have its dynamic instruments regularly checked or inspected (at least once a year).
64

18 General information and guarantee

18.1 General information

The device must be used by qualified professionals in compliance with the current legal provisions concerning occupational safety, health and accident prevention measures, and these instructions for use. In accordance with such requirements, the operators:
• must only use devices that are in perfect working order; in the event of irregular functioning, excessive vibration, abnormal heating or other signs that may indicate malfunction of the device, the work must be stopped immediately; in this case, contact a repair center that is approved by Bien-Air Dental SA;
• must ensure that the device is used only for the purpose for which it is intended, must protect themselves, their patients and third parties from any danger, and must avoid contamination through the use of the product.

18.2 Terms of guarantee

Bien-Air Dental SA grants the user a guarantee covering all functional defects, material or production faults.
The device is covered by this guarantee for:
• 12 months for the hose
• 12 months for the power supply
• 12months for integrable electronic boards
• 24 months for the iOptima/iOptima
• 36 months for series MX2 LED electric micromotors
from the date of invoicing.
INT
unit
Bien-Air Dental SA shall not be held responsible for damage or injury and the consequences thereof, resulting from:
• excessive wear and tear
•improper use
• non-observance of the instructions for installation, operation and maintenance
• unusual chemical, electrical or electrolytic influences
• poor connections, whether of the air, water or electricity supply.
The guarantee does not cover flexible “fiber optic” type light conductors, or any parts made of synthetic materials. The guarantee shall become null and void if the damage and its consequences are due to improper manipulation of the product, or modifications to the product carried out by persons not authorized by Bien-Air Dental SA.
Claims under the terms of the guarantee will be considered only on presentation, together with the product, of the invoice or the consignment note, on which the date of purchase, the product reference and the serial no. should be clearly indicated.
In case of justified claim, Bien-Air Dental SA or its authorized representative will fulfill the company’s obligations under this guarantee by repairing or replacing the product free of charge. Any other claims, of whatever nature, in particular in the form of a claim for damages and interest, are excluded.
65
Bien-Air Dental SA
Länggasse 60 Case postale 2500 Bienne 6, Switzerland Tel. +41 (0)32 344 64 64 Fax +41 (0)32 344 64 91 oce@bienair.com
Bien-Air Deutschland GmbH
Jechtinger Strasse 11 79111 Freiburg, Germany Tel. +49 (0)761 45 57 40 Fax +49 (0)761 47 47 28 ba-d@bienair.com
Bien-Air España, SAU
Entença, 169 Bajos 08029 Barcelona, Spain Tel. +34 934 25 30 40 Fax +34 934 23 98 60 ba-e@bienair.com
Bien-Air USA, Inc.
5 Corporate Park Suite 160 Irvine, CA 92606 USA Phone +1 800-433-2436 Phone +1 949-477-6050 Fax +1 949-477-6051 ba-usa@bienair.com
Bien-Air France Sàrl
19-21, rue du 8 Mai 1945 CS 30310 94113 Arcueil, France Tel. +33 (0)1 49 08 02 60 Fax +33 (0)1 46 64 86 58 ba-f@bienair.com
Bien-Air Italia S.r.l.
Via Vaina 3 20122 Milano, Italy Tel. +39 (02) 58 32 12 51 Fax +39 (02) 58 32 12 53 ba-i@bienair.com
Bien-Air UK Ltd
Arundel House Unit 1 - Ground Floor Amberley Court, Whitworth Road Crawley, RH11 7XL, England Tel. +44 (0)1293 550 200 Fax +44 (0)1293 520 481 ba-uk@bienair.com
Bien-Air Asia Ltd.
Nishi-Ikebukuro Daiichi-Seimei Bldg. 10F 2-40-12 Ikebukuro, Toshimaku Tokyo, 171-0014, Japan
Tel. +81 (3) 5954-7661 Fax +81 (3) 5954-7660 ba-asia@bienair.com
Beijing Bien-Air
Medical Instrument Technology Service Co. Ltd. Room 1415, Block B Lucky Tower, No 3 Dongsanhuan Beilu, Chaoyang District, Beijing 100027, China
Tel. +86 10 6567 0651 Fax +86 10 6567 8047 ba-beijing@bienair.com
▦℻ゑ㦬棂◉₫ₘ▦℻ゑ㦬棂◉₫ₘ 䘾▦恾⚆ㄇ䰞⮶䘾▦恾⚆ㄇ䰞⮶ %ㄶ⸳
%ㄶ⸳
REF 2100279-0004/2018.06 iOptima/iOptima
INT
© Bien-Air Dental
Loading...