These instructions are an essential part of the device. They must therefore be
kept in a suitable place near the device and should accompany the device if it
is transferred to other users.
For proper and safe use of this device it is essential that the following
warnings and safety instructions, as well as the instructions for use are read
and carefully observed by all users before first using the device.
It is the responsibility of those using the device to fully acquaint themselves
with its proper use and operation.
If a malfunction is suspected, the device is to be taken out of service
immediately and suitable warning signs should be attached to the device to
ensure that it cannot be used before necessary service and repair work has
been carried out.
DE - Edition 02 / 20073
Instructions for Use – BW 685 / BW 685 S BIEGLER
Table of contents
1 WARNINGS AND SAFETY INSTRUCTIONS .............................................. 5
•In the event of any suspected malfunction while in operation, the device
should be immediately removed from service and not used for infusions
or transfusions until appropriate investigations have demonstrated that
there has been no impairment.
•If the high temperature alarm is triggered the supply of liquid to the
patient must be immediately stopped by disconnecting the connection
tube to the patient. The medium being used in the device must no longer
be administered to the patient.
•The device may only be fastened to infusion stands, tripods or equipment
rails which are suitable due to their stability and load capacity.
•Only sterile consumables prescribed by BIEGLER may be used in
conjunction with the BW 685 / BW 685 S.
•The device must only be used in areas in which the electrical installations
are in accordance with the standards and regulations in force.
•Safe disconnection from the mains power supply can only be achieved
by unplugging the mains plug.
•The device may not be used in rooms subject to explosion hazard.
•Repairs and modifications to the device may only be carried out by
persons or service centres authorised by BIEGLER.
•The device must not be immersed in liquids or sterilised with steam or by
thermochemical methods.
•All extraneous influences, such as electromagnetic waves or high
temperatures are to be kept to a minimum.
•The system may malfunction due to the effect of strong electromagnetic
fields (e.g. by HF therapy or surgical devices). The warmer works
perfectly within the limit values of the Standard EN 60601-1-2. The
system can be affected by limit values that do not comply with the
Standard EN 60601-1-2.
•Avoid exerting force on the device or its accessories.
•If the device is dropped, damaged by force or if it shows a function
deviating from the instructions for use, stop using the device immediately
and return it to the service centre.
•The periodic technical safety inspections must be carried out (see
section: “Periodic inspections”)
DE - Edition 02 / 20075
Instructions for Use – BW 685 / BW 685 S BIEGLER
Supplementary safety instructions for the accessory tube warmer Biegler
Tubeflow BW 685 S:
•The Biegler Tubeflow may only be clamped to the Biegler blood
warmer BW 685 S with the fixing clamp according to the diagram in the
chapter "Initial operation".
•Do not kink, cover, warm or cool the Biegler Tubeflow. Do not cover
the Tubeflow with cloths or dressing material. Do not lead under blankets,
warm air blankets, etc. Do not expose to direct sunlight or heat radiation.
Lead the Tubeflow through the air unencumbered. Do not lead into or
through incubators. Overwarming of the warmed liquid or burns to the
patient cannot to be excluded in the case of non-compliance.
•Do not shorten or damage the Biegler Tubeflow.
The BW 685 / BW 685 S may not be used if:
• the housing is damaged or one of the front film layers becomes detached.
• the device has been exposed to a hard physical shock (e.g. dropped, hit or
shaken)
• the device has been immersed in water
• the device has triggered a high temperature alarm that was not caused by
external factors.
• the mains power cord or plug is damaged
• the device has given somebody an electric shock
• the fixing clamps are damaged and no longer assure safe clamping to the
infusion stand.
Should a malfunction be evident, suitable warning signs should be attached
to the device to ensure that it cannot be used before necessary service and
repair work has been carried out.
6DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
2DESCRIPTION
2.1 GENERAL DESCRIPTION
The BIEGLER BW 685 / BW 685 S is a warmer for infusions or transfusions
and operates on the basis of a continuous flow warmer, where the heat from
the heat exchanger is transferred via the extension tubing to the liquid flowing
within it.
The patented groove design enables several extension tubes to be used,
provided that each of the tubes placed in position can be warmed sufficiently.
The design of the housing allows rapid and simple fitting to all suitable
infusion stands and equipment rails. The temperature of the warmer element
can be set between 37°C and 41°C in steps of 0.5°C and is clearly displayed.
The preset temperature after switching on the BW 685 / BW 685 S is 38.5°C.
The alarm and self-test functions for high and low temperature that are
incorporated into the device assure safe operation.
Optional: The accessory tube warmer Biegler Tubeflow is available as a
warming system for transfusions and infusions between BW 685 S and the
patient. The system can be used in all cases in which warming of the
transfusion or infusion is necessary up to the patient.
2.2 SCOPE OF DELIVERY
Blood and infusion warmer BW 685 / BW 685 S and instructions for use
Optional accessory for BW 685 S:
Tube warmer Biegler Tubeflow Order no 12000
DE - Edition 02 / 20077
Instructions for Use – BW 685 / BW 685 S BIEGLER
2.3 CONSUMABLE MATERIALS
Various consumable materials are available according to requirements.
For use with BW 685 or BW 685 S without tube warmer:
Extension set3.5 mOrder no 35000
Extension set4.6 mOrder no 45000
Extension set with bubble trap 4.6 mOrder no 25000
For use with BW 685 S with tube warmer:
Extension set5.3 mOrder no 16000
Extension set with bubble trap 5.3 mOrder no 15000
8DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
3INITIAL OPERATION
Observe the instructions for use! Users must familiarise themselves
in detail with the contents of these instructions for use before putting
the system into operation.
3.1 INITIAL OPERATION OF THE DEVICE
Diag. 1
1Heat exchanger
2Temperature scale
3Control ON / STANDBY
4LED STANDBY
5Control for decreasing the
temperature
6Control for increasing the
temperature
7Extension tube
8Fixing clamp
9Entry of liquid
10 Exit of liquid
Flow direction
Fix the BIEGLER BW 685 / BW 685 S firmly to the infusion stand using the
clamps at the back (Diag.1/8). Only use infusion stands, tripods or equipment
rails that are sufficiently stable.
Connection of the optional tube warmer Biegler Tubeflow (to the BW 685 S):
The tube warmer Biegler Tubeflow must be inserted with light pressure as
indicated in diagram 3 into the intended fixture (Diag. 3/4) on the BW 685 S.
Remove the dust cap from the BW 685 S and attach the connector of the
Tubeflow to the female connector on the BW 685 S.
Connect the power cable to the power supply. Before connecting to the
mains power supply, check the voltage specified on the device label. The
device emits a short beep and the standby light (Diag.1/4) lights up.
DE - Edition 02 / 20079
Instructions for Use – BW 685 / BW 685 S BIEGLER
If a temperature other than 38.5°C is desired, it can be preset in the Standby
mode by using the controls and (Diag.1/5 and Diag.1/6). If an
adjustment control is pressed, the visual display indicates the existing preset
temperature. By repeated operation of the control or , the temperature
can be reset. Resetting of the temperature can only be performed in the
Standby mode.
The warmer of the BW 685 / BW 685 S can be started by pressing the
control (Diag.1/3). A self-test of safety relevant device functions is carried
out, which is terminated by a short acoustic signal. The BW 685 /
BW 685 S attains the set target temperature within 1 minute. The actual
temperature (+/- 0.5°C) is indicated on the display.
Select suitable consumable materials. See the section "Consumable
materials".
Prepare infusion or transfusion. Important: The length of the tube between
the BW 685 / BW 685 S and the patient must be at least 40 cm and the tube
must not be stretched.
Diag. 2
Position of the tube in the groove of the heat
exchanger
Beginning at the back of the heat exchanger, gently pull the extension tube
and coil it in a counter-clockwise direction towards the front. It is advisable
that the distance between the BW 685 / BW 685 S and the patient does not
exceed much more than 80 cm.
Important: The tube must be completely inserted into the groove (Diag. 2)
The flow direction specified in diagram 1 must be complied with.
10DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
Optional accessory tube warmer Biegler Tubeflow for BW 685 S:
The connector as per diagram 3/2 must be connected to the BW 685 S
before starting the optional tube warmer.
Insert the transfusion or infusion tube into the
groove of the tube warmer. Insert the tube over
the complete length of the Biegler Tubeflow.
Important: Overwarming can result, if the full
length of the tube warmer is not utilised, thereby
possibly harming the medium or patient, since
the reference temperature sensor is located at
the distal end.
Diag. 3
1Tube warmer
2Connector
3Control light
(Green – Operational / red – Alarm)
Fixture
Diag.3
DE - Edition 02 / 200711
Instructions for Use – BW 685 / BW 685 S BIEGLER
The Tubeflow is also activated when the BW 685 S is turned on; this is
indicated by a green control light on the Tubeflow.
3.2 ALARMS
The BW 685 / BW 685 S can trigger two types of temperature alarm:
The low temperature alarm is activated when the temperature of the heat
exchanger drops below 36.5°C. The audio low temperature alarm is
deactivated during the first 60 seconds after switching on.
The high temperature alarm is activated when the temperature of the heat
exchanger exceeds 42.0°C. In this event visual and audio alarm signals are
given and the heating is switched off. To reset the device or switch off the
alarm, disconnect the device from the power supply.
Important: The temperature alarm can also be set off externally e.g. by
exposure to sunlight.
3.3 SHUTTING DOWN THE DEVICE
After being used for treatment, the device is shut down as follows:
Switch the device into Standby mode using control (Diag.1/3).
Release pressure from the system by switching off possibly used pressure
cuffs or infusion pumps. Empty and disconnect the system as far as possible.
Remove the consumable materials from the heat exchanger (Diag.1/1) and, if
used, also from the optional tube warmer and dispose of in accordance with
the applicable local regulations
Disconnect the device from the power supply and clean and disinfect
according to the section "Cleaning and disinfection".
4MAINTENANCE
The BW 685 / BW 685 S is designed as a low-maintenance device. To
preserve the quality and functional safety, please comply with the following
points:
12DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
•Always keep the device clean (see the section: "Cleaning and
disinfection").
•The periodic technical safety inspections must be carried out (see the
section "Periodic inspections")
5CLEANING AND DISINFECTION
The BW 685 / BW 685 S and the optional tube warmer Tubeflow may only be
cleaned using a soft cloth with a water-soluble, non-aggressive cleaning
agent or a special cleaning agent for plastics.
For the purposes of disinfection, ready-made alcohol-based spray
disinfectants can be used.
Important: Before cleaning or disinfecting, the device must always be
disconnected from the mains power supply by pulling the plug.
DE - Edition 02 / 200713
Instructions for Use – BW 685 / BW 685 S BIEGLER
Do not disinfect the device with steam (i.e. in autoclaves), hot air or
thermochemical cleaning solutions.
6PERIODIC INSPECTIONS
The periodic technical safety inspections (according to the local standards in
force – e.g.. in Austria ÖVE/ÖNORM E 8751-1) must be carried out on the
BW 685 / BW 685 S at least every 12 months by persons able to carry out
such technical safety inspections based on their training, knowledge and
experience gained by practical activities.
The results of the periodic inspection are to be documented, together with the
date and the inspecting agency.
Important: Should a malfunction be evident, suitable warning signs should be
attached to the device to ensure that it cannot be used before necessary
service and repair work has been carried out.
CHECKING THE WARM-UP PERIOD
This is the time taken by the BW 685 / BW 685 S to heat up to 38.5°C from
room temperature. The device is malfunctioning if it takes much longer than
one minute.
CHECKING THE CONTROL TEMPERATURE
The control temperature is checked on the groove bed of the heat exchanger.
The sensor of a suitable contact thermometer (tolerance +/- 0.15°C) is fixed
to this position, e.g. by using a piece of infusion tubing. The examination is
performed at a temperature setting of 38.5°C.
The measured value is read after it has stabilised. The difference must not
exceed +/- 0.5°C. There is a malfunction if a difference from the control
temperature of greater than +/- 0.5°C is obtained.
CHECKING THE LOW TEMPERATURE ALARM
Preheat the device to 38.5°C, then disconnect the mains plug. Hold down the
control and reconnect the mains power plug. After the beep has ceased,
release the control . Press the switch . The device is now in an
operational state where all the alarms are active, but the heating is switched
off. The BW 685 / BW 685 S now slowly cools down. When the temperature
drops below the 36.5°C threshold, the low temperature alarm should trigger.
14DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
For reasons of safety, short beeping sounds are given at intervals of a
second in this operational mode and the STANDBY indicator (Diag.1/4)
flashes. There is a malfunction if the low temperature alarm is not triggered.
CHECKING THE HIGHTEMPERATURE ALARM
Preheat the device to 41°C and wait for the temperature to stabilise, then
disconnect the mains plug. Hold down the control and reconnect the
mains power plug. After the beep has ceased, release the control . Press
the switch . The device now slowly heats up to a target temperature of
42.5°C. Observe the temperature display carefully. The high temperature
alarm should be triggered at a temperature of 42°C.
For reasons of safety, short beeping sounds are given at intervals of a
second in this operational mode and the STANDBY indicator (Diag.1/4)
flashes. There is a malfunction if the high temperature alarm is not triggered.
VISUAL CHECK OF GENERAL CONDITION
The device should be checked for mechanical damage (general condition)
and for the completeness of the sticker information, particularly the device
label on the reverse. There is a malfunction if mechanical damage to the
device is evident which could be harmful or impair the functional operation of
the device.
ELECTRICAL SAFETY
All relevant electrical safety data should be checked, particularly the earth
conductor resistance (Target: < 0.3 Ohm) and leakage current (Target: < 0.75
mA). There is a malfunction if there is a measured value outside the
tolerances.
DE - Edition 02 / 200715
Instructions for Use – BW 685 / BW 685 S BIEGLER
7MANUFACTURER LIABILITY
The manufacturer and the supplier of the device reject all liability if:
•the device is not used in accordance with the instructions for use
•the user is not sufficiently informed about the functioning of the device on
the basis of the instructions for use and the safety instructions
•repairs are not performed exclusively by the manufacturer or by persons
and service centres expressly authorised by the manufacturer
•the device is used in places in which the electrical installations do not
comply with the applicable national standards, or if power supply during
the period of use of the device is not guaranteed
•original spare parts material are not used or the maintenance interval
is not complied with.
•disposal of the device and its accessories is not carried out in
accordance with the applicable local regulations.
8WARRANTY CONDITIONS
The manufacturer guarantees that all flaws of material and workmanship
which arise within 24 months from the date of purchase will be repaired free
of charge.
Claims are only accepted under the following terms:
•The manufacturer and/or supplier is informed immediately of the fault for
which the warranty claim is being made.
•The instructions of the manufacturer and/or supplier on storage or return
of the device are complied with.
•Presentation of a legible copy of the invoice for the device concerned,
showing the date of purchase.
•An exact description of the defects or malfunctions identified by the
customer.
16DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
The manufacturer's warranty will be void if it is found that the maintenance,
disinfection and inspection instructions have not been followed according to
the instructions for use, the device has been damaged by force or operating
error or has been used in any way contrary to the instructions for use and
safety instructions. The warranty will also be void if original Biegler materials
were not used as replacement parts, or measures for repair were undertaken
by persons not authorised by the manufacturer or supplier.
If the manufacturer is required to meet a warranty claim in accordance with
these terms, the customer shall bear the costs and risks of transport of the
device from and to the place of use.
The manufacturer and/or supplier shall under no circumstances assume
liability for slight negligence. The compensation for lost earnings and profits is
likewise excluded.
9RETURN OF DEVICES
Devices must be carefully cleaned and disinfected before being placed in the
original packaging for returning.
If the original packaging is no longer available, the product has to be suitably
packaged for the method of dispatch.
DE - Edition 02 / 200717
Instructions for Use – BW 685 / BW 685 S BIEGLER
10 ELECTROMAGNETIC COMPLIANCE
Table 201
Guidelines and manufacturer's declaration – electromagnetic emission
The BW685 is intended for use in the environment specified below. The customer or the user of the BW685 should assure
that it is used in such an environment.
Interference
emission
measurements
RF emissions acc.
to CISPR 11
RF emissions acc.
to CISPR 11
Emission of
harmonics acc. to
IEC 61000-3-2
Emission of
harmonics acc. to
IEC 61000-3-3
Complian
ce
Group 1The BW685 uses RF energy only for its internal function. Therefore, its RF emissions are
Class BThe BW685 is suitable for use in all establishments, including domestic establishments
Class A
Complies
Electromagnetic environment - guidelines
very low and are not likely to cause any interference in nearby electronic equipment.
and those directly connected to the public power supply network that also supplies
buildings used for domestic purposes.
18DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
Table 202
Guidelines and manufacturer's declaration – electromagnetic interference resistance
The BW685 is intended for use in the electromagnetic environment specified below. The customer or the user of the BW685
should assure that it is used in such an environment.
Interference
resistance test
Electrostatic discharge
(ESD) acc. to
IEC 61000-4-2
Fast transient
/electrical bursts acc.
to IEC 61000-4-4
Surge voltages
(surges) acc. to
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations of
the supply voltage
acc. to
IEC 61000-4-11
Power frequency (50
Hz/60 Hz) magnetic
field acc. to IEC
61000-4-8
IEC 60601test level
± 6 kV contact
discharge
± 8 kV air discharge
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV differential
mode
± 2 kV common
mode
< 5% UT (> 95% dip
in UT)
for ½ cycle
40% UT (60% dip in
UT)
for 5 cycles
70% UT (30% dip in
UT)
for 25 cycles
< 5% UT (> 95% dip
in UT)
for 5 s
3 A/m3 A/mPower frequency magnetic fields should be at
± 1 kV normal mode voltage
± 2 kV common mode
voltage
< 5% UT (> 95% dip in UT)
for ½ cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
< 5% UT (> 95% dip in UT)
for 5 s
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
The mains power supply quality should be that
of a typical commercial or hospital
environment.
The mains power supply quality should be that
of a typical commercial or hospital
environment.
The mains power supply quality should be that
of a typical commercial or hospital
environment.
If the use of the BW685 requires continued
operation during power mains interruptions, it
is recommended that the BW685 be powered
from an uninterruptible power supply or a
battery.
levels characteristic of a typical location in a
typical commercial or hospital environment.
NoteUT is the AC mains voltage prior to application of the test level.
DE - Edition 02 / 200719
Instructions for Use – BW 685 / BW 685 S BIEGLER
1V
P
V
d
∗
=
1
5,3
1E
P
E
d
∗
=
1
5,3
P
E
d
∗
=
1
7
Table 204
Guidelines and manufacturer's declaration – electromagnetic interference resistance to
The BW685 is intended for use in the electromagnetic environment specified below. The customer or the user of the BW685
should assure that it is used in such an environment.
Interference
resistance test
Conducted RF
disturbance variables
acc. to IEC 61000-4-6
Emitted RF
disturbance variables
acc. to
IEC 61000-4-3
IEC 60601test level
3 V
eff
150 kHz to 80
MHz
3 V/m
80 kHz to 2.5
GHz
Compliance
level
V/m
Electromagnetic environment - guidelines
Portable and mobile RF communications equipment should be
used no closer to any part of the BW685, including cables,
than the recommended protection distance calculated from the
equation applicable to the frequency of the transmitter:
Recommended protection distance:
in V
in
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended protection distance in metres (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site surveya should be less than the
compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked
with the following symbol:
for 80 MHz to 800 MHz
for 800 MHz to 2.5 GHz
NOTE 1At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by
a
b
Field strengths from stationary transmitters, such as base stations for radio (cellular /cordless) telephones and
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
absorption and reflection from structures, objects and people.
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to stationary RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which
the BW685 is used exceeds the applicable RF compliance level as specified above, the BW685 should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the BW685.
20DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
P
V
d
∗
=
1
5,3
P
E
d
∗
=
1
5,3
P
E
d
∗
=
1
7
Table 206
portable and mobile RF communications equipment and the BW685
The BW685 is intended for use in an electromagnetic environment in which radiated RF disturbance variables are controlled.
The customer or user of the BW685 can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the BW685 as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter W
0.010.120.120.23
0.10.370.370.74
11.171.172.33
103.693.697.38
10011.6711.6723.33
For transmitters rated at a maximum output power not listed above, the recommended distance can be determined using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
Note 1At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Recommended protection distances between
Protection distance according to transmitter frequency m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
11MANUFACTURER'S DECLARATION
The blood and infusion warmer BW 685 / BW 685 S is a medical product as
defined by Directive 93/42/EEC.
This is documented by attaching the CE mark.
Notified Body: TÜV Product-Service, Approval Number CE0123
DE - Edition 02 / 200721
Instructions for Use – BW 685 / BW 685 S BIEGLER
12SYMBOLS
Certifies compliance with the Directive 93/42/EEC
0123
Observe the instructions for use!
Do not dispose of this product as unsorted municipal waste
Control for increasing the temperature setting
Control for decreasing the temperature setting
IPX 4
- for optional tube warmer Tubeflow
22DE - Edition 02 / 2007
Humidity Classes
Degree of protection B
Control for switching On / Standby
OperationMalfunction
I
Do not coverDo not wrap up
Do not expose to direct
sunlight Do not cut
BIEGLER Instructions for Use – BW 685 / BW 685 S
13OPERATING AND STORAGE CONDITIONS
Permissible environmental conditions for transport and storage:
Temperature:10 – 40°C
Relative humidity:30 – 75%
Air pressure:700 – 1060 hPa
The ambient operating temperature must be in the range of 10 – 30°C.
Values higher or lower than the ranges specified above may cause damage
to the device or its accessories.
14TECHNICAL DATA
Device:Blood and infusion warmer
Type designation:BW 685 / BW 685 S
Voltage:230 V / 50Hz
Power consumption:1.5 A
Protection class:I
Degree of protection:B
Humidity protection:IPX4
Fuses:primary 2 x 1.6 AT secondary 500 mAT
Control temperature:37°C - 41°C adjustable in 0.5°C steps
Overtemperature shutdown:42°C / 42.5°C / 45°C±3°C
Max. system pressure:300 mm Hg
Classification:IIb according to Rule 9
Operational mode:Continuous operation
Dimensions BW 685:WxHxD 228 x 228 x 132 mm
Weight BW 685:1.9 kg
Dimensions BW 685 S:WxHxD 228 x 278 x 132 mm
Weight BW 685 S:2.0 kg
Tube warmer Tubeflow
for BW 685 S:Length 1340mm
DE - Edition 02 / 200723
Instructions for Use – BW 685 / BW 685 S BIEGLER
15MANUFACTURER
Biegler GmbH
Allhangstrasse 18a
3001 Mauerbach
Austria
Tel. +43 1 979 21 05
Fax +43 1 979 21 05 16
email: office@biegler.com
www.biegler.com
24DE - Edition 02 / 2007
BIEGLER Instructions for Use – BW 685 / BW 685 S
Space for notes:
DE - Edition 02 / 200725
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.