BHM-Tech Apollon Instructions For Use Manual

Instructions for use
3
Instructions for use
Thank you very much for your decision to use a quality product from BHM-Tech. The digital bodyworn hearing system apollon is produced in Austria according to medical device directive 93/42/EEC and has been developed and manufactured according to the latest audiological knowledge and state-of-the­art technical innovations. It has the CE-mark and follows therefore the newest requirement of quality and safety regulations.
If you follow the Instruction for use in detail, your digital bodyworn hearing system apollon will be your reliable partner for many years.
Your digital bodyworn hearing system apollon has been especially adapted to your individual hearing loss by a qualified hearing aid expert. Your hearing aid expert will also inform you about the correct use, handling and operation of apollon. Your hearing aid expert will also supply you with information on regular maintenance measures.
READ THIS INSTRUCTION FOR USE CAREFULLY BEFORE USING THE DEVICE. Please ensure that you have followed all the relevant procedures within the Instruction for use before you attempt to operate the device. The device should not be used if you suspect that it is not functioning or damaged. Please ensure that the hearing system does not have any mechanical problems. If the instrument has been dropped or otherwise damaged, please send it to your acoustician or ENT to review and repair.
Never pass on your digital bodyworn hearing system apollon to another person and prevent unauthorized use.
Intended use:
Hearing systems are developed to improve hearing. The essential function of a hearing system is to pick up sound, amplify it and reproduce the signal at the ear drum of the hearing-impaired person. Your hearing system is suitable for compensating sensorineural and combined hearing loss.
4
Instructions for use
Warning note for the hearing aid acoustician/ ENT specialist:
This hearing system can generate an output sound pressure level of 132 dBSPL. Particular care must be taken when adjusting the device as there is a risk of damage to the user´s residual hearing.
Swallowing of small parts of the hearing system or batteries may be hazardous to your health and may cause severe injuries and even death!
Index:
Instructions before use of the aid
• Parts of the aid
• Commissioning of apollon
• Fitting - Programming of apollon
Instruction during the use of the aid Handling of apollon
• Battery type
• Insertion of the batteries
• Volume control
• Mode control switch
• Microphone mode
• Telecoil mode
• Microphone-telecoil mode
• Switching off
• DAI (Direct Audio Input)
• Line-In 3.5mm
• Optical low battery warning - LED
• Summary of signal tones
Instructions after the use of the aid
• Maintenance of apollon
• Dispose of apollon and batteries
General Warnings General Information
5
Instructions for use
Technical data according IEC 60118-0
Max. output (dBSPL)
154
Max. gain (dB) 94
Current consumption (mA) 13,2 + - 10 % @ 3 V
Battery size AA
Average lifetime of the batteries (hours) (at capacity of 2500 mAh)
Appr. 189 hours
2
1
3
5
4
8
9
10
6
7
Instructions before use of the aid
Parts of the aid
1
Volume control
2
O-T-MT-M mode control switch
3
Low battery warning LED
4
Battery compartment
5
Paediatric lock ( / )
6
DAI
7
Line-In
(3,5 mm)
8
Shirt-clip
9
Receiver cable
10
Earphone
T O
M
MT
6
Instructions for use
Commissioning of apollon
Your ENT specialist or acoustician will instruct you in detail about the handling of the hearing system and about the accessories.
TO ENSURE SAFETY AND FUNCTIONALITY OF THE HEARING SYSTEM USE ONLY ACCESSORIES APPROVED BY BHM-TECH.
Cable - connections
Plug in the 2-pin end into the socket of the earphone and plug in the 3-pin end into the 3-pin socket at the aid-housing. Mind that all connections are polarized and must be plugged in right position! See fig.1 and 2.
Fig.1
Fig.2
7
Instructions for use
Assembling of the earmould
Your individual earmould should be assembled onto the earphone by the acoustician. Your ENT specialist or hearing aid acoustician will instruct you about the handling and daily maintenance of your earmould. In case of unpleasant comfort or skin irritation contact your ENT specialist or acoustician immediately.
Fitting – Programming of apollon
(For the ENT specialist or acoustician only!)
Important information:
The aid is pre-configured as monaural system by the factory!
1. Before programming
Remove the shirt-clip as shown in fig. 3a and 3b.
Attention: During programming, batteries must be inserted.
Fig.3a
Fig.3b
8
Instructions for use
Connect the programming cable to the aid and to the programming interface.
Attention: Ensure that the cable is connected proper according the markings on the cable plugs and according the markings on the sockets on the aid and on the programming interface (see fig. 4).
2. Programming software
To ensure correct function of the aid, use only the software and the equipment approved by BHM-Tech.
3. Post programming
Unplug the programming cable and mount the shirt­clip as shown in fig. 5 and 6.
Fig.4
9
Instructions for use
Instructions during the use of the aid
Handling of apollon
Battery type
Only use quality batteries Type AA, Batteries are not part of the delivery.
Insertion of the batteries
The apollon is equipped with a paediatric lock for the battery compartment. By sliding the red knob (5) to the right - the battery compartment (4) is locked; sliding to the left the battery compartment is unlocked (see fig. 7 and 8). Open the battery compartment by sliding it from the housing as shown in fig. 10 and 11. Insert the batteries polarized correctly. Hint: Remove the batteries when the aid is not in use for longer time. In case a 5 times low-pitch peep sounds and the low battery warning LED continuously lights up, change the batteries.
Fig.5
M
MT
T
0
1
2
3
4
Fig.6
10
Instructions for use
M
MT
T 0
Low battery warning LED
Fig.7
Fig.8
open
Fig.9
Fig.10
Fig.11
locked
11
Instructions for use
Volume control
With the volume control (1) (see fig. 12) you can adjust the volume for your convenience. Adjusting range: 1 … low volume MAX … maximum volume
Hint: Before switching on the aid, changing the modes, or connecting external devices to the aid, always turn the volume control to 1 (lowest volume). Afterwards adjust the volume according to your convenience. Recommendation: For your convenience, fit the volume of the aid in accordance with your environment.
Fig.12
12
Instructions for use
Mode Control Switch
apollon is equipped with a 4 position O-T-MT-M switch (2) (see fig.13). The chosen mode will be indicated with peep-tones. The signal-tones can be set and deactivated by using the “Setting” function in the fitting software.
Position O: Aid is switched OFF Position T: Aid is in telecoil mode (3 times peep-tone) Position MT: Aid is in telecoil & microphone – mode (2 times peep-tone) Position M: Aid is in microphone mode (1 time peep-tone)
Switching on
After switching on the aid, the battery-status will be checked and the LED-lamp (3) lights up and goes off when the batteries are in good condition.
Fig.13
13
Instructions for use
Microphone mode
Switching to the “M” position, the input signal from the internal microphone is used to receive acoustic signals.
Telecoil mode
Switching to the “T” position, the input signal from the internal Telecoil is used to receive inductive signals. The „T“-mode can be used in official buildings, cinemas, theatres, equipped with inductive­system or at home with inductive-accessories. It enables you to listen without disturbance.
Microphone-telecoil mode
Switching to the “MT” mode you are in the combined telecoil-microphone mode.
You can hear signals from the integrated microphone and from the integrated telecoil at the same time. This mode is useful in schools, and also in buildings equipped with inductive-system.
Switching OFF
To turn OFF the aid move the switch (2) to position “O”.
DAI (Direct Audio Input)
The apollon is equipped with a DAI (6) which is placed on top of the aid beside the microphone port. It enables you to use hearing aid accessories like blue-tooth link to your cell-phone, table microphone and so on. Consult your acoustician about the possibilities.
HINT: TO ENSURE SAFETY AND FUNCTIONALITY OF THE HEARING SYSTEM USE ONLY EXTERNAL DEVICES APPROVED BY AUTHORIZED HEARING AID MANUFACTURER OR BHM-TECH. THE USE OF UNAUTHORIZED DEVICES MAY BE HAZARDOUS TO YOUR HEALTH.
Line-In 3,5mm
apollon is equipped with an external audio input (7). Hint: Before put on an external device on the external
audio input, make sure that the volume control is on position1!
14
Instructions for use
Your acoustician or ENT-specialist will advise you about the opportunities of useful and available external accessory devices.
Hint for the use of external audio input and DAI­input:
In case an external accessory device is connected to external audio input or DAI, the hearing device apollon switch automatically to Line-In or DAI-mode.
A beneficial use in the same time of audio Input and DAI-Input is not possible! In case of trying to use both external inputs, the external audio Input will be automatically preferred.
Consult your acoustician or ENT-specialist about the opportunities of external accessory devices and their applications.
15
Instructions for use
Optical low battery warning - LED
apollon is equipped with an optical battery warning (3)
The batteries of apollon have to be replaced as soon as the optical battery warning lights up.
Receiver and receiver cable:
apollon is equipped with a receiver (10) and a receiver cable (9). Your hearing aid expert will advise you in handling and maintenance of receiver and receiver cable.
Hint: TO ENSURE SAFETY AND FUNCTIONALITY OF THE HEARING SYSTEM USE ONLY RECEIVERS AND RECEIVER CABLES APPROVED BY BHM-TECH. THE USE OF UNAUTHORIZED RECEIVERS AND RECEIVER CABLES MAY BE HAZARDOUS TO YOUR HEALTH.
Shirt-clip:
apollon is equipped with a shirt-clip (8). Your acoustician or ENT specialist will advise you about the wearing opportunities with this shirt-clip.
M
MT
T 0
Optical low battery warning LED
Fig.14
16
Instructions for use
Summary of signal tones
Changing the mode of O-T-MT-M switch up to 3-times high peep tone
Use of DAI 4-times high peep tone
Use of ext. audio input (Line-In) 5-times high peep tone
Instruction after the use of the aid
Maintenance of apollon
Perspiration and humidity should be removed daily from your hearing system. Use a cleaning brush or soft tissue to remove debris from around the switches, microphone and battery compartment. Always take off your hearing system before taking a
shower or washing your hair, before going to the sauna or in a steam bath.
For extra hygienic and skin-protective cleaning, we recommend special cleaning agents available at the relevant shops. Never use any chemical detergents, water, cleaning fluids or oil to clean the apollon.
Dispose of apollon and batteries
Your hearing system apollon, accessories and batteries are parts that belong to the group of products of modern technology which have to be especially disposed of according to environmental protection. Please make sure that you discard your hearing system properly as hazardous waste. Your hearing aid expert will be glad to assist you in this matter.
17
Instructions for use
General Warnings
Incorrect use of your hearing instrument apollon, accessories and batteries may be hazardous to your health. Please read following points carefully.
The basic adjustment of the hearing system apollon is carried out by an authorized hearing aid expert according to the hearing loss of the patient. The end user is kindly requested only to operate those control elements designed for his or her easily accessible use according to instructions.
Switch off your hearing system in mines or other areas where explosions could take place, unless your hearing system is especially certified for such use.
Always bear in mind that your hearing system apollon will not perform properly, if e.g. the battery is low. This is a especially hazardous, if you move in
traffic or are otherwise dependent on warning signals. Remove the batteries completely when apollon is not in use.
Store batteries and parts of the hearing system apollon out of reach of babies and small children. Never change batteries of the hearing system apollon in the presence of babies and small children.
Keep apollon and any parts of it, which you have taken off, out of reach of other persons or children!
Never try to recharge a disposable battery!
Swallowing of batteries may be hazardous to your health and may cause severe injuries and even death. For this reason, never store batteries together with medicine. Immediately consult a doctor, if a battery has been swallowed.
18
Instructions for use
Incorrect manipulation or rough handling of the hearing system apollon may cause detachment of small components of the hearing system. If any small parts are swallowed by mistake or inserted into the ear or nose please contact your doctor.
The hearing system apollon is able to transmit signals of very high volume. The fitting of the device must be carried out by an authorized hearing aid expert.
Never expose your hearing system to extreme weather conditions, like strong wind, hot sun (e.g. behind the windshield of a car), direct heat (do not put it on a radiator) or severe cold (overnight in a cold car) and avoid contact with chemicals and strong detergents.
If you are in an area that may have electromagnetic interference fields, such as a radio transmitter, the audio quality of your hearing system may be affected.
Consult a doctor in case of skin irritations of the ear or head.
Be aware that devices which are connected to the line-In Input of apollon has to comply with the safety requirements of EN 60601 or EN 60065.
General Information
It is absolutely necessary that a defective hearing aid must be repaired by an expert. Never try to open the hearing device yourself. Improper handling and operation will lead to the loss of guarantee. If your hearing system becomes defective, please consult your hearing aid expert. Regular maintenance and check-ups by your hearing aid expert may prevent the need for repairs.
19
Instructions for use
according to EC-Directive 93/42 EEC Quality certificate according to DIN EN ISO 13485
Version 1.5- 03/2014
Content may be changed without notice in order to improve product performance.
0297
BHM-Tech Produktionsgesellschaft mbH Grafenschachen 242 7423 Grafenschachen, Austria
www.bhm-tech.at
Made in Austria
-20°C
+60°C
10%
95%
Loading...