BEYES EndoPilot Instructions For Use Manual

Cordless Endodontic Handpiece
EndoPilot
Instruction for Use
2
Thank you for purchasing EndoPilot. Read this Operation Manual carefully before use for operation instructions and care and maintenance guidelines. Keep this Operation Manual for future reference.
Application
This product is a cordless handpiece used primarily for root canal enlargement.
Contraindication
It is prohibited to use in them such as patients with hemophilia or heart disease, Patients and doctors with pacemakers; Pregnant women and young children need to use caution.
User
Only qualied personnel is allowed to use the unit only in dentistry..
Prohibition
Do not use this motor handpiece for formation of an extremely bent root canal.
Do not use this for implants other than endodontic treatment or other dental
treatment.
Classication of Devices
Classication by type of protection against electric shock
— Class II devices
Classication by degree of protection against electric shock
— Applied part type B
Classication by sterilization or disinfection method allowed by the manufacturer — Refer to Sterilization
Classication by mode of operation
— Continuously operating device
Degree of protection against ingress of water: IPXO
Degree of safety of application in the presence of a ammable anesthetic mixture with
air or with oxygen or nitrous oxide: Equipment NOT suitable for use in the presence of a
ammable anesthetic mixture with air or with oxygen or nitrous oxide
3
This conforms to CE European Directive of “Medical equipment directive 93/42/EEC.”
Follow the waste of electric and electronic equipment (WEEE) Directive (2002/96/CE) to dispose of the product and accessories.
Class II equipment
Type B conforming component
Refer to the Operation Manual
Consult accompanying documents (user manual)
Manufacturer
Authorized representative in the European Community
0197
Symbols
4
Table of Contents
Cautions for handling and operation...................... 5-6
1. Features ......................................................... 7
2. Specication ................................................... 8-9
3. Name of each part .......................................... 9
4. Parts and its function ...................................... 10-12
5. Operations ...................................................... 13-19
6. Convenient Function ...................................... 19-21
7. Cleaning ......................................................... 22
8. Sterilization ..................................................... 23
9. Changing Battery............................................ 24
10. Error Code ...................................................... 25
11. Troubleshooting.............................................. 25-26
12. Warranty ......................................................... 27
13. Disposing Product .......................................... 27
14. Electromagnetic emissions and immunity ...... 27-31
Warranty Card ....................................................... 31
5
Cautions for handling and operation
Read these safety cautions thoroughly before use and operate the product properly.
These indicators are to allow you to use the product safely and prevent danger and harm to you and others. These are classied by degree of danger, damage and
seriousness. AII indicators concern safety, therefore always follow them.
Classication Degree of danger or damage and seriousness
Explains an instruction where death or serious injury may occur.
Explains an instruction where bodily injury or damage to device may occur.
Explains an instruction where possibility for minor to medium bodily injury or damage to device may exist.
Explains an instruction that should be observed for safety reasons.
DANGER
Use the specied batteries for this product. Never use any other batteries than United Dental species
This product is designed specically for use with rechargeable batteries. Do not
use manganese or alkaline batteries. Using and charging these by error may cause a liquid leakage or explosion.
High pressure from the outside will cause a liquid leakage or explosion
Keep the motor and battery away from the re, otherwise it will cause a
liquid leakage or explosion
Sterilizing the motor by heat or steam may cause a liquid leakage or explosion.
WARNING
Portable and mobile RF communications equipment can affect
Medical Electrical equipment. Do not use RF equipment outskirts for the product.
If the handpiece has not been used for long time, check it before use
keep your hands dry when operating or connecting the charger base
keep the handpiece away from water and chemical solution which may cause short circuit, re and other dangerous
Do not disassemble the motor handpiece. The motor can not be to be modied
If there’s liquid leakage on the battery, or handpiece deform or fade, immediately
stop using and contact with your dealer.
Do not charge the handpiece without battery.
DANGER
WARNING
CAUTION
NOTICE
6
Keep the charging interface away from something conductive such as copper scrap.
Please use original charger to charge the battery.
Connect the handpiece to the charger base, the handpiece LED will indicate the
charging state, if not, please refer to the trouble shooting.
There’s electronic circuit which control the torque (TORQUE LIMITER Function) to prevent les from breaking; however, les may still break due to metal fatigue
if the torque is conditioned to be higher. Please check the working instruction of
le before use
CAUTION
Fully training to use this product to guarantee the safety of patient.
The product is used for dental treatment only by qualied personnel.
Do not use a bent, damaged, deformed or non- ISO-conforming le. It may cause personal injury due to sudden breakage or ying off during rotation.
Does not use or leave the product in a high-temperature environment such as
under strong direct sunlight.
Check the product before use, if there’s abnormal loose, vibration, noise and temperature (heat generation). Immediately stop using and contact with your dealer.
Keep the handpiece away from dirt, it will affect the concentricity of contra angle and the judgment of the torque.
Before changing the contra angle or le or Maintenance of motor, turn off the
power of the motor.
Do not lubricate any part of the motor except the contra angle.
If use corrosive solutions during the operation, please clean the motor handpiece
immediately after operation.
This equipment is for indoor use only.
Follow the instruction of le manufacture to set the rotation and speed of the
motor.
NOTICE
The motor use the rechargeable li-ion battery.
The power consumption is low when the power is off, please fully charged after
each use, please charge at least 1 time a month if long-term use.
When the battery is low, the motor automatically shut down, leave it for a while and the motor can be turned on again, but this is not permitted, it may damage and reduce the battery life.
If the product is not used for a long period, please take the battery out.
If the power is not left completely, the battery may be damaged, please replace a
new one.
No need for the training of this product, please read the instruction before operating and maintenance.
7
1. Features
Ergonomic design, small size. The max diameter less than 28mm. The head holding
part is just 16mm.Therefore dentists can get better sight and more convenient operation
“Multi - position type head" allows direction of contra angle to be changed easily.
The LCD panel enhances user friendliness.
Operates continuously for more than 2 hour (Dependent upon rated load)
Memory for up to 5 programs. RESET bottom can be use once the error occurs.
The auto reverse function is activated depending on the Comparison between load
and set torque. A wide variety of functions such as “AUTO REVERSE", “AUTO STOP”,“AUTO REVERSE OFF", “REVERSE” and “AUTO RECIPROCATING” are available.
The motor can be turned on and off by pressing the ON/OFF key. Alternatively the unit can be operated by holding down the ON/OFF key during operation and letting go of the ON/OFF key to stop the handpiece.
The motor softly starts. Since the rotation briey stops before changing direction
there are no vibrations and shocks during the change of directions.
The motor will be turned into sleeping mode without any action in 10 minutes, click "power" key to wake up
The rotating speed and torque can be compensated perfectly by the controlling chip
Charging interface is designed as MICRO USB
90 minutes, battery will be fully charged.
The contra angle provided for this product can be autoclaved at 135°C (excluding
the motor).
8
2. Specication
Battery Adaptor
Adaptor Mode G DP06AH-0500500- EU
Input Voltage 100-240 - 50/60Hz 0.25A
Input Power 2.5vA
Output Voltage DC5V/0. 5A
Motor Handpiece
MODE EndoPilot
Input Voltage DC3.7V
Input Power Max7VA
Charging Time Approx.90. min
Use environment
Temperature 10-40 oC
Humidity 10-70%(Non condensing)
Atmospheric Pressure 1060hPa
Store/transport environment
Temperature -10- 50’
Humidity 10-80%(Non condensing)
Atmospheric pressure 500-1060hPa
Performance
Torque
Gear Min Torque(N.CM) Max Torque(N.CM)
4:1 0.1 0.7
10:1 0.2 2.0
16:1 0.3 3.0
20:1 0.4 4.0
9
Motor handpiece
Charging base
(
ALP 121)
Adaptor
(ALP 107)
Battery
(ALP 104)
Charging wire
(
ALP 108)
16:1 contra angle
(ALP101)
Silicone Sleeve
(ALP 119)
Spray nozzle
(ALP 111)
3. Name of each part
Rotation speed
Gear Min Rotation speed(rpm) Max Rotation speed(rpm)
4:1 500 2250
10:1 200 900
16:1 140 550
20:1 100 450
10
4. Parts and its function
POWER Key ( )
Holding down the POWER key for more than one second turns on the power and the LCD panel lights.
Holding down the POWER key for more than one second while the power is on turns off the power and the LCD display turns off.
ON/OFF ( )
Pressing this switch, when the power is on, rotates the motor handpiece, and de-pressing the switch again stops the hand piece. (Normal rotation operation) Pressing and holding the switch for approx. one second or longer rotates the motor handpiece and releasing the switch stops the handpiece. (Temporary rotation operation)
11
SELECT Key ( ) (When the motor is running, the key is invalid)
Press the SELECT key to ‘speed , torque or gear ratio setting', or ‘ Rotation Mode selection'
Press shortly the SELECT key, it will be displayed “speed, torque or gear ratio setting”
Enter “speed, torque or gear ratio setting” to make the change of the related setting. By the sequence of SPEED — GEAR RATIO- TORQU E-EXIT, in circle; adjust the value by press + or -
Press “P” a little long time to remember the setting, for future use
To quit the “speed, torque or gear ratio setting"; 1 , press “S” when setting torque; 2,
Press “S" a little long time ; 3, press ON/OFF ,
“P" or operation to quit.
Pressing and holding this SELECT Key for 1 second or longer, when the motor is
stopped can change the Rotation Mode.
Enter the Rotation Mode selection: the sequence will be, AUTO REVE RSE-AUTO STOP-AUTO REVERSE OFF- REVERSE- RECEPROCATIN G, in circle: adjust the value by press + or -
Press “P” a little long time to remember the setting, for future use
To quit the ‘Rotation Mode selection’; 1 , Press “S"; 2, press ON/OFF , P or
“operation" to quit.
PROGRAM Key ( ) (When the motor is running, the key is invalid)
A program can be selected. The present program number is displayed. There are 5 programs available, numbered from 1to 5.
Up to ve programs can be set, and program settings can be memorized by pressing
this key for one second or longer.
The following parameters can be set: Speed, torque, gear ratio and auto reverse setting.
+/- Key ( )
Use this key when adjusting the torque.
Unit with other function to adjust the value.
12
LCD panel
Auto Reverse
: If the load is removed after auto reverse rotation, it returns to the normal rotation again. : If the load is removed after auto reverse rotation, it stops. : Auto reverse rotation is not activated. (There will be no symbol displayed) : rotation is in anti-clockwise motion. :rotate as reciprocating mode, “REC” dis played
Battery Symbol The symbol indicating the battery status.The symbol will be animated when the batteries are being charged or in the refresh mode
:Full charge or nearly fulI charge : About 30-80 % remains : Less than about 30% remains : Batteries are d rained or in a remarkably low voltage. Charge the batteries.
: The symbol ashing with alarm. Please turn off the motor,
and charge the batteries at once
Notice
The symbol indicating the remaining amount of the batteries indicates a voltage. When load is applied to the motor handpiece, the symbol indicating the remaining amount of the battery charge appears to become lower
Alarm Symbol
: Max alarm : Lower alarm : OFF Alarm can be turn off (Refer to 6 Convenient Operation, Alarm Sound Volume Control)
13
5. Operations
5.1 Charging Batteries
1. Insert the power cord jack into the inlet at the back of the battery charger. (Fig. a)
2. Insert the power cord and plug in. Make sure you have the correct model/voltage. (Fig. b)
3. Turn on the power switch. At this time, check that the power lamp lights up.
4. Insert the motor handpiece into the battery charge. Charging starts with the charge
mark ashing on LCD.(Fig. c)
5. When the buzzer sounds and is displayed on the liquid crystal panel, charging is completed.
Notice
The handpiece can be charged directly by USB cable, even there is no power cord. But please ensure the handpiece be pIaced in dry condition.
Warning
When the buzzer does not sound and the charging animation is not displayed, please take off the handpiece, check the “ERROR, CODE” to nd out the
problem.
14
CAUTION
When the power lamp for the charger does not light up, the internal fuse may be
faulty. In this case please contact your dealer.
Be careful of handling when placing the motor handpiece into the charger. Do not
put handpiece into the WET or UNSAFE place.
Never use the battery charger from other manufacturer.
The charging normally takes approx. 90 minutes, but it depends on battery use
conditions, battery freshness, ambient temperature, etc.
Under the charging condition, the handpiece was not allowed to rotate.
The completely discharged will cause the damage of the battery. The user
should charge the battery completely once a month, or once a bi-month, if the handpiece will not be used for long term.
Do not put anything (dust, especially the metal debris) other than the handpiece
on the charger. Please protect the charger carefully.
PuII out the machine from the charge base up straightly after the battery fuIly
charged.
Charging may not be started in the following case.
Lower than approx. 0° C or higher than approx. 40° C
Battery voltage is sufcient
Battery voltage is ab normal
CAUTION
Turn OFF the power to remove or attach the contra angle.
Do not use the contra angle from other manufacturer
Check that the contra angIe is securely connected to the motor handpiece.
5.2 Changing Contra AngIe
The contra angIe can be connected with the motor handpiece at 6 adjustable head positions. Align the positioning pins of the contra angle with the positioning slots of the motor handpiece and insert the head untiI they click. When removing the contra angle head, puII it out axially.
Positioning convex
Contra angle head
Positioning groove
Motor handplece
15
CAUTION
When attaching and detaching the le, turn off the power beforehand.
After the le is locked in place, lightly pull out the le to make sure the le is
locked.
Always clean the shank of the le to be installed before use. Allowing dirt to enter
the chuck could cause loss of concentricity and deterioration of chucking force.
Do not exceed the rotation speed recommended by the le manufacturers.
5.3 Mounting and Removing File
Insert le to contra angle, lightly turn the le until it engages with the latch mechanism. Push it inward to click File Removal, depress the push- key and puII out the le.
5.4 Preparatory Operations
1. Hold down the POWER key for more than one second to turn on the power.
2. Keep holding d own the PROGRAM
3. key until program No. suitable for le to be used is displayed.
4. When changing the set vaIue of speed, torque, gear ratio or a uto reverse, press the SELECT key, select the setting item and press the+/- key to change.
16
5.5 Operation
If you press the ON/OFF key briey, the motor handpiece starts. If you re- press the
key, it stops. (Alternate operation) If you hold down the ON/OFF key for more than one second, the motor handpiece starts whiIe the key is pressed. If you release the key it stops. (Intermittent operation)
Auto Reverse Function Auto reverse setting can be changed by pressing the SELECT key for one second or
longer while the motor handpiece stops. As the auto reverse mark ashes while the
SELECT key is kept pressed, press the +/- key to adjust it.
: Auto Reverse : Auto Stop : Auto Reverse OFF : REVERSE : RECIP ROCATING
The alarm wilI sound when it becomes a value of half of a torque limit set up during motor hand piece rotation, and the alarm will change near the torque limit value. Motor will turn to other mode, which according to the value of the load.
Auto Reverse
When it continues applying a Ioad and it exceeds the torque limit value, it will
automatically turn to AUTO REVERSE mode, the le wit I rotate reversely.
Load lower than the set torque limit value
Load higher than the set torque limit value, anti-clockwise
Reverse rotation when load continues, stop and return to clockwise rotation when load is moved.
17
Load lower than the set torque limit value
Load lower than the set torque limit value
anticIockwise rotation
Load higher than the set torque limit value, anti-clockwise
Load higher than the set torque Iimit value, handpiece stopped
Stop, when the reverse
nished
Auto Stop
The motor hand piece starts in reverse. When a load is removed, the motor handpiece stops. If you want (forward-rotate) again, re-press the ON/OFF Key or re-step.
Auto Reverse Off
The motor handpiece stops without reverse rotation. If you want it to rotate (forward­rotate) again, re-press the ON/OFF Key or re-step.
Reverse
In the genera I condition, when the Endo Motor Rotate in rounter- clockwise motion wiII
be claim as reverse motion. the traditionaI rotary les withdrawing from the canal by
reverse motion. Therefore, beep sound wiII occur when using this function.
Notice
There will have no TORQUE CONTRAL when the motor in rounter-clockwise rotation.
18
Reciprocating
In the genera I condition, when the Endo Motor Rotate in rounter- clockwise motion wiII
be claim as reverse motion. the traditionaI rotary les withdrawing from the canal by
reverse motion. Therefore, beep sound wiII occur when using this function.
Load lower than the set torque Limit value, anticlockwise and Clockwise in turns (reciprocating)
load higher than the set torque limit value, clockwise rotation with ) alarm warning sound
Notice
Under the REC I P ROCAT I NG mode, just only 16:1 or 20: 1 is supported. The setting must according to the actual contra angle handpiece.
Speed, Torque can not be set or display
Caution
This function is desgined specially for the REC IP ROCAT I NG ROTARY FILES, the cutting side is on the anticlockwise motion,
the clockwise motion is for le to get out. If the load higher than the formatted torque (torque cannot be set), the motor will rotate clockwise continously to let the le get out
the canal. For next time use, only need to press ON/OFF key
19
Notice
The program cannot be memorize while the motor handpiece is in motion
The program is not memorized unless the PROGRAM key is held down for more
than one second.
If the program number is changed by the PROGRAM key, the initially memorized
preset values will be replaced.
5.6 After Use
When the treatment is completed, return the motor handpiece to the handpiece stand and hold down the POWER key for more than one second to turn off the power.
Auto Power Off If the key is not operated or approximately 10 minutes passes during rotation with no-load, the power is automatically turned OFF to save energy and prevent inn proper operation. However, if load is applied during rotation of the motor hand piece, the power is not turned OFF even during.
Last Memory The last setting used before the power is turned OFF is memorized. The setting when
the power is turned ON next is as that immediately used before the power has been turned OFF.
6. Convenient Function
6.1 Program
You can change any preset values and have them memorized to your desired settings
(rotation speed, torque limit value, gear ratio and auto reverse mode)
1. Press the PROGRAM key until it turns to the program number which you want to
have memorized (up to 5 programs)
2. Adjust the rotation speed, torque limit value, gear ratio and auto reverse mode by +/­key according to your needs.
3. Hold down the PROGRAM key for more than one second. When the alarm sounds
the settings have been memorized.
20
6.2 Initialization of Program (Factory-set conguration)
The program can be returned to the original state congured at the time of factory
shipment, if setup becomes confusing.
1. Turn off the power, when the power is on.
2. Press PROGRAM, and then press the POWER key for more than one second. Release the keys when the LCD panel indicates "RES”. The machine on hold.
3. Re-press the PROGRAM key more than 1 second, lease the key when display “FIN”.
4. after the long beep, reset success.
Caution
If this function is used, all programs will disappear and return to the originally set values. Record the present program details if required before you carry out this operation.
6.3 Calibration
This function is to decrease uctuation in the rotation speed of the motor handpiece and
the difference in torque by the contra angle.
1. Lubricate the contra angle head. (Refer to the “7 (1) Lubricating Contra Angle.")
2. Turn off the power. (If the motor power is on)
3. Press PROGRAM, and press the POWER key for more than one second. Release the keys when the LCD panel indicates “RES ” And then press “+” key three times.
4. The LCD panel displays “CAL" with the alarm sound. Attach the lubricated contra angle to the motor.
5. Put the handpiece into the charger, and then press ON/OFF key.
6. If the motor handpiece begins to rotate, leave it as it is until it stops.
7. This process ends, if the rotation stops and the LCD panel display returns to its original state.
8. If you want to stop this process, turn off the power key for more than 1 second.
21
Caution
This function does not work unless remaining battery capacity is sufcient.
Perform calibration after cleaning the contra angle. Residuals will affect the
calibration.
Do not attack any rotary les or other when calibration.
Do not shake the motor when calibration.
This function can achieve the perfectly match between the handpiece with contra
angle. However cannot match to more than 1 different contra angle.
6.4 Alarm sound setting
You can turn ON/OFF the alarm (load alarm) which sounds near the torque limit value when the motor handpiece rotates.
1. Turn off the power
2. Press the S key, and then keep pressing the POWER key for one second or longer. Release the keys when The alarm mark and ON or OFF on the LCD panel, and the sound volume changes.
3. Press the SELECT key to select OFF, MIDDIUM or MAX for alarm sound.
4. Press POWER key to make the set memorized, or the motor will re- start after 5 seconds, if there is no operation.
5. Display on the LCD returns to the original state if no operation is performed for a while.
22
7. Cleaning Maintenance
Insert the spray nozzle into the contra
angle head(insert to the part connect to the main unit)
Insert the lubricants spray nozzle to the provided spray nozzle loop, inject
the lubricants for 2-3 seconds till the outlet liquid from the head of the contra angle clean
7.1 Lubricatinq Contra AngIe
Use qualied SPRAY to lubricate or sterilization. " Lubricate the contra angle only.
“Apply SPRAY/SPRAY Plus after each use and/or before each calibration and autoclaving
7.2 Cleaning the Motor Handpiece
Wipe it off with a cotton cloth moistened with alcohol.
7.3 Cleaning charger
Wipe it off with a cotton cloth moistened with alcohol.
Warning
Do not lubricate the motor handpiece or the charger.
Before mounting the lubricated contra angle to the motor handpiece, wipe off
excess oil. Stand it on its end or lean it in the proper position for gravity draining. Mount it after excess oil has been completely drained.
Hold the contra angle securely to prevent it from ying off by the pressure of the
spray.
Hold the SPRAY (option) upright.
When cleaning the motor headpiece, do not use any solvent such as benzene
and thinner.
23
8. Sterilization
Sterilize the contra angle only.
For the sterilization method, we recommend the autoclaving.
Autoclave sterilization required after each patient as noted be low:
a) Brush the dirt off the surface of the contra angle, and wipe it off with a cotton cloth moistened with surgical spirit. Do not use a metal brush. b) Lubricate the head with the SPRAY/SPRAY Plus. (Refer to the “7 (1) Lubricating Contra Angle.”) c) Insert the head into an autoclave pouch and seal it. d) Autoclavable up to a max. 135°C (275° F). ex. Autoclave for 20 min. at 121° C (250° F), or 15 min. at 132° C (270° F). e) Keep the hand piece in the autoclave pouch to keep it clean until you use it.
* Sterilization at 12 1°C for more than 15 minutes is recommended by EN 13060 or
EN554.
Caution
Do not heat sterilize the motor handpiece.
Do not wipe with not clean or immerse in high acid water or sterilizing solutions
24
Caution
Do not open any part other than the battery cover.
Use only batteries as specied by UNITED DENTAL.
Designated Battery: 14500 cylinder shaper LI- ION battery, nominal 3. 7 V
Charging with these battery may cause damage
Once found the battery uid leak, please stop using at once.
Do not use any wet cloth, alcohol or other chemistry to wipe the surface of the battery.
Do not work with wet hands. Failure to do so may result in rust formation on battery, or cause the re.
Should the battery uid leak and get into your eyes, immediately wash with clean
water and ask for medical care.
9. Changing Batteries
The motor handpiece uses rechargeable battery. They can be recharged more than 300 times, depending on the use conditions of the motor handpiece.
“MEMORY EFFECT” is not applicable in li-ion battery. Recharge can be conduct
anytime.
300 times is the minimum workable recharging times, which will not affect battery
life
Elements affect battery life: a) Not in use for a long term. We recommend at least recharge 1 time a month. b) Do not recharge the battery, when the room temperature over 30 centigrade.
Change battery
1. Remove the battery cover by sliding it toward the arrow
2. Pull out the battery a little part, and then separate the cable from handpiece by hand carefully.
3. Replace with new battery
4. Close the battery cover
Slide the cover from the button to the top with a little force by the linger.
25
10. Error Code
11. Troubleshooting
Error ode Error Cause Check Remedy
When working
E-0
Self-Check error
Malfunction of circuit Contact your dealer
E-1 Overcurrent
The motor handpiece is locked. (auto reverse mode)
Remove payload
E-2 Overvoltage Malfunction of circuit Contact your dealer
When Charging
E-3
Charger failure
Malfunction of circuit Contact your dealer
E-4
Low voltage of battery
Battery out of service Chang Battery
E-5
High voltage of battery
Battery out of service Chang Battery
E-8
Low voltage of adaptor
Battery out of service Change adaptor
E-9
High voltage of adaptor
Battery out of service Change adaptor
When calibration
E-7
Calibration failure
Motor damaged or handpiece blocked
Replace the motor handpiece or motor
If something goes wrong, please check the following again before consulting your dealer. If none of these is applicable or the trouble is not remedied even after action has been taken, a failure of this product is suspected. Contact your dealer, do not
eld-repairable.
26
Problem Cause Solution
The power is unable to turn on.
Battery has run out. Recharge the battery.
No battery inserted. Insert battery.
The internal fuse has burnt. Contact your dealer.
Charge failure
Battery has been over used Replace with new battery.
The power cord plug is not inserted into the charger base well
plug the power cord or adaptor properly
The handpiece is not correctly match to the charger base
Reload the handpiece to get a good match
There is some residual Contamination on the charger base
remove the residual contamination
An error code is displayed. Refer to No. 10 Error Code.
Nothing is displayed on the panel of the handpiece even when it is loaded to the charger
Contact your dealer.
The charger base is broken
Charge the motor by cable, then contact your dealer
The motor hand piece does not rotate.
The contra angle has been blocked
Clean or reset the contra angle
The alarm sounds when the motor is working
The contra angle head has jammed up.
Contra rotation failure
Set to Auto Reverse mode, take off contra angle, there is NO click sound when working
Contact your dealer.
Set to Auto Reverse mode, take off contra angle, there is click sound
Clean the contra angle
The alarm sounds when the motor is working.
There is some residual contamination on the rotating shaft of the contra angle
Clean the contra angle
27
12. Warranty
13. Disposing Product
14. Electromaonetic
Manufacturer warrants its products to the original purchaser against defects in material and workmanship under normal practices of installation, use and servicing. Batteries etc., are disposable components, and are not covered by this warranty. Manufacture's information can be found in warranty card.
Please consuIt with the dealer from whom you purchased regarding waste disposal.
Please use the spare part specied by the manufacturer, unspecied spare
part such as adapter and cord may cause the system radiation increased or the interference reduced.
The used Li-ion batteries are recyclable, but their disposal may sometimes not be permitted by the respective country. Return them to your dealer.
The appliance is intended for use in the electromagnetic environment specied
below. The user of the device shouId assure that it is used in such an environment.
28
Guidance and manufacturer’s declaration — electromagnetic emissions
The EndoPilot is intended for use in the electromagnetic environment specied below.
The customer or the user of the EndoPilot shouId assure that is used in such an environ
-
ment.
Emission test Conformity
Emission test Conformity Electromagnetic Environment
- guidance
RF Emissions CISPR 1 Group 1
The appliance use RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions Cl SPR 1 Class B The EndoPilot is suitable for
use in at I establishments, including domestic establish
­ments and those directly connected to the public low­voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC61000-3-2
Class A
Voltage IIuctuations/
icker emissions
IEC 61000-3-3
Conform s
29
Guidance and manufacturer’s declaration — electromagnetic emissions
The EndoPilot is intended for use in the electromagnetic environment specied below. The
customer or the user of the EndoPilot should assure that is used in such an environment.
Immunity test IEC60601 test level Compliance Level
Electromagnetic en
-
vironment - guide
Electrostatic discharge(ESD) EN 61000-4-2
±6kV contact ± 8kV air
± 6kV contact ± 8kV air
Floors should be wood, concrete or
ceramic tile. If oors
are covered with synthetic material, the relative humidity shouId be at least 30%.
Electrical fast transient/burst, IEC 61000-4-4
± 2 kV for power supply lines ± 1 kV for input/ Output lines
± 2 kV for power supply lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
±1 kV line to line ±2 kV line to erath
±1 kV line to line
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage varia
­tions on power supply input lines IEC 61000-4-11
< 5 % UT ( > 95 % dip in UT) For 0.5 cycles 40% UT ( 60 % dip in UT) for 5 cycles < 5 % UT 70 % UT ( 30 % dip in UT) for 25 cycles < 5 % UT < 5 % UT ( > 95 % dip in UT) for 5 s
< 5 % UT (> 95 % dip in UT 400/UT ( 60 % dip in UT) for 5 cycles < 5 % UT 70 % UT ( 30 % dip in UT) for 25 cycles < 5 % UT < 5 % UT ( > 95 % dip in UT for 5s
Mains power quality should be that of a typical commercial or hospital environment. If the user of the EndoPilot requires continued operation during power mains interruptions, it is recommended that the EndoPilot be powered from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz) magnetic eld IEC
61000-4-8
3 A/m 3 A/m
Power frequency
magnetic elds
should be at levels characteristic of a typical location in a typical commercial or hospital environ
-
ment.
NOTE: UT is the ac. mains voltage prior to application of the test level.
30
Guidance and manufacturer’s declaration — electromagnetic emissions
The EndoPilot is intended for use in the electromagnetic customer or the user of the EndoPilot
shouId assure that environment specied below. The is used in such an environment.
Immunity test IEC60601 test level Compliance Level Electromagnetic environment - guide
Conducted RF IEC 61000-4-6 Radiated 3RF IEC 61000-4-
3 Vrms
150 kHz to
3V/m
80 MHz to
3 Vrms 3V/m
Portable and mobile RF communica
­tions equipment should be used no closer to any part of the EndoPilot, including cables, than the recommend
­ed separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance
80 MHz 800 MHz 800 MHz~2.5 G Hz
Where P is the maxi mum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommend
­ed separation distance in meters (m).
Field strengths from xed RF transmit
­ters as determined by an electromag­netic site survey, shouId be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 : At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 : These guidelines may not apply in at I situations. Electromagnetic propagation is af
-
fected by absorption and reection from structures objects and people.
a) Field strengths from xed transmitters, such as base stations for radio (celIular/ cord less)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broad
-
cast can not be predicted theoretically with accuracy. To assess the electromagnetic environ-
ment due to xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the EndoPilot is used exceeds the applicable
RF compliance level above, the EndoPilot should be observed to verify normal operation. If ab
­normal performance is observed, additional measures may be necessary, such as re-orienting or relocating the EndoPilot .
b) Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
31
Guidance and manufacturer’s declaration - electromagnetic emissions
The EndoPilot is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the EndoPilot can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF according to the maxi mum output power of the communications equipment.
Rated maximum output power of transmitter Watts [W]
Separation distance according to frequency of transmitter (in meters) Meters [m]
150 kHz 80 Mhz 80 MHz 800 Mhz 800 MHz 2. 5 GHz
0.01 0. 12 0. 12 0.23
0.1 0.38 0.38 0.23
1 1.2 1.2 2.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separa
­tion distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts(W) according to the transmitter manufacturer.
NOTE 1 : At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies. NOTE 2 : These guidelines may not apply in at I situations. Electromagnetic propagation
is affected by absorption and reection from structures, objects, and people.
1. The main engine of Cordless Endodontic Micromotor (including Motor Handpiece and Contra Angle Handpiece) enjoy one year free warranty.
2. The following circumstance does not belong to the scope of free warranty: a) Using the product did not follow the matters needing attentions in user`s manual; b) Disassembling the product by yourself; c) Altering the invoice or without the invoice.
3. Fill up the following information, then send it back to us with our products.
User’s Name:
Telephone Number:
Address:
Trouble Description:
(The information such as: When, Where and How it happened. How many times)
Warranty Card
Printed in Canada
ENI003R2
Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner licensed by the law of the states in which he or she practices to use or order the use of this device. Beyes Dental Canada Inc is not responsible for any typographical errors. *All brands are holders of their respective trademarks.
!
0197
Beyes Dental Canada Inc.
23-595 Middleeld Road
Toronto, Ontario, M1V 3S2 Canada
Tel: 1-855-603-1888 Fax: 1-855-720-1228 Email: info@beyes.ca Web: www.beyes.ca
Lotus NL B.V.
Koningin Julianaplein 10, Ie Verd, 2595AA, The Hague, Netherlands Tel: +31645171879 (English) +31626669008 (Dutch)
Loading...