Besantek BST-MG08, BST-MG09 Manual Instruction

BST-MG08/09 Multi-gas Detecting Alarm
Manual Instruction
1. Application: This product is designed to ensure sa fety for users who work in dangerous
places. It’s used for detecting CO, O2, H2S, CH4 simultaneously and alarm users by light,
sound and vibration. This device is portable, anti-explosion and intrinsically safe.
2.
Features:
Hydrogen Sulfide(H2S). Contact manufacturer f
or other gas combinations.
2. It will alarm users by sound, vibration, and light with different intensity if actual
concentration is higher or lower than preset alarm points.
3.
Operation:
1. 4 buttons included.
On/Off button; OK button; Up button; Down button
2. Long press
On/Off button to activate the device. It will show below information in turn:
A. Product brand
B. Low Alarm [L]
C. High Alarm [H]
D. STEL Alarm [S]
E. TWA Alarm [T]
F. Preset Calibration Value
G. Product number and next calibrate date, 30 seconds countdown
H. Auto Zero option
3. In auto zero option, after seeing “Adjust Zero?”, long press “OK” button to zero all
sensors, short press “
” button to exit. When the device is zeroed in unfresh air, the system will zero itself when the device is back to fresh air, n o need to worry about the
accuracy.
4. In main interface, it shows 4 alarm icons on top area: [L], [H], [S], [T], if there is any alarm,
the corresponding icon will turn red. Short press “” to switch among STEL, TWA, PEAK,
and MIN. No operation for 30 seconds will bring it back to main interface.
5. Long press “OK” button to enter function interface.
LOGO
Main Interface
Function interface
6. In f unction interface, short press “” or “” to select function, then short press “OK”
button to enter corresponding function.
7. In zero interface, short press “ ” or “ ” to select gas, long press “OK” button, the
background of selected gas will turn red, long press “OK” button again to c o mplete the
operation with display “set OK”. If there is no “set OK” showed, it means operation is
failed. Short press “
” to exit. Notice: Use 100%vol nitrogen gas to zero oxygen sensor, for other sensors, it can be done in
fresh air.
8. In alarm interface, short press “” or “ ” to select among “Set Low Alarm”, “Set High
Alarm”, “Set STEL Alarm”, and “Set TWA Alarm”. Short press “OK” bu tton to enter each setting “Alarm[L]”, “Alarm[H]”, “STEL[S]”, “TWA[T]”, short press “” or “” to set the
value, long press “OK” button to complete the operation with displ ay “set OK”. Short
press “
” to exit.
9. In time setting interface, short press “OK” button to select “Year”, “Month”, “Day”,
“Hour”, “Min”, then press “” or “” to adjust the value. Long press “OK” button to
complete this operation with display of “Set OK”. Short press “
” to exit.
10. In calibration setting interface, short press “” or “” to select between“Set Value”
and “Common Calibration”
A. “Set Value”: user should input available calibration gas concentration here, short
press “OK” button to select among gases and short press “” or “” to adjust gas
values, long press “OK” button to complete this setting with display of “Set OK”.
Zero Interface
Alarm Set
Low Alarm setting
Time Setting
Calibration Setting
Common Calibration
Important: user should make this setting before following operations. B. In “Co mmon Calibration” interface, short press “” or “” to select gas, long press
“OK” button, the background of selected gas will turn red, connect calibration gas
and after the value is stable, long press “OK” button to complete this operation with
display of “Set OK”. Short press “
” to exit.
Notice:
a. Before calibration, a ctivate this device for at least 6 minutes to allow it run stably
b. When the device is alarming for low power, stop calibration and charge it right away
c. Calibration gas must be standard gas used for calibration purpose
d. Zero sensors firstly before calibration
e. Make sure gas flow is stable before connecting calibration gas, calibration gas flow for CH4,
CO, O2 is 200ml/min, 300ml/min for H2S.
f. Suggested calibration gases are
as following:
CH4 1.00%vol, CO 500ppm, H2S 50ppm, oxygen can be calibrated in fresh air
11. In setting interface, short press “ ” or “ ” to select among “CH4”, “LEL”, “PPM”,
“MG/m³ ”, “Allow Calibration”, “” means item is selected. Long press “OK” button to
select a nd confirm. Short press “
” to exit.
Important: User can’t enter calibration i nterface without choosing “Allow Calibration”
4.
Technical Parameter:
Technical
Specification
Parameter
Measured gas Methane(CH4), Oxygen(O2),
Carbon Monoxide(CO), Hydrogen Sulfide(H2S)
Methane (CH4)
Measurement range (0.00~5.00)%vol
Measurement error
(0.00~1.00)%vol: ±0.10%CH4 (1.00~3.00)%vol: True value±10% (3.00-5.00)%vol: ±0.30%CH4
Respond time
20s (T90)
Resolution
0.01%vol Alarm point
1.0
0%vol (Adjustable for full
range)
Oxygen (O2)
Measurement range (0.0~30.0)%vol
Measurement error
(0.0~5.0)%vol: O2:±0.5%vol
(>5.0~30.0)%vol: O2:0.9%vol
Respond time
35s (T90)
Resolution
0.1%vol Alarm point
18.0%vol(Adjustable for full range)
25.0%vol(Adjustable for full range)
Carbon Monoxide (CO)
Measurement range (0~1000)ppm
Measurement error
(0~20)ppm:±2ppm
(>20~100)ppm:±4ppm