Bender LIM2000plus LIM2000-1, LIM2000plus LIM2000-1CB, LIM2000plus LIM2000-3CB Operating Manual

Operating Manual LIM2000plus™
Line Isolation Monitor LIM2000-1 LIM2000-1CB LIM2000-3CB
TGH 1337 Edition 01/06
Table of Contents
Operating manual ...................................
1. Safety Information ............................................... 5
1.1 Intended use ....................................................... 5
1.2 Warranty and liability ........................................... 5
1.3 Personnel ............................................................ 6
1.4 About the operating manual ................................ 6
1.5 Hazards when handling the Line Isolation Monitor
LIM2000-1, LIM2000-1CB and LIM2000-3CB ..... 6
1.6 Inspection, transport and storage ........................ 7
1.7 Important ............................................................. 7
1.8 Explanation of symbols and notes ...................... 8
1.9 Directives for installation ..................................... 9
2. General Information .......................................... 11
2.1 The fundamental functions ................................ 11
2.2 Display .............................................................. 12
2.3 Menu ................................................................. 13
2.4 Error Code Listing ............................................. 15
2.5 External Connections ........................................ 16
ISOTROL / BENDER Medical Products 700 Fox Chase Coatesville, PA 19320
Telephone: 1-610-383-9200 Fax: 1-610-383-7100 Tollfree: 1-800-833-6834
Rights to modification reserved
2
2.5.1 Cable Assembly (CP-LIM2000).......................... 17
2.5.2 Cable Assembly (CP-LIM2000CB) .................... 18
2.6 Accessories ....................................................... 19
2.6.1 Remote Indicator ............................................... 19
2.6.2 Digital Remote Indicator .................................... 20
3. Technical Description ........................................ 21
3.1 Technical Data .................................................. 22
4. Information Request .......................................... 23
3
1. Safety Information
1.1 Intended use The Line Isolation Monitor (LIM) calculates the total hazard
current in a single phase or three phase isolated power system. This is the maximum current that could flow through a person who made contact between an isolated conductor
and ground. Two seperate ground connections are provided for added safety
when wiring the LIM into an isolated power system. Each ground should be wired individually to the reference grounding bus. A break in either connection will cause the LIM to alarm (flashing red LED).
Any other use, or any use beyond the foregoing, is deemed to be improper. The BENDER companies shall not be liable for any loss and damaging arising therefrom.
Correct use also includes
• compliance with all instructions from the operating manual
• adherence to any inspection intervals.
As a basic principle, our “General conditions of Sale and Delivery“ shall apply. These are available to the operator not later than the time when the contract is concluded.
1.2 Warranty and liability
Warranty and liability claims in the event of injury to persons or damage to property are excluded if they can be attributed to one or more of the following causes:
• Improper use of the LIM.
• Improper assembly/fitting, commissioning, operation and
maintenance of the LIM.
• Failure to take note of the operating instructions concerning
transportation, commissioning, operation and maintenance of the LIM.
• Unauthorized structural modifications to the LIM.
• Failure to take note of the technical data.
4
5
. Safety Information
r
1. Safety Information
• Improperly performed repairs and the use of spare parts o accessories which are not recommended by the
manufacturer.
• Cases of disaster brought about by the effect of foreign bodies and force majeure.
• The assembly and installation of non-recommended
combinations of devices.
This operating manual, and in particular the safety information, must be noted by all persons who work with the LIM. In addition, it is essential to comply with the rules and regulations on accident prevention which are valid for the place of use.
1.3 Personnel
Only appropriately qualified personnel may work on the LIM. “Qualified“ means that such personnel are familiar with the
installation, commissioning and operation of the LIM, and that they have undergone training or instructions which is appropriate to the activity. The personnel must have read and understood the safety chapter and the warning information in these operating instructions.
1.4 About the operating manual
This operating manual has been compiled with the greatest possible care. Nevertheless, errors and mistakes cannot be entirely ruled out. BENDER companies assume no liability whatsoever for any injury to persons or damage to property which may be sustained as a result of faults or errors in these operating
instructions.
1.5 Hazards when handling the Line Isolation Monitor
LIM2000-1, LIM2000-1CB and LIM2000-3CB The LIM is constructed according to state of the art and the
recognised safety engineering rules. Nevertheless, when it is
6
being used, hazards may occur to the life and limb of the user or of third parties, or there may be adverse effects on the LIM or on other
valuable property. The LIM must only be used
• for the purpose for which it is intended
• when it is in perfect technical condition as far as safety is
concerned Any faults which may impair safety must be eliminated immediately.
Impermissible modifications and the use of spare parts and additional devices which are not sold or recommended by the manufacturer of the LIM may cause fire, electric shocks and
injuries. Unauthorized persons must not have access to or contact
with the LIM.
Warning signs must always be easily legible. Damaged or illegible signs must be replaced immediately.
1.6 Inspection, transportation and storage Inspect the dispatch packaging and the equipment packaging for
damage, and compare the contents of the package with the delivery documents. In the event of damage during transportation, please notify the BENDER company immediately.
The LIM must only be stored in rooms where they are protected against dust and moisture, and spraying or dripping water, and where the indicated storage temperatures are maintained.
1.7 Important Please check for correct system and supply voltage ! When insulation and voltage tests are to be carried out, the device
must be isolated from the system for the test period. In order to check the proper connection of the LIM, it is
recommended to carry out a functional test, before starting the LIM. Please check whether the basic setting of the devices complies with
the system requirements. Children or the public must not have access to the LIM.