Bender ISOMETER IRDH575 User Manual

ISOMETER® IRDH575

Insulation monitoring device for IT AC and DC sys­tems with integrated test generator and controller for EDS46..., EDS47... and EDS49... systems Software version: D0185 V1.80
IRDH575_D00089_05_M_XXEN/01.2020
Manual EN
Table of Contents
1. Important information .................................................................................... 8
1.1 How to use this manual ................................................................................. 8
1.2 Technical support: service and support ................................................... 9
1.2.1 First level support ............................................................................................. 9
1.2.2 Repair service ..................................................................................................... 9
1.2.3 Field service ..................................................................................................... 10
1.3 Training courses ............................................................................................. 10
1.4 Delivery conditions ....................................................................................... 10
1.5 Inspection, transport and storage ........................................................... 11
1.6 Warranty and liability ................................................................................... 11
1.7 Disposal ............................................................................................................ 12
2. Safety instructions ......................................................................................... 13
2.1 General safety instructions ........................................................................ 13
2.2 Work activities on electrical installations ............................................. 13
2.3 Device-specific safety information ......................................................... 14
2.4 Use for the intended purpose ................................................................... 16
3. Function ........................................................................................................... 17
3.1 General characteristics ................................................................................ 17
3.2 Essential functions of the ISOMETER® .................................................... 17
3.3 Essential functions of the Insulation fault location system (EDS) 18
3.4 Product description ...................................................................................... 18
3.5 Function ............................................................................................................ 19
3.5.1 Current output for external measuring instrument ......................... 20
3.5.2 Real-time clock ............................................................................................... 20
3.5.3 Insulation fault location .............................................................................. 20
3.5.4 Interconnected systems ............................................................................. 21
IRDH575_D00089_05_M_XXEN/01.2020
Table of Contents
3.5.5 Function input F1/F2 for connection or disconnection of IT systems
being monitored ........................................................................................... 22
3.5.6 ISOnet function for central control of the insulation monitoring
process when several IRDH575 are interconnected in IT systems ....
24
3.5.7 Self test .............................................................................................................. 26
3.5.8 Relay K3: device fault alarm and EDS common message ............... 28
4. Connection ..................................................................................................... 29
5. Commissioning flow chart ........................................................................... 33
5.1 Commissioning of the ISOMETER® function range (1) ..................... 33
5.2 Commissioning of the insulation fault location function (EDS) (1) ...
35
6. Operation and setting .................................................................................. 38
6.1 Operating features and displays IRDH575 ........................................... 38
6.1.1 Display in case of active EDS and detected fault ............................... 40
6.1.2 Display in the menu mode ......................................................................... 41
6.1.3 Function keys .................................................................................................. 41
6.2 Menu structure and menu mode ............................................................ 44
6.2.1 Diagram menu structure ............................................................................ 46
6.3 HISTORY INFO menu .................................................................................... 48
6.3.1 Diagram HISTORY INFO ............................................................................... 49
6.4 ISO SETUP menu: Setting of the basic ISOMETER® functions ........ 50
6.4.1 Response values Alarm 1 and Alarm 2 ................................................... 50
6.4.2 Starting the EDS system via the response values ALARM 1 and
ALARM 2 ........................................................................................................... 50
6.4.3 Operating principle of the alarm relays ................................................ 51
6.4.4 Memory setting (on/off) .............................................................................. 53
6.4.5 Current output for external measuring instruments ........................ 54
6.5 ISO ADVANCED menu: Setting of the extended functions ............ 55
6.5.1 External coupling device (AGH: no = factory setting) ..................... 55
6.5.2 Selecting the system leakage capacitance range .............................. 55
IRDH575_D00089_05_M_XXEN/01.2020
Table of Contents
6.5.3 Changing the measuring principle from AMP to DC
(Measure: AMP) .............................................................................................. 55
6.5.4 Setting the repetition time for automatic self tests
(Autotest: 24h) ................................................................................................ 55
6.5.5 Setting the real-time clock (Clock) .......................................................... 55
6.5.6 Setting the date (Date) ................................................................................ 56
6.5.7 Specifying the starting time of the automatic self test (Test) ....... 56
6.5.8 Diagram ISO ADVANCED ............................................................................ 57
6.6 EDS-SETUP menu: Settings for fault location ...................................... 58
6.6.1 EDS auto / on / off / pos / 1cycle ............................................................. 58
6.6.2 Diagram EDS-SETUP ..................................................................................... 59
6.6.3 System DC / 1 AC / 3 AC .............................................................................. 60
6.6.4 maxPuls 1 / 2.5 / 10 / 25 / 50 mA: ............................................................ 60
6.6.5 K3 alarm: ON ................................................................................................... 63
6.7 EDS460/490 menu ........................................................................................ 63
6.7.1 General .............................................................................................................. 63
6.7.2 Channel ............................................................................................................. 64
6.7.3 Relay ................................................................................................................... 67
6.7.4 EDS Test ............................................................................................................ 68
6.7.5 EDS Reset .......................................................................................................... 68
6.7.6 Diagram EDS46… /49… with Relay, EDS-Test and EDS-Reset ..... 69
6.8 EDS 470 menu ................................................................................................ 70
6.8.1 EDS Monitor .................................................................................................... 70
6.8.2 EDS Test ............................................................................................................ 70
6.8.3 EDS Reset .......................................................................................................... 70
6.8.4 Details about the menu points Relay, Memory and n-peak .......... 71
6.8.5 Diagram EDS 470 ........................................................................................... 72
6.8.6 Relay ................................................................................................................... 73
6.8.7 Memory ............................................................................................................ 73
6.8.8 CT Setup: .......................................................................................................... 73
6.8.9 n-peak: ............................................................................................................... 74
6.9 COM SETUP menu: Setting the BMS interface .................................... 75
IRDH575_D00089_05_M_XXEN/01.2020
Table of Contents
6.9.1 Bus address (Addr: ) ................................................................................... 75
6.9.2 ISO Monitor ...................................................................................................... 75
6.9.3 ISOnet ................................................................................................................ 76
6.9.4 Diagram COM SETUP ................................................................................... 76
6.10 PASSWORD menu ......................................................................................... 77
6.10.1 Activating and setting the password ..................................................... 77
6.10.2 Diagram PASSWORD .................................................................................... 77
6.11 Menu LANGUAGE ......................................................................................... 78
6.11.1 Setting the national language .................................................................. 78
6.11.2 Diagram Language ....................................................................................... 78
6.12 Menu SERVICE ................................................................................................. 79
6.13 Parameterization via Internet ................................................................... 79
7. Serial interfaces ............................................................................................. 80
7.1 RS-485 interface with BMS protocol ....................................................... 80
7.2 Topology RS-485 network .......................................................................... 81
7.2.1 Correct arrangement ................................................................................... 81
7.2.2 Wrong arrangement ..................................................................................... 81
7.2.3 Wiring ................................................................................................................ 81
7.3 BMS protocol (BMS) ...................................................................................... 82
7.3.1 BMS Master ...................................................................................................... 82
7.3.2 BMS-Slave ......................................................................................................... 83
7.3.3 BMS operation in the Standby mode ..................................................... 84
7.3.4 Combination with EDS46… devices ...................................................... 86
7.3.5 Commissioning of an RS-485 network with BMS protocol ............ 87
8. Factory settings ............................................................................................. 89
IRDH575_D00089_05_M_XXEN/01.2020
Table of Contents
9. Technical data IRDH575 ................................................................................ 91
9.1 Data in tabular form ..................................................................................... 91
9.2 Standards, approvals and certifications ................................................ 95
9.3 Characteristic curves .................................................................................... 96
9.3.1 Characteristic curves of the ISOMETER® ................................................ 96
9.3.2 Characteristic curves of the insulation fault locators EDS46… /
EDS49… .......................................................................................................... 100
9.3.3 Characteristic curves for the insulation fault location system
EDS470 ............................................................................................................ 102
9.4 Ordering details ........................................................................................... 111
9.4.1 Standard version ......................................................................................... 111
9.4.2 Protection against dust and moisture ................................................. 112
9.4.3 Adaptor for rail mounting ........................................................................ 112
9.4.4 Measuring instruments ............................................................................. 112
IRDH575_D00089_05_M_XXEN/01.2020

1. Important information

1.1 How to use this manual

This manual is intended for qualified personnel working in electrical engineering and electronics!
Always keep this manual within easy reach for future reference.
To make it easier for you to understand and revisit certain sections in this man­ual, we have used symbols to identify important instructions and information. The meaning of these symbols is explained below:
This signal word indicates that there is a high risk of danger that will result in electrocution or serious injury if not avoided.
This signal word indicates a medium risk of danger that can lead to death or serious injury if not avoided.
This signal word indicates a medium risk of danger that can lead to death or serious injury if not avoided.
This symbol denotes information intended to assist the user in making optimum use of the product.
8
IRDH575_D00089_05_M_XXEN/01.2020
Important information

1.2 Technical support: service and support

For commissioning and troubleshooting Bender offers you:

1.2.1 First level support

Technical support by phone or e-mail for all Bender products
 Questions concerning specific customer applications  Commissioning  Troubleshooting
Telephone: +49 6401 807-760* Fax: +49 6401 807-259
In Germany only: 0700BenderHelp (Tel. and Fax) E-mail: support@bender-service.de

1.2.2 Repair service

Repair, calibration, update and replacement service for Bender products
 Repairing, calibrating, testing and analysing Bender products  Hardware and software update for Bender devices  Delivery of replacement devices in the event of faulty or incorrectly
delivered Bender devices
 Extended guarantee for Bender devices, which includes an in-house
repair service or replacement devices at no extra cost
Telephone: +49 6401 807-780** (technical issues)
+49 6401 807-784**, -785** (sales)
Fax: +49 6401 807-789 E-mail: repair@bender-service.de
Please send the devices for repair to the following address:
Bender GmbH, Repair-Service, Londorfer Str. 65, 35305 Grünberg
IRDH575_D00089_05_M_XXEN/01.2020
9
Important information

1.2.3 Field service

On-site service for all Bender products
 Commissioning, configuring, maintenance, troubleshooting of Bender
products
 Analysis of the electrical installation in the building (power quality test,
EMC test, thermography)
 Training courses for customers
Telephone: +49 6401 807-752**, -762 **(technical issues)
+49 6401 807-753** (sales)
Fax: +49 6401 807-759 E-mail: fieldservice@bender-service.de Internet: www.bender-de.com
*Available from 7.00 a.m. to 8.00 p.m. 365 days a year (CET/UTC+1) **Mo-Thu 7.00 a.m. - 8.00 p.m., Fr 7.00 a.m. - 13.00 p.m

1.3 Training courses

Bender is happy to provide training regarding the use of test equipment. The dates of training courses and workshops can be found on the Internet at www.bender-de.com -> Know-how -> Seminars.

1.4 Delivery conditions

Bender sale and delivery conditions apply. For software products the "Softwareklausel zur Überlassung von Standard­Software als Teil von Lieferungen, Ergänzung und Änderung der Allgemeinen Lieferbedingungen für Erzeugnisse und Leistungen der Elektroindustrie" (software clause in respect of the licensing of standard software as part of de­liveries, modifications and changes to general delivery conditions for prod­ucts and services in the electrical industry) set out by the ZVEI (Zentralverband Elektrotechnik- und Elektronikindustrie e. V.) (German Electrical and Electron­ic Manufacturer's Association) also applies.
10
IRDH575_D00089_05_M_XXEN/01.2020
Important information
Sale and delivery conditions can be obtained from Bender in printed or elec­tronic format.

1.5 Inspection, transport and storage

Inspect the dispatch and equipment packaging for damage, and compare the contents of the package with the delivery documents. In the event of damage in transit, please contact Bender immediately. The devices must only be stored in areas where they are protected from dust, damp, and spray and dripping water, and in which the specified storage tem­peratures can be ensured.

1.6 Warranty and liability

Warranty and liability claims in the event of injury to persons or damage to property are excluded if they can be attributed to one or more of the follow­ing causes:
 Improper use of the device.  Incorrect mounting, commissioning, operation and maintenance of the
device.
 Failure to observe the instructions in this operating manual regarding
transport, commissioning, operation and maintenance of the device.
 Unauthorised changes to the device made by parties other than the
manufacturer.
 Non-observance of technical data.  Repairs carried out incorrectly and the use of replacement parts or
accessories not approved by the manufacturer.
 Catastrophes caused by external influences and force majeure.  Mounting and installation with device combinations not recom-
mended by the manufacturer.
This operating manual, especially the safety instructions, must be observed by all personnel working on the device. Furthermore, the rules and regulations that apply for accident prevention at the place of use must be observed.
IRDH575_D00089_05_M_XXEN/01.2020
11
Important information

1.7 Disposal

Abide by the national regulations and laws governing the disposal of this de­vice. Ask your supplier if you are not sure how to dispose of the old equip­ment. The directive on waste electrical and electronic equipment (WEEE directive) and the directive on the restriction of certain hazardous substances in electri­cal and electronic equipment (RoHS directive) apply in the European Commu­nity. In Germany, these policies are implemented through the "Electrical and Electronic Equipment Act" (ElektroG). According to this, the following applies:
 Electrical and electronic equipment are not part of household waste.  Batteries and accumulators are not part of household waste and must
be disposed of in accordance with the regulations.
 Old electrical and electronic equipment from users other than private
households which was introduced to the market after 13 August 2005 must be taken back by the manufacturer and disposed of properly.
For more information on the disposal of Bender devices, refer to our homepage at www.bender-de.com -> Service & support.
12
IRDH575_D00089_05_M_XXEN/01.2020

2. Safety instructions

2.1 General safety instructions

Part of the device documentation in addition to this manual is the enclosed "Safety instructions for Bender products".

2.2 Work activities on electrical installations

Only qualified personnel are permitted to carry out the work necessary to install, commission and run a device or system.
Risk of electrocution due to electric shock!
Touching live parts of the system carries the risk of:
 An electric shock  Damage to the electrical installation  Destruction of the device
Before installing and connecting the device, make sure
that the installation has been de-energised. Observe the rules for working on electrical installations.
If the device is used outside the Federal Republic of Germany, the applicable local standards and regulations must be complied with. The European stan­dard EN 50110 can be used as a guide.
IRDH575_D00089_05_M_XXEN/01.2020
13

2.3 Device-specific safety information

Children and unauthorised persons must not have access to or contact with the ISOMETER®.
Make sure that the operating voltage is correct!
Prior to insulation and voltage tests, the ISOMETER® must be disconnected from the IT system for the duration of the test. In order to check the correct connection of the device, a functional test has to be carried out before starting the system.
Make sure that the basic settings meet the requirements of the IT system.
In the event of an alarm message of the ISOMETER®, the insulation fault should be eliminated as quickly as possible.
If the ISOMETER® is installed inside a control cabinet, the insulation fault message must be audible and/or visible to attract attention.
Safety instructions
14
IRDH575_D00089_05_M_XXEN/01.2020
Safety instructions
When using ISOMETER®s in IT systems, make sure that only one active ISOMETER® is connected in each interconnected system. If IT systems are interconnected via coupling switches, make sure that ISOMETER®s not currently used are disconnected from the IT system and deactivated. IT systems coupled via diodes or capacitances may also influence the insulation monitoring process so that a central control of the different ISOMETER®s is required.
Prevent measurement errors!
When a monitored IT system contains galvanically coupled DC circuits, an insulation fault can only be detected correctly if the rectifier valves (e.g. rectifier diode, thyristors, IGBTs, frequency inverters, …) carry a minimum current of > 10 mA.
Unspecified frequency range
When connecting to an IT system with frequency components below the specified frequency range, the response times and response values may differ from the indicated technical data. However, depending on the application and the selected measurement method, continuous insulation monitoring is also possible in this frequency range. There is no influence on the insulation monitoring for IT systems with frequency components above the specified frequency range, e.g. within the range of typical switching frequencies of frequency inverters (2…20 kHz).
IRDH575_D00089_05_M_XXEN/01.2020
15
Safety instructions

2.4 Use for the intended purpose

The ISOMETER® is exclusively intended for:
 monitoring the insulation resistance of IT systems and  localization of insulation faults in combination with insulation fault
evaluators EDS4…
In order to meet the requirements of the applicable standards, customised pa­rameter settings must be made on the equipment in order to adapt it to local equipment and operating conditions. Please heed the limits of the range of application indicated in the technical data.
Any use other than that described in this manual is regarded as improper.
16
IRDH575_D00089_05_M_XXEN/01.2020

3. Function

AMP
Plus

3.1 General characteristics

 Four-line LC display  Automatic device self-test  Memory with real-time clock to store all alarm messages with date and
time stamp
 RS-485 interface (BMS protocol) for data exchange with other Bender
devices (RS-485 electrically isolated)
 Remote setting of certain parameters via the Internet (option;
FTC470XET additionally required)
 Option "W":
Improved shock and vibration resistance for use in ships, on rolling stock and in seismic environment.

3.2 Essential functions of the ISOMETER®

 ISOMETER® for IT AC systems with galvanically connected rectifiers and
for IT DC systems (isolated power)
 Automatic adaptation to the existing system leakage capacitances  measuring principle (European Patent: EP 0 654 673 B1)  Two adjustable response values in the range 1 kΩ …10 MΩ (Alarm 1/
Alarm 2)
 Connection monitoring  Internal disconnection of the ISOMETER® from the IT system to be mon-
itored (using a control signal; terminals F1/F2) , e.g. if several ISOMETER®s are interconnected.
 Current output 0(4)…20 mA (galvanically separated) in relation to the
measured insulation value.
IRDH575_D00089_05_M_XXEN/01.2020
17
Function

3.3 Essential functions of the Insulation fault location system (EDS)

 Generation of the test current necessary for selective insulation fault
location
 Indication of the insulation faults selectively localized by the EDS4…
systems
 Parameterization of EDS4… systems  Test function for EDS4… systems, incl. the connected measuring cur-
rent transformers.

3.4 Product description

The IRDH575 ISOMETER® monitors the insulation resistance of IT systems. It is suitable for universal use in 3NAC, AC/DC and DC systems. AC systems may in­clude extensive DC supplied loads, such as converters or thyristor-controlled DC drives. The device automatically adapts itself to the existing system leaka­ge capacitance. The IRDH575 ISOMETER® is fitted into an enclosure for panel mounting, 144 x 96 mm (WxH). For protection against dust and moisture, a panel seal (IP42) or a front-face transparent cover (IP65) are available as accessories, see page 112. The IRDH575 can be used in combination with a control and indicating device, PRC1470 version 2 or higher, for example, on the BMS (BMS = Bender Measu­ring Device Interface) bus.
18
IRDH575_D00089_05_M_XXEN/01.2020
Function
AMP
Plus

3.5 Function

The IRDH575 ISOMETER® is connected between the unearthed system and the protective conductor (PE). The response values and other function parameters are set via the function keys. The parameters are indicated on the LC display and are stored in a non-vola­tile memory (EEPROM) after the setting is completed. A microprocessor-controlled pulsating AC measuring voltage is superimpo­sed on the system ( measuring principle*). The measuring cycle consists of positive and negative pulses of the same amplitude. The period de­pends on the respective system leakage capacitances and the insulation resis­tance of the system to be monitored. An insulation fault between system and earth closes the measuring circuit. From the measured current value, the microprocessor calculates the insulati­on resistance which is indicated on the LC display or the external kΩ measu­ring instrument.
Prevent measurement errors!
When an AC system includes galvanically connected DC circuits, the following shall be considered: Insulation faults in DC circuits can only be monitored correctly when the rectifiers carry a continuous load of at least 510 mA.
The measuring time is determined by the system leakage capacitances, the in­sulation resistance, and the system-related interference disturbances. System leakage capacitances do not influence the measuring accuracy. If the reading is below the selected response values ALARM1/ALARM2, the as­sociated alarm relays respond and the alarm LEDs "ALARM1/2" light up and the measuring value is indicated on the LC display (in the event of DC insula­tion faults, the faulty supply line is indicated). The fault indication can be stored by bridging the terminals R1/R2 (external RESET button [NC contact]) or by setting the fault memory to "Memory:on" in the "ISO SETUP" menu. Pressing the external or internal RESET button, resets the fault message, pro­vided that the insulation resistance is at least 25 % above the preset value.
IRDH575_D00089_05_M_XXEN/01.2020
19
Function
AMP
Plus
R1/R2 bridged + Memory: on fault memory activated R1/R2 bridged + Memory: off fault memory activated R1/R2 not connected + Memory: on fault memory activated R1/R2 not connected + Memory: off fault memory deactivated
*) measuring principle "adaptive measuring pulse", a measuring principle developed by Bender.

3.5.1 Current output for external measuring instrument

The current output of IRDH575 provides 0(4)…20 mA. The current output is electrically isolated from the device electronics and the RS-485 interface. The ISO SETUP menu, on page 54, allows to switch over between 0…20 mA and 4…20 mA.

3.5.2 Real-time clock

The real-time clock serves as a time base for the historical memory and self test function. At first the correct time and date must be set in the menu "ISO ADVANCED". If time and date are not set, a "C" (clock) is flashing in the standard display. In the event of a supply voltage failure, time and dat e will be st ored for thir ty d ays. If t he 24 h se lf test is activated in the ISO ADVANCED me­nu, a special time of day can be selected for the execution of the self test in the menu "TEST: 12:00". Then a self test will be started automatically once a day exactly at the preset time. If the 1 h test has been selected, the self test is au­tomatically carried out every full hour.

3.5.3 Insulation fault location

Another function of the IRDH575 is selective insulation fault location. If the va­lue of the insulation resistance falls below the set response values ALARM 1 and ALARM 2, a certain test current is generated by the IRDH575. The maxi­mum value of the test current is determined by the maxPuls parameter, see page 60. In combination with an insulation fault evaluator EDS47…and the associated measuring current transformers connected to it, the insulation fault is selectively detected. The detected insulation fault is signalled to the
20
IRDH575_D00089_05_M_XXEN/01.2020
Function
IRDH575 via the RS-485 interface (BMS protocol) and is then indicated by an alarm LED and indicated on the display. In the Master mode (Addr. 1), this alarm message is indicated by the alarm relay K3 as common alarm. During the insulation fault location process the insulation monitoring func­tion is deactivated. If the test current falls below the value the EDS4… is capa­ble of measuring, the insulation fault location process will be stopped by IRDH575.
Risk of malfunctions due to excessive locating current on sensitive system parts!
An excessive locating current flowing be tween the IT system and earth may cause controller faults in sensitive system parts, such as the PLC or the relay. Ensure that the level of the locating current is compatible with the system to be monitored.

3.5.4 Interconnected systems

When using ISOMETER®s in IT systems care shall be taken that only one active ISOMETER® is connected in each interconnected system. If IT systems are interconnected via coupling switches, make sure that ISOMETER®s not cur­rently used are disconnected and deactivated via a control system. IT systems coupled via diodes or capacitances may also influence the insulation monito­ring process. Hence, also in this case a central control of the different ISOMETER®s is required.
IRDH575_D00089_05_M_XXEN/01.2020
21
Function
3.5.5 Function input F1/F2 for connection or disconnection of IT
systems being monitored
The ISOMETER® can be disconnected from the IT system and set to STANDBY mode with the function input F1/F2. When the input F1/F2 is bridged, the ter­minals L1/L2 are isolated from the measuring circuit via internal coupling re­lays and terminal L3 remains c onnected to the measuring circuit via a resistor of 10 MΩ. The measuring function is stopped and the message „STANDBY“ ap­pears on the display. When using software version 1.4 or higher, the latest measured insulation resistance is blanked and the value > 10 MΩ appears on the display. Furthermore, the alarm relays and alarm LEDs no longer provide alarm messages. Insulation faults already detected will be indicated by all EDS4…devices. After opening the function input F1/F2, first of all the connection to the IT sys­tem will be restored and a completely new measuring cycle for insulation mo­nitoring will be started. Using this function, selective disconnection of an IRDH575 in interconnected IT systems can be carried out via auxiliary contacts of the respective coupling switch. One coupling switch each in a line-type or ring-type arrangement can deactivate a subsequent IRDH575. This arrange­ment guarantees that only one ISOMETER® is active in each galvanically con­nected IT system. Theoretically speaking, in a ring-type arrangement with all coupling switches closed, all ISOMETER®s would be deactivated. In order to prevent this, a BMS Master (IRDH575 Adr1) monitors the condition of the function input F1/F2 of all Slave ISOMETER®s. When all Slave ISOMETER®s are in the STANDBY mode, the insulation monito­ring function of the Master ISOMETER® remains active, that means, the input F1/F2 of the Master is without function in this mode.
22
IRDH575_D00089_05_M_XXEN/01.2020
Function
BMS bus (A/B, RS485)
addr. 1 addr. 2
IT system 1
IT system 4 IT system 3
addr. 4 addr. 3
IT system 2
Example: Let as assume, in the before-mentioned ring-type arrangement, the associa­ted coupling switch of the Slave ISOMETER® 2 were open. The coupling swit­ches of the BMS Master (Addr. 1) and of the Slaves 3 and 4 were closed. In this case the ISOMETER® and EDS functions of the Master and the Slaves 3 and 4 would be deactivated. In spite of changing to the STANDBY mode, the Master status of the device with address 1 would remain. That means, if a paramete­rization is necessary, it has to be carried out via the IRDH575 with BMS address 1.
IRDH575_D00089_05_M_XXEN/01.2020
23
Function
3.5.6 ISOnet function for central control of the insulation
monitoring process when several IRDH575 are interconnected in IT systems
Up to 30 ISOMETER® can communicate with each other i n an ISOnet network. The ISOnet network can only be activated when a BMS bus is used for inter­connection. A typical method with four devices is shown as an example. The ISOnet function of all ISOMETER® in the ISOnet network must be activated in the COM SETUP „ISOnet=ON“ menu, refer to page 76.
The BMS master (BMS address 1) with activated ISOnet function controls the ISOnet slave devices via the BMS bus.
Address 1 must not be assigned to another BMS device on the respective BMS bus. When the master ISOMETER® has finished one measuring cycle, the authoriza­tion for insulation monitoring is passed on to the next higher BMS address. This authorization is returned by the slave with the highest BMS address to the BMS master after a cycle has been completed. During the insulation monitoring process, all other ISOMETER®, are in the STANDBY mode. In this way it is prevented that ISOMETER®s influence each other in interconnected IT systems. At f = 50Hz and C 12 s and then changes to the STANDBY mode. The maximum response time of the ISOnet device that made the last measurement, will be extended by the number of devices x 12s, in our example 48s. When an insulation fault is detected by an ISOMETER®, the insulation fault lo­cation process is started. The device stays in the measuring mode during this time. Only after completion of insulation fault location, the ISOnet device stops the monitoring mode and passes the authorization for insulation moni­toring on to the next device. In the STANDBY mode, the ISOnet device displays the last measured insulati­on resistance.
= 1 µF, an ISOnet device stays in the measuring mode for
e
24
IRDH575_D00089_05_M_XXEN/01.2020
Function
Each ISOnet slave checks the network for an ISOnet master. If there is no mas­ter available, the display will show the fault message “ISOnet Master?“ after approximately 1 hour. Additionally, the LED for device errors lights and the Relay K3 switches. With the ISOnet function activated, the function input F1/F2 is automatically deactivated.
All EDS devices in the IT systems being monitored must be operated with activated fault memory.
In comparison to a solution with coupling switches and function input F1/F2, the response time will be extended because the measurement is not perma­nently carried out.
BMS bus (A/B, RS485)
addr. 1 addr. 2
IT system 1
IT system 4 IT system 3
addr. 4 addr. 3
IRDH575_D00089_05_M_XXEN/01.2020
IT system 2
25
Function
This has the advantage that no auxiliary contacts of a coupling switch are re­quired. Furthermore, this solution is recommended for IT systems coupled via capacitances or diodes.

3.5.7 Self test

In order to guarantee high functional reliability, the ISOMETER® provides com­prehensive self test functions. After switching the supply voltage on, all inter­nal measuring functions, the components of the process control such as data and parameter memory as well as system and earth connections are checked using the self test functions. The progress of the self test is indicated on the display by a bar graph. Depending on the system conditions, the duration of the self test is 15… 20 seconds, then the device indicates "Test ok" for appro­ximately 2 seconds. Then the device returns to normal measuring mode and the current measuring value is displayed after the expiry of the reading time. When a fault is found the message "!Error!" appears on the display, the device fault relay K3 (31-32-34) drops, the device fault LED lights up and the respec­tive error message (see table) is indicated. If such a device fault occurs, a self test will be started every 60 seconds. If no more malfunction is detected, the fault message will automatically be deleted, the device fault LED extinguishes and the device fault relay K3 energizes again.
During operation, the self test function can be started by pressing the TEST button (internal or external) or automatically every hour or every 24 hours by selecting "ISO ADVANCED": Autotest:" in the menu. The alarm relays ALARM 1/ 2 can only switch after starting the self test function by pressing the TEST but­ton, that means if an automatic self test has been selected, the alarm relays do not switch.
26
If the on/off switching of the s upply voltage is not possible for technical reasons, a RESET of the device can be carried out by pressing the "INFO", "RESET" and "MENU" key si multaneously.
IRDH575_D00089_05_M_XXEN/01.2020
Function
Behaviour of the analogue output
Setting Manual test Automatic test
0-20 mA 20 mA
while test procedure
4-20 mA 20 mA
while test procedure
0 mA The current value depends on the insulation value
4 mA The current value depends on the insulation value
Error
message
System connection?
Connection PE?
Device error x Internal device
IRDH575_D00089_05_M_XXEN/01.2020
Meaning Steps to taken
No low-resistance connection of terminals L1, L2, L3 to the system
No low-resistance connection of the
terminals and KE to earth (PE)
error
1. Check the wiring of termi­nal L1, L2, L3 to the system
2. Press the TEST button
3. Switch the supply voltage on and off
4. Check the fuses
1. Check wiring of terminal
and KE to earth (PE)
2. Press the TEST button
3. Switch the supply voltage on and off
1. Press the TEST button
2. Switch the supply voltage on and off
3. Please contact Bender
27
Function

3.5.8 Relay K3: device fault alarm and EDS common message

Relay K3 is intended to signal device and connection errors of the ISOMETER®. In the Master mode with bus address 1, EDS alarm messages can be collectively indicated by the K3 relay as well as system fault messages. Coll­ective indication of the EDS alarms by K3 can be disabled by selecting „K3 Alarm: off“ from the "EDS Setup" menu. Refer to page 63. K3 is permanently set to N/C operation, with the contacts 31-34 normally closed, that means when a fault occurs, the relay deenergizes (K3: 31-32 con­nected). Further details are described in "Chapter 3.5.7 Self test". Even when a protocol converter of the FTC470… series is installed in a BMS system which is tem­porarily taking over the Master function is, the function collective EDS mes­sage will remain. This function is directly dependent on BMS address 1, the Master status is a secondary factor.
The settings for K3 are preset and cannot be adjusted.
28
IRDH575_D00089_05_M_XXEN/01.2020

4. Connection

Only qualified personnel are permitted to carry out the work necessary to install, commission and run a device or system.
Risk of electrocution due to electric shock!
Touching live parts of the system carries the risk of:
 An electric shock  Damage to the electrical installation  Destruction of the device
Before installing and connecting the device, make sure
that the installation has been de-energised. Observe the rules for working on electrical installations.
IRDH575 has plug-in terminals.
Connect the terminals A1/+ and A2/- to the supply voltage U with IEC 60364-4-43. The connections to the supply voltage shall be provided with protective devices to afford protection in the event of a short-circuit (a 6 A fuse is recommended). For UL and CSA applications, the use of 5 A fuses is mandatory.
Devices for protection against short-circuit in conformity with IEC 60364-4-43 for the IT system coupling L1/L2/L3 can be omitted if the wi­ring is carried out in such a manner as to reduce the risk of a short-circuited to a minimum (a short-circuit-proof and earth-fault-proof wiring is recommen­ded).
Use the accompanying terminal covers for terminal protection.
in accordance
S
29
IRDH575_D00089_05_M_XXEN/01.2020
Connection
Only one ISOMETER® may be connected to an external TEST or RESET button. The TEST and RESET inputs of different insulation monitoring devices must not be connected in parallel for collective testing.
Risk of property damage due to unprofessional installation!
If more than one insulation monitoring device is connected to a conductively connected system, the system can be damaged. If several devices are connected, the device does not function and does not signal insulation faults. Make sure that only one insulation monitoring device is connected in each conductively connected system.
Ensure disconnection from the IT system!
When insulation or voltage tests are to be carried out, the device shall be isolated from the system for the test period. Otherwise the device may be damaged.
IRDH575_D00089_05_M_XXEN/01.2020
30
Connection
31042
PE
6
1
5
L1 L2 L3 N
L+
L-
DC-System
L2L1
U
n
9
78
U
n
3(N) AC-System
L2L1
L1
L2
AC-System
L2L1
U
n
L3
L3
L3
mA
M+ M- T1 T2 R1 R2 F1 F2 A B 31 32 34 21 22 24 11 12 14
IRDH575
RS485
OFF
S2 S1
ON
K3
K2
K1
A1/+ A2/- L1 L2 L3 KE
6 A
11
12
13
14
31
IRDH575_D00089_05_M_XXEN/01.2020
Loading...
+ 90 hidden pages