Bender ATICS UMA710-2-63-ISO, ATICS UMA710-2-63-ISO-BP, ATICS UMA710-2-80-ISO, ATICS UMA710-2-80-ISO-BP User Manual

Manual
of ATICS® modules type
UMA710-2-63-ISO
UMA710-2-63-ISO-BP
UMA710-2-80-ISO
UMA710-2-80-ISO-BP
2-pole
automatic changeover and monitoring modules
with insulation monitoring
for medically used rooms
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 1
Imprint
Issued by: Bender GmbH & Co KG
Londorferstr. 65 35305 Grünberg, Germany
Tel: +49 6401-807-0 Fax: +49 6401-807-259
info@bender.de www.bender-de.com
All rights reserved. Subject to change without notice. Reprinting only with permission of the publisher.
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 2
Table of content
Imprint ............................................................................................................................................................. 2
Table of content ............................................................................................................................................... 3
1. How to use this operating manual effectively ...................................................................................... 5
1.1. How to use this manual ....................................................................................................................... 5
1.2. Explanations of symbols and notes ..................................................................................................... 5
2. Safety instructions ............................................................................................................................... 6
2.1. Intended use ........................................................................................................................................ 6
2.2. Skilled persons .................................................................................................................................... 6
2.3. Device-specific safety instruction ........................................................................................................ 6
2.4. General safety instructions .................................................................................................................. 7
2.5. Delivery conditions, guarantee, warranty and liability ......................................................................... 7
3. System description .............................................................................................................................. 8
3.1. MEDICS® ............................................................................................................................................ 8
3.2. Example applications ........................................................................................................................... 9
3.2.1. Example application operating theatre ................................................................................................ 9
3.2.2. Example intensive care unit ............................................................................................................... 10
3.3. Features of the UMA710… ................................................................................................................ 11
3.4. The automatic changeover module UMA710… ................................................................................ 12
3.4.1. The transfer switching device ............................................................................................................ 12
3.4.2. Insulation monitoring of the IT-System .............................................................................................. 14
3.4.3. Monitoring the device functions ......................................................................................................... 14
3.4.4. Power supply ..................................................................................................................................... 15
3.4.5. Manual mode ..................................................................................................................................... 15
3.4.6. Bypass mode (optional) ..................................................................................................................... 15
4. System components .......................................................................................................................... 16
4.1. Front view of atypical UMA710-2-xx-ISO-BP .................................................................................... 17
4.2. Typical mounting arrangement UMA710-2-xx-ISO-BP ..................................................................... 18
4.3. Automatic changeover and monitoring unit ATICS® ......................................................................... 19
4.4. Typical terminal setup ........................................................................................................................ 20
5. Installation of UMA710-2-xx-ISO-xx- ............................................................................................. 21
5.1. Safety Advice ..................................................................................................................................... 21
5.1.1. Short circuit protection ....................................................................................................................... 21
5.2. Changeover module .......................................................................................................................... 23
5.2.1. Typical connection ............................................................................................................................. 23
5.2.2. Wiring diagram UMA710-2-xx-ISO, …-BP ........................................................................................ 23
5.3. Notes for connecting the module ....................................................................................................... 24
5.3.1. Transformer ....................................................................................................................................... 24
5.3.2. Temperature sensor .......................................................................................................................... 24
5.3.3. Out-going branch circuit breakers (optional) ..................................................................................... 24
5.3.4. Remote alarm indicator ..................................................................................................................... 25
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 3
5.3.5. Building management and SCADA system ....................................................................................... 25
5.3.6. Bypass-switch (optional) .................................................................................................................... 25
6. Commissioning, settings and testing ................................................................................................. 26
6.1. Design and installation notes ............................................................................................................. 26
6.2. Setting and testing according to the checklist ................................................................................... 26
6.3. Avoiding errors ................................................................................................................................... 27
6.4. Example on how to set the addresses .............................................................................................. 28
6.5. Operation of the ATICS® unit ............................................................................................................ 29
6.6. Use of the bypass circuitry (optional) ................................................................................................ 29
7. Troubleshooting ................................................................................................................................. 31
7.1. Fault and alarm messages ATICS® .................................................................................................. 31
7.1.1. Alarm messages: ............................................................................................................................... 31
7.1.2. Messages with Error code or Service code ....................................................................................... 33
7.2. Replacements .................................................................................................................................... 34
7.2.1. Replacing the ATICS® unit ................................................................................................................ 34
7.2.2. Replacing the ATICS® unit when a bypass switch is installed ......................................................... 34
7.3. Fuses F1, F2 ...................................................................................................................................... 35
8. Periodic verification and Service ....................................................................................................... 36
8.1. Periodic verification ........................................................................................................................... 36
8.2. Commissioning and service ............................................................................................................... 37
8.3. Maintenance ...................................................................................................................................... 37
9. Data ................................................................................................................................................... 38
9.1. TÜV-Test report ................................................................................................................................. 38
9.2. Standards .......................................................................................................................................... 39
9.3. Manufacturer's certificate ................................................................................................................... 39
9.4. Technical data ................................................................................................................................... 40
9.4.1. Dimensions and weights .................................................................................................................... 42
9.5. Utilisation data ................................................................................................................................... 42
10. Manufacturers certificate, checklist, documentation .......................................................................... 43
11. Appendix ............................................................................................................................................ 43
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 4
1. How to use this operating manual effectively
1.1. How to use this manual
This operating manual describes how to operate the UMA710... changeover and monitoring module. It is designed for skilled persons working in electrical engineering and electronics; and in particular for those, installing and operating electrical equipment in the medical sector.
Before using the equipment, please read this operating manual, the supplement entitled "Important safety instructions for Bender Products" and the instruction leaflets supplied with the individual system components. This document must be kept in an easily accessible location near to the equipment.
Should you have any further questions, please contact our Technical Sales Department. We are also happy to provide on-site service. Please contact our Service Department for more information.
In this manual, the two redundant supplies for the power supply will be designated "preferred supply" or "line 1" and "second supply" or "line 2". Devices with displays use the terms "line 1" and "line 2" in the text they indicate.
This manual has been compiled with great care. Nevertheless errors and omissions cannot be entirely excluded. The Bender Group cannot accept any liability for injury to persons or damage to property resulting from errors or mistakes in this operating manual.
1.2. Explanations of symbols and notes
The following terms and symbols are used to denote hazards and instructions in Bender documentation:
This symbol indicates an immediate risk to life and limb. Failure to observe the associated instructions and take appropriate precautions will result in death, severe bodily injury or substantial damage to property.
This symbol indicates a potential risk to life and limb. Failure to observe these warnings and take appropriate precautions may result in death, severe bodily injury or substantial damage to property.
This symbol indicates a potentially dangerous situation. Failure to observe the associated instructions and take appropriate precautions may result in minor bodily injury or damage to property.
This symbol indicates important information about the correct use of the equipment purchased. Failure to observe the associated instructions can result in equipment malfunctioning or cause problems in the environment in which it is being used
This symbol indicates tips for using the equipment and particularly useful information. This type of information will help you to optimise your use of the equipment.
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 5
2. Safety instructions
2.1. Intended use
Changeover devices are used everywhere there is dependence on high availability from the power supply. The ATICS® transfer switching device is intended for the application described in the chapter "System description". When the preferred supply fails, the ATICS® automatically switches to the second supply.
Areas of application:
Group 1 and 2 medical locations according to DIN VDE 0100-710 and IEC 60364-7-710  Hospital main distribution boards (DIN VDE 0100-710)  Locations open to the public (DIN VDE 0100-718)  Emergency power supplies  Heating, air conditioning, ventilation, cooling  EDP, computer centres  Fire extinguisher and sprinkler systems
Several versions of the ATICS® are available. They differ for example by changing over from the two or four-pole systems or by the load current (see chapter "ATICS® tasks" on page 15). Please heed the limits of the area of application indicated in the technical specifications. Use which deviates from or is beyond the scope of these technical specifications is considered non-compliant. In order to meet the requirements of applicable standards, customised parameter settings must be made on the equipment in order to adapt it to local equipment and operating conditions. Intended use includes following all the instructions in the operating manual and complying with the test intervals.
2.2. Skilled persons
Only appropriately qualified personnel may work on Bender devices. Persons who are familiar with the assembly, commissioning and operation of the equipment and have undergone appropriate training are considered skilled persons. Such persons must have read this manual and understood all instructions relating to safety.
2.3. Device-specific safety instruction
Failure to adjust the settings may result in malfunction.
The settings must be changed in order to adapt the ATICS® transfer switching device to the existing equipment. When doing so, follow the instructions in chapter "5. Commissioning, settings and testing“.
Functional safety according to IEC 61508 can only be guaranteed when used properly.
Please follow the instructions given in this operating manual and in the check list. In particular note the information on the pages: 12, 13, 14, 20, 52 and 76 in the technical operating manual TGH1443 of the ATICS® device.
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 6
2.4. General safety instructions
Bender devices are designed and built in accordance with the state of the art and accepted rules in respect of technical safety. However, the use of such devices may introduce risks to the life and limb of the user or third parties and/or result in damage to Bender devices or other property.
Only use Bender equipment:
As intended In perfect working order In compliance with the accident prevention regulations and guidelines applicable at the
location of use
Eliminate all faults immediately which may endanger safety.  The device may not be opened.  Do not make any unauthorised changes and only use replacement parts and optional accessories
purchased from or recommended by the manufacturer of the equipment. Failure to observe this requirement can result in fire, electric shock and injury.
Information plates must always be clearly legible. Replace damaged or illegible plates immediately.  Make sure that the dimension of the UPS (special safety power supply source), the generator set
and the whole wiring is adequate. Abide by the relevant, applicable national and international standards. If there is an overload and short-circuit, this is the only way to guarantee the necessary safety and to ensure that the safety devices respond selectively.
If the device is overloaded by overvoltage or a short-circuit current load, it must be checked and
replaced if necessary.
2.5. Delivery conditions, guarantee, warranty and liability
The conditions of sale and delivery set out by Bender GmbH shall apply. Conditions of sale and delivery can be obtained from Bender GmbH in printed or electronic format.
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 7
3. System description
3.1. MEDICS®
The UMA710... change over and monitoring module is a part of the MEDICS® system. MEDICS® does not refer to a single product, but rather an intelligent system for safe power supplies in medical locations.
Example of a section of a hospital with the MEDICS® system
Table to the example
MK... Alarm indicator and test combination RCMS... Residual current monitoring system for TN-S systems SMI472 Signal converter for third-party technical equipment (e. g. med. gases, UPS) TM Alarm and control panel UMA710-2... Change over and monitoring module ATICS® for IT-Systems UMC107E.. Change over and monitoring module with contactors for IT-Systems UMA710-2... Change over and monitoring module ATICS® for IT-Systems with EDS... insulation fault
location system.
UFC107E... Change over and monitoring module for IT systems with EDS... insulation fault location
system UMC710D... Changeover and monitoring module for main distribution boards USC710D... Control module for changeover modules (preferably in main distribution boards)
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 8
MEDICS® includes::
AC and 3(N) AC changeover and monitoring modules. Examples of modules in the MEDICS®
system include UMC..., USC..., UFC... and EDS.... insulation fault location systems.
Display and operating units such as alarm indicator and operator panels or alarm indicator and test
combinations.
Communication between these components takes place via the BMS bus (two-wire connection).  The connection of third-party technical equipment by means of protocol converters (gateways), via
digital inputs and relay outputs.
The real strength of MEDICS® is to be found in communication between all involved components and the resulting information provided to the user. Readiness for operation is monitored constantly; and operating states, irregularities, faults and equipment failures are displayed. From the user's point of view, this means high operational reliability.
3.2. Example applications
3.2.1. Example application operating theatre
UMA710 Module, Changeover between the preferred and redundant line while monitoring the
medical IT system with transformer load and temperature monitoring
IR426-D47: Monitoring the operating theatre light IT system (optional)  MK2430 / MK800 / TM800: Alarm for at least two points with independent power supplies for
functional safety
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 9
3.2.2. Example intensive care unit
UMA Module, Changeover between the preferred and redundant line while monitoring the  medical IT system with transformer load and temperature monitoring including a bypass switch  EDS151 Insulation fault locator for fast insulation fault location (recommended)  ATICS-BP: Bypass switch for uninterrupted test / maintenance (recommended)  MK2430 / MK800 / TM800: Alarm for at least two points with independent power supplies for
functional safety
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 10
3.3. Features of the UMA710
The ATICS® transfer switching device has the following features:
Continuous monitoring of the internal electronic and the circuitry with automatic alarm  Preventive safety due to an automatic reminder of mandatory testing procedures  Maximum reliability of the switching
o Patented switching system with mechanical and electrical interlocking o Short circuit proof contacts with the mechanic of a moulded case circuit breaker o Insensitive to voltage fluctuation and vibration due to stable switch position and permanent
contact pressure
Easy to use and a good overview due to a clear menu structure and user interface  The right information at the right time with clear alarm text and a backlit graphic display as well as
via the BMS bus
Can manually be operated and can be locked for maintenance in position “0” Automatic logging of events e.g. switching, testing, parameter changes  External functional test or replacement of the ATICS® via an optional bypass switch  The ATICS® switch is pluggable  Electronic and switching elements all in one enclosure respectively on a mounting rack  Chang over and IT-System monitoring in one unit  Optional bypass switch  Communication of components with one another via BMS bus  Communication with remote alarm indicator and test combinations and TM alarm indicator and
operator panels via BMS bus
Can be used in systems in accordance with IEC 60364-7-710:2002-11, sections 556.5.2.2, 556.7
and 556.8 and DIN VDE 0100-710 (VDE 0100 Part 710):2002-11, section 710.564.6 with a changeover period = 15 s or = 0.5 s
Functional Safety according to IEC 61508 for use in special safety environment according to SIL2  Easy field wiring, completely prewired and type tested assembly
The UMA710 module is a configured unit and is only certified and tested in this assembly. Do not make any changes to the components, their password-protected settings or the wiring without consulting Bender first. In each case you should make the settings that are required for adaptation to the application case in question and local conditions.
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 11
3.4. The automatic changeover module UMA710
3.4.1. The transfer switching device
If the preferred supply fails, the ATICS® ensures the power supply is changed over safely. The switch contacts are offset on a rotating shaft. This design prevents simultaneous switching of line 1 and line 2. The switch has three positions:
1 Line 1 switched on 0 Both lines switched off 2 Line 2 switched on
Either line 1 or line 2 can be set as the preferred line
In the normal condition (fault-free operation) the preferred supply is connected. The ATICS® will
switch to the redundant line if:
the preferred line fails the "TEST" button is pressed a digital input is configured to "TEST" and this input is enabled the setting "Preferred supply" is reconfigured to the other line
The ATICS® switches from the redundant line back to the preferred line if
– the voltage on the preferred line is restored
- the return transfer delay time T(2->1) has expired and no switching back interlocking function is enabled
- or immediately after pressing the "RESET" button or when the redundant line fails (even when the switching-back interlocking function is enabled)
the setting "Preferred supply" is reconfigured to the other line the digital input is configured to "TEST" and this input is reset a transfer switching device test is enabled and the test time has expired
With the optional ATICS-ES energy storage device only: The ATICS® changes to switch position
"0" and stays there if the following conditions are simultaneously met:
Line1 and Line 2 have failed Automatic mode is set No short-circuit is present downstream of the transfer switching device The setting "Load disconnection" is set to "on" and the external ATICS-ES energy storage device
is connected
Malfunction if delay times are not adjusted
The response delay T(on), the return transfer delay time T(2->1), the delay on release T(off) and the dead time T(0) of the ATICS® are adjustable and must be adjusted to the requirements of the specific case, the short-circuit-calculation and the requirements of DIN VDE 0100-710 (VDE 0100 Part 710) Section 710.537.6 automatic changeover devices (see also chapter "Commissioning, settings and testing").
The factory settings guarantee a changeover period of t = 0.5 seconds and switching back within 10 seconds when voltage is restored on the preferred supply. Therefore, the ATICS® can be used in IT systems with a requirement for a changeover period t ≤ 0.5 s (IT systems with operating theatre lights, endoscopic field illumination in operating theatres or other essential sources of light, etc.).
When there is a short circuit downstream of the transfer switching device, the transfer switching device must not continually change back and forth between the two lines. This can occur when the short-circuit current is small and the transfer switching device switches faster than the short circuit breaker trips. The ATICS® monitors the load current downstream of the transfer switching device in order to detect a possible short circuit. If the preferred line fails and a short-circuit current is detected at the same time, the ATICS® does not change over immediately but only once the circuit breaker has tripped.
If the ATICS® detects a supply failure or a fault, an alarm appears on the LCD, the "ALARM" LED lights up, the alarm relay trips (if set) and this alarm is forwarded to other Bender devices (such as an alarm indicator and test combination) via the BMS bus.
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 12
Typical time diagrams can be found in the following chapters of the ATICS® switch user manual TGH1443 issued by Bender GmbH.
TGH1443, chapter 3.5.1.1 Time diagram: Changeover between the preferred and redundant line Example: Line 1 is set as the preferred line.
TGH1443, chapter 3.5.1.2 Time diagram: Staggered switching after total power failure The external ATICS-ES energy storage device (optional, see "Ordering information") is required for the staggered switching. If there is no voltage on either of the power supplies, the ATICS® switches to position "0". When the power is restored, the ATICS® switches on the supply again with the set delay time T(start). Example: Line 1 is set as the preferred line.
TGH1443, chapter 3.5.1.3 Time diagram: Changeover to generator mode
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 13
3.4.2. Insulation monitoring of the IT-System
Insulation monitoring
The integrated insulation monitoring device measures the insulation resistance on AC IT systems, which may also contain DC voltage components. Adaptation to the system leakage capacitances occurs automatically. The measuring time increases as the system leakage capacitances increase.
Load current measurement
The load current on the IT system is measured using an STW2 measuring current transformer.
Temperature monitoring
The temperature in the transformer winding is measured via PTC thermistors.
Evaluation
If any of the measured values does not fall within the limits, an alarm is triggered. A message appears on the LCD, the "ALARM" LED lights up, the alarm relay trips (if set) and this alarm is forwarded to other Bender devices (such as an alarm indicator and test combination) via the BMS bus.
Locating current injector
When an insulation fault is detected on an IT system, the integrated locating current injector generates a defined locating current signal to locate the insulation fault. The locating current is limited to 1 mA. This feature can be switched on or off (see chapter "7.4.4.4 Settings menu 4: IT system").
The locating current flowing between the IT system and earth can cause controller faults in sensitive parts of the system, such as the PLC or relay. It must therefore be ensured that the locating current is compatible with the monitoring
3.4.3. Monitoring the device functions
The control circuits are designed in such a way that, even though it is almost certain that a particular fault will occur, it cannot cause the power supply at the output of the automatic transfer switching device to fail. The ATICS® also continuously monitors:
The switch position of the switch and coils 1 and 2 of the switch  Power supplies 1 and 2, which supply the electronics from the systems concerned  Internal microcontrollers and memory modules  Important connecting wires, such as:
Measuring current transformer connection Temperature sensor connection Power supply and PE connection
For alarm and test combinations, and alarm indicator and operator panels, device failure
monitoring can also be programmed via the BMS bus (necessary for functional safety).
On systems with generator: total power failure possible
If the preferred line fails, the ATICS® starts the generator which is connected to the redundant line. If the generator does not start, the line downstream of the transfer switching device will be dead. When the generator is switched off, the ATICS® cannot check the redundant line. Therefore, regularly test the generator and the changeover function to make sure they are
working properly (see checklist in the Appendix to this manual).
Alle Rechte und Änderungen vorbehalten. Subject to change without notice. All rights reserved. © Bender GmbH & Co KG UMA710-2-xx-ISO-xx, Manual.docx Issued: 11.Jan.2016 14
Loading...
+ 30 hidden pages