BEI Sensors Programmable Hall Device User Manual

Page 1
PHD User Manual
Table of Contents
Overview....................................................................................................2
Install PHD Software.................................................................................3
Connect PHD Equipment..........................................................................5
Connect Sensor to the PHD Programmer.................................................6
Run PHD Application.................................................................................7
Verify Parameters and Program Sensor..................................................10
Saving and Loading Programs................................................................13
Troubleshooting .....................................................................................14
Page 2
PHD User Manual
Overview
BEI Sensors introduces the Programmable Hall Device (PHD), an easy way to take control of custom­izing BEI Hall effect sensors output. This compact system interfaces with your PC to BEI Hall effect
sensors and provides users with the ability to create output functions that meet specic requirements.
The system includes proprietary PHD software (compatible with Windows XP, 7 or Vista), the PHD programming unit, power cord and a USB computer interface cable. Cables to interface the PHD to
specic sensor connectors are sold separately. A positions xture is recommended if the sensors are not programmed after installation in the nal assembly
The PHD programming process uses simple 6-step procedure:
1. Install the PHD software.
2. Connect PHD equipment.
3. Connect the programmable sensor to the PHD Programmer using the Sensor Interconnect Cable.
4. Run PHD application and create the conguration prole on a PC.
5. Verify parameters and congure the sensor by manually rotating the sensor through the desired range of motion and selecting the programming points in the PHD application.
6. View output and verify the sensor is congured as desired.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
2
Page 3
PHD User Manual
Step 1: Install PHD Software
1. Insert USB ash drive provided with the PHD and install software. Desktop and start menu shortcuts will be created.
2. Click “Open folder to view les”.
3. Right-click “setup.exe” and select “Run as Administrator”.
4. Run through PHD install and click “Next>>” until the installation is complete.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
3
Page 4
PHD User Manual
5. Once the software has nished installing, the PHD Drivers will automatically be installed.
6. Click “Next>” to nish the Device Driver installation process.
7. Click “Finish”. You have successfully installed the PHD software and drivers.
8. Remove the USB ash drive.
9. Restart the computer.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
4
Page 5
PHD User Manual
Step 2: Connect PHD Equipment
1. Connect the USB-A end of the Computer Interface Cable to the USB port on your computer.
2. Connect the USB-B end of the Computer Interconnect Cable to the USB port on the Programmer.
3. Connect the AC Power Cord to the AC Adaptor.
4. Connect the AC Adaptor to the Programmer.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
5
Page 6
PHD User Manual
Step 3: Connect BEI Programmable Sensor to PHD Programmer
1. Connect the 25-pin connector end of the Sensor Interconnect Cable to the Programmer.
2. Connect the 6-pin or 3-pin connector end of the Sensor Interconnect Cable to the BEI Programmable Sensor.
3. Plug the AC Power Cord into an outlet and move the power switch on the Programmer to the
“1”or on position. The green and red lights should be lit.
Verify that the Windows Driver Install was successful.
A small bubble should appear alearting you that the PHD has been detected.
Click on the bubble to bring up the following screen.
After a moment of searching, Windows will locate the drivers and the PHD Programmer
will be ready to use.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
6
Page 7
PHD User Manual
Step 4: Run PHD Application and create the desired program
1. Run the PHD Application. It will automatically nd and connect to the PHD Programmer.
2. If the software cannot nd a PHD Programmer a “No PHD Device Found!” message will appear.
*Note: The user can continue without a PHD Device connected to create programming proles for later use. If there is no Programmer connected, a “Connect to PHD?” dialog will appear. Once you have connected the programmer, select “Connect” to continue.
3. Select the Unit Type: This is the series of BEI’s programmable sensor you will be programming.
*Note: Programmable versions of the 8360 and 9360 series will be available soon.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
7
Page 8
PHD User Manual
4. Select the Output Type: The output type is selected based on the “Unit Type” chosen in Step #4.
*NOTE: The desired programming will not be achieved if the incorrect Output Type is choosen.
Unit Type 9960–this sensor series has several different versions of programmable types.
0-5V Ratio – This is a 9960-5A Input/Output version, 0-5V Input for 0-5V Output.
PWM – This is a 9960-5X2 Input/Output version, 5V Input only.
0-5V Non- Ratiometric – This is a 9960-9X2 Input/Output version, 9-30V Input for 0-5V
Output
0-10V Non-Ratiometric – This is a 9960-15X Input/Output version, 15-30V Input for
0-10V Output
4-20mA – This is a 9960-9X1 Input/Output version, 9-30V Input for 4-20mA Output
5. Dene the Rotation Direction of the sensor. In the Rotation Direction box there are two options, Clockwise (CW) and Counterclockwise (CCW).
CW– Choose Clockwise for both outputs if the desired output slopes you are programming occur when the rotor is rotated clockwise (when viewed from the rotor side of the sensor).
CCW– Choose Counterclockwise for both outputs if the desired output slopes you are programming occur when the rotor is rotated counter clockwise (when viewed from the rotor side of the sensor).
6. Dene the Programming Points for Sensors 1 and 2 (Sensor 1 and Sensor 2 do not have to have the same number of programming points).
You are able to choose where in the rotational angle each of these points are as well as the output percentage at each point. For example, 3 Point options are most commonly dened as:
Point “A”– percentage of output at beginning of travel
Point “B”– between Points “A” and “C”, usually midway (50%)
Point “C”– percentage of output at end of travel
• The percentage output for Point “A” and Point “C” depends on your needs. For proper operation the % for Points ‘A”, “B” and “C” must be greater than the Clamp Low % and less than the Clamp High %.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
8
Page 9
PHD User Manual
7. Dene four additional parameters for the part:
Hysteresis– The Hysteresis option allows you to to
independently program the amount of hysteresis each output will have. A value of zero degrees will give the minimum amount
of hysteresis and 11 degrees will give the maximum amount of
hysteresis. This value is most commonly set to zero.
Travel (Deg)– The Travel option allows you to program the
rotational angle you want to resolve. Values from 1° to 360° are valid, however the linearity of the output may be negatively impacted for angles less than 15°. It is possible to program a separate angle for each of the sensor’s outputs; however
a slightly different programming sequence from this example is required.
Clamp Low and Clamp High– Allows you to independently set min/max percentage
limits for each output. These limits restrict each output from going beyond the limit
programmed. Clamp High and Low work exactly the same way for the Non-Ratio Metric and PWM units as it does for the Ratiometric. For example assuming 5VDC is supplied to
the sensor’s input, the output would be as follows:
Clamp Low
(%)
0 100 0/5
5 95 0.25/4.75
10 90 0.5/4.5
25 80 1.25/4
Clamp High
(%)
Min/Max Vout
(V)
8. If PWM was selected as the Output Type two additional parameters must be set:
PWM Frequency is the output frequency which the PWM unit will output
at, this value is CONSTANT and does not change with output location.
PWM Polarity determines how the Duty Cycle changes based on the position of the shaft.
Rising Edge– 0% High at 0% position, 100% High at 100% position.
Falling Edge– 0% High at 0% position, 100% High at 100% position.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
9
Page 10
PHD User Manual
Step 5: Verify Parameters and Program the BEI Sensor
1. Click “Verify Parameters” for either Sensor 1 or Sensor 2. This step checks to make sure that all the needed values have been entered correctly.
If there is an error the software will indicate what needs to be xed.
If there is no error the software will grey out the
sensor under program.
2. Click “Begin Programming” to start the programming process, this initializes the chip and prepares it for programming.
3. If you want to insure correlation between the two Sensors, select “Synchronous
Programming”.
*NOTE: This feature is only available if Sensor 1 and Sensor 2 have the same “Programming Points”
selected.
4. Position the sensor at the desired location for Point A and click “Program A” without moving the part.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
10
Page 11
PHD User Manual
5. Rotate the part in the appropriate direction (In this example Counter Clockwise) as selected in the “Rotation Direction” eld to the next programming point, Point B and click “Program B”
without moving the part from the desired position.
6. If you have selected something other than a “2 Points” programming option, repeat steps 4 & 5 for the remaining points. Otherwise, you may release the part and click “Finalize Programming” to nish the programming process.
7. You may now repeat the process for the other sensor. The procedure is the same if you have
selected “Synchronous Programming”.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
11
Page 12
PHD User Manual
Step 6: View output and verify conguration
1. If you wish to verify the output of the programmed sensor select the “View Output” button on the lower right hand section of the screen to bring up a real-time graph of the sensor output.
2. Rotate the part to observe how it’s output changes.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
12
Page 13
PHD User Manual
Saving and Loading Programs
Once you have selected the parameters to program you can choose to save those settings to use at another time using the File menu in the software.
1. Select “File” in the top menu barand then click “Save” and store the le on your computer for later use.
2. Select “Load” to open a previous saved program.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
13
Page 14
PHD User Manual
PHD Troubleshooting
Problem Solution
Software Does Not Install Correctly
The most likely cause of an invalid installation relates to computer security.
Uninstall the software and re-install being sure to right-
click the setup.exe le and select “Run as Administrator” Make sure that the PHD is properly connected to the
computer and is running the software via the supplied USB Computer Interface Cable.
Software Does Not Find the Hardware
Software Doesn’t See the Sensor
Sensor Programs “Backwards”
Sensor Programs Incorrectly
Sensor Outputs Don’t Correlate
Error Message “Part installed is not a BEI Sensors part.”
Verify that the red indicator light on the PHD is lit. This indicates that the PHD is receiving power.
Validate that the Drivers installed correctly by going to-
Start >> Control Panel >> Device Manager. There should be no devices with a yellow exclamation point.
Make sure that the sensor is connected correctly; re-
versed leads will prevent the communication process.
Verify that the right type of sensor has been selected. Verify that the correct rotation direction has been chosen.
Recall: The rotation direction is dened as the direction of
rotation when viewing the sensor from the shafted side.
Insure that the correct values have been selected for the programming points.
Insure that all connections to the sensor are solid and free of contaminates. An intermittent connection can cause issues.
Check all programming parameters for errors.
Check that the correct sensor type has been selected and the proper output type to correlate with that sensor
Output correlation can only be guaranteed by using the
“Simultaneous Programming” option. Even with this it is possible for slight variations to exist.
There are a few reasons that this message may appear:
Attempting to program a Non-BEI Part
Attempting to program a Non-Programmable BEI
Part
An error occurred at the factory and the part was
incorrectly coded.
If you feel this is the case, please contact BEI Sensors, Goleta for assistance.
7230 Hollister Avenue, Goleta, California 93117 | Tel: 805-968-0782 | Fax: 805-968-3154 | www.beisensors.com
Specication No.: 02135 Rev: 5-23-2013
14
Loading...