Step 3: Calibrate the Sensors
When you turn on the stabilizer, it should balance the camera in an upright position with the
horizon level. To ensure a proper alignment, the sensors in the system need to be calibrated
occasionally, particularly after changes in temperature or if you travel with the system. It is good
practice to calibrate the sensors before each day of shooting, as the process takes only 10
seconds.
1.
Set the stabilizer on the stand on a sturdy level surface.
2.
Attach and balance your camera.
3.
Turn on the stabilizer and wait for the start-up sequence to complete (three green lights
flash).
4.
Hold the stabilizer mode button down for two seconds.
5.
The lights will blink red, and the motors will turn off to indicate “calibration” mode.
6.
Gently tap the stabilizer button once to start the calibration sequence.
7.
The stabilizer will wait one second for the button shake to dissipate and will then calibrate
the sensors. It takes about 10 seconds. Do not bump or move the stabilizer during
calibration.
8.
When the calibration is complete, the lights will flash green, the motors will turn on, and the
stabilizer will return to normal operation.
TIPS FOR SENSOR CALIBRATION
*
Calibration is not necessary every time you turn on the stabilizer. However, variables like air
pressure changes, temperature changes, and long periods of time can cause the calibration to
drift slightly. For best performance, we recommend calibrating each time you set up the
stabilizer at a new shooting location.
*
It is possible to temporarily change the horizon point by turning follow mode off (three red
lights) when the camera is not in a neutral position. You can use this to your advantage, for
example, if you want to mount the camera when it is tilting up. To reset the neutral position in
this case, simply turn off the stabilizer and then turn it on again.