bedfront piCO+ Smokerlyzer Operating Manual

Operating manual
piCO
+TM
Smokerlyzer
®
breath analysis is the new blood test
piCO+ Operating Manual
English
Contents
1 Introduction and intended use 4
2 Quick start guide 5
3 Symbols and specification 6
4 Pack content list 7
5 Instrument layout 7
6 Display symbols 8
7 Warnings and maintenance 9
8 Operation 10-11
9 Calibration 12-14
10 Set-up – User profiles 15
11 Set-up – Sounder operation 16
12 Troubleshooting 17
13 COdata+ software 18
14 Frequently asked questions (FAQ) 19
15 Returns procedure 20
16 Spares and warranty 21
piCO+ Operating Manual
English
www.bedfont.com
Introduction
The piCO+ Smokerlyzer® is a breath carbon monoxide monitor intended for multi­patient use by healthcare professionals in smoking cessation programmes and research. It can also be used as an indicator of carbon monoxide poisoning in a healthcare environment.
Carbon monoxide is a toxic, odourless, colourless, tasteless gas. It is formed by incomplete combustion of organic material
at high temperatures, with an insufcient
oxygen supply.
When inhaled, CO displaces oxygen in the bloodstream to form carboxyhaemoglobin (COHb). This starves the body tissues of the oxygen vital to repair, regeneration and general living. CO can remain in the bloodstream for up to 24 hours, depending on factors such as physical activity, gender, and inhalation intensity. The half life is about 5 hours.
CO (ppm) and carboxyhaemoglobin
(%COHb) correlation:
Breath carbon monoxide is measured in parts per million (ppm) and blood carboxyhaemoglobin is measured in percentages (%COHb). The two are compatible and convertible: the CO reading relates to gas in the lungs and on the breath (i.e. the amount of poisonous CO that has been inhaled), and the COHb reading relates to the percentage of vital oxygen that has been replaced in the bloodstream. The piCO+ Smokerlyzer® displays both measurements. Clinical research has demonstrated that a useful relationship between carbon monoxide and carboxyhaemoglobin is obtained after a short period of breath-holding.
The cut-off point between smoker and non-smoker has been found to be 6ppm CO. The piCO+ Smokerlyzer® denes a non-smoker as 0-6ppm, a low-dependence smoker as 7-15ppm and strongly addicted smokers as over 15ppm.
The piCO+ Smokerlyzer® has been designed so that it can be used with young smokers. As their smoking habits and views are generally different from adult smokers, the cut-off points have changed: 0-4ppm represents a non-smoker, 5-6ppm a light or casual smoker, and over 7ppm a more frequent young smoker.
Other cut-off points can be adopted as
well as smoking classication at higher
CO levels. The piCO+ has a third prole which can be set by the user with COdata+
software to achieve complete exibility in
most circumstances
Intended use
piCO+ Operating Manual
English
Quick start guide
1 Turn on the monitor by holding down
the button for three seconds.
2 Attach a breath sampling D-piece and
new steribreath
TM
mouthpiece.
3 Inhale and double click button to start a
breath test.
4 Hold breath for the 15-second
countdown. If unable to hold breath for full 15 seconds, see ‘Operation’, page 10.
5 A beep will sound during the last three
seconds of the countdown.
6 Blow slowly into mouthpiece, aiming to
empty lungs completely.
7 The ppm and equivalent %
carboxyhaemoglobin (COHb) levels will rise and hold. The coloured LEDs will light accordingly. See the table on page 11 for an interpretation of the reading.
8 Remove the D-piece between tests to
purge the sensor with fresh air.
9 To repeat the breath test, double click the
button to return to main display and continue from step two, above
.
10 To switch off, hold button for 3 seconds.
Unit will auto power off after 5 minutes of inactivity.
piCO+ Operating Manual
English
www.bedfont.com
Specication
Concentration range: 0-100ppm
Display: LCD with 1ppm increments
Detection principle: Electrochemical sensor
Accuracy (repeatability of readings)
: ±2%
H2 cross-interference: <12%
Power: 2 × AA (LR6 or equivalent) alkaline batteries 1 x CR2032 Lithium coil cell (3v)
Response time: Typically <30 seconds to 90% FSD
Operating temperature range: 0-40°C (storage/transport 0-50°C)
Operating humidity: 10-90% (storage/transport 0-95%) non-condensing
Operating transport storage pressure 800-1200mbar
Sensor operating life: Approximately 5 years (5-year warranty)
Sensor sensitivity: 1ppm
Dimensions: Approximately 45 × 75 × 120 mm
Weight: Approximately 200g (including batteries)
Monitor construction:
Body: Polycarbonate/ABS blend with elastomeric overmould D-piece: Polypropylene
Symbols
Direct Current
Degree of protection against electric shock:
Type BF applied part
Type of protection against electric shock: Internally Powered Equipment
Please refer to the warnings and safety notes in the manual
Consult instructions for use
Degree of protection against ingress of liquid: IPXO - not protected against water ingress
Degree of safety application in the presence of a ammable anaeshetic mixture with air, oxygen or nitrous oxide: Equipment not suitable for use in the presence of ammable mixtures.
Environment
The piCO+ complies with the directive EN60601-1-2 electromagnetic compatability but can be affected by cellular
phones and by electromagnetic interference exceedding the levels specied in EN50082:1. This equipment should
be moved if necessary to avoid interference.
piCO+ Operating Manual
English
2
0
3
V
2
x
A
A
/L
R
6
B
edfo
nt
S
c
ie
ntific
L
td
M
E
1 3
Q
X
,
E
n
gla
nd
T
:
F
:
E
:
a
s
k
@
b
edfo
nt.c
om
w
w
w
.b
edfo
nt.c
o
m
0
0
8
6
1
2
3
4
5
7
6
10
9
8
2
12
11
Reset
+44(0)1622 851122 +44(0)1622 854860
13
12
Instrument layout
1 Display
2 Coloured LEDs
3 Button
4 Aperture for D-piece
5 Breath sampling D-piece
6 Securing lug on D-piece
7 Exhaust port for breath sample
8 Sounder
9 USB connector
10 Cardboard mouthpiece
11 Battery compartment
12 Reset button
13 Battery cover
5
2
Pack contents
1 Instrument
2 Carry pouch
3 User manual
4 Breath sampling D-piece
5 Steribreath
TM
mouthpieces
6 2 × AA batteries
7 Hand Foam
Infection control and maintenance guidelines
0086
Bedfont Scientic Ltd
ME17 1JA, England T: +44 (0)1622 851122 F: +44 (0)1622 854860 E: ask@bedfont.com www.bedfont.com
4
7
7
3
1
6
piCO+ Operating Manual
English
www.bedfont.com
1 Start up screens:
2 Battery condition: OK: Flat: Dead:
3 User prole: Adult: Adolescent: Custom:
4 Sounder: On: Mute:
5 Pointer:
6 Start breath test:
7 Enter calibration mode:
8 Enter set-up mode:
9 Inhale:
10 Breath-hold and countdown:
11 Exhale for breath test:
12 Peak CO reading (ppm and %COHb):
13 Set instrument zero in fresh air: (only if the reset button has been pressed)
14 Apply calibration gas at 20ppm:
15 Calibration pass/fail: Pass: Fail:
16 Calibration reminder:
17 User prole selection:
18 Select adult user prole:
19 Select adolescent user prole:
20 Select custom user prole (if set via COdata
+
software on PC):
21 Contact Bedfont or distributor for help:
22 D-piece reminder
23 Return:
9.2
Display symbols
piCO+ Operating Manual
English
Warnings and maintenance
Warnings
People with lung disease or chest ailments may not be able to achieve the 15-second breath-hold. In such cases, the user should inhale and hold their breath when the breath test is started, and exhale, if necessary, before the countdown has completed.
Steribreath mouthpieces should be replaced after every use.
Hands should be washed regularly in accordance with infection control practice.
Please do not attempt to modify the equipment in any way or use accessories
not specied by the manufacturer. Any
attempt to do so will invalidate the warranty and may compromise the safety of the device.
Bedfont will make available on request
service training to appropriately qualied
personnel
The PiCO+ should only be used by suitably
trained and qualied personnel typically
situated in a hospital or clinic environment.
The responsible organisation is advised that connection to other devices and
modications require evaluation to the
requirements of EN60601-1: 2006
Wipe the instrument and D-piece external
surfaces with a product specically
developed for this purpose. Bedfont provides instrument cleansing wipes.
Cleaning
Routine maintenance
Calibrate the monitor using Bedfont 20ppm CO in air calibration gas when prompted by the calibration reminder on the piCO+ (see page 12). Once calibration has been carried out a reminder will reoccur in 6 months.
Replace batteries when indicated – battery symbol either or .
Replace breath sampling D-piece when prompted by the unit every 30 days or if visibly soiled or contaminated.
The D-piece cannot be cleaned or sterilised.
Batteries
Batteries should be removed if the instrument is not likely to be used for some time.
Additional technical information can be made available on request; please contact Bedfont or its distributor.
The operator should not touch both the battery terminals and the patient simultaniously.
NEVER use alcohol or cleaning agents containing alcohol or other organic solvents as long term exposure to these vapours will damage the CO sensor inside.
Under no circumstances should the instrument be immersed in liquid or splashed with liquid.
10
piCO+ Operating Manual
English
www.bedfont.com
Ensure that two AA batteries are correctly located in battery compartment.
• The single button controls all functions:
• One long press of the button (at least three seconds) turns the unit on.
• A single click scrolls the pointer through the list of options.
• A quick double click selects the required option.
• One long press of the button (at least three seconds) turns the unit off.
Press and hold the button for three seconds until the display becomes active. Release the button. After a 10-second warm-up period, during which a start-up screen is displayed, the main menu will appear.
If a is shown rather than the main menu, it means that the instrument has failed to set a fresh air zero during start-up. This may be because the ambient air was not clear of carbon monoxide. Ensure that the instrument is in fresh air, then double click the button to repeat the start-up test. If it fails again, see ‘Troubleshooting’, page 17.
When the pointer is opposite the picture of an exhaling face, the piCO+ is ready to start a breath test. The green LEDs will double
ash to indicate that the instrument is ready.
Operation
Attach a breath sampling D-piece and a new SteribreathTM mouthpiece to the piCO+.
Check that all connections are pushed rmly
together.
To start a breath test, the patient should inhale and hold their breath, then a quick double click of the button will initiate a 15-second countdown display.
11
piCO+ Operating Manual
English
Ensure that the breath is held throughout the countdown. An audio beep will sound during the last three seconds of the countdown. Once the countdown reaches zero, exhale slowly but gently into the mouthpiece. Aim to empty the lungs as much as possible.
If the patient cannot hold their breath for the full 15 seconds, they should commence exhalation at a comfortable point, but exhale completely.
The display will show a rising ppm value, and an equivalent %COHb reading. The coloured LEDs will illuminate, according to
the selected user prole.
The sounder will beep at an increasing frequency, according to the concentration of CO measured. To temporarily mute the sounder, press the button once.
The highest level detected during the exhalation will remain on the display. The test is complete when the pointer and ‘return’ symbols appear at the bottom of the display.
At this point, a single click of the button will temporarily silence the sounder, if it is on.
To start another breath test, double click the button to return to the main display. A new mouthpiece should be used with every test.
Removing the D-piece between tests will allow fresh air to purge the sensor. It is good practice to wash your hands after removing the D-piece.
If no further tests are required, the piCO+ should be turned off by holding the button for three seconds. If left on, the piCO+ will
automatically turn off after ve minutes of
inactivity.
To re-calibrate the instrument, follow the instructions on page 13.
To change the user prole from adult to adolescent (or a custom prole, if one has
been set using COdata+ software), see page15.
To permanently silence the sounder during the breath-test, follow the instructions on page 16.
Operation (continued)
LED colour Description User prole 1: User prole 2: Adult (ppm) Adolescent (ppm)
Green Non-smoker 0-6 0-4
Yellow Danger zone 7-10 5-6
1 Red Smoker 11-15 7-10
2 Reds Frequent smoker 16-25 11-15
3 Reds Addicted smoker 26-35 16-25
4 Reds Heavily addicted smoker 36-50 26-35
4 Reds ashing
Dangerously addicted smoker
51+ 36+
12
piCO+ Operating Manual
English
www.bedfont.com
3
2
4
1
1.5 l/m
1.0 l/m
0.5 l/m
1
2
Calibration
● Instruments are calibrated before leaving
Bedfont.
● Calibration reminders will
prompt the user to calibrate the unit every six months
● Double click the
● The calibration gas required
is Bedfont 20ppm carbon monoxide in air.
Ensure that the ne control valve is in the
off position.
Screw the ne control valve and ow
indicator assembly to the gas can. This is best done by screwing the gas can into the valve.
● Double click on the cylinder
Icon
Bedfont calibration kit
1. Fine control valve
2. Flow indicator
3. Gas cylinder, 20ppm CO in air
4. Calibration adapter and tube
Immediately open the ne control valve
and allow the gas to ow at 1 litre per
minute.
To maintain this, adjust the ow so the
ball in the ow indicator remains at the
middle line.
Allow the gas to ow through the instrument for up to 1.5 minutes to ensure accurate calibration, again
monitoring the rate of ow.
As the 20ppm CO calibration gas is applied, the displayed ppm reading will climb. After
1.5 minutes, or until no further increase in the reading is obtained, the
nal measured value will
be shown and sampling will cease.
13
piCO+ Operating Manual
English
● The calibration gas required
is Bedfont 20ppm carbon monoxide in air.
Ensure that the ne control
valve is in the off position.
Screw the ne control valve
and ow indicator assembly
to the gas can. This is best done by screwing the gas can into the valve.
Double click to select the cylinder symbol.
Immediately open the ne control valve
and allow the gas to ow at 1 litre per
minute.
To maintain this, adjust the ow so the
ball in the ow indicator remains at the
middle line.
Allow the gas to ow through the instrument for up to 1.5 minutes to ensure accurate calibration, again monitoring the
rate of ow.
As the 20ppm CO calibration gas is applied, the displayed ppm reading will climb. After 1.5 minutes, or until no further increase in the reading is obtained, the
nal measured value will be shown and
sampling will cease.
If the displayed reading is between 5 and 40ppm, the calibration value will be automatically set in the instrument as 20ppm and a tick will be displayed to show a successful calibration. If the
Calibration (Manual)
displayed reading is outside these limits, the calibration fails, and a will be displayed (see ‘Troubleshooting’, page 17, if this happens).
The pointer and ‘return’ symbols re-appear at the bottom of the display to indicate that the process is complete and to allow the user to double click the button to return to the previous menu.
Turn off the gas ow, remove the D-piece sampling system, and disconnect the calibration adapter from the D-piece sampling system.
Unscrew the ne control valve and ow indicator from the gas can and store safely. If the valve is left attached to the can, the gas could escape.
The piCO+ is now calibrated and ready for use.
● The piCO+ will prompt the user to calibrate the unit again in six months.
● Turn the instrument on in the
normal way. When the main menu appears, scroll down to the symbol of the gas cylinder by single clicking the button. Double click to select the cylinder symbol.
14
piCO+ Operating Manual
English
www.bedfont.com
Single click the button to scroll down to the 20ppm gas cylinder symbol. Double click to select the symbol.
Immediately open the ne control valve and allow the
gas to ow at 1 litre per
minute.
To maintain this, adjust the
ow so the ball in the ow
indicator remains at the middle line.
If the displayed reading is between 5 and 40ppm, the calibration value will be automatically set in the instrument as 20ppm and a tick will be displayed to show a successful calibration. If the displayed reading is outside these limits, the calibration fails, and a will be displayed (see ‘Troubleshooting’, page 17, if this happens).
The pointer and ‘return’ symbols re- appear at the bottom of the display to indicate that the process is complete and to allow the user to double click the button to return to the previous menu.
Turn off the gas ow, remove the D-piece sampling system, and disconnect the calibration adapter from the D-piece
sampling system.
Unscrew the ne control valve and ow indicator from the gas can and store safely. If the valve is left attached to the can, the gas could escape.
The piCO+ is now calibrated and ready for use.
● The piCO+ will prompt the user to calibrate the unit again in six months.
Calibration (If Reset Button Pressed)
● After ‘Reset’ has been pressed the unit
must be switched off and on again. A light sequence will indicate it has been reset and the PiCO+ must be re-calibrated.
● The calibration gas required is Bedfont
20ppm carbon monoxide in air.
Ensure that the ne control valve is in
the off position.
Screw the ne control valve and ow
indicator assembly to the gas can. This is best done by screwing the gas can into the valve.
● Double click to select the
● Ensure that the instrument
is surrounded by fresh air. With the pointer opposite the ‘zeroing’ symbol, double click the button to set the instrument to zero. If the zeroing has been successful, a tick
will be displayed. If the zeroing fails, a will be displayed (see ‘Troubleshooting’, page 17, if this happens). Double click the button to return to the calibration menu.
Allow the gas to ow through the instrument for up to 1.5 minutes to ensure accurate calibration, again
monitoring the rate of ow.
As the 20ppm CO calibration gas is applied, the displayed ppm reading will climb. After 1.5 minutes, or until no further increase in the reading
is obtained, the nal
measured value will be shown and sampling will cease.
15
piCO+ Operating Manual
English
Adult 1
Adolescent 2
Custom 3
Set-up – User proles
In the set-up mode, the user can change the
selected user prole from adult to adolescent (or to a third custom prole, if one has been
set).
The user proles determine which coloured
LEDs are lit during a breath test. The default values for adults and adolescents are shown in the table in ‘Operation’, page 11. The
selected user prole is indicated by the small
symbol on the top line of the display:
To change the selected prole, scroll the
pointer to the set-up symbol on the main display, using single clicks of the button. Double click to select the set-up mode.
Scroll the pointer until it is opposite the ‘User Prole’ symbol . Double click to select this function.
Move the pointer until it is opposite the
desired prole, then double click to select.
The symbol on the top line of the display will
change to show the selected prole.
The custom prole will only be available
on the display if it has been previously set using the COdata+ software on a PC.
Exit this function by pointing at the ‘return’ symbol and double clicking. This returns you to the previous page of the menu.
16
piCO+ Operating Manual
English
www.bedfont.com
Even if the sounder has been turned off, it will continue to operate during the last three seconds of the breath hold countdown.
Exit this function by pointing at the ‘return’ symbol and double clicking. This returns you to the previous page of the menu.
Single click the button to temporarily silence the sounder during or immediately after the breath test, if required.
The set-up mode allows the sounder to be turned off during the breath test, although it will still operate during the breath-hold countdown.
The operation of the sounder during the exhalation phase of a breath test is indicated by the small symbol on the top line of the display.
To turn the sounder settings on or off, scroll the pointer to the set-up symbol on the main display using single clicks of the button. Double click to select the set-up mode – see previous page.
Scroll the pointer until it is opposite the sounder symbol ( ) then double click to switch the sounder on ( ) or off ( ). The large sounder symbol and the small symbol on the top line of the display will change to show whether the sounder is on or off.
Set-up – Sounder operation
17
piCO+ Operating Manual
English
shows a again, the unit will have to be reset or restored using the reset button located in the battery bay followed by the reset calibration procedure on page 13. If the problem persists, please return the unit to Bedfont or your local distributor for investigation and repair. In this case, the display shows a telephone and envelope symbol. The unit can be turned off by pressing the button for three seconds. It is possible to re-start the unit and use it with the previous calibration settings, or to attempt the complete calibration process again.
Troubleshooting
If the unit fails to turn on properly, or if either low battery symbols or are shown, replace both AA batteries. Ensure that the batteries are inserted the correct way around, matching the symbols moulded into the plastic.
If a breath test on a non-smoker gives an indication of CO being present, it may indicate a high ambient level of CO (possibly due to a faulty combustion appliance or exhaust fumes), or the effects of passive smoking. A false reading can, under certain circumstances, be caused by alcohol on a person’s breath, or the presence of hydrogen, which can be generated in the human digestive system.
If, after an attempted calibration with 20ppm CO gas, the display shows a
, the gas value was not within the permitted limits. The achieved reading is displayed underneath the . If the value is much lower than 20ppm, there may have been a problem with the supply of gas from the cylinder during the calibration process. If the displayed value is much higher than 20ppm, it is possible that the wrong concentration of calibration gas is being used. In either case, check the cylinder, connections
and ow rate before double clicking the
button to repeat the calibration process. If, after a third repeated attempt to calibrate the instrument, the display
18
piCO+ Operating Manual
English
www.bedfont.com
COdata+ software
Connecting to the PC
Place one end of the connection lead into the mini USB socket on the top of the piCO+. Connect the other end to the USB port on the PC.
Before starting the software, ensure that the piCO+ is connected to the PC and switched on. Once the sensor has stabilised, double click the COdata+ icon on the PC to start the programme.
Refer to the supplied documentation for how to operate COdata+ software.
The custom user prole can only be accessed from COdata+.
● The operator should not touch the patient whilst in contact with either the computer or USB connector parts. The piCO+ should only be connected to a computer that is manufactered in accordence with EN60950- safety of information technology equipment.
● Ensure the monitor is out of reach of the patient when in operation, at a distance of at least 1.5m
1 Note: the PiCO+ should not be used in
live test mode via COdata+. It should only be connected to the PC for data upload purposes.
19
piCO+ Operating Manual
English
Frequently asked questions (FAQ)
Frequently asked questions about breath carbon monoxide (CO) monitoring in smoking cessation
LED Colour Description User prole 1: User prole 2: Adult (ppm) Adolescent (ppm)
Green Non smoker 0-6 0-4
Yellow Danger zone 7-10 5-6
1 Red Smoker 11-15 7-10
2 Reds Frequent smoker 16-25 11-15
3 Reds Addicted smoker 26-35 16-25
4 Reds Heavily addicted smoker 36-50 26-35
4 Reds ashing Dangerously addicted smoker 51+ 36+
Q What is carbon monoxide?
A Carbon monoxide (CO) is a poisonous
gas that you cannot smell or see. When a smoker inhales smoke from a cigarette, CO is absorbed into their blood through their lungs. CO is dangerous because it binds to haemoglobin in red blood cells about 200 times more readily than oxygen, depriving the body of vital oxygen.
Q What does a breath CO test show?
A It shows the amount of carbon monoxide
in a person’s breath (ppm CO), which is a measure of blood carboxyhaemoglobin (%COHb). It also acts as an indicator of the possible level of around 4,000 toxic substances in cigarette smoke, 60 of which can cause cancer.
Q What does ppm and COHb mean?
A Parts per million; in this case, one part
of CO in one million parts of air (breath). This might not seem like very much, but ppm has a direct correlation with %COHb (the percentage of CO combined with haemoglobin in the blood).
Q How quickly does CO disappear from the
body when a person stops smoking?
A It takes about 5-6 hours to reduce the
original level by half. Usually, after a maximum period of 48 hours, an ex­smoker will have the same level of CO as a non-smoker in the same environment.
Q How long after having a cigarette
should the test be conducted?
A After a minimum of 10 minutes.
Q What level of breath CO do you expect
to see?
A Clinical research has shown that the
popular guidelines are:
Q When is the best time to take a CO
reading?
A The best time to take a CO reading is as
late as possible in the day.
If a patient gives a reading of 20ppm (that of a frequent adult smoker), it means that their oxygen carrying capacity is reduced by 5%. It also indicates a high level of addiction to nicotine.
20
piCO+ Operating Manual
English
www.bedfont.com
Returns procedure
If your equipment requires servicing, please contact Bedfont’s Customer Repairs Depatment before returning any goods. If you did not purchase your monitor directly from Bedfont, please contact your local distributor or supplier.
When you have supplied the Customer Repairs Department with the monitor serial number and description of the fault, you will be issued with a returns number.
Please state the returns number on a compliment slip or headed paper when returning the monitor, and ensure that your full details, including telephone and fax numbers, are clearly stated.
Bedfont advise that you use a courier service when returning monitors. This enables you to insure goods for loss or damage in transit.
When your goods are received, you will be sent an ‘order acknowledgement’.
The monitor will then be examined and you will be sent an ‘Engineer’s Report’ and a quotation for the repair, which will include an authorisation form.
If your monitor is still in warranty, Bedfont will repair it and return it to you with an ‘Engineer’s Report’, free of charge. If the monitor is found to simply require calibrating, you will be charged a fee for this service if you wish to proceed.
If the monitor is out of warranty and you wish to proceed with the repair or calibration, complete the authorisation form within the quotation, and ensure
that you include the ‘Ofcial Purchase
Order Number’. Return the unit to Bedfont. Please contact the Customer Repairs Department if you are unable
to supply an ‘Ofcial Purchase Order
Number’.
If you choose not to proceed with the repair, a handling fee will be charged. Ensure that you return the completed
authorisation form with an ‘Ofcial
Purchase Order Number’.
The equipment will be returned to you as soon as Bedfont has received all relevant paperwork. A carriage fee will be charged if the monitor is no longer in warranty.
A return ticket can also be opened at:
www.bedfont.com/support_ticket
21
piCO+ Operating Manual
English
Spares and warranty
Spares:
D-pieces D-PIECE-2
Disposable SteribreathTM mouthpieces STERIBREATH-MP
Calibration gas and kits 012-08-025II-V & 012-08-025IIK-U
Instrument cleansing wipes WIPE-U
AA alkaline batteries
Hand Foam HANDFOAM
Warranty:
Bedfont Scientic Limited warrants the
piCO
+TM
(excluding batteries) to be free
of defects in materials and workmanship
for a period of ve years from the date of
shipment. Bedfont’s sole obligation under this warranty is limited to repairing or replacing, at its choice, any item covered under this warranty when such an item is returned, intact and prepaid, to Bedfont
Scientic Limited or the local representative.
These warranties are automatically invalidated if the products are repaired, altered, or otherwise tampered with by unauthorised personnel, or have been subject to misuse, neglect or accident.
Never dispose of any electronic instrument or batteries in domestic waste. At the end of the product’s life, contact Bedfont or its distributor for disposal instructions.
Single use consumables and accessories should be disposed of inline with local chemical waste guidelines.
Bedfont Scientic Ltd Station Road, Harrietsham, Maidstone, Kent, ME17 1JA, England Tel: +44 (0) 1622 851122 Fax: +44 (0) 1622 854860 Email: ask@bedfont.com
Notes:
Bedfont Scientic Ltd
ISO 9001:2008
Cert No. FM 31664
ISO 13485:2003
Cert No. MD 502905
Station Road, Harrietsham, Maidstone, Kent, ME17 1JA, England Tel: +44 (0) 1622 851122 Fax: +44 (0) 1622 854860 Email: ask@bedfont.com
www.bedfont.com
Issue 14- June 2014 Part No: LAB220
© Bedfont Scientic Ltd Bedfont Scientic Limited reserves the right to change or update this literature without prior notice. Registered ofce: England and Wales. Registered No: 1289798