The Clay Adams CA6000 centrifuge is a multipurpose centrifuge designed for laboratory use in
performing separation and sedimentation procedures. Through the application of a Relative Centrifugal
Force (RCF) it allows the separation of substances made of different density elements. Potential
applications include hematology, chemistry, urinalysis, blood banking, microbiology, cytology, pharmacology, food processing, and agriculture.
Specific features include:
H Microprocessor controlled
H 5 Program (Memory) Settings
H Speeds from 300 6500 rpm in 10 rpm increments
H High or low acceleration (5 preprogrammed settings)
H High or low deceleration (4 preprogrammed settings plus inertial)
H Timer from 1 minute to 99 minutes plus infinite state
H Electronic-accelerometric imbalance encoder
H Direct drive system
H Brushless induction drive motor
MA01281
Page 7
CA6000 Centrifuge Users Manual
Drive System
The centrifuge is driven by a three phase asynchronous motor. It is situated in the center of a sealed
and armor plated centrifugation chamber.
Safety Interlock System
The CA6000 centrifuge is equipped with an interlock system that prevents opening of the centrifuge lid
when the rotor is spinning. The centrifuge will not operate until the lid is closed and latched in place.
The lid remains latched until the rotor stops spinning.
NOTE
If a power failure occurs, you can access the samples in the centrifuge by
following the Emergency Lid Unlock procedure in Section 4.7.
Imbalance Sensor
The centrifuge is equipped with a load imbalance detector. In case of excessive imbalance the machine
stops and displays the message IMBAL. After the lid has been opened to remove the cause of the
imbalance, the message disappears. Carefully balance the sample load as described in Section 4.2
to avoid activating the imbalance detection system.
Figure 1 Clay Adams CA6000 Centrifuge
MA01282
Page 8
CA6000 Centrifuge Users Manual
1.2Safety Devices and Notices
In order to ensure both mechanical and electrical operational safety, the centrifuge features the following
safety control systems:
F The centrifuge will not start unless the lid is locked. The lid cannot be opened while the rotor is
spinning, even in the event of power failure.
F Safety electronic circuits for self-diagnosis (with microprocessor control); any possible breakdown
of the circuits and of the sensors is shown on the display and neutralized, preventing unsafe
operating conditions.
F Load imbalance detection.
F Double metal protection casing; reinforced centrifugation chamber in AISI 304 stainless steel.
F Lid anti-fall mechanism.
F Main bipolar switch.
F Normalized filters to avoid Radio Frequency Interference (RFI).
F Individual certified test, with measurements and testing of the leakage currents, the ground
resistance, and the applied voltage.
F Identification plates; operating instructions.
1.3Hazards, Precautions, and Limitations of Use
The following CAUTIONS and WARNINGS represent dangerous operations and work conditions to avoid.
!Using the centrifuge if it has not been properly installed.
!Leaning on the machine.
!Depositing any object in the free area of the centrifuge bowl.
!Using the centrifuge with rotating equipment showing evident corrosion, wear marks, or cracking
on the rotating head or on the buckets.
!Using the centrifuge with rotating equipment and accessories not approved by Becton Dickinson.
!Using the centrifuge in explosion risk rooms, or with explosion risk products or chemical materials
subject to violent reaction.
!Incorrect assembly of the rotating head on the driving stud.
!Imbalanced load causing an excessive vibration of the centrifuge.
!Failing to insert all buckets into the rotating equipment, even for partial loads.
!Moving or shifting the machine during centrifugation.
WARNING
IF ANY DANGEROUS EVENT OCCURS INVOLVING THE CENTRIFUGE,
VACATE THE AREA AND SWITCH OFF THE MAIN POWER SOURCE.
MA01283
Page 9
CA6000 Centrifuge Users Manual
!Exceeding the maximum speed indicated in the max. rpm column of the performance table (see
4.2.2).
!Using old accessories on a new machine.
!Using tubes or bottles not suited for centrifugation.
!Centrifuging samples with higher density than allowed for the maximum speed, without reducing
the speed limit (see Section 4.2).
!Manipulating or tampering with the electronic and mechanical parts.
1.4Rotor and Accessory Precautions
1.4.1Corrosion Information
CA6000 rotors made of aluminum alloy are designed to operate at their rated RCF for many years. With
careful use they will resist corrosion, lessening the possibility of excessive imbalance, disruption and
subsequent damage to the instrument. The primary conditions for the initiation of corrosion exist in every
laboratory during daily use of the centrifuge. For this reason it is essential that care and attention be paid
to inspection and cleaning.
CHEMICAL CORROSION
This corrosion is characterized by chemical reactions due to the existence of any electrolyte liquid on
the surface of the rotating equipment. If these substances are allowed to remain on the surface, the
corrosion will almost certainly occur. This produces first a discoloration of the anodization and
eventually pitting of the metal.
Acidic and alkaline solutions sustaining their pH level will create problems of corrosion in aluminum
equipment. Chlorides, present in salts or even in skin contact with the rotating equipment, are among
the most aggressive and harmful substances commonly found in the laboratory.
Chemical products, which are the source of this corrosion do not necessarily originate from broken
tubes, for example they could come from:
H Chemical vapors present in the laboratory which are dissolved in the residual humidity, in
condensed water (refrigerated centrifuges) present at the base of the rotor pockets.
H Corrosive liquids originating from overfilled uncapped tubes (the liquid overflowing during
centrifugation)
H Inserts, adapters, racks, bottles whose exterior has been soiled by a chemical product or
poorly rinsed after decontamination (with bleach, for example)
NOTE
If the products are very corrosive, simply rinsing is insufficient. Residual
traces dissolve little by little with the humidity present in the bottom of the
rotor pocket.
CAUTION
BEWARE of the presence of solid particles beneath tubes, inserts,
racks, or adapters. These particles are crushed by the centrifugal
force and penetrate the protective, anodized layer of buckets and
rotors, thus creating easy pathways for corrosion.
MA01284
Page 10
CA6000 Centrifuge Users Manual
STRESS CORROSION
This term relates to the phenomenon of accelerated corrosion due to the effect of centrifugation when
a corrosive chemical is in contact with the alloy. From the time when the aluminum alloy has been
attacked by chemicals, stress corrosion begins to appear. Because the process begins on a
microscopic scale it is even more dangerous than a macroscopic process since it is invisible.
During centrifugation, chemicals responsible for corrosion are also submitted to the very high g force
which pushes them against the alloy. This close contact facilitates the chemical reaction which occurs
much faster than in a static situation. Moreover, centrifugal force is very directional thus corrosion
under stress creates, with a very small amount of corrosive product, straight microscopic fissures.
Each centrifugation run makes the chemical migrate further and further.
Fissures or cracks, although microscopic, are cuts in the metal, breaking the materials cohesion. As
one weak link in a chain allows the chain to break, so the microfissures break the chain of resistance
of the accessory to centrifugal force. As accessories are designed with high levels of safety, rupture
does not occur as soon as the first microfissures are produced.
Depending upon the location of the fissure, disruption may occur before it reaches the external surface
of the accessory. The fissure creates a weakness which makes the accessory less and less resistant
to mechanical fatigue. The corrosion by a small amount of corrosive product does not disrupt the
accessory but makes it mechanically weaker and weaker until disruption occurs due to both
centrifugal effort and the number of cycles.
Because stress corrosion is largely invisible, it is essential that rotating equipment be
scrutinized regularly paying particular attention to susceptible parts such as the bottom of
pockets, the outer edges and the base of the rotating equipment.
1.5Contamination Hazards
Clay Adams centrifuges are likely to be used in facilities where hazardous substances (including
radioactive chemicals) are frequently present.
WARNING
BLOOD AND BODY FLUIDS MAY CONTAIN THE HEPATITIS B VIRUS
(HBV), HEPATITIS C VIRUS (HCV), HUMAN IMMUNODEFICIENCY VIRUS
(HIV), OR OTHER DISEASE-CAUSING AGENTS. HANDLE ALL PATIENT
SPECIMENS AS POTENTIAL BIOHAZARDS CAPABLE OF TRANSMIT-
TING INFECTION. WEAR APPROPRIATE PERSONAL PROTECTIVE
EQUIPMENT, INCLUDING LABORATORY GLOVES, WHEN COLLECT-
ING, HANDLING, AND PROCESSING BLOOD AND BODY FLUIDS.
IN ADDITION TO WEARING GLOVES, THE USE OF DISPOSABLE
LAB COATS OR GOWNS AND PROTECTIVE GLASSES OR
GOGGLES IS RECOMMENDED WHEN WORKING AROUND THE
CENTRIFUGE.
IF A TUBE BREAKS IN THE UNIT, CAREFULLY REMOVE BROKEN
GLASS WITH A HEMOSTAT OR OTHER DEVICE, USING PUNCTURE
RESISTANT UTILITY GLOVES.
Always use the appropriate decontamination procedures when the rotating equipment is exposed to these
chemicals. Examples of commonly used techniques are outlined in Section 5 Maintenance. The
information is given as a guide only. Decontamination is the responsibility of the owner to use the most
suitable procedure.
MA01285
Page 11
CA6000 Centrifuge Users Manual
!
The rotating equipment should always be completely disassembled before being subjected to heat and after
external chemical cleaning. Seals, tubes, and plastic components should be decontaminated with the
method most suitable for them (which might not be the same as for the rotating equipment).
1.6Manual Structure
This users manual contains the following sections:
Section 1 Introduction provides an overview of the CA6000 centrifuge and its uses in the laboratory.
Section 2 Installation and Setup gives specifications for installation of the centrifuge and
instructions for setup.
Section 3 Controls and Indicators explains the use and meaning of all controls and indicators
Section 5 Maintenance explains all user system maintenance, including parts replacements.
Section 6 Troubleshooting provides a convenient guide for identifying and correcting system
malfunctions.
The Appendices contain warranty information, parts and accessories list, a listing of Becton Dickinson
international contacts, and an exploded view drawing.
1.7Use of this Manual
This users manual is designed as a reference tool for technologists, supervisors, and other personnel who
operate and maintain the CA6000 centrifuge on a regular basis. Every attempt has been made to include
all information which would be required during normal use and maintenance. Should a question arise which
is not answered in this manual, please contact the following parties (USA):
For assistance with mechanical or electrical performance problems:
( Field Service 18005447434
For procedural questions:
( Technical Services 18006318064
International contacts are listed in Appendix D.
1.8Manual Conventions
1.8.1Symbols Used on the Equipment
Figure 2 Symbols Used on the CA6000 Centrifuge
Left figure: Symbol for electrical hazard; Middle figure: Symbol for refer to accompanying
documentation (specifically, the users manual) for instructions; Right figure: Symbol for fuse.
MA01286
Page 12
CA6000 Centrifuge Users Manual
1.8.2Notes, Cautions, and Warnings
Throughout this manual, important information is presented in boxes offset from the regular text, and is
labeled as either a NOTE, CAUTION, or WARNING. These messages are formatted as shown below and
bear the following significance:
NOTE
Important information about system use worthy of special attention is
presented as a NOTE.
CAUTION
Information on an activity which potentially could cause damage to
the centrifuge is presented as a CAUTION.
WARNING
INFORMATION ON AN ACTIVITY WHICH POTENTIALLY COULD
CAUSE INJURY TO THE USER IS PRESENTED AS A WARNING.
MA01287
Page 13
Installation and Setup
2.1Transport, Packaging, Lifting
Due to the weight of the CA6000 centrifuge, all lifting and transporting must be carried out using proper handling
equipment (e.g., forklift) that complies with current regulations, and by people with the necessary training.
The centrifuge may be transported and stored in ambient temperatures from 20° C to 50° C, and relative
humidity up to 90% (non-condensing).
During transport the centrifuge is seated in a special packaging. Unpack the centrifuge, carefully removing
first the rotating equipment, any possible accessories and the material supplied for ordinary maintenance.
Check the contents of packaging against the packing list and note any damage to the shipping carton. If
damage is observed, notify the carrier immediately.
Keep the packaging until the centrifuge has been set up and tested.
WARNING
DUE TO ITS SIZE AND WEIGHT, AT LEAST TWO PEOPLE ARE
REQUIRED TO LIFT THE CENTRIFUGE FROM THE PALETTE AND
PLACE IT ON THE BENCH.
2.2Installation
2.2.1General Placement
The centrifuge should be installed in a dust and corrosion free room.
Place the centrifuge on a bench top capable of supporting the specified weight (see Section 2.3). Be sure
there is approximately 12 in. of clearance on all sides of the unit.
MA01288
Page 14
CA6000 Centrifuge Users Manual
2.2.2Inspection
Before installation, the centrifuge should be thoroughly inspected for corrosion and cleanness (see Section
5 Maintenance). The clip in the center of the rotor and the drive spindle should also be clean and
undamaged: these parts should be wiped over before each use.
Any failure to follow the above advice may have serious consequences for the safety of the appliance.
2.2.3Electrical Connection
Attach the power cord to the receptacle on the rear panel of the centrifuge. Plug the power cord into the
appropriate 3-wire grounded power receptacle. If an extension cord is required, use only a 3-wire grounded
cord rated for the correct voltage for your unit (120 VAC/15 A or 220 VAC/7.5 A).
WARNING
TO AVOID ELECTRICAL SHOCK, CONNECT THE POWER CORD ONLY
TO AN APPROVED POWER SOURCE SUCH AS A 3-WIRE GROUNDED
RECEPTACLE. WHERE A 2-WIRE RECEPTACLE IS ENCOUNTERED,
HAVE IT REPLACED WITH A PROPERLY GROUNDED 3-WIRE RECEP-
TACLE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE.
DO NOT, UNDER ANY CIRCUMSTANCES, REMOVE THE GROUND
PRONG FROM THE POWER PLUG.
SHOULD THE POWER CORD OR PLUG BECOME CRACKED, FRAYED,
BROKEN, OR OTHERWISE DAMAGED, REPLACE THE CORD OR
UNPLUG THE POWER CORD BEFORE ANY MAINTENANCE OR SERVICING.
NEVER ATTEMPT TO OVERRIDE THE ELECTRICAL SAFETY INTER-
2.2.4Lid Unlock
1Switch the centrifuge on by flipping the power switch on the right side of the machine to the on
position.
2When the centrifuge is switched on and not spinning the NOT IN USE LED will be lit.
3To open the lid, pull the lever on the right side of the machine towards the centrifuge front.
BEFORE PROCEEDING, REMOVE POWER TO THE CENTRIFUGE BY
4After opening the lid, remove any packaging around the driving spindle. Unscrew the rotor head nut
from the motor shaft. Carefully clean the inside of the centrifuge chamber removing any packing
residues. Due to the air-whirling, solid particles accidentally left in the centrifuge chamber would
create excessive wear of the chamber itself and of the rotors.
PLUG IMMEDIATELY.
LOCKS OF THE CENTRIFUGE.
WARNING
MOVING THE POWER SWITCH TO THE OFF POSITION.
2.2.5Assembly of the Rotating Head
The rotors drive is made by means of two pins inserted in the rotors shaft which must be aligned with the
milling of the rotor. For a correct mounting follow the instructions below:
MA01289
Page 15
CA6000 Centrifuge Users Manual
Ensure that both the spindle of the motor and the hole of the head are clean. It is suggested that
you apply a light coat of lubrication with a light oil or silicone spray.
Insert the head in the centrifuge protective bowl and slip it onto the spindle with the milling of the
collar face down.
Align the milling in the collar with the pins on the spindle.
Screw and slightly tighten (with the supplied key) the hexagonal nut on the threaded spindle of the motor
unit.
This section describes the meaning and use of the controls and indicators of the Clay Adams CA6000
centrifuge. The overall layout of the unit is shown in Figure 3. The keypad is shown in Figure 4.
WARNING
ALL SYSTEM USERS MUST BECOME THOROUGHLY FAMILIAR WITH
ALL CONTROLS AND INDICATORS BEFORE ATTEMPTING TO
OPERATE THE CENTRIFUGE.
3.2On/Off Switch
The On/Off switch is located on the right side of the centrifuge, at the bottom. When in the | (On) position,
power is applied to the centrifuge. When in the O (Off) position, power is removed from the centrifuge.
MA012812
Page 18
CA6000 Centrifuge Users Manual
Figure 3 CA6000 Centrifuge Layout
3.3Lid Release Latch
The Lid Release Latch is located on the right side of the centrifuge, at the top front. Pull this latch forward
to open the centrifuge lid.
Note that the Lid Release Latch does not operate when power to the unit has failed.
3.4Emergency Lid Release Access Screw
The emergency Lid Release Access Screw is located on the right side of the centrifuge, just in front of the
Lid Release Latch. In the event of a power failure, remove the screw to access the emergency lid release
function (see Section 4.7).
3.5CA6000 Keypad
The keypad is located at the center of the front panel. It presents all the keys and displays necessary to
program, start, and stop operation of the centrifuge.
3.5.1RPM Readout
When the centrifuge is stopped, the RPM Readout displays the speed setpoint (in revolutions per minute).
It displays the actual speed when the centrifuge is spinning. When the readout is flashing, it indicates that
the current speed/time setting can be entered into memory (Program 1 5).
3.5.2RPM Set Keys
Press p to increase or q to decrease the speed setting. The speed can be set from 300 to 6500 rpm
in 10 rpm increments.
MA012813
Page 19
CA6000 Centrifuge Users Manual
3.5.3NOT IN USE Indicator
When lit, this amber LED indicates that the rotor is not spinning and the lid can be opened.
3.5.4IN USE Indicator
When lit, this amber LED indicates that the rotor is spinning and the lid cannot be opened.
3.5.5Time (MIN) Readout
When the centrifuge is stopped, the time readout displays the time setpoint (in minutes). It displays the
remaining time when the centrifuge is spinning. When the readout is flashing, it indicates that the current
speed/time setting can be entered into memory (Program 1 5).
3.5.6Time Set keys
Press p to increase or q to decrease the time setting. The time can be set from 1 minute to 99 minutes.
The Hold (Hd) setting, located at the zero position, allows you to manually start and stop the centrifuge
without presetting a time (i.e., spin for an indeterminate or infinite time).
3.5.7SOFT ACCEL Key
This key is used to select soft acceleration modes. Press and hold the key until the RPM Readout shows
ACL flashing and a number from 1 5. Use p to increase or q to decrease the acceleration profile
number. Profile 1 is the slowest setting, increasing to Profile 5, the fastest setting.
3.5.8SOFT ACCEL Indicator
This green indicator flashes when the centrifuge is accelerating to its setpoint. The LED remains on while
the actual speed is at the setpoint.
Figure 4 CA6000 Centrifuge Keypad
MA012814
Page 20
CA6000 Centrifuge Users Manual
3.5.9SOFT BRAKE Key
This key is used to select soft deceleration modes. Press and hold the key until the RPM Readout shows
dCL flashing and a number from 0 4. Use p to increase or q to decrease the deceleration profile
number. Profile 0 represents a coast mode (no braking).
3.5.10 SOFT BRAKE Indicator
This green indicator flashes when the centrifuge is decelerating from its setpoint. The LED remains on while
the actual speed is at the setpoint.
3.5.11 START key
Press to start a centrifuge cycle.
3.5.11 STOP/PROGRAM LOCK Key
Press this key to stop the centrifuge during its spin cycle.
Press this key ALONG WITH a PROGRAM Key to lock the program setting. When a program is locked,
it cannot be changed unless it is unlocked.
3.5.12 PROGRAM Keys
When the RPM and Time (MIN) readouts are flashing, pressing one of the five PROGRAM keys saves the
current speed/time settings in memory. When RPM and Time readouts are not flashing, pressing
PROGRAM 1 5 recalls the speed/time settings stored in that memory location.
You can lock any of the programs while you are setting it by pressing the STOP/PROGRAM LOCK Key
along with the desired PROGRAM Key. When locked, a program cannot be changed. To unlock a program,
press the STOP Key along with the appropriate PROGRAM Key.
3.5.13 PROGRAM Indicators
When a preset program (1 5) is selected, the amber LED next to the PROGRAM number lights.
MA012815
Page 21
Operation
WARNING
USE ONLY THE ROTORS, BUCKETS, INSERTS, AND OTHER ACCES-
SORIES SPECIFIED IN APPENDIX B PARTS AND ACCESSORIES.
DO NOT OPERATE THE CENTRIFUGE WITH SPECIMENS OR MATE-
RIALS HAVING A DENSITY GREATER THAN 1200 KG/M3. DOING SO
MAY CAUSE DAMAGE TO THE CENTRIFUGE, AND MAY CAUSE
UNSTABLE OPERATION AND CAUSE THE CENTRIFUGE TO MOVE
OFF OF THE OPERATING SURFACE.
DO NOT OPERATE THE CENTRIFUGE IN ATMOSPHERES CONTAIN-
ING FLAMMABLE OR VOLATILE GASES. DO NOT USE THE CENTRI-
FUGE TO SPIN DOWN FLAMMABLE OR VOLATILE LIQUIDS.
4.1CA6000 Centrifuge Setup for Operation
Turn the On/Off Switch to the On (|) position.
The CA6000 will be in the same operative mode as the last time it was used.
Open the lid by pulling the Lid Release Latch (located on the upper right side of the centrifuge) toward the
front of the centrifuge.
MA012816
Page 22
CA6000 Centrifuge Users Manual
4.2Loading Balance
All rotating equipment is dynamically balanced at the source. Reference numbers and marks indicate the
exact respective positions of the various parts forming the equipment; when rotating, this should always
be whole and complete.
All buckets must be always inserted into the rotating fixed-angle equipment, even for partial loads.
WARNING
NEVER ATTEMPT TO BALANCE THE LOAD BY ADDING WEIGHTS,
MERCURY, OR METAL SHOT TO THE BOTTOMS OF TUBES OR
SHIELDS. ONLY THE PROPER BUCKETS, INSERTS, AND TUBES
SHOULD BE PLACED IN THE CENTRIFUGE.
Distribute the load correctly. Loads should be inserted two by two, of equal weight, at 180° from the rotation
axis and at equal distance (radius) from the center of rotation. Use matching inserts when balancing the rotor.
When spinning an odd number of samples, even the load by inserting a balance tube filled with water.
For partial loads of swing-out rotating equipment, it is advisable to distribute the tubes symmetrically also
with regard to the center of rotation.
Correct
Sample
Placement
in Insert
Incorrect
Sample
Placement
in Insert
Figure 5 Loading Balance
Avoid overloading the rotating equipment. To centrifuge liquids with density higher than 1200 kg/m3, the
maximum rpm should be reduced. Calculate the new allowable rpm using the following formula:
New allowed rpm = max rpm *1200
√
New Density
Lower the lid, then close it by pressing on the front edges until the hook in the lid clicks into the lid lock
mechanism. Only this operation will activate the centrifuge starting device.
When using the aerodynamic rotating heads, all tubes must be balanced with ± 2 grams, in order to obtain
the best separation.
The main factors of the centrifuge action in terms of correct analysis methodology are: the centrifugation
time and the gravity number xg (relative centrifugal force) to which the product is submitted. The formula
allows the determination, as a function of the radius r, of the relative centrifugal force value, expressed in
multiples of the gravity acceleration value g (9.9 m/sec
noitisopmoCdaoLseirosseccA
gnilaeS
paC
2
) for each rpm value between 150 and 20,000.
suidaR.xaMlufesU
)m(.ni
The relative centrifugal force and the RPM number are connected by the following relations:
RPM in thousands =
RCF
√
whereRCF = relative centrifugal force specified for the particular centrifugation procedure
r = Tip radius of tube (in inches) obtained from the chart above
28.4r
4.4Selecting Operating Modes
Instructions for setting the following parameters are provided in this section:
H Rpm centrifugation (300 rpm up to 6500 rpm)
H Centrifugation time (1 minute up to 99 minutes, plus Hold)
H Acceleration (Profiles 1-5)
H Deceleration (0-4 Coast-down; 0 inertial)
MA012818
Page 24
CA6000 Centrifuge Users Manual
Setting Acceleration
To select an acceleration profile, use the following procedure:
1Press and hold the SOFT ACCEL key for approximately 3 seconds, until the rpm display shows
ACL flashing and a number between 1 and 5.
SOFT
ACCEL
2Use the qp keys to select a profile. 1 is slowest.
High acceleration (LED off)
Low acceleration (LED on)
Setting Deceleration
It is possible to select various low coast-down deceleration curves, corresponding to several profiles
of decreasing speed.
1Press and hold the SOFT BRAKE key for approximately 3 seconds, until the rpm display shows
dCL flashing and a number between 0 and 4.
SOFT
BRAKE
2Use the qp keys to select a profile. 0 is coast mode (no braking).
High deceleration (LED off)
Low deceleration (LED on)
Setting Rpm, Time (MIN) Profiles
To set rpm or cycle time, press and hold the q or p key located under the respective parameter.
Release the key when the desired value is reached. The display will flash for 3 seconds after the
adjustment key is released and then stop, indicating that the new parameter is accepted.
How to Set and Memorize a Program
To store the parameters that appear on the display, press a PROGRAM key from 1 to 5 while the
parameter is flashing (the display flashes when setting rpm, minutes, soft acceleration, or deceleration). When the LED next to the program key is lit, the program has been memorized.
It is not possible to set a program with unspecified Hold (Hd) time: the timer display will continue
flashing; you must insert a determined time profile.
To restore manual mode, press again the selected program key; the LED goes off and you can make
new parameter selections.
How to Protect a Program
1To protect a program from overwriting, when you memorize a new program (pressing a PROGRAM
key during flashing phase) you must press simultaneously STOP/PROGRAM LOCK key.
2To cancel the program protection, press the PROGRAM key together with the STOP/PROGRAM
LOCK key.
MA012819
Page 25
CA6000 Centrifuge Users Manual
4.5Starting the Centrifuge
1Make sure the lid is closed.
2Press the START key.
3The Centrifuge accelerates to the set speed.
The yellow NOT IN USE LED goes off, and the IN USE LED lights.
The green SOFT ACCEL LED starts flashing. This LED stays on when the set speed equals the
actual speed. (Note that if the set rpm is higher than the maximum speed of the selected rotor (see
Section4.3), the LED continues flashing.)
4During the run time all displays show real parameters: Rpm, Time (min elapsed run time in
minutes counts down or remaining run time counts up if you have set a time of Hold). If you are
running a Program, you can show set values by pressing the corresponding arrow key.
If your parameters have been set manually it is possible to change them. To change the rpm or time,
use pq. To change the deceleration mode, press and hold the SOFT BRAKE key for about 3
seconds. The Readouts start to flash and show the set values for 3 seconds. While the Readout
flashes it is possible to change set values.
5When the run time is completely elapsed, deceleration will take place.
4.6Stopping the Centrifuge
Once the preset time has elapsed, the centrifuge automatically stops with the preset deceleration mode.
However, in any situation you can press the STOP key to stop the centrifuge.
Note that the cycle will be interrupted automatically for load imbalance, power failure, system error, or
breakdown.
If this occurs, the relevant information appears on the message display.
4.7Emergency Lid Unlock
In the event of a power failure, the lid can be opened using the special tools supplied with the centrifuge.
This operation should be performed with the power disconnected.
1Unscrew the nylon screw in the hole on the right side of the centrifuge, just in front of the Lid Release
Latch.
2Insert the unlocking tool supplied with the centrifuge in the hole.
3Pivot the unlocking tool downward, and at the same time pull the Lid Release Latch forward.
4Reinstall the nylon screw
WARNING
SHOULD A POWER FAILURE OCCUR DURING CENTRIFUGATION,
WAIT FOR TEN MINUTES. IF THE ROTOR IS STILL SPINNING,
CLOSE THE LID AND WAIT ANOTHER TEN MINUTES BEFORE RE-
PEATING THE OPERATION.
MA012820
Page 26
Maintenance
WARNING
DURING MAINTENANCE OPERATIONS, ALWAYS DISCONNECT THE
CENTRIFUGE FROM THE POWER SOURCE. OTHER THAN ROTOR
AND EXTERNAL FUSE REPLACEMENT, REFER SERVICING TO
QUALIFIED PERSONNEL.
5.1Disassembly of the Rotating Head
Remove the head nut securing the rotor to the motor shaft. Lift the rotor straight up.
5.2Rotating Equipment and Centrifugation Chamber Cleaning
Regular cleaning of the rotating heads and centrifugation chamber is vital for correct operation of the
centrifuge.
At a minumum the unit should be cleaned weekly in warm water containing a few drops of mild nonalkaline
detergent. The centrifuge should always be cleaned immediately after a spillage has occurred. Do not forget
to wash the core of the rotating head which comes into contact with the drive spindle. Each pocket of the
rotating heads, slots, and buckets must be washed thoroughly using a small nylon brush.
MA012821
Page 27
CA6000 Centrifuge Users Manual
CAUTION
Do not use metal wire brushes to clean the centrifuge.
Once the rotating equipment is clean, rinse it in running water, preferably distilled. Dry the rotating
equipment with a soft absorbent non-woven cloth or tissue. Drying may be finished off with warm air jet (e.g.,
a hair dryer).
MAKE CERTAIN THAT ROTATING HEADS, SLOTS, AND BUCKETS ARE WELL DRIED.
NEVER LEAVE damp rotating equipment on a metal surface, particularly stainless steel. An electrochemical reaction could take place with the aluminum or magnesium in the rotor.
For swing-out rotating equipment, be sure to clean the dirty grease from the studs and replace it with a small
amount of fresh grease. This will ensure that the buckets swing freely. The vast majority of apparent
imbalance problems arise from the failure of the user to clean and grease the studs, not from instrument
error.
Never apply grease if studs and balancing slots have not been carefully cleaned. Use only the same kind
of grease furnished in the rotor head packaging.
Once the rotating head has been removed, it is easy to clean the centrifugation chamber (do not use
corrosive solvents or detergents).
5.3Exterior Cleaning
For a correct cleaning a light domestic use detergent is recommended. A mixture of 50% isopropanol and
water is suggested because it will not damage the labels.
5.4Disinfection
Autoclave rotating equipment for 20 minutes at 120° C to destroy microorganisms.
Rotor lids must be disassembled from rotor bodies. O rings, autoclaved separately, should be replaced
when deformed.
Any part which has been subjected to temperatures above 130° C must be discarded.
Household bleach used at 10% concentration with 10 minutes immersion is effective against bacteria,
spores, and viruses but, as an oxidizing agent, is corrosive to metal alloys and must be thoroughly rinsedoff and dried. It should never be used if there is surface damage to the rotating equipment.
Note that the black color on the surface or the rotating equipment will be gradually washed out if the rotor
is regularly autoclaved or bleached. This does not necessarily denote a degradation on the anodization.
If there is a biological liquid overflow there may be the possibility of a discharged into the drain hose. Do
not forget to proceed to the decontamination with the same household bleach (10%).
For any other unscheduled cleaning, the customer must contact Becton Dickinson to verify that the
proposed method does not damage the centrifuge.
NOTE
Becton Dickinson makes no claims as to the effectiveness of proprietary
brands or decontaminating solutions.
MA012822
Page 28
CA6000 Centrifuge Users Manual
5.5Radioactive Decontamination
We recommend that all radioactive decontamination be referred to your Radioactivity Safety Officer.
After decontamination, the unit should be cleaned as described above (Section 5.4).
5.6Fuses
There are two fuses in the power entry module. The fuses are located in a drawer just below the power cord
plug.
These fuses are:
Diameter 5 x 20 mmF6A/250V
WARNING
BEFORE BEGINNING TO CHANGE THE FUSES, MAKE SURE THE
UNITS POWER IS TURNED OFF AND DISCONNECT THE CENTRI-
FUGE POWER CORD FROM THE POWER SOURCE.
In case of fuse blowing, do not replace them until the machine has been inspected by a maintenance
engineer.
MA012823
Page 29
Troubleshooting
6.1Alarm Messages
Below is a list of error messages and their meaning.
Please refer all repairs to a qualified Biomedical Engineer.
rodionelosdilevitcefeD
riapeRtnemurtsnInosnikciDnotceBllaC
MA012825
Page 31
Limited Warranty
f
Becton Dickinson warrants the Clay Adams CA6000 Centrifuge to be free from defects in workmanship and
materials for a period of 18 months from date of installation, provided the Centrifuge is operated in
accordance with this Users Manual. During such period, Becton Dickinson agrees to ship replacement
parts at no additional costs to the user which, in Becton Dickinsons sole judgment, are found to be
defective, provided the Centrifuge has not been subjected to misuse or abuse. The warranty stated herein
shall extend to the original consumer only and not to any subsequent consumer of the Centrifuge.
Becton Dickinson shall not be liable for any incidental or consequential damages. Becton Dickinson makes
no other warranties, expressed or implied, except as stated herein.
MA012826
Page 32
Parts and Accessories
g
metIrebmuNgolataC
stekcuB4htiw,tuOgniwS,rotoR10010624
paCgnilaeS20010624
ylnOrotoR11010624
ylnOtekcuB90010624
)8yticapaClatoT(lm05x2,tresnI30010624
)63yticapaClatoT(lm01/51x9,tresnI40010624
)61yticapaClatoT(lm51x4,tresnI50010624
)67yticapaClatoT(lm5x91,tresnI60010624
)84yticapaClatoT(lm7x21,tresnI70010624
)84yticapaClatoT(lm5x21,tresnI80010624
ecalP6elgnAdexiF,rotoR62010624
)6yticapaClatoT(lm05x1,tresnI72010624
)6yticapaClatoT(lm51/61x1,tresnI82010624
)6yticapaClatoT(lm5.1/0.2x1,tresnI92010624
MA012827
Page 33
CA6000 Centrifuge Users Manual
)rebmuNelbbuBweiVdedolpxE(metIrebmuNgolataC
)9(laeSlwoB/diL21010624
)8(laeSlwoB/rebmahCrotoM31010624
)21(rotoM41010624
)11(draoBCPyalpsiD51010624
)31(rosneSecnalabmIdaoL61010624
)7(ylbmessAkcaRylppuSrewoP71010624
)3(dapyeK81010624
)01(draoBCPrewoProssecorporciM91010624
)2(hctiwSrewoP12010624
)61(eludoMyrtnErewoP22010624
)4(gnirpSsaG32010624
)91(metsySgnikcoLdiL42010624
launaMs'resUegufirtneC0006ACsmadAyalC52010624
MA012828
Page 34
International Contacts
h
Becton Dickinson de
Mexico, S.A. de C.V.
Monte Pelvoux 111 9th Floor
Col. Lomas de Chapultepec
11000 Mexico D.F.
Voice: 52 5 237 1200 Fax: 52 5 237 1287