Becton Dickinson avea AVEA ventilator user guide critical care ventilation

Page 1
AVEA® ventilator user guide
Critical care ventilation
Page 2
Table of contents
AVEA® front panel..........................................1
AVEA rear panel .............................................2
Circuit assembly .............................................3
Disposable filter and water trap insertion ....... 6
Patient circuit assembly .................................. 7
Attaching flow sensors ................................8-9
User verification test ....................................10
Performing an EST test ........................11-13
Setting the ventilation mode ........................ 14
Graphics ......................................................15
Loops ........................................................... 16
Configure Loops screen ...............................17
Freeze ..........................................................18
Saving loops ................................................19
Create reference loops .................................20
Rear panel connections ................................ 27
Ventilator operations ....................................28
The AVEA ventilator User Guide is not intended as a replacement for the operator’s manual. You must become completely familiar with the AVEA ventilator operator’s manual before using the AVEA ventilator.
Connecting gas sources ............................... 37
Touch-Turn-Touch™/Touch-Turn-Accept™ .......38
New patient setup .......................................39
Setting the breath type and mode ................ 40
Infant mode select screen ............................41
Setting nCPAP/IMV ...................................... 42
Setting the breath type and mode ................ 44
Breath type and mode .................................45
Monitored parameters ................................. 46
The Monitor screen ......................................47
Graphics ......................................................48
Waveforms .................................................. 49
Events ..........................................................50
Trends ..........................................................51
Synchronized nebulizer ................................ 52
Alarms ......................................................... 53
Notes ......................................................55-57
i
Page 3
AVEA ventilator front panel
C
E
A
D
A. Membrane buttons
B. Data dial
C. Screen indicators
D. Monitored parameters
E. Screen graphics
F. Primary controls
G. Message window
H. Patient type
F
B
H
G
1
Page 4
AVEA ventilator rear panel
B
KC
A. AC power module
B. UIM connection
C. Analog input /output/ILV
D. Power on/off switch
A
E. Nurse’s call jack
J
F. Air/heliox smart connector
I
E
G. Oxygen sensor
D
H. Oxygen connection
I. External battery
H
connector
F
J. External battery fuse
K. Internal battery fuse
G
2
Page 5
Circuit assembly
Assembling the exhalation filter and water trap
1.
2.
3
Page 6
3. Align the locating ridge on the water trap assembly with the slot in the exhalation filter cartridge.
4. Slide the water trap/exhalation filter assembly into the cartridge.
4
Page 7
5. Open locking lever. 6. Insert exhalation filter.
5
Page 8
Disposable filter and water trap insertion
AVEA ventilator di spos able water tra p.
Inser t car tridge into filter cav ity. Close lever whe n fully in ser ted.
6
Cart ridge f ully ins erted with lever clo sed.
Page 9
Patient circuit assembly
The passive humidification system should be placed in-line in the patient circuit per the manufacturer’s instructions.
Patient circuit with active humidifier Patient circuit without active humidifier
Inspiratory limb of patient circuit
7
Page 10
Attaching flow sensors
The AVEA ventilator can accept either a hot wire or a variable orifice proximal flow sensor. The monitored values displayed for volume and flow is proximal values when a proximal sensor is in use. These are in addition to the instrument’s internal inspiratory flow sensor and heated expiratory flow sensor.
Variable orifice flow sensor Hot wire flow sensor
Retractable plastic collar
These are locking connectors. To attach, first pull back the plastic locking collar, then push firmly onto the ventilator receptacle. Then push the locking collar forward to lock the flow sensor in place.
To disconnect, first retract the plastic collar then firmly pull the connector away from the ventilator. Do not pull up or down as this can damage the connector.
8
Page 11
Attaching flow sensors (continued)
The standard hot wire flow sensor is suitable for neonatal and pediatric applications where the peak inspiratory flow rate is less than 30 L /min. This flow sensor is not active in adult applications.
NOTE: Hot wire flow sensors will not function with Heliox gas mixtures. During Heliox delivery, a variable orifice flow sensor should be used for monitoring delivered volumes at the proximal airway. Variable orifice flow sensors are also available on some AVEA ventilator models. The neonatal VariFlex flow sensor is compatible in neonatal and pediatric applications where the peak inspiratory flow rate is less than 30 L/min and is not active in adult applications. For adult and large pediatric applications, a pediatric / adult VariFlex flow sensor is available for use with patients whose flow requirements fall within the range of 1.2 to 180 L/min. Detailed information on the specifications of each flow sensor can be found in Appendix E: Sensor specifications and circuit resistance of the AVEA ventilator operator’s manual.
Hot wire flow sensor must be zeroed prior to use. Follow operator manual instructions.
9
Page 12
User verification test
The User Verification Test consists of three separate elements:
1. Post Test: Automatic on power up. Normal ventilation begins after completion.
2. Extended System Test (EST)
• Patient circuit leak test
• Patient circuit compliance measurement
• Two-point calibration of O2 sensor
3. Manual Alarms Testing: Detailed instructions are found in Ch. 2 of the AVEA ventilator operator’s manual.
10
Page 13
User verification test (continued)
Performing an Extended Systems Test (EST):
The EST is accessed from the Setup screen. To open this screen, press the Setup button on the lower left of the user interface. Select the appropriate Patient Size and proceed to second menu of Setup screen.
Press the EST button to access the EST screen
11
Page 14
Performing an Extended Systems Test (EST) (continued)
Remove the patie nt and bl ock the patient wye, then p ress the Continu e (Cont) but ton.
All tests are performed simult aneou sly. The m aximu m time for the EST is 9 0 seco nds. To restar t the ES T at any tim e, selec t the Cancel bu tton to return to the Setup screen.
Press the Conti nue button to return to the Setup screen.
When possible the EST should be performed with an oxygen supply connected. A two point calibration of the oxygen sensor is performed during the EST.
NOTE: If you do not connect the ventilator to an oxygen supply, the O
sensor calibration
2
will immediately fail.
CAUTION: Although failure of any of the above tests will not prevent the ventilator from functioning, it should be checked to make sure it is operating correctly before use on a patient.
12
Page 15
Performing an Extended Systems Test (EST) (continued)
When possible the EST should be performed with an oxygen supply connected. A two point calibration of the oxygen sensor is performed during the EST.
NOTE: If you do not connect the ventilator to an oxygen supply, the O2 sensor calibration will immediately fail.
CAUTION: Although failure of any of the above tests will not prevent the ventilator from functioning, it should be checked to make sure it is operating correctly before use on a patient.
13
Page 16
Setting the ventilation breath type and mode
Advanced settings
You can further refine delivery of the breath by accessing the Advanced Settings. Not all primary controls have advanced settings. Primary Controls that feature advanced settings will display a yellow triangle to the right of the control name.
Accessing the advanced settings
Press the Adv Se ttings membrane but ton.
Press the primary con trol to di splay th e advan ced settings available fo r that pa rticular cont rol.
14
Page 17
Graphics
Configure main screen graphics for scale and sweep speed
1. Touch the vertical or horizontal axis
to highlight.
2. Turn the data dial to adjust scale or
sweep speed. Touch the axis again or press the Accept button to confirm the changes.
15
Page 18
Loops
Accessing the Loops screen
To access the Loops screen, press the screen indicator or the Screens membrane button on the left of the UIM. Select Loop.
SCREEN SELECT
MAIN MONITOR
LOOP
STANDBYMANEUVERTRENDS
UTILITY
Choice of loops
The ventilator displays two loops in real time, selected from the following. Touch the Loop heading to open the scrollable menu. Press the Accept button or touch the Loop heading to confirm the change.
• Vt - Flow Flow / Volume loop.
• PAW - Vt Airway Pressure / Volume loop.
• PINSP - Vt Inspiratory Pressure / Volume loop.
16
Page 19
Configure Loops screen
Change vertical and horizontal scales with Touch-Turn-Touch or Touch-Turn-Accept techniques.
1. Touch the vertical or horizontal axis
to highlight.
250 500 750
2. Turn the data dial to adjust the scale. Touch the axis again or press the Accept button to confirm the change.
17
Page 20
Freeze
The Freeze button freezes the current screen and suspends real-time update of data until pressed again. When a screen is frozen you can use the data dial to scroll through displayed waveforms, loops or trends.
A Flow/ Volume loop in Freeze mode is shown below. As the dotted line cursor traces the Frozen loop, flags display the values along the curve of the loop.
Flow/ Volum e loop trac ing
18
270
0.0
250 500 750
Flags sho wing X and Y value s at variou s points a long the loop t racing
Cursor c urrentl y overlays th e “X” axis at zer o
Page 21
Saving loops
1. Press the Freeze button to freeze the
graphics display.
2. Press Save Loop.
3. Saved loops will appear with a time reference. A total of 4 loops can be saved at any time. If a 5 loop is selected, the oldest saved loop is removed.
th
19
Page 22
Create reference loops
1. Press the Freeze button to freeze the graphic display.
2. Touch the saved loop you want as a reference.
3. Press Ref Loop Off to toggle the Ref Loop On.
4. Press the Freeze button to return to live loops display over the reference loop.
To remove the reference loop, press the Freeze
button. Then press the Ref Loop On button to turn off the reference loop.
20
Page 23
Message Alarm condition Range Priority
High PPEAK Peak inspiratory pressure is greater tha n the
High nCPAP pressure
set high PP EAK. Inspiration is terminated and the circuit p ressure is allowed to return to baseline pressure + 5 ± 1.5 cmH next bre ath is deli vered.
Pressure during nCPAP exceeds the high airway pressure threshold for a perio d of greater tha n 20 seconds.
Pressure during nCPAP falls below the low airway pressure threshold for a perio d of greater tha n 20 seconds.
O before the
2
Normal breath rang e: 10 to 105 cmH Default 75 cmH 10 to 85 cmH Default 50 cmH (Inspirator y pressure +3 cmH rate not zero ) Sigh Breath Range: 1.5
Thresho ld: Set CPAP +c cmH ±0.5 cmH
Thresho ld: Set CPAP level -2 cmH CPAP 3 cmH (if set CPAP le vel <3 cmH
O (adult /pediatric)
2
O
2
O (neonate)
2
O nCPAP/ IMV 2 to 45 cmH20
2
0
2
0) or set CPAP level -1 cmH20
2
0, nCPAP/ IMV
2
0 tolerance
2
0)
2
0 (if set
2
High
High
High
21
Page 24
Message Alarm condition Range Priority
nCPAP Pressure Limit
Low Ppeak Low P peak is displayed an d a high prio rity
High Ppeak, SUST
Low PEEP Baseline pressure (Posi tive End E xpiratory
Activated when nCPAP pressure is greater than airway pressure for 3 seco nds. Deactivated wh en nCPAP pressure drops below 4.5 cmH
tone sounds, whenever the pe ak inspiratory pressure for a given bre ath is les s than the preset threshold fo r Low Ppeak.
Activates whenever the High Ppeak alarm has been ac tive for mo re than 5 seconds (i.e., if the circuit p ressure fails to return to PEEP + 5 cmH exhalation valves open and no breaths are delivered. The Safety Valve O pen alar m acti ­vates. Bia s flow is suspended w hile this alarm is active. PEEP may not b e maintained. This alarm will remain active until t he condition is resolv ed.
Pressure) is less than the set Low PEEP alarm threshold for a period greater than 0.25 ±
0.05 seconds. This alarm is of f if set to zero.
0.
2
0 within 5 se conds) . The safety and
2
Pressure Limit: 11 cmH20 (nCPAP only, rate set to off ) Set CPAP level + I nspirato ry pres sure + 3
0 (nCPAP/ IMV, rate not zero)
cmH
2
Range: 3 to 99 cmH Defaults: 8 cmH
0 (neonate) for nCPAP / IMV: 1-40
5 cmH
2
0 (neonatal –nCPAP/ IMV )
cmH
2
Limitat ions: N ot acti ve for spontan eous breaths
0
2
0 (adult /pediatric)
2
High
High
N/A High
0 to 60 cmH Defaults:3 cmH
0 (neonate)
1 cmH
2
0
2
0 (adult /pediatric)
2
High
22
Page 25
Message Alarm condition Range Priority
Low Ve Monitored exhaled minute volume (Ve) is less
High Ve Monitored exhaled minute volume (Ve) is
Flow Sens or Error (ne onatal volume guarantee o nly)
Low expired volume
than the set Low Ve alarm threshold.
greater tha n the set High Ve alarm threshold.
Occurs when 1) neonatal flow s ensor is in use 2) volume guarante e is enabled and
3) monitored Vti drops below 20% of net delivered volume. T he system will revert to operator set inspiratoy pressure.
When volume guarantee is active and monitore d expiratory tid al volume is less than the set threshold of the volume t arget.
Off ( 0), 1 to 50 L (adult ) Off(0), 0.1 to 30 L (ped iatric) Off ( 0), 0.01 to 5.00 L (neonate ) Default Off
0 to 75 L (adult)
0.0 to 30.0 L (pediatr ic)
0.00 to 5.0 0 L (neo nate) Defaults: 30.0 L (adult /pediatric)
Neonat al mode w ith volum e guarantee active, D elay is 3 breaths or 10 sec if greater, or 30 sec if l ess
Neonat al mode w ith volum e guarantee active, 3 0 sec or 10 breaths, whichever is greater
High
High
Medium
Medium
23
Page 26
Message Alarm condition Range Priority
High Vt The absolute m onitored exhaled tidal volume
Apnea Interval Active in A /C, SIMV, APRV / BiPhasic and CPAP/
High Rate The monitored total breath rate exceeds the set
I-Time Limit The inspiratory t ime for a brea th plus pause
Limit The I: E ratio for a mandatory breat h
Low FiO
2
High FiO
2
Circuit Disconnect
is greater than the set High Vt alar m thresho ld. (Inactive when using Volume Guarantee in Neonat al mode )
PSV modes if the ve ntilator d oes not detect a breath within the pre set Apnea time interval.
High Rate alarm threshold.
time exceeds 5.0 seconds for adult/ pediatric patient s and 3.0 seconds for ne onatal patients.
exceeds 4 :1. The inspirato ry phas e of the breath is terminated.
Delivered oxyge n percent age falls below the
minus 6% or 18% Fi O2, whichever
set FiO
2
is greater.
Delivered oxyge n percent age rise s above the
plus 6%.
set FiO
2
A high priority audible /visual alarm is a ctivated, and Circuit Disconnec t displayed, when ever the patient circuit becomes disconnec ted from the ventilator or patient. (See o perator ’s manual for nCPAP/ IMV di sconnect sensi tivit y)
0.10 to 3.00 L (adult) 25 to 1,000 mL ( pediat ric)
2.0 to 300.0 m L (neonate) Defaults: 3.00 L ( adult) 1,000 mL ( pediatric)
300.0 mL (neonate)
6 to 60 seconds Default 20 se conds
1 to 200 bpm Default 20 0 bpm
High
High
Medium
N/A Medium
Not active in APRV/ BiPhasic mode.
N/A
N/A
N/A
24
Page 27
Message Alarm condition Range Priority
Low Battery A high priority audible /visual alarm is a ctivated, and Low Batter y displayed,
Loss, AC Power A high priority audible /visual alarm is a ctivated and Los s, AC Power is
ILV Disconne ct A high priority audible /visual alarm is activated, and I LV Disconnect
Invalid Gas ID A medium priority audible /visual alarm shall be activated, and Invalid Ga s
Fan Failure A low prior ity audible /visual alarm is activated, and Fan Failure indicated,
Volume Guarantee Disabled
Volume Guarantee is only available in Pressure and TCPL
Set Vol Target will increase delivered press and volume
wheneve r the internal battery has be en depleted to a level that provides a minimum of t wo minutes of safe ope ration.
displayed, whenever the power switch is on and AC power has been removed from th e ventilator (i.e., powe r cord disconnect or loss of supply po wer)
displayed, whenever the master ventila tor becomes disco nnected from the slave ventilator durin g ILV.
ID shall be indicate d, whenever a defec tive gas ID connector is ins talled in the ventilator. When a defective Gas ID connector is detec ted, the ga s correc tions default to air.
wheneve r the fan has stopped rotating.
Proximal flow sensor disconnected or damage d (neonatal mode only) . N /A Message
Selec tion of Volume Guarante e on the mode scree n is not Pres sure or TCPL (neonatal mode only).
Volume Target set more than 20 % above current set ting (neonatal mode only).
N/A High
N/A High
N/A High
N/A Medium
N/A Low
N/A Message
N/A Message
25
Page 28
Message Alarm condition Range Priority
Set Vol Target will decrease delivered press and volume
High Ppeak <PEEP +7 cm H
Volume Guarantee pressure is limited
Volume Target set more than 20 % below current setting ( neonat al mode only).
When nCPAP / IMV mode is ac tive and breath
0
rate is not Of f, attempt to set High P peak
2
alarm limit or nCPAP such that High Ppe ak alarm limit settin g is less than +2 cmH
The pressure required to deliver the desired tidal volume is greate r than the Hi gh Ppeak alarm limit of -3 cmH
0.
2
N/A Message
N/A Message
0.
2
N/A Message
26
Page 29
Rear panel connections
Oxygen sensor
The oxygen sensor cell is located on the rear panel, between the two gas fittings. The oxygen sensor cable emerges from the rear panel directly above the sensor. Carefully align and then gently push the connector onto the oxygen sensor until it seats, then slide the cover down and push over the sensor.
27
Page 30
Ventilator operations
AVEA ventilator systems
The Membrane buttons are the UIM controls that surround the Touch Screen. Starting at the top right and moving clockwise around the UIM they are:
A. Alarm Silence (LED)
Pressing this button will disable the audible portion of an alarm for 2 minutes (± 1 second) or until the Alarm Silence button is pressed again. This button is not functional for a Vent Inop alarm.
NOTE: Pressing the Alarm Silence button will not prevent the audible alarms sounding again later for certain alarm conditions.
User interface module (English) showing button labels
28
Page 31
B. Alarm Reset
Cancels the visual indicator for alarms that are no longer active.
C. Alarm Limits
Opens the Alarm Limits screen for data entry or adjustment. Toggles the screen on and off.
NOTE: Pressing the Freeze button while the Alarm Limits window is open will automatically close the window and freeze the graphics.
D. Manual Breath
Pressing this button during the expiration phase of a breath delivers a single mandatory breath at current ventilator settings. No breath is delivered if the key is pressed during inspiration.
NOTE: The Manual Breath button is not active in APRV / BiPhasic.
E. Suction (LED)
Pressing this button initiates a Disconnect for Suction maneuver. The ventilator will:
• Enable an Increase % O
maneuver for 2 minutes (see Increase O2 below).
2
• While the Circuit Disconnect alarm is active, the ventilator will stop cycling and set a bias
flow. The ventilator will automatically detect the patient upon reconnection and resume normal ventilation.
• Silences alarms for 120 seconds.
29
Page 32
E. Suction (LED) (continued)
If the Suction key is pressed again during the 2 minutes that the disconnect for suction maneuver is active, the maneuver will be cancelled.
F. Increase O
2
When this key is pressed, the ventilator increases the oxygen concentration delivered to the patient for 2 minutes. If the ↑%O2 key is pressed again within this two-minute period, the maneuver is cancelled and the ventilator will return to prior settings.
Defaults : +20% neonatal; 79% adult /pediatric
Adult/pediatric: 79% above the set % O
2
Neonate: 20% above the set % O2 or 100%, whichever is less
,
To configure the Increase FiO
Increase FiO
2
access the configuration tab on the Utilities screen:
2
Configures the step increase used during the increase oxygen maneuver. Sets the amount of oxygen the ventilator will increase above the current set FiO2.
30
Page 33
Example: If the Increase FiO
is set at 20%
2
AND
The set FiO
is 40%
2
WHEN
The increase FiO
maneuver is activated the FiO2 will increase to 60% for two
2
minutes after which it will return to 40%.
The default setting for infants is 20% and 79% for pediatric and adult applications.
NOTE: The settings will be reset to default values when New Patient is selected in the setup.
NOTE: To achieve 100% delivered FiO
during the Increase O2 maneuver set the Increase FiO2
2
setting to its maximum of 79% .
WARNING: Heliox delivery will be interrupted for the time that either the Suction or the Increase O
buttons are pressed during administration of Heliox. Tidal volume may be affected
2
after the two-minute timeout period, or when the button is pressed, until the accumulator has been purged.
31
Page 34
G. Data dial
Changes the values for a selected field on the touch screen.
H. Accept
Accepts data entered into a field on the touch screen.
I. Cancel
Cancels data entered into a field on the touch screen. The ventilator will continue to ventilate at current settings.
J. Expiratory Hold
When the Exp Hold button is pressed, at the start of the next breath interval, the ventilator will not allow the patient to inspire or exhale for a maximum of 20 seconds (adult/pediatric) or 3 seconds (neonate). Expiratory Hold is NOT active in TCPL breaths.
K. Inspiratory Hold (Manual)
When the Insp Hold button is pressed, once the preset of a volume control or pressure control breath has been delivered, the patient will not be allowed to exhale for a maximum of 3.0 seconds (± 0.1 second).
32
Page 35
L. Nebulizer
The ventilator supplies blended gas to the nebulizer port at 10 ± 1.5 psig (0.7 bar) when an in-line nebulizer is attached and the Nebulizer key is pressed, provided that the calculated delivered flow is >15 L/min.
Delivery of the nebulized gas is synchronized with the inspiratory phase of a breath and lasts for 20 minutes. Press the Nebulizer key a second time to end the treatment prior to the end of the 20-minute period.
CAUTION: Use of an external flow source to power the nebulizer is not recommended.
WARNING: Using the nebulizer may impact delivered tidal volumes.
NOTE: Do not operate the nebulizer while using heliox
M. Patient Size
The Patient Size Indicators for adult, pediatric and neonate at the bottom of the UIM show which patient size is currently selected. These LED indicators have no associated membrane button on the UIM.
33
Page 36
M. Patient Size (continued)
NOTE: The ventilator will not allow patient size changes when the active mode of ventilation is not available in the new patient size selection. The ventilator will display a message instructing you to first change the ventilation mode. For example, in neonatal ventilation with TCPL active, you cannot change to a pediatric or adult patient size without first changing the mode to one available for those patients.
The ventilator will also not allow size changes if Machine Volume is active. A message displays indicating that Machine Volume must first be turned off before making a patient size change.
N. Panel Lock (LED)
The Lock key disables all front panel and screen controls except Manual Breath, Suction, ↑ %O Alarm Reset, Alarm Silence and Lock.
O. Print
The Print key outputs the contents of the currently displayed screen to a suitably connected parallel printer.
34
,
2
Page 37
P. Set-up
Opens the ventilator Set up screen.
NOTE: Pressing the Set-Up button a second time before accepting Set-Up will close the window and restore the previous settings. The Set-Up screen uses an on-screen accept button. To change patient size without selecting a new patient requires that patient Set-Up be accepted after selecting patient size.
Q. Advanced Settings (LED)
Opens the Advanced Settings screen for data entry or adjustment. Toggles the screen on and off.
NOTE: Pressing the Freeze button while the Advanced Setting window is open will automatically close the window and freeze the graphics.
R. Mode
Opens the Mode Select screen for data entry or adjustment toggles the screen on or off. Pressing the Mode indicator at the top of the touch screen will also access the screen.
NOTE: Pressing the Mode button a second time before accepting the Mode will close the window and restore the previous settings. The Mode screen uses an on screen accept button.
S. Event
Records an event for future reference. Some Events are recorded automatically others can be logged manually to display in this screen. See Chapter 4, Monitors and Displays, for a full list of Events.
35
Page 38
T. Freeze
The Freeze key freezes the current screen and suspends real-time update of screen data until pressed again. While the screen is frozen, a scrollable cursor appears. The data dial can be used to scroll the cursor through data points on waveform, loop or trend screens. To restore the screen to active, press the Freeze button a second time.
Figure 3.2 shows a flow/volume loop in freeze mode. The cursors trace the “frozen” loop curve along an X-Y plot line. The values along the curve of the loop are displayed as shown below.
Flags sho wing
270
Dashed c ursor li ne
0.0
X and Y value s at variou s points a long the loop tracing
250 500 750
Flow/ Volum e loop trac ing
Cursor c urrentl y overlays th e “X” axi s at zero
36
Page 39
Connecting gas sources
Gas fittings
To connect either the Air or Heliox (80/20) Smart connector, align the connector, seat gently onto the fitting and screw down the fitting until finger tight.
Each Smart connector is permanently tethered to the rear of the ventilator.
WARNING: Connection of a gas supply at the Helium-Oxygen mixture inlet that does not contain 20% oxygen can cause hypoxia or death.
Although an 80/20 mixture of Helium and Oxygen is marketed as medical grade gas, the Helium/Oxygen gas mixture is not labeled for any specific medical use.
In order to prevent delivery of excess tidal volume, allow 90 seconds for the accumulator to purge before initiating patient ventilation with Heliox gas.
37
Page 40
Touch-Turn-Touch/Touch-Turn-Accept techniques
Changes to controls and displays are accomplished with a three-step technique.
1. Touch the control or
display to be changed.
15
BPM Rate
500
ml
Volume
2. Turn the data dial to reach
15
the desired setting.
BPM Rate
500
ml
Volume
3. Touch the control again or press Accept to confirm change.
Changes not accepted by either method within 15 seconds revert to previous settings. Press Cancel prior to accepting proposed changes to return to previous settings.
38
Page 41
New patient setup
1. Power on the ventilator and perform User Verification
Tests as described in chapter 2 of your operator’s manual.
2. After the power is turned on, a single audible tone may
be heard when the internal capacitor linked to the alarm system reaches full charge.
3. Select New Patient. Select Resume Current or New
Patient. Touch Patient Accept. Selecting Resume Current resumes ventilation and trending at previous settings. Selecting New Patient clears all trends and saved loops, and sets all controls to defaults. The Patient Size Select screen appears as the first step of the new patient setup sequence.
4. Using Touch-Turn-Touch or
Touch-Turn-Accept technique, set controls found on the ventilation Setup screen as desired.
5. Touch Setup Accept to accept
this screen.
39
Patient Select
RESUME CURRENT
NEW
PATIENT
PATIENT ACCEPT
Page 42
Setting the ventilation breath type and mode
Adult/pediatric breath type and mode screen
40
Page 43
Infant mode select screen
Apnea Backup ventilation choices appear when CPAP / PSV or APRV / BiPhasic modes are selected. Apnea Backup is active in all Assist Control, SIMV, APRV / BiPhasic and CPAP/PSV modes.
41
Page 44
Setting nCPAP/IMV (neonatal mode only)
When mode is accepted, follow the on-screen instructions for flow characterization. See Operator’s Manual for a list of approved nCPAP interfaces.
42
Page 45
1. Once flow characterization is completed, the screen above is displayed.
bmp
2. Connect patient to ventilator, and set desired settings.
3. Alarm settings are only adjustable if rate is set other than Off. If nIMV is desired, then
set rate, inspiratory pressure and inspiratory time to desired settings.
4. Check disconnect sensitivity. See operator’s manual for detailed instructions.
Non-Invasive Support.
APNEA Backup Disabled.
Off 5 0.35 5 21
bmp
Rate
cmH20
Insp Pres
Non Alarm Limits in cCPAP when settings are disabled.
sec
Insp Time
cmH20
nCPAP
Fi02
43
%
Page 46
Setting the ventilation breath type and mode
1. Touch the Mode indicator or the Mode membrane
button to open the mode window (see pg. 1).
2. Select the desired mode and set the primary controls.
Touch Mode Accept to accept the new mode and control settings.
3. Advanced settings and alarm limits may also be
adjusted at this time.
44
Page 47
Breath type and mode
Setting Volume Guarantee (neonatal mode only)
1. Select Volume Guarantee with pressure or TCPL breath
type ONLY! Assure proximal flow sensor is zeroed and in place if needed.
2. Select desired settings. Volume will then
appear in place of inspiratory pressure. Set desired expired volume.
3. Under volume advanced settings, delivered
starting pressure limit will appear. Set desired starting inspiratory pressure.
4. Pressure will vary ± 3 cm maximum to achieved
set volume, but will give initial breath at set inspiratory pressure.
NOTE:
• If flow sensor becomes disconnected or non-functional, breath type will divert to preset
inspiratory pressure. Inspiratory pressure will be limited to High Pressure Limit alarm -3 cm.
• In TCPL breath type, flow and /or inspiratory time may limit volume delivery. Expiratory
volume alarm will be activated if expiratory volume falls below alarm threshold.
• Flow cycling is functional only in TCPL breath type.
45
Page 48
Monitored parameters
Main screen monitors
Five monitored parameters are continuously displayed on the left of the touch-screen display.
1. Touch the monitor you wish to set.
2. Scroll through the menu choices.
3. To accept your selection, press the
Accept button.
NOTE: The main screen monitored values may be different than the monitored values on the Loops screen or Trends screen.
46
Page 49
The Monitor screen
To open the Screen Select menu, press the Screen indicator or the Screens membrane button to the left of the touch screen. Select Monitor.
The Monitor screen can be configured in the same way as monitored parameters on the main or loops screen displays. For a complete list of monitored parameters, refer to chapter 4 of the AVEA ventilator operator’s manual.
47
SCREEN SELECT
MAIN MONITOR
LOOP
STANDBYMANEUVERTRENDS
UTILITY
Page 50
Graphics
Graphic displays are color coded to provide the clinician with helpful information. They may appear as red, blue, yellow, green or purple tracings.
• Red indicates the inspiratory portion of a mandatory breath
• Yellow indicates the inspiratory portion of an assisted or spontaneous breath (patient
assisted or spontaneous breaths are also denoted with a yellow demand indicator that appears in the left hand corner of the mode indicator)
• Blue represents the expiratory phase of a mandatory, assisted or spontaneous breath
• Green during the expiratory phase of a single breath indicates that a purge of the expiratory
flow sensor, or wye flow sensor (if attached) has occurred
• Purple indicates safety state, which occurs when the safety valve is open
48
Page 51
Waveforms
Three waveforms can be selected and simultaneously displayed on the main screen.
Scroll through the waveform choices. Make your selection, touch the heading again or press the Accept button to confirm the change.
49
Page 52
Events
Pressing the Event button opens a menu of selectable events that will appear in the trend buffer along with the monitored parameters. Scroll the menu and highlight the desired event. Press the Accept button adjacent to the data dial to place the event in the trend buffer. Events appear on the trend screen spreadsheet in green text. Certain events are automatically recorded by the ventilator. All events are listed below.
Selectable event
• Change of a primary or advanced control
• Powering the ventilator off or on
• Entering and exiting standby
• Activation of the nebulizer
• Activation of the inspiratory or expiratory hold
• A manual breath
• Activation of the suction button
• Activation of the increase O
button
2
• Selecting new patient
• Involuntary power loss and recovery
50
Selectable event
• Arterial blood gas
• Chest X-ray
• Suction
• Intubation
• Feeding
• Diagnostic procedure
• Therapeutic procedure
Page 53
Trends
1 2 3 4 5 6
1 2 3 4 5 6
120
P
peak
(cmH20
Press the Screens button or the Screen indicator. Press Trends on the screen select menu. Monitored parameters are trended as one minute averages over a 24-hour period.
NOTE: If left open, the Trends screen will refresh every 10 minutes.
09:29
09
09
09
09
90 60 30 0
120 90 60 30 0
T
i
me
:
30
:
31
:
32
:
33 57.157.1
)
1 2 3 4 5 6
mea
n
P
(cmH2O
)
1 2 3 4 5 6
V
te
V
e
PEEP
m
l
l
cm
H2O
0.8
53.6
56.7
52.4
57.0
6
0.9
6
0.9
6
1.0
6
1.1
6
Each histogram and column can be configured from the monitored parameters menu. Touch the title bar of any histogram or the heading of any column. Select the parameter to be displayed. Press the Accept button to accept the parameter.
Histograms can be scaled by touching either axis. Use the data dial to adjust the scale. Touch the axis again or press the Accept button to accept the change.
To look at histogram or spreadsheet trends over time, press the Freeze button and use the data dial to move the cursor through the timeline. The timeline is shown as yellow text on the spreadsheet. Event markers appear in green text.
51
20
PEEP
(cmH20 15 10 5 0
200 150 100 50 0
A
i
r Inlet
V
de
l
P
sig
m
l
94.0
93.6
93.4
93.1
92.32222
peak
cm
H2O
22
22
22
p
0
0
0
0
0
)
R
ate
(
bpm
)
P
PI
EE
P
F
R
PEF
cm
H2O
6
6
6
6
6
R
min
min
m
I/
l/cm
I/
41
41
41
41
41
40
41
41
41
41
C
dyn
H2O
3
4
3
4
4
Page 54
Synchronized nebulizer
An in-line nebulizer can be used with the AVEA ventilator. Nebulization is synchronized with inspiration, and will deliver gas at the set FiO2/FiHe for 20 minutes. Attach the nebulizer tubing to the fitting at the bottom of the front panel.
NOTE: To use the internal nebulizer, the AVEA ventilator must be connected to a high pressure Air or Heliox source. The nebulizer is not active while the AVEA ventilator is operating on its internal compressor. The nebulizer requires an inspiratory flow rate of at least 15 liters per minute to activate and is flow compensated to maintain set tidal volumes. The nebulizer is not active in neonatal applications.
CAUTION: When the internal nebulizer is used, the ventilator decreases the flow rate by 6 L/ min to compensate for the nebulizer output. However, since flow from the internal nebulizer can vary, using the internal nebulizer may impact the tidal volumes delivered to the patient.
52
Page 55
Alarms
Setting an alarm limit
Press the red Limits membrane button.
Use Touch-Turn-Touch/Touch-Turn­Accept technique to make changes to alarm controls.
NOTE: Red indicators appearing on the primary controls display the relative alarm settings of any associated alarm.
53
Page 56
Message Alarm condition Range Priority
Safety Valve Open
Vent Inop Ventilator failure due to non-recoverable conditio n. The safety valv e opens,
Loss, Air
Loss, O
2
Loss, Heliox The alarm is trigge red if heli ox is being used and the heliox gas supply to
Loss, Gas Supply All sources of gas fail; wall air, internal compressor (if ins talled ) and
Safety valve is open. N/A High
indicate d by a Safet y Valve Ope n messa ge, and the patient is allowed to breathe ro om air. PEEP is not maintaine d.
Wall air drops below 18.0 ps ig (1.2 bar) and no functional compre ssor is installed, or the compressor output is insufficient to m eet inst rument demand. Patient will continue to b e ventilated by O
Oxygen supply to the ventilator drops below 18.0 psig (1.2 bar) and
2
is set to > 21%. Pati ent will continue to be ventilated by the
the %O air supply o nly.
the ventila tor drops b elow 18.0 psig (1.2 bar). The patient co ntinues to b e ventilated by the oxygen suppl y only.
oxygen. The safety valve o pens, ind icated by a S afety Val ve Open message, and the patient is allowed to breat he room air. PEEP is not maintained.
2
supply only.
N/A High
N/A High
N/A Medium
N/A Low
N/A Message
54
Page 57
Notes
55
Page 58
Notes
56
Page 59
Notes
57
Page 60
WARNING—U.S. Federal Law restricts this device to sale by or on the order of a physician.
CareFusion 22745 Savi Ranch Parkway Yorba Linda, C A 92887
800.231.2466 toll-free
714.283. 2228 tel
714.283. 8493 fax
CareFusion Germany 234 GmbH Leibnizstrasse 7 97204 Hoechberg Germany
+49 931 4972-0 tel +49 931 4972-423 fax
carefusion.com
© 2011 CareFusi on Corp oration o r one of it s subsi diarie s. AVEA , Touch­Turn-Acce pt and Touc h-Turn-Touch are t radema rks or re giste red trad emark s of CareF usion Co rpora tion or on e of its su bsidia ries. A ll right s reser ved. RC260 5 (0611/2 000 ) L2242 Rev. B
Loading...