Beckman Coulter UniCel DxC Synchron 800, UniCel DxC Synchron 600 Instructions For Use Manual

A13914-AB

October 2005
®
UniCel® DxC Synchron
Clinical Systems
For In Vitro Diagnostic Use Only
This manual is intended for
UniCel UniCel® DxC 800
®
DxC 600
®
REPEC Beckman Coulter Ireland, Inc. Mervue Business Park, Mervue Galway, Ireland 353 91 774068
BECKMAN COULTER, INC. • 4300 N. Harbor Blvd., Fullerton, CA 92835 U.S.A.
© Copyright 2005 Beckman Coulter, Inc.

Table of Contents

CHAPTER 1 General Information.......................................................................................... 1-1
General Information....................................................................................... 1-1
Manual Conventions...................................................................................... 1-2
How to Use this Manual ................................................................................ 1-3
Summary of Hazards ..................................................................................... 1-4
Summary of Precautions................................................................................ 1-7
Symbols and Labels..................................................................................... 1-12
CHAPTER 2 System Description........................................................................................... 2-1
System Description........................................................................................ 2-1
Operational Conditions.................................................................................. 2-2
System Components ...................................................................................... 2-5
Sample Handling System............................................................................... 2-7
Modular Chemistry (MC) System ............................................................... 2-13
Cartridge Chemistry (CC) Reagent Handling System................................. 2-18
Cuvette Reaction System............................................................................. 2-21
Hydropneumatic System.............................................................................. 2-24
Operation and Control Components ............................................................ 2-26
Main Operator Screen and Program Structure............................................. 2-27
Theory of Operation..................................................................................... 2-37
Cartridge Chemistry: Calibration Theory.................................................... 2-38
Modular Chemistry: Calibration Theory ..................................................... 2-49
Cartridge Chemistry: Principles of Measurement ....................................... 2-50
CHAPTER 3 Preparing Samples for Analysis........................................................................ 3-1
Routine Operation Overview......................................................................... 3-1
Sample Programming and Processing ........................................................... 3-2
How to Use Reserved Racks.......................................................................... 3-8
CHAPTER 4 System Setup Options....................................................................................... 4-1
Overview........................................................................................................ 4-1
Password Setup.............................................................................................. 4-3
Auto Serum Index/ORDAC........................................................................... 4-5
Configuring the Chemistry Menu.................................................................. 4-7
Setting the Default Sample Type................................................................. 4-15
Date/Time Setup .......................................................................................... 4-16
Demographics Setup.................................................................................... 4-17
Patient Results – Immediate Reporting Setup ............................................. 4-18
Panels........................................................................................................... 4-19
Replicates..................................................................................................... 4-20
Report Setup ................................................................................................ 4-21
Reportable Ranges Setup............................................................................. 4-22
Reference/Critical Ranges Setup ................................................................. 4-24
UniCel DxC Systems Instructions For Use A13914 Table of Contents October 2005 Page 1 of 6
Sample Comments Setup............................................................................. 4-25
Special Calculations Definition ................................................................... 4-26
Timed Urine and Creatinine Clearance Results........................................... 4-29
Setup Summary............................................................................................ 4-30
Units/Precision Setup................................................................................... 4-31
User Defined Chemistries Setup.................................................................. 4-32
Bar Code Setup ............................................................................................ 4-33
Maximum Sample Program Age ................................................................. 4-34
Reserved Racks/Obstruct Detect ................................................................. 4-35
Disable Service Monitor .............................................................................. 4-36
Host Communications.................................................................................. 4-37
Language/Keyboard Setup........................................................................... 4-38
Printer Setup ................................................................................................ 4-39
Service Setup ............................................................................................... 4-40
System Configuration .................................................................................. 4-41
Version Upgrade.......................................................................................... 4-42
Status Alarm/Annunciator ........................................................................... 4-43
Chemistry Update ........................................................................................ 4-44
Auto Generation of Control ......................................................................... 4-45
CHAPTER 5 Reagent Load/Calibration................................................................................. 5-1
Reagent Load ................................................................................................. 5-1
System Calibration.......................................................................................... 5-13
System Calibration Introduction.................................................................. 5-13
Loading a Calibrator Diskette...................................................................... 5-14
Calibrator Assignment ................................................................................. 5-15
Calibration Status......................................................................................... 5-16
Reagent and Calibration Status Warnings ................................................... 5-17
Requesting a Calibration.............................................................................. 5-18
Calibration Failure Messages....................................................................... 5-20
Within-Lot Calibration ................................................................................ 5-21
Enzyme Validator ........................................................................................ 5-25
Calibration Override .................................................................................... 5-26
Chemistry Bypass ........................................................................................ 5-27
Extending Calibration Time......................................................................... 5-28
Calibration Acceptance Limits .................................................................... 5-29
Calibrator Set Point Modifications .............................................................. 5-30
Slope Offset Adjustment.............................................................................. 5-32
Reprint Calibration Reports......................................................................... 5-34
CHAPTER 6 Quality Control ................................................................................................. 6-1
Quality Control .............................................................................................. 6-1
Defining a Control ......................................................................................... 6-4
Control ID Assignments ................................................................................ 6-8
Running Control Samples.............................................................................. 6-9
Editing a Control Definition ........................................................................ 6-11
Reviewing a Control Definition................................................................... 6-12
Deleting a Control........................................................................................ 6-13
Table of Contents UniCel DxC Systems Instructions For Use A13914 Page 2 of 6 October 2005
Printing QC Ranges ..................................................................................... 6-14
QC File List ................................................................................................. 6-15
QC Summary ............................................................................................... 6-16
QC Chart (Levey-Jennings)......................................................................... 6-17
QC Log ........................................................................................................ 6-19
Archive QC.................................................................................................. 6-22
Restore and Review Archived Data............................................................. 6-24
CHAPTER 7 Sample Programming ....................................................................................... 7-1
Overview........................................................................................................ 7-1
Prior to Programming .................................................................................... 7-2
Identifying Samples ....................................................................................... 7-4
Selecting Tests ............................................................................................... 7-7
Programming Functions................................................................................. 7-8
Batch Programming ....................................................................................... 7-9
CHAPTER 8 Results Recall ................................................................................................... 8-1
Overview........................................................................................................ 8-1
Recalling Results by Sample ID .................................................................... 8-2
Recalling Results by Rack and Position ........................................................ 8-3
Recalling Results by Patient ID..................................................................... 8-5
Recalling Results by Run Date/Time............................................................. 8-6
Displaying Recalled Results .......................................................................... 8-7
Critical Result Rerun Editing......................................................................... 8-8
Printing Recalled Results............................................................................. 8-10
Sending Results to the Host......................................................................... 8-11
Absorbance Versus Time............................................................................. 8-12
Statistical Summary Report ......................................................................... 8-13
CHAPTER 9 User Defined Reagents ..................................................................................... 9-1
Overview........................................................................................................ 9-1
Requirements and Precautions....................................................................... 9-2
User-Defined Reagent Setup ......................................................................... 9-4
Chemistry Parameters.................................................................................... 9-5
Processing Parameters ................................................................................. 9-10
Error Detection Limits................................................................................. 9-16
Wavelength Selection .................................................................................. 9-18
Determination of Extinction Coefficients.................................................... 9-20
Exit Check Criteria ...................................................................................... 9-21
User Defined Reagent Removal .................................................................. 9-23
CHAPTER 10 Maintenance.................................................................................................... 10-1
Maintenance Overview................................................................................... 10-1
Overview...................................................................................................... 10-1
Electronic Maintenance Log........................................................................ 10-4
Weekly Maintenance ................................................................................... 10-8
Monthly Maintenance................................................................................ 10-18
Two-Month Maintenance .......................................................................... 10-37
UniCel DxC Systems Instructions For Use A13914 Table of Contents October 2005 Page 3 of 6
Three-Month Maintenance ........................................................................ 10-44
Four-Month Maintenance .......................................................................... 10-50
Six-Month Maintenance ............................................................................ 10-55
As-Needed/As-Required Maintenance ...................................................... 10-68
CTS Wick Replacement Procedure ......................................................... 10-121
CHAPTER 11 System Status and Commands........................................................................ 11-1
Overview...................................................................................................... 11-1
System Status.................................................................................................. 11-2
Status Summary ........................................................................................... 11-2
Status-Cycle Count ...................................................................................... 11-4
Temperatures ............................................................................................... 11-5
Power Subsystems ....................................................................................... 11-6
Hydropneumatics Subsystem....................................................................... 11-7
ICS/Smart Modules...................................................................................... 11-8
Cuvette Water Blank Status......................................................................... 11-9
CTS Tracking............................................................................................. 11-10
Instrument Commands.................................................................................. 11-11
Instrument Commands Summary .............................................................. 11-11
Home.......................................................................................................... 11-12
Pause .......................................................................................................... 11-13
Stop Print ................................................................................................... 11-15
Shutdown ................................................................................................... 11-16
System Power On/Boot.............................................................................. 11-19
Pause/Resume Waste B ............................................................................. 11-21
Enable/Disable Modules............................................................................ 11-22
Unload All.................................................................................................. 11-24
CHAPTER 12 Utilities............................................................................................................ 12-1
Overview...................................................................................................... 12-1
Prime............................................................................................................ 12-2
Maintenance................................................................................................. 12-7
Event Log..................................................................................................... 12-8
Alignment/Diagnostics/PVTs .................................................................... 12-14
Metering..................................................................................................... 12-15
Modem....................................................................................................... 12-16
Backup/Restore.......................................................................................... 12-17
Touch Screen Calibration .......................................................................... 12-20
CHAPTER 13 Troubleshooting Calibration and Result Errors.............................................. 13-1
Troubleshooting Calibration Errors................................................................ 13-1
Calibration Errors ........................................................................................ 13-1
MC Calibration ............................................................................................ 13-3
Linear Calibration........................................................................................ 13-8
Non-Linear and Multipoint Calibrations ................................................... 13-12
Troubleshooting Result Errors...................................................................... 13-17
Result Errors .............................................................................................. 13-17
Error Codes and Definitions ......................................................................... 13-18
Table of Contents UniCel DxC Systems Instructions For Use A13914 Page 4 of 6 October 2005
GLOSSARY
INDEX
Result Errors and Codes ............................................................................ 13-18
Error Code – Definitions............................................................................ 13-21
Common Error Messages and Corrective Actions .................................... 13-29
UniCel DxC Systems Instructions For Use A13914 Table of Contents October 2005 Page 5 of 6
Table of Contents UniCel DxC Systems Instructions For Use A13914 Page 6 of 6 October 2005

General Information

Intended Use

The UniCel® DxC Synchron® Clinical Systems are fully automated, computer­controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).

Scope of This Manual

This manual covers basic operating instructions and maintenance guidelines for UniCel DxC 600/800 Systems. Detailed operation, maintenance, and troubleshooting instructions are not included in this manual. In addition, medical and diagnostic interpretation, or the clinical significance of chemistries or assays are not discussed. Refer to the reference materials below for detailed information.

CHAPTER 1 General Information

General Information
Intended Use
1

Reference Manuals

Other manuals that accompany the system include:
• the UniCel DxC Synchron Clinical Systems Reference Manual which contains detailed system description, comprehensive operating instructions, theory of operation, system calibration, and programming procedures, and quality control information for the UniCel DxC Systems.
• the Synchron Clinical Systems Chemistry Information Manual which contains specific chemistry information for the full range of analytes available on the UniCel DxC and Synchron LX Systems.
• the UniCel DxC Synchron Clinical Systems Host Interface Specifications.
• the Synchron Clinical Systems Performance Verification Manual which will help you integrate the new UniCel DxC System into your daily laboratory routine.
• the UniCel DxC 600/800 Synchron Clinical System Operator Tips which provide additional information about your system, including screen names for general chemistries and Immunoassays.
• the Synchron LX®/UniCel DxC Clinical Systems Sample Template.
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-1

Manual Conventions

Manual Conventions

Manual Conventions
Manual Conventions
This Manual uses the following printed and visual cues to guide the user in responding to printed directions.
Table 1.1 Conventions Used in this Manual
Convention Description
Icon buttons and
Icon buttons and dialog box buttons are written in bold.
Dialog box buttons
Example:
Select the Samples icon.
Function buttons Function buttons are written in bold with the corresponding
function key in square brackets ([ ]).
Example:
Select Print [F10].
Command buttons (buttons with names
Buttons with names are bold and enclosed in angle brackets
< >. on a screen) or <> <↓>
Example:
Select <Cancel>. Instrument buttons Buttons on an instrument are in bold. May be all upper case or
initial caps.
Example:
Press the STOP button. Keyboard keys Keyboard keys are written in bold and enclosed in square
brackets [ ].
Examples:
[X], [], or [Tab] [Enter] Options button An Options button is a triangle enclosed in angle brackets < >.
Example:
Select the Options button <▼>. Text field Names in the text fields are bold followed by the word "field".
Example:
Type in the Patient ID in the Patient ID field. [Alt + X] Combination keys are written in bold and enclosed by square
brackets ([ ]) with a plus (+) sign between each key.
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-2 October 2005

How to Use this Manual

Manual Format

Information in this manual is presented in modular units. Each unit of information is described by a brief title in the left margin.
Many units consist of a table which presents a procedure, process, or description.

Procedure Tables

Procedure tables are the most common type of table in this manual. They list each step of a procedure by number with the corresponding action that is to be performed.
Occasionally, a decision must be made at a step within a procedure. A smaller decision table is then presented which describes the variable conditions in the left column and the appropriate action for each condition in the right column.
How to Use this Manual
Manual Format
1

Example of Procedure Table

The following table is an example of a procedure table that contains a decision table.
Step Action
1Select Rerun [F6]. 2
To enter... type...
individual Sample IDs, the Sample IDs for rerun in the Sample
a range of Sample IDs, the Sample ID at the beginning of the
3 Select a button from the bottom of the dialog box.
Read the decision table as a complete sentence, using the first heading to introduce the condition and the second heading to introduce the action. Step 2 of the table is read:
To enter individual Sample IDs, type the Sample IDs for rerun in the Sample IDs field.
IDs field.
range in the Range field.
To enter a range of Sample IDs, type the Sample ID at the beginning of the range in the Range field.
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-3

Summary of Hazards

Introduction

Summary of Hazards
Introduction
This section summarizes the hazards associated with the DxC System. Individual hazards associated with a specific procedure in this manual are included in Warning or Caution boxes within the procedures for that task. Please read this section and the following Summary of Precautions before operating the system.

Bar Code Reader Hazards

Do not tamper with or remove the housing of any bar code reader because of the laser­based nature of the readers and the potential hazard of looking directly at laser light. When the instrument is running, homing, or in diagnostics, the laser may be ON. At all other times the laser is OFF.

Biohazardous Materials Hazards

Observe all laboratory policies or procedures which pertain to handling of infectious and pathogenic materials.

Closed Tube Sampling (CTS) Cap Piercer Assembly Hazards/Biohazards

The Cap Piercer contains a razor sharp blade assembly that has been exposed to potentially biohazardous fluids. The points of the Blade are very sharp and extend below the Wash Tower. Stay away from the bottom of the Wash Tower. To prevent injury or exposure, do not touch the Points of the Blade and always wear gloves. Do not select <OK> when the cover is removed. This will move the Cap Piercer Carriage up and cause a pinching hazard.

CTS Auto-Gloss Handling Hazards

Be careful when handling the bottle of CTS Auto-Gloss. Prevent spills. This lubricant is extremely slippery and difficult to clean from a floor.

Electric Shock Hazards

Replacement or servicing of any components where contact with bare, live hazardous parts could occur, possibly resulting in electric shock, should only be performed by qualified service personnel.

Electrical Ground Hazards

Do not under any circumstances operate the system until an electrical ground is provided and the power cord is properly connected to the ground.

Flammable Materials Hazards

Do not use this system in the presence of flammable materials.
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-4 October 2005

Hazardous/Biohazardous Substances Hazards

When handling a spill of blood or other potentially hazardous substances, clean up the spill by using a 10% bleach solution, or use your laboratory decontamination solution. Then follow your laboratory procedure for disposal of hazardous materials. If the UniCel DxC system needs to be decontaminated, call your Beckman Coulter Service Representative for assistance.

ISE Module Hazards

When lowering the ISE module, keep fingers clear to avoid a pinching hazard.

Moving Parts Hazards

Do not place hands near any moving parts while the system is in operation.

No Foam Reagent Pressurization Hazard

The No Foam container is pressurized during system operation and must be properly depressurized prior to servicing to avoid sudden depressurization and potential exposure of the skin or eyes to the No Foam solution. To release the air pressure, disconnect the white quick connector located at the supply side of the container. If inhaled, move exposed individual to fresh air. If skin or eye contact with the solution occurs, flush the affected area thoroughly with water for at least 15 minutes. In both cases seek medical attention. Refer to the Antifoam Material Safety Data Sheets (MSDS) for additional information.
Hazardous/Biohazardous Substances Hazards
Summary of Hazards
1

Power Cord Hazards

Only use a three-pronged power cord to connect the instrument to a matching three­wire grounded outlet. Do not use an adapter to connect the power plug to a two­pronged outlet.

Service Procedures Hazards

Disconnect the power cord when performing service procedures such as replacing electronic or mechanical components.
The reagents and other chemical preparations used with the system will not normally cause adverse reactions; however, those persons with sensitive skin should wear protective rubber gloves before attempting to work with those substances.
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-5
Summary of Hazards

Sodium Azide Preservative Hazards

Sodium Azide Preservative Hazards
Reagents, calibrators and controls used with the system may contain small quantities (< 0.1%) of sodium azide preservative. Sodium azide preservative may form explosive compounds in metal drain lines. Refer to National Institute for Occupational Safety and Health Bulletin: Explosive Azide Hazards (8/18/76). Avoid skin contact with reagent. Use water to wash reagent from skin. Refer to the related Material Safety Data Sheets (MSDS) for additional information.

System Operations and Specifications Hazards

System operation should be consistent with the power requirements as stated in the
Summary of Hazards section of this chapter, and should always conform to the
procedures and safety warnings throughout this manual.
If the equipment is used in a manner not specified by Beckman Coulter, Inc., the protection provided by the equipment may be impaired.

Waste B Disposal Hazards/Biohazards

The Waste B collection bottle contents are considered biohazardous and should be handled appropriately.
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-6 October 2005

Summary of Precautions

Introduction

This section summarizes the precautions that should be taken when operating the DxC System. Individual precautions associated with a specific procedure in this manual are included in Caution boxes within the procedures for that task. Please read this section and the preceding Summary of Hazards before operating the system.

AccuSense™ Glucose Sensor Precautions

Whenever the sensor is replaced, reaction cup and stir bar cleaning is recommended. Remove the sensor prior to removing stir bar. To prevent damage to the glucose oxygen sensor membrane tip, do not insert stir bar removal tool, applicator stick, or any other object into the glucose reaction cup unless the sensor has been removed. Do not touch membrane tip of the AccuSense glucose oxygen sensor.

Air Filter Cleaning Precautions

Summary of Precautions
Introduction
1
Do not place a damp filter back on the system. Residual moisture may damage the system.

Alkaline Buffer Stability Precautions

The alkaline buffer reagent is stable for one month on the system. However, if a color change from pink to a lighter shade of pink should occur, replace the alkaline buffer with a fresh bottle of reagent.

Aqueous Calibrator Precautions

Repetitive refrigeration of Synchron aqueous calibrators may facilitate crystal formation. Once removed from refrigerated storage, these calibrators should remain at room temperature. The calibrator once opened is stable for the period claimed in the accompanying package insert.

Bar Code Label Precautions

A misread label can cause one sample ID to be read as another. The laboratory’s process for printing, placing, and meeting all bar code specifications is important to achieve highly accurate readings. Always follow the bar code label specifications to avoid misread labels.

Beckman Coulter Microtube™ Precautions

• Beckman Coulter Microtubes™ are designed for use on specific Synchron systems. Using the appropriate Microtube is essential for proper system operation.
• The sample height in the Microtube is critical for correct sample aspiration on all Synchron systems.
• The use of Array® Microtubes™ (P/N 448163 or P/N 448162) on Synchron Systems or the use of Synchron Microtubes (P/N 756776) on Array systems may result in short sampling, incorrect results, and/or sample probe damage.
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-7
Summary of Precautions

Biohazard Precautions

• The use of non-Beckman Coulter, third party Microtubes, which have not been designed and tested on Synchron Systems may result in system damage and/or short sampling.
Biohazard Precautions
All biohazard precautions should be observed when doing maintenance, service, or troubleshooting on the system. This includes but may not be limited to wearing gloves and eye shields, and washing hands after working on contaminated portions of the system.

Blood Barrier Collection Tube Precautions

When blood collection tubes that contain physical barriers are used, extra care should be exercised to ensure that the barrier is tightly packed. Loose particles from the barrier could coat or plug the sample probe, flow cell, chemistry modules, electrolyte injection cup (EIC), or cuvette wash station.

BUNm/UREAm Electrode Precautions

Loosen the BUNm/UREAm electrode retainer nut, approximately one-half turn, and withdraw electrode slightly before removing the stir bar and make sure the BUNm/ UREAm electrode is slightly withdrawn before reinstalling the stir bar. This prevents possible damage to the delicate electrode tip. After the stir bar is reinstalled, tighten the retainer nut to prevent leakage.

CC Subsystems Priming Precautions

If any two of the CC Subsystem items are selected, all three will prime automatically. Check to make certain that all three areas are ready to be primed if any two are selected.

CO2 Membrane Replacement Precautions

Do not touch the membrane surface when installing the quad-ring.

Covers, Doors and Shields Precautions

To ensure optimum performance of the system, operate the system with reagent doors and all shields and covers in place. To prevent possible motion errors, verify the proper positioning of any removed and reinstalled cover or shield.

CTS (Closed Tube Sampling) Cap Piercer Precautions

Do not use the Cap Piercer with foil-capped tubes. This will give level sense errors.
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-8 October 2005

CTS Tracking Loss Precautions

For systems with 1-Blade Thick CTS, if there is an unusual loss of network communication, then a message tells you that if you move a tube to a different instrument, remove its cap.
• A normal shutdown or reboot does NOT give this CTS message.
• When there is no CTS Tracking, if the cap were kept on a previously-pierced tube:
- it would be pierced again and
- pieces of rubber could fall into the sample.
• You will be notified when full CTS Tracking is restored.

Diethylamine HCL and the Calcium ISE Precautions

Do not use controls containing diethylamine HCL. This adversely affects the calcium Ion-Selective Electrode (ISE).

Electrostatic Discharge (ESD) Hazards

To prevent damage due to electrical static discharge (ESD), always wear the wrist ground strap when cleaning the probes, Flow Cell, ISE Drain or CC sample, Glass Cuvettes and reagent mixers; removing, installing or flushing the Flow Cell Electrodes; cleaning the EIC Ports; bleaching CO2 Alkaline Buffer Reagent Straw and
Line; removing the CO2 Measuring Electrode Membrane; replacing EIC Quad-Ring, Ratio Pump Quad-Rings, Alkaline Buffer Peri-Pump Tubing or Sample and Reagent
Syringes.
Summary of Precautions
CTS Tracking Loss Precautions
1

Environmental Conditions Precautions

Changes in ambient temperatures and environmental conditions may result in a "reference drift" message. In this case, the electrolyte chemistries must be recalibrated.

Fibrin Clots Precautions

Samples should be free of all visible fibrin. Clots could coat or plug the sample probes, flow cell, chemistry modules, electrolyte injection cup (EIC), or cuvette wash station leading to instrument malfunction and/or short sampling.

ISE Reagent Precautions

Failure to operate the system with sufficient ISE reagent will result in erroneous chemistry results. In some cases, results will be obtained without reagents. Therefore, before starting any run, verify that sufficient reagent is available to complete the run.

MC Reagent Stir Bar Precautions

When cleaning the MC Reagent Lines, Cups, and Stir Bars, The stir bar may rise in the cup due to air accumulation in the lines. Verify that the stir bar is positioned down into the bottom of the cup.
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-9
Summary of Precautions

Narrow Margin Bar Code Precautions

Narrow Margin Bar Code Precautions
The sample bar code reader on the DxC can read narrow-margin bar codes. Because of the sensitivity needed to read narrow-margin bar codes, the labels must be high quality. They must be free from smudges, spots or other imperfections. An imperfection could be read as part of the Sample ID. This could cause an inaccurate read of the bar code.

Racks Replacement Precautions

Racks should be replaced every five years. Damaged racks should not be used on the system or in the SPINCHRON™ Centrifuge.

Ratio Pump Quad-Ring Replacement Precautions

Be careful not to twist quad-rings or O-rings during installation, as this could result in reagent leakage or even in ratio pump failure.

Reagent Volume Precautions

Check reagent volumes before starting a run. Failure to operate with sufficient reagent will result in erroneous chemistry results. In some cases, results will be obtained without proper amounts of reagents in the modules. Therefore, before starting any run, verify that sufficient reagent is available to complete the run.
Modular chemistry reagent containers should not be handled while the system is performing modular chemistry measurements.
The use of expired reagents may cause erroneous results.

Residual Blood Contamination Precautions

Before you place your validated closed tubes on the DxC, check the top of the cap for any residual blood. Residual blood contamination into the sample could affect results. If blood is present, remove it by using a cotton-tipped applicator stick moistened with DI water. When running in the CTS mode, if tubes off-loaded from the UniCel DxC Systems have water or droplets of water on the caps, disable the CTS and call Beckman Coulter Technical Support.
Note: Oil on a cap is normal.

Reuse of Sample ID Precautions

If your LIS or normal workflow requires the reuse of sample IDs, the sample programming should be cleared from the DxC at a time interval that is less than the shortest time of sample ID reuse. Failure to heed this warning will result in new requests being merged with tests from incomplete samples that previously used that ID.
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-10 October 2005

Sample Bar Code Reader Precautions

Do not tamper with or remove the housing of the Sample Bar Code Reader.

Sample Syringe Replacement Precautions

When installing the syringes, do not mix the two different syringe sizes. The MC and CC sample syringes (100 µL) are located on the back wall of the instrument. The CC reagent syringe (500 µL) is located on the right side of the instrument.

Samples NOT received by Host Precautions

Do not clear samples until results are received at the host and printed. Clearing samples manually or through host programming, may cause results to be received at the host and printed at the DxC without the sample ID. Depending on the host implementation, this can cause lost sample results or sample results which merge with other sample results producing duplicate tests or added tests.

Samples Use Precautions

Do not use the same sample run on an DxC system for analysis of analytes for which a small quantity of carryover could greatly increase the results (for example, TBhCG).
Sample Bar Code Reader Precautions
Summary of Precautions
1
If your system has a Closed Tube Sampling (CTS) option, Do not run samples in cups on top of a sample tube. This would give motion errors. As an alternative, put the sample in a nesting cup on a tube in a reserved rack.

System Backup Precautions

After inserting a diskette into the floppy disk drive, selecting <OK> when prompted will prepare the floppy disk by ERASING it before copying data. Be sure the diskette does not contain critical data that is not available from another source.

System Configuration Change Precautions

Changes to the System Configuration Data should only be done at the request or at the direction of Beckman Coulter, Inc. Entry of incorrect information will lead to system errors.

System Restore Precautions

System Parameter and Alignment data can be restored from the backup floppy disks onto the system, however performing the Restore function deletes some or all files (depending on the areas restored) from the hard drive.

Urine Sample Precautions

After analysis of ten consecutive urine electrolytes, run one replicate of electrolytes on Synchron Calibrator Level 2 in the serum mode. This will minimize the potential for chloride drift due to matrix effects of urine samples.
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-11

Symbols and Labels

Introduction

Symbols and Labels
Introduction
The following is a list of symbols and labels used on the DxC Systems. You will find them affixed to the appropriate components of the system, as described briefly below.

Instrument Power Switch, ON

This symbol located on the main power switch indicates that the analyzer power is ON when this portion of the switch is in the down position.

Instrument Power Switch, OFF

This symbol, also located on the main power switch, indicates that the analyzer power is OFF when this side of the switch is in the down position.

Instrument or Printer Power Switch, ON

This symbol is located on the analyzer and printer power switch. When the portion of the switch with this symbol on it is in the down position, power to the monitor or printer is ON.

Monitor Switch, ON/OFF

This symbol is located on the monitor power switch. A green light to the left of this symbol indicates the power is ON.
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-12 October 2005

CPU Power OFF Switch

This symbol is located on the face of the Computer (CPU) unit and indicates the OFF state when pressed.

Primary Electrical Ground

This symbol is used to indicate an electrical ground.

Keyboard Connection

Symbols and Labels
CPU Power OFF Switch
1
This symbol is found above the connection between the computer and the keyboard.

Can Hold This Object Here

This black symbol, located on the bottom of each sample and reagent probe assembly, indicates that this area may be handled to rotate the probe.

Do Not Hold This Object Here

This red symbol, located on the top of each sample and reagent probe assembly, indicates that this area may not be handled.
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-13
Symbols and Labels

Mouse Port Connection

Mouse Port Connection
This symbol is found next to the connection between the computer and the mouse port.

High Voltage Electric Shock Risk

This symbol indicates high voltage is present and /or there is a risk of electric shock when working in this area.

CAUTION

This symbol indicates a caution message and is followed by an explanation or other symbols that define the caution (see examples below).

CAUTION Operate with All Covers in Place

This symbol is located on top of the work surface cover and the cover of an optional Cap Piercer. It indicates a caution to operate only with all covers in place to reduce risk of personal injury or biohazard.
456161-B

General Biohazard

These caution symbols indicate biohazardous risk from possible patient specimen contamination.
CAUTION
TO REDUCE RISK OF PERSONAL INJURY,
OPERATE ONLY WITH ALL COVERS IN PLACE.
A011459L.EPS
A011460L.EPS
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-14 October 2005

Laser Bar Code Caution

A label reading, "CAUTION. LASER LIGHT ACCESSIBLE. WHEN COVER IS OPEN OR REMOVED, DO NOT STARE INTO BEAM." is placed on the cover of any laser-based code reader. Do not stare into laser light beam when cover is open or removed.

Class II Laser Caution Warning

A label reading, "CAUTION. LASER LIGHT - DO NOT STARE INTO BEAM. 670nm - 1mW CLASS II LASER PRODUCT." is placed near any opening through which a bar code reading beam will emit. Do not stare into laser light beam.
Symbols and Labels
Laser Bar Code Caution
1

Laser

A label reading, "AVOID EXPOSURE. LASER LIGHT IS EMITTED FROM THIS APERTURE." is placed near any opening through which a bar code reading beam emits. Avoid exposure to laser light emitted.

Sharp Objects Caution

A label reading, "CAUTION. SHARP OBJECTS - REFER SERVICING AND MAINTENANCE TO QUALIFIED SERVICE PERSONNEL." is found on top of the optional cap piercing hardware which is located underneath the removable cover of the cap piercing tower.
CAUTION
SHARP OBJECTS - REFER
SERVICING AND MAINTENANCE TO
QUALIFIED SERVICE PERSONNEL.
A012936L.EPS
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-15
Symbols and Labels

ISE Cover Caution

ISE Cover Caution
A label reading, "THE ISE COVER SHOULD REMAIN IN PLACE DURING SYSTEM OPERATION." is placed on top of the ISE module frame under the ISE cover to indicate that the ISE cover should remain in place during system operation.
THE ISE COVER SHOULD REMAIN IN PLACE DURING SYSTEM OPERATION.
471830-AA

Moving Parts Caution

A label reading, "CAUTION PARTS MOVE AUTOMATICALLY" is placed inside the offload track, on the left side of the back wall.
A012937L.EPS
CAUTION PARTS MOVE AUTOMATICALLY

MC Door Caution

A label reading, "THIS DOOR SHOULD REMAIN CLOSED DURING SYSTEM OPERATION." is found on the top edge of the left hand (MC reagent) door and indicates that the door should remain closed during system operation.
THIS DOOR SHOULD REMAIN CLOSED DURING SYSTEM OPERATION.
471831-AA

Read Manual Caution

A label reading, "CAUTION - READ MANUAL BEFORE OPERATING." is found on the hydropneumatics behind the center door and recommends the operator read the manuals before operating the system.
A015047L.EPS
CAUTION - READ MANUAL
BEFORE OPERATING
ATTENTION - CONSULTER LA NOTICE
AVANT DE FAIRE FONCTIONNER.
270-455774-A
A012939L.EPS
A012938L.EPS
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-16 October 2005

Rack Loading Label

This label is found to the right of the Run and Priority Load buttons and indicates the correct position of a rack for placement into the autoloader.

Laser Certification Label

This label is found on the back, bottom edge of the system. It provides information about the laser.
Symbols and Labels
Rack Loading Label
1

Ethernet/Serial Port Label

This label is found on the right side of the system and identifies connections for the Ethernet and serial ports.
PRODUCT COMPLIES WITH 21 CFR CHAPTER I, SUBCHAPTER J
MANUFACTURED DECEMBER 2004
LABEL P/N 448229 AB
BECKMAN COULTER, INC
SERIAL PORT
MADE IN U.S.A. MARCA REG
ETHERNET
S
A011540L.EPS
A012942L.EPS
UniCel DxC Systems Instructions For Use A13914 General Information October 2005 Page 1-17
Symbols and Labels

Fluid Interface Label

Fluid Interface Label
This label is found on the center, back, bottom edge of the system. It identifies inlet and outlet ports on the system.
136

Recycling Label

This symbol is required in accordance with the Waste Electrical and Electronic Equipment (WEEE) Directive of the European Union. The presence of this marking on the product indicates:
1. the device was put on the European Market after August 13, 2005 and
2. the device is not be disposed via the municipal waste collection system of any
WASTE B
SENSOR
WASTE B OUTLET
D. I. WATER INLET
MAX. PRESS. 100 PSI (689 kPa)
WASTE OUTLET
member state of the European Union.
137128
D.I. FIBER
A012943L.EPS
A016608L.EPS
It is very important that customers understand and follow all laws regarding the proper decontamination and safe disposal of electrical equipment. For Beckman Coulter products bearing this label please contact your dealer or local Beckman Coulter office for details on the take back program that will facilitate the proper collection, treatment, recovery, recycling and safe disposal of device.
General Information UniCel DxC Systems Instructions For Use A13914 Page 1-18 October 2005

System Description

Introduction

The following chapter briefly describes the system components, operational theories, principles of measurement, programming structure, and operator controls. Detailed information is located in the UniCel DxC Synchron Clinical Systems Reference Manual.

CHAPTER 2 System Description

System Description
Introduction
2
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-1

Operational Conditions

Shipping Damage

Operational Conditions
Shipping Damage
Each DxC System is carefully examined and checked by Beckman Coulter, Inc. before it is shipped. When you receive your new DxC System, visually inspect the shipping container for damage. If there is damage, notify the Beckman Coulter Service Representative before he or she arrives at your facility to install your system.

Installation

Refer to the following table.
Table 2.1 Installation Requirements
Item Requirement
Installed by The DxC system is installed completely by Beckman
Coulter
Installation Category II

Clearances

The system requires the following clearances.
Table 2.2 System Clearances
Area Affected Clearance Needed
Left Side Minimum of 6 inches (15.2 cm) clearance or 12 inches
Right Side Minimum of 18 inches (45.7 cm) clearance when monitor
Back Zero inches. The venting design of the system allows for
Top Minimum of 22 inches (55.9 cm) from highest point of
Front Minimum of 25 inches (63.5 cm) to open doors.

Sunlight and Drafts

(30.5 cm) to access smart modules.
on swing arm is in use.
"0 inch" clearance at the back.
system.
Do not place the system in direct sunlight or in drafts. Both of these conditions may affect the temperature control of the system.

Drain

The system should be located near a sink or floor drain to accommodate the waste effluent at a minimum rate of 6 liters/hour (16 liters/hour continuous flow).
The drain must not be placed any higher than 36 inches (91.4 cm) above the floor.
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-2 October 2005

Power Requirements

Table 2.3 Power Requirements
Operating range 180–264 V AC, RMS
Frequency 47–63 Hz (50/60 Hz nominal) BTU generated DxC 600: 7,200 BTU/hour at nominal (estimated)
Power connector 20 A current rating. Nema L6-20R twistlock Uninterrupted power supply Refer to UPS manufacturer’s literature for
Item Requirement
220 V AC nominal, approximately 15 A at low line, exclusive of power on surge
DxC 800: 7,500 BTU/hour at nominal (estimated) DxC PC: 979 BTU/hour
specifications
Operational Conditions
Power Requirements
2
Notes on the System Power
The system can operate from any standard 3-wire electrical outlet and is wired as shipped from the factory to operate on 220V AC, 50/60 Hz.
Line Voltage from the electrical outlet should be free of spikes, fluctuations, and dropouts for protection of the electronic circuitry.
Only operate the system from a 3-wire power source. DO NOT use a 2-prong adapter or a 2-wire AC power source.

Environmental Conditions

Table 2.4 Temperature, Humidity and Elevation
Item Specification
Environment Indoor use only Ambient temperature +18°C to +32°C
NOTICE
CAUTION
Warm-up time 30 minutes (time to reach operating temperature) Relative humidity 20–85% relative, non-condensing Elevation Up to 1,280 m (4,200 ft.)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-3
Operational Conditions

Water Requirements

Water Requirements
Table 2.5 Water Requirements
Item Specification
Flow Rate 0.6 L/min peak flow rate
16 L/hr, minimum continuous flow rate Temperature +15°C to +25°C Water quality NCCLS Type II deionized water, except for total
bacteria count < 10 cfu/mL Water pressure Deionized water entering the system must be 30–90
psig.

Other System Specifications for IEC-1010 Compliance

Table 2.6 IEC-1010 Specifications
Item Specification
Pollution Degree 2 EN55011 Meets Class A Maximum Sound Pressure 65 dBA average over 8 hours with covers down at 1
meter away from the instrument at 25°C Maximum Leakage Current DxC 600: 222 µA at 240V, 50Hz
DxC 800: 240 µA at 240V, 50Hz
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-4 October 2005

System Components

DxC Systems

A UniCel DxC System can be divided into the following components:
• Sample Handling Components
• Modular Chemistry System
• Cartridge Chemistry Reagent Handling System
• Hydropneumatic System
• Operation and Control Components
A Closed Tube Sampling System has optional components listed below:
System Components
DxC Systems
2
• Large Particle Immuno Assay Module (LPIA)
• Closed Tube Sampling (CTS)
1
3
1. Modular Chemistry (MC) Section
2. Cartridge Chemistry (CC) Portion
3. Autoloader
2
54
4. Reaction Carousel
5. Dual Reagent Carousel
6. Operator Console
*
6
A015900P.EPS
Figure 2.1 UniCel DxC 600 Analyzer
*
Equivalent to Near Infrared Particle Immuno Assay (NIPIA).
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-5
System Components
DxC Systems
Figure 2.2 UniCel DxC 800 Analyzer
A015901P.EPS
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-6 October 2005

Sample Handling System

Introduction

The Sample Handling system is composed of the following components:
• Sample racks
• Autoloader/Offload track
• Priority load position
• Shuttle
• Bar code reader
• Cap piercer assembly (optional)
• Sample Carousel
• Sample probe/mixer assemblies
• Wash cup assemblies
• Obstruction detection assemblies
Sample Handling System
Introduction
2
The sample handling module is used to load samples onto the system, provide samples for analysis, and provide temporary storage of completed samples. A detailed description of each component is presented in the following paragraphs.

Sample Racks

A sample rack is a high-strength, plastic, centrifugable holder designed to house up to four samples. There are four sizes of racks with each size capable of holding one length/width combination of primary sample tubes in addition to accepting sample cups. (Refer to Figure 2.3.)
Sample racks accept the following tube and cup sizes:
Table 2.7 Sample Racks
NOTICE Adapters are provided to adapt various sized sample tubes (secondary tubes) to the short racks. These adaptors must only be used in racks designated as reserved. The reserved rack feature is described in CHAPTER 4,
Options.
System Setup
Rack Accepts These Cups and Tubes
13 × 75 12 × 75 mm tubes
13 × 75 mm tubes
0.5 mL cups
2.0 mL cups
(1 of 2)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-7
Sample Handling System

Rack ID Labels

Table 2.7 Sample Racks, continued
Rack Accepts These Cups and Tubes
16 × 75 16 × 75 mm tubes 13 × 100 13 × 100 mm tubes
16 × 100 16 × 100 mm tubes
0.5 mL cups
2.0 mL cups Capillary collection tubes (use with the capillary tube adapter) Beckman Coulter Microtubes™
16 × 92 mm tubes
16.5 × 100 mm tubes Beckman Coulter 0.5 mL Cup Insert (P/N 476399)
Rack ID Labels
Sheets of bar-coded rack ID labels are supplied with the system. They can be applied as shown in Figure 2.3.
(2 of 2)
1. Numeric Rack ID number
2. Rack size label
3. Bar coded rack ID label
Figure 2.3 Rack
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-8 October 2005

Autoloader/Offload Track

When viewed from the front of the system, the autoloader is on the left and holds up to 25 sample racks in preparation for presentation to the DxC 800 system. The DxC 600 system has room to load a maximum of 14 racks. (Refer to Figure 2.4.)
There is also space for 25 sample racks in the offload track as they are removed from the Sample Carousel upon completion. Combined, these two components allow for over one hour of uninterrupted sample processing, with no operator intervention involved.
When loading racks onto the autoloader, make sure that they are placed firmly down into the autoloader.
1
NOTICE
Sample Handling System
Autoloader/Offload Track
2
2 3 4 8765
1. Pushers
2. Autoloader
3. Run button
4. Priority load button
Figure 2.4 Sample Loading Area

Priority Load Position

Between the Autoloader and the Sample Gate is the Priority Load position. This position is used in conjunction with the PRIORITY LOAD button when a rack is to be loaded onto the Sample Carousel into a reserved priority position so that it can be run in a higher priority than other racks on the Autoloader. (Refer to Figure 2.4.)
A015902P.EPS
5. Sample gate
6. Shuttle
7. Offload track
8. Stop button
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-9
Sample Handling System

Priority Load Button

Priority Load Button
Typically, rack placement and removal is under microprocessor control. The operator may choose to use the reserved positions by pressing the PRIORITY LOAD button and placing the priority rack in the space provided by the system. (Refer to Figure 2.4.) The rack will load into one of the reserved positions on the Sample Carousel.
PRIORITY LOAD only prioritizes the loading of the rack. It does not alter the sample priority (STAT or routine) previously designated in Sample Programming.

Pushers

Pushers collect and move to the Sample Gate any racks loaded onto the system. They are activated when the RUN button is pressed. (Refer to Figure 2.4.)

Sample Gate

The Sample Gate is the mechanism that moves racks from the load tray to the shuttle during the load process. It also moves samples from the shuttle to the unload track during the unload process. (Refer to Figure 2.4.)

Shuttle

The Shuttle moves the rack from the gate area onto the Sample Carousel. (Refer to
Figure 2.4.)

Bar Code Reader (Sample)

Do not tamper with or remove the housing of the Sample Bar Code Reader.
The Bar Code Reader is a Class II fixed-beam laser scanner. It is used to read the rack bar code, the sample bar code (if present), and the background bar codes as the rack travels past. The rack bar code and sample bar code (if present) are used to identify the sample and link it to the appropriate sample programming.
There are two background bar codes that are used to determine whether a rack position is empty or occupied, and if occupied, whether the sample is in a cup or tube.
Refer to "Symbols and Labels of the CAUTION labels for the bar code reader.
CAUTION
" in CHAPTER 1, General Information, for a description
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-10 October 2005

1-Blade Thick CTS (Closed Tube Sampling) Cap Piercer Assembly (optional)

1-Blade Thick CTS (Closed Tube Sampling) Cap Piercer Assembly (optional)
CAUTION
This Cap Piercer contains a razor sharp blade assembly.
CAUTION
Do NOT use this Cap Piercer assembly with foil-capped tubes. This would give level sense errors.
This optional 1-Blade Thick CTS Cap Piercer assembly pierces thick-stoppered tubes allowing the sample probe access to the sample without the need for cap removal. Tubes with caps must be loaded in racks that have had the size correctly defined and that are not Reserved (Reserved Racks = No CTS). Both open and closed tubes can be in the same sample tube rack.
Table 2.8 Tubes Validated for Closed Tube Sampling
Sample Handling System
2
Cap Piercer
Tube Type Tube Size
Configuration
1-Blade Thick CTS
Becton Dickinson VACUTAINER®a with HEMOGARD™
Greiner VACUETTE
a
®a
13 × 75 mm 13 × 100 mm 16 × 100 mm
13 × 75 mm 13 × 100 mm
Sarstedt S-Monovette
®ab
75 × 15 mm
92 × 15 mm
a
All trademarks are the property of their respective owners.
b
This tube requires a special 5.5 mL rack (P/N A18642).
Refer to "Symbols and Labels" in CHAPTER 1, General Information, for a description of the CAUTION labels for the Cap Piercer.
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-11
Sample Handling System

Sample Carousel

Sample Carousel
The ten-rack position Sample Carousel is a motor-driven turntable. (Refer to Figure
2.5.) Under normal operation, eight of the Sample Carousel positions are available for
routine processing and two positions are reserved for priority racks.
1
2
3
4
1. Liquid level sense assembly
2. Sample probe (CC)
3. Collar wash
4. Sample Carousel
Figure 2.5 Sample Carousel Area
A015903P.EPS
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-12 October 2005

Modular Chemistry (MC) System

Introduction

The Modular Chemistry system consists of the following major assemblies:
• Reagent storage area
• Ratio pump
• Sample probe
• Electrolyte injection cup (EIC)
• Flow cell assembly
• Chemistry reaction modules
A detailed description of each component is presented in the following paragraphs.

Reagent Storage Area

Modular Chemistry (MC) System
Introduction
2
The reagent containers used to supply the modular chemistries are located behind the left front door of the system. The only exception is the CO2 alkaline buffer which is
located on the ISE module.
1
2
A015906P.EPS
1. Bar code reader (hand held)
2. Reagent storage area
Figure 2.6 Modular Chemistry Reagent Storage Area
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-13
Modular Chemistry (MC) System

Reagent Bar Code Reader (Modular)

Reagent Bar Code Reader (Modular)
Behind the left side door of the system there is a hand-held bar code reader. (Refer to
Figure 2.6.)
When this reader is held up to the label of one of the modular reagents and the trigger is pressed, the identity of the reagent, the lot number and reagent volume are automatically entered into the reagent load screen.

Ratio Pump

The Ratio Pump is a motor-driven, multicylinder, positive-displacement pump used to deliver the necessary reagents to the ISE flow cell. (Refer to Figure 2.7.) It consists of a three-step piston housed in three, stacked, independent cylinders.
54321
1. Solenoid valve (example)
2. Cylinder 1
3. Cylinder 2
Figure 2.7 DxC 800 Ratio Pump
6
A015904P.EPS
4. Cylinder 3
5. Outlet line (example)
6. Inlet line (example)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-14 October 2005

Electrolyte Injection Cup (EIC)

The EIC mixes the sample and buffer prior to delivery of the sample (now diluted) to the flow cell.
Modular Chemistry (MC) System
Electrolyte Injection Cup (EIC)
2
1. Waste outlet
2. Flow Cell outlet
3. Buffer inlet
4. Reference inlet
5. DI H
Figure 2.8 Electrolyte Injection Cup
O inlet
2
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-15
Modular Chemistry (MC) System

Flow Cell Assembly

Flow Cell Assembly
The flow cell assembly houses the seven electrodes that perform the analysis of sodium, potassium, chloride, carbon dioxide, and calcium. (Refer to Figure 2.9.)
1. Inlet port
2. CL electrode
3. K electrode
4. CO
5. CO
reference electrode
2
electrode
2
Figure 2.9 Flow Cell
6. Exit port for waste (large tube)
7. Exit port for internal reference
8. Na reference electrode
9. Na electrode
10. Ca electrode
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-16 October 2005

Chemistry Reaction Modules (Basic Components)

Each of the six Chemistry Reaction Modules have similarities in their design. These common elements are described below. (Refer to Figure 2.10.) Unique design elements of the modules are described under the specific module headings later in this section.
2
Chemistry Reaction Modules (Basic Components)
Modular Chemistry (MC) System
3
4
2
1
1. Circuit board (behind protective shield)
2. Reaction cup
3. Reaction cup housing
4. Mixer motor assembly
5. Reagent pump assembly
5
A015905P.EPS
Figure 2.10 Basic Components of Modules (Ex: Albumin)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-17

Cartridge Chemistry (CC) Reagent Handling System

Introduction

Cartridge Chemistry (CC) Reagent Handling System
Introduction
The Cartridge Chemistry Reagent Handling system is composed of the following components:
• Reagent cartridges
• Reagent carousel
• Reagent probe assembly
• Reagent mixer assembly
• Mixer wash cup
The Reagent Handling system is used to transfer reagent from the individual cartridges to the reaction cuvettes for processing and analysis of the requested chemistry tests.

Reagent Cartridges

Reagent cartridges are single use, recyclable plastic containers that house the individual liquid reagent components necessary to perform a chemistry test. The reagent carousel is capable of storing 59 cartridges on board.
1. A compartment
2. B compartment
3. C compartment
Figure 2.11 CC Reagent Cartridge
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-18 October 2005

Reagent Carousel

The Reagent Carousel Compartment provides an on-instrument storage area for the individual reagent cartridges. A total of 59 reagent cartridges can be stored in the carousel at one time. (Refer to Figure 2.12.)
The storage compartment is refrigerated and fan-cooled to maintain a temperature of +5°C (±3°C).
Cartridge Chemistry (CC) Reagent Handling System
1 2
3
4
Reagent Carousel
2
1. Top positions 31–59
2. Top Bar Code Reader
3. Bottom positions 1–30
4. Bottom Bar Code Reader
Figure 2.12 CC Dual Carousels with Two Bar Code Readers

Reagent Bar Code Reader (CC)

The Bar Code Reader, (refer to Figure 2.12), situated near the front of the Cartridge Chemistry reagent access door, scans each label during the loading of reagent cartridges. Only the reader for the selected carousel is active. An audible signal acknowledges successful reading of the label.
The Reagent Probe assembly consists of a mechanical structure that supports two moveable cranes. Attached to each crane is a pickup probe. (Refer to Figure 2.13
A007408P.EPS
.)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-19
Cartridge Chemistry (CC) Reagent Handling System

Reagent Mixer Assembly

1
1. Reagent Mixer
2. CC Reagent Probe A
3. CC Reagent Probe B
4. Collar Wash
2
4
3
A015908P.EPS
Figure 2.13 CC Reagent Probe Area
Reagent Mixer Assembly
This assembly consists of a mechanical structure that supports a single, moveable crane. Attached to the crane is a high-speed mixer. (Refer to Figure 2.13.)

Reagent Mixer Wash Cup

The Mixer Wash Cup sprays the mixer with diluted wash solution while the mixer moves up and down in the cup. (Refer to Figure 2.13.)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-20 October 2005

Cuvette Reaction System

Introduction

The Cuvette Reaction system consists of the following components:
• Reaction carousel assembly
• Photometer assembly
• LPIA (Large Particle Immuno Assay) or NIPIA (Near-Infrared Particle Immuno Assay) module (optional)
• Cuvette wash station
The Cuvette Reaction system involves the process of obtaining absorbance readings from each cuvette during the analysis cycle.
Following the completion of each chemistry test, the cuvettes are processed through a wash station in preparation for the next chemistry.
Cuvette Reaction System
Introduction
2
A description of each component is presented in the following paragraphs.

Reaction Carousel

The reaction carousel assembly, (refer to Figure 2.14), supports a total of 125 cuvettes. Each cuvette is glass with a 0.5 cm path length and is approximately 30 mm high. The cuvettes are non-disposable and have an indefinite life-span on the instrument. Cuvettes only need replacement if they are damaged. (They remain under warranty for two years.)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-21
Cuvette Reaction System

Photometer Assembly

1. LPIA module
2. Reaction carousel
3. Photometer
2
3
1
A015909P.EPS
Figure 2.14 Reaction Carousel Area (Typical – Cover Removed)
Photometer Assembly
Attached to the reaction carousel support frame is the Photometer assembly. This consists of a xenon pulse lamp, a discrete 10-position silicon-diode detector array, a monochromator housing unit, and associated electronic circuitry. (Refer to Figure
2.14.)
As each cuvette passes through this optics station during a spin cycle, the xenon lamp is flashed and the resulting light beam travels through the opposing sides of the square cuvette.

LPIA (Large Particle Immuno Assay) or NIPIA (Near-Infrared Particle Immuno Assay) Module (optional)

The optional LPIA module uses a photometric detection system for large particle immuno assays. This assembly is attached to the reaction carousel support frame to the left of the photometer assembly, near the sample carousel. (Refer to Figure 2.14.) It has two printed circuit boards, an LED (light-emitting diode) and a photodetector. Communication with the system software is through fiber optic cables.
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-22 October 2005

Cuvette Wash Station

The Cuvette Wash Station, (refer to Figure 2.15), consists of four coaxial probes, an elevator assembly, and the associated tubing.
A motor controls the vertical motion required by the elevator to raise and lower the probes during the wash stage.
Cuvette Reaction System
Cuvette Wash Station
2
1. Probe #1
2. Probe #2
3. Probe #3
Figure 2.15 Cuvette Wash Station
4. Probe #4
5. Wash station
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-23

Hydropneumatic System

Introduction

Hydropneumatic System
Introduction
The main components of the Hydropneumatic System are mounted on a slide-out drawer that allows for easier operator access. (Refer to Figure 2.16 and Figure 2.17.) When fully extended, the drawer locks open. To close, lift up on the metal tabs, located on each side of the bottom runner of the hydropneumatic unit, and push the drawer inward.
6
5

Function

4
3
2
1
A016497P.EPS
1. Wash concentrate solution
2. No foam reagent
3. DI water reservoir
4. Dilute wash solution reservoir
5. Wash concentrate reservoir
6. Auto-Gloss
Figure 2.16 DxC 800 Hydropneumatics (right side)
The function of the hydropneumatic system is to provide the following media to the different functional areas of the instrument:
• Vacuum
• Compressed air
• Diluted wash solution
• Deionized water
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-24 October 2005
Hydropneumatic System
567
4
3
1
2
A015911P.EPS
Function
2
1. Waste exit
2. Waste B exit
3. DI water inlet On/Off
4. DI water reservoir
5. Vacuum accumulator
6. Waste B canister
7. Waste canister
Figure 2.17 DxC 800 Hydropneumatics (left side)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-25

Operation and Control Components

Operator Controls

Operation and Control Components
Operator Controls
The operator interfaces with various control devices such as the keyboard, monitor and push-button controls during a routine run. Basic operating functions are controlled and reviewed from the monitor. Calibration functions are also controlled from the monitor. Information is selected and entered into the system through touch screen monitors, from a mouse, and/or at a keyboard.
Push-button controls (refer to Figure 2.4) are used to start the process, to prioritize a sample run, or to stop the process under certain conditions as described in Table 2.9 below.
Table 2.9 DxC Push-Button Controls
Push-Button
Primary Function
Control Type
Run To start the test process. Priority To prioritize the loading of a rack by creating a space in front of
the autoloader for loading the next rack into a priority position on the sample carousel. It does not alter the sample priority (STAT or routine) previously defined in sample programming.
Stop To stop the process. The stop button should be used only under
the following conditions:
• To stop instrument motions
• To conduct a maintenance/repair activity
• To home and realign mechanical components without rebooting
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-26 October 2005

Main Operator Screen and Program Structure

Main Operator Screen

DxC System operating and programming functions are initiated from the main operator screen at the DxC analyzer (refer to Figure 2.18). In addition, the screen provides status information to help determine the present state of the system.
Main Operator Screen and Program Structure
Main Operator Screen
2
1
CTS
3
1. CTS Indicator
2. Host Indicator
3. System Status Indicator
4
4. Function Selection Icons
5. Sample Status Icons
6. Sample Status Icon Legend
2
Figure 2.18 UniCel DxC 800 Main Operator Screen
6
5
E016482S.EPS

Accessing the Help System

Help is available online in six supported languages (the default is English). To reach the on-line Help system, select the Question Mark icon ( ) on the right side of the
Function Selection Icon menu. The system will offer three options:
• Install an On-Line Instructions For Use manual (IFU)
• View the installed On-Line Instructions For Use manual
• View the License Agreement
When you access the On-Line Instructions For Use manual screen, the currently loaded language manual version will be show in the Choose a Language dropdown field. To view the IFU Manual in that language, press (on the touch screen) or select (with the mouse) the <View Manual> button. The first page of the IFU Manual will appear.
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-27
Main Operator Screen and Program Structure

Printing Data From a Screen

To load an alternate language, select the desired language from the Choose a Language dropdown list and select the <Install Manual> button. You can then view the IFU Manual in the selected language as described in the preceding paragraph.
To view the License Agreement, simply select the <License> button on the screen.
Printing Data From a Screen
Many screens contain data that may be printed by selecting Print [F10]. Pressing the [Control] and [Print Screen] key simultaneously on the keyboard will also cause the current screen to print, regardless of whether the Print [F10] option is available.

Status Functions

The following Table depicts the status information available from the operator screens of the DxC analyzer.
Table 2.10 Main Operator Screen Status Functions
Status Indicator Status Description
CTS This indicator (refer to item 1 on Figure 2.18) appears in the blue
bar at the top left side of the Operator screens when the Closed Tube Sampling (CTS) option is installed. When "CTS" appears, Closed Tube Sampling is enabled. When CTS appears within the
international "No" symbol ( ), the feature is not enabled. When the indicator is absent, the CTS option is not installed. Enabling and disabling Closed Tube Sampling is accomplished using the Setup procedures described in the "Reserved Racks/
Obstruct Detect" topic of CHAPTER 4, System Setup Options.
NOTICE CTS is an optional feature of the system, which allows the system to pierce primary sample tubes. The operator will only be offered the enable/disable CTS option if a CTS assembly is installed, otherwise that selection will be grayed out on the setup screen. The default for CTS sampling is "OFF."
Host This indicator (refer to item 2 on Figure 2.18) appears on the blue
bar at the top right side of the Operator screens. The indicator to the right of the label shows communication activity between the instrument and the host computer. A blue bar indicates that the host is sending information. A green bar indicates that the host is receiving information.
(1 of 2)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-28 October 2005
Main Operator Screen and Program Structure

Function Selection Icons and Program Structure

Table 2.10 Main Operator Screen Status Functions, continued
Status Indicator Status Description
System Status This indicator appears just below the Function Selection Icons
on the left side of the Operator screens (refer to item 3 on Figure
2.18). It shows the current state of the system: either Running or
Standby. When both the Modular Chemistry (MC) and the Cartridge Chemistry (CC) functions have the same status, a single status indicator appears. When the statuses of these components are different, the MC Status is shown on the left, and the CC Status appears on the right. For example, Running/Standby would indicate that the MC side is Running, and the CC side is in
Standby mode.
(2 of 2)
Sample Status Indicators
When monitoring sample status, a sample status icon (refer to item 5 on Figure 2.22) appears in front of listed sample. Samples are listed in a rack status box directly below the sample status icon legend (6). The sample statuses shown are as follows:
2
Table 2.11 Sample Status Indicators
Status Indicator Status Description
Not Programmed Indicates a sample has been loaded without any programming
associated with that Sample ID.
Query Pending Sample is waiting for specific program information to be
downloaded from the host. In Progress Sample has been identified and is currently being processed. Aspirated Indicates whether an initial aspiration of the sample has been
accomplished. Incomplete Sample has some tests that are still pending.
Function Selection Icons and Program Structure
Near the top of the Operator screens, a series of icons on the touch screen provide access to each of the major functional areas of the system (refer to item 4 on Figure
2.18). The overall program structure is shown in beginning on the following page.
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-29
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
Main
Samples
F1 Results F2 Unload F8 Log F9 Pre Run F10 Post Run
F2 Demog
F3 Options
F4 Batch
F5 Control
F6 Rerun
F7 Clear
F1 Sample F2 Edit F3 Urine F10 Next
Clear OK Smp ID Only
F3 Options F5 Control F6 Remove F8 Rgt Cart F9 Cancel F10 Save
Select All Batch Cancel
Sample IDs Rack(s)/Position(s) Date/Time (From/To) OK Cancel
System Replicates Sample Replicate Serum Index Off-line Dilution Factor Manual ORDAC Test Replicates OK Cancel
Rack End
System Replicates Sample Replicate Serum Index Off-line Dilution Factor Manual ORDAC Test Replicates OK Cancel
F3 Options F7 Rgt Cart F8 Remove F9 Cancel F10 Save
F1 Clear F9 Cancel F10 Program
Sample ID(s) From/To Rack(s)/Position(s) Date/Time (From/To) Status Print Display Cancel
F1 Disk F2 Recall F3 ABS F4 Edit F6 Host F7 Print F8 Prev F9 Next F10 Done
Results
F8 Racks
F9 List
F10 Next
Sample ID(s) Rack(s)/Pos(s) Patient ID Multi-Sample Report Run Date/Time (From/To) F1 Results F7 Sum F8 Host F10 Print
Figure 2.19 Program Structure (Main, Samples and Results)
Sample Required In Progress Complete Incomplete Rerun Manual Assign Cancel
Plot Ta bl e Both Print Next Prev Done
E007323L.EPS
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-30 October 2005
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
2
Rgts/Cal
QC
F1 Load F2 Params F3 No Load F4 Cal F5 List F6 Options
F7 Assign F8 No Cal F9 Sort (CC List) F10 Print
F1 Review
F1 Clear F10 Done
Load calibrator diskette Calibrator acceptance limits Modify set points Slope/Offset adjustment Within lot calibration Enzyme Validator Calibration override Bypass chemistry Extend calibration time Print cal report Cancel
Avail Racks Next Prev Close
Position Chemistry Tests Left Days Left Cal Time Left OK
F2 Control ID F3 Delete F9 Print F10 Done
Current Cal Report Previous Cal Report(s)
F2 Define
F3 Delete
F4 List
F5 Log
F6 Summary
F7 Chart
F8 Archive
F10 Print
F1 Chems F2 Control ID F3 Delete F10 Done
F1 All Chem F2 All File F10 Done
QC Date Range OK Cancel F8 Inter-Lab F9 Print F10 Done
QC Date Range OK Cancel F1 Chems F2 Filenum F7 Prev F8 Next F9 Print F10 Done
Option 1 Archive to Disk Option 2 Review Disk Cancel
QC Date Range OK Cancel F1 Chem F2 Rgt Lot F3 Delete F4 Action F9 Print F10 Done
E007324L.EPS
Figure 2.20 Program Structure (Rgts/Cal and QC)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-31
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
Setup 1. Auto Serum Index/ORDAC F1 Restore
2. Chem Config/Sample Type
3. Date/Time
4. Demographics Setup
5. Immediate Report
6. Panels
7. Replicates
8. Report Setup
9. Reportable Ranges
10. Reference/Critical Ranges
F10 Done
F1 Date F2 Time F3 Restore F10 Done
F1 Restore F10 Done
MC STAT Host/Printer MC ALL Host/Printer Immediate Report of Reruns OK Cancel
F1 Define F2 Delete F3 Default F10 Done
F1 Restore F10 Done
Report Header Patient Format Control Format Inter-Laboratory Information Print Patient and Control Report F1 Restore F10 Done
Define/Edit Cancel
Define/Edit Clear Ranges
F1 Chems F2 UDR F3 Define F4 Clear F5 Insert F6 Delete F10 Done
F1 Restore F8 Prev F9 Next F10 Done
F1 Delete F2 Clear All F3 Default F8 Previous F9 Next F10 Done
Serum CSF Plasma Random Urine Timed Urine Blood Other
E007409L.EPS
11. Sample Comments
Summary by Sample Type Ranges Summary Cancel
F10 Done
Figure 2.21 Program Structure (Setup – Page 1)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-32 October 2005
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
2
Setup, continued
12. Special Calculations F1 View
13. Setup Summary
14. Units/Precision
15. User-Defined Chemistries
16. Barcode
17. Reserved Racks/Obstruct Detect
18. Disable Service Monitor
19. External Water Monitor
20. Host Communications
F2 Define F3 Delete F10 Done
F10 Done
F1 Restore F10 Done
F1 Define F2 Delete F4 Config F9 Print F10 Done
Code 39 Parameters Codabar Parameters Interleaved 2 of 5 Parameters Code 128 Barcode Mode Rack/Position Mode Maximum Sample Program Age (Days/Hours) F10 Done
1-Blade Thick CTS Reserved Racks 75mm Racks 100mm Racks Obstruction Detection Reserved Racks for HbA1c Reserved Racks for IBCT Restore Defaults OK Cancel
F1 Restore F2 Diag F10 Done
F1 Restore F9 Cancel F10 Done
F1 Restore F2 Host Com F10 Done
21. Language/Keyboard
22. Printer Setup
23. Service Setup
24. System Configuration
25. Version Upgrade
26. Status Alarm/Annunciator
27. Chem Update
28. Password Setup
29. Auto Generation of Control
Language Keyboard OK Cancel
Restore Default Printer Type Paper Size OK Cancel
Insert CD
Alarm Test OK Cancel
System
Media
Control Sample Multiple Cartridges F10 Done
DxC
CD Diskette
E007326L.EPS
Figure 2.22 Program Structure (Setup – Page 1 continued and Page 2)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-33
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
Utils 1. Prime F1 Prime
F2 CC
F3 Hydro
F4 MC
Start Prime Stop Prime F10 Exit
All CC Subsystems All Hydropneumatic Subsystems All MC Subsystems Prime DxC Prime CTA Prime CTA Auto-Gloss CTS Auto-Gloss CTS Blade Wash
Reagent Delivery Subsystem Sample Delivery Subsystem Cuvette Wash
Fill Canisters and reservoirs CC Drain waste sump CC Drain waste B sump Drain gravity drain sump
Cup Selection: ALBm (DxC 800 only) CREm (DxC 800 only) GLUCm (DxC 600 only) PHOSm (DxC 800 only) TPm (DxC 800 only) BUNm (DxC 800 only) Prime Cup(s) with: Reagent DI Water Sample Delivery Subsystem ISE All Electrolyte Buffer Electrolyte Reference EIC Wash ISE CO2 Alkaline Buffer Reference and Acid
1. 1-Blade Thick CTS Blade Replacement
2. Maintenance
2. Cartridge Chemistry Probe Cleaning
3. ISE Service
4. Cup Maintenance
5. Hydropneumatic Maintenance
6. Wash All Cuvettes
7. Syringe Rod Replacement
8. CUPs Lamp/Sensor Calibration
9. CC Reagent Wash All Cuvettes
10. Clean Flow Cell and Cups F1 Action F2 Log F10 Exit
Figure 2.23 Program Structure (Utils)
1. Albumin (DxC 800 only)
2. BUNm (DxC 800 only)
3. Creatinine (DxC 800 only)
4. Glucose (DxC 600 only)
5. Phosphorus (DxC 800 only)
6. Total Protein (DxC 800 only) Prime20 Prime5 Drain Rinse Cancel
E007327L.EPS
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-34 October 2005
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
2
Utils, continued
3. Event Log Display
4. Diagnostics
5. Alignment
Copy Time Clear Print Done
1. Chemistry Errors
2. Motion Errors
3. Status Monitor Errors
4. Other Instrument Errors
5. Instrument Events
6. LIS Comm. Errors
7. Other Console Errors
8. Input Device Events
9. Other Console Events
10. Sample Processing Events
11. Deleted Results
Functional Areas
Sub-functional Areas
Test Selection F10 Exit
Align All Reaction Carousel Cuvette Wash Station Reagent Carousel Sample Carousel Sample Rack Handling Sample Rack Barcode Reader Cap Piercer CC Sample Crane or Probe CC Sample Crane Smart Module (B2) MC Sample Crane or Probe MC Sample Crane Smart Module (52) Reagent A Crane or Probe Reagent A Crane Smart Module (C2) Reagent B Crane or Probe Reagent B Crane Smart Module (C3) CC Sample Mixer Crane or Body CC Sample Mixer Smart Module (B3) Reagent Mixer Crane or Body Reagent Mixer Smart Module (C1) CTA Assemblies Procedure Selection - Based on Functional Area selected F10 Exit
Sample System Reagent System Reaction System Hydropneumatic System ISE Module System Modular Cup Chemistry Module Valve and Fluidics
Based on Functional Area selected
Based on Functional Area selected
E007328L.EPS
Figure 2.24 Program Structure (Utils – continued)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-35
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
Utils, continued
Status
Instr Cmd
Help
6. PVT
7. Modem
8. Backup/Restore
9. Metering
10. Touch Screen Calibration
F1 Summary F2 Count F3 Temp F4 Power F5 Hydro F6 Smrt Mdl F7 Cuv Stat F8 CTS F10 Print
1. Home
2. Pause
3. Stop Print
4. Shutdown
5. Pause Waste B
6. Resume Waste B
7. Enable/Disable Modules
8. Unload All
Choose a Language: English (default) Italian Spanish French German Japanese Install Manual View Manual License
Performance Verification Tests Sub-Functional Area Test Selection - Based on Functional Area Selected F10 Exit
System Parameters Alignment Data Files Backup Restore Cancel
F1 Disk F2 Replen F3 Report F4 Off F5 Modem F9 Cancel F10 Done
E007329L.EPS
Figure 2.25 Program Structure (Utils – continued 2, Status, Instr Cmd, and Help)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-36 October 2005

Theory of Operation

Introduction

The UniCel DxC Synchron Clinical Systems are microprocessor-controlled, random access clinical analyzers capable of processing a wide variety of operator-selected chemistries in a single run.

Cartridge Chemistries (CC)

The optical system of the DxC enables rate, endpoint, and nonlinear analyses to be performed simultaneously. These analyses are referred to as cartridge chemistries because the reagents are stored in cartridges.

Cartridge Chemistry (CC): Sample and Reagent Processing

During operation, a number of events occur simultaneously and are under direct control of the instrument microprocessors.
Theory of Operation
Introduction
2
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-37

Cartridge Chemistry: Calibration Theory

Introduction

Cartridge Chemistry: Calibration Theory
Introduction
Calibration determines the relationship between measured reaction responses and known concentrations. Calibration factors are derived from this relationship. These factors are used to convert the measured reaction responses to final concentration results.
Calibrated chemistries include endpoint and first-order rate chemistries, drugs, DATs, and specific proteins. Zero-order rate chemistries include enzymes, which are precalibrated and require no routine calibration. Enzyme verification can be performed on some of the enzymes to conform to International Federation of Clinical Chemistry (IFCC) guidelines at 37°C.

Endpoint and First-Order Chemistries

Calibration of endpoint and first-order rate chemistries involve the use of a single-level calibrator solution or a two level calibrator kit. Each analyte in the calibrator solution has a known concentration value associated with it. With each new lot of calibrator solution, the values are transferred from disk and stored in memory for later use in the calibration procedure.
For most calibrated cartridge chemistries, the system will set calibration factors based on four calibrator replicates per calibrator level. The instrument will determine and discard the highest and lowest of the four replicates. The remaining two values are called the usable calibrator replicates. All four replicates will appear on the report but the average of the two usable replicates is used to determine the calibration factor.
For other chemistries, calibration is based on two calibrator replicates per calibrator level. No replicates will be discarded. The average value of the calibrator replicates is used to determine the calibration factor.
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-38 October 2005

Endpoint and First-Order Calibration Formulas

The calibration factor is determined by using one of the following equation sets where reaction and blank are used from the usable replicates.
Table 2.12 Calculation of Calibration Factors for Endpoint and Rate
Chemistries
Type Formula
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
2
Nonblanked Endpoint Chemistries
FOR HIGH CALIBRATOR LEVEL:
Reaction ABS = ABS Reaction ABS = ABS (ABS
rep1
+ ABS
rep1 rep2
) × 0.5 = ABS
rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
Reaction ABS = ABS Reaction ABS = ABS (ABS
+ ABS
rep1
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × ABS
rep1 rep2
) × 0.5 = ABS
rep2
(lo)
avg
Cal Set Point (hi) – Cal Set Point (lo)
ABS
(hi) – ABS
avg
avg
avg
(hi)]
(lo)
Sample values are calculated by the following equation: (Reaction ABS × Cal Factor) + offset = sample value
E014416L.EPS
NOTICE For single point linear calibration, the low calibrator is a fixed zero point and the offset is equal to zero.
(1 of 6)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-39
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
Table 2.12 Calculation of Calibration Factors for Endpoint and Rate
Chemistries, continued
Type Formula
Blanked Endpoint Chemistries
FOR HIGH CALIBRATOR LEVEL:
(Reaction ABS - Blank ABS) = Delta ABS (Reaction ABS - Blank ABS) = Delta ABS (Delta ABS
+ Delta ABS
rep1
) × 0.5 = Delta ABS
rep2
rep1 rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
(Reaction ABS - Blank ABS) = Delta ABS (Reaction ABS - Blank ABS) = Delta ABS (Delta ABS
Cal Factor (Slope) =
+ Delta ABS
rep1
) × 0.5 = Delta ABS
rep2
Cal Set Point (hi) – Cal Set Point (lo)
Delta ABS
Offset = Cal Set Point (hi) - [Cal Factor × Delta ABS
rep1 rep2
(hi) – Delta ABS
avg
avg
avg
(lo)
(lo)
avg
E014417L.EPS
(hi)]
Sample values are calculated by the following equation: [(Reaction ABS - Blank ABS) × Cal Factor] + offset = sample value
NOTICE For single point linear calibration, the low calibrator is a fixed zero point and the offset is equal to zero.
(2 of 6)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-40 October 2005
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
Table 2.12 Calculation of Calibration Factors for Endpoint and Rate
Chemistries, continued
Type Formula
2
Blanked Endpoint Chemistries (with Volume Correction)
FOR HIGH CALIBRATOR LEVEL:
Blank Correction Factor =
Volume of Reagent(s) (and Sample) at Blank Read
=
Volume of Total Reagent and Sample at Reaction Read
Blank Volume
Total Reaction Volume
E007123L.EPS
[Reaction ABS - (Blank ABS × Blank Correction Factor)] = Delta ABS
rep1
[Reaction ABS - (Blank ABS × Blank Correction Factor)] = Delta ABS
(Delta ABS
rep2
+ Delta ABS
rep1
) × 0.5 = Delta ABS
rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
[Reaction ABS - (Blank ABS × Blank Correction Factor)] = Delta ABS
rep1
[Reaction ABS - (Blank ABS × Blank Correction Factor)] = Delta ABS (Delta ABS
rep2
+ Delta ABS
rep1
) × 0.5 = Delta ABS
rep2
avg
(lo)
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × Delta ABS
Cal Set Point (hi) – Cal Set Point (lo)
Delta ABS
(hi) – Delta ABS
avg
avg
E014417L.EPS
(hi)]
avg
(lo)
Sample values are calculated by the following equation: [[[Reaction ABS - (Blank ABS × Blank Correction Factor)] * Cal Factor]] + offset = sample value
NOTICE For single point linear calibration, the low calibrator is a fixed zero point and the offset is equal to zero.
(3 of 6)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-41
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
Table 2.12 Calculation of Calibration Factors for Endpoint and Rate
Chemistries, continued
Type Formula
Nonblanked Rate Chemistries
FOR HIGH CALIBRATOR LEVEL:
Reaction Rate = Rate Reaction Rate = Rate (Rate
rep1
+ Rate
rep1 rep2
) × 0.5 = Rate
rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
Reaction Rate = Rate Reaction Rate = Rate (Rate
+ Rate
rep1
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × Rate
rep1
rep2
) × 0.5 = Rate
rep2
(lo)
avg
Cal Set Point (hi) – Cal Set Point (lo)
Rate
(hi) – Rate
avg
avg
avg
(hi)]
(lo)
Sample values are calculated by the following equation: (Reaction Rate × Cal Factor) + offset = sample value
E014418L.EPS
NOTICE For single point linear calibration, the low calibrator is a fixed zero point and the offset is equal to zero.
(4 of 6)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-42 October 2005
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
Table 2.12 Calculation of Calibration Factors for Endpoint and Rate
Chemistries, continued
Type Formula
2
Blanked Rate Chemistries
FOR HIGH CALIBRATOR LEVEL:
(Reaction Rate - Blank Rate) = Delta Rate (Reaction Rate - Blank Rate) = Delta Rate (Delta Rate
+ Delta Rate
rep1
) × 0.5 = Delta Rate avg(hi)
rep2
rep1 rep2
FOR LOW CALIBRATOR LEVEL:
(Reaction Rate - Blank Rate) = Delta Rate (Reaction Rate - Blank Rate) = Delta Rate (Delta Rate
Cal Factor (Slope) =
+ Delta Rate
rep1
) × 0.5 = Delta Rate
rep2
Cal Set Point (hi) – Cal Set Point (lo)
Delta Rate
Offset = Cal Set Point (hi) - [Cal Factor × Delta Rate
rep1 rep2
(hi) – Delta Rate
avg
avg
(lo)
avg
(lo)
avg
E014419L.EPS
(hi)]
Sample values are calculated by the following equation: [(Reaction Rate - Blank Rate) × Cal Factor] + offset = sample value
NOTICE For single point linear calibration, the low calibrator is a fixed zero point and the offset is equal to zero.
(5 of 6)
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-43
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
Table 2.12 Calculation of Calibration Factors for Endpoint and Rate
Chemistries, continued
Type Formula
Blanked Rate Chemistries (with Volume Correction)
FOR HIGH CALIBRATOR LEVEL:
Blank Correction Factor =
Volume of Reagent(s) (and Sample) at Blank Read
=
Volume of Total Reagent and Sample at Reaction Read
Blank Volume
Total Reaction Volume
E007123L.EPS
[Reaction Rate - (Blank Rate × Blank Correction Factor)] = Delta Rate
rep1
[Reaction Rate - (Blank Rate × Blank Correction Factor)] = Delta Rate
(Delta Rate
rep2
+ Delta Rate
rep1
) × 0.5 = Delta Rate
rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
[Reaction Rate - (Blank Rate × Blank Correction Factor)] = Delta Rate
rep1
[Reaction Rate - (Blank Rate × Blank Correction Factor)] = Delta Rate (Delta Rate
rep2
+ Delta Rate
rep1
) × 0.5 = Delta Rate
rep2
avg
(lo)
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × Delta Rate
Cal Set Point (hi) – Cal Set Point (lo)
Delta Rate
(hi) – Delta Rate
avg
avg
avg
E014419L.EPS
(hi)]
Sample values are calculated by the following equation: [[Reaction Rate - (Blank Rate × Blank Correction Factor)] × Cal Factor] + offset = sample value
NOTICE For single point linear calibration, the low calibrator is a fixed zero point and the offset is equal to zero.
(lo)
(6 of 6)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-44 October 2005

Non-Linear Chemistries

Non-linear chemistries include drugs and specific protein assays. Unlike the first-order rate and endpoint chemistries, which exhibit a linear response to increasing concentration, the calibration curves for non-linear chemistries exhibit logarithmic (S-shaped) or other nonlinear relationships. For this reason, curve fitting interpolation techniques are employed to construct the calibration curve.
For some non-linear calibrations, the curve parameters for a reagent lot are calculated during manufacturing. The curve parameters are encoded in bar code form, shipped on a card in the reagent box and loaded onto the system. A single point calibration is then run to adjust for instrument-to- instrument variation. If necessary, the calibration includes the sample diluent (DIL1), used as a blank, that is subtracted from all calibrator or sample responses.
• Multi-point chemistry calibration consists of five or six different levels of calibrators. These chemistries set calibration based on single replicates of each calibrator level.
Cartridge Chemistry: Calibration Theory
Non-Linear Chemistries
2
• Single-point chemistry calibration consists of one or two levels of calibrators. These chemistries set calibration based on two to four replicates of each calibrator level.

Non-Linear Calibration Formulas

The standard curve is determined by use of one of several nonlinear math models. The system uses an iterative technique to calculate the curve parameters. A modified Newton iteration is used to choose values. The best-fitting calibration curve is determined by minimizing the sum of the difference between the observed response and the calculated response of each standard.
The following symbols are used in the math models presented below:
R = sample response Conc = standard or sample concentration R
K
= calculated response for a zero sample
O
= scale parameter
C
a, b, c = parameters which define the nonlinear elements of the math model
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-45
Cartridge Chemistry: Calibration Theory
Non-Linear Calibration Formulas
Table 2.13 Math Models for Non-Linear Chemistries
Type Formula
Model #1 Math Model #1 is the four-parameter log-logit function most
commonly used with reagents that use antibodies.
R = R0 + K
c
1 + e
1
–a–b1n(conc)
E014420L.EPS
Sample values are determined using the calculated curve parameters and the math model. Values may be calculated directly as this model can be solved for concentration.
Model #2 Math Model #2 is a five-parameter logit function.
R = R0 + K
c
1 + e
1
–a–b1n(conc) – c(conc)
E014421L.EPS
This function cannot be solved directly for concentration. The instrument uses an iterative method to determine the sample value.
Model #3 Math Model #3 is a five-parameter exponential function.
R = R0 + K
a1n(conc) + b1n(conc)2 + c1n(conc)
e
c
3
E014422L.EPS
This function cannot be solved directly for concentration. The instrument uses an iterative method to determine the sample value.
Models #4
These Models are reserved for future development.
through #7 Model #8 Math Model #8 is an alternative to model #2, the five-parameter logit
function.
c
E014423L.EPS
R = R0 + K
c
1 + e
1
–a–b1n(conc)
This function cannot be solved directly for concentration. The instrument uses an iterative method to determine the sample value.
(1 of 2)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-46 October 2005
Cartridge Chemistry: Calibration Theory

Drugs of Abuse Testing (DAT) Chemistries

Table 2.13 Math Models for Non-Linear Chemistries, continued
Type Formula
Model #9 Math Model #9 is an extension to model #1, the four-parameter log-
logit function.
2
Polynomial Exponential (PXP)
Quadratic (POLY2)
Lorentz
Double Inflection Model DP4
R = R0 + K
c
1 + c e
1
–a–b1n(conc)
E014424L.EPS
The "c" is allowed to be either +1 or -1. If c = +1, then this is equivalent to model #1. If c = -1, an alternative function is being used.
This function cannot be solved directly for concentration. The instrument uses an iterative method to determine the sample value.
R = b0 + Pn(x)e Pn(x) = Polynomial of degree N
R = b0 + b1∗(conc) + b2∗(conc)
K
c
R = R0 + π {arctan(c∗conc + a) + 2 }
R =
1 +
-a(conc)
E014476L.EPS
E014477L.EPS
K
c1
a
1
Conc Conc
+
1 +
K
c2
a
2
A011537L.EPS
2
π
E014478L.EPS
(2 of 2)
Drugs of Abuse Testing (DAT) Chemistries
The Drugs of Abuse Testing (DAT) assays require three levels of calibrators. The calibration measures the separation between calibrators to measure reagent integrity. The calibration factor generated is non-functional for sample result calculation.
The cutoff value for each DAT chemistry represents the mean reaction rate of the low calibrator, reported in mA/min units on patient and control reports. The reaction rate of the samples is compared to the reaction rate of the low (cutoff) calibrator and reported out as POSITIVE or NEGATIVE. Cutoff values are stored in memory until the next successful calibration.
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-47
Cartridge Chemistry: Calibration Theory

Enzyme Verification

Enzyme Verification
Enzyme verification is a means of adjusting enzyme chemistry reporting units to IFCC methods. This feature is available for ALP, ALT-, AST-, CHE, CK-, GGT and LD. Verification also allows results to be adjusted for country specific correlation needs.
Verification is similar to calibration except that normalization factors are applied to the sample result in the form of a slope and offset adjustment, whereas calibration factors would be applied to the reaction response.
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-48 October 2005

Modular Chemistry: Calibration Theory

Calibration Theory

Modular chemistries are calibrated using two to three levels of calibrator (chemistry dependent). Four replicates per level are assayed. Data from two middle replicates of each level is used to set the system response. The highest and lowest replicates are discarded. Error checks are performed on the two middle replicates to verify successful calibration.

Calibration Error Detection

The analog signals generated by the calibrator measurements are converted to digital form. The resulting ADC values are compared to pre-programmed back-to-back, span and range limits to determine the calibration acceptability.

Modular Chemistries (MC)

The UniCel DxC contains seven chemistry modules (see chart below), each of which is used in the determination of eleven modular chemistries (MC), as follows.
Modular Chemistry: Calibration Theory
Calibration Theory
2
Table 2.14 Methodology and Modules Used with Modular Chemistries
Chemistry Methodology Module
Sodium Ion selective electrode (ISE) ISE Flow cell Potassium Ion selective electrode ISE Flow cell Chloride Ion selective electrode ISE Flow cell Carbon Dioxide pH electrode ISE Flow cell Calcium Ion selective electrode ISE Flow cell
Urea Nitrogen
Phosphorus
Creatinine
a
a
a
Conductivity electrode Urea Nitrogen
Colorimetric Phosphorus
Colorimetric Creatinine
Glucose Oxygen sensor Glucose
Total Protein
Albumin
a
DxC 800 only.
a
a
Colorimetric Total Protein
Colorimetric Albumin
UniCel DxC Systems Instructions For Use A13914 System Description October 2005 Page 2-49

Cartridge Chemistry: Principles of Measurement

Spectrophotometric Methods

Cartridge Chemistry: Principles of Measurement
Spectrophotometric Methods
Spectrophotometric methods rely on the principle that a sample, such as a patient sample, a control, or a calibrator, when mixed with one or more appropriate chemical reagents, produces a substance that has the ability to absorb light at specific wavelengths. This substance is referred to as a chromophore.

Beer's Law

According to Beer's Law, the amount of light absorbed by the chromophore is proportional to the concentration of the constituent being measured. The system can measure this as an endpoint or a rate of formation.
A = abc
where
A = absorbance of the chromophore a = absorptivity of the absorbing substance at the specific measuring
wavelength(s) b = cuvette light pathlength (cm) c = constituent concentration (M)
System Description UniCel DxC Systems Instructions For Use A13914 Page 2-50 October 2005

CHAPTER 3 Preparing Samples for Analysis

Routine Operation Overview

Daily Procedure

The following table shows an example of daily work flow using the DxC Synchron Clinical Chemistry System.
This procedure assumes that the initial system setup has been completed.
Step Action
1 If necessary, start the system. 2 Check reagent status. Load reagent as necessary.
NOTICE
Routine Operation Overview
Daily Procedure
3
3 Check calibration. Program or load calibrators as needed. 4 Program or load controls, if required. 5 If the system is not already running, press the RUN button on the system. 6 Check the control results to verify system operation. 7 Program or load patient samples. 8 If the system has gone into Standby, press the RUN button on the system. 9 Review the patient results.
10 Return to Step 7 if more samples need to be run. 11 The system will automatically return to Standby when all testing has been
completed.
UniCel DxC Systems Instructions For Use A13914 Preparing Samples for Analysis October 2005 Page 3-1

Sample Programming and Processing

Introduction

Sample Programming and Processing
Introduction
Sample programming provides the ability to identify samples, select tests to run, describe samples, and designate how to run samples. Samples are programmed through a host computer or Laboratory Information System (LIS or at the analyzer). The minimum information required to save a sample program includes:
• a sample ID or a rack and position assignment
• and one test selection
Test selections are assigned by the use of panels or by the selection of individual tests. Samples may be described and defined through sample type, sample comment, patient ID, and patient demographic functions. A sample can be programmed as a control sample or as a STAT for priority processing. Once programmed, samples are placed on the autoloader for processing.
If your LIS or normal workflow requires the reuse of sample IDs, the sample programming should be cleared from the DxC at a time interval that is less than the shortest time of sample ID reuse. If this warning is not followed, results from the new request will be merged with tests from incomplete samples that previously used that ID.

Program Samples and Controls

Refer to this chapter for Sample information and CHAPTER 6, Quality Control for Control programming information. If auto QC is enabled, reuse the same control ID bar code labels.
CAUTION
Preparing Samples for Analysis UniCel DxC Systems Instructions For Use A13914 Page 3-2 October 2005

Prepare Samples

Minimum Sample Volume
A minimum sample volume is required to run tests. To determine what volume of sample to use, refer to the Synchron LX/UniCel DxC Clinical Systems Sample Template.
Sample Racks
Sample racks accept the following tube and cup sizes:
Table 3.1 Sample Racks
13 × 75 12 × 75 mm tubes
Sample Programming and Processing
Rack Accepts these cups and tubes
13 × 75 mm tubes
0.5 mL cups
2.0 mL cups
Prepare Samples
3
16 × 75 16 × 75 mm tubes 13 × 100 13 × 100 mm tubes
0.5 mL cups
2.0 mL cups Capillary collection tubes (use with the capillary tube adapter)
16 × 100 16 × 100 mm tubes
16.5 × 92 mm tubes Beckman Coulter 0.5 mL Cup Insert (P/N 476399)
NOTICE Adapters are provided to adapt various sized sample tubes (secondary tubes) to the short racks. These adaptors must only be used in racks designated as reserved. The reserved rack feature is described in this chapter.
UniCel DxC Systems Instructions For Use A13914 Preparing Samples for Analysis October 2005 Page 3-3
Sample Programming and Processing

Sample Preparation by Container Type

Sample Preparation by Container Type
Table 3.2 shows how to prepare different sample containers.
Table 3.2 Preparation of Sample Containers
If running a sample from a... Then...
Primary Tube • Use the sample template in the Appendix to determine
adequate sample volume.
• Remove the stopper.
• For CTS systems, remove the stopper, if not a validated closed tube.
Secondary Tube • Remove the cap.
• Determine sufficient volume.
• Check for fibrin or other materials resulting from storage.
Beckman Coulter Synchron Microtube™
Autoanalyzer Cup
BD Microtainer
®
0.5 mL 2.0 mL
E014498L.EPS
• Pipette the sample into a Synchron Microtube™.
• Verify there are no bubbles at the bottom of the tube.
• Place into a 13 × 100 "reserved" rack.
• Place the cup into a rack.
OR
• Place in 15 × 85 tube (with sample bar code on tube).
• Verify there are no bubbles in sample.
• If cup is placed in a tube with a Bar Coded label a "reserved rack" must be used.
• Place Microtainer in adapter P/N 472987.
(refer to figure to the right)
• Verify there are no bubbles in sample.
• If cup is placed in a tube with a Bar Coded label a "reserved rack" must be used.
0.5 mL Cup Insert (P/N 476399) (reusable)
• Place the metal Cup Insert into a 16 × 100 mm rack (P/N 471921 – blue or 474463 – purple).
• Insert a 0.5 mL Autoanalyzer cup into the Cup Insert.
• Run in the "reserved rack" mode.
A011538L.EPS
Preparing Samples for Analysis UniCel DxC Systems Instructions For Use A13914 Page 3-4 October 2005

Sample Tubes Validated for CTS

Introduction
Rubber cap tubes can be run directly on the system without removing the cap. The CTS assembly cuts a small hole in the cap and the sample probe aspirates a sample directly from the tube. With this procedure, samples are processed faster and safer.
CTS Validated Sample Tubes
Only validated sample tubes can be run on UniCel DxC Systems with CTS. However, capped and uncapped tubes can be placed on the same reserved rack. Each tube is scanned to determine if it needs piercing. Refer to the table below for validated tubes.
Table 3.3 Tubes Validated for Closed Tube Sampling
Sample Programming and Processing
Sample Tubes Validated for CTS
3
CTS Configuration Tubes Validated
for Cap Piercing
1-Blade Thick CTS
Becton Dickinson VACUTAINERa
with HEMOGARD
Greiner VACUETTE
Sarstedt S-Monovette
a
All trademarks are property of their respective owners.
b
This tube requires a special 5.5 mL rack (P/N A18642).
CAUTION
Before placing validated closed tubes on the UniCel DxC Systems, check the top of the cap for any residual blood. Residual blood can contaminate the sample and affect results. If blood is present, remove it by using a cotton­tipped applicator stick moistened with DI water. When running in the CTS mode, if tubes off-loaded from the UniCel DxC Systems have water or droplets of water on the caps, disable the CTS and call Beckman Coulter Technical Support. Note: Oil on a cap is normal.
a
a
®ab
Size
13 × 75mm 13 × 100 mm 16 × 100 mm
13 × 75mm 13 × 100 mm
75 × 15 mm 92 × 15 mm

Reserved Racks for CTS Samples

The Reserved Racks/Obstruct Detect feature allows the user access to the following features:

Closed Tube Sampling (CTS)

This is an optional feature that allows the system to pierce primary sample tubes. If a CTS assembly is installed, the operator can enable/disable the CTS assembly. A CTS Tracking database monitors pierced tubes. If you run a tube again, it will not be pierced again. The sample ID stays in this database for seven days for a tube that has been pierced.
UniCel DxC Systems Instructions For Use A13914 Preparing Samples for Analysis October 2005 Page 3-5
Sample Programming and Processing

Reserved Racks

The 16 × 100 tubes must be run in a non-reserved 16 × 100 mm rack.
If a sample ID is manually cleared, the CTS tracking information is cleared on that instrument only. If a sample ID is reused and the tube requires piercing, the sample ID must be cleared on the instrument that is loaded first.
Do not clear the Sample ID manually. Before you run a sample whose cap has been pierced, be sure that its sample ID has not been cleared.
• If you clear a sample ID at the Host, the sample ID is NOT cleared in the CTS Tracking database.
• If you were to manually clear the sample ID for a tube, you would also clear the sample ID in the CTS Tracking database. The instrument would think it has a new tube and would pierce its cap again. Follow the steps below to verify sample ID clearing status.
NOTICE
Step Action
Reserved Racks
If a rack number is entered into this field, any sample containers in this rack will not be cap pierced even if the Cap Piercing feature is enabled. Level sensing on samples in the rack will be set to its most sensitive level to detect small sample volumes. This type of reserved rack may be used for uncapped, primary or secondary tubes that have a small volume of sample, for example: nested cups on primary tubes, Microtubes™ and pediatric capillary collection tubes.
1 Select the Samples icon from the menu bar. 2 Type in the sample ID. 3 Press [Enter]. 4 Does the sample have its initial program?
If… Then… The sample has its initial
program,
The sample does NOT have its initial program (the ID has been
• Do NOT remove its stopper.
• Run the sample.
• Remove its stopper.
• Run the sample.
cleared),
Preparing Samples for Analysis UniCel DxC Systems Instructions For Use A13914 Page 3-6 October 2005

Reserved Racks for HbA1c, or IBCT

There are two separate fields for specific tests. One for HbA1c and one for IBCT. If a rack number is entered into one of these fields, any samples run in that rack will only function for that specific test. For example, Rack 66 is entered into the Reserved Racks for HbA1c field. Any sample placed in that rack can only have an HbA1c run on it. The reason for this is that each test in this group requires some sort of sample preparation. If these tests are part of a group of other tests that do not require preparation, they cannot be run at the same time as the others. Using these features allows the prepared sample to be run in one of the reserved racks and allows the result of the prepared sample to be merged with the original sample report when all tests are complete.
Even if the Auto Serum Index feature is enabled, no serum indices will be run for the pre-treated HbA1c and pre-treated IBCT samples run in these reserved racks.
NOTICE
Sample Programming and Processing
Reserved Racks for HbA1c, or IBCT
3
The default for each of these features is:
• CTS = OFF,
• Reserved racks = none reserved,
• Obstruction detection = On,
• Reserved racks for HbA1c or IBCT = none reserved.
These features can only be modified when the system is in the Standby or Stopped state.
UniCel DxC Systems Instructions For Use A13914 Preparing Samples for Analysis October 2005 Page 3-7

How to Use Reserved Racks

When NOT to Use a Reserved Rack

How to Use Reserved Racks
When NOT to Use a Reserved Rack
When you do CTS (Closed Tube Sampling) do NOT run a closed tube in a reserved rack.

Recommended Use of Reserved Racks

Green Racks
Series 350–400 to run:
• nesting cups
• capillary collection containers
• Beckman Coulter Microtubes™
Gray Racks
Series 601 to run:
• "Barcoded" Calibrators and Controls
Purple Racks
• Series 401–420 to run HbA1c
• 0.5 mL Cup Insert (P/N 476399)
Brown Racks
• Series 451–499 to run IBCT
Blue Racks
0.5 mL Cup Insert (P/N 476399)
To identify a reserved rack, put a red ® on the top of the rack.
NOTICE
Preparing Samples for Analysis UniCel DxC Systems Instructions For Use A13914 Page 3-8 October 2005
Recommended Use of Reserved Racks
Assigning or Reassigning Reserved Racks
Step Action
1 Select the Setup icon from the menu bar. 2 Select the <Page Down> button on the right side of the screen. 3 Select <17> Reserved Racks/Obstruction Detection.
OR
Type 17 in the Option Number field and press [Enter].
The following screen appears.
How to Use Reserved Racks
3
E015928S.EPS
Figure 3.1
4 Type the rack numbers to assign as reserved racks in the HbA1c and IBCT
fields.
AND/OR
Type the rack numbers to assign as reserved racks.
5Select <OK> to assign the racks.
OR
Select <Cancel> to exit without assigning the racks.
UniCel DxC Systems Instructions For Use A13914 Preparing Samples for Analysis October 2005 Page 3-9
How to Use Reserved Racks

Bar Code Labeling

Bar Code Labeling
The use of bar code labels is a highly accurate and efficient method for identifying and processing laboratory samples. However, the system must be able to identify and read every bar code label to process each sample correctly. The following paragraphs provide some basic information pertaining to bar code labels. Additional bar code information can be found in the UniCel DxC Synchron Clinical Systems Reference Manual.
Bar Code Label Placement
Bar code labels must be applied to each sample tube in the correct location so that the bar code reader can read the bar code. The following diagram (Figure 3.2) describes how to place the label on a sample tube and how to place the tube into a sample rack.
CAUTION
A misread label can cause one sample ID to be read as another. The laboratory’s process for printing, placing, and meeting all bar code specifications is important to achieve highly accurate reading. Follow the bar code label specifications to keep the rate of misread labels to a minimum.
1
2
3
4
1. 14 mm (0.55 inch) Minimum
2. Label Placement Area
3. 20mm (0.78 inch) Minimum
4. 7.5 Degree Maximum
Figure 3.2 Bar Code Label Placement
4
A014600L.EPS
Preparing Samples for Analysis UniCel DxC Systems Instructions For Use A13914 Page 3-10 October 2005

Processing Bar Coded Samples

Generally, previous sample programming will not need to be cleared before placing a bar coded sample on the instrument. If the lab reuses Sample IDs, previous programming must be cleared.
Step Action
1 Load tubes in a rack with the bar code labels visible through the slot on the
same side of the rack as the rack bar code label.
2 For closed tube sampling systems, check the top of the validated closed
tube for visible blood. If blood is present, clean the rubber stopper with a cotton tipped applicator stick.
For non-CTS Systems or tubes not validated for closed tube sampling, remove the cap.
3 For a Routine Sample or if system is in Standby: Priority samples
should be placed on the autoloader first.
• Place the rack in the autoloader with the rack bar code label to the right and
How to Use Reserved Racks
Processing Bar Coded Samples
3
•Press RUN.
For a STAT Sample: If the system is running and there are other racks on
the autoloader,
•Press PRIORITY. The rack pusher moves back one space so the STAT rack can be placed in front of the other racks.
•Press RUN.
UniCel DxC Systems Instructions For Use A13914 Preparing Samples for Analysis October 2005 Page 3-11
How to Use Reserved Racks

Processing Unreadable Bar Coded Samples

Processing Unreadable Bar Coded Samples
If a sample bar code is unreadable and the LIS needs to be queried, use manual assignment. A manual assignment is rack and position and a Sample ID with no other information. The DxC will query the LIS for tests and demographics.
If... Then...
Individual Samples, • Select the Samples icon from the menu bar,
Several samples in the same rack, • Select the Samples icon from the menu bar,
• Type rack and position,
• Type Sample ID,
Save [F10].
• Select Rack [F8],
• Select the rack,
• Type Sample ID(s),
• Select <OK>.
Note: If previous programming appears, it may be necessary to clear it.

Processing Samples Manually

Manual programming is used when there is no LIS or the LIS is not available. Samples may or may not be bar coded. The sample is identified by a Sample ID and/ or rack and position. The user selects tests for programming.
The following items require manual programming:
• samples without bar codes
• samples with bar codes and no sample programming
Step Action
1 Select the Samples icon from the menu bar. 2 Before programming, check for previous programming.
•Select Racks [F8].
• Select the rack number and select <OK>.
• If the rack has programming that is no longer required, select Clear [F1].
• Type the Sample IDs into the Sample ID(s) field.
•Select <OK>. A second screen will open. Select <OK> again to clear the data.
3Select Program [F10].
(1 of 2)
Preparing Samples for Analysis UniCel DxC Systems Instructions For Use A13914 Page 3-12 October 2005
Step Action, continued
4 Identify samples using the table below:
If… Then…
How to Use Reserved Racks
Processing Samples Manually
3
Sample has a readable bar code,
• Type in the sample ID.
• If previous programming appears, it may be necessary to clear it.
• Go to Step 2, or Step 5 if no previous program appeared.
Sample has no bar code or bar code can not be read,
• Type in the sample ID.
• Select the Rack field.
• Type in a rack number that is not programmed.
• Go to Step 5.
5 If the sample is stat, select the STAT check box. 6Select Sample Type and Sample Comment if provided. 7 To enter Demographics, select Demog [F2]. Type in the information
provided and select Next [F10] to return to the Program Sample screen.
8 Select each chemistry and/or select a panel. To access additional
chemistries, use the up and down arrows to move to other pages.
9Select Next [F10].
10 To program additional samples, repeat Steps 2-8. 11 Get the rack(s) listed on the screen and place the samples into assigned
rack positions as shown on the load list.
12 For a Routine Sample or if system is in Standby: Priority samples
should be placed on the autoloader first.
• Place the rack in the autoloader with the rack bar code label to the right and,
•Press RUN.
For a STAT Sample: If the system is running and there are other racks on the autoloader,
•Press PRIORITY. The rack pusher moves back one space so the STAT rack can be placed in front of the other racks.
•Press RUN.
(2 of 2)
UniCel DxC Systems Instructions For Use A13914 Preparing Samples for Analysis October 2005 Page 3-13
How to Use Reserved Racks

Adding Tests or Rerunning a Sample

Adding Tests or Rerunning a Sample
Step Action
1 Select the Samples icon from the menu bar. 2Select Rerun [F6]. 3 Type the sample ID, rack, or rack and position for rerun. One or more
samples can be programmed to rerun.
4 Select one of the following buttons:
• To add or rerun specific tests press the <Select> button. All previous
programming will be highlighted. Deselect chemistries that are not to be run.
• To rerun all completed tests press the <All> button. Rerun all completed
tests for all requested samples. Go to Step 6.
5Select Next [F10]. 6 For a Routine Sample or if system is in Standby: Priority samples

Clearing Samples

Once the test program has been processed, if all positions were used, the rack is not available for programming again until it has been cleared. Only the programming information is deleted; the results are stored and are accessible through the recall function using the appropriate recall option.
Calibrator IDs or rack/position assignments are not cleared from within Sample Programming. Refer to CHAPTER 5, Reagent Load/Calibration, "Calibrator
Assignment", for information on calibrator identification.
should be placed on the autoloader first.
• Place the rack in the autoloader with the rack bar code label to the right and,
•Press RUN.
For a STAT Sample: If the system is running and there are other racks on the autoloader,
•Press PRIORITY. The rack pusher moves back one space so the STAT rack can be placed in front of the other racks.
•Press RUN.
NOTICE When clearing data, to verify exactly what will be cleared, first view the load list for the intended range of samples. The samples recalled on the load list are identical to the samples that will be cleared.
Preparing Samples for Analysis UniCel DxC Systems Instructions For Use A13914 Page 3-14 October 2005
How to Use Reserved Racks

Range of Sample IDs

When to Clear Samples
Samples should be cleared as follows:
If... Then...
Sample programming has been completed. Clear sample programming. A sample ID has to be reused. Clear that sample ID first.
3
A sample rack can not be programmed due to previous programming. (This occurs when a sample ID does not match, or when a rack number shows previous information.)
Range of Sample IDs
The system retrieves a range of sample IDs in either Alphanumeric or Numeric form. The system automatically selects one of the two, depending on the type of sample ID entered. These same methods of retrieval are used for Load List, Clear Samples and Recall.
Sample IDs containing only Numeric Data:
• Samples are listed in numeric order. For example 1, 2, 3, 4...
• Include leading zeros in the Range field. For example, range 1 thru 1000 would not
include sample 0099, but range 0001 thru 1000 would.
Sample IDs containing only Alpha or Alphanumeric Data:
• Samples are listed in dictionary order. For example A, AB, B...
• Numbers come before letters. For example 1, 900, 1A, A...
Clear that programming.
• Include all characters in the To field. For example, range A thru Z would not
include ZEBRA, but range A thru ZZZZZZZZZZZZZZ would.
NOTICE Do not mix a pure numeric entry with an alphanumeric entry. For example, 0 thorough 1A will delete all numeric.
UniCel DxC Systems Instructions For Use A13914 Preparing Samples for Analysis October 2005 Page 3-15
How to Use Reserved Racks

Procedure for Clearing Samples

Procedure for Clearing Samples
Step Action
1 Select the Samples icon from the menu bar. 2Select Clear [F7]. 3 Sample may be cleared by Sample ID(s), Rack/Position, or by Date/Time
created. Type in the specific criteria for deletion.
Clearing Rack/Position does not clear programming for sample(s) programmed in those positions with associated Sample IDs. Sample must still be cleared by Sample ID. However, clearing by Sample ID does clear rack/position in which the Sample ID was located.
NOTICE
Deletion Criteria Procedure for Sample Clearing
Sample ID • Type in the Sample ID(s) to be
cleared. Single and multiple IDs may be cleared. Use a comma (,) to separate a list of IDs.
Range of Sample IDs Refer to "Range of Sample
IDs" in this chapter for
additional information
• Type the first Sample ID in the
Range field.
• Type the last Sample ID in the
Thru field.
Time/Date Programs Created Type the time and date range of
samples to clear.
Rack/Position(s) • Type Rack/Position(s) to be
cleared. Single and multiple racks may be entered.
If only Rack is entered, programming in all positions in that rack will be cleared.
4Select <OK> to continue clearing. A confirmation screen will appear.
OR
Select <Cancel> to retain sample programs.
Preparing Samples for Analysis UniCel DxC Systems Instructions For Use A13914 Page 3-16 October 2005

Overview

Introduction

Overview
Introduction

CHAPTER 4 System Setup Options

This chapter summarizes the 29 System Setup options depicted on the Setup screens shown in Figure 4.1 and Figure 4.2 below:
4
Figure 4.1 Setup Screen (scrolled to the top)
E015930S.EPS
E015931S.EPS
Figure 4.2 Setup Screen (scrolled to the bottom)
UniCel DxC Systems Instructions For Use A13914 System Setup Options October 2005 Page 4-1
Overview
Introduction
For detailed step-by-step instructions on using the System Setup option, refer to the UniCel DxC Synchron Clinical Systems Reference Manual.
System Setup Options UniCel DxC Systems Instructions For Use A13914 Page 4-2 October 2005

Password Setup

Introduction

The Password Setup option allows the operator to:
• Assign up to 100 user names and their passwords
• Assign Administrator or Operator level privileges to each user name
• Define/edit or delete user name/password setup
• Enable/disable security for accessibility to certain system functions and setups
The following features may be password secured:
• Results, Edit
• Rgts/Cal, Modify Set Points
• Rgts/Cal, Slope/Offset Adjustment
Password Setup
Introduction
4
• Rgts/Cal, Within-Lot Calibration
• Rgts/Cal, Enzyme Validator
•QC
•Setup
• Utils, Clear Event Log

Defining/Editing Password Setup

To define, edit or delete user names, passwords, privilege levels and accessibility levels, follow the steps below.
Step Action
1 Select the Setup icon from the menu bar. 2Select <28> Password Setup from the Setup screen 3 Type an Administrator password in the Administrator Password dialog
box.
Select <OK>.
Select <Cancel> to return to the System Setup menu.
OR
Note: After a software install, the default password is beckman.
(1 of 2)
UniCel DxC Systems Instructions For Use A13914 System Setup Options October 2005 Page 4-3
Password Setup
Defining/Editing Password Setup
Step Action, continued
4 The Password Setup dialog box shows Page 1 (of 10) of the users with
5 In the Define/Edit Password dialog box, type the information requested.
password security. If necessary, select <Page Down> on the right side of the screen to show the user desired.
Select New [F1] to enter the password setup for a new user.
OR
To edit the password setup for a user, select the user’s password number.
OR
Type the user’s password number in the Numbers field and press [Enter].
Select Define [F2].
NOTICE
Do NOT type beckman in the Password field.
Select Administrator in the Title field to allow the user to modify password setup.
OR
Select Operator in the Title field to NOT allow the user to modify password setup.
Select <OK>.
OR
Select <Cancel> to return to the Password Setup screen.
6 The new or edited user is now in the Password Setup screen in alphabetical
order by last name.
(2 of 2)
Refer to the UniCel DxC Synchron Clinical Systems Reference Manual for additional information on deleting users from the Password List, and enabling and disabling password security for functions.
System Setup Options UniCel DxC Systems Instructions For Use A13914 Page 4-4 October 2005

Auto Serum Index/ORDAC

Auto ORDAC

Auto ORDAC permits the enabling or disabling of the automatic Overrange Detection and Correction (ORDAC) function for specified chemistries (refer to the Synchron Clinical Systems Chemistry Information Manual for a list of chemistries offering ORDAC). When a chemistry result exceeds the instrument printable range and Auto ORDAC is enabled, the sample will automatically rerun with either:
• a smaller sample size or
• an on-line sample dilution (chemistry dependent).
When Auto ORDAC for Ig-A and Haptoglobin is enabled, the URDAC feature is also enabled. Ig-A and Haptoglobin URDAC is used to analyze samples with concentrations below the analytical range. In this case, the system takes a larger sample volume.
Auto Serum Index/ORDAC
Auto ORDAC
4
(The manual ORDAC function in Sample Programming is used for samples which are known to exceed the usable range. Chemistries designated with manual ORDAC at the time of programming are run at the ORDAC sample volume.)
From the Setup Screen, select <1> Auto Serum Index/ORDAC to enable or disable ORDAC. The default for Auto ORDAC is OFF.
The Automatic ORDAC screen may be viewed at any time, however the system must be in Standby to modify the ORDAC selection.
NOTICE The analytical ranges for each analyte are system limits found in the respective CISs. These are the ranges that Beckman Coulter has verified can be achieved by the system. There is no flagging associated with values exceeding these limits.
The instrument printable ranges for each analyte are internal system limits. These ranges actually exceed the analytical ranges by a certain limit. This allows for precision variations and still permits a result to print even though it exceeds the analytical range slightly. Results outside this range will be suppressed. The suppressed results will be flagged OIR HI or OIR LO (Out of Instrument Range).
UniCel DxC Systems Instructions For Use A13914 System Setup Options October 2005 Page 4-5
Auto Serum Index/ORDAC

Auto Serum Index

Auto Serum Index
The Auto Serum Index function, when enabled, automatically analyzes every sample for the detection of hemolysis, icerus and lipemia and numeric values (index) for the relative concentrations (range) are included in the report.
The values are printed below the Special Calculations area of a patient report. The values are printed with the test results for a control sample.
Auto Serum Index is intended for sample integrity assessment only; not for patient diagnosis.
For additional information on Serum Indexing, refer to the Synchron Clinical Systems Chemistry Information Manual.
NOTICE When running the system with Automatic Serum Index enabled, remember to manually disable the serum index feature when choosing options to set the replicates. This prevents running indices needlessly.
From the Setup screen, select <1> Auto Serum Index/ORDAC to enable/disable Auto Serum Index.
To enable or disable Auto Serum Index, the system must be in Standby or Stopped.
The default for Auto Serum Index is OFF.
System Setup Options UniCel DxC Systems Instructions For Use A13914 Page 4-6 October 2005

Configuring the Chemistry Menu

Introduction

• The Chemistry Configuration function accommodates the installation of desired tests from a comprehensive list of available chemistries.
• Of the total tests available, 180 chemistries can be selected and positioned onto the menu at one time and customized to match the test order on the lab request forms.
• Modular chemistries (MC) are permanently configured on the system.
• These chemistries may be relocated on the configuration screen, but not removed.
• BUNm and UREAm both show (DxC 800 only) as available chemistries; only one may be configured at a time. To replace one with the other, overwrite the existing chemistry acronym with the replacement acronym.
• The Chemistry Configuration screen may be viewed at any time, but modifications can only be made when the system status is Standby or Stopped.
Configuring the Chemistry Menu
Introduction
4

Configuring a Beckman Coulter Chemistry

The chemistry menu, available in sample programming, quality control, panel definition and other screens, is defined by the user. To define Beckman Coulter chemistries, select <2> Chemistry Configuration/Sample Type from the Setup screen. Place the cursor in an open field in the Configuration Chemistry screen and press Chems [F1]. Select the DxC chemistry from the pop-up list.

Configuring a User Defined Chemistry (UDR)

The chemistry menu available in sample programming, quality control, panel definition and other screens is defined by the user. After a User Defined Chemistry (UDR) has been defined by selecting <15> User-Defined Chemistries from the Setup screen, you can Configure the UDR by selecting <2> Chemistry Configuration/ Sample Type from the Setup screen. After the Chemistry Configuration/Sample Type screen appears, selecting UDR [F2] presents a list of all User Defined Chemistries on the system. Refer to CHAPTER 9, User Defined Reagents in this manual for additional information on application of user-defined parameters.

Defining Chemistry Print Name

The Define Print Names function permits definition of the chemistry name that appears on the chartable report. Each chemistry may be given a name of a maximum of 15 characters. To define the chemistry print name, select <2> Chemistry Configuration/Sample Type from the Setup screen. Select Define [F3] from the Configured Chemistry screen to use the Define Print Names screen.
UniCel DxC Systems Instructions For Use A13914 System Setup Options October 2005 Page 4-7
Configuring the Chemistry Menu

Inserting a Chemistry

Inserting a Chemistry
The Insert function inserts a blank position for adding a chemistry to the menu. After a blank position is inserted, the chemistries following that position will be incremented by one position. If 180 is exceeded, the last one will be dropped. From the Setup screen, select <2> Chemistry Configuration/Sample Type. Refer to UniCel DxC Synchron Clinical Systems Reference Manual for additional information on this procedure.

Deleting a Chemistry

The Delete Chemistry function requires that you first clear the calibration, sample programming and control (QC) information from the instrument before deleting the chemistry from the system. Where it is used, Beckman Coulter recommends that you remove chemistries from the system following the sequence and procedures below to prevent problems with deleting chemistries from the QC definition:
Control samples in the "sample required," "incomplete" or "rerun" status may prevent the removal of the chemistry from the QC definition.
NOTICE
1. Remove the chemistry from Reagent/Calibration (Example: DIG)
Step Action
1 Select the Rgts/Cal icon from the menu bar. 2 Ensure the cal status of the chemistry DIG is not "Requested." To cancel a
calibration request:
• Select the position for the appropriate requested and assigned chemistry to be canceled.
•Select No Cal [F8] to cancel the calibration request. Chemistries currently selected may be canceled only if the calibration is not yet in progress.
3 Remove calibrator assignments. To clear calibrator bar code ID and/or rack
and position:
•Select Assign [F7].
•Select <Next> or <Prev> to locate the calibrator of interest.
OR
• Select the Calibrator Name options button <> at the top of the Assign Barcode/Rack dialog box to view the List of Calibrators.
• Clear the fields by selecting the Calibrator ID, Rack and Position fields for each level of calibrator and pressing the [Delete] key on the keyboard.
•Select <Close> to exit and save information.
(1 of 2)
System Setup Options UniCel DxC Systems Instructions For Use A13914 Page 4-8 October 2005
Loading...