Common Error Messages and Corrective Actions .................................... 13-29
UniCel DxC Systems Instructions For Use A13914Table of Contents
October 2005Page 5 of 6
Table of ContentsUniCel DxC Systems Instructions For Use A13914
Page 6 of 6October 2005
General Information
Intended Use
The UniCel® DxC Synchron® Clinical Systems are fully automated, computercontrolled clinical chemistry analyzers designed for the in vitro determination of a
variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can
be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood
(sample type is chemistry dependent).
Scope of This Manual
This manual covers basic operating instructions and maintenance guidelines for
UniCel DxC 600/800 Systems. Detailed operation, maintenance, and troubleshooting
instructions are not included in this manual. In addition, medical and diagnostic
interpretation, or the clinical significance of chemistries or assays are not discussed.
Refer to the reference materials below for detailed information.
CHAPTER 1 General Information
General Information
Intended Use
1
Reference Manuals
Other manuals that accompany the system include:
• the UniCel DxC Synchron Clinical Systems Reference Manual which contains
detailed system description, comprehensive operating instructions, theory of
operation, system calibration, and programming procedures, and quality control
information for the UniCel DxC Systems.
• the Synchron Clinical Systems Chemistry Information Manual which contains
specific chemistry information for the full range of analytes available on the UniCel
DxC and Synchron LX Systems.
• the UniCel DxC Synchron Clinical Systems Host Interface Specifications.
• the Synchron Clinical Systems Performance Verification Manual which will help
you integrate the new UniCel DxC System into your daily laboratory routine.
• the UniCel DxC 600/800 Synchron Clinical System Operator Tips which provide
additional information about your system, including screen names for general
chemistries and Immunoassays.
• the Synchron LX®/UniCel DxC Clinical Systems Sample Template.
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-1
Manual Conventions
Manual Conventions
Manual Conventions
Manual Conventions
This Manual uses the following printed and visual cues to guide the user in responding
to printed directions.
Table 1.1Conventions Used in this Manual
ConventionDescription
Icon buttons and
Icon buttons and dialog box buttons are written in bold.
Dialog box buttons
Example:
Select the Samples icon.
Function buttonsFunction buttons are written in bold with the corresponding
function key in square brackets ([ ]).
Example:
Select Print [F10].
Command buttons
(buttons with names
Buttons with names are bold and enclosed in angle brackets
< >.
on a screen)
or <↑> <↓>
Example:
Select <Cancel>.
Instrument buttonsButtons on an instrument are in bold. May be all upper case or
initial caps.
Example:
Press the STOP button.
Keyboard keysKeyboard keys are written in bold and enclosed in square
brackets [ ].
Examples:
[X], [→], or [Tab] [Enter]
Options buttonAn Options button is a triangle enclosed in angle brackets < >.
Example:
Select the Options button <▼>.
Text fieldNames in the text fields are bold followed by the word "field".
Example:
Type in the Patient ID in the Patient ID field.
[Alt + X]Combination keys are written in bold and enclosed by square
brackets ([ ]) with a plus (+) sign between each key.
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-2October 2005
How to Use this Manual
Manual Format
Information in this manual is presented in modular units. Each unit of information is
described by a brief title in the left margin.
Many units consist of a table which presents a procedure, process, or description.
Procedure Tables
Procedure tables are the most common type of table in this manual. They list each step
of a procedure by number with the corresponding action that is to be performed.
Occasionally, a decision must be made at a step within a procedure. A smaller
decision table is then presented which describes the variable conditions in the left
column and the appropriate action for each condition in the right column.
How to Use this Manual
Manual Format
1
Example of Procedure Table
The following table is an example of a procedure table that contains a decision table.
StepAction
1Select Rerun [F6].
2
To enter...type...
individual Sample IDs,the Sample IDs for rerun in the Sample
a range of Sample IDs,the Sample ID at the beginning of the
3Select a button from the bottom of the dialog box.
Read the decision table as a complete sentence, using the first heading to introduce the
condition and the second heading to introduce the action. Step 2 of the table is read:
To enter individual Sample IDs, type the Sample IDs for rerun in the Sample IDs
field.
IDs field.
range in the Range field.
To enter a range of Sample IDs, type the Sample ID at the beginning of the range in the
Range field.
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-3
Summary of Hazards
Introduction
Summary of Hazards
Introduction
This section summarizes the hazards associated with the DxC System. Individual
hazards associated with a specific procedure in this manual are included in Warning
or Caution boxes within the procedures for that task. Please read this section and the
following Summary of Precautions before operating the system.
Bar Code Reader Hazards
Do not tamper with or remove the housing of any bar code reader because of the laserbased nature of the readers and the potential hazard of looking directly at laser light.
When the instrument is running, homing, or in diagnostics, the laser may be ON. At
all other times the laser is OFF.
Biohazardous Materials Hazards
Observe all laboratory policies or procedures which pertain to handling of infectious
and pathogenic materials.
Closed Tube Sampling (CTS) Cap Piercer Assembly Hazards/Biohazards
The Cap Piercer contains a razor sharp blade assembly that has been exposed to
potentially biohazardous fluids. The points of the Blade are very sharp and extend
below the Wash Tower. Stay away from the bottom of the Wash Tower. To prevent
injury or exposure, do not touch the Points of the Blade and always wear gloves. Do
not select <OK> when the cover is removed. This will move the Cap Piercer Carriage
up and cause a pinching hazard.
CTS Auto-Gloss Handling Hazards
Be careful when handling the bottle of CTS Auto-Gloss. Prevent spills. This lubricant
is extremely slippery and difficult to clean from a floor.
Electric Shock Hazards
Replacement or servicing of any components where contact with bare, live hazardous
parts could occur, possibly resulting in electric shock, should only be performed by
qualified service personnel.
Electrical Ground Hazards
Do not under any circumstances operate the system until an electrical ground is
provided and the power cord is properly connected to the ground.
Flammable Materials Hazards
Do not use this system in the presence of flammable materials.
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-4October 2005
Hazardous/Biohazardous Substances Hazards
When handling a spill of blood or other potentially hazardous substances, clean up the
spill by using a 10% bleach solution, or use your laboratory decontamination solution.
Then follow your laboratory procedure for disposal of hazardous materials. If the
UniCel DxC system needs to be decontaminated, call your Beckman Coulter Service
Representative for assistance.
ISE Module Hazards
When lowering the ISE module, keep fingers clear to avoid a pinching hazard.
Moving Parts Hazards
Do not place hands near any moving parts while the system is in operation.
No Foam Reagent Pressurization Hazard
The No Foam container is pressurized during system operation and must be properly
depressurized prior to servicing to avoid sudden depressurization and potential
exposure of the skin or eyes to the No Foam solution. To release the air pressure,
disconnect the white quick connector located at the supply side of the container. If
inhaled, move exposed individual to fresh air. If skin or eye contact with the solution
occurs, flush the affected area thoroughly with water for at least 15 minutes. In both
cases seek medical attention. Refer to the Antifoam Material Safety Data Sheets
(MSDS) for additional information.
Hazardous/Biohazardous Substances Hazards
Summary of Hazards
1
Power Cord Hazards
Only use a three-pronged power cord to connect the instrument to a matching threewire grounded outlet. Do not use an adapter to connect the power plug to a twopronged outlet.
Service Procedures Hazards
Disconnect the power cord when performing service procedures such as replacing
electronic or mechanical components.
The reagents and other chemical preparations used with the system will not normally
cause adverse reactions; however, those persons with sensitive skin should wear
protective rubber gloves before attempting to work with those substances.
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-5
Summary of Hazards
Sodium Azide Preservative Hazards
Sodium Azide Preservative Hazards
Reagents, calibrators and controls used with the system may contain small quantities
(< 0.1%) of sodium azide preservative. Sodium azide preservative may form explosive
compounds in metal drain lines. Refer to National Institute for Occupational Safety
and Health Bulletin: Explosive Azide Hazards (8/18/76). Avoid skin contact with
reagent. Use water to wash reagent from skin. Refer to the related Material Safety
Data Sheets (MSDS) for additional information.
System Operations and Specifications Hazards
System operation should be consistent with the power requirements as stated in the
Summary of Hazards section of this chapter, and should always conform to the
procedures and safety warnings throughout this manual.
If the equipment is used in a manner not specified by Beckman Coulter, Inc., the
protection provided by the equipment may be impaired.
Waste B Disposal Hazards/Biohazards
The Waste B collection bottle contents are considered biohazardous and should be
handled appropriately.
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-6October 2005
Summary of Precautions
Introduction
This section summarizes the precautions that should be taken when operating the DxC
System. Individual precautions associated with a specific procedure in this manual are
included in Caution boxes within the procedures for that task. Please read this section
and the preceding Summary of Hazards before operating the system.
AccuSense™ Glucose Sensor Precautions
Whenever the sensor is replaced, reaction cup and stir bar cleaning is recommended.
Remove the sensor prior to removing stir bar. To prevent damage to the glucose
oxygen sensor membrane tip, do not insert stir bar removal tool, applicator stick, or
any other object into the glucose reaction cup unless the sensor has been removed. Do
not touch membrane tip of the AccuSense glucose oxygen sensor.
Air Filter Cleaning Precautions
Summary of Precautions
Introduction
1
Do not place a damp filter back on the system. Residual moisture may damage the
system.
Alkaline Buffer Stability Precautions
The alkaline buffer reagent is stable for one month on the system. However, if a color
change from pink to a lighter shade of pink should occur, replace the alkaline buffer
with a fresh bottle of reagent.
Aqueous Calibrator Precautions
Repetitive refrigeration of Synchron aqueous calibrators may facilitate crystal
formation. Once removed from refrigerated storage, these calibrators should remain at
room temperature. The calibrator once opened is stable for the period claimed in the
accompanying package insert.
Bar Code Label Precautions
A misread label can cause one sample ID to be read as another. The laboratory’s
process for printing, placing, and meeting all bar code specifications is important to
achieve highly accurate readings. Always follow the bar code label specifications to
avoid misread labels.
Beckman Coulter Microtube™ Precautions
• Beckman Coulter Microtubes™ are designed for use on specific Synchron systems.
Using the appropriate Microtube is essential for proper system operation.
• The sample height in the Microtube is critical for correct sample aspiration on all
Synchron systems.
• The use of Array® Microtubes™ (P/N 448163 or P/N 448162) on Synchron
Systems or the use of Synchron Microtubes (P/N 756776) on Array systems may
result in short sampling, incorrect results, and/or sample probe damage.
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-7
Summary of Precautions
Biohazard Precautions
• The use of non-Beckman Coulter, third party Microtubes, which have not been
designed and tested on Synchron Systems may result in system damage and/or short
sampling.
Biohazard Precautions
All biohazard precautions should be observed when doing maintenance, service, or
troubleshooting on the system. This includes but may not be limited to wearing gloves
and eye shields, and washing hands after working on contaminated portions of the
system.
Blood Barrier Collection Tube Precautions
When blood collection tubes that contain physical barriers are used, extra care should
be exercised to ensure that the barrier is tightly packed. Loose particles from the
barrier could coat or plug the sample probe, flow cell, chemistry modules, electrolyte
injection cup (EIC), or cuvette wash station.
BUNm/UREAm Electrode Precautions
Loosen the BUNm/UREAm electrode retainer nut, approximately one-half turn, and
withdraw electrode slightly before removing the stir bar and make sure the BUNm/
UREAm electrode is slightly withdrawn before reinstalling the stir bar. This prevents
possible damage to the delicate electrode tip. After the stir bar is reinstalled, tighten
the retainer nut to prevent leakage.
CC Subsystems Priming Precautions
If any two of the CC Subsystem items are selected, all three will prime automatically.
Check to make certain that all three areas are ready to be primed if any two are
selected.
CO2 Membrane Replacement Precautions
Do not touch the membrane surface when installing the quad-ring.
Covers, Doors and Shields Precautions
To ensure optimum performance of the system, operate the system with reagent doors
and all shields and covers in place. To prevent possible motion errors, verify the proper
positioning of any removed and reinstalled cover or shield.
CTS (Closed Tube Sampling) Cap Piercer Precautions
Do not use the Cap Piercer with foil-capped tubes. This will give level sense errors.
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-8October 2005
CTS Tracking Loss Precautions
For systems with 1-Blade Thick CTS, if there is an unusual loss of network
communication, then a message tells you that if you move a tube to a different
instrument, remove its cap.
• A normal shutdown or reboot does NOT give this CTS message.
• When there is no CTS Tracking, if the cap were kept on a previously-pierced tube:
- it would be pierced again and
- pieces of rubber could fall into the sample.
• You will be notified when full CTS Tracking is restored.
Diethylamine HCL and the Calcium ISE Precautions
Do not use controls containing diethylamine HCL. This adversely affects the calcium
Ion-Selective Electrode (ISE).
Electrostatic Discharge (ESD) Hazards
To prevent damage due to electrical static discharge (ESD), always wear the wrist
ground strap when cleaning the probes, Flow Cell, ISE Drain or CC sample, Glass
Cuvettes and reagent mixers; removing, installing or flushing the Flow Cell
Electrodes; cleaning the EIC Ports; bleaching CO2 Alkaline Buffer Reagent Straw and
Line; removing the CO2 Measuring Electrode Membrane; replacing EIC Quad-Ring,
Ratio Pump Quad-Rings, Alkaline Buffer Peri-Pump Tubing or Sample and Reagent
Syringes.
Summary of Precautions
CTS Tracking Loss Precautions
1
Environmental Conditions Precautions
Changes in ambient temperatures and environmental conditions may result in a
"reference drift" message. In this case, the electrolyte chemistries must be
recalibrated.
Fibrin Clots Precautions
Samples should be free of all visible fibrin. Clots could coat or plug the sample
probes, flow cell, chemistry modules, electrolyte injection cup (EIC), or cuvette wash
station leading to instrument malfunction and/or short sampling.
ISE Reagent Precautions
Failure to operate the system with sufficient ISE reagent will result in erroneous
chemistry results. In some cases, results will be obtained without reagents. Therefore,
before starting any run, verify that sufficient reagent is available to complete the run.
MC Reagent Stir Bar Precautions
When cleaning the MC Reagent Lines, Cups, and Stir Bars, The stir bar may rise in the
cup due to air accumulation in the lines. Verify that the stir bar is positioned down into
the bottom of the cup.
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-9
Summary of Precautions
Narrow Margin Bar Code Precautions
Narrow Margin Bar Code Precautions
The sample bar code reader on the DxC can read narrow-margin bar codes. Because
of the sensitivity needed to read narrow-margin bar codes, the labels must be high
quality. They must be free from smudges, spots or other imperfections. An
imperfection could be read as part of the Sample ID. This could cause an inaccurate
read of the bar code.
Racks Replacement Precautions
Racks should be replaced every five years. Damaged racks should not be used on the
system or in the SPINCHRON™ Centrifuge.
Ratio Pump Quad-Ring Replacement Precautions
Be careful not to twist quad-rings or O-rings during installation, as this could result in
reagent leakage or even in ratio pump failure.
Reagent Volume Precautions
Check reagent volumes before starting a run. Failure to operate with sufficient reagent
will result in erroneous chemistry results. In some cases, results will be obtained
without proper amounts of reagents in the modules. Therefore, before starting any run,
verify that sufficient reagent is available to complete the run.
Modular chemistry reagent containers should not be handled while the system is
performing modular chemistry measurements.
The use of expired reagents may cause erroneous results.
Residual Blood Contamination Precautions
Before you place your validated closed tubes on the DxC, check the top of the cap for
any residual blood. Residual blood contamination into the sample could affect results.
If blood is present, remove it by using a cotton-tipped applicator stick moistened with
DI water. When running in the CTS mode, if tubes off-loaded from the UniCel DxC
Systems have water or droplets of water on the caps, disable the CTS and call
Beckman Coulter Technical Support.
Note: Oil on a cap is normal.
Reuse of Sample ID Precautions
If your LIS or normal workflow requires the reuse of sample IDs, the sample
programming should be cleared from the DxC at a time interval that is less than the
shortest time of sample ID reuse. Failure to heed this warning will result in new
requests being merged with tests from incomplete samples that previously used that
ID.
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-10October 2005
Sample Bar Code Reader Precautions
Do not tamper with or remove the housing of the Sample Bar Code Reader.
Sample Syringe Replacement Precautions
When installing the syringes, do not mix the two different syringe sizes. The MC and
CC sample syringes (100 µL) are located on the back wall of the instrument. The CC
reagent syringe (500 µL) is located on the right side of the instrument.
Samples NOT received by Host Precautions
Do not clear samples until results are received at the host and printed. Clearing
samples manually or through host programming, may cause results to be received at
the host and printed at the DxC without the sample ID. Depending on the host
implementation, this can cause lost sample results or sample results which merge with
other sample results producing duplicate tests or added tests.
Samples Use Precautions
Do not use the same sample run on an DxC system for analysis of analytes for which a
small quantity of carryover could greatly increase the results (for example, TBhCG).
Sample Bar Code Reader Precautions
Summary of Precautions
1
If your system has a Closed Tube Sampling (CTS) option, Do not run samples in cups
on top of a sample tube. This would give motion errors. As an alternative, put the
sample in a nesting cup on a tube in a reserved rack.
System Backup Precautions
After inserting a diskette into the floppy disk drive, selecting <OK> when prompted
will prepare the floppy disk by ERASING it before copying data. Be sure the diskette
does not contain critical data that is not available from another source.
System Configuration Change Precautions
Changes to the System Configuration Data should only be done at the request or at the
direction of Beckman Coulter, Inc. Entry of incorrect information will lead to system
errors.
System Restore Precautions
System Parameter and Alignment data can be restored from the backup floppy disks
onto the system, however performing the Restore function deletes some or all files
(depending on the areas restored) from the hard drive.
Urine Sample Precautions
After analysis of ten consecutive urine electrolytes, run one replicate of electrolytes on
Synchron Calibrator Level 2 in the serum mode. This will minimize the potential for
chloride drift due to matrix effects of urine samples.
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-11
Symbols and Labels
Introduction
Symbols and Labels
Introduction
The following is a list of symbols and labels used on the DxC Systems. You will find
them affixed to the appropriate components of the system, as described briefly below.
Instrument Power Switch, ON
This symbol located on the main power switch indicates that the analyzer power is ON
when this portion of the switch is in the down position.
Instrument Power Switch, OFF
This symbol, also located on the main power switch, indicates that the analyzer power
is OFF when this side of the switch is in the down position.
Instrument or Printer Power Switch, ON
This symbol is located on the analyzer and printer power switch. When the portion of
the switch with this symbol on it is in the down position, power to the monitor or
printer is ON.
Monitor Switch, ON/OFF
This symbol is located on the monitor power switch. A green light to the left of this
symbol indicates the power is ON.
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-12October 2005
CPU Power OFF Switch
This symbol is located on the face of the Computer (CPU) unit and indicates the OFF
state when pressed.
Primary Electrical Ground
This symbol is used to indicate an electrical ground.
Keyboard Connection
Symbols and Labels
CPU Power OFF Switch
1
This symbol is found above the connection between the computer and the keyboard.
Can Hold This Object Here
This black symbol, located on the bottom of each sample and reagent probe assembly,
indicates that this area may be handled to rotate the probe.
Do Not Hold This Object Here
This red symbol, located on the top of each sample and reagent probe assembly,
indicates that this area may not be handled.
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-13
Symbols and Labels
Mouse Port Connection
Mouse Port Connection
This symbol is found next to the connection between the computer and the mouse port.
High Voltage Electric Shock Risk
This symbol indicates high voltage is present and /or there is a risk of electric shock
when working in this area.
CAUTION
This symbol indicates a caution message and is followed by an explanation or other
symbols that define the caution (see examples below).
CAUTION Operate with All Covers in Place
This symbol is located on top of the work surface cover and the cover of an optional
Cap Piercer. It indicates a caution to operate only with all covers in place to reduce
risk of personal injury or biohazard.
456161-B
General Biohazard
These caution symbols indicate biohazardous risk from possible patient specimen
contamination.
CAUTION
TO REDUCE RISK OF PERSONAL INJURY,
OPERATE ONLY WITH ALL COVERS IN PLACE.
A011459L.EPS
A011460L.EPS
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-14October 2005
Laser Bar Code Caution
A label reading, "CAUTION. LASER LIGHT ACCESSIBLE. WHEN COVER IS
OPEN OR REMOVED, DO NOT STARE INTO BEAM." is placed on the cover of
any laser-based code reader. Do not stare into laser light beam when cover is open or
removed.
Class II Laser Caution Warning
A label reading, "CAUTION. LASER LIGHT - DO NOT STARE INTO BEAM.
670nm - 1mW CLASS II LASER PRODUCT." is placed near any opening through
which a bar code reading beam will emit. Do not stare into laser light beam.
Symbols and Labels
Laser Bar Code Caution
1
Laser
A label reading, "AVOID EXPOSURE. LASER LIGHT IS EMITTED FROM THIS
APERTURE." is placed near any opening through which a bar code reading beam
emits. Avoid exposure to laser light emitted.
Sharp Objects Caution
A label reading, "CAUTION. SHARP OBJECTS - REFER SERVICING AND
MAINTENANCE TO QUALIFIED SERVICE PERSONNEL." is found on top of the
optional cap piercing hardware which is located underneath the removable cover of the
cap piercing tower.
CAUTION
SHARP OBJECTS - REFER
SERVICING AND MAINTENANCE TO
QUALIFIED SERVICE PERSONNEL.
A012936L.EPS
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-15
Symbols and Labels
ISE Cover Caution
ISE Cover Caution
A label reading, "THE ISE COVER SHOULD REMAIN IN PLACE DURING
SYSTEM OPERATION." is placed on top of the ISE module frame under the ISE
cover to indicate that the ISE cover should remain in place during system operation.
THE ISE COVER SHOULD REMAIN IN PLACE DURING SYSTEM OPERATION.
471830-AA
Moving Parts Caution
A label reading, "CAUTION PARTS MOVE AUTOMATICALLY" is placed inside
the offload track, on the left side of the back wall.
A012937L.EPS
CAUTION
PARTS MOVE
AUTOMATICALLY
MC Door Caution
A label reading, "THIS DOOR SHOULD REMAIN CLOSED DURING SYSTEM
OPERATION." is found on the top edge of the left hand (MC reagent) door and
indicates that the door should remain closed during system operation.
THIS DOOR SHOULD REMAIN CLOSED DURING SYSTEM OPERATION.
471831-AA
Read Manual Caution
A label reading, "CAUTION - READ MANUAL BEFORE OPERATING." is found
on the hydropneumatics behind the center door and recommends the operator read the
manuals before operating the system.
A015047L.EPS
CAUTION - READ MANUAL
BEFORE OPERATING
ATTENTION - CONSULTER LA NOTICE
AVANT DE FAIRE FONCTIONNER.
270-455774-A
A012939L.EPS
A012938L.EPS
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-16October 2005
Rack Loading Label
This label is found to the right of the Run and Priority Load buttons and indicates the
correct position of a rack for placement into the autoloader.
Laser Certification Label
This label is found on the back, bottom edge of the system. It provides information
about the laser.
Symbols and Labels
Rack Loading Label
1
Ethernet/Serial Port Label
This label is found on the right side of the system and identifies connections for the
Ethernet and serial ports.
PRODUCT COMPLIES WITH
21 CFR CHAPTER I, SUBCHAPTER J
MANUFACTURED DECEMBER 2004
LABEL P/N 448229 AB
BECKMAN COULTER, INC
SERIAL PORT
MADE IN U.S.A. MARCA REG
ETHERNET
S
A011540L.EPS
A012942L.EPS
UniCel DxC Systems Instructions For Use A13914General Information
October 2005Page 1-17
Symbols and Labels
Fluid Interface Label
Fluid Interface Label
This label is found on the center, back, bottom edge of the system.
It identifies inlet and outlet ports on the system.
136
Recycling Label
This symbol is required in accordance with the Waste Electrical and Electronic
Equipment (WEEE) Directive of the European Union. The presence of this marking
on the product indicates:
1. the device was put on the European Market after August 13, 2005 and
2. the device is not be disposed via the municipal waste collection system of any
WASTE B
SENSOR
WASTE B
OUTLET
D. I. WATER INLET
MAX. PRESS. 100 PSI (689 kPa)
WASTE OUTLET
member state of the European Union.
137128
D.I. FIBER
A012943L.EPS
A016608L.EPS
It is very important that customers understand and follow all laws regarding the proper
decontamination and safe disposal of electrical equipment. For Beckman Coulter
products bearing this label please contact your dealer or local Beckman Coulter office
for details on the take back program that will facilitate the proper collection, treatment,
recovery, recycling and safe disposal of device.
General InformationUniCel DxC Systems Instructions For Use A13914
Page 1-18October 2005
System Description
Introduction
The following chapter briefly describes the system components, operational theories,
principles of measurement, programming structure, and operator controls. Detailed
information is located in the UniCel DxC Synchron Clinical Systems Reference Manual.
CHAPTER 2 System Description
System Description
Introduction
2
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-1
Operational Conditions
Shipping Damage
Operational Conditions
Shipping Damage
Each DxC System is carefully examined and checked by Beckman Coulter, Inc. before
it is shipped. When you receive your new DxC System, visually inspect the shipping
container for damage. If there is damage, notify the Beckman Coulter Service
Representative before he or she arrives at your facility to install your system.
Installation
Refer to the following table.
Table 2.1Installation Requirements
ItemRequirement
Installed byThe DxC system is installed completely by Beckman
Coulter
Installation CategoryII
Clearances
The system requires the following clearances.
Table 2.2System Clearances
Area AffectedClearance Needed
Left SideMinimum of 6 inches (15.2 cm) clearance or 12 inches
Right SideMinimum of 18 inches (45.7 cm) clearance when monitor
BackZero inches. The venting design of the system allows for
TopMinimum of 22 inches (55.9 cm) from highest point of
FrontMinimum of 25 inches (63.5 cm) to open doors.
Sunlight and Drafts
(30.5 cm) to access smart modules.
on swing arm is in use.
"0 inch" clearance at the back.
system.
Do not place the system in direct sunlight or in drafts. Both of these conditions may
affect the temperature control of the system.
Drain
The system should be located near a sink or floor drain to accommodate the waste
effluent at a minimum rate of 6 liters/hour (16 liters/hour continuous flow).
The drain must not be placed any higher than 36 inches (91.4 cm) above the floor.
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-2October 2005
The system can operate from any standard 3-wire electrical outlet and is wired as
shipped from the factory to operate on 220V AC, 50/60 Hz.
Line Voltage from the electrical outlet should be free of spikes, fluctuations,
and dropouts for protection of the electronic circuitry.
Only operate the system from a 3-wire power source. DO NOT use a 2-prong
adapter or a 2-wire AC power source.
Environmental Conditions
Table 2.4Temperature, Humidity and Elevation
ItemSpecification
EnvironmentIndoor use only
Ambient temperature+18°C to +32°C
NOTICE
CAUTION
Warm-up time30 minutes (time to reach operating temperature)
Relative humidity20–85% relative, non-condensing
ElevationUp to 1,280 m (4,200 ft.)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-3
Operational Conditions
Water Requirements
Water Requirements
Table 2.5Water Requirements
ItemSpecification
Flow Rate0.6 L/min peak flow rate
16 L/hr, minimum continuous flow rate
Temperature+15°C to +25°C
Water qualityNCCLS Type II deionized water, except for total
bacteria count < 10 cfu/mL
Water pressureDeionized water entering the system must be 30–90
psig.
Other System Specifications for IEC-1010 Compliance
Table 2.6IEC-1010 Specifications
ItemSpecification
Pollution Degree2
EN55011Meets Class A
Maximum Sound Pressure≤ 65 dBA average over 8 hours with covers down at 1
meter away from the instrument at 25°C
Maximum Leakage CurrentDxC 600: 222 µA at 240V, 50Hz
DxC 800: 240 µA at 240V, 50Hz
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-4October 2005
System Components
DxC Systems
A UniCel DxC System can be divided into the following components:
• Sample Handling Components
• Modular Chemistry System
• Cartridge Chemistry Reagent Handling System
• Hydropneumatic System
• Operation and Control Components
A Closed Tube Sampling System has optional components listed below:
System Components
DxC Systems
2
• Large Particle Immuno Assay Module (LPIA)
• Closed Tube Sampling (CTS)
1
3
1. Modular Chemistry (MC) Section
2. Cartridge Chemistry (CC) Portion
3. Autoloader
2
54
4. Reaction Carousel
5. Dual Reagent Carousel
6. Operator Console
*
6
A015900P.EPS
Figure 2.1UniCel DxC 600 Analyzer
*
Equivalent to Near Infrared Particle Immuno Assay (NIPIA).
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-5
System Components
DxC Systems
Figure 2.2UniCel DxC 800 Analyzer
A015901P.EPS
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-6October 2005
Sample Handling System
Introduction
The Sample Handling system is composed of the following components:
• Sample racks
• Autoloader/Offload track
• Priority load position
• Shuttle
• Bar code reader
• Cap piercer assembly (optional)
• Sample Carousel
• Sample probe/mixer assemblies
• Wash cup assemblies
• Obstruction detection assemblies
Sample Handling System
Introduction
2
The sample handling module is used to load samples onto the system, provide samples
for analysis, and provide temporary storage of completed samples. A detailed
description of each component is presented in the following paragraphs.
Sample Racks
A sample rack is a high-strength, plastic, centrifugable holder designed to house up to
four samples. There are four sizes of racks with each size capable of holding one
length/width combination of primary sample tubes in addition to accepting sample
cups. (Refer to Figure 2.3.)
Sample racks accept the following tube and cup sizes:
Table 2.7Sample Racks
NOTICE
Adapters are provided to adapt various sized sample tubes (secondary tubes)
to the short racks. These adaptors must only be used in racks designated as
reserved. The reserved rack feature is described in CHAPTER 4,
Options.
System Setup
RackAccepts These Cups and Tubes
13 × 7512 × 75 mm tubes
13 × 75 mm tubes
0.5 mL cups
2.0 mL cups
(1 of 2)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-7
Sample Handling System
Rack ID Labels
Table 2.7Sample Racks, continued
RackAccepts These Cups and Tubes
16 × 7516 × 75 mm tubes
13 × 10013 × 100 mm tubes
16 × 10016 × 100 mm tubes
0.5 mL cups
2.0 mL cups
Capillary collection tubes (use with the capillary tube adapter)
Beckman Coulter Microtubes™
16 × 92 mm tubes
16.5 × 100 mm tubes
Beckman Coulter 0.5 mL Cup Insert (P/N 476399)
Rack ID Labels
Sheets of bar-coded rack ID labels are supplied with the system. They can be applied
as shown in Figure 2.3.
(2 of 2)
1. Numeric Rack ID number
2. Rack size label
3. Bar coded rack ID label
Figure 2.3Rack
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-8October 2005
Autoloader/Offload Track
When viewed from the front of the system, the autoloader is on the left and holds up to
25 sample racks in preparation for presentation to the DxC 800 system. The DxC 600
system has room to load a maximum of 14 racks. (Refer to Figure 2.4.)
There is also space for 25 sample racks in the offload track as they are removed from
the Sample Carousel upon completion. Combined, these two components allow for
over one hour of uninterrupted sample processing, with no operator intervention
involved.
When loading racks onto the autoloader, make sure that they are placed firmly
down into the autoloader.
1
NOTICE
Sample Handling System
Autoloader/Offload Track
2
2348765
1. Pushers
2. Autoloader
3. Run button
4. Priority load button
Figure 2.4Sample Loading Area
Priority Load Position
Between the Autoloader and the Sample Gate is the Priority Load position. This
position is used in conjunction with the PRIORITY LOAD button when a rack is to be
loaded onto the Sample Carousel into a reserved priority position so that it can be run
in a higher priority than other racks on the Autoloader. (Refer to Figure 2.4.)
A015902P.EPS
5. Sample gate
6. Shuttle
7. Offload track
8. Stop button
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-9
Sample Handling System
Priority Load Button
Priority Load Button
Typically, rack placement and removal is under microprocessor control. The operator
may choose to use the reserved positions by pressing the PRIORITY LOAD button
and placing the priority rack in the space provided by the system. (Refer to Figure 2.4.)
The rack will load into one of the reserved positions on the Sample Carousel.
PRIORITY LOAD only prioritizes the loading of the rack. It does not alter the sample
priority (STAT or routine) previously designated in Sample Programming.
Pushers
Pushers collect and move to the Sample Gate any racks loaded onto the system. They
are activated when the RUN button is pressed. (Refer to Figure 2.4.)
Sample Gate
The Sample Gate is the mechanism that moves racks from the load tray to the shuttle
during the load process. It also moves samples from the shuttle to the unload track
during the unload process. (Refer to Figure 2.4.)
Shuttle
The Shuttle moves the rack from the gate area onto the Sample Carousel. (Refer to
Figure 2.4.)
Bar Code Reader (Sample)
Do not tamper with or remove the housing of the Sample Bar Code Reader.
The Bar Code Reader is a Class II fixed-beam laser scanner. It is used to read the rack
bar code, the sample bar code (if present), and the background bar codes as the rack
travels past. The rack bar code and sample bar code (if present) are used to identify the
sample and link it to the appropriate sample programming.
There are two background bar codes that are used to determine whether a rack position
is empty or occupied, and if occupied, whether the sample is in a cup or tube.
Refer to "Symbols and Labels
of the CAUTION labels for the bar code reader.
CAUTION
" in CHAPTER 1, General Information, for a description
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-10October 2005
1-Blade Thick CTS (Closed Tube Sampling) Cap Piercer Assembly (optional)
1-Blade Thick CTS (Closed Tube Sampling) Cap Piercer Assembly (optional)
CAUTION
This Cap Piercer contains a razor sharp blade assembly.
CAUTION
Do NOT use this Cap Piercer assembly with foil-capped tubes. This would
give level sense errors.
This optional 1-Blade Thick CTS Cap Piercer assembly pierces thick-stoppered tubes
allowing the sample probe access to the sample without the need for cap removal.
Tubes with caps must be loaded in racks that have had the size correctly defined and
that are not Reserved (Reserved Racks = No CTS). Both open and closed tubes can be
in the same sample tube rack.
Table 2.8Tubes Validated for Closed Tube Sampling
Sample Handling System
2
Cap Piercer
Tube TypeTube Size
Configuration
1-Blade Thick CTS
Becton Dickinson VACUTAINER®a
with HEMOGARD™
Greiner VACUETTE
a
®a
13 × 75 mm
13 × 100 mm
16 × 100 mm
13 × 75 mm
13 × 100 mm
Sarstedt S-Monovette
®ab
75 × 15 mm
92 × 15 mm
a
All trademarks are the property of their respective owners.
b
This tube requires a special 5.5 mL rack (P/N A18642).
Refer to "Symbols and Labels" in CHAPTER 1, General Information, for a description
of the CAUTION labels for the Cap Piercer.
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-11
Sample Handling System
Sample Carousel
Sample Carousel
The ten-rack position Sample Carousel is a motor-driven turntable. (Refer to Figure
2.5.) Under normal operation, eight of the Sample Carousel positions are available for
routine processing and two positions are reserved for priority racks.
1
2
3
4
1. Liquid level sense assembly
2. Sample probe (CC)
3. Collar wash
4. Sample Carousel
Figure 2.5Sample Carousel Area
A015903P.EPS
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-12October 2005
Modular Chemistry (MC) System
Introduction
The Modular Chemistry system consists of the following major assemblies:
• Reagent storage area
• Ratio pump
• Sample probe
• Electrolyte injection cup (EIC)
• Flow cell assembly
• Chemistry reaction modules
A detailed description of each component is presented in the following paragraphs.
Reagent Storage Area
Modular Chemistry (MC) System
Introduction
2
The reagent containers used to supply the modular chemistries are located behind the
left front door of the system. The only exception is the CO2 alkaline buffer which is
located on the ISE module.
1
2
A015906P.EPS
1. Bar code reader (hand held)
2. Reagent storage area
Figure 2.6Modular Chemistry Reagent Storage Area
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-13
Modular Chemistry (MC) System
Reagent Bar Code Reader (Modular)
Reagent Bar Code Reader (Modular)
Behind the left side door of the system there is a hand-held bar code reader. (Refer to
Figure 2.6.)
When this reader is held up to the label of one of the modular reagents and the trigger
is pressed, the identity of the reagent, the lot number and reagent volume are
automatically entered into the reagent load screen.
Ratio Pump
The Ratio Pump is a motor-driven, multicylinder, positive-displacement pump used to
deliver the necessary reagents to the ISE flow cell. (Refer to Figure 2.7.) It consists of
a three-step piston housed in three, stacked, independent cylinders.
54321
1. Solenoid valve (example)
2. Cylinder 1
3. Cylinder 2
Figure 2.7DxC 800 Ratio Pump
6
A015904P.EPS
4. Cylinder 3
5. Outlet line (example)
6. Inlet line (example)
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-14October 2005
Electrolyte Injection Cup (EIC)
The EIC mixes the sample and buffer prior to delivery of the sample (now diluted) to
the flow cell.
Modular Chemistry (MC) System
Electrolyte Injection Cup (EIC)
2
1. Waste outlet
2. Flow Cell outlet
3. Buffer inlet
4. Reference inlet
5. DI H
Figure 2.8Electrolyte Injection Cup
O inlet
2
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-15
Modular Chemistry (MC) System
Flow Cell Assembly
Flow Cell Assembly
The flow cell assembly houses the seven electrodes that perform the analysis of
sodium, potassium, chloride, carbon dioxide, and calcium. (Refer to Figure 2.9.)
1. Inlet port
2. CL electrode
3. K electrode
4. CO
5. CO
reference electrode
2
electrode
2
Figure 2.9Flow Cell
6. Exit port for waste (large tube)
7. Exit port for internal reference
8. Na reference electrode
9. Na electrode
10. Ca electrode
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-16October 2005
Chemistry Reaction Modules (Basic Components)
Each of the six Chemistry Reaction Modules have similarities in their design. These
common elements are described below. (Refer to Figure 2.10.) Unique design
elements of the modules are described under the specific module headings later in this
section.
2
Chemistry Reaction Modules (Basic Components)
Modular Chemistry (MC) System
3
4
2
1
1. Circuit board (behind protective shield)
2. Reaction cup
3. Reaction cup housing
4. Mixer motor assembly
5. Reagent pump assembly
5
A015905P.EPS
Figure 2.10Basic Components of Modules (Ex: Albumin)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-17
Cartridge Chemistry (CC) Reagent Handling System
Introduction
Cartridge Chemistry (CC) Reagent Handling System
Introduction
The Cartridge Chemistry Reagent Handling system is composed of the following
components:
• Reagent cartridges
• Reagent carousel
• Reagent probe assembly
• Reagent mixer assembly
• Mixer wash cup
The Reagent Handling system is used to transfer reagent from the individual cartridges
to the reaction cuvettes for processing and analysis of the requested chemistry tests.
Reagent Cartridges
Reagent cartridges are single use, recyclable plastic containers that house the
individual liquid reagent components necessary to perform a chemistry test. The
reagent carousel is capable of storing 59 cartridges on board.
1. A compartment
2. B compartment
3. C compartment
Figure 2.11CC Reagent Cartridge
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-18October 2005
Reagent Carousel
The Reagent Carousel Compartment provides an on-instrument storage area for the
individual reagent cartridges. A total of 59 reagent cartridges can be stored in the
carousel at one time. (Refer to Figure 2.12.)
The storage compartment is refrigerated and fan-cooled to maintain a temperature of
+5°C (±3°C).
Cartridge Chemistry (CC) Reagent Handling System
1
2
3
4
Reagent Carousel
2
1. Top positions 31–59
2. Top Bar Code Reader
3. Bottom positions 1–30
4. Bottom Bar Code Reader
Figure 2.12CC Dual Carousels with Two Bar Code Readers
Reagent Bar Code Reader (CC)
The Bar Code Reader, (refer to Figure 2.12), situated near the front of the Cartridge
Chemistry reagent access door, scans each label during the loading of reagent
cartridges. Only the reader for the selected carousel is active. An audible signal
acknowledges successful reading of the label.
The Reagent Probe assembly consists of a mechanical structure that supports two
moveable cranes. Attached to each crane is a pickup probe. (Refer to Figure 2.13
A007408P.EPS
.)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-19
Cartridge Chemistry (CC) Reagent Handling System
Reagent Mixer Assembly
1
1. Reagent Mixer
2. CC Reagent Probe A
3. CC Reagent Probe B
4. Collar Wash
2
4
3
A015908P.EPS
Figure 2.13CC Reagent Probe Area
Reagent Mixer Assembly
This assembly consists of a mechanical structure that supports a single, moveable
crane. Attached to the crane is a high-speed mixer. (Refer to Figure 2.13.)
Reagent Mixer Wash Cup
The Mixer Wash Cup sprays the mixer with diluted wash solution while the mixer
moves up and down in the cup. (Refer to Figure 2.13.)
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-20October 2005
Cuvette Reaction System
Introduction
The Cuvette Reaction system consists of the following components:
The Cuvette Reaction system involves the process of obtaining absorbance readings
from each cuvette during the analysis cycle.
Following the completion of each chemistry test, the cuvettes are processed through a
wash station in preparation for the next chemistry.
Cuvette Reaction System
Introduction
2
A description of each component is presented in the following paragraphs.
Reaction Carousel
The reaction carousel assembly, (refer to Figure 2.14), supports a total of 125 cuvettes.
Each cuvette is glass with a 0.5 cm path length and is approximately 30 mm high. The
cuvettes are non-disposable and have an indefinite life-span on the instrument.
Cuvettes only need replacement if they are damaged. (They remain under warranty for
two years.)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-21
Cuvette Reaction System
Photometer Assembly
1. LPIA module
2. Reaction carousel
3. Photometer
2
3
1
A015909P.EPS
Figure 2.14Reaction Carousel Area (Typical – Cover Removed)
Photometer Assembly
Attached to the reaction carousel support frame is the Photometer assembly. This
consists of a xenon pulse lamp, a discrete 10-position silicon-diode detector array, a
monochromator housing unit, and associated electronic circuitry. (Refer to Figure
2.14.)
As each cuvette passes through this optics station during a spin cycle, the xenon lamp
is flashed and the resulting light beam travels through the opposing sides of the square
cuvette.
The optional LPIA module uses a photometric detection system for large particle
immuno assays. This assembly is attached to the reaction carousel support frame to the
left of the photometer assembly, near the sample carousel. (Refer to Figure 2.14.) It
has two printed circuit boards, an LED (light-emitting diode) and a photodetector.
Communication with the system software is through fiber optic cables.
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-22October 2005
Cuvette Wash Station
The Cuvette Wash Station, (refer to Figure 2.15), consists of four coaxial probes, an
elevator assembly, and the associated tubing.
A motor controls the vertical motion required by the elevator to raise and lower the
probes during the wash stage.
Cuvette Reaction System
Cuvette Wash Station
2
1. Probe #1
2. Probe #2
3. Probe #3
Figure 2.15 Cuvette Wash Station
4. Probe #4
5. Wash station
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-23
Hydropneumatic System
Introduction
Hydropneumatic System
Introduction
The main components of the Hydropneumatic System are mounted on a slide-out
drawer that allows for easier operator access. (Refer to Figure 2.16 and Figure 2.17.)
When fully extended, the drawer locks open. To close, lift up on the metal tabs,
located on each side of the bottom runner of the hydropneumatic unit, and push the
drawer inward.
6
5
Function
4
3
2
1
A016497P.EPS
1. Wash concentrate solution
2. No foam reagent
3. DI water reservoir
4. Dilute wash solution reservoir
5. Wash concentrate reservoir
6. Auto-Gloss
Figure 2.16DxC 800 Hydropneumatics (right side)
The function of the hydropneumatic system is to provide the following media to the
different functional areas of the instrument:
• Vacuum
• Compressed air
• Diluted wash solution
• Deionized water
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-24October 2005
Hydropneumatic System
567
4
3
1
2
A015911P.EPS
Function
2
1. Waste exit
2. Waste B exit
3. DI water inlet On/Off
4. DI water reservoir
5. Vacuum accumulator
6. Waste B canister
7. Waste canister
Figure 2.17DxC 800 Hydropneumatics (left side)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-25
Operation and Control Components
Operator Controls
Operation and Control Components
Operator Controls
The operator interfaces with various control devices such as the keyboard, monitor and
push-button controls during a routine run. Basic operating functions are controlled and
reviewed from the monitor. Calibration functions are also controlled from the monitor.
Information is selected and entered into the system through touch screen monitors,
from a mouse, and/or at a keyboard.
Push-button controls (refer to Figure 2.4) are used to start the process, to prioritize a
sample run, or to stop the process under certain conditions as described in Table 2.9
below.
Table 2.9DxC Push-Button Controls
Push-Button
Primary Function
Control Type
RunTo start the test process.
PriorityTo prioritize the loading of a rack by creating a space in front of
the autoloader for loading the next rack into a priority position on
the sample carousel. It does not alter the sample priority (STAT or
routine) previously defined in sample programming.
StopTo stop the process. The stop button should be used only under
the following conditions:
• To stop instrument motions
• To conduct a maintenance/repair activity
• To home and realign mechanical components without
rebooting
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-26October 2005
Main Operator Screen and Program Structure
Main Operator Screen
DxC System operating and programming functions are initiated from the main
operator screen at the DxC analyzer (refer to Figure 2.18). In addition, the screen
provides status information to help determine the present state of the system.
Main Operator Screen and Program Structure
Main Operator Screen
2
1
CTS
3
1. CTS Indicator
2. Host Indicator
3. System Status Indicator
4
4. Function Selection Icons
5. Sample Status Icons
6. Sample Status Icon Legend
2
Figure 2.18UniCel DxC 800 Main Operator Screen
6
5
E016482S.EPS
Accessing the Help System
Help is available online in six supported languages (the default is English). To reach
the on-line Help system, select the Question Mark icon () on the right side of the
Function Selection Icon menu. The system will offer three options:
• Install an On-Line Instructions For Use manual (IFU)
• View the installed On-Line Instructions For Use manual
• View the License Agreement
When you access the On-Line Instructions For Use manual screen, the currently
loaded language manual version will be show in the Choose a Language dropdown
field. To view the IFU Manual in that language, press (on the touch screen) or select
(with the mouse) the <View Manual> button. The first page of the IFU Manual will
appear.
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-27
Main Operator Screen and Program Structure
Printing Data From a Screen
To load an alternate language, select the desired language from the Choose a
Language dropdown list and select the <Install Manual> button. You can then view
the IFU Manual in the selected language as described in the preceding paragraph.
To view the License Agreement, simply select the <License> button on the screen.
Printing Data From a Screen
Many screens contain data that may be printed by selecting Print [F10]. Pressing the
[Control] and [Print Screen] key simultaneously on the keyboard will also cause the current screen to print, regardless of whether the Print [F10] option is available.
Status Functions
The following Table depicts the status information available from the operator screens
of the DxC analyzer.
Table 2.10Main Operator Screen Status Functions
Status IndicatorStatus Description
CTSThis indicator (refer to item 1 on Figure 2.18) appears in the blue
bar at the top left side of the Operator screens when the Closed
Tube Sampling (CTS) option is installed. When "CTS" appears,
Closed Tube Sampling is enabled. When CTS appears within the
international "No" symbol (), the feature is not enabled.
When the indicator is absent, the CTS option is not installed.
Enabling and disabling Closed Tube Sampling is accomplished
using the Setup procedures described in the "Reserved Racks/
Obstruct Detect" topic of CHAPTER 4, System Setup Options.
NOTICE
CTS is an optional feature of the system, which allows the
system to pierce primary sample tubes. The operator will
only be offered the enable/disable CTS option if a CTS
assembly is installed, otherwise that selection will be grayed
out on the setup screen. The default for CTS sampling is
"OFF."
HostThis indicator (refer to item 2 on Figure 2.18) appears on the blue
bar at the top right side of the Operator screens. The indicator to
the right of the label shows communication activity between the
instrument and the host computer. A blue bar indicates that the
host is sending information. A green bar indicates that the host is
receiving information.
(1 of 2)
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-28October 2005
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
Table 2.10Main Operator Screen Status Functions, continued
Status IndicatorStatus Description
System StatusThis indicator appears just below the Function Selection Icons
on the left side of the Operator screens (refer to item 3 on Figure
2.18). It shows the current state of the system: either Running or
Standby. When both the Modular Chemistry (MC) and the
Cartridge Chemistry (CC) functions have the same status, a single
status indicator appears. When the statuses of these components
are different, the MC Status is shown on the left, and the CC
Status appears on the right. For example, Running/Standby would
indicate that the MC side is Running, and the CC side is in
Standby mode.
(2 of 2)
Sample Status Indicators
When monitoring sample status, a sample status icon (refer to item 5 on Figure 2.22)
appears in front of listed sample. Samples are listed in a rack status box directly below
the sample status icon legend (6). The sample statuses shown are as follows:
2
Table 2.11Sample Status Indicators
Status IndicatorStatus Description
Not ProgrammedIndicates a sample has been loaded without any programming
associated with that Sample ID.
Query PendingSample is waiting for specific program information to be
downloaded from the host.
In ProgressSample has been identified and is currently being processed.
AspiratedIndicates whether an initial aspiration of the sample has been
accomplished.
IncompleteSample has some tests that are still pending.
Function Selection Icons and Program Structure
Near the top of the Operator screens, a series of icons on the touch screen provide
access to each of the major functional areas of the system (refer to item 4 on Figure
2.18). The overall program structure is shown in beginning on the following page.
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-29
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
Main
Samples
F1 Results
F2 Unload
F8 Log
F9 Pre Run
F10 Post Run
F2 Demog
F3 Options
F4 Batch
F5 Control
F6 Rerun
F7 Clear
F1 Sample
F2 Edit
F3 Urine
F10 Next
Clear
OK
Smp ID Only
F3 Options
F5 Control
F6 Remove
F8 Rgt Cart
F9 Cancel
F10 Save
Select
All
Batch
Cancel
Sample IDs
Rack(s)/Position(s)
Date/Time (From/To)
OK
Cancel
System Replicates
Sample Replicate
Serum Index
Off-line Dilution Factor
Manual ORDAC
Test Replicates
OK
Cancel
Rack
End
System Replicates
Sample Replicate
Serum Index
Off-line Dilution Factor
Manual ORDAC
Test Replicates
OK
Cancel
Fill Canisters and reservoirs
CC Drain waste sump
CC Drain waste B sump
Drain gravity drain sump
Cup Selection:
ALBm (DxC 800 only)
CREm (DxC 800 only)
GLUCm (DxC 600 only)
PHOSm (DxC 800 only)
TPm (DxC 800 only)
BUNm (DxC 800 only)
Prime Cup(s) with:
Reagent
DI Water
Sample Delivery Subsystem
ISE All
Electrolyte Buffer
Electrolyte Reference
EIC Wash
ISE CO2 Alkaline Buffer
Reference and Acid
1. 1-Blade Thick CTS Blade Replacement
2. Maintenance
2. Cartridge Chemistry Probe Cleaning
3. ISE Service
4. Cup Maintenance
5. Hydropneumatic Maintenance
6. Wash All Cuvettes
7. Syringe Rod Replacement
8. CUPs Lamp/Sensor Calibration
9. CC Reagent Wash All Cuvettes
10. Clean Flow Cell and Cups
F1 Action
F2 Log
F10 Exit
Figure 2.23Program Structure (Utils)
1. Albumin (DxC 800 only)
2. BUNm (DxC 800 only)
3. Creatinine (DxC 800 only)
4. Glucose (DxC 600 only)
5. Phosphorus (DxC 800 only)
6. Total Protein (DxC 800 only)
Prime20
Prime5
Drain
Rinse
Cancel
E007327L.EPS
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-34October 2005
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
2
Utils,
continued
3. Event LogDisplay
4. Diagnostics
5. Alignment
Copy
Time
Clear
Print
Done
1. Chemistry Errors
2. Motion Errors
3. Status Monitor Errors
4. Other Instrument Errors
5. Instrument Events
6. LIS Comm. Errors
7. Other Console Errors
8. Input Device Events
9. Other Console Events
10. Sample Processing Events
11. Deleted Results
Functional Areas
Sub-functional Areas
Test Selection
F10 Exit
Align All
Reaction Carousel
Cuvette Wash Station
Reagent Carousel
Sample Carousel
Sample Rack Handling
Sample Rack Barcode Reader
Cap Piercer
CC Sample Crane or Probe
CC Sample Crane Smart Module (B2)
MC Sample Crane or Probe
MC Sample Crane Smart Module (52)
Reagent A Crane or Probe
Reagent A Crane Smart Module (C2)
Reagent B Crane or Probe
Reagent B Crane Smart Module (C3)
CC Sample Mixer Crane or Body
CC Sample Mixer Smart Module (B3)
Reagent Mixer Crane or Body
Reagent Mixer Smart Module (C1)
CTA Assemblies
Procedure Selection - Based on Functional Area selected
F10 Exit
Sample System
Reagent System
Reaction System
Hydropneumatic System
ISE Module System
Modular Cup Chemistry Module
Valve and Fluidics
Based on Functional Area selected
Based on Functional Area selected
E007328L.EPS
Figure 2.24Program Structure (Utils – continued)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-35
Main Operator Screen and Program Structure
Function Selection Icons and Program Structure
Utils,
continued
Status
Instr
Cmd
Help
6. PVT
7. Modem
8. Backup/Restore
9. Metering
10. Touch Screen Calibration
F1 Summary
F2 Count
F3 Temp
F4 Power
F5 Hydro
F6 Smrt Mdl
F7 Cuv Stat
F8 CTS
F10 Print
1. Home
2. Pause
3. Stop Print
4. Shutdown
5. Pause Waste B
6. Resume Waste B
7. Enable/Disable Modules
8. Unload All
Choose a Language:
English (default)
Italian
Spanish
French
German
Japanese
Install Manual
View Manual
License
Performance Verification Tests
Sub-Functional Area
Test Selection - Based on Functional Area Selected
F10 Exit
System Parameters
Alignment Data Files
Backup
Restore
Cancel
F1 Disk
F2 Replen
F3 Report
F4 Off
F5 Modem
F9 Cancel
F10 Done
E007329L.EPS
Figure 2.25Program Structure (Utils – continued 2, Status, Instr Cmd, and Help)
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-36October 2005
Theory of Operation
Introduction
The UniCel DxC Synchron Clinical Systems are microprocessor-controlled, random
access clinical analyzers capable of processing a wide variety of operator-selected
chemistries in a single run.
Cartridge Chemistries (CC)
The optical system of the DxC enables rate, endpoint, and nonlinear analyses to be
performed simultaneously. These analyses are referred to as cartridge chemistries
because the reagents are stored in cartridges.
Cartridge Chemistry (CC): Sample and Reagent Processing
During operation, a number of events occur simultaneously and are under direct
control of the instrument microprocessors.
Theory of Operation
Introduction
2
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-37
Cartridge Chemistry: Calibration Theory
Introduction
Cartridge Chemistry: Calibration Theory
Introduction
Calibration determines the relationship between measured reaction responses and
known concentrations. Calibration factors are derived from this relationship. These
factors are used to convert the measured reaction responses to final concentration
results.
Calibrated chemistries include endpoint and first-order rate chemistries, drugs, DATs,
and specific proteins. Zero-order rate chemistries include enzymes, which are
precalibrated and require no routine calibration. Enzyme verification can be performed
on some of the enzymes to conform to International Federation of Clinical Chemistry
(IFCC) guidelines at 37°C.
Endpoint and First-Order Chemistries
Calibration of endpoint and first-order rate chemistries involve the use of a
single-level calibrator solution or a two level calibrator kit. Each analyte in the
calibrator solution has a known concentration value associated with it. With each new
lot of calibrator solution, the values are transferred from disk and stored in memory for
later use in the calibration procedure.
For most calibrated cartridge chemistries, the system will set calibration factors based
on four calibrator replicates per calibrator level. The instrument will determine and
discard the highest and lowest of the four replicates. The remaining two values are
called the usable calibrator replicates. All four replicates will appear on the report but
the average of the two usable replicates is used to determine the calibration factor.
For other chemistries, calibration is based on two calibrator replicates per calibrator
level. No replicates will be discarded. The average value of the calibrator replicates is
used to determine the calibration factor.
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-38October 2005
Endpoint and First-Order Calibration Formulas
The calibration factor is determined by using one of the following equation sets where
reaction and blank are used from the usable replicates.
Table 2.12Calculation of Calibration Factors for Endpoint and Rate
Chemistries
TypeFormula
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
2
Nonblanked
Endpoint
Chemistries
FOR HIGH CALIBRATOR LEVEL:
Reaction ABS = ABS
Reaction ABS = ABS
(ABS
rep1
+ ABS
rep1
rep2
) × 0.5 = ABS
rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
Reaction ABS = ABS
Reaction ABS = ABS
(ABS
+ ABS
rep1
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × ABS
rep1
rep2
) × 0.5 = ABS
rep2
(lo)
avg
Cal Set Point (hi) – Cal Set Point (lo)
ABS
(hi) – ABS
avg
avg
avg
(hi)]
(lo)
Sample values are calculated by the following equation:
(Reaction ABS × Cal Factor) + offset = sample value
E014416L.EPS
NOTICE
For single point linear calibration, the low calibrator is a fixed zero
point and the offset is equal to zero.
(1 of 6)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-39
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
Table 2.12Calculation of Calibration Factors for Endpoint and Rate
Offset = Cal Set Point (hi) - [Cal Factor × Delta ABS
Cal Set Point (hi) – Cal Set Point (lo)
Delta ABS
(hi) – Delta ABS
avg
avg
E014417L.EPS
(hi)]
avg
(lo)
Sample values are calculated by the following equation:
[[[Reaction ABS - (Blank ABS × Blank Correction Factor)]
* Cal Factor]] + offset = sample value
NOTICE
For single point linear calibration, the low calibrator is a fixed zero
point and the offset is equal to zero.
(3 of 6)
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-41
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
Table 2.12Calculation of Calibration Factors for Endpoint and Rate
Chemistries, continued
TypeFormula
Nonblanked
Rate
Chemistries
FOR HIGH CALIBRATOR LEVEL:
Reaction Rate = Rate
Reaction Rate = Rate
(Rate
rep1
+ Rate
rep1
rep2
) × 0.5 = Rate
rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
Reaction Rate = Rate
Reaction Rate = Rate
(Rate
+ Rate
rep1
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × Rate
rep1
rep2
) × 0.5 = Rate
rep2
(lo)
avg
Cal Set Point (hi) – Cal Set Point (lo)
Rate
(hi) – Rate
avg
avg
avg
(hi)]
(lo)
Sample values are calculated by the following equation:
(Reaction Rate × Cal Factor) + offset = sample value
E014418L.EPS
NOTICE
For single point linear calibration, the low calibrator is a fixed zero
point and the offset is equal to zero.
(4 of 6)
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-42October 2005
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Calibration Formulas
Table 2.12Calculation of Calibration Factors for Endpoint and Rate
Offset = Cal Set Point (hi) - [Cal Factor × Delta Rate
Cal Set Point (hi) – Cal Set Point (lo)
Delta Rate
(hi) – Delta Rate
avg
avg
avg
E014419L.EPS
(hi)]
Sample values are calculated by the following equation:
[[Reaction Rate - (Blank Rate × Blank Correction Factor)]
× Cal Factor] + offset = sample value
NOTICE
For single point linear calibration, the low calibrator is a fixed zero
point and the offset is equal to zero.
(lo)
(6 of 6)
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-44October 2005
Non-Linear Chemistries
Non-linear chemistries include drugs and specific protein assays. Unlike the first-order
rate and endpoint chemistries, which exhibit a linear response to increasing
concentration, the calibration curves for non-linear chemistries exhibit logarithmic
(S-shaped) or other nonlinear relationships. For this reason, curve fitting interpolation
techniques are employed to construct the calibration curve.
For some non-linear calibrations, the curve parameters for a reagent lot are calculated
during manufacturing. The curve parameters are encoded in bar code form, shipped on
a card in the reagent box and loaded onto the system. A single point calibration is then
run to adjust for instrument-to- instrument variation. If necessary, the calibration
includes the sample diluent (DIL1), used as a blank, that is subtracted from all
calibrator or sample responses.
• Multi-point chemistry calibration consists of five or six different levels of
calibrators. These chemistries set calibration based on single replicates of each
calibrator level.
Cartridge Chemistry: Calibration Theory
Non-Linear Chemistries
2
• Single-point chemistry calibration consists of one or two levels of calibrators. These
chemistries set calibration based on two to four replicates of each calibrator level.
Non-Linear Calibration Formulas
The standard curve is determined by use of one of several nonlinear math models. The
system uses an iterative technique to calculate the curve parameters. A modified
Newton iteration is used to choose values. The best-fitting calibration curve is
determined by minimizing the sum of the difference between the observed response
and the calculated response of each standard.
The following symbols are used in the math models presented below:
R= sample response
Conc= standard or sample concentration
R
K
= calculated response for a zero sample
O
= scale parameter
C
a, b, c = parameters which define the nonlinear elements of the math model
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-45
Cartridge Chemistry: Calibration Theory
Non-Linear Calibration Formulas
Table 2.13Math Models for Non-Linear Chemistries
TypeFormula
Model #1Math Model #1 is the four-parameter log-logit function most
commonly used with reagents that use antibodies.
R = R0 + K
c
1 + e
1
–a–b∗1n(conc)
E014420L.EPS
Sample values are determined using the calculated curve parameters
and the math model. Values may be calculated directly as this model
can be solved for concentration.
Model #2Math Model #2 is a five-parameter logit function.
R = R0 + K
c
1 + e
1
–a–b∗1n(conc) – c(conc)
E014421L.EPS
This function cannot be solved directly for concentration. The
instrument uses an iterative method to determine the sample value.
Model #3Math Model #3 is a five-parameter exponential function.
R = R0 + K
a∗1n(conc) + b∗1n(conc)2 + c∗1n(conc)
e
c
3
E014422L.EPS
This function cannot be solved directly for concentration. The
instrument uses an iterative method to determine the sample value.
Models #4
These Models are reserved for future development.
through #7
Model #8 Math Model #8 is an alternative to model #2, the five-parameter logit
function.
c
E014423L.EPS
R = R0 + K
c
1 + e
1
–a–b∗1n(conc)
This function cannot be solved directly for concentration. The
instrument uses an iterative method to determine the sample value.
(1 of 2)
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-46October 2005
Cartridge Chemistry: Calibration Theory
Drugs of Abuse Testing (DAT) Chemistries
Table 2.13Math Models for Non-Linear Chemistries, continued
TypeFormula
Model #9Math Model #9 is an extension to model #1, the four-parameter log-
logit function.
2
Polynomial
Exponential
(PXP)
Quadratic
(POLY2)
Lorentz
Double
Inflection
Model DP4
R = R0 + K
c
1 + c ∗ e
1
–a–b∗1n(conc)
E014424L.EPS
The "c" is allowed to be either +1 or -1.
If c = +1, then this is equivalent to model #1.
If c = -1, an alternative function is being used.
This function cannot be solved directly for concentration. The
instrument uses an iterative method to determine the sample value.
R = b0 + Pn(x)e
Pn(x) = Polynomial of degree N
R = b0 + b1∗(conc) + b2∗(conc)
K
c
R = R0 + π {arctan(c∗conc + a) + 2 }
R =
1 +
-a∗(conc)
E014476L.EPS
E014477L.EPS
K
c1
a
1
Conc Conc
+
1 +
K
c2
a
2
A011537L.EPS
2
π
E014478L.EPS
(2 of 2)
Drugs of Abuse Testing (DAT) Chemistries
The Drugs of Abuse Testing (DAT) assays require three levels of calibrators. The
calibration measures the separation between calibrators to measure reagent integrity.
The calibration factor generated is non-functional for sample result calculation.
The cutoff value for each DAT chemistry represents the mean reaction rate of the low
calibrator, reported in mA/min units on patient and control reports. The reaction rate
of the samples is compared to the reaction rate of the low (cutoff) calibrator and
reported out as POSITIVE or NEGATIVE. Cutoff values are stored in memory until
the next successful calibration.
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-47
Cartridge Chemistry: Calibration Theory
Enzyme Verification
Enzyme Verification
Enzyme verification is a means of adjusting enzyme chemistry reporting units to IFCC
methods. This feature is available for ALP, ALT-, AST-, CHE, CK-, GGT and LD.
Verification also allows results to be adjusted for country specific correlation needs.
Verification is similar to calibration except that normalization factors are applied to
the sample result in the form of a slope and offset adjustment, whereas calibration
factors would be applied to the reaction response.
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-48October 2005
Modular Chemistry: Calibration Theory
Calibration Theory
Modular chemistries are calibrated using two to three levels of calibrator (chemistry
dependent). Four replicates per level are assayed. Data from two middle replicates of
each level is used to set the system response. The highest and lowest replicates are
discarded. Error checks are performed on the two middle replicates to verify
successful calibration.
Calibration Error Detection
The analog signals generated by the calibrator measurements are converted to digital
form. The resulting ADC values are compared to pre-programmed back-to-back, span
and range limits to determine the calibration acceptability.
Modular Chemistries (MC)
The UniCel DxC contains seven chemistry modules (see chart below), each of which
is used in the determination of eleven modular chemistries (MC), as follows.
Modular Chemistry: Calibration Theory
Calibration Theory
2
Table 2.14Methodology and Modules Used with Modular Chemistries
UniCel DxC Systems Instructions For Use A13914System Description
October 2005Page 2-49
Cartridge Chemistry: Principles of Measurement
Spectrophotometric Methods
Cartridge Chemistry: Principles of Measurement
Spectrophotometric Methods
Spectrophotometric methods rely on the principle that a sample, such as a patient
sample, a control, or a calibrator, when mixed with one or more appropriate chemical
reagents, produces a substance that has the ability to absorb light at specific
wavelengths. This substance is referred to as a chromophore.
Beer's Law
According to Beer's Law, the amount of light absorbed by the chromophore is
proportional to the concentration of the constituent being measured. The system can
measure this as an endpoint or a rate of formation.
A = abc
where
A= absorbance of the chromophore
a= absorptivity of the absorbing substance at the specific measuring
System DescriptionUniCel DxC Systems Instructions For Use A13914
Page 2-50October 2005
CHAPTER 3 Preparing Samples for Analysis
Routine Operation Overview
Daily Procedure
The following table shows an example of daily work flow using the DxC Synchron
Clinical Chemistry System.
This procedure assumes that the initial system setup has been completed.
StepAction
1If necessary, start the system.
2Check reagent status. Load reagent as necessary.
NOTICE
Routine Operation Overview
Daily Procedure
3
3Check calibration. Program or load calibrators as needed.
4Program or load controls, if required.
5If the system is not already running, press the RUN button on the system.
6Check the control results to verify system operation.
7Program or load patient samples.
8If the system has gone into Standby, press the RUN button on the system.
9Review the patient results.
10Return to Step 7 if more samples need to be run.
11The system will automatically return to Standby when all testing has been
completed.
UniCel DxC Systems Instructions For Use A13914Preparing Samples for Analysis
October 2005Page 3-1
Sample Programming and Processing
Introduction
Sample Programming and Processing
Introduction
Sample programming provides the ability to identify samples, select tests to run,
describe samples, and designate how to run samples. Samples are programmed
through a host computer or Laboratory Information System (LIS or at the analyzer).
The minimum information required to save a sample program includes:
• a sample ID or a rack and position assignment
• and one test selection
Test selections are assigned by the use of panels or by the selection of individual tests.
Samples may be described and defined through sample type, sample comment, patient
ID, and patient demographic functions. A sample can be programmed as a control
sample or as a STAT for priority processing. Once programmed, samples are placed on
the autoloader for processing.
If your LIS or normal workflow requires the reuse of sample IDs, the sample
programming should be cleared from the DxC at a time interval that is less
than the shortest time of sample ID reuse. If this warning is not followed,
results from the new request will be merged with tests from incomplete
samples that previously used that ID.
Program Samples and Controls
Refer to this chapter for Sample information and CHAPTER 6, Quality Control for
Control programming information. If auto QC is enabled, reuse the same control ID
bar code labels.
CAUTION
Preparing Samples for AnalysisUniCel DxC Systems Instructions For Use A13914
Page 3-2October 2005
Prepare Samples
Minimum Sample Volume
A minimum sample volume is required to run tests. To determine what volume of
sample to use, refer to the Synchron LX/UniCel DxC Clinical Systems Sample Template.
Sample Racks
Sample racks accept the following tube and cup sizes:
Table 3.1Sample Racks
13 × 7512 × 75 mm tubes
Sample Programming and Processing
RackAccepts these cups and tubes
13 × 75 mm tubes
0.5 mL cups
2.0 mL cups
Prepare Samples
3
16 × 7516 × 75 mm tubes
13 × 10013 × 100 mm tubes
0.5 mL cups
2.0 mL cups
Capillary collection tubes (use with the capillary tube adapter)
16 × 10016 × 100 mm tubes
16.5 × 92 mm tubes
Beckman Coulter 0.5 mL Cup Insert (P/N 476399)
NOTICE
Adapters are provided to adapt various sized sample tubes (secondary tubes)
to the short racks. These adaptors must only be used in racks designated as
reserved. The reserved rack feature is described in this chapter.
UniCel DxC Systems Instructions For Use A13914Preparing Samples for Analysis
October 2005Page 3-3
Sample Programming and Processing
Sample Preparation by Container Type
Sample Preparation by Container Type
Table 3.2 shows how to prepare different sample containers.
Table 3.2Preparation of Sample Containers
If running a sample from a...Then...
Primary Tube• Use the sample template in the Appendix to determine
adequate sample volume.
• Remove the stopper.
• For CTS systems, remove the stopper, if not a validated
closed tube.
Secondary Tube• Remove the cap.
• Determine sufficient volume.
• Check for fibrin or other materials resulting from
storage.
Beckman Coulter
Synchron
Microtube™
Autoanalyzer
Cup
BD Microtainer
®
0.5 mL2.0 mL
E014498L.EPS
• Pipette the sample into a Synchron Microtube™.
• Verify there are no bubbles at the bottom of the tube.
• Place into a 13 × 100 "reserved" rack.
• Place the cup into a rack.
OR
• Place in 15 × 85 tube (with sample bar code on tube).
• Verify there are no bubbles in sample.
• If cup is placed in a tube with a Bar Coded label a
"reserved rack" must be used.
• Place Microtainer in adapter P/N 472987.
(refer to figure to the right)
• Verify there are no bubbles in sample.
• If cup is placed in a tube with a Bar Coded
label a "reserved rack" must be used.
0.5 mL Cup Insert
(P/N 476399)
(reusable)
• Place the metal Cup Insert into a 16 × 100 mm rack
(P/N 471921 – blue or 474463 – purple).
• Insert a 0.5 mL Autoanalyzer cup into the Cup Insert.
• Run in the "reserved rack" mode.
A011538L.EPS
Preparing Samples for AnalysisUniCel DxC Systems Instructions For Use A13914
Page 3-4October 2005
Sample Tubes Validated for CTS
Introduction
Rubber cap tubes can be run directly on the system without removing the cap. The
CTS assembly cuts a small hole in the cap and the sample probe aspirates a sample
directly from the tube. With this procedure, samples are processed faster and safer.
CTS Validated Sample Tubes
Only validated sample tubes can be run on UniCel DxC Systems with CTS. However,
capped and uncapped tubes can be placed on the same reserved rack. Each tube is
scanned to determine if it needs piercing. Refer to the table below for validated tubes.
Table 3.3Tubes Validated for Closed Tube Sampling
Sample Programming and Processing
Sample Tubes Validated for CTS
3
CTS ConfigurationTubes Validated
for Cap Piercing
1-Blade Thick CTS
Becton Dickinson VACUTAINERa
with HEMOGARD
Greiner VACUETTE
Sarstedt S-Monovette
a
All trademarks are property of their respective owners.
b
This tube requires a special 5.5 mL rack (P/N A18642).
CAUTION
Before placing validated closed tubes on the UniCel DxC Systems, check the
top of the cap for any residual blood. Residual blood can contaminate the
sample and affect results. If blood is present, remove it by using a cottontipped applicator stick moistened with DI water. When running in the CTS
mode, if tubes off-loaded from the UniCel DxC Systems have water or
droplets of water on the caps, disable the CTS and call Beckman Coulter
Technical Support. Note: Oil on a cap is normal.
a
a
®ab
Size
13 × 75mm
13 × 100 mm
16 × 100 mm
13 × 75mm
13 × 100 mm
75 × 15 mm
92 × 15 mm
Reserved Racks for CTS Samples
The Reserved Racks/Obstruct Detect feature allows the user access to the following
features:
Closed Tube Sampling (CTS)
This is an optional feature that allows the system to pierce primary sample tubes. If a
CTS assembly is installed, the operator can enable/disable the CTS assembly. A CTS
Tracking database monitors pierced tubes. If you run a tube again, it will not be
pierced again. The sample ID stays in this database for seven days for a tube that has
been pierced.
UniCel DxC Systems Instructions For Use A13914Preparing Samples for Analysis
October 2005Page 3-5
Sample Programming and Processing
Reserved Racks
The 16 × 100 tubes must be run in a non-reserved 16 × 100 mm rack.
If a sample ID is manually cleared, the CTS tracking information is cleared on
that instrument only. If a sample ID is reused and the tube requires piercing,
the sample ID must be cleared on the instrument that is loaded first.
Do not clear the Sample ID manually. Before you run a sample whose cap has been
pierced, be sure that its sample ID has not been cleared.
• If you clear a sample ID at the Host, the sample ID is NOT cleared in the CTS
Tracking database.
• If you were to manually clear the sample ID for a tube, you would also clear the
sample ID in the CTS Tracking database. The instrument would think it has a new
tube and would pierce its cap again. Follow the steps below to verify sample ID
clearing status.
NOTICE
StepAction
Reserved Racks
If a rack number is entered into this field, any sample containers in this rack will not be
cap pierced even if the Cap Piercing feature is enabled. Level sensing on samples in
the rack will be set to its most sensitive level to detect small sample volumes. This
type of reserved rack may be used for uncapped, primary or secondary tubes that have
a small volume of sample, for example: nested cups on primary tubes, Microtubes™
and pediatric capillary collection tubes.
1Select the Samples icon from the menu bar.
2Type in the sample ID.
3Press [Enter].
4Does the sample have its initial program?
If…Then…
The sample has its initial
program,
The sample does NOT have its
initial program (the ID has been
• Do NOT remove its stopper.
• Run the sample.
• Remove its stopper.
• Run the sample.
cleared),
Preparing Samples for AnalysisUniCel DxC Systems Instructions For Use A13914
Page 3-6October 2005
Reserved Racks for HbA1c, or IBCT
There are two separate fields for specific tests. One for HbA1c and one for IBCT. If a
rack number is entered into one of these fields, any samples run in that rack will only
function for that specific test. For example, Rack 66 is entered into the Reserved
Racks for HbA1c field. Any sample placed in that rack can only have an HbA1c run
on it. The reason for this is that each test in this group requires some sort of sample
preparation. If these tests are part of a group of other tests that do not require
preparation, they cannot be run at the same time as the others. Using these features
allows the prepared sample to be run in one of the reserved racks and allows the result
of the prepared sample to be merged with the original sample report when all tests are
complete.
Even if the Auto Serum Index feature is enabled, no serum indices will be run
for the pre-treated HbA1c and pre-treated IBCT samples run in these reserved
racks.
NOTICE
Sample Programming and Processing
Reserved Racks for HbA1c, or IBCT
3
The default for each of these features is:
• CTS = OFF,
• Reserved racks = none reserved,
• Obstruction detection = On,
• Reserved racks for HbA1c or IBCT = none reserved.
These features can only be modified when the system is in the Standby or Stopped
state.
UniCel DxC Systems Instructions For Use A13914Preparing Samples for Analysis
October 2005Page 3-7
How to Use Reserved Racks
When NOT to Use a Reserved Rack
How to Use Reserved Racks
When NOT to Use a Reserved Rack
When you do CTS (Closed Tube Sampling) do NOT run a closed tube in a reserved
rack.
Recommended Use of Reserved Racks
Green Racks
Series 350–400 to run:
• nesting cups
• capillary collection containers
• Beckman Coulter Microtubes™
Gray Racks
Series 601 to run:
• "Barcoded" Calibrators and Controls
Purple Racks
• Series 401–420 to run HbA1c
• 0.5 mL Cup Insert (P/N 476399)
Brown Racks
• Series 451–499 to run IBCT
Blue Racks
0.5 mL Cup Insert (P/N 476399)
To identify a reserved rack, put a red ® on the top of the rack.
NOTICE
Preparing Samples for AnalysisUniCel DxC Systems Instructions For Use A13914
Page 3-8October 2005
Recommended Use of Reserved Racks
Assigning or Reassigning Reserved Racks
StepAction
1Select the Setup icon from the menu bar.
2Select the <Page Down> button on the right side of the screen.
3Select <17> Reserved Racks/Obstruction Detection.
OR
Type 17 in the Option Number field and press [Enter].
The following screen appears.
How to Use Reserved Racks
3
E015928S.EPS
Figure 3.1
4Type the rack numbers to assign as reserved racks in the HbA1c and IBCT
fields.
AND/OR
Type the rack numbers to assign as reserved racks.
5Select <OK> to assign the racks.
OR
Select <Cancel> to exit without assigning the racks.
UniCel DxC Systems Instructions For Use A13914Preparing Samples for Analysis
October 2005Page 3-9
How to Use Reserved Racks
Bar Code Labeling
Bar Code Labeling
The use of bar code labels is a highly accurate and efficient method for identifying and
processing laboratory samples. However, the system must be able to identify and read
every bar code label to process each sample correctly. The following paragraphs
provide some basic information pertaining to bar code labels. Additional bar code
information can be found in the UniCel DxC Synchron Clinical Systems Reference Manual.
Bar Code Label Placement
Bar code labels must be applied to each sample tube in the correct location so that the
bar code reader can read the bar code. The following diagram (Figure 3.2) describes
how to place the label on a sample tube and how to place the tube into a sample rack.
CAUTION
A misread label can cause one sample ID to be read as another. The
laboratory’s process for printing, placing, and meeting all bar code
specifications is important to achieve highly accurate reading. Follow the
bar code label specifications to keep the rate of misread labels to a
minimum.
1
2
3
4
1. 14 mm (0.55 inch) Minimum
2. Label Placement Area
3. 20mm (0.78 inch) Minimum
4. 7.5 Degree Maximum
Figure 3.2Bar Code Label Placement
4
A014600L.EPS
Preparing Samples for AnalysisUniCel DxC Systems Instructions For Use A13914
Page 3-10October 2005
Processing Bar Coded Samples
Generally, previous sample programming will not need to be cleared before placing a
bar coded sample on the instrument. If the lab reuses Sample IDs, previous
programming must be cleared.
StepAction
1Load tubes in a rack with the bar code labels visible through the slot on the
same side of the rack as the rack bar code label.
2For closed tube sampling systems, check the top of the validated closed
tube for visible blood. If blood is present, clean the rubber stopper with a
cotton tipped applicator stick.
For non-CTS Systems or tubes not validated for closed tube sampling,
remove the cap.
3For a Routine Sample or if system is in Standby: Priority samples
should be placed on the autoloader first.
• Place the rack in the autoloader with the rack bar code label to the right
and
How to Use Reserved Racks
Processing Bar Coded Samples
3
•Press RUN.
For a STAT Sample: If the system is running and there are other racks on
the autoloader,
•Press PRIORITY. The rack pusher moves back one space so the STAT
rack can be placed in front of the other racks.
•Press RUN.
UniCel DxC Systems Instructions For Use A13914Preparing Samples for Analysis
October 2005Page 3-11
How to Use Reserved Racks
Processing Unreadable Bar Coded Samples
Processing Unreadable Bar Coded Samples
If a sample bar code is unreadable and the LIS needs to be queried, use manual
assignment. A manual assignment is rack and position and a Sample ID with no other
information. The DxC will query the LIS for tests and demographics.
If...Then...
Individual Samples,• Select the Samples icon from the menu bar,
Several samples in the same rack, • Select the Samples icon from the menu bar,
• Type rack and position,
• Type Sample ID,
• Save [F10].
• Select Rack [F8],
• Select the rack,
• Type Sample ID(s),
• Select <OK>.
Note: If previous programming appears, it may be necessary to clear it.
Processing Samples Manually
Manual programming is used when there is no LIS or the LIS is not available.
Samples may or may not be bar coded. The sample is identified by a Sample ID and/
or rack and position. The user selects tests for programming.
The following items require manual programming:
• samples without bar codes
• samples with bar codes and no sample programming
StepAction
1Select the Samples icon from the menu bar.
2Before programming, check for previous programming.
•Select Racks [F8].
• Select the rack number and select <OK>.
• If the rack has programming that is no longer required, select Clear [F1].
• Type the Sample IDs into the Sample ID(s) field.
•Select <OK>. A second screen will open. Select <OK> again to clear
the data.
3Select Program [F10].
(1 of 2)
Preparing Samples for AnalysisUniCel DxC Systems Instructions For Use A13914
Page 3-12October 2005
StepAction, continued
4Identify samples using the table below:
If…Then…
How to Use Reserved Racks
Processing Samples Manually
3
Sample has a readable bar
code,
• Type in the sample ID.
• If previous programming appears,
it may be necessary to clear it.
• Go to Step 2, or Step 5 if no
previous program appeared.
Sample has no bar code or bar
code can not be read,
• Type in the sample ID.
• Select the Rack field.
• Type in a rack number that is not
programmed.
• Go to Step 5.
5If the sample is stat, select the STAT check box.
6Select Sample Type and Sample Comment if provided.
7To enter Demographics, select Demog [F2]. Type in the information
provided and select Next [F10] to return to the Program Sample screen.
8Select each chemistry and/or select a panel. To access additional
chemistries, use the up and down arrows to move to other pages.
9Select Next [F10].
10To program additional samples, repeat Steps 2-8.
11Get the rack(s) listed on the screen and place the samples into assigned
rack positions as shown on the load list.
12For a Routine Sample or if system is in Standby: Priority samples
should be placed on the autoloader first.
• Place the rack in the autoloader with the rack bar code label to the right
and,
•Press RUN.
For a STAT Sample: If the system is running and there are other racks on
the autoloader,
•Press PRIORITY. The rack pusher moves back one space so the STAT
rack can be placed in front of the other racks.
•Press RUN.
(2 of 2)
UniCel DxC Systems Instructions For Use A13914Preparing Samples for Analysis
October 2005Page 3-13
How to Use Reserved Racks
Adding Tests or Rerunning a Sample
Adding Tests or Rerunning a Sample
StepAction
1Select the Samples icon from the menu bar.
2Select Rerun [F6].
3Type the sample ID, rack, or rack and position for rerun. One or more
samples can be programmed to rerun.
4Select one of the following buttons:
• To add or rerun specific tests press the <Select> button. All previous
programming will be highlighted. Deselect chemistries that are not to be
run.
• To rerun all completed tests press the <All> button. Rerun all completed
tests for all requested samples. Go to Step 6.
5Select Next [F10].
6For a Routine Sample or if system is in Standby: Priority samples
Clearing Samples
Once the test program has been processed, if all positions were used, the rack is not
available for programming again until it has been cleared. Only the programming
information is deleted; the results are stored and are accessible through the recall
function using the appropriate recall option.
Calibrator IDs or rack/position assignments are not cleared from within Sample
Programming. Refer to CHAPTER 5, Reagent Load/Calibration, "Calibrator
Assignment", for information on calibrator identification.
should be placed on the autoloader first.
• Place the rack in the autoloader with the rack bar code label to the right
and,
•Press RUN.
For a STAT Sample: If the system is running and there are other racks on
the autoloader,
•Press PRIORITY. The rack pusher moves back one space so the STAT
rack can be placed in front of the other racks.
•Press RUN.
NOTICE
When clearing data, to verify exactly what will be cleared, first view the load list
for the intended range of samples. The samples recalled on the load list are
identical to the samples that will be cleared.
Preparing Samples for AnalysisUniCel DxC Systems Instructions For Use A13914
Page 3-14October 2005
How to Use Reserved Racks
Range of Sample IDs
When to Clear Samples
Samples should be cleared as follows:
If...Then...
Sample programming has been completed.Clear sample programming.
A sample ID has to be reused.Clear that sample ID first.
3
A sample rack can not be programmed due to
previous programming. (This occurs when a
sample ID does not match, or when a rack number
shows previous information.)
Range of Sample IDs
The system retrieves a range of sample IDs in either Alphanumeric or Numeric form.
The system automatically selects one of the two, depending on the type of sample ID
entered. These same methods of retrieval are used for Load List, Clear Samples and
Recall.
Sample IDs containing only Numeric Data:
• Samples are listed in numeric order. For example 1, 2, 3, 4...
• Include leading zeros in the Range field. For example, range 1 thru 1000 would not
include sample 0099, but range 0001 thru 1000 would.
Sample IDs containing only Alpha or Alphanumeric Data:
• Samples are listed in dictionary order. For example A, AB, B...
• Numbers come before letters. For example 1, 900, 1A, A...
Clear that programming.
• Include all characters in the To field. For example, range A thru Z would not
include ZEBRA, but range A thru ZZZZZZZZZZZZZZ would.
NOTICE
Do not mix a pure numeric entry with an alphanumeric entry. For example, 0
thorough 1A will delete all numeric.
UniCel DxC Systems Instructions For Use A13914Preparing Samples for Analysis
October 2005Page 3-15
How to Use Reserved Racks
Procedure for Clearing Samples
Procedure for Clearing Samples
StepAction
1Select the Samples icon from the menu bar.
2Select Clear [F7].
3Sample may be cleared by Sample ID(s), Rack/Position, or by Date/Time
created. Type in the specific criteria for deletion.
Clearing Rack/Position does not clear programming for sample(s)
programmed in those positions with associated Sample IDs. Sample
must still be cleared by Sample ID. However, clearing by Sample ID
does clear rack/position in which the Sample ID was located.
NOTICE
Deletion CriteriaProcedure for Sample Clearing
Sample ID• Type in the Sample ID(s) to be
cleared. Single and multiple IDs
may be cleared. Use a comma (,)
to separate a list of IDs.
Range of Sample IDs
Refer to "Range of Sample
IDs" in this chapter for
additional information
• Type the first Sample ID in the
Range field.
• Type the last Sample ID in the
Thru field.
Time/Date Programs CreatedType the time and date range of
samples to clear.
Rack/Position(s)• Type Rack/Position(s) to be
cleared. Single and multiple racks
may be entered.
If only Rack is entered,
programming in all positions in
that rack will be cleared.
4Select <OK> to continue clearing. A confirmation screen will appear.
OR
Select <Cancel> to retain sample programs.
Preparing Samples for AnalysisUniCel DxC Systems Instructions For Use A13914
Page 3-16October 2005
Overview
Introduction
Overview
Introduction
CHAPTER 4 System Setup Options
This chapter summarizes the 29 System Setup options depicted on the Setup screens
shown in Figure 4.1 and Figure 4.2 below:
4
Figure 4.1Setup Screen (scrolled to the top)
E015930S.EPS
E015931S.EPS
Figure 4.2Setup Screen (scrolled to the bottom)
UniCel DxC Systems Instructions For Use A13914System Setup Options
October 2005Page 4-1
Overview
Introduction
For detailed step-by-step instructions on using the System Setup option, refer to the
UniCel DxC Synchron Clinical Systems Reference Manual.
System Setup OptionsUniCel DxC Systems Instructions For Use A13914
Page 4-2October 2005
Password Setup
Introduction
The Password Setup option allows the operator to:
• Assign up to 100 user names and their passwords
• Assign Administrator or Operator level privileges to each user name
• Define/edit or delete user name/password setup
• Enable/disable security for accessibility to certain system functions and setups
The following features may be password secured:
• Results, Edit
• Rgts/Cal, Modify Set Points
• Rgts/Cal, Slope/Offset Adjustment
Password Setup
Introduction
4
• Rgts/Cal, Within-Lot Calibration
• Rgts/Cal, Enzyme Validator
•QC
•Setup
• Utils, Clear Event Log
Defining/Editing Password Setup
To define, edit or delete user names, passwords, privilege levels and accessibility
levels, follow the steps below.
StepAction
1Select the Setup icon from the menu bar.
2Select <28> Password Setup from the Setup screen
3Type an Administrator password in the Administrator Password dialog
box.
Select <OK>.
Select <Cancel> to return to the System Setup menu.
OR
Note: After a software install, the default password is beckman.
(1 of 2)
UniCel DxC Systems Instructions For Use A13914System Setup Options
October 2005Page 4-3
Password Setup
Defining/Editing Password Setup
StepAction, continued
4The Password Setup dialog box shows Page 1 (of 10) of the users with
5In the Define/Edit Password dialog box, type the information requested.
password security. If necessary, select <Page Down> on the right side of
the screen to show the user desired.
Select New [F1] to enter the password setup for a new user.
OR
To edit the password setup for a user, select the user’s password number.
OR
Type the user’s password number in the Numbers field and press [Enter].
Select Define [F2].
NOTICE
Do NOT type beckman in the Password field.
Select Administrator in the Title field to allow the user to modify
password setup.
OR
Select Operator in the Title field to NOT allow the user to modify
password setup.
Select <OK>.
OR
Select <Cancel> to return to the Password Setup screen.
6The new or edited user is now in the Password Setup screen in alphabetical
order by last name.
(2 of 2)
Refer to the UniCel DxC Synchron Clinical Systems Reference Manual for additional
information on deleting users from the Password List, and enabling and disabling
password security for functions.
System Setup OptionsUniCel DxC Systems Instructions For Use A13914
Page 4-4October 2005
Auto Serum Index/ORDAC
Auto ORDAC
Auto ORDAC permits the enabling or disabling of the automatic Overrange Detection
and Correction (ORDAC) function for specified chemistries (refer to the Synchron
Clinical Systems Chemistry Information Manual for a list of chemistries offering
ORDAC). When a chemistry result exceeds the instrument printable range and Auto
ORDAC is enabled, the sample will automatically rerun with either:
• a smaller sample size or
• an on-line sample dilution (chemistry dependent).
When Auto ORDAC for Ig-A and Haptoglobin is enabled, the URDAC feature is also
enabled. Ig-A and Haptoglobin URDAC is used to analyze samples with
concentrations below the analytical range. In this case, the system takes a larger
sample volume.
Auto Serum Index/ORDAC
Auto ORDAC
4
(The manual ORDAC function in Sample Programming is used for samples which are
known to exceed the usable range. Chemistries designated with manual ORDAC at the
time of programming are run at the ORDAC sample volume.)
From the Setup Screen, select <1> Auto Serum Index/ORDAC to enable or disable
ORDAC. The default for Auto ORDAC is OFF.
The Automatic ORDAC screen may be viewed at any time, however the system must
be in Standby to modify the ORDAC selection.
NOTICE
The analytical ranges for each analyte are system limits found in the respective
CISs. These are the ranges that Beckman Coulter has verified can be achieved by
the system. There is no flagging associated with values exceeding these limits.
The instrument printable ranges for each analyte are internal system limits.
These ranges actually exceed the analytical ranges by a certain limit. This allows
for precision variations and still permits a result to print even though it exceeds
the analytical range slightly. Results outside this range will be suppressed. The
suppressed results will be flagged OIR HI or OIR LO (Out of Instrument Range).
UniCel DxC Systems Instructions For Use A13914System Setup Options
October 2005Page 4-5
Auto Serum Index/ORDAC
Auto Serum Index
Auto Serum Index
The Auto Serum Index function, when enabled, automatically analyzes every sample
for the detection of hemolysis, icerus and lipemia and numeric values (index) for the
relative concentrations (range) are included in the report.
The values are printed below the Special Calculations area of a patient report. The
values are printed with the test results for a control sample.
Auto Serum Index is intended for sample integrity assessment only; not for patient
diagnosis.
For additional information on Serum Indexing, refer to the Synchron Clinical Systems
Chemistry Information Manual.
NOTICE
When running the system with Automatic Serum Index enabled, remember to
manually disable the serum index feature when choosing options to set the
replicates. This prevents running indices needlessly.
From the Setup screen, select <1> Auto Serum Index/ORDAC to enable/disable
Auto Serum Index.
To enable or disable Auto Serum Index, the system must be in Standby or Stopped.
The default for Auto Serum Index is OFF.
System Setup OptionsUniCel DxC Systems Instructions For Use A13914
Page 4-6October 2005
Configuring the Chemistry Menu
Introduction
• The Chemistry Configuration function accommodates the installation of desired
tests from a comprehensive list of available chemistries.
• Of the total tests available, 180 chemistries can be selected and positioned onto the
menu at one time and customized to match the test order on the lab request forms.
• Modular chemistries (MC) are permanently configured on the system.
• These chemistries may be relocated on the configuration screen, but not removed.
• BUNm and UREAm both show (DxC 800 only) as available chemistries; only one
may be configured at a time. To replace one with the other, overwrite the existing
chemistry acronym with the replacement acronym.
• The Chemistry Configuration screen may be viewed at any time, but modifications
can only be made when the system status is Standby or Stopped.
Configuring the Chemistry Menu
Introduction
4
Configuring a Beckman Coulter Chemistry
The chemistry menu, available in sample programming, quality control, panel
definition and other screens, is defined by the user. To define Beckman Coulter
chemistries, select <2> Chemistry Configuration/Sample Type from the Setup
screen. Place the cursor in an open field in the Configuration Chemistry screen and
press Chems [F1]. Select the DxC chemistry from the pop-up list.
Configuring a User Defined Chemistry (UDR)
The chemistry menu available in sample programming, quality control, panel
definition and other screens is defined by the user. After a User Defined Chemistry
(UDR) has been defined by selecting <15> User-Defined Chemistries from the Setup
screen, you can Configure the UDR by selecting <2> Chemistry Configuration/Sample Type from the Setup screen. After the Chemistry Configuration/Sample Type
screen appears, selecting UDR [F2] presents a list of all User Defined Chemistries on
the system. Refer to CHAPTER 9, User Defined Reagents in this manual for additional
information on application of user-defined parameters.
Defining Chemistry Print Name
The Define Print Names function permits definition of the chemistry name that
appears on the chartable report. Each chemistry may be given a name of a maximum
of 15 characters. To define the chemistry print name, select <2> Chemistry Configuration/Sample Type from the Setup screen. Select Define [F3] from the
Configured Chemistry screen to use the Define Print Names screen.
UniCel DxC Systems Instructions For Use A13914System Setup Options
October 2005Page 4-7
Configuring the Chemistry Menu
Inserting a Chemistry
Inserting a Chemistry
The Insert function inserts a blank position for adding a chemistry to the menu. After a
blank position is inserted, the chemistries following that position will be incremented
by one position. If 180 is exceeded, the last one will be dropped. From the Setup
screen, select <2> Chemistry Configuration/Sample Type. Refer to UniCel DxC
Synchron Clinical Systems Reference Manual for additional information on this
procedure.
Deleting a Chemistry
The Delete Chemistry function requires that you first clear the calibration, sample
programming and control (QC) information from the instrument before deleting the
chemistry from the system. Where it is used, Beckman Coulter recommends that you
remove chemistries from the system following the sequence and procedures below to
prevent problems with deleting chemistries from the QC definition:
Control samples in the "sample required," "incomplete" or "rerun" status may
prevent the removal of the chemistry from the QC definition.
NOTICE
1. Remove the chemistry from Reagent/Calibration (Example: DIG)
StepAction
1Select the Rgts/Cal icon from the menu bar.
2Ensure the cal status of the chemistry DIG is not "Requested." To cancel a
calibration request:
• Select the position for the appropriate requested and assigned chemistry
to be canceled.
•Select No Cal [F8] to cancel the calibration request. Chemistries
currently selected may be canceled only if the calibration is not yet in
progress.
3Remove calibrator assignments. To clear calibrator bar code ID and/or rack
and position:
•Select Assign [F7].
•Select <Next> or <Prev> to locate the calibrator of interest.
OR
• Select the Calibrator Name options button <▼> at the top of the Assign
Barcode/Rack dialog box to view the List of Calibrators.
• Clear the fields by selecting the Calibrator ID, Rack and Position
fields for each level of calibrator and pressing the [Delete] key on the
keyboard.
•Select <Close> to exit and save information.
(1 of 2)
System Setup OptionsUniCel DxC Systems Instructions For Use A13914
Page 4-8October 2005
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.