Beckman Coulter UniCel DxC 600, UniCel DxC 600i, UniCel DxC 800 Instructions For Use Manual

Page 1

Instructions For Use

Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821

Volume 1

UniCel
®
DxC Synchron®
Clinical Systems
For In Vitro Diagnostic Use
This manual is intended for UniCel® DxC 600 UniCel® DxC 800 UniCel® DxC 600i
A13914AF April 2010
Page 2
Instructions For Use
Beckman Coulter Ireland, Inc. Mervue Business Park, Mervue Galway, Ireland 353 91 774068
Beckman Coulter do Brasil Com e Imp de Prod de Lab Ltda Estr dos Romeiros, 220 - Galpao G3 - Km 38.5 06501-001 - Sao Paulo - SP - Brasil CNPJ: 42.160.812/0001-44
製造販売元 : ベ ッ ク マ ン ・ コ ール タ ー株式会社 東京都江東区有明三丁目 5 番 7 号 TOC 有明ウエ ス ト タ ワー
贝克曼库尔特有限公司, 美国加利福尼亚州,Brea 市,S. Kraemer 大街 250 号, 邮编:92821 电话:(001) 714-993-5321
UniCel DxC Synchron Clinical Systems
PN A13914AF (April 2010)
Copyright © 2010 Beckman Coulter, Inc.
Trademarks
Following is a list of Beckman Coulter trademarks
AccuSense
Array
®
®
Microtube™
SPINCHRON™
Synchron
Synchron LX
UniCel
®
®
®
All other trademarks are the property of their respective owners.
Find us on the World Wide Web at:
www.beckmancoulter.com
Page 3
Initial Issue, A13914AA, 12/04
Software version 1.0
A13914AB, 10/05
Software version 1.4
A13914AC, 6/07
Software version 2.0
A13914AD, 2/08
Software version 3.0
A13914AE, 12/08
Software version 4.0
A13914AF, 4/2010
Software version 4.0
Changes:

Revision History

Safety Notice:
Added statement to Environmental Conditions Precautions
Added Patient Results
Added Quality Control
Added Sample Integrity
CHAPTER 3, System Setup Options:
Added footnotes to Table 3.3, UniCel DxC 600 Predefined Special Calculation Formulas and
Table 3.4, UniCel DxC 800 Predefined Special Calculation Formulas
CHAPTER 9, Maintenance:
Added Twice Weekly bullet to Maintenance Schedule
Added Twice Weekly Maintenance
Added bullets to Weekly Maintenance
Added Twice Weekly Maintenance procedure
Added cleaning statement to Weekly Maintenance
Added statement to Clean Flow Cell, Cups and CC Probes/Mixers (Automated)
Deleted CC Probes/Mixers Cleaning (Manual Program) procedure
A13914AF
iii
Page 4
Revision History
iv
A13914AF
Page 5
Summary of Hazards
A011460L.EPS
Introduction
This section summarizes the hazards associated with the DxC System. Individual hazards associated with a specific procedure in this manual are included in for that task. Please read this section and the following Summary of Precautions before operating the system.
Bar Code Reader Hazards
Do not tamper with or remove the housing of any bar code reader because of the laser-based nature of the readers and the potential hazard of looking directly at laser light. When the instrument is running, homing, or in diagnostics, the laser may be ON. At all other times the laser is OFF.

Safety Notice

Warnings or Cautions within the procedures
Biohazardous Materials Hazards
Observe all laboratory policies or procedures which pertain to handling of infectious and pathogenic materials.
Closed Tube Sampling (CTS) Cap Piercer Assembly Hazards/Biohazards
The Cap Piercer contains a razor sharp blade assembly that has been exposed to potentially biohazardous fluids. The points of the Blade are very sharp and extend below the Wash Tower. Stay away from the bottom of the Wash Tower. To prevent injury or exposure, do not touch the Points of the Blade and always wear gloves.
CTS Auto-Gloss Handling Hazards
Be careful when handling the bottle of CTS Auto-Gloss. Prevent spills. This lubricant is extremely slippery and difficult to clean from the floor.
Electrical Ground Hazards
Do not under any circumstances operate the system until an electrical ground is provided and the power cord is properly connected to the ground.
A13914AF
v
Page 6
Safety Notice
456161-B
TO REDUCE RISK OF PERSONAL INJURY,
OPERATE ONLY WITH ALL COVERS IN PLACE.
CAUTION
A011459L.EPS
Summary of Hazards
Electric Shock Hazards
Replacement or servicing of any components where contact with bare, live hazardous parts could occur, possibly resulting in electric shock, should only be performed by qualified service personnel.
Flammable Materials Hazards
Do not use this system in the presence of flammable materials.
Hazardous/Biohazardous Substances Hazards
When handling a spill of blood or other potentially hazardous substances, clean up the spill by using a 10% bleach solution, or use your laboratory decontamination solution. Then follow your laboratory procedure for disposal of hazardous materials. If the UniCel DxC system needs to be decontaminated, call your Beckman Coulter Service Representative for assistance.
ISE Module Hazards
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
Moving Parts Hazard
Do not place hands near any moving part while the system is operating. Lower and/or close protective guards and covers during operation.
No Foam Reagent Pressurization Hazard
The No Foam container is pressurized during system operation and must be properly depressurized prior to servicing to avoid sudden depressurization and potential exposure of the skin or eyes to the No Foam solution. To release the air pressure, disconnect the white quick connector located at the supply side of the container. If inhaled, move exposed individual to fresh air. If skin or eye contact with the solution occurs, flush the affected area thoroughly with water for at least 15 minutes. In both cases seek medical attention. Refer to the No Foam Material Safety Data Sheets (MSDS) for additional information.
vi
A13914AF
Page 7
Power Cord Hazards
Only use a three-pronged power cord to connect the instrument to a matching three-wire grounded outlet. Do not use an adapter to connect the power plug to a two-pronged outlet.
Service Procedures Hazards
Disconnect the power cord when performing service procedures such as replacing electronic or mechanical components.
Always wear appropriate personal protective equipment when handling reagents and other chemical preparations used with the system.
Sodium Azide Preservative Hazards
Reagents, calibrators and controls used with the system may contain small quantities (< 0.1%) of sodium azide preservative. Sodium azide preservative may form explosive compounds in metal drain lines. Refer to National Institute for Occupational Safety and Health Bulletin: Explosive Azide Hazards (8/18/76). Avoid skin contact with reagent by using personal protective equipment. If contact on skin occurs, use water to wash reagent from skin. Refer to the related Material Safety Data Sheets (MSDS) for additional information.
Safety Notice
Summary of Hazards
System Motors Hazards
To prevent possible injury, press the STOP button on the DxC System to disable the motors before attempting to clear any jams.
System Operations and Specifications Hazards
System operation should be consistent with the power requirements as stated in the Summary of
Hazards section of this chapter, and should always conform to the procedures and safety warnings
throughout this manual.
If the equipment is used in a manner not specified by Beckman Coulter, Inc., the protection provided by the equipment may be impaired.
Waste B Disposal Hazards/Biohazards
The Waste B collection bottle contents are considered biohazardous and should be handled appropriately.
A13914AF
vii
Page 8
Safety Notice
Summary of Precautions
Summary of Precautions
Introduction
This section summarizes the precautions that should be taken when operating the DxC System. Individual precautions associated with a specific procedure in this manual are included in boxes within the procedures for that task. Please read this section and the preceding Summary of
Hazards before operating the system.
AccuSense Glucose Sensor Precautions
Whenever the sensor is replaced, reaction cup and stir bar cleaning is recommended. Remove the sensor prior to removing stir bar. To prevent damage to the glucose oxygen sensor membrane tip, do not insert the stir bar removal tool, applicator stick, or any other object into the glucose reaction cup unless the sensor has been removed. Do not touch membrane tip of the AccuSense glucose oxygen sensor.
Caution
Air Filter Cleaning Precautions
Do not place a damp filter back on the system. Residual moisture may damage the system.
Alkaline Buffer Stability Precautions
The alkaline buffer reagent is stable for one month on the system. However, if a color change from pink to a lighter shade of pink should occur, replace the alkaline buffer with a fresh bottle of reagent.
Aqueous Calibrator Precautions
Repetitive refrigeration of Synchron aqueous calibrators may facilitate crystal formation. When removed from refrigerated storage, these calibrators should remain at room temperature. After the calibrator is opened it is stable for the period claimed in the accompanying package insert.
Bar Code Label Precautions
A misread label can cause one sample ID to be read as another. The laboratory’s process for printing, placing, and meeting all bar code specifications is important to achieve highly accurate readings. Always follow the bar code label specifications to avoid misread labels.
viii
A13914AF
Page 9
Beckman Coulter Microtube Precautions
Beckman Coulter Microtubes are designed for use on specific Synchron systems. Using the
appropriate Microtube is essential for proper system operation.
The sample height in the Microtube is critical for correct sample aspiration on all Synchron systems.
The use of Array Microtubes (PN 448163 or PN 448162) on Synchron Systems or the use of Synchron Microtubes (PN 756776) on Array systems may result in short sampling, incorrect results, and/or sample probe damage.
The use of non-Beckman Coulter, third party Microtubes, which have not been designed and tested on Synchron Systems may result in system damage and/or short sampling.
Biohazard Precautions
All biohazard precautions should be observed when doing maintenance, service, or troubleshooting on the system. Always wear appropriate personal protective equipment, and wash hands after working on contaminated portions of the system.
Safety Notice
Summary of Precautions
Blood Barrier Collection Tube Precautions
When blood collection tubes that contain physical barriers are used, extra care should be exercised to ensure that the barrier is tightly packed. Loose particles from the barrier could coat or plug the sample probe, flow cell, chemistry modules, electrolyte injection cup (EIC), or cuvette wash station.
BUNm/UREAm Electrode Precautions
To prevent damage to the BUNm/UREAm electrode, do not insert the stir bar removal tool, applicator stick, or any other object into the BUNm/UREAm cup unless the electrode has been removed. Remove the electrode before you remove the stir bar.
CC Subsystems Priming Precautions
If any two of the CC Subsystem items are selected, all three are primed automatically. Make sure that all three areas are ready to be primed.
CO2 Membrane Replacement Precautions
Do not touch the membrane surface when installing the quad-ring.
Covers, Doors and Shields Precautions
To ensure optimum performance of the system, operate the system with reagent doors and all shields and covers in place. To prevent possible motion errors, verify the proper positioning of any removed and reinstalled cover or shield.
A13914AF
ix
Page 10
Safety Notice
Summary of Precautions
CTS (Closed Tube Sampling) Cap Piercer Precautions
Use only validated sample containers with the CTS to avoid level sense errors.
CTS Tracking Loss Precautions
For systems with 1-Blade CTS, if there is an unusual loss of network communication, follow the instructions in the message that appears.
A normal shutdown or reboot does NOT give this CTS message.
When there is no CTS Tracking, if the cap were kept on a previously-pierced tube: it would be pierced again and pieces of rubber could fall into the sample.
A notification appears when full CTS Tracking is restored.
Diethylamine HCL and the Calcium ISE Precautions
Do not use controls containing diethylamine HCL. This adversely affects the calcium Ion-Selective Electrode (ISE).
Electrostatic Discharge (ESD) Precautions
To prevent damage due to electrical static discharge (ESD), always wear the wrist ground strap when directed to in a procedure.
Environmental Conditions Precautions
Changes in ambient temperatures and environmental conditions may result in a "reference drift" message. In this case, the electrolyte chemistries must be recalibrated.
Studies have shown that NA (sodium) recovery could drift as much as 0.8 mmol/L for each degree Celsius change in laboratory room temperature, from calibration to the time the sample is tested. Therefore, laboratory temperature fluctuations need to be minimized.
Fibrin Clots Precautions
Samples should be free of all visible fibrin. Clots could coat or plug the sample probes, flow cell, chemistry modules, electrolyte injection cup (EIC), or cuvette wash station leading to instrument malfunction and/or short sampling.
x
A13914AF
Page 11
ISE Reagent Precautions
Failure to operate the system with sufficient ISE reagent causes erroneous chemistry results. In some cases, results are obtained without reagents. Therefore, before starting a run, make sure that sufficient reagent is available to complete the run.
MC Reagent Stir Bar Precautions
When cleaning the MC Reagent Lines, Cups, and Stir Bars, the stir bar may rise in the cup due to air accumulation in the lines. Verify that the stir bar is positioned down into the bottom of the cup.
Motion Error Precautions
If the same motion error occurs repeatedly, refer to the instructions in the correct section of CHAPTER 5, Troubleshooting, of the UniCel DxC Synchron Clinical Systems Reference Manual or CHAPTER 12, Troubleshooting Calibration and Result Errors of this manual. If the motion error continues, contact your Beckman Coulter representative.
Safety Notice
Summary of Precautions
Narrow Margin Bar Code Precautions
The sample bar code reader on the DxC System can read narrow-margin bar codes. Because of the sensitivity needed to read narrow-margin bar codes, the labels must be high quality. They must be free from smudges, spots or other imperfections. An imperfection could be read as part of the Sample ID. This could cause an inaccurate read of the bar code.
Obstruction Detection and Correction (ODC) Precautions
Disabling ODC inactivates sample handling safeguards and may compromise sample integrity and cause erroneous results.
Patient Results
Patient results should be reviewed using delta checks. Please be aware of and question sequential abnormal results.
Printed Reports Precautions
Based on the system’s units/precision setup, a result may be printed as either “high” or “low” even though the value on the report is within the defined limits. This is due to the rounding of results. For example, a potassium result of 5.14 mmol/L may be printed as 5.1 mmol/L in the “High” Result column in a patient report when its reference range is defined as 3.5 to 5.1 mmol/L. The potassium result of 5.14 mmol/L is greater than the upper reference range limit of 5.10 mmol/L but the value printed is rounded to 5.1 mmol/L due to the units/precision set up on the system.
A13914AF
xi
Page 12
Safety Notice
Summary of Precautions
Quality Control
Laboratory Quality Control practices should be commensurate with laboratory operations. Refer to assay Instructions for Use and CLIA '88 (update 1/24/2004 Section 493.1256).
Racks Replacement Precautions
Racks should be replaced every five years. Damaged racks should not be used on the system or in the SPINCHRON Centrifuge.
Ratio Pump Quad-Ring Replacement Precautions
Be careful not to twist quad-rings or O-rings during installation, as this could result in reagent leakage or ratio pump failure.
Reagent Volume Precautions
Check reagent volumes before you start a run. Failure to operate with sufficient reagent causes erroneous chemistry results. In some cases, results are obtained without proper amounts of reagents in the modules. Therefore, before you start a run, make sure that sufficient reagent is available to complete the run.
Modular chemistry reagent containers should not be handled while the system is performing modular chemistry measurements.
The use of expired reagents may cause erroneous results.
Residual Blood Contamination Precautions
Before you place your validated closed tubes on the DxC System, check the top of the cap for any residual blood. Residual blood contamination into the sample could affect results. If blood is present, remove it by using a cotton-tipped applicator stick moistened with DI water. When running in the CTS mode, if tubes off-loaded from the UniCel DxC Systems have water or droplets of water on the caps, disable the CTS and the contact the Beckman Coulter Support Center.
NOTE
Oil on a cap is normal.
Reuse of Sample ID Precautions
xii
If your LIS or normal workflow requires the reuse of sample IDs, the sample programming should be cleared from the DxC at a time interval that is less than the shortest time of sample ID reuse. Failure to observe this warning causes new requests to be merged with tests from incomplete samples that previously used that ID.
A13914AF
Page 13
Sample Bar Code Reader Precautions
Do not tamper with or remove the housing of the Sample Bar Code Reader.
Sample Integrity
To assure good sample integrity, review pre-analytical sample procedures with your nursing, phlebotomy and laboratory staff. Follow your tube manufacturer's handling procedures. Contact the sample tube manufacturer for educational materials and training.
Sample Syringe Replacement Precautions
When installing the syringes, do not mix the two different syringe sizes. The MC and CC sample syringes (100 μL) are located on the back wall of the instrument. The CC reagent syringe (500 μL) is located on the right side of the instrument.
Safety Notice
Summary of Precautions
Samples NOT Received by Host Precautions
Do not clear samples until results are received at the host or printed. Clearing samples manually or through host programming, may cause results to be received at the host and printed at the DxC without the sample ID. Depending on the host implementation, this can cause lost sample results or sample results which merge with other sample results producing duplicate tests or added tests.
Sample Use Precautions
Do not use the same sample run on a DxC system for analysis of analytes for which a small quantity of carryover could greatly increase the results (for example, TBhCG).
If your system has a Closed Tube Sampling (CTS) option, place low volume samples in a nesting cup in a tube using a reserved rack to prevent motion errors.
System Backup Precautions
After inserting a diskette into the disk drive, selecting OK when prompted prepares the diskette by ERASING it before copying data. Be sure the diskette does not contain critical data that is not available from another source.
System Configuration Change Precautions
Changes to the System Configuration Data should only be done at the request or at the direction of Beckman Coulter, Inc. Entry of incorrect information leads to system errors.
A13914AF
xiii
Page 14
Safety Notice
Summary of Precautions
System Restore Precautions
System Parameter and Alignment data can be restored from the backup diskettes onto the system; however, performing the Restore function deletes some or all files (depending on the areas restored) from the hard drive.
Urine Sample Precautions
After analysis of ten consecutive urine electrolytes, run one replicate of electrolytes on Synchron Calibrator Level 2 in the serum mode. This minimizes the potential for chloride drift due to matrix effects of urine samples.
xiv
A13914AF
Page 15
Hardware Symbols and Labels
Introduction
This section briefly describes symbols and labels used on the DxC Systems. They are affixed to the appropriate components of the system.
Instrument Power Switch, ON
This symbol located on the main power switch indicates that the analyzer power is ON when this portion of the switch is in the down position.
Safety Notice
Hardware Symbols and Labels
Instrument Power Switch, OFF
This symbol, also located on the main power switch, indicates that the analyzer power is OFF when this side of the switch is in the down position.
Instrument or Printer Power Switch, ON
This symbol is located on the analyzer and printer power switch. When the portion of the switch with this symbol on it is in the down position, power to the monitor or printer is ON.
A13914AF
xv
Page 16
Safety Notice
Hardware Symbols and Labels
Monitor Switch, ON/OFF
This symbol is located on the monitor power switch. A green light to the left of this symbol indicates the power is ON.
CPU Power OFF Switch
This symbol is located on the face of the Computer (CPU) unit and indicates the OFF state when pressed.
Primary Electrical Ground
This symbol is used to indicate an electrical ground.
Keyboard Connection
This symbol is found above the connection between the computer and the keyboard.
xvi
A13914AF
Page 17
Can Hold This Object Here
This black symbol, located on the bottom of each sample and reagent probe assembly, indicates that this area may be handled to rotate the probe.
Do Not Hold This Object Here
This red symbol, located on the top of each sample and reagent probe assembly, indicates that this area may not be handled.
Safety Notice
Hardware Symbols and Labels
Mouse Port Connection
This symbol is found next to the connection between the computer and the mouse port.
High Voltage Electric Shock Risk
This symbol indicates high voltage is present and /or there is a risk of electric shock when working in this area.
A13914AF
xvii
Page 18
Safety Notice
456161-B
TO REDUCE RISK OF PERSONAL INJURY,
OPERATE ONLY WITH ALL COVERS IN PLACE.
CAUTION
A011459L.EPS
A011460L.EPS
Hardware Symbols and Labels
CAUTION
This symbol indicates a caution message and is followed by an explanation or other symbols that define the caution (see examples below).
CAUTION Operate with All Covers in Place
This symbol is located on top of the work surface cover and the cover of an optional Cap Piercer. It indicates a caution to operate only with all covers in place to reduce risk of personal injury or biohazard.
General Biohazard
These caution symbols indicate biohazardous risk from possible patient specimen contamination.
Laser Bar Code Caution
A label reading, "CAUTION. LASER LIGHT ACCESSIBLE. WHEN COVER IS OPEN OR REMOVED, DO NOT STARE INTO BEAM." is placed on the cover of any laser-based code reader. Do not stare into laser light beam when cover is open or removed.
xviii
A13914AF
Page 19
Class II Laser Caution Warning
A012936L.EPS
CAUTION
SHARP OBJECTS - REFER
SERVICING AND MAINTENANCE TO
QUALIFIED SERVICE PERSONNEL.
A label reading, "CAUTION. LASER LIGHT - DO NOT STARE INTO BEAM. 670 nm - 1mW CLASS II LASER PRODUCT." is placed near any opening through which a bar code reading beam is emitted. Do not stare into laser light beam.
Laser
A label reading, "AVOID EXPOSURE. LASER LIGHT IS EMITTED FROM THIS APERTURE." is placed near any opening through which a bar code reading beam emits. Avoid exposure to laser light emitted.
Safety Notice
Hardware Symbols and Labels
Sharp Objects Caution
A label reading, "CAUTION. SHARP OBJECTS - REFER SERVICING AND MAINTENANCE TO QUALIFIED SERVICE PERSONNEL." is found on top of the optional cap piercing hardware which is located underneath the removable cover of the cap piercing tower.
A13914AF
xix
Page 20
Safety Notice
A012937L.EPS
THE ISE COVER SHOULD REMAIN IN PLACE DURING SYSTEM OPERATION.
471830-AA
CAUTION PARTS MOV E AUTOMATICALLY
A015047L.EPS
A012938L.EPS
THIS DOOR SHOULD REMAIN CLOSED DURING SYSTEM OPERATION.
471831-AA
Hardware Symbols and Labels
ISE Cover Caution
A label reading, "THE ISE COVER SHOULD REMAIN IN PLACE DURING SYSTEM OPERATION." is placed on top of the ISE module frame under the ISE cover to indicate that the ISE cover should remain in place during system operation.
Moving Parts Caution
A label reading, "CAUTION PARTS MOVE AUTOMATICALLY" is placed inside the offload track, on the left side of the back wall.
MC Door Caution
A label reading, "THIS DOOR SHOULD REMAIN CLOSED DURING SYSTEM OPERATION." is found on the top edge of the left hand (MC reagent) door and indicates that the door should remain closed during system operation.
xx
A13914AF
Page 21
Read Manual Caution
A012939L.EPS
270-455774-A
CAUTION - READ MANUAL
BEFORE OPERATING
ATTENTION - CONSULTER LA NOTICE
AVANT DE FAIRE FONCTIONNER.
A011540L.EPS
BECKMAN COULTER, INC
MADE IN U.S.A. MARCA REG
PRODUCT COMPLIES WITH 21 CFR CHAPTER I, SUBCHAPTER J
MANUFACTURED DECEMBER 2004
LABEL P/N 448229 AB
S
A label reading, "CAUTION - READ MANUAL BEFORE OPERATING." is found on the hydropneumatics behind the center door and recommends the operator read the manuals before operating the system.
Rack Loading Label
This label is found to the right of the Run and Priority Load buttons and indicates the correct position of a rack for placement into the autoloader.
Safety Notice
Hardware Symbols and Labels
Laser Certification Label
This label is found on the back, bottom edge of the system. It provides information about the laser.
A13914AF
xxi
Page 22
Safety Notice
A012942L.EPS
ETHERNET
SERIAL PORT
A012943L.EPS
WASTE B OUTLET
WASTE OUTLET
WASTE B
SENSOR
D. I. WATER INLET
MAX. PRESS. 100 PSI (689 kPa)
136
137128
D.I. FIBER
Hardware Symbols and Labels
Ethernet/Serial Port Label
This label is found on the right side of the system and identifies connections for the Ethernet and serial ports.
Fluid Interface Label
This label is found on the center, back, bottom edge of the system. It identifies inlet and outlet ports on the system.
xxii
A13914AF
Page 23
Recycling Label
A016608L.EPS
This symbol is required in accordance with the Waste Electrical and Electronic Equipment (WEEE) Directive of the European Union. The presence of this marking on the product indicates:
1. the device was put on the European Market after August 13, 2005 and
2. the device is not to be disposed of via the municipal waste collection system of any member
state of the European Union.
It is very important that customers understand and follow all laws regarding the proper decontamination and safe disposal of electrical equipment. For Beckman Coulter products that have this label please contact your dealer or local Beckman Coulter office for details on the take back program that will facilitate the proper collection, treatment, recovery, recycling and safe disposal of device.
Safety Notice
Hardware Symbols and Labels
A13914AF
xxiii
Page 24
Safety Notice
Hardware Symbols and Labels
Restriction of Hazardous Substances (RoHS) Labels
These labels and materials declaration table (the Table of Hazardous Susbtance's Name and Concentration) are to meet People's Republic of China Electronic Industry Standard SJ/T11364-2006 "Marking for Control of Pollution Caused by Electronic Information Products" requirements
RoHS Caution Label
This logo indicates that this electronic information product contains certain toxic or hazardous elements, and can be used safely during its environmental protection use period. The number in the middle of the logo indicates the environmental protection use period for the product. The outer circle indicates that the product can be recycled. The logo also signifies that the product should be recycled immediately after its environmental protection use period has expired. The date on the label indicates the date of manufacture.
RoHS Environmental Label
This logo indicates that the product does not contain any toxic or hazardous substances or elements. The "e" stands for electrical, electronic and environmental electronic information products. This logo indicates that this electronic information product does not contain any toxic or hazardous substances or elements, and is green and is environmental. The outer circle indicates that the product can be recycled. The logo also signifies that the product can be recycled after being
discarded, and should not be casually discarded.
xxiv
A13914AF
Page 25
Documentation Symbols
Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate instrument. Do not attempt to perform any procedure before carefully reading all instructions. Always follow product labeling and manufacturer’s recommendations. If in doubt as to how to proceed in any situation, contact your Beckman Coulter representative.
Alerts for Warning, Caution, Important, and Note
WARNING
WARNING indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. May be used to indicate the possibility of erroneous data that could result in an incorrect diagnosis (does not apply to all products).
Safety Notice
Documentation Symbols
CAUTION
CAUTION indicates a potentially hazardous situation, which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. May be used to indicate the possibility of erroneous data that could result in an incorrect diagnosis (does not apply to all products).
IMPORTANT
Following the advice in the Important Notice adds benefit to the performance of a piece of equipment or to a process.
NOTE NOTE is used to call attention to notable information that should be followed during installation, use,
or servicing of this equipment.
IMPORTANT is used for comments that add value to the step or procedure being performed.
A13914AF
xxv
Page 26
Safety Notice
Documentation Symbols
xxvi
A13914AF
Page 27
Revision History, iii
Safety Notice, v
Introduction, xxxiii
CHAPTER 1: System Description, 1-1
System Description, 1-1

Contents

Operational Conditions, 1-1
System Components, 1-5
Sample Handling System, 1-6
Modular Chemistry (MC) System, 1-12
Cartridge Chemistry (CC) Reagent Handling System, 1-17
Cuvette Reaction System, 1-21
Hydropneumatic System, 1-24
Operation and Control Components, 1-26
Main Screen and Program Structure, 1-27
Theory of Operation, 1-43
Cartridge Chemistry: Calibration Theory, 1-43
Modular Chemistry: Calibration Theory, 1-53
Cartridge Chemistry: Principles of Measurement, 1-54
CHAPTER 2: Preparing Samples for Analysis, 2-1
Routine Operation Overview, 2-1
Preparing Samples for Analysis, 2-2
How to Use Reserved Racks, 2-7
xxvii
Page 28
Contents
CHAPTER 3: System Setup Options, 3-1
Overview, 3-1
Password Setup, 3-2
Auto Serum Index/ORDAC, 3-4
Configuring the Chemistry Menu, 3-5
Setting the Default Sample Type, 3-13
Date/Time Setup, 3-13
Demographics Setup, 3-14
Patient Results – Immediate Reporting Setup, 3-14
Panels, 3-15
Replicates, 3-16
Report Setup, 3-16
Reportable Ranges Setup, 3-17
Reference/Critical Ranges Setup, 3-19
Sample Comments Setup, 3-19
Special Calculations Definition, 3-20
Timed Urine and Creatinine Clearance Results, 3-23
Version Information, 3-23
Units/Precision Setup, 3-24
User Defined Chemistries Setup, 3-24
Bar Code Setup, 3-25
Maximum Sample Program Age, 3-26
Reserved Racks/Obstruct Detect, 3-26
Disable Service Monitor, 3-27
Host Communications, 3-28
Language/Keyboard Setup, 3-29
Printer Setup, 3-30
Service Setup, 3-30
xxviii
System Configuration, 3-31
Version Upgrade, 3-31
Status Alarm/Annunciator, 3-32
Chemistry Update, 3-32
Auto Generation of Control, 3-33
Page 29
CHAPTER 4: Reagent Load/Calibration, 4-1
Reagent Load, 4-1
System Calibration, 4-12
Load a Calibrator Diskette, 4-12
Calibrator Assignment, 4-13
Calibration Status, 4-14
Reagent and Calibration Status Warnings, 4-15
Request a Calibration, 4-16
Calibration Failure Messages, 4-18
Within-Lot Calibration, 4-19
Enzyme Validator, 4-23
Calibration Override, 4-24
Contents
Chemistry Bypass, 4-25
Extend Calibration Time, 4-26
Calibration Acceptance Limits, 4-27
Calibrator Set Point Modifications, 4-28
Slope Offset Adjustment, 4-29
Reprint Calibration Reports, 4-31
CHAPTER 5: Quality Control, 5-1
Quality Control, 5-1
Define a Control, 5-4
Control ID Assignments, 5-7
Run Control Samples, 5-7
Edit a Control Definition, 5-9
Review a Control Definition, 5-12
Delete a Control, 5-12
Print QC Ranges, 5-13
QC File List, 5-14
QC Summary, 5-15
QC Chart (Levey-Jennings), 5-16
QC Log, 5-18
Quality Assurance Program (QAP) “Copy To Disk” Feature, 5-21
Archive QC, 5-23
xxix
Page 30
Contents
Review Archived Data, 5-24
CHAPTER 6: Sample Programming and Processing, 6-1
Overview, 6-1
Prior to Programming, 6-2
Identify Samples, 6-4
Sample Programming and Processing, 6-6
Additional Programming Information, 6-10
Clear Samples, 6-12
CHAPTER 7: Results Recall, 7-1
Overview, 7-1
Recall Results by Sample ID, 7-2
Recall Results by Rack and Position, 7-3
Recall Results by Patient ID, 7-4
Recall Results by Run Date/Time, 7-4
Display Recalled Results, 7-5
Edit Critical Rerun Result, 7-6
Print Recalled Results, 7-7
Send Results to the Host, 7-8
Absorbance Versus Time, 7-8
Statistical Summary Report, 7-10
CHAPTER 8: User Defined Reagents, 8-1
Overview, 8-1
Requirements and Precautions, 8-1
User-Defined Reagent Setup, 8-3
Chemistry Parameters, 8-4
Processing Parameters, 8-8
Error Detection Limits, 8-13
xxx
Wavelength Selection, 8-15
Determination of Extinction Coefficients, 8-16
Exit Check Criteria, 8-17
User Defined Reagent Removal, 8-19
Expanded User Defined Chemistry Feature, 8-20
Page 31
CHAPTER 9: Maintenance, 9-1
Overview, 9-1
Electronic Maintenance Log, 9-4
Twice Weekly Maintenance, 9-7
Weekly Maintenance, 9-9
Check Chloride Calibration Span, 9-17
Monthly Maintenance, 9-17
Two-Month Maintenance, 9-36
Three-Month Maintenance, 9-43
Four-Month Maintenance, 9-48
Six-Month Maintenance, 9-52
As-Needed/As-Required Maintenance, 9-64
Contents
CHAPTER 10: System Status and Commands, 10-1
Overview, 10-1
System Status, 10-1
Status-Cycle Count, 10-2
Temperatures, 10-3
Show Temperature Status, 10-3
Power Subsystems, 10-4
Hydropneumatics Subsystem, 10-4
ICS/Smart Modules, 10-5
Cuvette Water Blank Status, 10-6
CTS Tracking, 10-7
Instrument Commands, 10-8
Home, 10-8
Pause, 10-9
Stop Print, 10-11
Shutdown, 10-12
System Power On/Boot, 10-14
Pause/Resume Waste B, 10-15
Enable/Disable Modules, 10-16
Unload All, 10-18
xxxi
Page 32
Contents
CHAPTER 11: Utilities, 11-1
Overview, 11-1
Prime, 11-1
Maintenance, 11-6
Event Log, 11-6
Alignment/Diagnostics/PVTs, 11-12
Metering, 11-12
Modem, 11-12
Backup/Restore, 11-13
Touch Screen Calibration, 11-16
CHAPTER 12: Troubleshooting Calibration and Result Errors, 12-1
Calibration Errors, 12-1
MC Calibration, 12-2
Linear Calibration, 12-7
Non-Linear and Multipoint Calibrations, 12-10
Troubleshooting Result Errors, 12-13
Error Codes and Definitions, 12-14
Error Code – Definitions, 12-17
Common Error Messages and Corrective Actions, 12-1
Glossary, Glossary-1
Index, Index-1
Related Documents
xxxii
Page 33
Intended Use
The UniCel DxC Synchron Clinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).
Scope of This Manual

Introduction

This manual covers basic operating instructions and maintenance guidelines for UniCel DxC 600/ 800 Systems. Detailed operation, maintenance, and troubleshooting instructions are not included in this manual. In addition, medical and diagnostic interpretation, or the clinical significance of chemistries or assays are not discussed. Refer to the reference materials in the Related Documents section for detailed information.
Manual Conventions
This manual uses the following printed and visual cues to guide the user in responding to printed directions.
Table 1 Conventions Used in this Manual
Convention Description
Combination Keys Keyboard characters that when pressed, in conjunction with the
Command buttons (buttons with names on a screen)
Control or Alt key, invoke a command. They are enclosed in + sign between each key.
Examples:
+
Alt
(
Buttons with names are bold and use a SansSerif font.
Examples:
Select OK. Select
)
or
(X)
Cancel.
+
(
Ctrl
+
Alt
)
Delete
(
)
(
)
(
)
with a
A13914AF
xxxiii
Page 34
Introduction
Manual Conventions
Table 1 Conventions Used in this Manual (Continued)
Function buttons Function buttons are bold and use a SansSerif font.
Convention Description
Example:
Select Print F10.
Icon buttons Icon buttons are bold and use a SansSerif font.
Example:
Select Samples from the menu bar.
Instrument buttons Buttons on an instrument are bold and use a SansSerif font. They may
be all upper case or initial caps.
Example:
Press the STOP button.
Keyboard keys
Keyboard keys are enclosed in
Examples:
(X), (→), (
Ta b
), (
Enter
)
.
(
)
Pull-down menus Use the pull-down menu to see a list of options you can select.
Example:
Select the Calibrator Name pull-down menu at the top of the Assign
Barcode/Rack
dialog box to view the list of calibrators.
Text field Names in the text fields are bold and use a SansSerif font, followed by
the word "field".
Example:
Type the Patient ID in the Patient ID field.
xxxiv
A13914AF
Page 35
How to Use this Manual
Manual Format
Information in this manual is presented in modular units. Each unit of information is described by a brief title in the left margin.
Many units consist of a numbered list which presents a procedure, process, or description.
Procedure Lists
Procedure lists are the most common type of lists in this manual. Each step of a procedure is listed by number with the corresponding action that is to be performed.
Occasionally, a decision must be made at a step within a procedure. An indented decision list is then presented which describes the variable conditions in the first bullet and the appropriate action for each condition in the subordinate bullets.
Introduction
How to Use this Manual
Example of a Procedure
The following is an example of a procedure that contains a decision list.
1
Select Rerun F6.
To enter individual Sample IDs, Type the Sample IDs for rerun in the
To enter a range of Sample IDs, Type the Sample ID at the beginning of the range in the
2
Select a button from the bottom of the dialog box.
Read the decision list as a complete sentence, using the first bullet to introduce the condition and the second level bullet to introduce the action.
Sample IDs field.
Range field.
A13914AF
xxxv
Page 36
Introduction
How to Use this Manual
xxxvi
A13914AF
Page 37

System Description

Introduction
This chapter describes the system components, operational theories, principles of measurement, programming structure, and operator controls. Detailed information is located in the UniCel DxC Synchron Clinical Systems Reference Manual.

CHAPTER 1

System Description

Operational Conditions

Shipping Damage
Each DxC System is carefully examined and checked by Beckman Coulter, Inc. before it is shipped. When you receive your new DxC System, visually inspect the shipping container for damage. If there is damage, notify the Beckman Coulter Service representative before he or she arrives at your facility to install your system.
Installation
Table 1.1 Installation Requirements
Item Requirement
Installed by The DxC system is installed completely by Beckman Coulter
Installation Category II
A13914AF
1-1
Page 38
System Description
Operational Conditions
Clearances
Table 1.2 System Clearances
Area Affected Clearance Needed
Left Side Minimum of 6 inches (15.2 cm) clearance or 12 inches (30.5 cm) to access smart
Right Side Minimum of 18 inches (45.7 cm) clearance when monitor on swing arm is in use.
Back Zero inches. The venting design of the system allows for "0 inch" clearance at the
Top Minimum of 22 inches (55.9 cm) from highest point of system.
Front Minimum of 25 inches (63.5 cm) to open doors.
Sunlight and Drafts
Do not place the system in direct sunlight or in drafts. Both of these conditions may affect the temperature control of the system.
modules.
back.
Drain
The system should be located near a sink or floor drain to accommodate the waste effluent at a minimum rate of 6 liters/hour (16 liters/hour continuous flow).
The drain must not be placed any higher than 36 inches (91.4 cm) above the floor.
Power Requirements
Table 1.3 Power Requirements – DxC 600 or 800 Analytic Unit
Operating range 200–240 VAC ± 10% (180–264 VAC)
Frequency 50/60 Hz
BTU generated 10,500 BTU/hour
Power connector 20 A current rating, NEMA L6-20R twistlock in-line connector
Item Requirement
14 A at low line, exclusive of power on surge
1-2
A13914AF
Page 39
Table 1.4 Power Requirements – DxC Console (PC System and Monitor)
Item Requirement
Operating range 100–120 VAC ± 10% (90–132 VAC); 4A
OR 200–240 VAC ± 10% (180–264 VAC); 2A
Frequency 50/60 Hz
BTU generated 1,500 BTU/hour
Power connector 15 A current rating, IEC 320 standard connector
Table 1.5 Power Requirements – Okidata B4350 LED Printer
Item Requirement
Operating range 110–127 VAC ± 10% (99–140 VAC); 3A
OR 220–240 VAC ± 10% (198–264 VAC); 1.6A
Frequency 50/60 Hz
System Description
Operational Conditions
1
BTU generated 1228 BTU/hour (printing); 34 BTU/hour (standby power save)
Power connector 15 A current rating, IEC 320 standard connector
Notes on the System Power
The system can operate from any standard 3-wire electrical outlet and is wired as shipped from the factory to operate on 220 VAC, 50/60 Hz.
IMPORTANT
protection of the electronic circuitry.
Line Voltage from the electrical outlet should be free of spikes, fluctuations, and dropouts for
CAUTION
Only operate the system from a 3-wire power source. DO NOT use a 2-prong adapter or a 2-wire AC power source.
Environmental Conditions
Table 1.6 Temperature, Humidity and Elevation
Item Specification
Environment Indoor use only
Ambient temperature +18°C to +32°C
A13914AF
Warm-up time 30 minutes (time to reach operating temperature)
Relative humidity 20–85% relative, non-condensing
Elevation Up to 4,200 ft (1280 m)
1-3
Page 40
System Description
Operational Conditions
Water Requirements
Table 1.7 Water Requirements
Item Specification
Flow Rate 0.6 L/min peak flow rate, 16 L/hr minimum continuous flow rate
Temperature +15°C to +25°C
Water pressure Deionized water entering the system must be 30–90 psi.
Table 1.8 Water Quality Requirements
CLSI (CLRW) 4TH Ed. C03 --A4
Formerly NCCLS (Type I & II)
Notes Replaces Type I & II.
4th Ed. C03 --A4
CLSI
CLRW
Organic Impurities
Total Organic Carbon (TOC)
Microbiological Impurities
Maximum microbial content colony forming unit (CFU/mL)
Ionic Impurities
Minimum resistivity, megohm.centimeter (MΩ. Cm@25C)
Particulate & Colloid Content Purification process
pH Not Applicable Not Appiicable
Maximum silicate (mg/mL) SiO
500 ng/g TOC (Total Organic Carbon) parts per billion (ppb)
10 CFU/mL 10 CFU/mL
10 MΩ. Cm 1.0 MΩ.Cm
requirement only: water filtration using 0.22 μm pore size to remove microorganisms and particulates
Not Applicable Not Applicable
2
Beckman Coulter Requirements
Not Applicable
Purification process requirement only: water filtration using 0.22 μm pore size to remove microorganisms and particulates
Other System Specifications for IEC-1010 Compliance
Table 1.9 IEC-1010 Specifications
Item Specification
Pollution Degree 2
EN55011 Meets Class A
Maximum Sound Pressure ≤ 65 dBA average over 8 hours with covers down at 1 meter away from
the instrument at +25°C
1-4
A13914AF
Page 41
Table 1.9 IEC-1010 Specifications
Item Specification
Maximum Leakage Current DxC 600: 222 μA at 240V, 50Hz

System Components

DxC Systems
A UniCel DxC System can be divided into the following components:
Sample Handling Components
Modular Chemistry System
DxC 800: 240 μA at 240V, 50Hz
System Description
System Components
1
Cartridge Chemistry Reagent Handling System
Hydropneumatic System
Operation and Control Components
A Closed Tube Sampling System has optional components listed below:
Large Particle Immuno Assay Module (LPIA)
Closed Tube Sampling (CTS)
*
A13914AF
* Equivalent to Near Infrared Particle Immuno Assay (NIPIA).
1-5
Page 42
System Description
A011869P.EPS
4
3
1
2
5

Sample Handling System

Figure 1.1 UniCel DxC 600/800 Analyzer (600 shown)
1. Modular Chemistry (MC) Section
2. Cartridge Chemistry (CC) Portion
3. Autoloader
Sample Handling System
Introduction
The Sample Handling system is composed of the following components:
Sample racks
Autoloader/Offload track
Shuttle
Bar code reader
Cap piercer assembly (optional)
Sample Carousel
Sample probe/mixer assemblies
4. Dual Reagent Carousel
5. Operator Console
The sample handling module is used to load samples onto the system, provide samples for analysis, and provide temporary storage of completed samples. A detailed description of each component is presented in the following paragraphs.
1-6
A13914AF
Page 43
Sample Racks
A sample rack is a high -strengt h, plastic , centrifugable ho lder desig ned to h ouse up to f our sa mples. There are four sizes of racks with each size capable of holding one length/width combination of primary sample tubes in addition to accepting sample cups. Refer to Figure 1.2.
Sample racks accept the following tube and cup sizes:
System Description
Sample Handling System
1
IMPORTANT
racks. These adaptors must only be used in racks designated as reserved. The reserved rack feature is described in CHAPTER 3, System Setup Options.
Table 1.10 Sample Racks
13 × 75 12 × 75 mm tubes
16 × 75 16 × 75 mm tubes
13 × 100 13 × 100 mm tubes
16 × 100 16 × 100 mm tubes
Adapters are provided to adapt various sized sample tubes (secondary tubes) to the short
Rack Accepts These Cups and Tubes
13 × 75 mm tubes
0.5 mL cups
2.0 mL cups
0.5 mL cups
2.0 mL cups
Capillary collection tubes (use with the capillary tube adapter)
Beckman Coulter Microtubes
16 × 92 mm tubes
A13914AF
16.5 × 100 mm tubes
Beckman Coulter 0.5 mL Cup Insert (PN 467406)
1-7
Page 44
System Description
Sample Handling System
Rack ID Labels
Sheets of bar-coded rack ID labels are supplied with the system. They can be applied as shown in
Figure 1.2.
Figure 1.2 Rack
1. Numeric Rack ID Number
2. Rack Size Label
3. Bar Coded Rack ID Label
1-8
A13914AF
Page 45
Autoloader/Offload Track
A011870P.EPS
2 3 764 5
1
When viewed from the front of the system, the autoloader is on the left and holds up to 25 sample racks in preparation for presentation to the DxC 800 system. The DxC 600 system has room to load a maximum of 14 racks. Refer to Figure 1.3.
There is also space for 25 sample racks in the offload track as they are removed from the Sample Carousel upon completion.
System Description
Sample Handling System
1
IMPORTANT
autoloader.
Figure 1.3 Sample Loading Area
1. Pushers
2. Run Button
3. Priority Load Button
4. Autoloader
When loading racks onto the autoloader, make sure that they are placed firmly down into the
5. Shuttle
6. Offload Track
7. Stop Button
Priority Load Position
Between the Autoloader and the Sample Gate is the Priority Load position. This position is used in conjunction with the PRIORITY LOAD button when a rack is to be loaded onto the Sample Carousel into a reserved priority position so that it can be run in a higher priority than other racks on the Autoloader. Refer to Figure 1.3.
A13914AF
1-9
Page 46
System Description
Sample Handling System
Priority Load Button
Typically, rack placement and removal is under microprocessor control. The operator may use the reserved positions in the sample carousel by pressing the PRIORITY LOAD button and placing the priority rack in the space provided by the system. The rack loads into one of the reserved positions on the Sample Carousel.
PRIORITY LOAD only prioritizes the loading of the rack. It does not alter the sample priority (STAT or routine) previously designated in Sample Programming.
Pushers
Pushers collect and move to the Sample Gate any racks loaded onto the system. They are activated when the RUN button is pressed. Refer to Figure 1.3.
Sample Gate
The Sample Gate is the mechanism that moves racks from the load tray to the shuttle during the load process. It also moves samples from the shuttle to the unload track during the unload process.
Shuttle
The Shuttle moves the rack from the gate area onto the Sample Carousel. Refer to Figure 1.3.
Bar Code Reader (Sample)
CAUTION
Do not tamper with or remove the housing of the Sample Bar Code Reader.
The Bar Code Reader is a Class II fixed-beam laser scanner. It is used to read the rack bar code, the sample bar code (if present), and the background bar codes as the rack travels past. The rack bar code and sample bar code (if present) are used to identify the sample and link it to the appropriate sample programming.
There are two background bar codes that are used to determine whether a rack position is empty or occupied, and if occupied, whether the sample is in a cup or tube.
Refer to Documentation Symbols in the Safety Notice section, for a description of the CAUTION labels for the bar code reader.
1-10
A13914AF
Page 47
System Description
Sample Handling System
1-Blade Thick CTS (Closed Tube Sampling) or 1-Blade Narrow CTS Cap Piercer Assembly (optional)
CAUTION
This Cap Piercer contains a razor sharp blade assembly.
CAUTION
To avoid damage to the blade, do NOT use this Cap Piercer assembly with foil-capped tubes.
This optional 1-Blade Thick CTS or 1-Blade Narrow CTS Cap Piercer assembly pierces capped tubes allowing the sample probe access to the sample without the need for cap removal. Tubes with caps must be loaded in racks that have had the size correctly defined and that are not Reserved (Reserved Racks = No CTS). Both open and closed tubes can be in the same sample tube rack.
Table 1.11 Tubes Validated for Closed Tube Sampling
1
Cap Piercer Configuration Tube Type Tube Size
1-Blade Thick CTS Becton Dickinson VACUTAINER with
HEMOGARD
Greiner VACUETTE 13 × 75 mm
1-Blade Narrow CTS Sarstedt S-Monovette
a. This tube requires a special 5.5 mL rack (PN A18642).
13 × 75 mm 13 × 100 mm 16 × 100 mm
13 × 100 mm
75 × 15 mm
92 × 15 mm
a
Refer to Documentation Symbols in the Safety Notice section, for a description of the CAUTION labels for the Cap Piercer.
A13914AF
1-11
Page 48
System Description
A015903P.EPS
1
2
3
4

Modular Chemistry (MC) System

Sample Carousel
The ten-rack position Sample Carousel is a motor-driven turntable. Refer to Figure 1.4. Under normal operation, eight of the Sample Carousel positions are available for routine processing and two positions are reserved for priority racks.
Figure 1.4 Sample Carousel Area
1. Liquid Level Sense Assembly
2. Sample Probe (MC)
Modular Chemistry (MC) System
Introduction
The Modular Chemistry system consists of the following major assemblies:
Reagent storage area
Ratio pump
Sample probe
Electrolyte injection cup (EIC)
Flow cell assembly
Chemistry reaction modules
A detailed description of each component is presented in the following paragraphs.
3. Collar Wash
4. Sample Carousel
1-12
A13914AF
Page 49
Reagent Storage Area
The reagent containers used to supply the modular chemistries are located behind the left front door of the system. The only exception is the CO
Figure 1.5 Modular Chemistry Reagent Storage Area
System Description
Modular Chemistry (MC) System
alkaline buffer which is located on the ISE module.
2
1
1. Bar Code Reader (hand held)
2. Reagent Storage Area
Reagent Bar Code Reader (Modular)
Behind the left side door of the system there is a hand-held bar code reader. Refer to Figure 1.5.
When this reader is held up to the label of one of the modular reagents and the trigger is pressed, the identity of the reagent, the lot number and reagent volume are automatically entered into the reagent load screen.
A13914AF
1-13
Page 50
System Description
A015904P.EPS
6
54321
Modular Chemistry (MC) System
Ratio Pump
The Ratio Pump is a motor-driven, multicylinder, positive-displacement pump used to deliver the necessary reagents to the ISE flow cell. Refer to Figure 1.6. It consists of a three-step piston housed in three, stacked, independent cylinders.
Figure 1.6 DxC 800 Ratio Pump
1. Solenoid Valve (example)
2. Cylinder 1
3. Cylinder 2
4. Cylinder 3
5. Outlet Line (example)
6. Inlet Line (example)
1-14
A13914AF
Page 51
Electrolyte Injection Cup (EIC)
The EIC mixes the sample and buffer prior to delivery of the sample (now diluted) to the flow cell.
Figure 1.7 Electrolyte Injection Cup
System Description
Modular Chemistry (MC) System
1
1. Waste Outlet
2. Flow Cell Outlet
3. Buffer Inlet
4. Reference Inlet
5. DI H
O Inlet
2
A13914AF
1-15
Page 52
System Description
Modular Chemistry (MC) System
Flow Cell Assembly
The flow cell assembly houses the seven electrodes that perform the analysis of sodium, potassium, chloride, carbon dioxide, and calcium.
Figure 1.8 Flow Cell
1. Inlet Port
2. CL Electrode
3. K Electrode
4. CO
5. CO
Reference Electrode
2
Electrode
2
6. Exit Port for Waste (large tube)
7. Exit Port for Internal reference
8. Na Reference Electrode
9. Na Electrode
10. Ca Electrode
1-16
A13914AF
Page 53
Chemistry Reaction Modules (Basic Components)
A015905P.EPS
3
4
5
1
2
Each of the six Chemistry Reaction Modules have similarities in their design. These common elements are described below. Refer to Figure 1.9. Unique design elements of the modules are described under the specific module headings later in this section.
Figure 1.9 Basic Components of Modules (Ex: Albumin)
System Description

Cartridge Chemistry (CC) Reagent Handling System

1
1. Circuit Board (behind protective shield)
2. Reaction Cup
3. Reaction Cup Housing
4. Mixer Motor Assembly
5. Reagent Pump Assembly
Cartridge Chemistry (CC) Reagent Handling System
Introduction
The Cartridge Chemistry Reagent Handling system is composed of the following components:
Reagent cartridges
Reagent carousel
Reagent probe assembly
Reagent mixer assembly
Mixer wash cup
The Reagent Handling system is used to transfer reagent from the individual cartridges to the reaction cuvettes for processing and analysis of the requested chemistry tests.
A13914AF
1-17
Page 54
System Description
Cartridge Chemistry (CC) Reagent Handling System
Reagent Cartridges
Reagent cartridges are single use, recyclable plastic containers that house the individual liquid reagent components necessary to perform a chemistry test. The reagent carousel is capable of storing 59 cartridges on board.
Figure 1.10 CC Reagent Cartridge
1. A Compartment
2. B Compartment
3. C Compartment
1-18
A13914AF
Page 55
Reagent Carousel and Reagent Bar Code Readers
The Reagent Carousel Compartment provides an on-instrument storage area for the individual reagent cartridges. A total of 59 reagent cartridges can be stored in the carousel at one time. Refer to Figure 1.11.
The storage compartment is refrigerated and fan-cooled to maintain a temperature of +5°C (±3°C).
The Bar Code Reader, (refer to Figure 1.11), situated near the front of the Cartridge Chemistry reagent access door, scans each label during the loading of reagent cartridges. Only the reader for the selected carousel is active. An audible signal acknowledges successful reading of the label.
Figure 1.11 CC Dual Carousels with Two Bar Code Readers
1
2
System Description
Cartridge Chemistry (CC) Reagent Handling System
1
1. Top Positions 31–59
2. To p Ba r Co d e Re a d e r
3
4
A007408P. E P S
3. Bottom Positions 1–30
4. Bottom Bar Code Reader
A13914AF
1-19
Page 56
System Description
A015908P. E P S
2
4
1
3
Cartridge Chemistry (CC) Reagent Handling System
Reagent Probe Assembly
The Reagent Probe assembly consists of a mechanical structure that supports two moveable cranes. Attached to each crane is a pickup probe. Refer to Figure 1.12.
Figure 1.12 CC Reagent Probe Area
1. Reagent Mixer
2. CC Reagent Probe A
Reagent Mixer Assembly
This assembly consists of a mechanical structure that supports a single, moveable crane. Attached to the crane is a high-speed mixer. Refer to Figure 1.12.
Reagent Mixer Wash Cup
The Mixer Wash Cup sprays the mixer with diluted wash solution while the mixer moves up and down in the cup.
3. CC Reagent Probe B
4. Collar Wash
1-20
A13914AF
Page 57

Cuvette Reaction System

Introduction
The Cuvette Reaction system consists of the following components:
Reaction carousel assembly
Photometer assembly
LPIA (Large Particle Immuno Assay) or NIPIA (Near-Infrared Particle Immuno Assay) module
(optional)
Cuvette wash station
The Cuvette Reaction system involves the process of obtaining absorbance readings from each cuvette during the analysis cycle.
Following the completion of each chemistry test, the cuvettes are processed through a wash station in preparation for the next chemistry.
System Description
Cuvette Reaction System
1
A description of each component is presented in the following paragraphs.
Reaction Carousel
The reaction carousel assembly, (refer to Figure 1.13), supports a total of 125 cuvettes. Each cuvette is glass with a 0.5 cm path length and is approximately 30 mm high. The cuvettes are non-disposable and have an indefinite life-span on the instrument. Cuvettes only need replacement if they are damaged. (They remain under warranty for two years.)
A13914AF
1-21
Page 58
System Description
A015909P.EPS
2
1
3
Cuvette Reaction System
Figure 1.13 Reaction Carousel Area (Typical – Cover Removed)
1. LPIA Module (optional)
2. Reaction Carousel
3. Photometer
Photometer Assembly
Attached to the reaction carousel support frame is the Photometer assembly. This consists of a xenon pulse lamp, a discrete 10-position silicon-diode detector array, a monochromator housing unit, and associated electronic circuitry. Refer to Figure 1.13.
As each cuvette passes through this optics station during a spin cycle, the xenon lamp is flashed and the resulting light beam travels through the opposing sides of the square cuvette.
LPIA (Large Particle Immuno Assay)
The optional LPIA module uses a photometric detection system for large particle immuno assays. This assembly is attached to the reaction carousel support frame to the left of the photometer assembly, near the sample carousel. Refer to Figure 1.13. It has two printed circuit boards, an LED (light-emitting diode) and a photodetector. Communication with the system software is through fiber optic cables.
1-22
A13914AF
Page 59
Cuvette Wash Station
The Cuvette Wash Station, (refer to Figure 1.14), consists of four coaxial probes, an elevator assembly, and the associated tubing.
A motor controls the vertical motion required by the elevator to raise and lower the probes during the wash stage.
Figure 1.14 Cuvette Wash Station
System Description
Cuvette Reaction System
1
1. Probe #1
2. Probe #2
3. Probe #3
4. Probe #4
5. Wash Sta tion
A13914AF
1-23
Page 60
System Description
A016497P.EPS
6
1
2
3
4
5

Hydropneumatic System

Hydropneumatic System
Introduction
The main components of the Hydropneumatic System are mounted on a slide-out drawer that allows for easier operator access. Refer to Figure 1.15 and Figure 1.16. When fully extended, the drawer locks open. To close, lift up on the metal tabs, located on each side of the bottom runner of the hydropneumatic unit, and push the drawer inward.
Figure 1.15 DxC 800 Hydropneumatics (right side)
1. Wash Concentrate Solution
2. No Foam Reagent
3. DI Water Canister
1-24
4. Wash Solution Canister
5. Wash Concentrate Re servoir
6. Auto-Gloss
A13914AF
Page 61
Function
A015911P.EPS
2
4
5 6 7
1
3
System Description
Hydropneumatic System
The function of the hydropneumatic system is to provide the following media to the different functional areas of the instrument:
Vac uum
Compressed air
Diluted wash solution
Deionized water
Figure 1.16 DxC 800 Hydropneumatics (left side)
1
1. Waste B Exit Sump
2. Waste Exit Sump
3. DI Water Inlet On/Off
4. Waste Sump
A13914AF
5. Waste B Sump
6. DI Water Reservoir
7. Vacuum Accumulator
1-25
Page 62
System Description

Operation and Control Components

Operation and Control Components
Operator Controls
The operator interfaces with various control devices such as the keyboard, monitor and push­button controls during a routine run. Basic operating functions are controlled and reviewed from the monitor. Calibration functions are also controlled from the monitor. Information is selected and entered into the system through touch screen monitors, from a mouse, and/or at a keyboard.
Push-button controls (refer to Figure 1.3) are used to start the process, to prioritize a sample run, or to stop the process under certain conditions as described in Table 1.12.
Table 1.12 DxC Push-Button Controls
Push-Button Control Type
Run To start the test process.
Priority To prioritize the loading of a rack by creating a space in front of the autoloader for
loading the next rack into a priority position on the sample carousel. It does not alter the sample priority (STAT or routine) previously defined in sample programming.
Stop To stop the process. The stop button should be used only under the following
conditions:
To stop instrument motions
To conduct a maintenance/repair activity
To home and realign mechanical components without rebooting
Primary Function
1-26
A13914AF
Page 63

Main Screen and Program Structure

E011950S.EPS
5
6
7
4
3
1 2
Main Screen
DxC System operating and programming functions are initiated from the Main screen at the DxC analyzer (refer to Figure 1.17). In addition, the screen provides status information to help determine the present state of the system.
Figure 1.17 UniCel DxC 800 Main Screen
System Description
Main Screen and Program Structure
1
1. CTS Indicator
2. Host Indicator
3. System Status Indicator
A13914AF
4. Menu Bar
5. Sample Status Icons
6. Rack Status Area
7. Function Bar
1-27
Page 64
System Description
Main Screen and Program Structure
Status Functions
The following Table depicts the status information available from the operator screens of the DxC analyzer.
Table 1.13 Main Screen Status Functions
Status Indicator Status Description
CTS This indicator (1) in Figure 1.17, appears in the blue bar at the top left side of the
Host This indicator (2) in Figure 1.17, appears on the blue bar at the top right side of the
Operator screens when the Closed Tube Sampling (CTS) option is installed. When "CTS" appears, Closed Tube Sampling is enabled. When CTS appears within the
international "No" symbol ( ), the feature has been installed, but it is not enabled. When the indicator is absent, the CTS option is not installed. Enabling and disabling Closed Tube Sampling is accomplished using the Setup procedures described in the Reserved Racks/Obstruct Detect topic of CHAPTER 3, System Setup Options.
IMPORTANT CTS is an optional feature of the system, which allows the system to
pierce primary sample tubes. The operator is only offered the enable/disable CTS option if a CTS assembly is installed, otherwise that selection is grayed out on the Setup screen. The default for CTS sampling is "OFF."
Operator screens. The indicator to the right of the label shows communication activity between the instrument and the host computer. A blue bar indicates that the host is sending information. A green bar indicates that the host is receiving information.
Printing The indicator appears at the middle of the operator screens. When the indicator
appears, you must reboot the DxC system to restart printing.
System Status This indicator (3) appears on the left side of the Main screen in Figure 1 .17 just
above the Menu Bar (4). It shows the current state of the system: either Running, Standby or Stopped. When both the Modular Chemistry (MC) and the Cartridge Chemistry (CC) functions have the same status, a single status indicator appears. When the statuses of these components are different, the MC Status is shown on the left, and the CC Status appears on the right. For example, Running/Standby would indicate that the MC side is Running, and the CC side is in Standby mode.
1-28
A13914AF
Page 65
System Description
Main Screen and Program Structure
Sample Status Indicators
Refer to Figure 1.17. When monitoring sample status, a sample status icon appears in front of listed samples. Samples are listed in a rack status box directly below the sample status icon legend (5). The sample statuses shown are as follows:
Table 1.14 Sample Status Indicators
Status Indicator Status Description
Not Programmed Indicates a sample has been loaded without any programming associated with that
Sample ID.
Query Pending Sample is waiting for specific program information to be downloaded from the host.
In Progress Sample has been identified and is currently being processed.
Aspirated Indicates whether an initial aspiration of the sample has been accomplished.
Incomplete Sample has some tests that are still pending.
Complete All tests for sample have been completed.
1
A13914AF
1-29
Page 66
System Description
Main Screen and Program Structure
Menu Bar Icons and Program Structure
Near the top of the Operator screens, a series of icons on the touch screen provide access to each of the major functional areas of the system (4) (refer to Figure 1.17).
Figure 1.18 Program Structure (Main, Samples)
1-30
A13914AF
Page 67
Figure 1.19 Program Structure (Results, Rgts/Cal)
System Description
Main Screen and Program Structure
1
A13914AF
1-31
Page 68
System Description
Main Screen and Program Structure
Figure 1.20 Program Structure (QC, Setup)
1-32
A13914AF
Page 69
Figure 1.21 Program Structure (Setup - continued)
System Description
Main Screen and Program Structure
1
A13914AF
1-33
Page 70
System Description
Main Screen and Program Structure
Figure 1.22 Program Structure (Setup - continued)
1-34
A13914AF
Page 71
Figure 1.23 Program Structure (Utils)
System Description
Main Screen and Program Structure
1
A13914AF
1-35
Page 72
System Description
Main Screen and Program Structure
Figure 1.24 Program Structure (Utils – continued)
1-36
A13914AF
Page 73
Figure 1.25 Program Structure (Utils – continued, Status, Instr Cmd, and Help)
System Description
Main Screen and Program Structure
1
A13914AF
1-37
Page 74
System Description
Main Screen and Program Structure
Recall Results for On-board Samples
To preview the results of tests In Progress, before the rack is unloaded from the system, perform the following procedure.
1
Select the sample for the desired results. The sample is highlighted.
2
Select Results F1. The results for any completed tests for the requested sample are shown along with the status of any pending tests.
3
Select one of the function keys:
Print F8 - to send completed results to the printer.
Host F6 - to send completed results to the host.
Accessing the Help System
Help is available online in multiple supported languages. To reach the online Help system,
Select the Help icon , or
Use the combination keys
The manual opens up full-screen in the language selected in System Setup. You can use the combination keys the online manual.
Online Manual Links
The system may display an “Error -- The page cannot be displayed” pop-up message when you select the
Event ID button to link to the online manual from an instrument pop-up error message, or when
you select a link in the online manual. To continue, record the Event ID, select pop-up message, and then use the search feature of the online manual to find the topic you were seeking.
NOTE
To view the instrument pop-up error Event ID number, use the (ALT) + (F1) key combination to return
to the instrument screen.
(
ALT
+
ALT
(
+
to toggle between the active screen and the last accessed page of
)
(F1)
from any screen.
)
(F1)
OK from the Error
Print Data From a Screen
Many screens contain data that may be printed by selecting Print F10. Pressing the the keyboard also causes the current screen to print, regardless of whether the
available.
1-38
PrtScn
(
Print F10 option is
key on
)
A13914AF
Page 75
Rack Status Area
These boxes, Figure 1.17 (6), represent the ten possible rack locations on the Sample Carousel. As each rack is loaded, the rack number is displayed at the top of the Rack Status box. All sample IDs on the rack are listed below the rack number. The Sample Status icons described earlier indicate sample status.
STAT Highlighting
Any sample programmed as a STAT is highlighted in a yellow rack status box on the Main screen.
Additional Function Keys
Other functions may also be accessed through the Main screen.
Results F1 - Retrieves results for samples that are In Progress on the system.
Unload F2 - Requests particular racks to be unloaded at the next possible opportunity. Log F8 - Lists errors detected on samples, for example, incomplete tests, host query timeout,
duplicate sample ID, programming conflict. If the error is logged, the in yellow. The yellow highlight disappears when the key is selected. The error will stay in the Sample Log screen for 12 hours.
System Description
Main Screen and Program Structure
Log F8 key is highlighted
1
Pre Run Summary F9 - Lists reagent and calibration status, as well as sample programming
needs.
Post Run Summary F10 - Lists error conditions which have occurred on the system, including
suppressed results and no sample detected.
These functions are covered more thoroughly in the following sections.
Request a Sample for Unload
To unload racks that contain specific samples perform the following procedure.
1
Select the rack to unload. The rack is highlighted.
2
Select Unload F2. The rack is offloaded at the earliest opportunity.
Sample Log
The sample log feature provides a method to review the status of any incomplete patient and control samples that have been off the system for less than 12 hours.
A13914AF
1-39
Page 76
System Description
Main Screen and Program Structure
The samples in the log are arranged in a first in, first out sequence. When a new sample comes in at the top of the list, the other sample(s) move down one space. Samples will remain on the list until their status changes to "Complete" or they are on the list for more than 12 hours.
Request the Log
Table 1.15 Operations with the Sample Log
See the log Log F8 from the Main screen
To... Select...
See more pages
Print the log (present list of sample history)
Page Up
(
PRINT
)
or
Page Down
(
NOTE This printout contains the same information as the
body of the window. This window remains in view after the print button is selected.
Exit and return to the Main screen OK
Description of Log Fields
Table 1.16 Sample Log Fields
Field Description
Rack The rack in which the sample was run.
Position (Pos) The position of the sample in the rack when it was run.
Sample ID There are two ways to get this sample ID:
1. The system reads the bar code.
2. Manually input a sample program.
If there is no sample ID, there are two possible fields:
If there is no sample ID then "NO ID" appears.
If there is no sample ID and it is required, then a status of
Not Programmed appears.
)
1-40
IMPORTANT The Not Programmed status is used for samples for which:
A bar code was read, but no programming was found.
A bar code was not read, and no manual sample programming was
found for that rack and position.
Updated This is the time and date that the sample entry was last updated.
The time and date is displayed in the format selected in System Setup (12 or 24-hour clock and month/day or day/month).
A13914AF
Page 77
System Description
Main Screen and Program Structure
Table 1.16 Sample Log Fields (Continued)
Field Description
Status If there are two results because of a Critical Sample Rerun, the log entry
shows:
Review Results
If an incomplete sample is removed from the sample carousel, the log entry
shows one of the following status indications:
Incomplete
Sample Required
Not Programmed
RemovedRerun
In Progress
IMPORTANT "In Progress" means a sample load error. Refer to the list
below for more status indications for some of these errors.
1
If some CC results are still incubating, the sample status shows Removed
when it is first taken off the sample carousel. In this case, the sample is deleted from the list when it goes to Complete. If it does not go to Complete, more status strings are added, as applicable. Possible statuses are listed below:
Duplicate Sample IDEditing Sample
Programming Conflict
Barcode ID Too LongInvalid Barcode Char(acter)
Host Query Timeout
The Status updates when the sample is:
loaded and an error condition is detected
removed from the sample carouselcompleted with all its tests
A13914AF
1-41
Page 78
System Description
Main Screen and Program Structure
Pre Run Summary
The Pre Run Summary is a printed report containing summary information about the programmed tests for each chemistry. This information helps in determining the status of the reagents to verify that the system is in a condition to complete the requested tests.
Information includes:
reagent volume/tests available for resident reagents
calibration time remaining and cal status
lot and serial number of cartridge
reagent status/on-board life
number of tests programmed versus available for on-board reagents
number of test programmed for non-resident reagents
Request a Pre Run Summary
1
If not currently in the Main screen, select the Main icon from the menu bar.
2
Select Pre Run F9. The Pre Run summary prints automatically. It may take up to 25 seconds to start to print.
Post Run Summary
The Post Run Summary option provides a printed list of the samples that are pending or incomplete, with an explanation of the status. Incomplete or Suppressed tests are generated due to a reagent or calibration situation. To run a Pending test, simply reload the rack or bar code labeled tube. Results for reloaded, incomplete tests replace results that were previously incomplete. Reloaded sample results are collated with the existing reported tests in the sample report.
Request a Post Run Summary
1
If not currently in the Main screen, select the Main icon from the menu bar.
1-42
2
Select Post Run F10. The Post Run Summary prints automatically.
NOTE
The Post Run Summary may require up to 7 minutes for printing.
A13914AF
Page 79

Theory of Operation

Introduction
The UniCel DxC Synchron Clinical Systems are microprocessor-controlled, random access clinical analyzers capable of processing a wide variety of operator-selected chemistries in a single run.
Cartridge Chemistries (CC)
The optical system of the DxC enables rate, endpoint, and nonlinear analyses to be performed simultaneously. These analyses are referred to as cartridge chemistries because the reagents are stored in cartridges.
Cartridge Chemistry (CC): Sample and Reagent Processing
System Description
Theory of Operation
1
During operation, a number of events occur simultaneously and are under direct control of the instrument microprocessors.

Cartridge Chemistry: Calibration Theory

Introduction
Calibration determines the relationship between measured reaction responses and known concentrations. Calibration factors are derived from this relationship. These factors are used to convert the measured reaction responses to final concentration results.
Calibrated chemistries include endpoint and first-order rate chemistries, drugs, DATs, and specific proteins. Zero-order rate chemistries include enzymes, which are precalibrated and require no routine calibration. Enzyme verification can be performed on some of the enzymes to conform to International Federation of Clinical Chemistry (IFCC) guidelines at +37°C.
A13914AF
1-43
Page 80
System Description
Cartridge Chemistry: Calibration Theory
Endpoint and First-Order Chemistries
Calibration of endpoint and first-order rate chemistries involve the use of a single-level calibrator solution or a two-level calibrator kit. Each analyte in the calibrator solution has a known concentration value associated with it. With each new lot of calibrator solution, the values are transferred from disk and stored in memory for later use in the calibration procedure.
For most calibrated cartridge chemistries, the system will set calibration factors based on four calibrator replicates per calibrator level. The instrument will determine and discard the highest and lowest of the four replicates. The remaining two values are called the usable calibrator replicates. All four replicates will appear on the report but the average of the two usable replicates is used to determine the calibration factor.
For other chemistries, calibration is based on two calibrator replicates per calibrator level. No replicates are discarded. The average value of the calibrator replicates is used to determine the calibration factor.
Endpoint and First-Order Calibration Formulas
The calibration factor is determined by using one of the following equation sets where reaction and blank are used from the usable replicates.
1-44
A13914AF
Page 81
Cartridge Chemistry: Calibration Theory
Table 1.17 Calculation of Calibration Factors for Endpoint and Rate Chemistries
Ty pe F or m ul a
System Description
1
Nonblanked Endpoint Chemistries
Blanked Endpoint Chemistries
FOR HIGH CALIBRATOR LEVEL:
Reaction ABS = ABS Reaction ABS = ABS
rep1
+ ABS
rep2
(ABS
rep1
rep2
) × 0.5 = ABS
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
Reaction ABS = ABS Reaction ABS = ABS (ABS
rep1
+ ABS
rep2
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × ABS
rep1
rep2
) × 0.5 = ABS
Cal Set Point (hi) – Cal Set Point (lo)
avg
ABS
(lo)
avg
(hi) – ABS
(hi)]
avg
avg
(lo)
E014416L.EPS
Sample values are calculated by the following equation: (Reaction ABS × Cal Factor) + offset = sample value
IMPORTANT For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
FOR HIGH CALIBRATOR LEVEL:
(Reaction ABS - Blank ABS) = Delta ABS (Reaction ABS - Blank ABS) = Delta ABS (Delta ABS
+ Delta ABS
rep1
FOR LOW CALIBRATOR LEVEL:
(Reaction ABS - Blank ABS) = Delta ABS (Reaction ABS - Blank ABS) = Delta ABS (Delta ABS
+ Delta ABS
rep1
rep1
rep2
) × 0.5 = Delta ABS
rep2
rep1
rep2
) × 0.5 = Delta ABS
rep2
avg
avg
(hi)
(lo)
A13914AF
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × Delta ABS
Cal Set Point (hi) – Cal Set Point (lo)
Delta ABS
(hi) – Delta ABS
avg
avg
(hi)]
(lo)
avg
E014417L.EPS
Sample values are calculated by the following equation: [(Reaction ABS - Blank ABS) × Cal Factor] + offset = sample value
IMPORTANT For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
1-45
Page 82
System Description
E007123L.EPS
Blank Correction Factor =
Total Reaction Volume
Blank Volume
=
Volume of Total Reagent and Sample at Reaction Read
Volume of Reagent(s) (and Sample) at Blank Read
Cartridge Chemistry: Calibration Theory
Table 1.17 Calculation of Calibration Factors for Endpoint and Rate Chemistries (Continued)
Ty pe F or m ul a
Blanked Endpoint Chemistries (with Volume Correction)
FOR HIGH CALIBRATOR LEVEL:
[Reaction ABS - (Blank ABS × Blank Correction Factor)] = Delta ABS [Reaction ABS - (Blank ABS × Blank Correction Factor)] = Delta ABS (Delta ABS
+ Delta ABS
rep1
) × 0.5 = Delta ABS
rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
[Reaction ABS - (Blank ABS × Blank Correction Factor)] = Delta ABS [Reaction ABS - (Blank ABS × Blank Correction Factor)] = Delta ABS (Delta ABS
Cal Factor (Slope) =
+ Delta ABS
rep1
) × 0.5 = Delta ABS
rep2
avg
(lo)
Cal Set Point (hi) – Cal Set Point (lo)
Delta ABS
(hi) – Delta ABS
avg
(lo)
avg
E014417L.EPS
rep1
rep2
rep1
rep2
Offset = Cal Set Point (hi) - [Cal Factor × Delta ABS
Sample values are calculated by the following equation: [[Reaction ABS - (Blank ABS × Blank Correction Factor)] × Cal Factor] + offset = sample value
IMPORTANT For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
avg
(hi)]
1-46
A13914AF
Page 83
System Description
Cartridge Chemistry: Calibration Theory
Table 1.17 Calculation of Calibration Factors for Endpoint and Rate Chemistries (Continued)
Ty pe F or m ul a
1
Nonblanked Rate Chemistries
FOR HIGH CALIBRATOR LEVEL:
Reaction Rate = Rate Reaction Rate = Rate (Rate
rep1
+ Rate
rep2
rep1
rep2
) × 0.5 = Rate
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
Reaction Rate = Rate Reaction Rate = Rate (Rate
rep1
+ Rate
rep2
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × Rate
rep1
rep2
) × 0.5 = Rate
avg
(lo)
Cal Set Point (hi) – Cal Set Point (lo)
Rate
(hi) – Rate
avg
avg
(hi)]
avg
(lo)
E014418L.EPS
Sample values are calculated by the following equation: (Reaction Rate × Cal Factor) + offset = sample value
IMPORTANT For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
A13914AF
1-47
Page 84
System Description
Cartridge Chemistry: Calibration Theory
Table 1.17 Calculation of Calibration Factors for Endpoint and Rate Chemistries (Continued)
Ty pe F or m ul a
Blanked Rate Chemistries
FOR HIGH CALIBRATOR LEVEL:
(Reaction Rate - Blank Rate) = Delta Rate (Reaction Rate - Blank Rate) = Delta Rate (Delta Rate
+ Delta Rate
rep1
) × 0.5 = Delta Rate
rep2
rep1
rep2
(hi)
avg
FOR LOW CALIBRATOR LEVEL:
(Reaction Rate - Blank Rate) = Delta Rate (Reaction Rate - Blank Rate) = Delta Rate (Delta Rate
+ Delta Rate
rep1
) × 0.5 = Delta Rate
rep2
Offset = Cal Set Point (hi) - [Cal Factor × Delta Rate
rep1
rep2
(lo)
avg
(hi)]
avg
Sample values are calculated by the following equation: [(Reaction Rate - Blank Rate) × Cal Factor] + offset = sample value
IMPORTANT For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
1-48
A13914AF
Page 85
System Description
E007123L.EPS
Blank Correction Factor =
Total Reaction Volume
Blank Volume
=
Volume of Total Reagent and Sample at Reaction Read
Volume of Reagent(s) (and Sample) at Blank Read
Cartridge Chemistry: Calibration Theory
Table 1.17 Calculation of Calibration Factors for Endpoint and Rate Chemistries (Continued)
Ty pe F or m ul a
1
Blanked Rate Chemistries (with Volume Correction)
FOR HIGH CALIBRATOR LEVEL:
[Reaction Rate - (Blank Rate × Blank Correction Factor)] = Delta Rate [Reaction Rate - (Blank Rate × Blank Correction Factor)] = Delta Rate (Delta Rate
+ Delta Rate
rep1
) × 0.5 = Delta Rate
rep2
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
[Reaction Rate - (Blank Rate × Blank Correction Factor)] = Delta Rate [Reaction Rate - (Blank Rate × Blank Correction Factor)] = Delta Rate (Delta Rate
+ Delta Rate
rep1
) × 0.5 = Delta Rate
rep2
avg
(lo)
rep1
rep2
rep1
rep2
Offset = Cal Set Point (hi) - [Cal Factor × Delta Rate
Sample values are calculated by the following equation: [[Reaction Rate - (Blank Rate × Blank Correction Factor)] × Cal Factor] + offset = sample value
IMPORTANT For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
avg
(hi)]
A13914AF
1-49
Page 86
System Description
Cartridge Chemistry: Calibration Theory
Non-Linear Chemistries
Non-linear chemistries include drugs and specific protein assays. Unlike the first-order rate and endpoint chemistries, which exhibit a linear response to increasing concentration, the calibration curves for non-linear chemistries exhibit logarithmic (S-shaped) or other nonlinear relationships. For this reason, curve fitting interpolation techniques are employed to construct the calibration curve.
For some non-linear calibrations, the curve parameters for a reagent lot are calculated during manufacturing. The curve parameters are encoded in bar code form, shipped on a card in the reagent box and loaded onto the system. A single point calibration is then run to adjust for instrument-to-instrument variation. If necessary, the calibration includes the sample diluent (DIL1), used as a blank, that is subtracted from all calibrator or sample responses.
Multi-point chemistry calibration consists of five or six different levels of calibrators. These chemistries set calibration based on single replicates of each calibrator level.
Single-point chemistry calibration consists of one or two levels of calibrators. These chemistries set calibration based on two to four replicates of each calibrator level.
Non-Linear Calibration Formulas
The standard curve is determined by use of one of several nonlinear math models. The system uses an iterative technique to calculate the curve parameters. A modified Newton iteration is used to choose values. The best-fitting calibration curve is determined by minimizing the sum of the difference between the observed response and the calculated response of each standard.
The following symbols are used in the math models presented below:
R = sample response
Conc = standard or sample concentration
R
0
K
c
a, b, c = parameters which define the nonlinear elements of the math model
= calculated response for a zero sample
=scale parameter
1-50
A13914AF
Page 87
System Description
Cartridge Chemistry: Calibration Theory
Table 1.18 Math Models for Non-Linear Chemistries
Ty pe F or m ul a
Model #1 Math Model #1 is the four-parameter log-logit function most commonly used with
reagents that use antibodies.
Sample values are determined using the calculated curve parameters and the math model. Values may be calculated directly as this model can be solved for concentration.
Model #2 Math Model #2 is a five-parameter logit function.
E014421L.EPS
1
This function cannot be solved directly for concentration. The instrument uses an iterative method to determine the sample value.
Model #3 Math Model #3 is a five-parameter exponential function.
This function cannot be solved directly for concentration. The instrument uses an iterative method to determine the sample value.
Models #4 through #7
Model #8 Math Model #8 is an alternative to model #2, the five-parameter logit function.
These Models are reserved for future development.
This function cannot be solved directly for concentration. The instrument uses an iterative method to determine the sample value.
A13914AF
1-51
Page 88
System Description
E014478L.EPS
Cartridge Chemistry: Calibration Theory
Table 1.18 Math Models for Non-Linear Chemistries (Continued)
Ty pe F or m ul a
Model #9 Math Model #9 is an extension to model #1, the four-parameter log-logit function.
The "c" is allowed to be either +1 or -1.
If c = +1, then this is equivalent to model #1.
If c = -1, an alternative function is being used.
This function cannot be solved directly for concentration. The instrument uses an iterative method to determine the sample value.
Polynomial Exponential (PXP)
E014476L.EPS
Quadratic (POLY2)
Lorentz
Double Inflection Model DP4
R =
1 +
E014477L.EPS
K
c1
a
1
+
1 +
K
c2
a
2
concconc
E011537L.EPS
1-52
A13914AF
Page 89
Drugs of Abuse Testing (DAT) Chemistries
The Drugs of Abuse Testing (DAT) assays require three levels of calibrators. The calibration measures the separation between calibrators to measure reagent integrity. The calibration factor generated is non-functional for sample result calculation.
The cutoff value for each DAT chemistry represents the mean reaction rate of the low calibrator, reported in mA/min units on patient and control reports. The reaction rate of the samples is compared to the reaction rate of the low (cutoff) calibrator and reported out as POSITIVE or NEGATIVE. Cutoff values are stored in memory until the next successful calibration.
Enzyme Verification
Enzyme verification is a means of adjusting enzyme chemistry reporting units to IFCC/DGKCh methods. This feature is available for ALP, ALT-, AST-, CHE, CK-, GGT and LD. Verification also allows results to be adjusted for country specific correlation needs.
Verification is similar to calibration except that normalization factors are applied to the sample result in the form of a slope and offset adjustment, whereas calibration factors would be applied to the reaction response.
System Description

Modular Chemistry: Calibration Theory

1
Modular Chemistry: Calibration Theory
Calibration Theory
Modular chemistries are calibrated using two to three levels of calibrator (chemistry dependent). Four replicates per level are assayed. Data from two middle replicates of each level is used to set the system response. The highest and lowest replicates are discarded. Error checks are performed on the two middle replicates to verify successful calibration.
Calibration Error Detection
The analog signals generated by the calibrator measurements are converted to digital form. The resulting ADC values are compared to pre-programmed back-to-back, span and range limits to determine the calibration acceptability.
Modular Chemistries (MC)
The UniCel DxC contains seven chemistry modules (see chart below), each of which is used in the determination of eleven modular chemistries (MC), as follows.
A13914AF
1-53
Page 90
System Description

Cartridge Chemistry: Principles of Measurement

Table 1.19 Methodology and Modules Used with Modular Chemistries
Chemistry Methodology Module
Sodium Ion selective electrode (ISE) ISE Flow cell
Potassium Ion selective electrode ISE Flow cell
Chloride Ion selective electrode ISE Flow cell
Carbon Dioxide pH electrode ISE Flow cell
Calcium Ion selective electrode ISE Flow cell
Urea Nitrogen
Phosphorus
Creatinine
Glucose Oxygen sensor Glucose
To t al P r o t ei n
Albumin
a. DxC 800 only.
a
a
a
a
a
Conductivity electrode Urea Nitrogen
Colorimetric Phosphorus
Colorimetric Creatinine
Colorimetric Total Protein
Colorimetric Albumin
Cartridge Chemistry: Principles of Measurement
Spectrophotometric Methods
Spectrophotometric methods rely on the principle that a sample, such as a patient sample, a control, or a calibrator, when mixed with one or more appropriate chemical reagents, produces a substance that has the ability to absorb light at specific wavelengths. This substance is referred to as a chromophore.
Beer's Law
According to Beer's Law, the amount of light absorbed by the chromophore is proportional to the concentration of the constituent being measured. The system can measure this as an endpoint or a rate of formation.
A = abc
where
A = absorbance of the chromophore
a = absorptivity of the absorbing substance at the specific measuring wavelength(s)
b = cuvette light pathlength (cm)
1-54
c = constituent concentration (M)
A13914AF
Page 91

Routine Operation Overview

Daily Procedure
The following procedure shows an example of daily work flow using the UniCel DxC Synchron Clinical System.

CHAPTER 2

Preparing Samples for Analysis

IMPORTANT
1
If necessary, start the system.
2
Check reagent status. Load reagent as necessary.
3
Check calibration. Program or load calibrators as needed.
4
Program or load controls, if required.
5
If the system is not already running, press RUN on the system.
6
Check the control results to verify system operation.
7
Program or load patient samples.
This procedure assumes that the initial system setup has been completed.
A13914AF
8
If the system has gone into Standby, press RUN on the system.
9
Review the patient results.
2-1
Page 92

Preparing Samples for Analysis

Preparing Samples for Analysis
10
Return to Step 7 if more samples need to be run.
11
The system automatically returns to Standby when all testing is completed.
Preparing Samples for Analysis
Minimum Sample Volume
A minimum sample volume is required to run tests. To determine what volume of sample to use, refer to the Synchron LX/UniCel DxC Clinical Systems Sample Template.
Sample Racks
Sample racks accept the following tube and cup sizes:
Table 2.1 Sample Racks
13 × 75 12 × 75 mm tubes
16 × 75 16 × 75 mm tubes
13 × 100 13 × 100 mm tubes
16 × 100 16 × 100 mm tubes
IMPORTANT
Rack Accepts these cups and tubes
13 × 75 mm tubes
0.5 mL cups
2.0 mL cups
0.5 mL cups
2.0 mL cups Capillary collection tubes (use with the capillary tube adapter)
16.5 × 92 mm tubes Beckman Coulter 0.5 mL Cup Insert (PN 467406)
Adapters are provided to adapt various sized sample tubes (secondary tubes) to the short racks. These adaptors must only be used in racks designated as reserved. The reserved rack feature is described in this chapter.
2-2
A13914AF
Page 93
Preparing Samples for Analysis
E014498L.EPS
0.5 mL 2.0 mL
Preparing Samples for Analysis
Sample Preparation by Container Type
Table 2.2 shows how to prepare different sample containers.
Table 2.2 Preparation of Sample Containers
If running a sample from a... Then...
Primary Tube Use the Synchron LX/UniCel DxC Clinical Systems Sample Template to
determine adequate sample volume.
Remove the cap if not using CTS.
For CTS systems, remove the cap, if not a validated closed tube.
Secondary Tube Determine sufficient volume.
Check for fibrin or other materials resulting from storage.
2
Beckman Coulter Synchron Microtube
Sample Cup
0.5 mL (PN 651412)
2.0 mL (PN 652730 or
81902)
BD Microtainer Place Microtainer in adapter (PN 472987).
0.5 mL Cup Insert (PN 467406) (reusable)
Place into a 13 × 100 mm rack.
Pipette the sample into a Synchron Microtube.
Make sure there are no bubbles at the bottom of the tube.
A "reserved rack" must be used.
Place the cup into a 13 × 75 mm, or a 13 × 100 mm rack.
OR
Place the cup in 15 × 85 mm tube into a 16 × 100 mm rack.
Make sure there are no bubbles in sample.
If cup is placed in a 15 × 85 tube with a Bar Coded label, a "reserved rack"
must be used.
(Refer to figure to the right.)
Make sure there are no bubbles in sample.
A "reserved rack" must be used.
Place the metal Cup Insert into a 16 × 100 mm rack.
Insert a 0.5 mL cup into the Cup Insert.
A "reserved rack" must be used.
A13914AF
A011538L.EPS
2-3
Page 94
Preparing Samples for Analysis
2
3
4
4
A014600L.EPS
1
Preparing Samples for Analysis
Bar Code Labeling
The use of bar code labels is a highly accurate and efficient method for identifying and processing laboratory samples. However, the system must be able to identify and read every bar code label to process each sample correctly. The following paragraphs provide some basic information pertaining to bar code labels. Additional bar code information can be found in the UniCel DxC Synchron Clinical Systems Reference Manual.
CAUTION
A misread label can cause one sample ID to be read as another. The laboratory’s process for printing, placing, and meeting all bar code specifications is important to achieve highly accurate reading. Follow the bar code label specifications to keep the rate of misread labels to a minimum.
Bar Code Label Placement
Bar code labels must be applied to each sample tube in the correct location so that the bar code reader can read the bar code. The following diagram (Figure 2.1) describes how to place the label on a sample tube and how to place the tube into a sample rack.
Figure 2.1 Bar Code Label Placement
1. 14 mm (0.55 inch) Minimum
2. Label Placement Area
3. 20 mm (0.78 inch) Minimum
4. 7.5 Degree Maximum
2-4
A13914AF
Page 95
Sample Tubes Validated for CTS
Capped tubes can be run directly on the system without removing the cap. The CTS assembly cuts a small hole in the cap and the sample probe aspirates a sample directly from the tube. With this procedure, samples are processed faster and more safely.
CTS Validated Sample Tubes
Only validated sample tubes can be run on UniCel DxC Systems with CTS. However, capped and uncapped tubes can be placed on the same rack. Refer to the table below for validated tubes.
Table 2.3 Tubes Validated for Closed Tube Sampling
CTS Configuration Tubes Validated for Cap Piercing Size
1-Blade Thick CTS Becton Dickinson VACUTAINER with HEMOGARD 13 × 75mm
1-Blade Narrow CTS Sarstedt S-Monovette
a. This tube requires a special 5.5 mL rack (PN A18642).
Preparing Samples for Analysis
Preparing Samples for Analysis
13 × 100 mm 16 × 100 mm
Greiner VACUETTE 13 × 75mm
13 × 100 mm
75 × 15 mm 92 × 15 mm
2
a
CAUTION
Before placing validated closed tubes on the UniCel DxC Systems, check the top of the cap for any residual blood. Residual blood can contaminate the sample and affect results. If blood is present, remove it by using a cotton-tipped applicator stick moistened with DI water. When running in the CTS mode, if tubes off-loaded from the UniCel DxC Systems have water or droplets of water on the caps, disable the CTS and contact Beckman Coulter Support Center. Note: Oil on a cap is normal.
Closed Tube Sampling (CTS)
This is an optional feature that allows the system to pierce primary sample tubes. If a CTS assembly is installed, the operator can enable/disable the CTS assembly. A CTS Tracking database monitors pierced tubes. If you run a tube again, it will not be pierced again. The sample ID stays in this database for seven days for a tube that has been pierced.
When a particular Sample ID is manually cleared through Sample Programming, the CTS Tracking information is also cleared for that Sample ID. If the tube with this particular Sample ID was previously pierced, reloading this sample tube on any tracking networked instrument will cause it to be repierced.
NOTE
If a sample ID is cleared at the Host (LIS), the sample ID is NOT cleared in the CTS Tracking database.
A13914AF
If a previously pierced sample needs to be rerun, do not clear the Sample ID; use the rerun procedure.
If a Sample ID is manually cleared and reprogrammed for a previously pierced sample, remove the cap to prevent repiercing.
2-5
Page 96
Preparing Samples for Analysis
Preparing Samples for Analysis
If a Sample ID is reused for an unpierced sample, clear the Sample ID on the instrument, the cap does not need to be removed.
If you need to clear the CTS database (for example, if the host system's counter rolls over and uses the same Sample IDs again), clear the Sample IDs at any of the instruments connected by the tracking network.
Follow the steps below to verify Sample ID clearing status.
1
Select Samples from the menu bar.
2
Type the sample ID.
3
Press
Enter
(
.
)
4
Does the sample have its initial program?
If the sample has its initial program:
Do NOT remove its cap. Run the sample.
If the sample does NOT have its initial program (the ID has been cleared):
Remove its cap. Run the sample.
2-6
A13914AF
Page 97

How to Use Reserved Racks

When NOT to Use a Reserved Rack
When you use CTS (Closed Tube Sampling), do NOT run a closed tube in a reserved rack.
Reserved Racks
If a rack number is entered into this field, any sample containers in this rack will not be cap pierced even if the Cap Piercing feature is enabled. Level sensing on samples in the rack is set to the most sensitive level to detect small sample volumes. This type of reserved rack may be used for uncapped, primary or secondary tubes that have a small volume of sample, for example: nested cups on primary tubes, Microtubes and pediatric capillary collection tubes.
Preparing Samples for Analysis
How to Use Reserved Racks
2
IMPORTANT
To identify a reserved rack, put a red ® on the front of the rack.
Reserved Racks for HbA1c, or IBCT
There are two separate fields for specific tests. One for HbA1c and one for IBCT. If a rack number is entered into one of these fields, any samples run in that rack will only function for that specific test. For example, Rack 66 is entered into the Reserved Racks for HbA1c field. Any sample placed in that rack can only have an HbA1c/HbA1c2 run on it. The reason for this is that each test in this group requires some sort of sample preparation. If these tests are part of a group of other tests that do not require preparation, they cannot be run at the same time as the others. Using these features allows the prepared sample to be run in one of the reserved racks and allows the result of the prepared sample to be merged with the original sample report when all tests are complete.
IMPORTANT
HbA1c and pre-treated IBCT samples run in these reserved racks.
The default for each of these features is:
CTS = OFF,
Reserved racks = none reserved,
Obstruction detection = On,
Reserved racks for HbA1c or IBCT = none reserved.
Even if the Auto Serum Index feature is enabled, no serum indices are run for the pre-treated
A13914AF
These features can only be modified when the system is in Standby, Stopped, Startup, Instrument Down, or Homing state.
2-7
Page 98
Preparing Samples for Analysis
E015928S.EPS
How to Use Reserved Racks
Assigning or Reassigning Reserved Racks
1
Select Setup from the menu bar.
2
Select Page Down on the right side of the screen.
3
Select 17 Reserved Racks/Obstruction Detection. The following screen appears.
Figure 2.2 Reserved Racks/Obstruction Detection Setup Dialog Box
4
Type the rack numbers to assign as reserved racks in the HbA1c and IBCT fields.
AND/OR Type the rack numbers to assign as reserved racks.
5
Select OK to assign the racks.
2-8
A13914AF
Page 99

Overview

Introduction
This chapter summarizes the 29 System Setup options depicted on the Setup screens shown in
Figure 3.1 and Figure 3.2 below:
Figure 3.1 Setup Screen (scrolled to the top)

CHAPTER 3

System Setup Options

A13914AF
E011951S.EPS
3-1
Page 100
System Setup Options

Password Setup

Figure 3.2 Setup Screen (scrolled to the bottom)
E015931S.EPS
For detailed step-by-step instructions on using the System Setup option, refer to the UniCel DxC Synchron Clinical Systems Reference Manual.
Password Setup
Introduction
The Password Setup option allows the operator to:
Assign up to 100 user names and their passwords
Assign Administrator or Operator level privileges to each user name
Define/edit or delete user name/password setup
Enable/disable security for accessibility to certain system functions and setups
The following features may be password secured:
Results, Edit
Rgts/Cal, Modify Set Points
Rgts/Cal, Slope/Offset Adjustment
3-2
Rgts/Cal, Within-Lot Calibration
Rgts/Cal, Enzyme Validator
QC
Setup
Utils, Clear Event Log
A13914AF
Loading...