Beckman Coulter S120 Instructions For Use Manual

Page 1
Instructions for Use
StatSpin CritSpin Digital Reader Model Number S120
For In Vitro Diagnostic Use
This manual is intended for
CSD2 Digital Reader -with tube adapter for 75mm tubes, and with a universal power supply.
Printed in U.S.A.
55-003823-001CC September 2018
Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A.
Page 2
Instructions for Use StatSpin CritSpin Digital Reader Model Number S120
PN 55-003823-001CC (September 2018)
©
2018 Beckman Coulter, Inc.
All Rights Reserved.
Trademarks
Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc. in the United States and other countries.
All other trademarks are the property of their respective owners.
Find us on the World Wide Web at:
www.beckmancoulter.com
Rx Only Original Instructions
Page 3

Revision History

55-003823-001 CC, 09/2018
• Moved: Symbol/Regulatory Mark and a link to the website in the California Proposition 65 statement
55-003823-001 CB, 03/2018
• Converted the CritSpin Digital Reader Operations Manual to a Beckman Coulter Instructions for Use (IFU) Manual, and made general clarification to the IFU
• Added: Revision History, Safety Notice, Symbols and Definitions table, Alerts for Warning, Caution, Important, Note, and Tip explanations, and Warning and Cautions
• Updated: Logo, Manufacturer address, and Limited Warranty statement
• Deleted: EC Rep
55-003823-001CC
iii
Page 4
Revision History
iv 55-003823-001CC
Page 5
Read all product manuals and consult with Beckman Coulter-trained personnel before you operate the system. Do not perform any procedure before you carefully read all instructions. Always follow the product labels and the manufacturer’s recommendations. If you have any questions:
• Visit http://www.beckmancoulter.com.
• US customers: Contact Beckman Coulter Customer Support at 1-800-854-3633.
• International customers: Contact your local distributor.

Alerts for Warning, Caution, Important, Note, and Tip

Warning
Warning indicates a potentially hazardous situation which, if not avoided, could cause death or serious injury. Warning can indicate the possibility of erroneous data that could cause an incorrect diagnosis.

Safety Notice

Caution
Caution indicates a potentially hazardous situation which, if not avoided, can cause minor or moderate injury. Caution can also alert against unsafe practices, or indicate the possibility of erroneous data that could cause an incorrect diagnosis.
Important
Important indicates important information to follow.
Note
Note indicates notable information to follow.
Tip
Tip indicates information to consider.

Warning and Cautions

Pay close attention to the instructions that accompany the notes and symbols and the standard laboratory procedures outlined by your facility and local regulatory agencies.
55-003823-001CC
v
Page 6
Safety Notice
Warning and Cautions
Warning
Perform system operations with caution.
Wear Personal Protective Equipment (PPE) such as gloves, eye shields, and lab coats.
Wash hands thoroughly after contact with sample media and all maintenance activities.
Observe all laboratory policies and procedures related to the handling of biohazardous materials.
Refer to the applicable sources (such as Material Safety Data Sheets) for specific hazard information.
Warning
If the equipment is used in a manner not specified by Beckman Coulter, the protection provided by the equipment may be impaired.
Warning
Outside of North America: do not use the power cord supplied. Use power cord for at least 1.0 Amp with an IEC320/CEE22 female connector and male connector suitable for the power outlet to be used.
Warning
Electromagnetic Wave and Noise
The system generates, uses, and can radiate radio frequency energy. If the system is not installed and operated correctly, this energy can cause interference with other equipment. In addition, other equipment can radiate radio frequency energy to which the system is sensitive. If you suspect interference between the system and other equipment, Beckman Coulter recommends the following actions to correct the interference:
This equipment complies with the emission and immunity requirements described in this
part of the EN/IEC 61326 -1 As to emission, this system has been designed and tested to CISPR 11 Class A, so in a
domestic environment, it may cause radio interference, in which case, you may need to take measure to mitigate the interference.
It is recommended to evaluate the electromagnetic environment prior to operations of
the system. Do not use this system in close proximity to sources of strong electromagnetic radiation
(for example, unshielded intentional RF sources). As they can interfere with the proper operation.
vi 55-003823-001CC
Page 7
Do not use mobile or cordless telephones and transceivers in the same room as the
system. Do not use medical equipment that can be susceptible to malfunctions caused by Electric
Magnetic Field (EMF) near the system.
Please use the instrument as intended. Improper use may cause damage to the instrument, inaccurate results, or potentially nullify warranties.

Symbols and Definitions

CritSpin Symbols GlossaryTable 1
Symbol Description
Safety Notice
Symbols and Definitions
Consult instructions for use
Indicates the need for the user to consult the instructions for use.
ISO 15223-1. Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General Requirements. #5.4.3
Caution
Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot, for a variety of reasons, be presented on the medical device itself.
ISO 15223-1. Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General Requirements. #5.4.4
Direct current
To indicate on the rating plate that the equipment is suitable for direct current only; to identify relevant terminals.
IEC 60417: Graphical symbols for use on equipment - Overview and application, #5031
Manufacturer
Indicates the medical device manufacturer as defined in EU Directives 90/385/ EEC, 93/42/EEC and 98/79/EC.
ISO 15223-1. Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General Requirements. #5.1.1
Supplemental Product-Specific Manufacturer Information
This symbol identifies who the legal manufacturer of the product is.
55-003823-001CC vii
Page 8
Safety Notice
Symbols and Definitions
Table 1
CritSpin Symbols Glossary (Continued)
Symbol Description
Date of Manufacture
To indicate the date when the medical device was manufactured.
ISO 15223-1. Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General Requirements. #5.1.3
In vitro diagnostic medical device
Indicates a medical device that is intended to be used as an in vitro diagnostic medical device.
ISO 15223-1: Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. General requirements, clause 5.5.1
Catalogue Number
Indicates the manufacturer's catalogue number so that the medical device can be identified.
ISO 15223-1. Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General Requirements. #5.1.4
Serial number
Indicates the manufacturer's serial number so that a specific medical device can be identified.
ISO 15223-1. Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General Requirements. #5.1.7
Country of Origin Symbol
This symbol indicates the country that the product was manufactured.
Polarity of d.c. power connector
To identify the positive and negative connections (the polarity) of a d.c. power supply, or the positive and negative connections on a piece of equipment to which a d.c. power supply may be connected.
IEC 60417: Graphical symbols for use on equipment - Overview and application, #5926
viii 55-003823-001CC
Page 9
Safety Notice
Symbols and Definitions
Table 1
CritSpin Symbols Glossary (Continued)
Symbol Description
Warning; Biological hazard
To warn of a biological hazard.
IEC 60878. Graphical Symbols for electrical equipment in medical practices. #7010-W009
Supplemental Product-Specific Manufacturer Information
This label indicates a caution to operate only with all covers in position to decrease risk of personal injury or biohazard.
This label indicates the use of biohazardous materials in the area. Use caution when working with possible infectious samples.
Wear Personal Protective Equipment (PPE) such as gloves, eye shields, and lab coats. Handle and dispose of biohazardous materials according to your laboratory procedures.
California Proposition 65
This product can expose you to chemicals known to the State of California to cause Cancer and Reproductive Harm. For more information go to
https://www.P65Warnings.ca.gov.
55-003823-001CC ix
Page 10
Safety Notice
Symbols and Definitions
x 55-003823-001CC
Page 11
Revision History , iii
Safety Notice, v
CHAPTER 1: System Overview, 1-1
Intended Use, 1-1
Description, 1-1
CHAPTER 2: Operating Instructions, 2-1
Operation, 2-1
Read Spun Hematocrit Tubes inside Rotor, 2-1
Use 75 mm Adapter, 2-3
Readout Setting, 2-3
Cleaning, 2-4
Troubleshooting and Maintenance, 2-4
APPENDIX A: Specifications, A-1
Specifications , A-1
Accessories List, A-1

Contents

APPENDIX B: References, B-1
References, B-1
55-003823-001CC
xi
Page 12
Contents
xii 55-003823-001CC
Page 13

Intended Use

The StatSpin CritSpin Digital Reader assists the operator in measuring spun hematocrits from the StatSpin Centrifuges. A tube adapter is also provided to measure spun hematocrits in 75 mm tubes from other hematocrit centrifuges.
For in vitro diagnostic use.

Description

The CritSpin Digital Reader operates from the switching power supply (PN X01-003553-001) provided with the digital reader. Plug the power supply into the digital reader (Digital Reader (PN CSD2).)

CHAPTER 1

System Overview

Important
Use only the AC Power Adapter supplied with the digital reader. Use of another power supply will damage the Reader and void the warranty.
Figure 1.1
Reader Configuration Product No.: CSD2
55-003823-001CC
1-1
Page 14
System Overview
Description
1-2 55-003823-001CC
Page 15

Operation

To use this digital reader, hematocrit tubes are filled with blood, sealed and centrifuged in the RH12 hematocrit rotor. The procedure and step-by-step drawings for filling hematocrit tubes can be found in your centrifuge Instructions for Use manual. Normal values are located in your Instructions for Use manual, read the manual before performing a hematocrit. Follow laboratory procedures when performing any procedure on body fluids. When tubes are centrifuged, the hematocrit rotor is removed from the centrifuge and placed into the digital reader.

Read Spun Hematocrit Tubes inside Rotor

The digital reader has a tilt bar on the underside. Use the tilt bar to provide a tilted position from which to read tubes. You can use either position for reading.
1
Install the rotor in the digital reader:
a. Move the pointer all the way to the right.

CHAPTER 2

Operating Instructions

Tip
To move the pointer, gently press it down to allow a smooth gliding motion.
b. Place the rotor in the center of the digital reader.
2
Turn the rotor so that the first tube to be read is at the 3-o’clock position.
3
To initiate the first reading, press Set. The LED next to the label Move pointer to bottom of cells. Set "0" flashes.
Figure 2.1 Press and Set Buttons
55-003823-001CC
2-1
Page 16
Operating Instructions
Read Spun Hematocrit Tubes inside Rotor
1. Press button 2. Set button
Figure 2.2 LEDs that Flash
1. LED that flashes for step 3. 2. LED that flashes for step 4.
4
Move the pointer to the bottom of the red blood cells (top of the clay). Then press Set. The system displays 0 in the digital display.
The LED next to the label Move pointer to top of plasma. Set "100" flashes.
5
Move the pointer to the top of the plasma (plasma and air interface). Then press Set. The system displays 100 in the digital display.
The digital display becomes active, and numbers change as you move the pointer.
6
Move the pointer slowly to the plasma and red cell interface, and then press Press.
Note
To avoid false readings, do not press the membrane on either side of the gliding pointer arm.
7
Read the digital display, and record the number.
8
Turn the rotor for the next tube to be read, and press Set to initiate the next reading.
9
Continue until all tubes are read and recorded.
2-2 55-003823-001CC
Page 17

Use 75 mm Adapter

1
Move the pointer completely to the right. Place the tube adapter in the slot in the center of the digital reader.
2
Place the spun 75 mm hematocrit tube in tube adapter (clay on the right). Slide the hematocrit tube completely to the right. Follow steps 1 to 5 in the Read Spun Hematocrit Tubes inside Rotor to read the tube. Record the result.

Readout Setting

The digital display can be set to read either percent or decimal figures. When the digital reader is shipped, it is set in the percent mode. The mode can be changed from a percent to a decimal figure.
Operating Instructions
Use 75 mm Adapter
2
1
Remove the base plate on the underside of the digital reader by removing four Phillips screws. Set the base plate and screws aside.
2
Move the jumper (small black piece) from pins 1 + 2 to pins 2 + 3.
Figure 2.3
1. Detail A
2. Jumper
3. Percent Mode
4. Pin 3
5. Negative
6. Bulb
7. Positive
8. Fiber Optic Bundle
55-003823-001CC 2-3
Page 18
Operating Instructions

Cleaning

Cleaning
The instruction overlay on the digital reader has a protective coating. To clean the instrument, dampen an absorbent tissue with warm water and a mild detergent and wipe all surfaces.

Troubleshooting and Maintenance

The error indicator "flashing zero" indicates an error in the reading sequence. Repeat the reading. Select Set to reset and begin the reading process.
The light is the only user-serviceable part of the digital reader. To replace the light, remove the bottom plate by releasing the four screws on the underside. For more information, refer to Readout Setting figure. Gently dislodge the bulb by pulling out the bulb. Remove bulb and snap in new bulb. Replace the component to the original position.
The StatSpin CritSpin Digital Reader has no other user-serviceable parts. If a problem occurs, the operator must contact Beckman Coulter Customer support 1-800-854-3633. Disassembly of the unit voids the warranty.
2-4 55-003823-001CC
Page 19

Specifications

Product No. CSD2 (supplied with the digital reader)
Model No. S120
Electrical 24 Volts DC, 1.7 amps.
Dimensions Height: 2 3/8” 6.0 cm
Environmental Indoor use

APPENDIX A

Specifications

Power supplied by switching power supply for 100-240 VAC, 50/60 Hz.
Width: 6 7/8" 17.5 cm
Length: 6 3/8” 16.2 cm
Weight: 1.25 lbs. 0.6 kg
Altitude up to 2000m

Accessories List

X01-003553-001 Power Supply
X01-003877-001 Light Bulb for Digital Reader
X00-002227-001 RH12-1 12-position Hematocrit Rotor
Plastic and glass tubes are available from your local dealer, or Contact Beckman Coulter Customer Support 1-800-854-3633.
Operating temperature: 15°C to 32°C
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C.
Main supply voltage fluctuations not to exceed +/- 10% of the nominal voltage.
Transient overvoltages according to installation category II
Pollution degree 2
CritSpin Digital Reader AccessoriesTable A.2
PN Description
55-003823-001CC
A-1
Page 20
Specifications
Accessories List
A-2 55-003823-001CC
Page 21

References

APPENDIX B

References

1
Clinical Laboratory Standards Institute. Proposed Guideline, Protection of Laboratory Workers from Instrument Biohazards, Document I 17-P, Vol. 11, No. 15, Appendix A, Pg.
61.
2
CDC. Recommendations for prevention of HIV transmission in health care settings. MMWR (Suppl. No. 2S):2S-18S, 1987.
3
CDC. Update: Universal precautions for prevention of transmission of human immunodeficiency virus, hepatitis B virus and other blood borne pathogens in health care settings. MMWR 37:377-387, 1988.
4
Clinical Laboratory Standards Institute. Quantities and Units: SI; Committee Report. CLSI Publication C11-CR. CLSI, Villanova, PA 1983.
5
Clinical Laboratory Standards Institute. Procedure for Determining Packed Cell Volume by the Microhematocrit Method. Approved Standard. CLSI Publication H7-A. NCCLS,
Villanova, PA, 1985.
6
Miale, JB, Laboratory Medicine: Hematology, 5th Edition, C.V. Mosby Co., St. Louis, 1977.
7
Clinical Laboratory Standards Institute. Procedures for the Collection of Blood by Venipuncture, 2nd Edition; Approved Standard. CLSI Publication H3-A2. CLSI, Villanova,
PA, 1984.
8
Fairbanks, VF, Nonequivalence of Automated and Manual Hematocrit and Erythrocytic Indices, Am. J. Clin. Path. 73:55-62, 1980.
9
Meites, S., Pediatric Clinical Chemistry: 3rd Edition, Amer. Assoc. for Clin. Chem., Washington, DC. 1989.
55-003823-001CC
B-1
Page 22
References
References
B-2 55-003823-001CC