Published by the Centrifuge Instrument Systems Development Center of Beckman Coulter, Inc., Palo Alto, California 94304
Symbol
Symbol
Symbole
Símbolo
Simbolo
!
Title / Titel / Titre / Titulo / Titolo /
Dangerous voltage
Gefährliche elektrische Spannung
Courant haute tension
Voltaje peligroso
Pericolo: alta tensione
Attention, consult accompanying documents
Achtung! Begleitpapiere beachten!
Attention, consulter les documents joints
Atención, consulte los documentos adjuntos
Attenzione: consultare le informazioni allegate
On (power)
Ein (Netzverbindung)
Marche (mise sous tension)
Encendido
Acceso (sotto tensione)
Off (power)
Aus (Netzverbindung)
Arrêt (mise hors tension)
Apagado
Spento (fuori tensione)
Protective earth (ground)
Schutzleiteranschluß
Liaison à la terre
Puesta a tierra de protección
Collegamento di protezione a terra
Earth (ground)
Erde
Terre
Tierra
Scarica a terra
!
This safety notice summarizes information basic to the safe operation of the equipment
described in this manual. The international symbol displayed above is a reminder that all
safety instructions should be read and understood before installation, operation, maintenance, or repair of this instrument. When you see the symbol on other pages, pay special
attention to the safety information presented. Observance of safety precautions will also
help to avoid actions that could damage or adversely affect the performance of the
equipment.
Other symbols may also be displayed on the equipment. These are reproduced and
described on the inside of the front cover.
Safety During Installation and/or Maintenance
SAFETY NOTICE
This instrument is designed to be installed by a Beckman Coulter Field Service
representative. Installation by anyone other than authorized Beckman Coulter personnel
invalidates any warranty covering the instrument. Also, if the instrument needs to be
moved, a Beckman Coulter Field Service representative must reinstall and relevel the
instrument in its new location.
Any servicing of this equipment that requires removal of any covers can expose parts
which involve the risk of electric shock or personal injury. Make sure that the power
switch is turned off and the instrument is disconnected from the main power source,
and refer such servicing to qualified personnel.
Electrical Safety
To reduce the risk of electrical shock, this instrument uses a three-wire electrical cord and
plug to connect this equipment to earth-ground. Make sure that the matching wall outlet
receptacle is properly wired and earth-grounded.
Do not place containers holding liquid on or near the chamber door. If they spill, liquid
may get into the instrument and damage electrical or mechanical components.
Safety Against Risk of Fire
Fuses protect certain electrical circuits within this instrument against overcurrent conditions. For continued protection against the risk of fire, replace only with the same type and
rating specified.
This instrument is not designed for use with materials capable of developing flammable
or explosive vapors. Do not centrifuge such materials (for example, chloroform or ethyl
alcohol) in this instrument nor handle or store them near the centrifuge.
.
Mechanical Safety
For safe operation of the equipment, observe the following:
• Use only the Beckman Coulter rotors and accessories designed for use in this
instrument.
• Do not exceed the maximum rated speed of the rotor in use.
• NEVER attempt to slow or stop a rotor by hand.
• Do not move the centrifuge while the drive motor is spinning.
• In the event of a power failure, do not attempt to retrieve the sample from the instrument for at least one hour. Then follow the instructions for recovery of the sample
under
TROUBLESHOO TING AND MAINTENANCE
Chemical and Biological Safety
Normal operation may involve the use of solutions and test samples that are pathogenic,
toxic, or radioactive. Such materials should not be used in this instrument, however, unless
all necessary safety precautions are taken.
• Handle body fluids with care because they can transmit disease. No known test offers
complete assurance that they are free of micro-organisms. some of the most virulent—
Hepatitis (B and C) and HIV (I–V) viruses, atypical mycobacterium, and certain
systemic fungi—further emphasize the need for aerosol protection. Handle other infectious samples according to good laboratory procedures and methods to prevent spread
of disease. Because spills may generate aerosols, observe proper safety precautions for
aerosol containment. Do not run toxic, pathogenic, or radioactive materials in a rotor
without taking appropriate safety precautions. Biosafe containment should be used
when Risk Group II materials (as identified in the World Health Organization
tory Biosafety Manual
) are handled; materials of a higher group require more than one
Labora-
level of protection.
• Dispose of all waste solutions according to appropriate environmental health and safety
guidelines.
It is your responsibility to decontaminate the instrument and accessories before
requesting service by Beckman Coulter Field Service.
cations, operation, and routine operator care and maintenance.
• The following introductory pages contain the instrument specifica-
• Section 1 provides a brief description of the instrument, the
• Section 2 explains how to use the control panel to enter run
• Section 3 covers operating procedures for both kinds of runs.
• Section 4 lists possible malfunctions, together with probable
We recommend that you read this entire manual, especially the
SAFETY NOTICE
or performing maintenance on this instrument.
tions and preinstallation requirements.
operating controls, and indicators.
parameters for standard and programmed operation.
causes and corrective actions, and maintenance procedures.
and all safety-related information, before operating
vii
Introduction
CONVENTIONS
➠
NOTE
If the Optima LE-80K is used in a manner other
than specified in this manual, the safety and
performance of this instrument could be
impaired. Further, the use of any equipment
other than that intended for use by Beckman
Coulter has not been evaluated for safety. Use of
any equipment not specifically recommended in
this manual is the sole responsibility of the user.
NOTES, CAUTIONS, AND WARNINGS
Certain symbols are used throughout this manual to call attention to
safety-related and other important information. The symbols used and
their definitions are as follows:
➠
!
NOTE
Used to call attention to information that should
be followed during installation, use, and/or
servicing of the equipment.
CAUTION
Used to indicate a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury and/or mechanical damage. It is
also used to alert against unsafe practices.
viii
!
WARNING
Used whenever an action or condition may
potentially cause serious personal injury or
loss of life. Mechanical damage may also
result.
TYPOGRAPHIC CONVENTIONS
ix
Introduction
WARNING
Indicates high voltage or risk of electric
shock. Refer servicing of all areas displaying
this symbol to service personnel.
Certain typographic conventions are used throughout this manual to
distinguish names of user interface components, such as keys and
displays.
RADIO INTERFERENCE
•Key names (for example,
_78%68a
ENTER/RECALL
or ) appear in
capital letters within boxes.
•
Display names
(for example,
TEMP°C
or
SPEED-RPM
) appear
in bold type.
This instrument has been tested and found to comply with the limits
for a Class A digital device, pursuant to Part 15 of FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environ-
ment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with
this instruction manual, may cause interference to radio communica-
tions. Operation of this equipment in a residential area may cause
interference, in which case the user will be required to correct the
interference at his or her own expense.
Introduction
CANADIAN REGULATIONS
x
This digital apparatus does not exceed the Class A limits for radio
noise emissions from digital apparatus as set out in the radio interfer-
ence regulations of the Canadian Department of Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques
dépassant les limites applicables aux appareils numériques de Classe
A prescrites dans le reglement sur le brouillage radioelectrique édicté
par le Ministère des Communications du Canada.
xi
t
Introduction
M
U
U
C
A
V
89
MP–˚C
TE
ME–HR:MIN
I
T
M
P
–R
D
E
E
P
S
µ
–
M
U
U
C
A
V
ROGHOLD
P
2
t
w
CEL
E
D
EL
C
C
A
7
D
E
E
P
S
6
5
4
START
123
ME
I
T
CE
0
STOP
STOP
L
L
A
C
TEMP
E
/R
R
E
T
EN
E
V
SA
SPECIFICATIONS
CONTROL FEATURES
X 10,000
POWER OFF
Only values with tolerances or limits are guaranteed data. Values
without tolerances are informative data, without performance
guarantee.
Speed
Set speed . . . . . . . . . . . . . . . . . . . . . 1000 rpm to 80 000 rpm in increments of 100 rpm
Speed control . . . . . . . . . . . . . . . . . . actual rotor speed will be ±50 rpm of the set speed (above 1000 rpm)
Vacuum. . . . . . . . . . . . . . . . . . . . . . . . diffusion pump in series with a mechanical pump reduces chamber pressure
to below 5 microns (0.7 Pa)
Instrument Classification . . . . . . . . . R (uses all Beckman Coulter preparative rotors except Types 35 and
42.1 rotors with serial numbers 1299 or lower); not equipped for zonal or
continuous-flow operation
INSTALLATION REQUIREMENTS
!
CAUTION
Do not attempt to install or turn on the power
to the instrument. Its purchase price includes
installation by Beckman Coulter Field Service.
Installation by anyone other than authorized
Beckman Coulter personnel invalidates the
warranty.
xii
PREINSTALLATION REQUIREMENTS
Preinstallation requirements have been sent prior to shipment of the
instrument. (Copies are also attached to the shipping container.) The
following information is provided in case the ultracentrifuge must be
relocated. Contact your Beckman Coulter Field Service representa-
tive to adjust and level the instrument if it must be moved. (The pads
on each leveling leg are designed to prevent possible rotation of the
instrument in case of a rotor mishap.)
Electrical Data
Instrument rating: 220 to 240 VAC, 50 Hz, 20 A
Power line range: 180 to 264 VAC, 60 or 50 Hz (single-phase), 30 A
200 to 240 VAC, 50/60 Hz, 20 A
Introduction
To reduce the risk of electrical shock, this equipment uses a three-
wire electrical cord (1.8 m; 6 ft) and plug to connect the equipment
to earthground. In regions where the instrument is supplied with
an unterminated cord, a plug that meets local electrical and safety
requirements must be supplied. (Contact your local Beckman Coulter
office for specific information regarding these requirements.) See the
Table below for the required wire connections. Make sure that the
matching wall outlet receptacle is properly wired and earth-grounded.
This section describes major instrument components.
STANDBY MODE
A circuit breaker, labeled I (on) and O (off), is located on the right
side of the instrument and controls electrical power to the instrument.
Power may be left ON at all times (see STANDBY MODE below)
except in the case of an emergency or when maintenance is required.
A red POWER OFF button is provided on the neck of the control
head for your convenience (see Figure 1-1). Pressing the POWER
OFF button trips the circuit breaker to the off position. To return
power to the instrument, the circuit breaker must be returned to the
up position.
When the instrument is at rest, it automatically goes into standby
mode. In this condition, only minimal power is used by the instrument. To restore the instrument to fully operational mode, press
any key.
CONTROL PANEL
Figure 1-1 shows the Optima LE-80K control panel, which includes a
video display and keys for entering run information.
1-1
Description
Location of
Diagnostic
Messages
Parameter
Keys
SPEED–RPMTIME–HR:MINTEMP–˚C
VACUUM–µ
2
tDECELACCEL
X 10,000
SAVEPROGHOLDw
Figure 1-1. The Control Panel
Parameter
KeysKeypad
SPEED
TIME
TEMP
789
456
123
0
ENTER/RECALL
POWER OFF
Activation
Keys
VACUUM
START
CE
STOP
HARDKEYS
DISPLAYS
•Parameter hardkeys are used to enter the run conditions. They are
_74))(a, _8-1)a, and _8)14a to the left of the keypad and _%'')0a,
_()')0a, , _,30(a, _463+a, and _7%:)a along the bottom of the
2
ω
t
display area.
• Activation keys control specific ultracentrifuge functions. These
are
_:%'991a,_78%68a and _7834a, located to the right of the
keypad.
The keypad is used to enter numerical values. It consists of
the numbers 0 through 9, a
ENTER/RECALL
key.
_')a (clear entry) key, and an
Digital displays indicate rotor speed, run time, rotor temperature, the
2
ω
t value (when in that mode), and the selected program number (if
used). The displays serve a dual purpose.
1-2
Actual (Current) Values
Set V alues
Description
When the power is on, the displays show actual operating conditions
at all times except when the instrument is in editing mode or when
ENTER/RECALL
has been pressed (see below).
Set values are the run conditions entered by the user. When you press
a parameter key, the associated display blinks to indicate that a value
can be entered or modified. This is referred to as the “editing” mode.
The display continues to blink until you press another parameter key
ENTER/RECALL
or . If you press another parameter key, the associated
display begins to blink and the set values continue to be displayed.
If you press , the instrument displays the set values for
ENTER/RECALL
5 seconds, then exits the editing mode and returns to showing actual
conditions. Set values remain in memory until new v alues are entered.
RUN STATUS INDICATORS
DIAGNOSTIC MESSAGES
Whenever you press , the instrument displays the set
ENTER/RECALL
values for 5 seconds, then returns to showing actual conditions.
Run indicators are provided in the form of green and red LEDs to
indicate the status of the instrument. Figure 1-1 shows the location of
the LEDs on the control panel and Table 1-1 describes the meaning of
each. (Unless otherwise noted, the LEDs appear above their respective keys.)
Diagnostic messages appear as red LEDs under the SPEED display
(see Figure 1-1) to alert you to conditions that may need attention.
The indicators are not visible unless an abnormal condition occurs.
The meanings of the diagnostic messages, as well as appropriate
actions to take, are discussed in Section 4,
MAINTENANCE.
TROUBLESHOO TING AND
1-3
Description
Table 1-1. Run Status Indicators
Key
Pressed
STARTA green LED on the _78%68a key lights when the key has been pressed. It blinks until the rotor
reaches set speed and then remains lit until the run ends or_7834a is pressed.
STOPA green LED on the_7834a key lights when either the key is pressed or the rotor begins to decel-
erate. It blinks until the rotor comes to a stop.
ACCEL“MAX” or “SLOW” indicates which acceleration profile has been selected.
DECEL“MAX,” “SLOW,” or “NO BRAKE” indicates which deceleration profile has been selected.
Description of Run Status Indicators
2
ω
t“
HOLD“HOLD” indicates that the ultracentrifuge is in the HOLD mode.
PROGA number (1 through 9) above the _463+a (program) key indicates the number of the stored
SAVE“SAVE” blinks to indicates that you may save the program values just entered. Press _7%:)a and
VACUUM“750,” “200,” and “<20” (microns)— located in the upper display—indicate the approximate
2
ω
t ” indicates that the ultracentrifuge is in the
program that has been selected for the run.
the values will be saved in memory under the assigned program number.
chamber pressure as the chamber is being evacuated. Only one LED will be on continuously at
any given time. Once below 20 microns the “<20” LED shines continuously until the _:%'991a key
is pressed to vent the chamber. (The instrument will typically draw vacuum to 5 microns or less.)
2
ω
t mode.
DOOR
The chamber door is made of high-strength structural steel. A solenoid interlock prevents it from being opened during operation. The
door can be opened only by using the door handle and only if the
power is on and the vacuum is off, with the chamber at atmospheric
pressure. Refer to Section 4 for instructions on accessing the chamber
to retrieve a sample in case of a power outage.
1-4
ROTOR CHAMBER
The rotor chamber is made of aluminum, coated with a chemicalresistant epoxy finish. The rotor drive spindle, radiometer,
photoelectric devices, and safety plate are visible in the bottom
of the chamber (Figure 1-2).
Description
Drive
Spindle
Safety
Plate
VACUUM SYSTEM
Photoelectric
Devices
Radiometer
Figure 1-2. The Rotor Chamber
The vacuum system is automatically activated when
and
_78%68a are pressed, or it may be turned on directly by pressing
the
_:%'991a key. Three vacuum LEDs, labeled 750, 200, and <20,
ENTER/RECALL
are located in the middle of the upper display to indicate the level of
chamber pressure in microns.
At the end of a run, the chamber vacuum must be vented (by pressing
_:%'991a) before the door can be opened. After the door is opened,
the chamber is brought to approximate room temperature to prevent
condensation from collecting in the chamber. To help keep the
chamber dry and clean, the door should be kept closed whenever
possible.
A purge system removes significant amounts of moisture from the
vacuum pump and chamber. If it takes a long time to pull a vacuum,
however, it is likely caused by excess moisture in the system. Refer
to Section 4 for information on correcting this condition.
1-5
Description
TEMPERATURE SENSING
TEMPERATURE CONTROL SYSTEM
Rotor temperature is monitored by a radiometer mounted in the
bottom of the rotor chamber (see Figure 1-2) when the chamber pressure is below 100 microns. Above 100 microns (or if the radiometer
fails), chamber temperature is measured by a thermistor mounted in
the chamber.
The instrument uses a solid-state thermoelectric refrigeration and
heating system, which eliminates the need for a conventional refrigeration system and heater. Neither Freon
1
nor water is needed; the only
coolant required is forced air from the cooling fans.
OVERSPEED SYSTEM
After the power is turned on, the temperature control system is activated when the door is closed and the vacuum system is turned on.
The rotor temperature is controlled to ±1°C of the set value. If the
temperature system malfunctions, the TEMP diagnostic message will
appear and the rotor will decelerate to a stop.
The overspeed system is a safety feature designed to ensure that the
rotor does not exceed its maximum allowable speed. This system
includes photoelectric sensors in the rotor chamber next to the drive
spindle (see Figure 1-2) and an overspeed disk on the bottom of the
rotor. Individual rotor manuals provide information on the correct
overspeed disks to be used with each rotor.
The overspeed disk has alternating light and dark sectors. As the rotor
spins, the passage of reflecting and nonreflecting sectors over the photoelectric device generates a series of pulses that are detected by electronic circuitry and processed by the software.
1-6
1
Trademark of E. I. Du Pont de Nemours & Company.
After the rotor reaches 1000 rpm, the set speed is checked against the
overspeed disk. If the set speed is greater than the maximum speed
permitted by the disk, the speed setting is automatically lowered to
the disk’s maximum speed, but the run continues without interruption. An error message is displayed to alert you to the change. (In
the
ω
ROTOR ENERGY SYSTEM
As the rotor accelerates between 15 000 and 21 000 rpm, rotor inertia
is measured and the rotor energy is calculated for the set speed. If the
calculated rotor energy is determined to be excessive, a SPEED diagnostic message will appear and the instrument will immediately shut
down, thus providing a redundant safety check.
2
ω
t mode, the time setting is recalculated to give an equivalent
2
t value.)
Description
DRIVE
NAME RATING PLATE
The instrument uses a frequency-controlled, air-cooled, direct-drive
induction motor that requires no gears or brushes. In addition, the
drive does not require an oil vacuum seal, external oil reservoir, or
continuously operating damper.
2
Externally cooled by forced air and
internally cooled by oil, the drive delivers ultra-smooth, quiet performance, with high tolerance of rotor imbalance.
A name rating plate is affixed to the rear of the instrument. Always
mention the serial number (located on the chamber door) and model
number when contacting Beckman Coulter regarding your Optima
LE-80K.
2
U.S. Pat. No. 4,846,773.
1-7
Run Preparation
This section explains how to enter run information.
2
ENTERING INFORMATION
Entering or modifying information is referred to as being in the
editing mode.
• Press the appropriate hardkey to enter or modify information or
•When you press a parameter key, the appropriate field on the
•While a field is still blinking, you can retrieve the original value
• Press
to access a specific function.
screen will begin to blink, indicating that you can enter or modify
the value in that field.
after having modified it by pressing that parameter key again.
The original value appears but the field continues to blink so that
you can either enter a new value or press
original one.
_)28)6a or another parameter key to save the entry. If the
entry is valid the field will stop blinking. If the entry is invalid,
the field will continue to blink, an error tone will sound, and a
message will be displayed to explain the error . Enter an acceptable
value as prompted.
_)28)6a to retain the
2-1
Run Preparation
STANDARD OPERATION
KEYPAD
aa
____''''))))aa
(CLEAR ENTRY)
Use the number keys, in sequence, to make numerical entries. As you
enter each digit, the digits already entered move one place to the left
in the display.
The
_')a key can clear a value after a parameter k ey has been pressed
(if the instrument is in the editing mode), or it can clear a diagnostic
message.
ENTER/RECALL
• Press
_')a once to clear a value while in the editing mode or to
clear a diagnostic message.
• Press
_')a a second time to exit from the editing mode. The field
will stop blinking and the actual values will reappear.
• Press to enter a setting into memory.
• Press and
• Press during a run to display the settings for
ENTER/RECALL
ENTER/RECALL
ENTER/RECALL
_78%68a within 5 seconds to start a run.
5 seconds. After 5 seconds, the displays show the actual conditions
once again.
• Press and
ENTER/RECALL
_7834a within 5 seconds to display the
STOP values (the values that existed when the rotor began
decelerating.)
2-2
PARAMETER KEYS
Follow the instructions below for using the parameter keys and
keypad to enter run settings.
Run Preparation
____77774444))))))))((((aa
aa
Run speed can be set between 1000 and 80 000 rpm in increments
of 100 rpm. Speeds between 0 and 1000 rpm are displayed in increments of 10 (for example, 98
displayed in increments of 100 (for example, 45 6
0 rpm). Speeds above 1000 rpm are
00 rpm).
A delayed start program function automatically sets the speed at
“0” rpm. This function is described at the end of this section.
To enter or modify the set speed:
1. Press
_74))(a. The SPEED field blinks.
2. Use the keypad to enter or modify the speed. (The last two digits
are fixed as zeros and cannot be changed by keypad entries.) If
you try to enter an invalid number—such as 98 000 rpm—the
instrument will beep twice. Press
_')a to clear the value, then
enter a valid speed.
3. Press another parameter key or to save the set
ENTER/RECALL
speed.
Example: To enter 60 000 rpm as the set speed:
Press
_74))(a.
_a, _a, _a in that order.
Press
Press another parameter key or .
ENTER/RECALL
At 1000 rpm, the set speed is checked against the overspeed disk. If
the speed setting is greater than that allowed by the overspeed disk,
the run will continue, but the rotor speed will be lowered to the maximum speed permitted by the disk. The TIME setting is also
changed to provide an equivalent
2
ω
t value, if in the
2
ω
t mode. A
SPEED diagnostic message will appear , indicating that these changes
have been made. Press
_')a to clear the message.
You may change the set speed at any time by repeating the steps in the
example. The rotor will accelerate or decelerate to the new speed.
The rotor will not accelerate beyond 3000 rpm until chamber pressure
drops below 750 microns. At that point, the rotor accelerates to set
speed.
If a speed-related malfunction occurs, the SPEED diagnostic
message will appear on the screen. Consult the Diagnostic Troubleshooting Chart in Section 4 for further details.
2-3
Run Preparation
____8888----1111))))aa
aa
The run termination method depends on the run mode—TIME,
2
ω
t,
or HOLD. The information that appears in the TIME-HR: MIN
display also depends upon the mode selected.
Run time can be set up to 99 hours and 59 minutes. During the run,
the display indicates the time remaining in hours and minutes.
(When is pressed, however, the set time is shown for
ENTER/RECALL
5 seconds.)
To enter or modify the TIME setting (in TIME mode):
1. Press
_8-1)a. The TIME field blinks.
2. Use the keypad to enter or modify the run time. As each digit is
entered, the previous digit moves one place to the left.
3. Press another parameter key or to sav e the run time.
ENTER/RECALL
Example: To enter 16 hours and 30 minutes as the
run time:
Press
_8-1)a.
_a, _a, _a, _a in that order.
Press
Press another parameter key or .
ENTER/RECALL
If you entered a number that is between 60 and 99 minutes, the
instrument automatically recalculates the time in hours and minutes.
Example: If you enter
_a _a as your run time, the
instrument will change the entry to read
1:30.
After you press and
ENTER/RECALL
_78%68a to begin the run, the
instrument checks to make sure that a time entry has been made. (The
instrument cannot proceed without a valid entry for TIME,
2
ω
t, or
HOLD.) If you have not made an entry, the TIME field blinks and you
must enter one before the run can proceed.
The time begins counting down when the rotor begins to accelerate.
When the time remaining reaches zero, the rotor automatically decelerates to a stop.
2-4
2
ω
t
(Accumulated
Centrifugal Effect)
2
The
ω
t display indicates the accumulated centrifugal effect in
radians squared per second, to three significant digits, using exponential notation. The
per second. The
2
ω
t can be set for up to 9.99 × 1014radians squared
2
ω
t begins accumulating when and
ENTER/RECALL
Run Preparation
_78%68ahave been pressed and the rotor be gins spinning; deceleration
begins when the set value is reached.
There is no separate
2
ω
t display—the
SPEED display when the instrument is in the
2
ω
t values are shown in the
2
ω
t mode and the
key is pressed.
To enter or modify the
1. Press . The SPEED field blinks.
2
ω
t
2. Use the keypad to enter an
2
ω
t setting:
2
ω
t value. An “E” remains stationary in
the hundreds place to indicate that the last two places represent the
exponent. If you enter an invalid number, the instrument will beep
twice. Press
3. Press another parameter key or to save the
value. The
Example: To enter 3.24 × 10
_')a to clear the value, then enter a valid number.
2
ω
t indicator will light.
Press .
Press
2
ω
t
_a, _a, _a, _a, _a in that order . (The display
ENTER/RECALL
12
as the
ω
2
t value:
ω
will show “3.24E12.”)
Press another parameter key or .
ENTER/RECALL
2
ω
t
2
t
____,,,,33330000((((aa
When in the
2
set
ω
t value is reached) is shown in the TIME display. (If the calcu-
2
ω
t mode, a calculation of the time remaining (until the
lation exceeds 99 hours and 59 minutes, four dashes will appear in the
TIME display.)
After the run, pressing ,
ENTER/RECALL
will show the accumulated value for
_7834a, and in succession
2
ω
t (in the SPEED display)
2
ω
t
when the rotor began to decelerate. After 5 seconds, the display
returns to showing current conditions.
2
The
ω
t LED will remain on until a dif ferent run mode is selected as a
reminder that the instrument is still in the
2
ω
t from the previous run will automatically return to zero when
ENTER/RECALL
aa
The HOLD mode is used for runs of unspecified lengths. When
ENTER/RECALL
and
_78%68a are pressed to begin a new run.
is pressed while the instrument is in this mode, the
2
ω
t mode. The accumulated
word HOLD appears in the TIME display for 5 seconds. The display
then returns to showing the time elapsed since the beginning of
the run.
2-5
Run Preparation
To select the HOLD mode:
____8888))))11114444aaaa
(Temperature)
1. Press
_,30(a. The LED above the _,30(a key lights to indicate
that the instrument is in HOLD mode, and the word HOLD
appears in the TIME display.
2. Press the next parameter key or .
After and
ENTER/RECALL
_78%68a have been pressed, the run will
continue indefinitely until you press
ENTER/RECALL
_7834a to end it. If the run
exceeds 99 hours and 59 minutes, dashes appear in the TIME display.
The temperature can be set between 0 and 40°C in increments of 1°C.
If no other value has been entered, the instrument automatically
selects 25°C as the default value for the operating temperature.
To enter or modify the TEMP setting:
1. Press
_8)14a. The TEMP field blinks.
2. Use the keypad to enter or modify the temperature setting. If you
make an invalid entry, the instrument will beep twice. Press
_')a
to clear the entry, then enter a valid temperature setting.
3. Press another parameter key or to save the
ENTER/RECALL
temperature setting.
____%%%%''''''''))))0000aaaa
(Acceleration)
Example: To enter 20°C as the run temperature:
Press _8)14a.
_a, _a in that order.
Press
Press another parameter key or .
ENTER/RECALL
You may change the set temperature at any time by repeating the
steps in the example. The rotor temperature will be adjusted
accordingly.
Actual rotor temperature, after equilibration, is controlled to ±1°C
of the set value. If the temperature control system malfunctions, the
TEMP diagnostic message will light and the rotor will decelerate
to a stop.
The instrument has two acceleration profiles: maximum and slow.
(The default setting is MAX.) To select slow acceleration, press
_%'')0a. The SLOW run status indicator will light. The key acts as a
toggle switch; press it again to select MAX acceleration. Each time
you press the
_%'')0a key, the alternate choice will light.
2-6
Run Preparation
____(((())))''''))))0000aaaa
(Deceleration)
The instrument has three deceleration profiles: maximum, slow, and
no brake. (The default setting is MAX.) To select slow deceleration,
press
as a toggle switch; press it again to select NO BRAKE and again to
select MAX deceleration. Each time you press the
next choice in the series will light.
PROGRAMMED OPERATION
and
aa
(Program)
aa
The instrument internal memory is capable of holding up to nine sets
of user-specified run conditions, referred to as programs. Each
program is stored by number and can be recalled using the keypad.
Programs are retained in memory even when the power is turned off.
You can select an existing program by either pressing the program
number (1 to 9) using the keypad, or by pressing the
repeatedly until the desired set of parameters appears. (This is useful
if you do not remember the contents of a particular program or to scan
through available programs.)
____444466663333++++aa
____7777%%%%::::))))aa
_()')0a. The SLOW run status indicator will light. The k ey acts
_()')0a key, the
_463+a key
Follow the instructions below to create a new program or to modify
an existing one.
To create or modify a program:
1. Press
_463+a. If no program has been previously selected, a dash
will appear above the
_463+a key. Actual values remain in the
other displays.
2. Use the keypad to select a program number, which replaces
the dash in the display. (Or continue pressing
_463+a until the
program number appears.) The program number blinks to indicate
that you are editing the program.
3. Use the parameter keys and keypad to enter the run settings. After
all parameters have been entered, press . The SAVE
ENTER/RECALL
indicator blinks indicating that you may save the program.
4. Press
_7%:)a and the program will be stored in memory under the
assigned number. Actual values will reappear in the displays.
2-7
Run Preparation
To select a program for use:
ACTIVATION KEYS
____::::%%%%''''999999991111aa
aa
1. Press
_463+a.
2. Press the appropriate program number.
3. Press and
ENTER/RECALL
_78%68a.
To modify a parameter during the run without affecting the program,
change the setting in the normal manner, then press .
Do NOT press
_7%:)a. The instrument will run using the modified
ENTER/RECALL
settings, but the program itself will remain unchanged. The number
above the
_463+a key will go out to indicate that the instrument is
no longer running from program memory.
Use the activation keys to control the following ultracentrifuge
functions.
The vacuum system is activated automatically when you press
ENTER/RECALL
and
_78%68a to begin a run. To evacuate the chamber
at another time, make sure the chamber door is closed, then press
_:%'991a.
As the vacuum system begins to evacuate the chamber, a series of
three LEDs (labeled 750, 200, and <20 microns) blink, remain
lighted, or go off to allow you to monitor the change in pressure (see
Table 2-1). When the pressure drops below 20 microns, the < 20 LED
will remain lit until
_:%'991a is pressed to vent the chamber at the
end of the run.
Table 2-1. Vacuum LEDs
Pressure750 LED200 LED<20 LED
750 or above
220 to 750
20 to 200
<20
blinking
lit
off
off
off
blinking
lit
off
off
off
blinking
lit
2-8
Run Preparation
When the run begins, the rotor will not accelerate past 3000 rpm until
the chamber pressure drops below 750 microns. At that time, the rotor
accelerates to set speed.
____77778888%%%%66668888aa
____7777888833334444aa
aa
To vent the chamber, press
_:%'991a. The chamber door can be
opened only when the vacuum has been completely vented and the
rotor is at rest.
aa
Press and
_78%68a, the instrument will not be activated. If this occurs, press
After you have pressed and
ENTER/RECALL
ENTER/RECALL
_78%68a to begin a run. If you press only
and
_78%68a again to begin the run.
ENTER/RECALL
_78%68a, the instrument
checks for a setting in the TIME field. If there is no entry, the instrument beeps twice and the TIME display begins blinking. Enter a
value and press . Then press and
ENTER/RECALLENTER/RECALL
_78%68a
again to begin the run.
As the rotor begins to accelerate, the green LED on the
_78%68a key
begins to blink. When the rotor reaches set speed, the LED stops
blinking and remains on until the run ends or
_7834a is pressed.
Use the _7834a key to do the following:
• Press
_7834a at any time to terminate the run. A green LED on the
key will blink to indicate that the rotor is decelerating; a tone will
sound when the rotor comes to a stop and the green LED will
turn off.
• Press and
ENTER/RECALL
run conditions that existed when the rotor began decelerating.
These are called the STOP values.
• Press ,
accumulated
ENTER/RECALL
2
ω
t at the time the rotor began decelerating. The
value is displayed in the SPEED display.
DELAYED START PROGRAM FUNCTION
The delayed start program function allows you to delay the start of a
run so that it begins and ends at the time you require—without you
being there to start the run. For example, if you require a four-hour
run to end at 8:00 am, you can program the instrument to hold the
rotor at a specified temperature and time at zero speed and then automatically begin the run at 4:00 am.
_7834a within 5 seconds to display the
_7834a, and in succession to display the
2
ω
t
2
ω
t
2-9
Run Preparation
ENTERING PROGRAM 0
To use the delayed start program function, a special program (called
program “0”) is created, followed by a second program that contains
the run conditions for the actual run. Follow the instructions below.
The first program—program 0—is used to delay the start of the run
until the required time. Enter the settings as follows.
1. Press
_463+a then _a on the keypad to access program 0. A
blinking zero appears above the
for program 0 are displayed.
2. Use the parameter keys and keypad to enter time and temperature
settings as required. You cannot change the speed setting, which is set at zero. (If you try to change it, the instrument will signal
an error.)
3. Press . The SAVE message blinks to indicate that
ENTER/RECALL
you can save the program settings.
4. Press
_7%:)a to save program 0.
5. To exit from the program mode, press
RUNNING PROGRAM 0 WITH A SECOND PROGRAM
To run the delayed start program function, do the following.
_463+a key. The existing settings
_')a twice.
2-10
1. Press
_463+a then _a on the keypad to access program 0. A
blinking zero appears above the
_463+a key, and the display shows
the program 0 settings.
2. Use the keypad to select a second program from the stored
programs. The selected program settings appear in the display
and the program number appears in the
ENTER/RECALL
.
3. To begin the delayed start program function, press
and
_78%68a.
_463+a display. Press
ENTER/RECALL
The word “delay” appears in the SPEED display and the second
program number you selected appears in the
_463+a display.
Run Preparation
When the elapsed time in program 0 reaches zero, the second
program begins automatically.
Example: To run programs 0 and 3:
➠
_463+a, _a, _a, and .
Press
Press and
NOTE
ENTER/RECALL
If you change a run setting during program 0, the
second program is automatically canceled. The
instrument will stop at the end of
program 0.
ENTER/RECALL
_78%68a.
2-11
Run Procedure
This section summarizes standard and programmed operation as well
as the delayed start function. Use any of the acceptable Beckman
Coulter rotors, consulting the applicable rotor manual and Rotors and Tubes for Preparative Ultracentrifuges (LR-IM) for instructions
on preparing the rotor and accessories for centrifugation. The
Optima LE-80K is not equipped for zonal or continuous-flow operation. For fast temperature equilibration, refrigerate or warm the
rotor to the required temperature before the run.
3
!
WARNING
!
WARNING
Do not use the Optima LE-80K in the vicinity
of flammable liquids or vapors, and do not
run such materials in the instrument. Do not
lean on the instrument or place items on it
while it is operating.
Normal operation may involve the use of solutions and test samples that are pathogenic,
toxic, or radioactive. Operator error or tube
failure may generate aerosols. Do not run
toxic, pathogenic, or other hazardous
materials in this instrument unless you take
all appropriate safety precautions. Ask your
laboratory safety officer to advise you about
the level of containment required for your
application and the proper decontamination
or sterilization procedures to follow in the
event that fluids escape from containers.
3-1
Run Procedure
STANDARD OPERATION
The power must be turned on and the vacuum system turned off in
order to open the chamber door.
1. With the power on, open the chamber door, install the rotor, and
close the chamber door. Press
_:%'991a if you want to evacuate
the chamber now.
2. Enter the required run conditions (SPEED, TEMP, and TIME,
HOLD, or
2
ω
t).
3. Select ACCEL and DECEL profiles if required. Otherwise, the
instrument automatically selects maximum acceleration and
deceleration rates.
4. Press and
Press
for any reason). Runs in the timed and
ENTER/RECALL
ically activated now, unless you pressed
_7834a to terminate a run in the HOLD mode (or to stop any run
_78%68a. (The vacuum system is automat-
_:%'991a earlier.)
2
ω
t modes terminate automatically and the instrument sounds a tone to indicate the end of the run.
After the rotor has stopped, press
_:%'991a to vent the chamber.
Open the chamber door and remove the rotor. To keep the rotor
chamber clean and dry, keep the door closed between runs.
To repeat a run, press and
ENTER/RECALL
_78%68a.
3-2
PROGRAMMED OPERATION
1. With the power on, open the chamber door, install the rotor, and
close the chamber door. Press
the chamber now.
2. Press
press
3. Press and
system is automatically activated now, unless you pressed
_:%'991a earlier.)
_:%'991a if you want to evacuate
_463+a and use the keypad to select a program number. (Or
_463+a until the desired program appears.)
ENTER/RECALL
_78%68a to begin the run. (The vacuum
Run Procedure
Any of the run conditions can be changed while a programmed run is
in progress. However, such changes will affect only the current run.
To change a parameter during the run:
1. Press the appropriate parameter key (for example,
2. Enter the new value.
3. Press .
ENTER/RECALL
The run in progress will be modified accordingly.
Runs in the timed and
ω
set value is reached. After the rotor has stopped, press
vent the chamber.
Open the chamber door, remov e the rotor, and close the door. To keep
the rotor chamber clean and dry, keep the door closed between runs.
DELAYED START PROGRAM FUNCTION
1. With the power on, open the chamber door, install the rotor, and
close the chamber door.
_74))(a).
2
t modes terminate automatically when the
_:%'991a to
POINTS TO REMEMBER
2. Press
3. Press and
_463+a. Use the keypad to press _a and then a second
program number. Press .
ENTER/RECALL
ENTER/RECALL
_78%68a.
The word “delay” appears in the SPEED display and the second
program number appears in the
_463+a display. When the elapsed
time in program 0 reaches zero, the second program begins
automatically.
• Keep the chamber door closed whenever possible. This helps to
keep the chamber clean, as well as assuring optimum performance
of the vacuum system.
• For fast temperature equilibration, cool or warm the rotor to the
required temperature before the run.
3-3
Run Procedure
• Refer to the applicable rotor manual for complete instructions on
rotor preparation, use, and care.
• Refer to Section 4,
TROUBLESHOO TING AND MAINTENANCE, if
a diagnostic message flashes on the control panel during the run.
3-4
Troubleshooting
and Maintenance
This section lists possible malfunctions, together with probable
causes and corrective actions. Maintenance procedures are also
described. Refer to the applicable rotor manual for instructions on
the care of rotors and accessories. Troubleshooting and maintenance
not covered in this manual should be handled by Beckman Coulter
Field Service.
4
NOTE
➠
It is your responsibility to decontaminate the
instrument, as well as any rotors and/or accessories, before requesting service by Beckman
Coulter Field Service.
IN CASE OF POWER FAILURE DURING THE RUN
If a power failure occurs during the run, the rotor begins to decelerate
with the brake off. If the rotor is still spinning above 1000 rpm when
power is restored, the instrument will resume operation and the
rotor will return to set speed with maximum acceleration. A PWR
diagnostic message will appear to alert you that a power outage
occurred.
If the rotor is spinning below 1000 rpm or has stopped when the
power is restored, the run automatically restarts. If the instrument is
in the TIME mode, it will automatically reset the TIME display to the original set time. In the HOLD mode, the instrument will return to set
speed and elapsed time will begin to accumulate again from zero. A PWR diagnostic message will appear to alert you that a power outage
occurred and that the run has been restarted automatically.
4-1
Troubleshooting and Maintenance
RETRIEVING YOUR SAMPLE
WARNING
!
WARNING
Any maintenance procedure requiring
removal of a panel exposes the operator to
the possibility of electrical shock and/or
mechanical injury. Theref ore, turn the power
off and disconnect the instrument from the
main power source, and refer such maintenance to service personnel.
NEVER attempt to slow or stop the rotor
by hand.
!
WARNING
The following procedure should be implemented only when absolutely necessary and
only by qualified service personnel.
If a power failure lasts for several hours, it may be necessary to
retrieve the sample from the rotor. (A rotor decelerating without the
brake may take hours to come to a complete stop.) To gain access to
the rotor, it will be necessary to remove the control head cover and
the front panel to disengage the door lock.
1. Locate the two metal latches (see Figure 4-1) at the lower outside
corners on the back of the control head. Use your fingers to
depress the latches to disengage the cover from the control head.
With both hands, lift it slightly and pull it toward you until it is
free. Set the cover aside in a safe place.
2. To remove the front panel, use a long, thin tool (about 1.5 mm in
diameter) to depress the latch (see Figure 4-2). While the latch is
depressed, use your free hand to lift the top panel until it stops,
remove the tool, then continue lifting the top panel to free the front
panel. Let the front panel tilt toward you.
4-2
3. Lift the front panel to free the bottom tabs that secure it in place.
Troubleshooting and Maintenance
S
P
E
E
D
7
8
9
VA
C
U
U
M
45
6
TIME
12
3
ST
A
R
T
0
CE
T
E
MP
E
N
T
E
R
/
R
E
C
A
L
L
S
TOP
Cover Latch
Access to
Cover Latch
Figure 4-1. Removing the Control Head Cover
!
WARNING
After removing the panel, listen carefully
for any sounds coming from the drive. Then
touch the fan housing (Figure 4-3) to feel if it
is vibrating and listen again. Do not proceed
if any sound or vibration is emitted from the
housing.
4-3
Troubleshooting and Maintenance
Top Panel Tabs
M
U
9
U
C
A
V
8
7
D
E
E
P
S
6
5
4
T
R
A
T
S
3
2
1
E
M
I
T
E
C
0
P
O
T
S
L
L
P
ECA
M
E
/R
T
R
TE
N
E
Front Panel Access Port
Front Panel Tabs
Figure 4-2. Depressing the Latch to Loosen the Front Panel
4. Vent the vacuum chamber by slowly turning the small setscrew on
the vacuum solenoid valve to the right (clockwise) until you hear
the air released into the chamber. If you hear a whining noise,
close the valve and wait, because the rotor is still spinning.
5. About 10 seconds after the hissing noise stops, turn the screw to
the left (counterclockwise) until it stops.
6. Push down on the interlock pin (Figure 4-3) with your finger while
opening the chamber door with your free hand. (If the rotor is still
spinning, close the door and wait.) Be very careful not to touch
any wires or circuitry.
4-4
Fan housing
Figure 4-3. Interior View of Instrument
Troubleshooting and Maintenance
Spring-loaded
Interlock Pin
Door
Lock
System
Vacuum
Solenoid
Screw
7. After retrieving the sample, close the chamber door, then replace
the front panel on the instrument. Insert the tabs at the bottom of
the front panel onto the lip at the base of the instrument, being
sure to align the front panel with the side panels (see Figure 4-4).
8. Lift the front edge of the top panel a few inches and insert the
upper edge of the front panel under it. Push back gently to engage
the tabs, then down to activate the latch.
9. Replace the control head by sliding it along the small ridges provided, making sure to clear the emergency PO WER OFF switch in
front. When the control head cover is correctly positioned, push it
until the latches snap shut and the cover is aligned with the rest of
the instrument panels.
!
WARNING
Do not attempt to run the instrument before
returning the front panel to its correct
position.
4-5
Troubleshooting and Maintenance
Align Panel
M
U
9
U
C
VA
8
7
D
E
E
P
S
6
45
T
R
A
T
S
3
2
1
E
M
TI
E
C
0
TOP
S
L
CAL
MP
RE
E
/
T
R
E
T
N
E
Front Panel Tabs
DIAGNOSTIC MESSAGES
Figure 4-4. Reinstalling the Front Panel
Diagnostic messages appear as red LEDs at the left side of the upper
display (see Figure 4-5) to alert you to conditions that may need your
attention. A tone will sound and the appropriate message will blink
until you press the
_')a key. The diagnostic messages will reappear if
you attempt to restart the instrument before the problem is corrected.
Some of the messages are intended to provide cautionary information
that will not shut down a run in progress. Others indicate a user error.
For example, if you left the chamber door open when you pressed
_78%68a, the DOOR message would appear to let you kno w it must be
closed.
If the associated display is flashing when a diagnostic message
appears, a shut-down malfunction has occurred. The run will come
to a stop.
See Table 4-1 to determine the nature of the error or problem,
possible causes, and recommended corrective actions. If no user
action is indicated, or the error persists, call Beckman Coulter Field
Service for assistance.
4-6
Troubleshooting and Maintenance
SPEED
IMBAL
TEMP
DOOR
Diagnostic Messages
SPEED–RPMTIME–HR:MINTEMP–˚C
DRIVE
PWR
VAC
CPU
SPEED
VACUUM–µ
2
tDECELACCEL
X 10,000
SAVEPROGHOLDw
TIME
TEMP
789
456
123
0
CE
ENTER/RECALL
POWER OFF
VACUUM
START
STOP
Figure 4-5. Location of Diagnostic Messages on the Control Panel
CLEANING
ROTOR CHAMBER
Keep instrument surfaces clean by wiping them with a cloth dampened with a mild detergent such as Beckman Solution 555™ (diluted
10 to 1 with water). If salts or other corrosive materials are used, or if
spillage occurs, wash all affected areas immediately. Do not allow
corrosive materials to dry on the instrument. (Be careful not to spill
liquid on the instrument where electrical or mechanical components
could get damaged.)
The rotor chamber is coated with epoxy resin paint, and the chamber
door O-ring is Buna N rubber. The chamber can be wiped clean with
a cloth dampened with a mild detergent such as Solution 555. The
O-ring should be cleaned with a tissue or soft cloth every 3 or 4
months. Replace the O-ring whenever it becomes worn or damaged.
Lightly coat the O-ring with silicone vacuum grease (335148) before
reinstalling it to ensure an optimum vacuum seal. (Instrument O-rings
have not been designed as bioseals for aerosol containment.)
4-7
Troubleshooting and Maintenance
Diagnostic
Message
Table 4-1. Diagnostic Troubleshooting Chart
Possible CauseUser Action
SPEEDSpeed has been set above the
maximum allowable for the rotor; or
wrong, damaged, or missing overspeed
disk on the installed rotor
TEMPTemperature control or vacuum system
malfunctioning
DRIVEAbnormal change in drive speed or
overheated drive
VACVacuum not being drawn properly• Check door O-ring for damage and dirt.
IMBALRotor imbalance (at low speeds)Check for proper rotor loading.
DOORDoor is open when the _78%68a key is
pressed
PWRLoss of power during runCheck TIME display; the run may need to be
Check set speed; check the rotor for clean, undamaged, and correct overspeed disk (see the Rotors and Tubes manual [LR-IM] for replacement
instructions).
Check the air inlet (at bottom of the front panel) for
obstructions. Call Beckman Coulter Field Service if
problem persists.
Be sure a rotor is properly installed on the spindle; if
power has failed, wait for 5 minutes for drive to cool;
check for air inlet obstruction.
• Check rotor lid O-rings for possible leakage.
• Check the vacuum oil. If milky in color, run the
vacuum system for several hours or overnight until
the oil is clear.
Be sure door is closed.
restarted or aborted.
CPUMicroprocessor malfunction or loss of
program memory
NOTE
➠
No user action. Call Beckman Coulter Field Service.
Keep the chamber door closed between runs to
keep the chamber clean and dry. If necessary,
wipe off the chamber walls with a dry cloth
before operation.
Do not place containers holding liquid near
the, chamber door. Spilled liquid can damage
electrical or mechanical components.
4-8
DECONTAMINATION
If the instrument and/or accessories are contaminated with radioactive
or pathogenic solutions, follow appropriate decontamination
procedures as determined by your laboratory safety officer. Refer
to Chemical Resistances (publication IN-175), or contact Beckman
Coulter Field Service to ensure that the decontamination method
does not damage any part of the instrument (or accessories).
STERILIZA TION AND DISINFECTION
Troubleshooting and Maintenance
The top working surface is finished with urethane paint; the sides are
finished with general purpose paint. Ethanol (70%)
both these surfaces. See Chemical Resistances for more information
regarding chemical resistance of instrument and accessory materials.
While Beckman Coulter has tested these methods and found that they
do not damage the instrument, no guarantee of sterility or disinfection
is expressed or implied. When sterilization or disinfection is a
concern, consult your laboratory safety officer regarding proper
methods to use.
STORA GE AND TRANSPORTATION
Contact Beckman Coulter Field Service for specific instructions and/
or assistance in preparing the instrument for transport or long-term
storage. Temperature and humidity requirements for storage should
meet the environmental requirements described under
TIONS at the front of this manual.
1
may be used on
SPECIFICA-
1
Flammability hazard. Do not use in or near operating ultracentrifuges.
4-9
Troubleshooting and Maintenance
SUPPLY LIST
NOTE
➠
To obtain copies of referenced publications,
contact Beckman Coulter, Inc., Technical Publications Department, 1050 Page Mill Road, Palo
Alto, CA 94304, U.S.A. (Telephone 650-8591753; Fax 650-859-1375).
Call Beckman Coulter Sales (1-800-742-2345 in the United States;
worldwide offices are listed at the back of this manual) or see Ultra-centrifuge Rotors, Tubes & Accessories (BR-8101, available at
www .beckmancoulter .com) for detailed information on ordering parts
and supplies. For your convenience, a partial list is given below.
Subject to the exceptions and upon the conditions specified
below, Beckman Coulter, Inc., agrees to correct, either by
repair, or, at its election, by replacement, any defects of
material or workmanship which develop within one (1) year
after delivery of the Optima Ultracentrifuge (the product), to
the original Buyer by Beckman Coulter, or by an authorized
representative, provided that investigation and factory inspection by Beckman Coulter discloses that such defect developed
under normal and proper use.
Some components and accessories by their nature are not
intended to and will not function for as long as one (1) year. If
any such component or accessory fails to give reasonable
service for a reasonable period of time, Beckman Coulter will
repair or, at its election, replace such component or accessory.
What constitutes either reasonable service and a reasonable
period of time shall be determined solely by Beckman Coulter.
REPLACEMENT
Any product claimed to be defective must, if requested by
Beckman Coulter be returned to the factory, transportation
charges prepaid, and will be returned to Buyer with the transportation charges collect unless the product is found to be
defective, in which case Beckman Coulter will pay all transportation charges.
Beckman Coulter makes no warranty concerning products or
accessories not manufactured by it. In the event of failure of
any such product or accessory, Beckman Coulter will give
reasonable assistance to the Buyer in obtaining from the
respective manufacturer whatever adjustment is reasonable in
light of the manufacturer’s own warranty.
Damage to the instrument while operating a rotor not of
Beckman Coulter manufacture is not covered by warranty or
service contract terms. Further, Beckman Coulter shall be
released from all obligations under all warranties either
expressed or implied, if the product covered hereby is repaired
or modified by persons other than its own authorized service
personnel, unless such repair is made by others who meet
qualifications similar to those required of Beckman Coulter’s
service personnel, or unless such repair in the sole opinion of
Beckman Coulter is minor, or unless such modification is
merely the installation of a new Beckman Coulter plug-in
component for such product.
SPECIAL DRIVE WARRANTY
below. During the drive’s second through tenth year of use
there is a prorated drive replacement price based on years of
use if the drive unit is installed, serviced, and operated in
accordance with the conditions listed below.
Drive replacement price for units not under service contract*
= current drive exchange price
years of use
----------------------------
×
10
labor and travel.+
CONDITIONS
1. The drive has been operated only within its rated speed
and temperature ranges.
2. The drive unit has not been subjected to unequal loading,
improper rotor installation, corrosion from material
spilled onto the hub or accumulated in the chamber of the
instrument.
3. The drive unit has not been disassembled, modified, or
repaired, except by Beckman Coulter personnel.
4. The drive unit was installed by a Beckman Coulter Field
Service representative.
5. The instrument in which the drive unit has been used and
operated, and its associated rotors, were manufactured by
Beckman Coulter and serviced only by Beckman Coulter
Field Service representatives.
If the above conditions are not met, the full appropriate
exchange price for the drive will be charged.
DISCLAIMER
IT IS EXPRESSLY AGREED THAT THE ABOVE
WARRANTY SHALL BE IN LIEU OF ALL WARRANTIES OF FITNESS AND OF THE WARRANTY OF MERCHANTABILITY AND THAT BECKMAN COULTER,
INC. SHALL HAVE NO LIABILITY FOR SPECIAL OR
CONSEQUENTIAL DAMAGES OF ANY KIND WHATSOEVER ARISING OUT OF THE MANUFACTURE,
USE, SALE, HANDLING, REPAIR, MAINTENANCE, OR
REPLACEMENT OF THE PRODUCT.
During the instrument warranty period (one year), there will
be no charge for drive replacement if the drive unit is installed,
serviced, and operated in accordance with the conditions listed
*For details of drive coverage with a service contract, contact your
local Beckman Coulter service representative.
Beckman Coulter Worldwide Life Science Research Division Offices
AUSTRALIA
Beckman Coulter Australia Pty Ltd
Unit D, 24 College St.
Gladesville, NSW 2111
Australia