Beckman Coulter, the stylized logo, and the Beckman
Coulter product and service marks mentioned herein are
trademarks or registered trademarks of Beckman Coulter,
Inc. in the United States and other countries.
All other trademarks, service marks, products, or services
are trademarks or registered trademarks of their
respective holders.
Contact Us
If you have any questions, contact our Customer Support
Center.
•Worldwide, find us via our website at
www.beckman.com/support/technical.
•In the USA and Canada, call us at 1-800-369-0333.
•Outside of the USA and Canada, contact your local
Beckman Coulter Representative.
Find us on the World Wide Web at: www.beckman.com
May be covered by one or more pat. - see
www.beckman.com/patents
Glossary of Symbols is available at beckman.com/
techdocs (PN C24689).
Page 3
Revision History
This document applies to the latest software listed and higher versions. When a subsequent software
version changes the information in this document, a new issue will be released.
Initial Issue, 02/08
Issue AA, 03/14
Updates were made to the document format. Changes were made to Specifications, Installing the
Zonal Interlock Accessory, Run Procedure For Pelleting Sample (Continuous Flow Cores), Standard Zonal
Core, and Zonal Bracket Kit Contents.
Issue AB, 06/19
Changes were made to CHAPTER 3, J2 and J6 Series Centrifuges, CHAPTER 6, Supply List.
PN JCFZ-IM-12AB
iii
Page 4
Revision History
iv
PN JCFZ-IM-12AB
Page 5
Safety Notice
Read all product manuals and consult with Beckman Coulter-trained personnel before attempting
to operate instrument. Do not attempt to perform any procedure before carefully reading all
instructions. Always follow product labeling and manufacturer’s recommendations. If in doubt as
to how to proceed in any situation, contact your Beckman Coulter Representative.
Alerts for Danger, Warning, Caution, Important, and Note
WARNING
WARNING indicates a potentially hazardous situation which, if not avoided, could
result in death or serious injury. May be used to indicate the possibility of
erroneous data that could result in an incorrect diagnosis.
CAUTION
CAUTION indicates a potentially hazardous situation, which, if not avoided, may
result in minor or moderate injury. It may also be used to alert against unsafe
practices. May be used to indicate the possibility of erroneous data that could
result in an incorrect diagnosis.
IMPORTANT
Following the advice in the Important adds benefit to the performance of a piece of equipment or to a
process.
NOTE NOTE is used to call attention to notable information that should be followed during installation, use,
or servicing of this equipment.
This safety notice summarizes information basic to the safe operation of the rotor described in this
manual. The international symbol displayed above is a reminder to the user that all safety
instructions should be read and understood before installation, operation, maintenance, or repair
of this equipment is attempted. When you see the symbol on other pages of this manual, pay special
attention to the specific safety information presented. Observance of safety precautions will also
help to avoid actions that could damage or adversely affect the performance of the rotor. This rotor
was developed, manufactured, and tested for safety and reliability as part of a Beckman Coulter
centrifuge/rotor system. Its safety or reliability cannot be assured if used in a centrifuge not
of Beckman Coulter’s manufacture or in a Beckman Coulter centrifuge that has been modified
without Beckman Coulter’s approval.
IMPORTANT is used for comments that add value to the step or procedure being performed.
PN JCFZ-IM-12AB
v
Page 6
Safety Notice
Alerts for Danger, Warning, Caution, Important, and Note
Handle body fluids with care because they can transmit disease. No known test offers complete
assurance that such fluids are free of micro-organisms. Some of the most virulent —Hepatitis (B and
C)) viruses, HIV (I–V, atypical mycobacteria, and certain systemic fungi—further emphasize the
need for aerosol protection. Handle other infectious samples according to good laboratory
procedures and methods to prevent spread of disease. Because spills may generate aerosols,
observe proper safety precautions for aerosol containment. Do not run toxic, pathogenic, or
radioactive materials in a rotor without taking appropriate safety precautions. Biosafe containment
should be used when Risk Group II materials (as identified in the World Health Organization
Laboratory Biosafety Manual) are handled; materials of a higher group require more than one level
of protection.
The rotor and accessories are not designed for use with materials capable of developing flammable
or explosive vapors. Such materials (such as chloroform or ethyl alcohol) should not be centrifuged
in or handled near the centrifuge.
Always connect a pressure gauge between the pump and the rotor. If pressure exceeds 25 psi during
acceleration or rises gradually during centrifugation, there is an air block in the system.
Always turn the pump off before clamping a line.
Always keep two lines to the rotor unclamped when the pump is on.
vi
PN JCFZ-IM-12AB
Page 7
Safety Notice
Alerts for Danger, Warning, Caution, Important, and Note
WARNING
The processing of large volumes of pathogenic materials in the JCF-Z rotors can
produce extremely potent concentrations. Agents that are relatively harmless in
their naturally occurring dilute state may become infectious when concentrated.
Biological hazards must be carefully assessed and precautions taken.
Consideration must be given to the air flow in the room; consult your laboratory
safety officer.
WARNING
Protect your hands when handling bowl and lid threads, which become very sharp
with prolonged use. Do not touch the rotor while the centrifuge is operating.
WARNING
Never attempt to block the open end of any pressurized tubing or coupling with
the figures or hand, as injection of liquid under the skin could occur.
CAUTION
The rotor lid is designed to have liquid support. Never run an empty or partially
empty rotor to full speed. Do not run the rotor above 10 000 rpm unless it is
completely full of liquid.
CAUTION
The seal assembly (used with the standard zonal and all continuous flow cores)
requires liquid lubrication for cooling. Do not stop liquid flow into the rotor for
longer than 2 minutes during centrifugation. Do not run the seal assembly above
3000 rpm without fluid.
PN JCFZ-IM-12AB
vii
Page 8
Safety Notice
Alerts for Danger, Warning, Caution, Important, and Note
viii
PN JCFZ-IM-12AB
Page 9
Contents
Revision History,iii
Safety Notice,v
Alerts for Danger, Warning, Caution, Important, and Note,v
CHAPTER 1:The Process of Continuous Flow and Zonal Centrifugation,1-1
Continuous Flow Centrifugation,1-1
ZONAL CENTRIFUGATION,1-3
CHAPTER 2:Description,2-1
Rotor Components,2-1
Specifications,2-6
CHAPTER 3:Assembly and Installation,3-1
Installing the Zonal Bracket (Avanti J Series Centrifuges Only),3-1
Installing the Zonal Interlock Accessory,3-1
Tools and Supplies,3-1
Tools Provided,3-1
Supplies Required But Not Provided,3-2
Determining The Flow Rate: Large And Standard Pellet Cores,4-2
Setting The Temperature,4-5
Avanti Series Centrifuges,4-5
Microprocessor-controlled J2 And J6 Series Centrifuges (Models J2-
MI, J6-MI, J2-21M, J2-MC, J6-MC, and J2-21 M/E),4-5
Analog J2 And J6 Series Centrifuges (Models J2-HS, J2-21, J2-HC, J6-
HC, and J-6B),4-6
Determining The Liquid Pressure,4-6
Determining Appropriate Gradients,4-6
CHAPTER 5:Operation,5-1
Operating Precautions,5-1
Continuous Flow Operation,5-2
Pelleting Technique,5-2
Run Procedure For Pelleting Sample (Continuous Flow Cores),5-3
Purging Air From The System (Continuous Flow Rotor Only),5-5
Density Gradient Operation (Continuous Flow Cores),5-6
Gradient Or Cushion Separation Techniques,5-6
Forming an Air Block,5-7
Run Procedure for Separating in a Gradient or Cushion
(Continuous Flow Cores),5-9
Zonal Operating Techniques And Run Procedures,5-9
Standard Zonal Core,5-9
Reorienting Gradient (Reograd) Zonal Core,5-13
Ending The Run,5-14
Continuous Flow Core Containing Pellet,5-14
Continuous Flow Core Containing Gradient or Cushion,5-18
Standard Zonal Core,5-19
Reograd Zonal Core,5-20
Flushing The Tubing Lines,5-21
CHAPTER 6:Care and Maintenance,6-1
Maintenance,6-1
x
Page 11
Cleaning,6-1
Decontamination,6-2
Sterilization And Disinfection,6-3
Inspecting The Bearings,6-3
Smoothing (Lapping) The Rotating Seal,6-4
Returning a Rotor,6-4
Supply List,6-5
CHAPTER 7:Troubleshooting,7-1
Leakage,7-1
Rotating Seal/Stationary Seal Interface,7-1
Scavenger Line,7-1
Other Areas,7-1
Mixing of Inlet and Outlet Liquid,7-1
Contents
Vibration,7-2
APPENDIX A:Assembling the Rotating Seal,A-1
Assembling the Rotating Seal,A-1
APPENDIX B:Using the JCF-Z Continuous Flow/Zonal
Rotor Bracket (Avanti J Series Centrifuges Only),B-1
Description,B-1
Zonal Bracket Kit Contents,B-2
Assembly Instructions,B-3
Step 1. Install the Rotor,B-3
Step 2. Attach the Zonal Bracket to the Centrifuge Chamber
Wall,B-3
Step 3. Insert the Tubing Through the Door Gasket,B-3
Size 16 Tubing,B-3
Sizes 14 and 15 Tubing,B-4
Step 4. Attach the Tubing to the Zonal Bracket,B-5
Step 5. Attach Tubing to the JCF-Z Seal Assembly Housing,B-6
Step 6. Attach a Scavenger Line (Optional),B-7
JCF-Z Rotor Warranty And Seal Assembly Warranty
xi
Page 12
Illustrations
Illustrations
1.1Cross Section of a Continuous Flow Rotor,1-2
1.2Cross Section of Zonal Rotor with Standard Zonal Core,1-3
2.1Rotor Components.,2-2
2.2Seal and Transfer Assemblies,2-3
2.3Identifying O-Rings. O-rings requiring occasional replacement
are shown, actual size.,2-5
3.1Tools Provided in Tool Kit 335143.,3-2
4.1Speed and Flow Rate Nomogram, Standard Pellet Core.,4-4
4.2Radius Versus Volume for Standard Pellet Core, Standard Zonal
Core, and Reograd Core.,4-7
5.1Pelleting Technique, Continuous Flow Cores,5-2
5.2Rotor Installation,5-3
5.3Line Connections, Continuous Flow Cores. Using this tubing
arrangement for basic pelleting against the rotor wall.,5-4
5.4Gradient or Cushion Separation Technique, Continuous Flow
Cores.
Technique B, described in Table 5.1, is shown.,5-8
5.5Sucrose Gradient after One Hour, Monitored by Refractometer.
Diffusion between the discontinuous layers of the step
gradient results in a near-linear gradient.,5-9
5.6Line Connections, Standard Zonal Core,5-10
5.7Operating Procedure, Standard Zonal Core,5-11
5.8Removing and Reinserting Fittings and Liquid Lines,5-12
5.9Line Connections, Reograd Zonal Core,5-13
5.10Operating Procedure, Reograd Zonal Core,5-15
5.11Lowering Rotor onto Plug. Rotor contents will spill out
as soon as the core is removed if the plug and plug O-ring
are not in place.,5-16
5.12Disassembling the Rotor. Disassemble as shown to remove
sample
(arrows show direction of tool use). See CHAPTER 6, Care and Maintenance
for instructions on inspecting, cleaning, and replacing the
bearings.,5-17
5.13Removing Small Pellet. Lift out canoe-shaped pellet container
and
scrape out pellet.,5-18
xii
Page 13
Illustrations
A.1Transfer Tube Assembly Components,A-1
A.2O-ring (870656) and Silicone Vacuum Grease (335148),A-2
A.3Fully Assembled Transfer Tube Assembly,A-3
A.4Transfer Tube Assembly and Rotating Seal
(Ready for Assembly into Rotor),A-3
A.5Transfer Tube Beveled End,A-4
A.6Correct Position of the Rotating Seal,A-4
A.7Incorrect Position of the Rotating Seal (Wrong Spring
Tension),A-5
A.8Bearing Housing Assembled,A-5
A.9Stationary Seal Housing Mated to Bearing Housing
(Seal/Bearing Housing),A-6
A.10Pushing Down on Seal/Bearing Assembly to Tension Spring,A-7
A.11Finishing Rotor Assembly,A-7
B.1Avanti J Series Centrifuge Chamber with JCF-Z Rotor and Zonal
Bracket Installed,B-1
B.2Attaching the Bracket to the Chamber Wall,B-3
B.3Inserting Size 16 Tubing through the Gasket,B-4
B.4Attaching Tubing to the Adapter,B-5
B.5Inserting the Standard-Flow Feed Fitting into the Tubing,B-5
B.6Attaching the Tubing Clamps to the Bracket,B-6
B.7Inserting Tubing and Feed Fittings into the JCF-Z Seal
Assembly,B-6
B.8Attaching a Scavenger Line,B-8
xiii
Page 14
Tables
Tables
1.1JCF-Z Continuous Flow Applications,1-2
1.2JCF-Z Zonal Applications,1-4
2.1Specifications,2-6
2.2Different Core Specifications,2-7
4.1J2 and J6 Series Centrifuge Temperature Settings. All settings
are approximate.,4-6
5.1Two Techniques for Forming an Air Block During Density
Gradient Operation,5-7
The JCF-Z rotor, which can be used in Beckman Coulter J2, J6, and Avanti J series centrifuges, has
interchangeable cores that enable it to perform both continuous flow and zonal centrifugation. This
section describes these processes and their applications.
Continuous Flow Centrifugation
In continuous flow centrifugation, solutions are pumped through the rotor continuously while it is
spinning in the centrifuge (Figure 1.1). Particles sediment out of the flowing stream, which then
emerges from the rotor as a particle-depleted effluent. This process continues until the particlecontaining capacity of the rotor is reached, or until the starting material (sample) is completely
processed. Thus the amount of sample that can be processed in one run is determined by the
sample’s volume and concentration of particles.
CHAPTER 1
Centrifugation
Continuous flow processing in the JCF-Z rotor allows efficient processing (pelleting, purification, or
concentration) of large sample volumes containing particles with 600 S or higher sedimentation
coefficients
large subcellular particles such as membranes, and mitochondria. It is also frequently used to clear
virus-containing culture media of cellular debris before final isolation of the virus.
*
(Table 1.1). Continuous flow processing is suitable for sedimenting bacteria, yeast,
PN JCFZ-IM-12AB
* The sedimentation coefficient is usually expressed in Svedberg units (S), where 1 S = 1 ´ 10-13 seconds.
1-1
Page 16
The Process of Continuous Flow and Zonal Centrifugation
Continuous Flow Centrifugation
Figure 1.1 Cross Section of a Continuous Flow Rotor
1. Edgeline Outlet
2. To Center Inlet
3. Rotating Seal Assembly
4. Lid
5. Milled Slot
7. Bowl Wall
8. Core Taper Volume
9. Core
10. Channel Leading to Central Passage
11. Bowl
6. Upper Radial Channel
Particles may be concentrated in one of three ways: by pelleting onto the wall of the rotor bowl, by
sedimenting onto a cushion of dense liquid such as concentrated sucrose, or by banding in a
gradient. Use of a cushion gradient is appropriate if your solution contains biologically active
particles that could be damaged by pelleting against the rotor wall. Banding in a gradient is
appropriate when it is necessary to separate a certain particle from contaminants of different
sedimentation velocities in the solution.
Table 1.1 JCF-Z Continuous Flow Applications
Core TypeTypical Sample Type
Large pellet
large viruses (s 600 S),
pellet, concentrate400 to 800
mitochondria,
tissue homogenates, bacteria
Standard pelletlarge viruses,
mitochondria,
pellet, purify, or
concentrate
tissue homogenates, bacteria
Small pelletmitochondria, bacteria, algae,
pellet50 to 200
clay
a. Pellet refers to sedimenting particles onto the rotor wall. Concentrate refers to sedimenting particles onto a cushion of dense
solution (e.g., sucrose). Purify refers to sedimenting particles through a step gradient.
Result
a
Recommended Pellet
Size Range (mL)
200 to 400
1-2
PN JCFZ-IM-12AB
Page 17
The Process of Continuous Flow and Zonal Centrifugation
ZONAL CENTRIFUGATION
The amount of pellet anticipated or the necessity to sediment onto a cushion or into a step gradient
determines which of the three available cores is appropriate for your application.
1
•The
•The
•The
standard core rotor configuration has a total capacity of 660 mL. Of this volume, about 400
mL can be occupie d by pellet, step gradient, or cushion. This co re is suitable if no more than 400
mL of pellet is anticipated.
large pellet core rotor configuration has a total capacity of 1250 mL, of which
approximately 800 mL can be occupied by pellet. This core is well suited for centrifuging fluids
with high solid-to-liquid ratios, when a large pellet is expected. Note that when processing a
solution of high solid-to-liquid ratio, the core fills with particles more quickly and must be
unloaded more often than during processing of a less concentrated solution.
small pellet core rotor configuration has a total capacity of 240 mL, of which about 200 mL
can be occupied by pellet. It is designed for processing large amounts of material with a low
solid-to-liquid ratio. In order to minimize resuspension of sedimented materials and to
facilitate sample recovery, this core has six individual cavities, each holding removable
containers in which the pellets are collected.
ZONAL CENTRIFUGATION
In density gradient separation, sample capacity is limited by the area in the sample cavity over
which the sample spreads. The only practical way to significantly increase the sample capacity is to
increase the area of the sample cavity. When density gradient separation is performed in a zonal
rotor, gradient and sample are loaded directly into the rotor cavity; no tubes or bottles are used
(Figure 1.2). Cylindrical or bowl-shaped zonal rotors have a capacity 50 to 100 times that of the
typical swinging bucket rotor; thus zonal centrifugation enables large quantities of sample to be
processed in much less time than in other types of rotors.
In the JCF-Z rotor, the zonal technique is useful in separating large particles, such as mammalian
viruses and subcellular organelles, from plant or animal tissue homogenates (see Table 1.2).
Figure 1.2 Cross Section of Zonal Rotor with Standard Zonal Core
1. Rotating Seal Assembly
2. Lid
3. Core
4. Bowl
PN JCFZ-IM-12AB
1-3
Page 18
The Process of Continuous Flow and Zonal Centrifugation
ZONAL CENTRIFUGATION
Two zonal cores are available with the JCF-Z: standard and reorienting gradient (also known as
reograd). These cores are divided into sector-shaped compartments by vanes attached to the core;
the rotor is enclosed by a threaded lid. A rotating seal assembly allows fluid to be pumped in and
out of the standard zonal core cavity (for loading and unloading only) while the rotor is spinning.
Table 1.2 JCF-Z Zonal Applications
Core TypeTypical Sample Type
Standard zonaltissue homogenates, small or
large viruses, chloroplasts,
bacteria, algae, whole cells
Reorienting gradient
(reograd) zonal
a. Density gradient may refer either to sedimentation velocity (rate zonal) centrifugation or to equilibrium sedimentation
(isopycnic banding) centrifugation. Concentrate refers to sedimenting particles onto a cushion of dense solution (e.g,
sucrose).
•The
standard zonal core has a capacity of 1900 mL. When this core is used, the rotor is loaded
tissue homogenates, small or
large viruses, chloroplasts,
bacteria, algae, whole cells, and
especially fragile material
density gradient
concentrate
density gradient
concentrate
Result
a
Recommended Sample
Volume (mL)
59 to 500
(1400 to 1850 mL
is gradient)
1650 (250 mL
is cushion)
50 to 500
(1250 to 1700 mL
is gradient)
1500 (250 mL
is cushion)
and unloaded at a speed of 2000 rpm. Gradient solution is pumped in at the periphery of the
rotor; the sample and overlay solutions are pumped in through the center. After
centrifugation, the rotor contents are unloaded by introducing heavy displacement fluid at the
periphery and collecting fractions at the center.
•The
reograd zonal core has a capacity of 1750 mL. When this core is used, the rotor is loaded and
unloaded at rest. To prevent mixing of the rotor contents, slow acceleration to and deceleration
from 1000 rpm is recommended. This core is recommended for use with long-stranded, fragile
materials that might be damaged by passing through a rotating seal assembly.
1-4
PN JCFZ-IM-12AB
Page 19
This section describes the JCF-Z rotor components and lists rotor specifications.
Rotor Components
The basic components of the JCF-Z rotor are the rotor bowl and lid, the rotor core, the transfer tube,
and the rotating seal assemblies. The cores are interchangeable. Continuous flow cores are available
in the following sizes: standard pellet, large pellet, and small pellet. Two types of zonal cores are
available: standard and reorienting gradient (also known as reograd). Each core, except the
continuous flow small pellet, has four baffles dividing the rotor interior into four quadrants (the
small pellet core has six cavities). Baffles in continuous flow cores are removable; in zonal cores
they are not. All of the cores have holes drilled through them to enable fluid to flow into or out of
the rotor. The transfer tube separates the sample and effluent channels of fluid flow and transfers
these liquids into and out of the rotor.
CHAPTER 2
Description
The rotor bowl and lid are made of titanium; the rotor cores are made of Noryl plastic. (Components
and materials used in the rotor assemblies are detailed in Figure 2.1 and Figure 2.2) Tygon tubing
feed lines attach to the rotor for fluid introduction and removal. A drive pin in each core seats it
into the centrifuge drive hub to prevent core slippage during acceleration. Depending on which
core is used, the fully loaded JCF-Z rotor weighs 12 to 14 kg (26 to 31 lb). The seal assembly (except
bearings and rotating seal) is warranted for 1 year; other rotor components are warranted for
7 years. See the Warranty for details.
PN JCFZ-IM-12AB
2-1
Page 20
2-2
Figure 2.1 Rotor Components.
Description
Rotor Components
PN JCFZ-IM-12AB
The seal and transfer assemblies are described in Figure 2.2. Letters indicate O-rings made
of Buna N or of Viton.
Buna N: (a) 815473, (d) 870315, (e) 870688, (f) 854519 or 870528 (Viton), and (g) 812715.
Viton: (b) 870526 x 4 and (c) 817033. See also Figure 2.3.
1. Seal Assembly
(Standard 335142; high flow
335134)
2. Transfer Assembly (part of 350585)
3. Lid (336282)
4. Small Pellet Core (350601)
5. Standard Core (335130)
6. Large Pellet Core (350641)
7. Standard Zonal Core
(335141)
8. Reorienting Gradient
Core (350585)
9. (354006)
10. (354005)
11. (354005)
12. Continuous Flow
Rotors
13. Zonal Rotors
Page 21
Figure 2.2 Seal and Transfer Assemblies
Description
Rotor Components
2
PN JCFZ-IM-12AB
2-3
Page 22
Description
Rotor Components
1. Standard Flow (SF) Seal Housing
2. Feed Fitting (Standard Flow) (335146)
3. Feed Fitting O-ring (824464)
4. Manifold
a
5. Fitting Retaining Clamp (344108)
6. Manifold Screw (870690)
7. Washer (853118)
8. Standard Flow Housing (336080)
9. Scavenger Drain Tube (336286)
10.High Flow (HF) Seal Housing Assembly
(335133)
11.Manifold O-ring (870655)
12.Manifold O-ring (868638)
13.Stationary Seal (335145)
14.Rotating Seal (335144)
15.Rotating Seal O-ring (824412)
16.Rotating Seal O-ring
(870656; in bottom of seal)
17.Sample Transfer Tube
19. Spring (870654))
20.Retaining Ring (870689)
21. Flat Washer (853391)
22.Retaining Seal Mount (336284)
23.Fitting Retaining Clamp (344108)
24.Retaining Ring (870689)
25.Feed Fitting O-ring (824464)
26.Feed Fitting (Std. Flow)(335146)
27.Housing
28.Transfer Tube Assy O-ring (870656)
29.Retaining Ring (870689)
30.Transfer Assembly (reograd only)
31. Seal Assembly (all other cores)
32.Bearing Housing (336079)
33.Flat Washer (878279, nylon)
34.Bearings screw (804063)
35.Spring Washer (870658)
36.Bearing (366190)
37.Bearing Spacer (336285)
(335120 [SF] or 335132 [HF]; and 306653)
18.Transfer Tube Assy O-ring (870656)
a. Standard-Flow Manifold consists of part numbers 870690, 853118, 336080, 870655, 868638, and 335145. High-Flow
Manifold is part of High-Flow Seal Housing Assembly, which consists of part numbers 870690, 853118, 338672,
2 x 335136, and 336286.
2-4
PN JCFZ-IM-12AB
Page 23
Figure 2.3 Identifying O-Rings. O-rings requiring occasional replacement are shown, actual size.
Description
Rotor Components
2
1. Edge of Rotor Bowl (seals lid) (870688)
2. Bottom of Rotor Bowl (854519 or 870528 [Viton])
3. Zonal Core (870315)
4. Continuous Flow Core (817033)
5. Rotor Plug (812715)
PN JCFZ-IM-12AB
6. Rotating Seal (824412)
7. Rotor Lid Stem (815473)
8. Transfer Tube Assembly and Bottom of Rotating Seal (870656)
Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data,
without guarantee.
Table 2.1 Specifications
Maximum rated speed
Minimum speed
NOTE Necessary for adequate vertical positioning of the rotor contents,
except when the reograd core is used. The reograd core is loaded and
unloaded while the rotor is at rest and is accelerated and decelerated
slowly to prevent mixing of the contents.
Materials
rotor bowl and lid
rotor cores
Maximum permissible average density of rotor contents at maximum
speed
NOTE When liquids of higher average density than this are centrifuged, the
maximum rotor speed must be reduced. Calculate the speed reduction
as follows:
Speed reductions may also be required when sedimenting unusually
heavy pellets, such as liquids containing metal or clay particles.
Maximum continuous flow rates:
with standard-flow (SF) seal assembly
with high-flow (HF) seal assembly
Allowable pH range for liquids
20 000 rpm
2 000 rpm
titanium
Noryl plastic
1.45 g/mL
45 liters/hour
100 liters/hour
pH 4 to 10
a
Acceleration/deceleration time
Maximum gauge pressure tolerated by the rotating seal
Vacuum trap for scavenger line
J2 and J6 series centrifuges
Avanti J series centrifuges
Rotor code (for microprocessor-controlled J2 and J6 series centrifuges) 20.9
a. Maximum pressure limit may be reached before maximum continuous flow rates are achieved at higher rotor speeds.
2-6
6 to 7 minutes each
30 psig (210 kPa)
a
5 psia (35 kPa)
3 psia (20 kPa)
PN JCFZ-IM-12AB
Page 25
Table 2.2 Different Core Specifications
RCF
rω
2
g
---------=
RCF1.12r
RPM
1000
-------------
2
=
Description
Specifications
2
Rotor
Type
Cont.
Flow
Core
Type
large
pellet
standard
pellet
small
r
min
(mm)
top: 56
bottom: 51
top: 76
bottom: 71
52
r
max
(mm)
89
89
81
Max RCF
(x g)*
at r
max
39 900
39 900
36 300
Rotor
Capacity
(mL)
1250
660
240
Pellet
Capacity
(mL)
800
400†
204
k Factors
(outermost r
at 20 000 rpm)
293
100
281
min
pellet
Zonal
Standard
Zonal
reograd
top: 20
bottom: 29
26
89
89
39 900
39 000
1900
1750
filled
rotor
—
—
710
779
1/2-filled
rotor‡
260
277
* Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius and
speed (rω 2) to the standard acceleration of gravity (g) according to the following formula:
where r is the radius in millimeters, ω is the angular velocity in radians per second (2 π RPM /60),
and g is the standard acceleration of gravity (9807 mm/s
2
). After substitution:
†If you are using gradients rather than pelleting, about 500 mL of the capacity will be gradient.
‡If you are only partly filling the rotor with gradient, be sure to fill the balance of the rotor with plain
buffer. Do not run an empty or partially empty rotor at full speed.
PN JCFZ-IM-12AB
2-7
Page 26
Description
Specifications
2-8
PN JCFZ-IM-12AB
Page 27
Assembly and Installation
This section describes JCF-Z components and accessories for both continuous flow and zonal operation,
and the appropriate assembly and installation procedures for each.
Installing the Zonal Bracket (Avanti J Series Centrifuges Only)
When the JCF-Z rotor is used in any of the Avanti J series centrifuges, a zonal bracket must be
installed. See APPENDIX B, Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series
Centrifuges Only) for complete information.
Installing the Zonal Interlock Accessory
CHAPTER 3
Before using the standard zonal or reograd core in a J2-HS, J2-21, J6-HC, or J-6B centrifuge, you must
install a zonal key safety interlock (343348). Instructions for installation are shipped with the
interlock.
Tools and Supplies
Tools Provided
A set of rotor assembly tools ships with the rotor (tool kit 335143). The components of this kit are
shown in Figure 3.1 and their use is described in this section. Other supplies provided by Beckman
Coulter are listed in CHAPTER 6, Care and Maintenance.
PN JCFZ-IM-12AB
3-1
Page 28
Assembly and Installation
Tools and Supplies
Figure 3.1 Tools Provided in Tool Kit 335143.
1. Rotor vise (335127)
2. Plug (335149) and O-ring (812715)
3. Large spanner wrench (870667)
NOTE See Section 3 for complete instructions on when and how to use these tools.
Supplies Required But Not Provided
•Tubing connectors for 3/16-in. tubing
•Clamps or hemostats
•Auxiliary timer
•vacuum trap for scavenger line, 5 psia (35kPa)
•Pressure gauge
•Pump that meets the following requirements: (a) has a continuously adjustable flow rate from
10 mL to 1,000 mL/min for #15 tubing or 17 mL to 1,700 mL/min for #24 tubing; (b) has a flow
rate controllable to within ±1.5%; (c) has a flow rate independent of back pressure up to 40 psi;
and (d) has pulsation less than 5% of flow rate.
We recommend the Cole-Parmer Masterflex L/S pump head model EW-07518-12 (accepts L/S 15
and L/S 24 tubing), 10 ft (3 m) of Tygon LFL L/S 24 tubing part number EW-06440-24 (flow range of
28 to 1700 mL/min), and 10 to 600 rpm standard digital drive model
EW-07523-40 (115 VAC) or EW-07523-47 (230 VAC).
4. Small spanner wrench (870666)
5. 3/16-in. T-bar hex wrench (001311)
6. Etched glass plate (870671)
3-2
Pump and pump head model numbers can change, so contact Cole-Parmer directly to receive
current information before ordering. See the Cole Parmer website at www.coleparmer.com for
additional information.
PN JCFZ-IM-12AB
Page 29
Assembling the Rotor
Step 1
Assembly is much the same for continuous flow and zonal rotor configurations; assembly steps for
both configurations are covered in steps 1 through 13 below. Some steps are for particular cores only and are so identified. Before beginning assembly, make sure that
•all parts are clean and dry, and
•the rotor vise is bolted to a sturdy table.
Step 1a
Large and Standard pellet— Put the rotor in the centrifuge. Install the four baffles in the core,
flat end up.
and install the core O-ring (817033; see Figure 2.1 and Figure 2.3).
Large pellet only: lubricate and install the four O-rings (870526; see Figure 2.3). Lubricate
Assembly and Installation
Assembling the Rotor
3
PN JCFZ-IM-12AB
3-3
Page 30
Assembly and Installation
Assembling the Rotor
Step 1b
Small pellet only— Install and align the canoe-shaped pellet containers.
NOTE
During the run, the small pellet containers may rotate slightly. This rotation can often be prevented
by applying a small patch or two of silicone vacuum grease (335148) to the outside of the container
before installing it in the core. Rotation may cause slightly less pellet to be retained, but any difference
in pellet size from container to container is not sufficient to cause rotor imbalance.
Step 1c
Standard Zonal only— Lubricate and install the core O-ring (870315, shown in Figure 2.1).
3-4
PN JCFZ-IM-12AB
Page 31
Assembly and Installation
Assembling the Rotor
Step 1d
Reorienting Gradient only— Fit the core and base together, aligning the pin in the core with the hole
in the base.
3
PN JCFZ-IM-12AB
3-5
Page 32
Assembly and Installation
Assembling the Rotor
Step 2
Coat both of the bowl O-rings with silicone vacuum grease and install them in the rotor as shown:
one in the bottom of the rotor and one around the upper edge. Set the rotor into the rotor vise
(which has been bolted to a sturdy surface). Apply Spinkote lubricant to the metal threads around
the outside of the bowl.
1. O-ring (870688)
2. O-ring (854519)
3. Vise
3-6
PN JCFZ-IM-12AB
Page 33
Step 3
Assembly and Installation
Assembling the Rotor
Install the core in the rotor bowl. Note that the “up” side of each core has a small hole in the center.
The “down” side has a large hole in the center with a metal drive pin in or across it.
3
PN JCFZ-IM-12AB
3-7
Page 34
Assembly and Installation
Assembling the Rotor
Step 4
Screw the lid onto the rotor and tighten it with the large spanner wrench. Align index marks on the
lid and rotor (the lid may be tightened past the bowl mark).
1. Large Spanner Wrench (870667)
2. Index Marks
3-8
PN JCFZ-IM-12AB
Page 35
Step 5
Assembly and Installation
Assembling the Rotor
Coat the O-rings on the transfer tube or transfer tube assembly (reograd core) with silicone vacuum
grease. Check the Tygon tubing on the transfer tube to be sure it is clean. Replace this tubing as
necessary.
3
1. Transfer Tube Assembly
2. O-ring
3. Transfer Tube
4. Tygon Tubing (306653)
5. O-ring (870656)
PN JCFZ-IM-12AB
3-9
Page 36
Assembly and Installation
Assembling the Rotor
Step 6
Step 6a
1. Spanner Wrench (870666)
2. Seal Mount (336284)
3. Bearing Assembly
4. O-ring (815473)
5. Bearing Housing (336079)
6. Nylon Washer (878279)
Standard zonal and continuous flow rotors. Assemble and align the bearing assembly
as follows:
1
Unscrew the bearing housing with the T-bar hex wrench.
2
Loosen and remove the seal mount so that you can remove the bearings.
3
Inspect the bearings and replace them, if necessary (see CHAPTER 6, Inspecting The Bearings).
4
Place the two bearings, the spacer, and the spring washer on the rotor in the order shown in
the illustration below. Install the bearing housing.
7. Bearing Screw (804063)
8. 3/16-in. T-bar Hex Wrench (001311)
9. Bearing (366190)
10. Spring Washer (870658)
11. Spacer (336285)
12. Bearing (366190)
3-10
PN JCFZ-IM-12AB
Page 37
Assembly and Installation
Assembling the Rotor
5
Screw the seal mount into place counterclockwise (it has a left-hand thread) and tighten it
with a spanner wrench (870666).
6
Tighten the bearing screw using T-bar hex wrench (001311).
CAUTION
To avoid damaging the bearings, never press on the bearing assembly with your
hand when tightening the bearing screw.
7
Loosen the seal mount slightly, spin the assembly several times, and then retighten the seal
mount firmly.
8
Loosen the bearing screw with the hex wrench and spin the assembly around again several
times. Retighten. Do not press down on the housing while tightening it.
3
NOTE
Proper alignment of the bearing assembly is essential for good rotor performance.
PN JCFZ-IM-12AB
3-11
Page 38
Assembly and Installation
Assembling the Rotor
Step 6b
Reograd zonal rotor
Coat the rotor lid stem O-ring with silicone vacuum grease and slip it onto the rotor lid shaft. Apply
Spinkote lubricant to the threads of the shaft. Insert the transfer tube into the shaft and turn it
counterclockwise onto the lid until it is firmly seated. Go to Step 9b.
1. O-ring (815473)
3-12
PN JCFZ-IM-12AB
Page 39
Step 7
Assembly and Installation
Assembling the Rotor
Standard zonal and continuous flow rotors.
Before rotor assembly can be completed, the proper seal housing assembly should be selected and
prepared. Your choice depends on the required flow rate: maximum rate for the standard assembly
is 45 liters/hour, and for the high flow, 100 liters/hour. Except for the fitting retaining clamp, it is
a permanent assembly and does not ordinarily require disassembly. (If it is damaged, replace with
new parts. See Figure 2.2
standpipe. To effectively overcome the chamber vacuum, connect the standpipe to a 5-psia (35-kPa)
vacuum trap (with J2 and J6 series centrifuges) or to a 3 psia (20-kPa) vacuum trap (with Avanti J
series centrifuges). Lubricate the threads on the seal housing base with Spinkote.
footnote for part numbers.) Use a screwdriver to remove or install the
Standard Seal Housing
Standard Zonal
Rotate retainer to remove fittings before centrifugation or to attach after. Standpipe must be
removed.
3
1. Fitting Retaining Clamp
2. Feed Fitting
3. Feed Fitting
Continuous Flow
Standpipe must be hooked up to a vacuum to keep bearings dry.
1. Standpipe
PN JCFZ-IM-12AB
3-13
Page 40
Assembly and Installation
Assembling the Rotor
High Flow Seal Housing
Continuous Flow
Standpipe must be hooked up to a vacuum to keep bearings dry.
1. Standpipe
Step 8
Standard zonal and continuous flow rotors
Check the rotating seal to be sure it is not damaged; the sealing surface should be smooth and shiny.
Lubricate O-ring (824412) with silicone vacuum grease and put it around the rotating seal. Lubricate
the two stationary seal O-rings (870656 and 868638) with silicone vacuum grease and place them in
the inside grooves of the upper seal housing. The upper seal housing containing the stationary seal
is then screwed to the left (counterclockwise) onto the lower bearing assembly.
1. Undamaged
2. O-ring (824412)
3-14
PN JCFZ-IM-12AB
Page 41
Step 9
Assembly and Installation
Assembling the Rotor
Step 9a
Standard zonal and continuous flow rotors
If the seal assembly is already assembled and aligned, assemble the rotor as shown here. Be sure that
all O-rings, as well as the threads and center shaft of the lid, are lightly coated with silicone vacuum
grease. Use the small spanner wrench to tighten the seal mount counterclockwise onto the rotor.
3
NOTE
Proper assembly of the rotating seal is critical to successful rotor operation. See APPENDIX A,
Assembling the Rotating Seal for detailed instructions.
PN JCFZ-IM-12AB
1. Rotating Seal
and O-rings
2. Small Spanner Wrench
(870666)
3. Lid and Bowl
4. Seal Housing
(from Step 7)
5. Transfer Tube and O-ring
6. Spring
7. Metal Washer
8. Seal Mount
9. Bearing Assembly (from Step 6a)
10. O-ring
11. Vise
3-15
Page 42
Assembly and Installation
Assembling the Rotor
Step 9b
Reograd zonal rotor
Install the assembled rotor on the centrifuge drive hub.
CAUTION
The centrifuge drive hub can be bent if the rotor is dropped onto it or forced
sideways. Install the rotor by centering it over the hub and carefully lowering it
straight down.
3-16
1. Zonal Seal Cap (part of assembly 350585)
PN JCFZ-IM-12AB
Page 43
Step 10
Assembly and Installation
Assembling the Rotor
Load the rotor with gradient and sample according to the appropriate procedure in CHAPTER 5,
Operation. Then unscrew the transfer tube assembly from the rotor and replace it with the zonal
seal cap.
3
1. Drive Hub
PN JCFZ-IM-12AB
3-17
Page 44
Assembly and Installation
Assembling the Rotor
Step 11
Avanti J Series Centrifuges
Refer to APPENDIX B, Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges
Only) for complete information.
J2 and J6 Series Centrifuges
Remove the access panels on the sides of the centrifuge with a Phillips screwdriver. Remove the
rubber stoppers from two or three portholes, as necessary.
Step 12
3-18
1. Rubber Stoppers
Avanti J Series
A zonal bracket is needed to support the tubing lines in Avanti J series centrifuges. Refer to
APPENDIX B, Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges Only) for
complete installation instructions.
J2 and J6 Series Centrifuges
Thread 1/8-in. inside-diameter Tygon tubing lines (144664) to the rotor through the ports as shown
(J2-21 porthole configuration shown). Tubing lines should be long enough to reach all necessary
pumps and facilities, but no longer than necessary.
PN JCFZ-IM-12AB
Page 45
Assembly and Installation
Assembling the Rotor
Use the #1, one-hole rubber stoppers provided to hold the tubing in the ports as shown. Connectors
for 3/16-in. tubing (for a tight fit) are needed on the lines as shown.
3
Step 13
1. This port not needed for JCF-Z use.
2. Tubing Connector
Assemble and install the pressure gauge between the pump and rotor.
3. Scavenger line (for continuous flow rotors only)
4. Vacuum (psia)
PN JCFZ-IM-12AB
3-19
Page 46
Assembly and Installation
Assembling the Rotor
3-20
PN JCFZ-IM-12AB
Page 47
Set system variables according to the parameters presented in this section.
RPM20 000
1.45 g/mL
density of contents
------------------------------------------ ----=
Determining the Rotor Speed
The maximum allowable run speed for the JCF-Z rotor is 20 000 rpm, as long as the mean liquid
density in t he rotor is 1 .45 g/mL or l ess. If the m ean solution density in the rotor is greater than 1.45
g/mL, the maximum allowable rotor speed must be reduced according to equation (1):
Equation (1)
Whether particles of a certain size (and all larger particles) are retained in the rotor or not depends
on a combination of flow rate and rotor speed. The speed determined with equation (1) is the
maximum speed at which the rotor may be run, and should not be exceeded. In addition, the rotor
should not be used at speeds slower than 2000 rpm (the minimum speed required for adequate
vertical positioning of the rotor contents).
CHAPTER 4
Preparing for Operation
PN JCFZ-IM-12AB
4-1
Page 48
Preparing for Operation
Fπshω
2
r
max
2
r
i
2
–
1n r
maxri
⁄()
------------------------------
,=
r
i
rtrb–
1n r
trb
⁄()
------------------------=
Determining The Flow Rate: Large And Standard Pellet Cores
Determining The Flow Rate: Large And Standard Pellet Cores
The equations below may be used to determine approximate flow rate F or rotor speed in RPM for
specific samples. Figure 4.1 has been generated from equation (3a) for the standard pellet core.
Equation (2)
whereπ = 3.412
h (height) = 8.2 cm
ω = 0.10472 x RPM
r
= 8.9 cm
max
For the standard pellet core r
= 7.6 cm, and rb = 7.1 cm, reducing ri to 7.35 cm and Equation (2) to
t
Equation (3a):
Equation (3a)
F = 2.23 x 10
-10
srRPM
2
For the large pellet core, rt = 5.6 cm and rb = 5.1 cm. Thus ri becomes 5.35 cm and Equation (2) reduces
to Equation (3b):
Equation (3b)
F = 1.69 x 10
-10
srRPM
2
F is in milliliters per minute; sr, in Svedberg units, is the sedimentation coefficient adjusted for run
conditions. Equation (3a) is illustrated in Figure 4.1for ready calculation of flow rate in the standard
pellet core.
NOTE
The above equations assume that the density and viscosity of the liquid in which the particles are
suspended are similar to water at 20°C. If this is not so, an adjusted sedimentation coefficient (sr) should
be calculated.
4-2
PN JCFZ-IM-12AB
Page 49
Preparing for Operation
s
r
D2ρpρr–()
18η
r
------------------------------=
srs
20 w,
ρpρr–
ρ
pρ20 w,
–
--------------------------
η
20 w,
η
r
--------------
=
Determining The Flow Rate: Large And Standard Pellet Cores
Before you use the nomogram or equation (3), however, sr must be calculated from the known
sedimentation coefficient or the known diameter of the particle of interest. If the diameter (D) is
known, use equation (4):
Equation (4)
whereρp = density of the particles in g/mL
ρ
= density of the liquid containing your particles in g/mL
r
η
= viscosity of the liquid in mPa•s or in cp
r
4
If the sedimentation coefficient (for example, s
) of your particle is known, it may be used —once
20,w
adjusted to the conditions of your run—to calculate sr.
Equation (5)
Equation (5) is used to adapt any known sedimentation coefficient to the conditions of your
centrifuge run. Substitute the known values for the “20,w” variables in the equation.
Once s
is known, it may be used in Figure 4.1 to determine flow rate.
r
PN JCFZ-IM-12AB
4-3
Page 50
Preparing for Operation
Determining The Flow Rate: Large And Standard Pellet Cores
Figure 4.1 Speed and Flow Rate Nomogram, Standard Pellet Core.
4-4
Connect two known scale values with a straight-edge. Read the desired information where it intersects with
the third scale. Values vary with viscosity and density: use equation (4) or (5) to determine sr before using
this figure to estimate flow rate with the standard core.
PN JCFZ-IM-12AB
Page 51
Setting The Temperature
To ensure that the rotor reaches the required temperature during the run, follow the appropriate
instructions below for the model of J centrifuge being used.
Precooling the rotor is recommended whenever temperature control of the sample is critical. Place
the rotor, with the core installed, in the centrifuge and run it at 3000 rpm for 30 minutes. For J2 and
J6 series centrifuges, use the temperature and compensation settings shown in Table 4.1. At the end
of this time, remove the rotor, complete the assembly procedures, and return it to the centrifuge.
Gradient material should also be precooled if the sample is temperature sensitive.
Temperature control will also depend on ambient conditions when the JCF-Z is run above 18 000
rpm. The lowest temperatures which the rotor can be controlled are about 5°C at 19 000 rpm and
15°C at 20 000 rpm.
Avanti Series Centrifuges
Preparing for Operation
Setting The Temperature
4
Enter the run temperature according to the instructions in your centrifuge instruction manual. No
additional input is required.
Microprocessor-controlled J2 And J6 Series Centrifuges (Models J2-MI, J6-MI,
J2-21M, J2-MC, J6-MC, and J2-21 M/E)
Enter rotor code 20.9, then follow the steps below.
1
Press the
sample temperature.
2
Find the compensation value in Table 4.1 that corresponds with the set temperature and run
speed.
NOTE
3
Press
flashes.
TEMP
key on the centrifuge control panel and then use the keypad to enter the
(
)
The temperature compensation settings in Table 4.1 are based on zonal operation. For
continuous-flow operation at low flow rates, add one to two units to these figures to compensate for
heat generated in the bearings.
COMP ADJ
(
. The word COMP flashes below the TEMPERATURE display and the display
)
PN JCFZ-IM-12AB
4
Use the keypad to enter the required compensation value. Press the
sign; pressing it again will remove the minus sign.
5
Check the temperature display. (If the entry is incorrect, press
key to enter a minus
(±)
and reenter the digits.)
(CE)
4-5
Page 52
Preparing for Operation
Determining The Liquid Pressure
6
When the entry is correct, press
NOTE
To clear a COMP ADJ entry, press (COMP ADJ), (0), and (ENTER/RECALL).
Table 4.1 J2 and J6 Series Centrifuge Temperature Settings. All settings are approximate.
ENTER/RECALL
(
.
)
Speed
(rpm)
20 000
18 000
15 000
10 000
–20°–10°2°5°10°20°40°
N
N
N
N
N
N
N
N
Required Rotor Temperature (°C)
–max
–max
–10
–6
–max
–max
–9
–6
–max
–10
–8
–5
–max
–9
–7
–4
–10
–8
–6
–3
Analog J2 And J6 Series Centrifuges (Models J2-HS, J2-21, J2-HC, J6-HC, and J-6B)
Enter the required run temperature and the appropriate temperature compensation units (see
Table 4.1) on the centrifuge control panel as follows.
1
Turn the SET knob to the required sample temperature.
2
Find the compensation value that corresponds to the required temperature and run speed. Set
the COMP dial to that setting. (Interpolate if intermediate values are required.)
Determining The Liquid Pressure
A line that delivers liquid to the rotor should be set up with a pressure gauge between the pump and
the centrifuge. At flow rates up to 250 mL/min, pressure may be as high as 10 psi.
Above 500 mL/min, 15 to 20 psi is allowable. Unless your liquid is quite viscous, higher pressures
probably indicate that air is trapped in the passages of the rotor. Trapped air can block liquid flow
and cause tubing to separate from connectors when the pump is operating. The flow rate should not
decrease with increased gauge pressure. (See CHAPTER 5, Purging Air From The System (Continuous
Flow Rotor Only) for instructions on removing an air block.)
Determining Appropriate Gradients
Figure 4.2 provides an aid to banding sample in a gradient using the standard pellet core or either
zonal core. According to the figure, the radius at which a zone bands can be determined by
4-6
PN JCFZ-IM-12AB
Page 53
Preparing for Operation
Determining Appropriate Gradients
measuring the liquid volume it takes to displace the sample out of the rotor. The illustration relates
each recovered fraction to its radial position in the rotor. For example, with the standard zonal
core, if the desired band begins to emerge when 1000 mL of heavy displacement liquid has been
pumped into the rotor, then the inner edge of the band is at a radius of 67.5 mm. The centrifugal
force (RCF) at this point can be calculated by inserting the run speed and the radius (mm) into the
equation RCF = 1.12r(RPM/1000)
Figure 4.2 Radius Versus Volume for Standard Pellet Core, Standard Zonal Core, and Reograd Core.
2
.
4
PN JCFZ-IM-12AB
4-7
Page 54
Preparing for Operation
Determining Appropriate Gradients
4-8
PN JCFZ-IM-12AB
Page 55
This section describes operating procedures for the rotor and each of its cores. It contains general
information as well as items specific to either continuous flow or zonal operation. Some steps pertain only
to a particular core or cores and are so noted; follow only those steps appropriate to the core you are using.
If a problem arises during operation, see CHAPTER 7, Troubleshooting. Use this manual in combination
with the appropriate centrifuge instruction manual for complete operating information.
Operating Precautions
WARNING
The processing of large volumes of pathogenic materials in the JCF-Z rotors can
produce extremely potent concentrations. Agents that are relatively harmless in
their naturally occurring dilute state may become infectious when concentrated.
Biological hazards must be carefully assessed and precautions taken.
Consideration must be given to the air flow in the room; consult your laboratory
safety officer.
CHAPTER 5
Operation
WARNING
Protect your hands when handling bowl and lid threads, which become very sharp
with prolonged use. Do not touch the rotor while the centrifuge is operating.
WARNING
Never attempt to block the open end of any pressurized tubing or coupling with
the figures or hand, as injection of liquid under the skin could occur.
CAUTION
The rotor lid is designed to have liquid support. Never run an empty or partially
empty rotor to full speed. Do not run the rotor above 10 000 rpm unless it is
completely full of liquid.
PN JCFZ-IM-12AB
5-1
Page 56
Operation
Continuous Flow Operation
CAUTION
The seal assembly (used with the standard zonal and all continuous flow cores)
requires liquid lubrication for cooling. Do not stop liquid flow into the rotor for
longer than 2 minutes during centrifugation. Do not run the seal assembly above
3000 rpm without fluid.
CAUTION
While the rotor is moving, do not remove any portion of the seal assembly except
the fitting retaining clamp and the two push-pull feed fittings.
NOTE
The rotor will automatically stop at the end of a run if the centrifuge timer is set to anything but HOLD.
Some JCF-Z uses require unloading while the rotor is spinning. To prevent inadvertent rotor shutdown,
always set the centrifuge to HOLD and time centrifugation with an auxiliary timer.
Continuous Flow Operation
Pelleting Technique
Although the JCF-Z can also perform separations in a gradient or cushion, the rotor is most
commonly used for pelleting. Below is a description of pelleting technique which can be used for
each of the three continuous flow cores. This technique is illustrated in Figure 5.1.
Run Procedure For Pelleting Sample (Continuous Flow Cores)
1
Calibrate the pump flow rate in the range to be used for separation.
CAUTION
If you use your pump for more than one application, always disconnect and rinse
the line in the pump and check that you have set the correct flow rate before
installing and using the pump.
Operation
Continuous Flow Operation
5
NOTE
2
Assemble the rotor according to CHAPTER 3, Assembly and Installation. Then refrigerate or
warm the rotor as appropriate prior to use.
3
Prepare buffer and sample, refrigerating or warming as appropriate.
When automatically calibrating pumps with digital controllers, flow must not be interrupted for
more than 15 seconds if the rotor is running above 2,000 RPM.
CAUTION
Do not allow rotor contents to freeze, or threads between bowl and lid may be
damaged.
4
Install the rotor in the centrifuge: set it straight down on the drive hub so that the rotor base is
flush with the lip of the drive spindle (Figure 5.2). Hook up lines according to Figure 5.3 or
Figure 5.6, and label the lines with masking tape: lower or edge fitting, upper or center fitting,
and standpipe.
Figure 5.2 Rotor Installation
PN JCFZ-IM-12AB
1. Drive spindle lip must be
even with bottom of rotor
5-3
Page 58
Operation
Continuous Flow Operation
Figure 5.3 Line Connections, Continuous Flow Cores. Using this tubing
arrangement for basic pelleting against the rotor wall.
• To fill rotor, clamp C
• To load sample, clamp B
1. Lower Fitting
2. Upper Fitting
3. Buffer
4. Sample
5. Supernatant
6. Standard and Large Pellet Cores
5
Set all centrifuge controls, including setting the timer on HOLD* and the speed to 1000 rpm. If
you are using a J2 or J6 microprocessor-controlled centrifuge, enter rotor code
20.9. See
CHAPTER 4, Preparing for Operation for establishing run parameters. (See your centrifuge
instruction manual for complete centrifuge operating instructions.)
Use the pump to pull sample up through the valve that splits the sample inlet line and the buffer
NOTE
inlet line. Sample should be drawn through the valve so that when flow is switched from buffer to
sample, air in the valve is not pumped into the rotor.
6
Begin to pump buffer into the rotor, at the flow rate to be used during the separation, as you
accelerate to 1000 rpm.
7
When you reach 1000 rpm, if air bubbles are no longer coming out of the rotor, reverse the
direction of the flow (most pumps have a flow direction switch).
8
When bubbles no longer exit the rotor, accelerate the rotor to 5000 rpm.
9
Switch the flow direction back to the forward direction.
5-4
There may be a slight amount of back pressure (approximately 5 psi) if flow rates are high. If
NOTE
the pressure is greater than 7 psi, repeat steps 7 through 9.
* Use an auxiliary timer to time centrifugation runs.
PN JCFZ-IM-12AB
Page 59
Operation
Purging Air From The System (Continuous Flow Rotor Only)
If gauge pressure exceeds 25 psi (175 kPa) during acceleration or rises gradually during
centrifugation, you have at least a partial air block. To remove the block:
a. Start decelerating the rotor to 2000 rpm.
b. Change the exit line to collect effluent, because at lower rpm sample particles will no
longer be retained.
c. Save the effluent to reintroduce after the air block has been cleared.
d. If lowering the speed does not clear the air block, reverse the flow direction as described in
Purging Air From The System (Continuous Flow Rotor Only), below.
You may want to switch flow from sample to buffer during the air cleaning operation. Before
changing the flow direction, make sure that the inlet tubing is submerged in a reservoir of
liquid (sample or buffer).
10
When the rotor reaches run speed, switch to sample flow.
11
See Ending The Run, later in this section.
5
Purging Air From The System (Continuous Flow Rotor Only)
Air in the system can cause a rise in system pressure which may eventually interfere with fluid flow.
Air can enter the system accidentally through an inlet tubing line, or can accumulate in the rotor
as a result of outgassing of the sample under centrifugal force.
At the beginning of the run, when the rotor is loaded and ready to be accelerated to operating
speed, follow the steps below to verify that there is no air in the tubing lines:
1
Accelerate the rotor to 1000 rpm.
2
Pump buffer into the center line at 50 mL/min and let the buffer circulate through the system;
then reverse the pump direction
opposite direction.
3
Switch the pump back to its original direction and let the buffer recirculate through the system.
4
Accelerate the rotor to 5000 rpm and repeat steps 2 and 3.
*
and let the buffer circulate back through the system in the
PN JCFZ-IM-12AB
* See the instructions accompanying the pump.
5-5
Page 60
Operation
Density Gradient Operation (Continuous Flow Cores)
Be sure that a pressure gauge is connected between the pump and the rotor to monitor system
pressure while the centrifuge and rotor are operating. If gauge pressure exceeds 25 psi during
acceleration or rises gradually during centrifugation, the system has an air block which is affecting
fluid flow.
To remove the block:
1
Turn valve to switch reservoirs from sample to buffer.
2
Start decelerating the rotor to 2000 rpm. (Continue with steps 3 through 5 as the rotor is
decelerating.)
3
Collect the outflow in a separate receptacle so that it can be reprocessed after the air block is
removed.
4
Pump buffer into the center line at 50 mL/min and let the buffer circulate through the system;
then reverse the pump direction and let the buffer circulate through the system in the opposite
direction.
5
Switch the pump back to its original direction, and let the buffer recirculate through the
system.
6
Begin accelerating back up to run speed.
To prevent air from entering the inlet lines:
•
Do weight the end of the tubing that sits in the reservoir so that the end doesn’t come out
accidentally.
•
Don’t let a buffer or sample reservoir run dry.
•
Don’t lift the inlet tubing line out of the buffer/sample reservoir.
Density Gradient Operation (Continuous Flow Cores)
Gradient Or Cushion Separation Techniques
The JCF-Z can separate particles in a gradient or cushion. These techniques are described in
Table 5.1 and technique B is illustrated in Figure 5.4.
5-6
PN JCFZ-IM-12AB
Page 61
Operation
Density Gradient Operation (Continuous Flow Cores)
Forming an Air Block
During unloading, the rotor contents are displaced toward the center by a very dense liquid
pumped to the rotor wall. An air block is needed to cause the displacing solution to enter at the
periphery instead of at the top of the gradient. This bubble is extremely difficult to dislodge and
thus forces the displacing liquid to enter the passages across the top of the core to the periphery of
the rotor.
Before forming the air block, begin decelerating the rotor to 2000 rpm. Then perform the following
steps:
1
Cut a piece of 1/8-in. inside-diameter tubing approximately 40 inches (1 meter) long. This
tubing must be dry on the inside
disconnect the line from the core edge, at the rotor and at the T-connector. Replace with the
dry tubing.
2
Pump dense solution at 50 mL/min. to form the air block. Watch the gauge as you pump. If it
registers 15 to 17 psi, the air block is successful. Continue pumping to displace the rotor
contents (the pressure will drop slightly to about 12 to 15 psi). If an air block did not form, stop
the pump and repeat step 1 with dry tubing.
. When the rotor reaches 2000 rpm, clamp the tubing and
5
3
Continue pumping dense liquid until the rotor contents have been displaced.
Table 5.1 Two Techniques for Forming an Air Block During Density Gradient Operation
Step 1Step 2Step 3Step 4
Technique ABuffer in, verify no air in
system, then cushion in
Technique BBuffer in, then gradient
in, light end first
In Figure 5.4, the arrow from the left indicates that solution enters the rotor from the left; the arrow
from the right indicates liquid entry from the right. Technique B is illustrated.
sample inair block (8 mL)dense solution
sample inair block (8 mL)dense solution
PN JCFZ-IM-12AB
5-7
Page 62
Operation
Density Gradient Operation (Continuous Flow Cores)
Figure 5.4 Gradient or Cushion Separation Technique, Continuous Flow Cores.
Technique B, described in Table 5.1, is shown.
5-8
PN JCFZ-IM-12AB
Page 63
Operation
Zonal Operating Techniques And Run Procedures
Run Procedure for Separating in a Gradient or Cushion (Continuous Flow Cores)
Follow steps 1 through 11 in Run Procedure For Pelleting Sample (Continuous Flow Cores), earlier in this
section, noting the following changes in steps 3 and 6 that pertain to separation in a gradient or
cushion:
STEP 3: Be sure to cool or warm the gradient or cushion as appropriate.
STEP 6: Begin to pump buffer into the rotor, at the flow rate to be used during the separation, as the
rotor accelerates to 1000 rpm. If you are using a gradient or cushion, switch from buffer to the
gradient or cushion and pump at 30 mL/min. (A typical step gradient is shown in Figure 5.5.)
Figure 5.5 Sucrose Gradient after One Hour, Monitored by Refractometer. Diffusion between the
discontinuous layers of the step gradient results in a near-linear gradient.
5
Zonal Operating Techniques And Run Procedures
Zonal operation of the JCF-Z is described below and illustrated in Figures Figure 5.7 and Figure 5.8.
WARNING
Practice the procedure shown in Figure Figure 5.8 before using this core.
Standard Zonal Core
1
Calibrate the pump flow rate in the range to be used for operation.
PN JCFZ-IM-12AB
5-9
Page 64
Operation
Zonal Operating Techniques And Run Procedures
2
Assemble the rotor according to CHAPTER 3, Assembly and Installation. Then refrigerate or
warm the rotor and solutions as appropriate prior to use.
3
Install the rotor in the centrifuge. Set it straight down on the drive hub (see Figure 5.2) so that
the rotor base is flush with the lip of the drive spindle. Hook up lines according to Figure 5.6,
and label the lines with masking tape: upper or center line, and lower or edge line.
Figure 5.6 Line Connections, Standard Zonal Core
4
Set the centrifuge to ZONAL mode using either the centrifuge keyswitch or touchscreen
interface (see your centrifuge manual for details). On analog J2 and J6 models (J2-HS, J2-21, J6HC, and J-6B), the keyswitch is located on the zonal key safety interlock accessory.
*
Earlier J-21 centrifuges have no key lock; the door may be opened when the rotor is turning by
removing a stopper from one of the chamber portholes, which releases chamber vacuum. A
stopper must also be removed to open J2 series centrifuge doors.
5
Set all centrifuge controls, including setting the timer on HOLD,† the speed to 2000 rpm, and the
temperature to 30°C. If you are using a J2 or J6 microprocessor-controlled centrifuge, enter
rotor code
6
Pump in gradient at 30 to 50 mL/min (see Figure 5.7). Then pump in the sample solution at 15
20.9.
to 25 mL/min. About 50 mL of an overlay solution (buffer) may be layered over the sample if
desired, to move the sample away from the center of rotation. Pump the overlay in at 15 to 25
mL/min.
* If you have not already installed a zonal key safety interlock, contact Beckman Coulter to order one.
Publication J-TB-025 accompanying the interlock provides complete installation instructions.
† Use an auxiliary timer to time centrifugation runs.
‡ See Griffith, O. M. and Wright, H. Resolution of components from rat liver homogenate in reorienting density gradients.
Anal. Biochem, Vol. 47, No. 2, 575-583 (June 1972) for a description of a discontinuous gradient. (Reprints available
from Spinco Business Center of Beckman Coulter.
‡
5-10
PN JCFZ-IM-12AB
Page 65
Figure 5.7 Operating Procedure, Standard Zonal Core
Operation
Zonal Operating Techniques And Run Procedures
5
PN JCFZ-IM-12AB
7
Stop the pump and clamp the lines.
5-11
Page 66
Operation
Zonal Operating Techniques And Run Procedures
8
Rotate the fitting retaining clamp as described in Figure 5.8 and remove the fittings and liquid
lines from the chamber. Reinstall stoppers in all portholes. Reset the centrifuge controls as
necessary.
Figure 5.8 Removing and Reinserting Fittings and Liquid Lines
5-12
1. Fitting Retaining Clamp (344108)
Before Accelerating to Run Speed:
Rotate the fitting retaining clamp counterclockwise to unlock the fittings and slip them out. Lift off the
retaining clamp.
After Decelerating from Run Speed:
After the seal assembly comes to a stop, slide the retaining clamp down over the seal housing and reinstall
the fittings. Turn the retaining clamp clockwise to lock it in place.
9
Close the centrifuge door.
10
Accelerate the rotor to speed and set the auxiliary timer.
11
See Ending The Run, later in this section.
PN JCFZ-IM-12AB
Page 67
Reorienting Gradient (Reograd) Zonal Core
1
Calibrate the pump flow rate in the range of 15 to 50 mL/min.
2
Assemble the rotor according to CHAPTER 3, Assembly and Installation. Then refrigerate or
warm the rotor and solutions as appropriate prior to use.
3
The rotor may be filled either in the centrifuge or in a cold room. Hookup the lines as shown in
Figure 5.9. Pump in gradient first (Figure 5.10) at 30 to 50 mL/min. Then pump in sample at 15
to 25 mL/min.
Figure 5.9 Line Connections, Reograd Zonal Core
*
Operation
Zonal Operating Techniques And Run Procedures
5
PN JCFZ-IM-12AB
4
Remove the lines and transfer tube and cap the rotor with the zonal seal cap. Install the rotor
in the centrifuge: set it straight down on the drive hub so that the rotor base is flush with the
lip of the drive spindle (see Figure 5.2). Then close the centrifuge door.
5
Set all centrifuge controls as required. If you are using a J2 or J6 microprocessor-controlled
centrifuge, enter rotor code
6
Accelerate the rotor to speed using the slowest acceleration rate possible for the centrifuge.
7
See Ending The Run, below.
* See Griffith and Wright, op. cit.
20.9.
5-13
Page 68
Operation
Ending The Run
Ending The Run
Continuous Flow Core Containing Pellet
1
Push the
2
Turn the pump flow rate down.
CAUTION
Do not turn off the pump completely until the rotor has stopped. Without some
flow through the spinning rotor, the bearing housing may overheat and shorten
the life of the bearings. Overheating the bearing housing may also dislocate the
feed fitting, resulting in significant leakage.
When the rotor has stopped, turn off the pump. Flush the tubing lines (see Flushing The Tubing
Lines, later in this section).
4
Clamp the lines. Remove the rotor from the centrifuge and set it in the vise.
5
Remove the rotating seal assembly.
6
Loosen the rotor lid one half turn, using the large spanner wrench (870667). It may be necessary
to tap the wrench handle with a hammer to loosen the lid. Remove the rotor lid.
5-15
Page 70
Operation
Ending The Run
7
8
9
Figure 5.11 Lowering Rotor onto Plug. Rotor contents will spill out
Take the rotor off the vise.
Put the plug on the vise. (If the plug is not in the vise, the rotor contents will leak as soon as you
take the core out.)
Lower the rotor onto the plug (see Figure 5.11). The plug prevents leakage from the rotor bowl,
and raises the core for easier removal.
as soon as the core is removed if the plug and plug O-ring
are not in place.
5-16
1. Plug (335149)
2. Plug O-ring (812705)
PN JCFZ-IM-12AB
Page 71
Operation
Ending The Run
10
Siphon out the liquid. Take the core out of the rotor. Scrape the pellet from the core and bowl
(small pellet, see Figure 5.13). Don’t use metal tools to lift the pellet container or to scrape out
the pellet, as they can scratch the rotor.
Figure 5.12 Disassembling the Rotor. Disassemble as shown to remove sample
(arrows show direction of tool use). See CHAPTER 6, Care and Maintenance
for instructions on inspecting, cleaning, and replacing the bearings.
5
PN JCFZ-IM-12AB
1. Small Spanner Wrench
2. Large Spanner Wrench
3. Plug/O-ring
4. Vise
5-17
Page 72
Operation
Ending The Run
Figure 5.13 Removing Small Pellet. Lift out canoe-shaped pellet container and
scrape out pellet.
1. Toothpick
11
Finish disassembling the rotor (see Figure 5.12).
12
See CHAPTER 6, Care and Maintenance for cleaning instructions.
Continuous Flow Core Containing Gradient or Cushion
1
When the auxiliary timer indicates that the run is over, turn the pump off and clamp the sample
inlet line.
2
Begin decelerating the rotor to 2000 rpm. Switch the pump to a reservoir of dense displacement
solution; purge the pump and fill lines with displacement solution. Stop the pump again and
install air-block tubing as described in Forming an Air Block, earlier in this section.
5-18
PN JCFZ-IM-12AB
Page 73
Operation
Ending The Run
3
While the rotor is spinning at 2000 rpm, pump dense displacement solution at 30 mL/min into
the rotor. This first forces the air from the dry tubing into the rotor, creating an air block as
shown in Figure 5.1. The dense solution then forces the rotor contents out. Collect fractions,
manually or with a fraction collector, as liquid exits the rotor.
4
When sample has been collected, turn off the pump. Flush the tubing lines (see Flushing The
Tubing Lines, later in this section).
5
Clamp and disconnect the lines, and stop the rotor. Remove the rotor from the chamber and set
it on the plug in the rotor vise (Figure 5.11)
6
Disassemble the rotor (Figure 5.12) and see CHAPTER 6, Care and Maintenance for cleaning
instructions.
5
Standard Zonal Core
1
Reattach feed fittings and tubing lines. Attach one line to a pump and reservoir of heavy
displacement liquid, and pump the line and fitting full of liquid. Place collection vessels at the
end of the other line.
2
At the end of the run, decelerate the rotor to 2000 rpm.
3
Remove stoppers from chamber portholes to release the chamber vacuum.
4
Open the chamber door (the centrifuge must be in ZONAL mode).
5
Install inlet and outlet tubing lines (step 1 above) and fittings by passing them through the
portholes into the chamber. Use holed stoppers to hold them just within reach of the seal
assembly.
PN JCFZ-IM-12AB
5-19
Page 74
Operation
Ending The Run
6
7
8
After the seal housing comes to a stop, install and rotate the plastic fitting retainer so that
fittings and liquid lines can be reattached.
CAUTION
Make sure there is no air in the displacement liquid line.
Attach the liquid-filled line and fitting into the lower fitting hole and the collection line into the
upper fitting hole. Rotate the plastic retainer on the lower fitting hole back so that the fittings
are locked in place. Let go of the seal housing.
Unclamp the pump line and begin pumping displacement liquid at 30 mL/min. Collect sample
as it exits.
9
When all fractions have been collected, turn off the pump. Flush the tubing lines (see Flushing
The Tubing Lines, later in this section).
10
Clamp the lines and stop the rotor. Detach tubing lines and remove the rotor from the chamber,
setting it onto the plug in the rotor vise (Figure 5.11).
11
Disassemble the rotor (Figure 5.12) and see CHAPTER 6, Care and Maintenance for cleaning
instructions.
Reograd Zonal Core
1
Decelerate the rotor (it should take approximately 5 minutes to accelerate to and decelerate
from 1000 rpm) and then push the centrifuge
2
When the rotor has stopped, remove the zonal seal cap and reinstall the transfer tube assembly.
Then attach tubing and pump to the vertical fitting and collect rotor contents through that
fitting.
STOP
(
button.
)
5-20
3
Flush the tubing lines (see Flushing The Tubing Lines, below).
4
Unscrew the transfer tube assembly and complete disassembly as appropriate (see Figure 5.12).
See CHAPTER 6, Care and Maintenance for cleaning instructions.
PN JCFZ-IM-12AB
Page 75
Flushing The Tubing Lines
After every run, flush the tygon tubing lines according to the following procedure. The rotor can
then be disassembled and the components can be cleaned and sterilized according to the
procedures in CHAPTER 6, Care and Maintenance. (After a continuous flow run in which pellet is
collected, flush the tubing before clamping the lines. After a continuous flow run using gradient or
cushion, or after a zonal run [standard or reograd core], flush the tubing after all fractions have
been collected.)
1
Set the pump at approximately 30 mL/min. Pump at least 200 mL of Solution 555, diluted
approximately 1 to 10 with water, through the system.
2
To remove the detergent, pump at least one liter of sterile distilled water through the rotor.
Stop the pump.
3
Remove the rotor from the centrifuge. Remove the lid and core (see Figure 5.13) and rinse all
parts thoroughly with distilled water to remove any remaining detergent residue. See
CHAPTER 6, Care and Maintenance for detailed cleaning instructions.
Operation
Ending The Run
5
PN JCFZ-IM-12AB
5-21
Page 76
Operation
Ending The Run
5-22
PN JCFZ-IM-12AB
Page 77
This section describes procedures for handling, inspecting, cleaning, and sterilizing the rotor. It also tells
how to return a rotor to the factory for repair, if necessary.
Maintenance
NOTE Do not use sharp tools on the rotor that could cause scratches in the rotor surface. Corrosion begins
in scratches and may open fissures in the rotor with continued use.
CHAPTER 6
Care and Maintenance
WARNING
Protect your hands when handling the bowl and lid threads, which become very
sharp with prolonged use. Do not touch the rotor while the centrifuge is operating.
Cleaning
•Periodically (at least monthly) inspect the rotor for rough spots or pitting, white powder
deposits (frequently aluminum oxide), or heavy discoloration. If any of these signs are evident,
do not run the rotor. Contact your Beckman Coulter representative for information about the
Field Rotor Inspection Program and the rotor repair center.
•Regularly lubricate the metal threads in the rotor with a thin, even coat of Spinkote lubricant
(306812). Failure to keep these threads lubricated can result in damaged threads.
•Regularly apply silicone vacuum grease (335148) to the O-rings. Replace O-rings about twice a
year or whenever worn or damaged.
•Routinely apply Spinkote lubricant (306812) to the centrifuge drive spindle hub to prevent the
rotor from sticking (not required on Avanti J series centrifuges).
Refer to Chemical Resistances (publication IN-175) for chemical compatibilities of rotor and
accessory materials. Your Beckman Coulter representative provides contact with the Field Rotor
Inspection Program and the rotor repair center.
Wash the rotor body, cores, and other rotor components immediately if salts or other corrosive materials
are used or if spillage has occurred. Do not allow corrosive materials to dry on the rotor.
Under normal use, wash the rotor frequently (at least weekly) to prevent buildup of residues.
PN JCFZ-IM-12AB
6-1
Page 78
Care and Maintenance
Decontamination
1
With the plug in place (Figure 5.11) to prevent the rotor from leaking when the core is removed,
disassemble the rotor according to Figure 5.12.
NOTE
2
After a pelleting run, remove the liquid and scrape out the pellet. See Figure 5.13 for
instructions on removing the pellet from the small pellet core. After gradient separation, the
liquid contents may be siphoned out.
3
Clean the emptied rotor with Solution 555 (diluted 10 to 1 with water). Inject the solution into
all liquid passages of the seal assembly and the core. Air-dry components thoroughly before
reassembling.
4
After each run, dismantle the rotating seal assembly, thoroughly rinse the components with
distilled water, and air-dry.
The bearing screw must be removed before the seal housing can be removed. Remove O-rings
by slipping a narrow spatula along their edges. Reograd core: unscrew the transfer tube assembly.
Clean metal threads every 6 months, or as necessary. Use a brush and concentrated Solution 555.
Rinse and dry thoroughly, then lubricate lightly but evenly with Spinkote to coat all threads.
Periodically remove the O-rings and wipe clean as necessary, then reapply a light film of silicone
vacuum grease. Clean the O-ring grooves with a cotton-tipped swab.
Decontamination
If the rotor (and/or accessories) becomes contaminated with radioactive material, decontaminate
it using a solution that will not damage the anodized surfaces. Beckman Coulter has tested a number
of solutions and found two that do not harm anodized aluminum: RadCon Surface Spray or IsoClean
Solution (for soaking),
IsoClean can cause fading of colored anodized surfaces. use it only when necessary and remove it
promptly from surfaces.
While Beckman Coulter has tested these methods and found that they do not damage components,
no guarantee of decontamination is expressed or implied. Consult your laboratory safety officer
regarding the proper decontamination methods to use.
*
and Radiacwash.†
6-2
* In U.S.A., contact Nuclear Associates (New York); in Eastern Europe and Commonwealth States, contact Victoreen
GmbH (Munich);
in South Pacific, contact Gammasonics Pty, Ltd. (Australia); in Japan, contact Toyo Medic Co. Ltd. (Tokyo).
† In U.S.A., contact Biodex Medical Systems (Shirley, NY); internationally, contact the U.S. office to find the dealer nearest
you.
PN JCFZ-IM-12AB
Page 79
Sterilization And Disinfection
WARNING
Flammability Hazard. Do not use Ethanol near operating centrifuges.
The rotor— except for the baffles on the standard and large pellet cores—can be autoclaved at 121°C
for up to 1 hour. However, autoclaving the Noryl plastic cores is not recommended. We recommend
cold sterilization methods only, such as Ethanol (70%).
the cores and of the bearings
To autoclave, empty all liquid from the rotor. Set the rotor and lid (separately) upside down in the
autoclave. Run a dry cycle at the end of the autoclave cycle.
While Beckman Coulter has tested these methods and found that they do not damage components,
no guarantee of sterility or disinfection is expressed or implied. When sterilization or disinfection
is a concern, consult your laboratory safety officer regarding proper methods to use.
.
Care and Maintenance
Sterilization And Disinfection
Repeated autoclaving will shorten the life of
6
Inspecting The Bearings
Before each use, inspect the bearings by disassembling the bearing assembly using the
3/16-in. T-bar hex wrench (001311). Hold each bearing between your thumb and forefinger and
spin it. If a bearing is noisy or spins roughly, it should be replaced. To reassemble the bearings,
follow the alignment procedure in CHAPTER 3, Step 6a.
Some JCF-Z rotors with new or older bearings may be misidentified in Avanti J series centrifuges. If
the rotor is misidentified, stop the run and follow one the steps below. (JCF-Z rotors manufactured
after March, 1997 have factory-installed magnets that prevent misidentification.)
On the first few uses of a new JCF-Z rotor, or in an older JCF-Z rotor with newly replaced bearings,
•
run the rotor from 0 to 5000 rpm and back to 0 rpm three times. This procedure will distribute
lubricant around the bearings in the rotating seal assembly, reducing drag.
•
In older JCF-Z rotors with bearings that have not been replaced recently, follow the same
procedure as for a new JCF-Z rotor (perform three runs from 0 to 5000 rpm and back to 0 rpm)
to ensure that the bearings are properly lubricated. If the rotor is misidentified after the third
run, when the bearings are lubricated, this indicates that the bearings are worn and need to be
replaced. Replace the bearings following the instructions CHAPTER 3, Step 6a.
PN JCFZ-IM-12AB
6-3
Page 80
Care and Maintenance
Smoothing (Lapping) The Rotating Seal
Smoothing (Lapping) The Rotating Seal
With normal use, the rotating seal (335144) will give several thousand hours of service. However,
the seal is made of carbon graphite and can be nicked, scratched, or unevenly worn in the presence
of abrasive samples; such damage may eventually allow leakage between the inlet and the outlet
lines. To repair damage, rub the damaged surface in a figure-eight motion against the etched glass
plate (870671) provided with the rotor. Rotate the seal as you press it lightly. The raised rings (lands)
on the seal mu st be at lea st 1.5-mm ( 1/16-in.) high for s urfaces to seal prope rly. If the l ands are wo rn
beyond this, replace the seal. Properly lapped lands should reflect light smoothly and evenly. Note
that the glass plate will eventually wear out and should be replaced.
Returning a Rotor
Before returning a rotor or accessory for any reason, prior permission (a Returned Goods
Authorization form) must be obtained from Beckman Coulter, Inc. The RGA form may be obtained
from your local Beckman Coulter sales office. It should contain the following information:
•rotor serial number,
•history of use (approximate frequency of use),
•reason for the return,
•original purchase order number, billing number, and shipping number, if possible,
•name and phone number of the person to be notified upon receipt of the rotor or accessory at
the factory, and
•name and phone number of the person to be notified about repair costs, etc.
To protect our personnel, it is the customer’s responsibility to ensure that the parts are free from
pathogens and/or radioactivity. Sterilization and decontamination must be done before returning
the parts. Smaller items (such as tubes, bottles, etc.) should be enclosed in a sealed plastic bag.
All parts must be accompanied by a note, plainly visible on the outside of the box or bag, stating that they
are safe to handle and that they are not contaminated with pathogens or radioactivity.
this notification will result in return or disposal of the items without review of the reported problem.
Use the address label printed on the RGA form when mailing the rotor and/or accessories.
Customers located outside the United States should contact their local Beckman Coulter office.
Failure to attach
6-4
PN JCFZ-IM-12AB
Page 81
Supply List
Many of the parts listed are pictured in Figure 2.1, Figure 2.2, and Figure 2.3. Refer to these Figures
for easy identification. Refer to APPENDIX B, Using the JCF-Z Continuous Flow/Zonal Rotor Bracket
(Avanti J Series Centrifuges Only) for parts required for the zonal bracket.
Table 6.1 Supply List
Bearing housing (anodized aluminum)336079
Bearing (pressed steel)355161
Bearing (stainless steel)366190
Bearing screw, hex head (stainless steel)804063
Bearing spacer (stainless steel)336285
Core Kit, continuous flow, standard pellet335130
Core Kit, continuous flow, large pellet350641
Core Kit, continuous flow, small pellet350601
Core, reorienting gradient, and sample transfer assembly350585
Care and Maintenance
Supply List
6
Core Kit, standard zonal335141
Feed fitting, high flow (stainless steel)
Feed fitting, standard flow (stainless steel)335146
Fitting retaining clamp (Noryl)334108
Housing, high flow stationary seal (stainless steel)338672
Housing, standard flow stationary seal (stainless steel)336080
Key interlock, zonal343348
Mount, rotating seal (stainless steel)336284
O-ring (Viton), continuous flow lid870526
O-ring (Viton), continuous flow lid817033
O-ring, rotating seal, standard flow manifold868638
O-ring, rotating seal824412
O-ring, rotor bowl bottom (minimum order 6)854519
O-ring, rotor bowl edge870668
O-ring, rotor lid stem (minimum order 6)815473
O-ring, rotor plug (minimum order 12)812715
O-ring, seal housing feed fitting824464
O-ring, standard flow manifold (minimum order 6)870655
a
335136
PN JCFZ-IM-12AB
O-ring (Viton), transfer tube assembly and rotating seal870656
O-ring, zonal lid870315
Retaining ring (stainless steel)870689
Rotating seal (carbon-graphite)335144
6-5
Page 82
Care and Maintenance
Supply List
Table 6.1 Supply List
Sample transfer tube, standard flow seal assembly
335120
(stainless steel)
Sample transfer tube, high flow seal assembly (stainless steel)335132
Scavenger drain tube (stainless steel)336286
Screw, manifold (stainless steel)870690
Seal assembly, high flow335133
Seal assembly kit, high flow (for continuous-flow operation)335134
Seal assembly kit, standard flow (for zonal and continuous-flow operation)335142
Rotor vise
Plug
Plug O-ring
Spanner wrench, large
Spanner wrench, small
Hex wrench, 3/16-in. T-bar
Glass plate
a. Use loctite grade “T” primer and #35 retaining compound to bond fitting to seal housing.
335127
335149
812715
870667
870666
001311
870671
6-6
PN JCFZ-IM-12AB
Page 83
This section lists possible malfunctions, together with probable causes and corrective actions required. If
a problem persists, contact Beckman Coulter (1-800-742-2345 in the United States; customers outside the
U.S. should contact their local Beckman Coulter office).
Leakage
Rotating Seal/Stationary Seal Interface
Leakage from the rotating seal/stationary seal interface can be caused by a damaged rotating seal
or by insufficient spring pressure on the rotating seal. Inspect the rotating seal for signs of wear or
damage and replace if necessary. See APPENDIX A, Assembling the Rotating Seal for detailed
instructions on assembling the rotating seal.
CHAPTER 7
Troubleshooting
Scavenger Line
Excessive leakage from the scavenger line can result from either excessive back pressure in the
system or a damaged rotating seal. Check the gauge to see if there is excessive system pressure
(about 30 psi); if so, follow the instructions in CHAPTER 5, Operation to purge air from the system.
Inspect the rotating seal for damage and replace, if necessary.
Other Areas
Leakage from under the rotor lid indicates that the lid is not fastened down properly. Check
O-rings and resecure the lid to the rotor body, making sure that the lid has been tightened to or past
the index marks (see CHAPTER 3, Step 4).
Leakage from the rotor bottom indicates that the bottom O-ring (854519) is missing. Reinstall the
O-ring, making sure that it is lightly coated with silicone vacuum grease.
Mixing of Inlet and Outlet Liquid
Such mixing is due to cross-leakage over the face of the rotating seal. Remove the rotating seal and
inspect all of its surfaces. See CHAPTER 6, Care and Maintenance for instructions on smoothing the
seal.
PN JCFZ-IM-12AB
7-1
Page 84
Troubleshooting
Vibration
Vibration
Rotor vibration can be caused by defective or misaligned bearings (see CHAPTER 3, Step 6a for
bearing alignment instructions) or by an improperly assembled rotor. Review assembly steps 1
through 13 in CHAPTER 3, Assembling the Rotor, if necessary.
7-2
PN JCFZ-IM-12AB
Page 85
Assembling the Rotating Seal
Use the following assembly procedures when assembling the rotating seal. They incorporate
instructions for proper tensioning of the rotating seal and ensure proper operation of the seal at
high speeds and high operating pressures. Following these procedures will avoid rotating seal
leakage problems, which are the main cause of reduced operational efficiency and performance of
this rotor. At properly tensioned spring pressures, the rotating seal is capable of up to two liters per
minute of sample throughput in the JCF-Z.
1
Figure A.1 illustrates the components of the transfer tube assembly: the transfer tube (335120,
standard flow; 335132, high flow), the tygon tubing (306653), the spring (870654), the spring
washer (853391), and the O-ring (870656).
Figure A.1 Transfer Tube Assembly Components
APPENDIX A
Assembling the Rotating Seal
PN JCFZ-IM-12AB
1. Tygon Tubing (306653
2. Transfer Tube
Standard Flow (335120)
High Flow (335132)
2
Before each use, lubricate the O-ring sparingly with silicone vacuum grease (Figure A.2).
3. O-ring (870656)
4. Spring Washer (853391)
5. Spring (870654)
A-1
Page 86
Assembling the Rotating Seal
Assembling the Rotating Seal
Figure A.2 O-ring (870656) and Silicone Vacuum Grease (335148)
A-2
PN JCFZ-IM-12AB
Page 87
Assembling the Rotating Seal
Assembling the Rotating Seal
3
After lubricating the O-ring, assemble the transfer tube assembly for use in the rotor as shown
in Figure A.3.
Figure A.3 Fully Assembled Transfer Tube Assembly
4
The transfer tube and rotating seal are shown in Figure A.4. Place the transfer tube in the rotor
stem, inserting the spring end in the down position.
A
Figure A.4 Transfer Tube Assembly and Rotating Seal
(Ready for Assembly into Rotor)
PN JCFZ-IM-12AB
A-3
Page 88
Assembling the Rotating Seal
Assembling the Rotating Seal
5
The beveled end of the transfer tube should protrude out of the seal housing as shown in
Figure A.5.
Figure A.5 Transfer Tube Beveled End
6
Place the rotating seal on the transfer tube assembly as shown in Figure A.6. Push down on the
rotating seal until the transfer tube spring bottoms out and then release pressure on the
rotating seal. It should immediately return to its original position as shown in Figure A.6.
Figure A.6 Correct Position of the Rotating Seal
A-4
PN JCFZ-IM-12AB
Page 89
Assembling the Rotating Seal
Assembling the Rotating Seal
7
If the rotating seal remains in the seal housing (Figure A.7), it is not properly tensioned. (This
condition has been virtually eliminated by the new design transfer tube assembly which has a
pre-tensioned spring.) Repeat steps 1 through 6 to ensure that the seal is pushed out of the seal
housing when spring tension is released (see Figure A.6).
Figure A.7 Incorrect Position of the Rotating Seal (Wrong Spring Tension)
A
8
When proper spring tension is achieved, assemble the bearing housing as shown in Figure A.8.
Figure A.8 Bearing Housing Assembled
PN JCFZ-IM-12AB
A-5
Page 90
Assembling the Rotating Seal
Assembling the Rotating Seal
9
Assemble the stationary seal housing together with the bearing housing and check to see that
the entire seal/bearing housing assembly sits on the rotating seal and rides up on the bearings.
This happens because the transfer tube spring supports the weight of the seal/bearing housing
when it is properly tensioned (Figure A.9).
Figure A.9 Stationary Seal Housing Mated to Bearing Housing (Seal/Bearing Housing)
1. Note height
A-6
PN JCFZ-IM-12AB
Page 91
Assembling the Rotating Seal
Assembling the Rotating Seal
10
Push down on the seal/bearing assembly until the transfer tube spring bottoms out
(Figure A.10).
Figure A.10 Pushing Down on Seal/Bearing Assembly to Tension Spring
A
11
While continuing to bear down on the seal/bearing housing, tighten the seal/bearing assembly
using the T-handle wrench supplied with the rotor (001311). Properly tightened, the seal/
bearing assembly will remain in the bottomed-out position, keeping tension on the transfer
tube spring (Figure A.11).
Figure A.11 Finishing Rotor Assembly
PN JCFZ-IM-12AB
A-7
Page 92
Assembling the Rotating Seal
Assembling the Rotating Seal
A-8
PN JCFZ-IM-12AB
Page 93
Description
The zonal bracket is designed for use in Avanti J series centrifuges when the JCF-Z Continuous Flow/
Zonal Rotor is used. The bracket provides support for the inlet and outlet tubing lines, keeping them
stable and out of the way of the spinning rotor during high-speed operation. The complete
installation is shown in Figure B.1.
Figure B.1 Avanti J Series Centrifuge Chamber with JCF-Z Rotor and Zonal Bracket Installed
APPENDIX B
Using the JCF-Z Continuous Flow/Zonal
Rotor Bracket (Avanti J Series Centrifuges
Only)
PN JCFZ-IM-12AB
1. Avanti J Centrifuge Chamber
2. Zonal Bracket
3. JCF-Z Rotor
B-1
Page 94
Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges Only)
Description
Zonal Bracket Kit Contents
Two zonal bracket kits are available for use with standard Cole-Parmer size 16 tubing
(6.4-mm, 1/4-in. O.D.). Kit (363843) is used with Avanti JXN-30, J-30I, and J-25 series centrifuges; kit
(366431) is used with Avanti JXN-26, J-26S XP, J-26 XP, and J-20 series centrifuges.
Table B.1 Basic Kit 363843 or 366431
ItemQuantityPart Numbers
Zonal Bracket1363818
(Avanti JXN-30, J-30I,
and J-25 series)
366430
(Avanti JXN-26, J-26S XP,
J-26 XP, and J-20 series)
Thumbscrew, stainless steel,
M4 (thread size) x 19 mm
Thumbscrew, stainless steel,
M4 (thread size) x 12 mm
Cable Clamp, nylon,
6.4-mm (1/4-in.) I.D.
2893412
2893411
10000499
Customers using Cole-Parmer size 14 (4.8-mm, 3/16-in. O.D.) or size 15 (9.5-mm, 3/8-in. O.D.) tubing
must use one of the other two kits described below in addition to the basic kit 363843 or 366431.
Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges Only)
Assembly Instructions
Step 1. Install the Rotor
Install the assembled JCF-Z rotor into the Avanti J chamber. See CHAPTER 3, Assembly and
Installation, for instructions.
Step 2. Attach the Zonal Bracket to the Centrifuge Chamber Wall
Insert the two 19-mm thumbscrews (the longer ones) through the holes in the bracket into the
rivnuts in the chamber wall (Figure B.2). Finger-tighten until secure.
Figure B.2 Attaching the Bracket to the Chamber Wall
Assembly Instructions
B
Step 3. Insert the Tubing Through the Door Gasket
Size 16 Tubing
You will need two pieces of tubing, one for the inlet line and one for the outlet line. Size 16 tubing
can be inserted directly through the tubing holes in the left side of the door gasket, resulting in a
snug fit.
1
Remove the plugs from the two center tubing holes in the door gasket.
PN JCFZ-IM-12AB
B-3
Page 96
Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges Only)
Assembly Instructions
2
To make it easier to insert the tubing, first carefully cut the end at an angle with scissors or a
sharp blade. Insert a piece of tubing through one of the tubing holes, from
centrifuge (Figure B.3). After the tubing is inserted, recut the tubing straight across.
Pull about eleven inches of tubing into the centrifuge chamber. Leave enough tubing outside
the centrifuge to reach your sample and buffer containers.
Figure B.3 Inserting Size 16 Tubing through the Gasket
1. Door Gasket
outside to inside the
3
Attach the other tubing end to a vacuum trap outside the centrifuge.
Sizes 14 and 15 Tubing
Sizes 14 and 15 tubing require the use of tubing adapters, which fit snugly inside the
tubing holes. The adapter ends are sized to fit the smaller- and larger-diameter tubing.
1
Remove the plugs from the two center tubing holes in the door gasket.
2
Insert an adapter through one of the holes until it is centered, with equal lengths of adapter on
either side of the tubing holes (Figure B.4). Repeat this step with the second adapter in the other
hole.
B-4
PN JCFZ-IM-12AB
Page 97
Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges Only)
Figure B.4 Attaching Tubing to the Adapter
1. Door Gasket
Assembly Instructions
B
Step 4. Attach the Tubing to the Zonal Bracket
1
Sizes 16 and 14 tubing: insert a standard-flow feed fitting into each tubing end, as shown in
Figure B.5. (For the larger size 15 tubing, which is used with the JCF-Z high-flow seal assembly,
feed fittings are not used; see Figure B.7.)
Figure B.5 Inserting the Standard-Flow Feed Fitting into the Tubing
1. Standard Flow Feed Fitting
2. Tubing Clamp
PN JCFZ-IM-12AB
2
Place one tubing clamp over each piece of tubing, about one inch behind the end of the fitting
(Figure B.5). The raised numbering on the clamp, indicating the O.D. of tubing used with the
clamp (in inches), should face up.
B-5
Page 98
Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges Only)
Assembly Instructions
3
Place one of the tubing clamps, with tubing pointing inward, over one of the holes in the
bracket (Figure B.6). Insert one of the shorter (12-mm) thumbscrews through the hole and turn
slightly to fasten. Do not tighten. Repeat with the other piece of clamped tubing. Leave both
thumbscrews slightly fastened. You will tighten them securely in step 5, number 2.
Figure B.6 Attaching the Tubing Clamps to the Bracket
Step 5. Attach Tubing to the JCF-Z Seal Assembly Housing
1
Sizes 16 and 14 tubing: press one standard-flow feed fitting into each of the holes in the side of
the seal housing.
Size 15 tubing: press one tubing end over each fitting. Size 15 tubing is used with the JCF-Z high-
flow seal assembly, in which the fittings are welded to the housing (Figure B.7).
Figure B.7 Inserting Tubing and Feed Fittings into the JCF-Z Seal Assembly
1. Standard Flow Assembly
2. High Flow Assembly
3. Seal Housing
B-6
PN JCFZ-IM-12AB
Page 99
Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges Only)
2
Make sure that the two sections of tubing from the gasket to the rotor are of even length, so
that the rotor is not being pulled to either side. Then finger-tighten the thumbscrews (shown
in Figure B.6) down securely.
You are now ready to complete setting up the JCF-Z flow system as described in this manual.
To attach a scavenger line (optional), follow the instructions in step 6 below.
Step 6. Attach a Scavenger Line (Optional)
1
Make sure that the seal assembly is oriented with the scavenger port facing the centrifuge
gasket portholes (see Figure B.8).
2
Remove the plug from the gasket hole nearest the front of the centrifuge.
Assembly Instructions
B
3
Route a piece of tubing through the hole, as shown in Figure B.8, from outside to inside the
centrifuge. This tubing must be long enough to reach from the scavenger port to a
20-kPa (3-psi) vacuum trap outside the centrifuge. Carefully cut the tubing at an angle to make
insertion easier, and then cut it straight across after insertion.
CAUTION
Do not route the scavenger line through the hole with the slit in the rubber
(nearest the rear of the centrifuge), or the door seal will be compromised,
affecting the vacuum level inside the chamber.
PN JCFZ-IM-12AB
B-7
Page 100
Using the JCF-Z Continuous Flow/Zonal Rotor Bracket (Avanti J Series Centrifuges Only)
Assembly Instructions
Figure B.8 Attaching a Scavenger Line
1. Scavenger Line
2. Scavenger Port
4
Attach the tubing to the scavenger port. Gently pull the tubing back through the porthole until
there is no excess tubing inside the centrifuge.
5
Attach the other tubing end to a vacuum trap outside the centrifuge.
B-8
PN JCFZ-IM-12AB
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.