COULTER HmX Hematology Analyzer
COULTER HmX Hematology Analyzer with
Autoloader
Operator’s Training
Guide
B
ECKMAN
OULTER
C
ECKMAN
B
C
OULTER
AVOID EXPOSURE, LASER RADIATION
EMITTED FROM THIS APERTURE
HmX
COULTER
®
HmX
HmX
COULTER
®
HmX
PN 4237530BB (March 2011)
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea, CA 92821
Page 2
WARNINGS AND PRECAUTIONS
READ ALL PRODUCT MANUALS AND CONSULT WITH BECKMAN COULTER-TRAINED PERSONNEL BEFORE ATTEMPTING
TO OPERATE INSTRUMENT. DO NOT ATTEMPT TO PERFORM ANY PROCEDURE BEFORE CAREFULLY READING ALL
INSTRUCTIONS. ALWAYS FOLLOW PRODUCT LABELING AND MANUFACTURER’S RECOMMENDATIONS. IF IN DOUBT AS
TO HOW TO PROCEED IN ANY SITUATION, CONTACT YOUR BECKMAN COULTER REPRESENTATIVE.
HAZARDS AND OPERATIONAL PRECAUTIONS AND LIMITATIONS
WARNINGS, CAUTIONS, and IMPORTANTS alert you as follows:
WARNING - Can cause injury.
CAUTION - Can cause damage to the instrument.
IMPORTANT - Can cause misleading results.
BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY
STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO,
PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR
ANY OTHER AUTOMATED LABORATORY ANALYZER.
WARNING Risk of operator injury if:
r All doors, covers and panels are not closed and secured in place prior to and during instrument operation.
r The integrity of safety interlocks and sensors is compromised.
r Instrument alarms and error messages are not acknowledged and acted upon.
r You contact moving parts.
r You mishandle broken parts.
r Doors, covers and panels are not opened, closed, removed and/or replaced with care.
r Improper tools are used for troubleshooting.
To avoid injury:
r Keep doors, covers and panels closed and secured in place while the instrument is in use.
r Take full advantage of the safety features of the instrument. Do not defeat safety interlocks and sensors.
r Acknowledge and act upon instrument alarms and error messages.
r Keep away from moving parts.
r Report any broken parts to your Beckman Coulter Representative.
r Open/remove and close/replace doors, covers and panels with care.
r Use the proper tools when troubleshooting.
CAUTION System integrity might be compromised and operational failures might occur if:
r This equipment is used in a manner other than specified. Operate the instrument as instructed in the Product Manuals.
r You introduce software that is not authorized by Beckman Coulter into your computer. Only operate your system’s
computer with software authorized by Beckman Coulter.
r You install software that is not an original copyrighted version. Only use software that is an original copyrighted
version to prevent virus contamination.
IMPORTANT If you purchased this product from anyone other than Beckman Coulter or an authorized Beckman Coulter
distributor, and, if it is not presently under a Beckman Coulter service maintenance agreement, Beckman Coulter cannot
guarantee that the product is fitted with the most current mandatory engineering revisions or that you will receive the most
current information bulletins concerning the product. If you purchased this product from a third party and would like
further information concerning this topic, call your Beckman Coulter Representative.
Page 3
REVISION STATUS
Initial Issue, A 6/99
Software version 1.0
Revision B, 3/01
Software version 1.3
Information concerning the IVD parameters, MRV and IRF, was added to this training
guide.
®
Changed pages: Cover, iii, ix, 3-5, 3-6, 3-7, 4-10, 4-14, 4-22, 6-26, 7-39, Running 5C
Control summary page 2 of 2, and Control File Management summary page 3 of 3.
Note: Changes that are part of the most recent revision are indicated in text by a bar in the
margin of the amended page.
Revision BA, 5/10
Software Version 1.3.
Updates were made to the company corporate address.
Cell
Note: Changes that are part of the most recent revision are indicated in text by a bar in the
margin of the amended page.
Revision BB, 03/11
Software Version 1.3.
Changes were made to pages, 6-3, 6-22.
Note: Changes that are part of the most recent revision are indicated in text by a bar in the
margin of the amended page.
PN 4237530BB
This document applies to the latest software listed and higher versions. When a subsequent software version
changes the information in this document, a new issue will be released to the Beckman Coulter website. For
labeling updates, go to www.beckmancoulter.com and download the most recent manual or system help for
your instrument.
iii
Page 4
REVISION STATUS
iv
PN 4237530BB
Page 5
LEGAL NOTICES
REVISION STATUS, iii
CONTENTS, v
INTRODUCTION, vii
1GETTING TO KNOW YOUR INSTRUMENT, 1-1
2STARTUP / SHUTDOWN, 2-1
3SET UP OPTIONS, 3-1
4QUALITY CONTROL, 4-1
5CALIBRATION, 5-1
6SAMPLE ANALYSIS, 6-1
7DATA REVIEW, 7-1
CONTENTS
8BASIC TROUBLESHOOTING, 8-1
SUMMARY PAGES
TRAINING CHECKLIST
TRADEMARKS
PN 4237530BB
v
Page 6
CONTENTS
vi
PN 4237530BB
Page 7
PURPOSE OF THIS DOCUMENT
To provide an authorized trainer with corporately accepted training materials for instructing a
new operator in the proper use and operation of a COULTER HmX Hematology Analyzer
with Autoloader or a HmX Hematology Analyzer.
DOCUMENTATION
Several manuals and related documents are shipped with a new instrument. It is important
that you familiarize yourself with these materials as soon as possible. Since these materials
may be used during your training, please make sure the following manuals and related
documents are readily available before the training session begins.
HmX Hematology Analyzer with Autoloader
rReference Manual, PN 4237523, provides in-depth information about what the
instrument does, the methods it uses, its specifications, and information on installation,
safety, and software options.
rOperator’s Guide, PN 4237521, provides step-by-step instructions for the day-to-day
running of your instrument.
rSpecial Procedures and Troubleshooting Manual, PN 4237522, provides in-depth
information about how to run a calibration, how to clean, replace, or adjust a component
of the instrument, and provides troubleshooting tables as diagnostic tools.
rHost Specifications, PN 4237518, defines the contents of records transmitted from a
HmX Hematology Analyzer with Autoloader to a host computer.
rMaster Index, PN 4237525, is a combined index for the Operator’s Guide, Reference,
and Special Procedures and Troubleshooting manuals. Each manual also has an
individual index.
INTRODUCTION
HmX Hematology Analyzer
rReference Manual, PN 4237523, provides in-depth information about what the
instrument does, the methods it uses, its specifications, and information on installation,
safety, and software options.
rOperator’s Guide, PN 4237519, provides step-by-step instructions for the day-to-day
running of your instrument.
rSpecial Procedures and Troubleshooting Manual, PN 4237520, provides in-depth
information about how to run a calibration, how to clean, replace, or adjust a component
of the instrument, and provides troubleshooting tables as diagnostic tools.
rHost Specifications, PN 4237518, defines the contents of records transmitted from a
HmX Hematology Analyzer to a host computer.
rMaster Index, PN 4237524, is a combined index for the Operator’s Guide, Reference,
and Special Procedures and Troubleshooting manuals. Each manual also has an
individual index.
PN 4237530BB
vii
Page 8
INTRODUCTION
ABOUT THIS TRAINING GUIDE
ABOUT THIS TRAINING GUIDE
Scope
This training guide provides the trainer with the opportunity to customize the training
experience to best fit the individual trainee and laboratory. Since trainee experience levels and
laboratory needs differ, this training outline is designed with flexibility in mind.
This document is designed to support both HmX Hematology Analyzer with Autoloader and
HmX Hematology Analyzer training. Material that is unique to the autoloader or the
single-specimen, rotary cap-pierce analyzer is clearly identified, usually with the analyzer’s
name at the top of the page.
rHmX Hematology Analyzer with Autoloader headers indicate the autoloader version of
the analyzer.
rHmX Hematology Analyzer headers indicate the single-specimen, rotary cap-pierce
version of the analyzer.
rUsing the term instrument or HmX instrument indicates the material applies to both.
Topics
The materials in this guide support the training of eight different topics:
rTopic 1 GETTING TO KNOW YOUR INSTRUMENT
rTopic 2 STARTUP / SHUTDOWN
rTop ic 3SE T U P O PT I ON S
rTopic 4 QUALITY CONTROL
rTop ic 5CA LI BR AT IO N
rTopic 6 SAMPLE ANALYSIS
rTop ic 7DATA RE VI EW
rTopic 8 BASIC TROUBLESHOOTING
Each topic consists of several related subjects that may be used, as needed, in your training.
Organization
The eight individual topics contained in the Operator’s Training Guide are part of a larger
organization that includes four basic types of information:
IntroductionExplains how to use this training guide
TopicsHeart of this document, each topic consists of multiple subjects and
operational summaries relating to the overall topic
viii
Summary PagesMaster copy for use by the Key Operator in their laboratory once
training is complete
Training ChecklistMethod for documenting the specific training an operator receives
PN 4237530BB
Page 9
INTRODUCTION
RESPONSIBILITIES
Summary Pages
Routine operational tasks detailed in the various instrument manuals are abbreviated in this
training document. These abbreviated instructions are referred to as either summary pages or
operational summaries. Each summary page is an individual, stand-alone document that is
designed for use in your laboratory as a training aid and/or as a reference tool for a trained
operator. A master copy of each summary page is located behind the
These summary pages contain essential operational instructions. As a result, it is important
that you become proficient in the use of this tool during your training. To eliminate the need
to flip back and forth between the topic and the summary masters, the appropriate
operational summaries are inserted within each topic. When an operational summary is
inserted in the training material, a rotated chevron (
Page title. For example, at the point of insertion, the Startup summary would begin with
➤) appears to the left of the Summary
Summary Masters tab.
➤ Startup Summary
Since this is a stand-alone summary for use after training is complete, some of the
instructions discussed in the training topic may be repeated in the summary. Two rotated
chevrons (
➤➤) mark the end of the inserted summary.
Training Checklist
This checklist provides the trainer with an official method to document the completion of
training on the COULTER HmX Hematology Analyzer with Autoloader or the HmX
Hematology Analyzer.
When the training process is complete, we recommend that both the trainer and trainee sign
the Training Checklist to show agreement that the trainee is sufficiently trained and able to
meet the objectives listed for each topic.
RESPONSIBILITIES
Beckman Coulter’s Responsibility
Beckman Coulter is responsible for instructing a Key Operator in the proper use and
operation of a COULTER HmX Hematology Analyzer with Autoloader or a HmX Hematology
Analyzer.
Our goal is to provide your laboratory with high quality training that best fits your
laboratory’s needs. This means the emphasis given to each topic may differ, with some topics
needing to be covered more thoroughly than others.
Key Operator’s Responsibility
It is the Key Operator’s responsibility to train all other operators in their laboratory. To assist
you in this process, a master copy of each operational summary is located at the end of the
Operator’s Training Guide. Once trained by an authorized Beckman Coulter trainer, the Key
Operator is authorized to copy these masters for use in training other laboratory personnel.
PN 4237530BB
ix
Page 10
INTRODUCTION
CONVENTIONS
CONVENTIONS
This training guide uses the following conventions:
rItalics font indicates screen text and/or messages displayed by the instrument such as
rNote (in bold font) prefaces information that is important to remember or helpful in
r
RESET THE SYSTEM or Press any key.
performing a procedure.
Bold indicates
ta menu item such as
tor a function such as F3 Run.
rë indicates a key such as Û.
rÌ indicates the operator needs to press and release the F4 key.
rë ë indicates to press and release the first key listed, then press and release the next
key listed.
rë + ë indicates to press and hold the first key listed, then press the next key.
rSoftware navigation to access a needed function or screen appears as a sequence of menu
items separated by double arrow heads. For example, the path to the screen for setting
up reagents is
contains one letter displayed in black.
rTo select a particular menu item,
tUse the æ or ç keys to highlight the menu item then press Û
or
tNote the black letter within the desired menu item and press the corresponding
alphabetic key. (It is not necessary to press the Û key.)
rA rotated chevron (
summary inserted within the training materials. For example,
Special Functions tt Set Up tt System set up tt Reagents. Each menu item
Run Samples
➤) placed to the left of a Summary Page title identifies an operational
➤ Changing Reagents Summary
GRAPHICS
x
rTwo rotated chevrons (➤ ➤) designate the end of an operational summary.
rHmX Hematology Analyzer with Autoloader or HmX Analyzer with Autoloader indicates
an autoloader unit.
rHmX Hematology Analyzer or HmX Analyzer indicates a single-specimen, rotary
cap-pierce unit.
rInstrument or HmX instrument indicates information concerning both the autoloader
and the single-specimen, rotary cap-pierce versions of the analyzer.
All graphics, including screens and printouts, are for illustration purposes only and must not
be used for any other purposes.
PN 4237530BB
Page 11
OBJECTIVES
When the subject is complete, you will be able to . . .
Data Management System (DMS)
rLocate and name the three major components of the peripheral computer system.
rLocate and explain the function of specific keys on the keyboard.
rQuickly access a commonly used area of software using the Access screen.
rSelect a desired option from the DMS Main Menu or submenu.
rOn the Status line, explain the meaning of the
rReplace the ribbon or ink cartridge and paper in the printer.
Main Unit
rLocate and name the hardware controls and indicators.
rGiven an instrument that appears to be without power,
GETTING TO KNOW YOUR INSTRUMENT
1
tSymbols
tBackground colors
tUp arrows versus down arrows.
1
tTell whether it is powered down or in Standby.
tBring the instrument back into operation.
rReset the system.
rLocate, name, and explain the major function of each module.
Modes of Operation
rState the four modes of operation available on HmX instruments.
rBriefly overview the Primary mode of operation versus the Secondary mode of operation.
rList four situations when the Secondary mode can be used.
rDemonstrate how to silence an alarm.
Reagents
rName the reagents used on a HmX instrument.
rState if the reagent is used for CBC analysis, DIFF analysis, or both.
rState if the reagent is used for Retic analysis.
rState the open container stability of each reagent used on a HmX instrument.
rExplain how to change reagents.
rExplain how to replace a waste container, if applicable.
PN 4237530BB
1-1
Page 12
GETTING TO KNOW YOUR INSTRUMENT
ACCESS SCREEN
ACCESS SCREEN
COULTER(R) HmX HEMATOLOGY ANALYZER
F1
RUN
SAMPLES
F4
DATA BASE
QUERY
This program is protected by U.S. and International laws as described
in the manual. (C) Copyright Beckman Coulter Inc., 1999
02/28/99 10:14 OPR DMS PR HC DB XB WL HWL QC
F5
WORKLIST
F2
RUN
CONTROLS
F6
HOST
WORKLIST
F3
CLEAN
F9
MAIN
MENU
SELECT FUNCTION
rFirst screen to appear after a system reset or power up
rScreen provides operator with quick access to the most commonly used areas of the
software
1-2
rDisplay automatically returns to the Access screen upon exit when this screen is used to
access any of the designated software areas
rMay also display this screen from the Main Menu by pressing É
PN 4237530BB
Page 13
GETTING TO KNOW YOUR INSTRUMENT
SOFTWARE MENU TREE - HmX Hematology Analyzer with Autoloader
SOFTWARE MENU TREE - HmX Hematology Analyzer with Autoloader
1
Sample Analysis
Run Samples
Data Base Query
Worklist
Host Worklist
XB
Current XB batch
XB batch means
XB graphs
F3 from Run Samples screen
SAMPLE MODE?
F2
F3
F4
F5
F6
F7
F8
F9
DIFF: ON
BLOOD DET: ON
Select to change/ESC to continue
F5 from Run Samples screen
F2 XB: ON N=2 IN
F4 DB: ON
F5 Print: NONE
F6 Host: OFF
F7 Display only: OFF
F8 Operator: OPR
F11 B&W screen print
F12 Color screen print
Autoloader Home
Clear the Bed/
Autoloader Home
Rock the Bed
Right Elevator Up/Down
Left Elevator Up/Down
Autoloader Test Routine
XB limits
Definitive flag limits
High/low flag limits
Laboratory Normal Ranges
Screen Labels
Parameter Selection
Reporting Units
Auto Print Format
Ticket Options
Spooler Priority
Graphics Options
Optional Printer
Host Computer Definition
PN 4237530BB
1-3
Page 14
GETTING TO KNOW YOUR INSTRUMENT
SOFTWARE MENU TREE - HmX Hematology Analyzer
SOFTWARE MENU TREE - HmX Hematology Analyzer
Sample Analysis
Run Samples
Data Base Query
Worklist
Host Worklist
XB
Current XB batch
XB batch means
XB graphs
F3 from Run Samples screen
SAMPLE MODE?
F2
F3
F4
F5
F6
F7
F8
F9
DIFF: ON
BLOOD DET: ON
Select to change/ESC to continue
F5 from Run Samples screen
F2 XB: ON N=2 IN
F3 CAP PIERCER 10mm-13mm
F4 DB: ON
F5 Print: NONE
F6 Host: OFF
F7 Display only: OFF
F8 Operator: OPR
F11 B&W screen print
F12 Color screen print
Auto Print Format
Ticket Options
Spooler Priority
Graphics Options
Optional Printer
Host Computer Definition
1-4
PN 4237530BB
Page 15
GETTING TO KNOW YOUR INSTRUMENT
STATUS LINE TABLE
07/13/99 19:34OPRDMSPRP2HCDBXBWLHWLQC
STATUS LINE TABLE
1
SymbolRefers to
DMSData
Management
System
PR*Graphics
Printer
HCHost
Computer
DBData BaseStore is ON.Store is OFF.Data base is not
XB
WLWorklistPreassigned
HWLHost WorklistPreassigned
QCQuality Control Auto-Stop is
P2Additional
% Analysis
Graphics
Printer
↑↓
Connected to
Analyzer.
Autoprint is set
to ALL,
ABNORMALS or
NORMALS.
Auto
transmission
ON.
XB is ON.XB is OFF.Last completed
entries pending
on Worklist.
entries are on
the Host
Worklist.
ON.
Autoprint is ON. Autoprint is OFF. Printer is
Not connected
to Analyzer.
Autoprint is set
to NONE.
Auto
transmission
OFF.
No preassigned
entries on the
Worklist.
No preassigned
entries on the
Host Worklist.
Auto-Stop is
OFF.
RedYellowWhite
Not
communicating
with Analyzer.
Printer is
off-line OR
Printer is out of
paper.
Not connected
to host.
functional.
System stops.
Reset the
system and
rerun last 2
specimens.
batch was OUT
(outside the XB
Action Limits).
3 consecutive or
10 total error
messages are in
the status field.
Host Worklist is
full.
Last control run
had an error
message.
off-line OR
printer is out of
paper.
DMS busy or
receiving data.
Printer is
printing.
Sending to host
computer.
Data base is
storing data.
N/ALast completed
The Worklist is
full (300
preassigned
samples).
DMS is
receiving
preassigned
samples from
the host
computer.
Receiving a
control run.
Printer is
printing.
DMS is OK.
Printer and DMS
are connected.
Host and DMS
are connected.
Data base is OK.
batch was IN
(within the XB
Action Limits).
Worklist is OK.
Host Worklist is
OK.
Results of last
control run are
OK.
Printer and DMS
are connected.
PN 4237530BB
* Changes to MA for manual printing, BA for batch printing, and AU for autoprinting.
1-5
Page 16
GETTING TO KNOW YOUR INSTRUMENT
DATA MANAGEMENT SYSTEM (DMS)
DATA MANAGEMENT SYSTEM (DMS)
Functions of the Data Management System (DMS)
rControls and monitors instrument operation
rDisplays, stores, outputs, and allows recall of sample results
rStores and graphs control results for the QC program
rAutomatically calculates new calibration factors
rAllows bidirectional communication with a host computer
rConsists of three major components: the computer, the monitor, and the printer
Major Components
Computer
rMust not use this system as a personal computer
rConsists of hardware components and software
rHardware is the term used to describe the physical components of a computer system
tLocate the following hardware components:
- Power ON/OFF button
- Power ON/OFF indicator light
- Hard drive indicator and symbol
- Diskette drive
rSoftware refers to a sequence of detailed instructions (called a program) that directs the
computer to perform certain actions
tData Management System or DMS refers to the software used in the computer
Monitor
rComputer uses the monitor to visually display information
rLocate the Power ON/OFF switch and the indicator light that lights when power is on
rLocate screen adjustment controls
Printer
rUsed to either print a copy of the screen displayed on the monitor or of data stored in the
computer’s data base
rUp to two printers can be used with your system
rLocate the Power ON/OFF switch
1-6
rExplain use of the ONLINE/OFFLINE button
rDemonstrate how to load the paper
rDemonstrate how to change the ribbon or ink cartridge
Keyboard
rDirect line of communication between the operator and the peripheral computer
containing the DMS software program
PN 4237530BB
Page 17
GETTING TO KNOW YOUR INSTRUMENT
DATA MANAGEMENT SYSTEM (DMS)
Layout
rAlphanumeric keys
tConfigured in a standard typewriter keyboard layout
tConsists of all standard alphanumeric keys plus a handful of special computer keys
and symbols
rÚ
tWorks like the Shift Lock key on a typewriter
tWith
rNumeric keypad
tKeypad contains the numbers 0 through 9, plus the period, an Û key, and various
tCan expedite manual entry of patient ID numbers and certain quality control entries
Caps Lock on,
- Pressing any alphabet key on the keyboard produces an uppercase letter
- Pressing the Ü key plus a letter produces a lowercase letter
- Only shifts the letter keys, all other keys on the keyboard remain the same
mathematical symbols
1
rê
tControls operation of the numeric keypad
tLocate light that illuminates when
tWith
tDMS keyboard is designed to be used with
tWith
rCursor control keys
tè æ ç é
tä and å
tâ and ã
tà and á keys are not used with the DMS program
Num Lock on, the numeric keypad produces numbers
Num Lock off, the numeric keypad on a typical keyboard doubles as a cursor
keypad; however, the DMS program is designed for use with the cursor control keys
- Also called arrow keys
- Directional keys for moving the cursor on the screen
- Used to highlight menu items, scroll up and down screens, or move to a field on
a screen to enter or edit data
- Commonly used to page through stored sample and control data
- Generally used to locate the beginning or the end of stored data
Num Lock is turned on
Num Lock on
PN 4237530BB
- When appropriate, a function key is designated as a delete key
rÉ through Ô Function Keys
tPressing a Function Key activates the function assigned to that key
tFor example, if a screen displays
displayed data and exits the screen display at the same time
rÕ key is not used with the DMS program
tWhen appropriate, a function key is designated as a print key
F10-Save/Esc as an option, pressing Ò saves the
1-7
Page 18
GETTING TO KNOW YOUR INSTRUMENT
DATA MANAGEMENT SYSTEM (DMS)
rÛ
tMainly used to confirm a menu or submenu selection which then activates a change
in the screen display
tDo not use Û to bring up the DMS display as pressing this key may activate a
command
tKeyboard has two Û keys
- Both work identically
- Û key was placed by the numeric keypad to facilitate rapid entry of numbers
rÈ
tUsed to back out of an option one screen at a time
tUse Ò
Save/Esc instead of È to store data you just entered or edited
rß
tUsed to toggle among options when a choice is required
tUse the ßto bring up the DMS display when the screen is dark
rÝ and Þ
tModifier keys that have no action by themselves but can be used in conjunction
with other keys to generate special characters or operations not readily available at
the keyboard
Hot Keys
rAlso called shortcut keys because they allow the operator to do certain functions quickly
rIncludes the following:
ÉDisplays the Access screen
(only available when the Main Menu is displayed)
ÌPrints
ÑExits the current screen and returns to either the Main Menu or the
Access screen depending on point of origination
Exception → if the F3 Run window is displayed, then the function of
Ñ is Stop
ÒSaves and/or returns to the previous screen
Þ + ãStops instrument beeping and removes the error message at the
bottom of the screen
Ý + ÊMoves from the current screen to the Error file and back to the
original screen
Ý + wMoves from Sample Analysis screen to Worklist and back when a
submenu or window is not displayed
1-8
Ý + ÑStops the cycle (HmX Hematology Analyzer with Autoloader only)
Ý + C Clears the bed (HmX Hematology Analyzer with Autoloader only)
PN 4237530BB
Page 19
GETTING TO KNOW YOUR INSTRUMENT
DATA MANAGEMENT SYSTEM (DMS)
Screen Conventions
rDMS uses several conventions that either reduce keystrokes or aid the operator in
obtaining the proper screen needed to complete a desired task
Menu Item Conventions
rAre displayed in blue text
rEach menu item contains one black letter
tPressing the corresponding alphabetic key is one way of selecting that particular
menu item
tPressing Û is not necessary
rMain Menu items have designated colors
1
tSample Analysis
tControls
tDiluter Functions
tSpecial Functions
rDesignated color coding appears on
tSubmenu borders
tScreen titles
tTitles that are part of a bordered display box
Display Screen Conventions
rAs a general rule, black text indicates a screen label or a field that cannot be edited;
blue text indicates a field that can be edited
rNumeric data stored in a table generally has a turquoise background
Instructional and Informational Message Convention
rMost, but not all, instructional or informational messages appear as black text with a
black border
blue
green
magenta
orange
PN 4237530BB
To continue HmX Hematology Analyzer training, proceed to page 1-10.
To continue HmX Hematology Analyzer with Autoloader training, go to page 1-18.
1-9
Page 20
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer
MAIN UNIT - HmX Hematology Analyzer
Front View of the Main Unit
POWER On/Off
rocker switch
C
ECKMAN
B
OULTER
READY
indicator
light
COMPUTER AC
POWER
AC INPUT
STANDBY
AVOID EXPOSURE, LASER RADIATION
EMITTED FROM THIS APERTURE
indicator
light
Red LED
Green LED
Standby/Reset
rocker switch
Exit
tray
HmX
Entry
port
®
HmX
COULTER
Bar-code
reader
Aspirator probe
Sample bar
IMPORTANT Risk of misleading results. Electrical interference can produce erratic parameter results. Operate
this instrument with all the doors closed to minimize external electrical interference.
Controls and Indicators
1-10
POWER On/Off Rocker Switch
rMain POWER On/Off switch located on the back of the Main Unit
rRecommend the Main Power be left on at all times
tProlongs the life of the lasers
tIncreases stability of electronic components
Standby/ Reset Rocker Switch
rLocated on the left side of the lower front door, near the bottom
rUsed to place instrument in a Ready or Standby state
tSymbol for indicating the Ready state is “
|”
tSymbol for indicating the Standby state is “O”
PN 4237530BB
Page 21
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer
rInstrument state is determined by the position of this rocker switch
tIn the Ready state, the back portion of the rocker switch is flush with the
instrument’s base; this is position
tIn the Standby state, the front portion of the rocker switch is flush with the
instrument’s base; this is position
rAlso used to reset the system or reestablish communication between components
|
O
1
tTo reset the system, place the Standby/Reset rocker switch in Standby (position
for 15 seconds then flip the switch back to Ready (position
Ready Indicator Light
rUsed to quickly identify instrument status
rWhen the light is ON, the Main Power is ON and the instrument is ready for operation
rWhen the Standby (
is OFF
Standby Indicator Light
rUsed to quickly identify instrument status
rWhen the light is ON, the Main Power is ON and the instrument is in the Standby state
rTo return to the Ready state, the
rStandby/Reset rocker switch must be moved to Ready, position
rIn the Standby state, voltages are still being applied to a memory location in the Analyzer
but everything else is powered down
rWhen the Standby (
is OFF
Entry Port
rPort for inserting closed-vial specimen tubes
O) and Ready (|) indicator lights are OFF (not lighted), Main Power
O) and Ready (|) indicator lights are OFF (not lighted), Main Power
|)
|
O)
PN 4237530BB
Exit Tray
rOutput tray for specimen tubes processed using the carousel in the Rotary Cap-Pierce
module in the Closed-Vial mode
Aspirator Probe
rUsed to aspirate from an open vial containing a whole-blood specimen, prediluted
specimen, or a retic preparation
rAlso used to aspirate from an open bottle of LATRON Primer or Control
Sample Bar
rPress to start aspiration from an open vial
1-11
Page 22
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer
Bar-code Reader
rScanner for reading the bar-code label on a specimen tube before either closed-vial or
open-vial sampling
Green LED
rWhen scanning a bar-code label, a glowing green LED indicates the barcode was read
successfully - the operator can process the specimen
Red LED
rGlowing red LED means wait
rWhen scanning a bar-code label, a glowing red LED indicates the barcode was not read
successfully
MODES OF OPERATION
HmX Hematology Analyzer . . . a Single-Specimen, Rotary Cap-Pierce System
Primary Mode
rThe lower door of the Main Unit must remain closed while the instrument is cycling
rAlso referred to as the Closed-Vial or Automatic mode
rCarousel rotates the specimen tube from the entry port to the cap-piercing station then
when sampling is complete, carousel rotates again to deposit the tube onto the exit tray
for removal by the operator
rAspirates 185 µL of sample from each tube
rSpecimen tubes must contain at least 1.0 mL of whole blood (minimum volume with the
proper proportion of blood to anticoagulant)
rTube and tube devices can be used if they fall within the following ranges:
diameter of 10 to 16 mm / length of 47 to 77 mm / capacity of 2 to 7 mL
tBeckman Coulter control and calibration tubes can also be used
tBefore inserting a specimen tube through the entry port, it may be necessary to
change the size of the carousel tube slot from the 10mm-13mm tube slot to the
16mm tube slot, or vice versa
tSee Appendix A of the Reference Manual for a listing of tested collection devices
rMay run specimen tubes with or without bar-code labels
1-12
PN 4237530BB
Page 23
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer
Secondary Mode
rAlso referred to as the Open-Vial or Manual mode
rAspirates 125 µL of sample from each tube via the aspirator probe
rOpening for aspiration is located on the back side of the aspirator probe’s tip
rMode used for processing
tMicrosamples
tRetics
tLATRON Primer and Control
tAny specimen tube that does not fit into the carousel
tAny specimen tube with insufficient sample volume for processing in the Primary
(Closed-Vial) mode of operation (that is, tubes containing <1.0 mL of whole blood)
tAny samples requiring dilution
Predilute Mode
rAspirator probe also used for running samples in the Predilute mode
rCBC mode only
1
rRequires specimen to be diluted 1:3 (minimum 50 µL of blood and 100 µL of diluent)
rAspirates 125 µL of the prediluted sample via the aspirator probe
rFinal results are calculated automatically
Retic Mode
rAspirator probe also used for running samples in the Retic mode
rSample preparation is required before processing
rAspirates approximately 2 mL of the sample dilution via the aspirator probe
PN 4237530BB
1-13
Page 24
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer
MAIN UNIT - HmX Hematology Analyzer
Modules on the Front of the Main Unit
Analyzer
module
Electronic
ower Supply
CBC module
BSV module
1-14
Rotary
Cap-Pierce
module
Pump module
PN 4237530BB
Page 25
INSTRUMENT MODULES
Modules on the Front of the Main Unit
Analyzer Module
rAlso referred to as the Analyzer/ Systems Control module
rContains the various circuit boards needed to operate the instrument
rControls the timing and sequencing of the operating cycles
rReceives pulses and raw data from the CBC and VCS modules
rCounts, measures, and computes CBC parameter results
rSends final CBC parameter results and VCS data to the DMS
Electronic Power Supply
rSupplies the necessary power for all instrument operations
Rotary Cap-Pierce Module
rSingle specimen tube transport and cap-piercing system
rWhen the specimen tube is inserted through the entry port, the carousel rotates the tube
to the cap-piercing station where the needle pierces the tube’s stopper so that 185 µL of
whole blood can be aspirated from the specimen tube into the instrument for analysis
rWhen the aspiration is complete, the carousel rotates the specimen tube away from the
needle and deposits it onto the exit tray for removal by the operator
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer
1
CBC Module
rContains all the components necessary for CBC analysis
BSV (Blood Sampling Valve) Module
rContains components associated with the sampling of blood including the aspiration
pumps for pulling sample from the specimen tube into the instrument
rAlso contains the sheath tank that supplies the diluent needed for proper sample flow for
differential analysis using VCS technology
Pump Module
rContains the two Erythrolyse II reagent pumps needed for differential analysis using
VCS technology
PN 4237530BB
1-15
Page 26
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer
Right Side of the Main Unit
041993
Main Diluter
module
Mixing
module
LASER
ON
Flow Cell
module
Modules on the Right Side of the Main Unit
Main Diluter Module
rContains various components relating to sample dilution and cycling as well as vacuum
and pressure regulators
Flow Cell Module
rContains the components necessary for analyzing the WBC Diff and Retics using VCS
technology
rAlso referred to as the Triple Transducer Module (TTM)
rFlow cell and laser are inside a protective housing
Mixing Module
rAssociated with the VCS portion of the system
rUsed to mix the whole blood sample with the reagents needed to prepare the sample for
WBC differential analysis using VCS technology
1-16
PN 4237530BB
Page 27
Back View of the Main Unit
Reagent
manifold
and
connectors
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer
Fans
Power ON/OFF
rocker switch
Plugs for
computer
1
Pneumatic
Power
Supply
Module on the Back of the Main Unit
Pneumatic Power Supply
rSource of the vacuums and pressures used throughout the instrument
rThe compressor/ vacuum pump produces one service adjustable level of air pressure,
60 psi and one nonadjustable level of vacuum, >22" Hg
To continue HmX Hematology Analyzer training, go to page 1-26.
PN 4237530BB
1-17
Page 28
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer with Autoloader
MAIN UNIT - HmX Hematology Analyzer with Autoloader
Front View of the Main Unit
POWER On/Off
rocker switch
POWER
COMPUTER AC
AC INPUT
Emergency
Stop button
Standby/Reset
rocker switch
C
B
ECKMAN
OULTER
HmX
Loading bay
COULTER
READY
indicator
light
STANDBY
indicator
light
Aspirator probe
Sample bar
®
HmX
1-18
IMPORTANT Risk of misleading results. Electrical interference can produce erratic parameter results. Operate
this instrument with all the doors closed to minimize external electrical interference.
Controls and Indicators
POWER On/Off Rocker Switch
rMain POWER On/Off switch located on the back of the Main Unit
rRecommend the Main Power be left on at all times
tProlongs the life of the lasers
tIncreases stability of electronic components
PN 4237530BB
Page 29
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer with Autoloader
Standby / Reset Rocker Switch
rLocated on the left side of the lower front door, near the bottom
rUsed to place instrument in a Ready or Standby state
1
tSymbol for indicating the Ready state is “
tSymbol for indicating the Standby state is “O”
rInstrument state is determined by the position of this rocker switch
tIn the Ready state, the back portion of the rocker switch is flush with the
instrument’s base; this is position
tIn the Standby state, the front portion of the rocker switch is flush with the
instrument’s base; this is position
rAlso used to reset the system or reestablish communication between components
tTo reset the system, place the Standby/Reset rocker switch in Standby (position
for 15 seconds then flip the switch back to Ready (position
Ready Indicator Light
rUsed to quickly identify instrument status
rWhen the light is ON, the Main Power is ON and the instrument is ready for operation
rWhen the Standby (
is OFF
Standby Indicator Light
rUsed to quickly identify instrument status
rWhen the light is ON, the Main Power is ON and the instrument is in the Standby state
rTo return to the Ready state, the Standby/Reset rocker switch must be moved to Ready,
position
rIn the Standby state, voltages are still being applied to a memory location in the Analyzer
but everything else is powered down
rWhen the Standby (
is OFF
|
O) and Ready (|) indicator lights are OFF (not lighted), Main Power
O) and Ready (|) indicator lights are OFF (not lighted), Main Power
|
O
|”
O)
|)
PN 4237530BB
Loading Bay
rCassettes are loaded here for processing in the Primary (Closed-Vial) mode
Aspirator Probe
rUsed to aspirate from an open vial containing a whole-blood specimen, prediluted
specimen, or a retic preparation
rAlso used to aspirate from an open bottle of LATRON Primer or Control
Sample Bar
rPress to start aspiration from an open vial
Emergency Stop Button
rPress this button to immediately stop the autoloader mechanism
1-19
Page 30
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer with Autoloader
MODES OF OPERATION
HmX Hematology Analyzer with Autoloader
Primary Mode
rThe lower door of the Main Unit must remain closed while the instrument is cycling
rAlso referred to as the Closed-Vial or Automatic mode
rAutoloader uses the rocker bed and sensors to move specimen tube holders called
cassettes from the loading bay to a cap-piercing station, then to the unloading bay
rAspirates 185 µL of sample from each tube
rSpecimen tubes must contain at least 1.0 mL of whole blood (minimum volume with the
proper proportion of blood to anticoagulant)
rUp to five specimen tubes of various types and sizes can be loaded into a cassette
rTube and tube devices can be used if they fall within the following ranges:
diameter of 10 to 16 mm / length of 47 to 77 mm / capacity of 2 to 7 mL
tBeckman Coulter control and calibration tubes can also be used
tSee Appendix A of the Reference Manual for a listing of tested collection devices
rProper size cassette must be used
tUse only 10 to 13 mm o.d. tubes in the universal cassette with gray inserts
tUse only 16 mm o.d. tubes in the cassette with black inserts
tWhen loading tubes into a cassette, make sure each specimen tube fits securely
inside the cassette to prevent blood spills that may occur if a tube falls or a cap is
pierced off center
rMay run specimen tubes with or without bar-code labels
rStop button is for emergency use only
Secondary Mode
rAlso referred to as Open-Vial or Manual mode
rAspirates 125 µL of sample from each specimen tube via the aspirator probe
rOpening for aspiration is located on the back side of the aspirator probe’s tip
rMode used for processing
tMicrosamples
tRetics
tLATRON Primer and Control
tAny specimen tube that does not fit into the cassette
tAny specimen tube with insufficient sample volume for processing in the Primary
(Closed-Vial) mode of operation (that is, tubes containing <1.0 mL of whole blood)
1-20
tAny samples requiring dilution
PN 4237530BB
Page 31
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer with Autoloader
Predilute Mode
rAspirator probe also used for running samples in the Predilute mode
rCBC mode only
rRequires specimen to be diluted 1:3 (minimum 50 µL of blood and 100 µL of diluent)
rAspirates 125 µL of the prediluted sample via the aspirator probe
rFinal results are calculated automatically
Retic Mode
rAspirator probe also used for running samples in the Retic mode
rSample preparation is required before processing
1
PN 4237530BB
1-21
Page 32
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer with Autoloader
MAIN UNIT - HmX Hematology Analyzer with Autoloader
Modules on the Front of the Main Unit
Analyzer
module
Electronic
Power Supply
CBC module
BSV module
DO NOT STARE INTO BEAM
CAUTION
LASER LIGHT
1.0 MILLIWATT MAXIMUM CLASS II LASER PRODUCT
670 nm DIODE LASER
1-22
Autoloader
module
Pump module
PN 4237530BB
Page 33
INSTRUMENT MODULES
Modules on the Front of the Main Unit
Analyzer Module
rAlso referred to as the Analyzer/ Systems Control module
rContains the various circuit boards needed to operate the instrument
rControls the timing and sequencing of the operating cycles
rReceives pulses and raw data from both the CBC and VCS modules
rCounts, measures, and computes CBC parameter results
rSends final CBC parameter results and VCS data to the DMS
Electronic Power Supply
rSupplies the necessary power for all instrument operations
Autoloader Module
rCassette transport and cap-piercing system
rContains the robotic mechanisms to move the specimen tube holders called cassettes
from the loading bay to the cap-piercing station and then to the unloading bay
rUp to five specimen tubes can be loaded into a cassette which is then placed in the
loading bay
rOnce operation is initiated, the bottom cassette is lowered onto the rocker bed where the
cassette is rocked back and forth while it moves along the bed to the cap-pierce station
rEach time a tube reaches the cap-pierce station, the needle pierces its stopper so that
185 µL of whole blood can be aspirated from the specimen tube into the instrument for
analysis
rWhen all the specimen tubes are pierced, the rocker bed moves the cassette to the
unloading bay for removal by the operator
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer with Autoloader
1
PN 4237530BB
CBC Module
rContains all the components necessary for CBC analysis
BSV (Blood Sampling Valve) Module
rContains components associated with the sampling of blood including the aspiration
pumps for pulling sample from the specimen tube into the instrument
rAlso contains the sheath tank that supplies the diluent needed for proper sample flow for
differential analysis using VCS technology
Pump Module
rContains the two Erythrolyse II reagent pumps needed for differential analysis using
VCS technology
1-23
Page 34
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer with Autoloader
Right Side of the Main Unit
041993
Main Diluter
module
Mixing
module
LASER
ON
Flow Cell
module
Modules on the Right Side of the Main Unit
Main Diluter Module
rContains various components relating to sample dilution and cycling as well as vacuum
and pressure regulators
Flow Cell Module
rContains the components necessary for analyzing the WBC Diff and Retics using VCS
technology
rAlso referred to as the Triple Transducer Module (TTM)
rFlow cell and laser are inside a protective housing
Mixing Module
rAssociated with the VCS portion of the system
rUsed to mix the whole blood sample with the reagents needed to prepare the sample for
WBC differential analysis using VCS technology
1-24
PN 4237530BB
Page 35
Back View of the Main Unit
Reagent
manifold
and
connectors
GETTING TO KNOW YOUR INSTRUMENT
MAIN UNIT - HmX Hematology Analyzer with Autoloader
Fans
Power ON/OFF
rocker switch
Plugs for
computer
1
Pneumatic
Power
Supply
Module on the Back of the Main Unit
Pneumatic Power Supply
rSource of the vacuums and pressures used throughout the instrument
rThe compressor/ vacuum pump produces one service adjustable level of air pressure,
60 psi and one nonadjustable level of vacuum, >22" Hg
PN 4237530BB
1-25
Page 36
GETTING TO KNOW YOUR INSTRUMENT
REAGENTS
REAGENTS
Reagents Connected to the Main Unit
ReagentRecommendedFunctionCBC DIFFRetic
Open Container
Stability
DiluentISOTON III diluent r Isotonic solution that
dilutes the sample
r Stabilizes cell
membranes
r Conducts aperture
current
r Carries and focuses the
VCS sample stream
r Rinses the system
between samples
CBC Lytic
Reagent
Diff Lytic Reagent Erythrolyse II erythrocyte
Diff Leukocyte
Preservative
Cleaning AgentCOULTER CLENZ
LYSE S III diff lytic
reagent
lytic reagent (also called
PAK LYSE)
One of the HmX PAK
reagents
StabiLyse leukocyte
preservative (also called
PAK PRESERVE)
One of the HmX PAK
reagents
cleaning agent
r Disrupts erythrocytes
r Frees hemoglobin
r Reduces cellular debris
r Converts hemoglobin to
a stable cyanidecontaining pigment
r Dilutes the sample for
the diff
r Lyses red blood cells
r Reduces cellular debris
r Preserves white blood
cells in their near-native
state for differentiation
using VCS technology
r Cleans and rinses the
diluter parts
r Prevents protein buildup
r Eliminates routine
aperture bleaching
Same as shelf life
(date printed on
the container)
60 Days
60 Days
60 Days
90 Days
1-26
PN 4237530BB
Page 37
GETTING TO KNOW YOUR INSTRUMENT
Special Reagents Used in Sample Preparation for Retic Analysis
ReagentRecommendedFunctionCBC DIFFRetic
REAGENTS
Open Container
Stability
1
ReticPrep
Reagent A
ReticPrep
Reagent B
New Methylene Blue dye,
special formulation
Clearing reagent
(very dilute sulfuric acid)
r Stains any reticulum
present in the RBCs
r Removes hemoglobin
from the cells without
removing the
precipitated stain-RNA
complex
r Keeps the cell
membrane intact
60 Days
60 Days
Storage and Handling
rAll reagents should be stored and used at the laboratory’s ambient room temperature
rActive components may separate (layer) if the reagent has been frozen during transport
or storage → inconsistency of components affects parameter results
rTo ensure reagent integrity, treat each reagent container as if it were frozen sometime
during transport or delivery to your institution
rBefore using a reagent
tMake sure the reagent is completely thawed
tGently mix by inversion
tAllow sufficient time for any microbubbles to dissipate
Note: Microbubbles may increase the background count and affect parameter
results.
➤ Changing Reagents Summary
1 Record the New Reagent Information
rFrom the Main Menu, select Special Functions tt Set Up tt System Set Up tt Reagents
rEnter the lot number and the expiration date or open container stability date,
as applicable
tThe date opened automatically changes to today’s date when the lot number is
entered (this is the reason the lot number should always be typed even if it is the
same as the last container)
tSix digits must be entered for the expiration date or open container stability date
Note: Open container stability is stated on the reagent container, when applicable.
rPress Ò to save the information then Ñ to return to the Main Menu
PN 4237530BB
1-27
Page 38
GETTING TO KNOW YOUR INSTRUMENT
REAGENTS
2 Replace the Reagent Container
rTurn the compressor off
From the Main Menu, select
Fluidics Tests tt Compressor On/Off then press Û; press the ßto select Off and
press Û again
rOpen the new reagent container
rRemove the pickup tube assembly from the old container then transfer it directly to the
new container and tighten
Note: If the lower part of the assembly touches you or anything outside the container,
flood that lower part of the assembly with distilled water then wipe it with a lint-free
tissue.
3 Prime the Lines
rFrom the Main Menu, select Diluter Functions tt Prime Reagents
rSelect the reagent that was changed then press Û, the compressor turns on and the
Prime Reagents program activates
Special Functions tt Diagnostics tt Operator Options tt
rOnce priming is complete, press Ñ to return to the Main Menu and continue normal
operation
➤ ➤
➤ Replacing a Waste Container Summary
1 Put the Instrument into Standby
rMove the Standby/Reset rocker switch to Standby, position O
2 Turn OFF the Main Power
rUse the Main POWER On/Off rocker switch located on the back of the Main Unit
3 Replace the Waste Container
rUsing biohazardous precautions,
tRemove the waste assembly from the full container
tTransfer the waste assembly directly to an empty waste container and tighten
rDispose of the biohazardous waste according to you laboratory’s protocol
1-28
4 Turn ON the Main Power
rUse the Main POWER On/Off rocker switch located on the back of the Main Unit
5 Put the Instrument into Ready
rMove the Standby/Reset rocker switch to Ready, position |
➤ ➤
PN 4237530BB
Page 39
OBJECTIVES
When the subject is complete, you will be able to . . .
Startup
rInitiate a Startup.
rRecognize an out-of-limit result on the Startup Summary or System Status Values screen.
rExplain what to do if a Startup result is out of limits.
Clean Cycle
rInitiate a Clean Cycle.
rState the difference between a Clean Cycle and a Shutdown.
Shutdown
rInitiate a Shutdown.
rState how often a Shutdown must be performed.
rState the minimum time the COULTER CLENZ cleaning agent should stay in the
instrument while in Shutdown.
STARTUP / SHUTDOWN
2
2
PN 4237530BB
2-1
Page 40
STARTUP / SHUTDOWN
OBJECTIVES
NOTES
2-2
PN 4237530BB
Page 41
STARTUP
rMust be completed before running patient specimens or controls anytime cleaning agent
has replaced diluent in the aperture baths
rDuring a Clean Cycle, Startup is automatically initiated 30 minutes after completion of
the Shutdown cycle
rPerform quality control checks before running patient specimens
STARTUP SCREENS
Summary Screen
STARTUP / SHUTDOWN
STARTUP
2
PN 4237530BB
rAlso referred to as the First Start Up screen because it is the first display that
automatically appears and prints
rShows a reagent summary on the left and a system check summary on the right
rAutoloader Check prompt only appears on the Summary screen of a HmX Hematology
Analyzer with Autoloader
Reagent Summary
rThe reagent summary lists each reagent lot number and expiration date (may be the
open container expiration date, if applicable)
rIf a date appears in red numbers, the expiration date is exceeded
System Check Summary
rShows results on a Pass/Fail basis
rA check that fails appears in red
2-3
Page 42
STARTUP / SHUTDOWN
STARTUP
System Status Values Screen
rAlso called the Second Start Up screen
rAllows an operator to review the detailed system check results
rCan only be viewed by pressing Ê when the Summary screen is displayed
Background Values
rIf any Background result is outside of its limits,
tResult is displayed in red
tPress Ë
Repeat Background to redo the background counts
tIf the condition still exists, refer to the Special Procedures and Troubleshooting
(SPT) Manual
System Values
rIf any of the electronic readings, pressures, vacuums, or temperature readings are outside
of limits,
tValue is displayed in red
tPerform a
Special Functions tt Diagnostics tt Operator Options tt System Test
System Test to confirm
Press Û then Ë Run to initiate the test
tIf the condition still exists, refer to the Special Procedures and Troubleshooting
(SPT) Manual
2-4
PN 4237530BB
Page 43
➤ STARTUP SUMMARY
1 Check Current Instrument Status
rIf the power is off, turn it on
rIf the Main unit is in Standby, switch it to Ready
rIf the DMS monitor is dark, press ß
rIf Startup results are already displayed as the result of a Clean Cycle, go to step 3; if not,
enter your three character Operator ID (optional)
tFrom the Main Menu, press Í
tType your three character Operator ID
tPress È to return to the Main Menu
2 Perform Start Up
rFrom the Main Menu, select Diluter Functions tt Start Up
rPress Û
Options
STARTUP / SHUTDOWN
STARTUP
2
3 Review Results
rOnce the startup cycles are complete, review the Summary screen and, if necessary, the
System Status Values screen
tIf the Background Check fails, press Ê to view the System Status Values screen
and then press Ë
tIf Fail appears for any other check, press Ê to view the System Status Values
screen then perform a
Special Functions
Press Û then Ë Run to initiate the test then follow the prompts
rStartup results are not saved, but do print automatically
rSave the Summary and System Status Values screen printouts for documentation
purposes
Repeat Background to redo the background counts
System Test
tt Diagnostics tt Operator Options tt System Test
4 Perform Quality Control Checks
rPerform quality control checks before running patient specimens
➤ ➤
PN 4237530BB
2-5
Page 44
STARTUP / SHUTDOWN
STARTUP
NOTES
2-6
PN 4237530BB
Page 45
SHUTDOWN
rAllowing cleaning agent to remain in the instrument a minimum of 30 minutes
rPerform either a Clean Cycle or a Shutdown once every 24 hours that the instrument is
Clean Cycle
rConsists of a Shutdown cycle followed 30 minutes later by a Startup cycle
rPerform quality control checks before running patient specimens
Shut Down
rWhen the Shutdown cycle is complete, leave the power ON → every 24 hours, the system
rStartup must be completed before running patient specimens or controls
STARTUP / SHUTDOWN
SHUTDOWN
minimizes protein buildup in the instrument
in use
does an autopurge cycle to purge the flow cell and the sample lines with diluent
Note: If you turn the power OFF and the instrument is going to be idle for more than
48 hours, do the Prolonged Shutdown Procedure in the Operator’s Guide, Chapter 5.
2
➤ CLEAN CYCLE SUMMARY
1 Perform the Clean Cycle
rFrom the Access screen, press Ë CLEAN then press Û
2 Let the Instrument Stand
rCleaning agent remains in the instrument
rLeave instrument power on
rAfter 30 minutes, a Startup automatically begins
3 Review Results
rOnce the startup cycles are complete, review the Summary screen and, if necessary, the
System Status Values screen
tIf the Background Check fails, press Ê to view the System Status Values screen
and then press Ë
tIf Fail appears for any other check, press Ê to view the System Status Values
screen then perform a
Special Functions
Press Û then Ë Run to initiate the test then follow the prompts
rStartup results are not saved, but do print automatically
rSave the Summary and System Status Values screen printouts for documentation
purposes
Repeat Background to redo the background counts
System Test
tt Diagnostics tt Operator Options tt System Test
PN 4237530BB
4 Perform Quality Control Checks
rPerform quality control checks before running patient specimens
➤ ➤
2-7
Page 46
STARTUP / SHUTDOWN
SHUTDOWN
➤ SHUTDOWN SUMMARY
1 Perform Shut Down
rVerify SELECT FUNCTION is displayed on the status line
rFrom the Main Menu, select
rPress Û
Diluter Functions tt Shut Down
2 Let the Instrument Stand
rAllow cleaning agent to remain in the instrument a minimum of 30 minutes
rLeave instrument power on
Note: If the power is turned off and the instrument is going to be idle for more than
48 hours, do the Prolonged Shutdown Procedure in the Operator’s Guide, Chapter 5.
rPerform the Startup procedures before running patient specimens or controls
➤ ➤
2-8
PN 4237530BB
Page 47
OBJECTIVES
When the subject is complete, you will be able to . . .
rSet up control files using the
Control set up option.
SET UP OPTIONS
3
3
rEnter data under the
formats, and print options.
rEnter data under the
host computer is in use.
Sample analysis set up option to include flagging limits, display
System set up option including the communications definition if a
PN 4237530BB
3-1
Page 48
SET UP OPTIONS
OBJECTIVES
NOTES
3-2
PN 4237530BB
Page 49
SET UP MENU OPTIONS
rFrom the Main Menu, select Special Functions tt Set Up to access this menu
r
Set Up submenu consists of three options
tControl set up
tSample analysis set up
tSystem set up
rAll the options available in the Set Up area of the DMS software are listed here
tEach menu item is listed with a short description
tFor details, see Chapter 6 of the Operator’s Guide
CONTROL SET UP
rFrom the Main Menu, select Special Functions tt Set Up tt Control set up
rFor additional details, see Chapter 6 of the Operator’s Guide
rUsed to set up any of the 20 control files
rSets up a variety of control files
tCBC/DIFF file
SET UP OPTIONS
SET UP MENU OPTIONS
3
tLatex file
tCBC file
tRETIC file
rAlso used to enable or disable the Auto-Stop function
tAlthough available on both HmX instruments, the Auto-Stop function is more
typically used with a HmX Hematology Analyzer with Autoloader
t
Auto-Stop option ON
Instrument stops and a beeping alarm sounds when one of the following control
error messages occur:
Control Out of Range
Control Expired
File Full
File not Found (CBC/DIFF only)
Disk Drive C: Full
t
Auto-Stop option OFF
Instrument continues analysis when a control error occurs
Setting Up Your Laboratory’s Control Files
rUse the instructions under the Initial Set Up of Control Files Summary heading to set up
your laboratory’s control files
PN 4237530BB
3-3
Page 50
SET UP OPTIONS
CONTROL SET UP
➤ INITIAL SET UP OF CONTROL FILES SUMMARY
Any Control File can be set up as a:
rLATRON control file (Latex file)
r5C cell control file (CBC/DIFF file)
rRetic-C cell control file (RETIC file)
Setting Up a LATRON Control File
1 Select the File
rFrom the Main Menu, select
rMove the cursor to a NOT SETUP file then press Û
2 Type Manual Entries
rEnter the name of the file and your Operator ID
rEnter the Lot # and expiration date from the package insert
Special Functions tt Set Up tt Control set up tt Latex file
3 Verify Host Setting
rMake sure the Host transmit setting correlates with your laboratory’s protocol
rTo change, use the ß to toggle between ON and OFF
4 Verify All Information
rCheck all entries are correct then press Ò to save and exit the screen
Setting Up a 5C Cell Control File
1 Select the File
rFrom the Main Menu, select
CBC/DIFF file
rMove the cursor to a NOT SETUP file then press Û
2 Upload the Assay Values
rInsert the 5C cell control diskette into the A: drive of the computer
rPress Í
rPress the function key for the desired level of control
3 Type Manual Entries
rSelect the Shift and manually type the Operator ID
Upload Assay Values
Special Functions tt Set Up tt Control set up tt
3-4
4 Verify Host Setting
rMake sure the Host transmit setting correlates with your laboratory’s protocol
rTo change, use the ß to toggle between ON and OFF
5 Verify All Information
rCheck all entries are correct then press Ò to save and exit the screen
PN 4237530BB
Page 51
6 Set Up Another Level of Control, if applicable
rMove the cursor to a
rRepeat steps 2 through 6
7 Remove the Diskette from Drive A: (after all needed set ups are complete)
NOT SETUP file then press Û
Setting Up a RETIC File
1 Select the File
rFrom the Main Menu, select
rMove the cursor to a NOT SETUP file then press Û
2 Type Entries from the Table of Expected Results
rEnter the Lot #, Expiration Date, Shift, and your Operator ID
rEnter the RBC and Retic % values from the assay sheet
Note: If you report Retic number (RET #), it is important that you enter the correct
RBC value from the assay sheet. If you enter the wrong number and then correct it
after running controls, the DMS does not recalculate the incorrect RET # control
results that are stored in the RETIC file.
rIf your instrument has the MRV and IRF parameters enabled, enter 99.9 as the
assigned values and expected ranges for these parameters. Use these values until
you establish your own.
Special Functions tt Set Up tt Control set up tt RETIC file
SET UP OPTIONS
CONTROL SET UP
3
3 Verify Host Setting
rMake sure the Host transmit setting correlates with your laboratory’s protocol
rTo change, use the ß to toggle between ON and OFF
4 Verify All Information
rCheck all entries are correct then press Ò to save and exit the screen
➤ ➤
Note: This summary is only for initial set up of control files; therefore, it is not included in
the Summary Pages. The next time you set up a control file, use the
Using the Default Limits/ Expected Ranges from the Package Insert Summary
to maintain customized expected ranges in a 5C control file. In this case, use the
Management Using Your Laboratory’s Custom Ranges Summary
Control File Management
unless you wish
Control File
.
PN 4237530BB
3-5
Page 52
SET UP OPTIONS
SAMPLE ANALYSIS SET UP
SAMPLE ANALYSIS SET UP
rFrom the Main Menu, select Special Functions tt Set Up tt Sample analysis set up
rFor additional details, see Chapter 6 of the Operator’s Guide
Action limits
XB limits
rSets Target Values and Limits, user defined
Definitive flag limits
rThese word messages are generated based on user defined limits
High/low flag limits
Note: MRV and IRF parameters appear on this screen when these parameters are enabled.
rThese flags are generated based on user defined limits
Laboratory Normal Ranges
Note: MRV and IRF parameters appear on this screen when these parameters are enabled.
rThese limits are the default normal ranges for the HmX instruments
rThe laboratory’s normal ranges should be entered here for printing on the report
Location list
rCreates a list of common locations
rOperator may then choose from this list when assigning demographics in the Worklist or
when editing in the Data Base
Physician list
rCreates a list of common physicians
rOperator may then choose from this list when assigning demographics in the Worklist or
when editing in the Data Base
Display formats
Screen Labels
Note: MRV and IRF parameters appear on this screen when these parameters are enabled.
3-6
rMakes Primary Identifier selections and allows the operator to change the name of the
label identifiers
PN 4237530BB
Page 53
SET UP OPTIONS
SAMPLE ANALYSIS SET UP
Parameter Selection
rEnables or disables the Pct, PDW, and RET#
rEnables or disables the MRV and IRF
Reporting Units
Note: MRV and IRF parameters appear on this screen when these parameters are enabled.
rSelects the appropriate reporting units for the laboratory
Delete database
rDeletes all samples in the data base and resets the system
Delete host spooler
rClears the buffer of results waiting to be transmitted to the host computer
Clear printer spooler queue
rClears the buffer of results waiting to be transmitted to the printer
3
Print options
Auto Print Format
rSelects ticket or graphics format
Ticket Options
rSelects the options available for the ticket format (if applicable)
rThese options include selecting the units (% or 10
(WBC or RBC) or if Suspect and Definitive flags will print
Spooler Priority
rSets which will print first (Manual, Auto, or Batch)
Graphics Options
rSelects the options available for the graphics format
rThese options include the page format (width /font), which scatterplots will print, and
determines if demographics, suspect and definitive flags, units, normal ranges and the
diff box will print
Optional Printer
rSets the options for a second printer
3
/µL), normal ranges, parameter labels
PN 4237530BB
3-7
Page 54
SET UP OPTIONS
SYSTEM SET UP
SYSTEM SET UP
rFrom the Main Menu, select Special Functions tt Set Up tt System set up
rFor additional details, see Chapter 6 of the Operator’s Guide
Shift
rUsed to separate the control files into different shifts
Reagents
rUpdates the reagent log
Institution
rSets up the banner for the top of each printout
Communication def
rHost computer definition
rConfigures the communication format for transfer of sample information to a host
computer
IQAP ID#
rAfter entering the lab’s participant number in this field, it will attach to the control files
for IQAP reference
Set Date/Time
rAllows the operator to adjust the current date and/or time (for example, daylight savings
time)
Supervisor Password
rAllows the operator to change the supervisor password
Optimize Hard Disk
rSets the option to initiate the clean up of the DMS hard drive upon power up or resetting
the unit
rSee the Special Procedures and Troubleshooting (SPT) Manual for more information
3-8
PN 4237530BB
Page 55
OBJECTIVES
When the subject is complete, you will be able to . . .
LATRON Primer and Control
rExplain the storage and handling requirements.
rState the open-vial stability.
rPrepare and run LATRON primer and control.
rRecognize an out-of-limits result on the LATRON screen.
rUsing the Operator’s Guide, explain what to do if a LATRON primer or control result is
rExplain the statistics associated with the control files and graphs.
5C Cell Control
rExplain the storage and handling requirements.
rState the open-vial stability.
rDefine an event.
rState indications of instability or deterioration.
rPrepare and run 5C cell control.
out-of-limits.
QUALITY CONTROL
4
4
rRecognize an out-of-limit result on the 5C control screen.
rUsing the Operator’s Guide, explain what to do if a 5C cell control result is out-of-limits.
rExplain the statistics associated with the control files and graphs.
Retic-C Cell Control
rExplain the storage and handling requirements.
rState the open-vial stability.
rDefine an event.
rState indications of instability or deterioration.
rPrepare and run Retic-C cell control.
rRecognize an out-of-limit result on the Retic-C control screen.
rUsing the Operator’s Guide, explain what to do if a Retic-C control result is out-of-limits.
rExplain the statistics associated with the control files and graphs.
Interlaboratory Quality Assurance Program (IQAP)
rExplain the purpose of the Interlaboratory Quality Assurance Program (IQAP).
PN 4237530BB
4-1
Page 56
QUALITY CONTROL
OBJECTIVES
Control File Management
rReview numeric and graphic control data for errors and make appropriate corrections.
rPrint and/or download control file data.
rDelete individual control runs.
rDelete a control file.
rProperly set up control files for running LATRON, 5C Cell Control, and/or Retic-C Cell
rExplain how to make changes to an existing control file.
Additional QC Checks
rExplain the purpose of Mode-to-Mode Comparison.
rExplain the purpose of a linearity study.
rExplain the purpose of a diff comparison.
XB Analysis
rExplain the purpose of XB Analysis.
Control.
rExplain why only RBC indices are used in the X
Analysis.
B
rDefine the terms batch, target values, action limits, non-reportable results, and
non-random sampling as they apply to the X
rState the minimum number of samples for a valid X
rExplain when an X
rDescribe the three areas of the X
batch mean table prints automatically.
B
graph.
B
rIdentify the problem parameter (or parameters) causing an X
Analysis.
B
batch.
B
batch to be out-of-control.
B
4-2
PN 4237530BB
Page 57
LATRON PRIMER AND CONTROL
rDaily quality control check to verify performance of the VCS technology TTM
components in both the DIFF and Retic modes
rMust be at room temperature prior to use
tMay be stored in the refrigerator but not necessary
tIf stored in the refrigerator product must be brought to room temperature before
running
rInstrument aspirates approximately 1.5 mL
LATRON Primer
rContains a surfactant to remove bubbles and clean the sample tubing and pathway from
the tip of the aspirate probe to and through the flow cell (located inside the TTM)
rDoes not need to be mixed
tVigorous mixing creates microbubbles that may cause increased counts
rResults are automatically saved in the Latex control file
QUALITY CONTROL
LATRON PRIMER AND CONTROL
4
rCan manually print the results if desired
LATRON Control
rContains specific size latex particles with a predetermined Volume, Conductivity and
Laser Light Scatter (measurements determined using VCS technology)
rControl is used to verify
tFlow cell alignment
tGains for flow cell volume, conductivity and laser light scatter
tCV for flow cell volume, conductivity and laser light scatter
rLatex particles settle to the bottom of the vial
tOperator must gently invert the vial to resuspend particles
tVigorous mixing creates microbubbles that may affect results
rControl must not be exposed to direct sunlight or temperatures above 30°C (86°F);
keep the vials away from windows and heaters
rResults are automatically saved in the Latex control file
rCan manually print the results if desired
PN 4237530BB
4-3
Page 58
QUALITY CONTROL
LATRON PRIMER AND CONTROL
Reviewing LATRON Control Results
On the Control Run Screen
rMean Channel result for Volume, Conductivity, and Scatter
tAn
H or L is displayed when a result is outside the limits (flag does not appear in red
as its does on the Startup screen)
rCV results for Volume, Conductivity, and Scatter
tAn
H is displayed when a result is above the upper limit
Using Î Graph . . . a Control Run Screen Option
rSemi-quantitative graphs
rMean channel and %CV graphs of the last 10 runs
rPress Î as many times as necessary to scroll to the graphs you want to view
Using Review or Report, Latex . . . a Control Menu Option
rComprehensive display of all runs, up to 100 runs per file
rAllows operator to view the Mean Channel and CV data
tAn
tAn
H or L is displayed when a Mean Channel result is outside the limits
H is displayed when a CV percentage result is above the upper limit
rUse the é and è keys to move between DIFF and Retic data
rUse the
F6-Remove/Restore option as a data management tool
Using Graphs . . . another Control Menu Option
rLevey-Jennings graphs
rPlot point for each result, up to 100 runs per file
rIdeal for spotting trends and shifts
4-4
rUse Î
Additional Graphs to move between DIFF and Retic graphs
PN 4237530BB
Page 59
➤ Running LATRON Primer and Control Summary
1 Status for Proceeding
rEnsure Startup is complete
rLATRON Primer and Control must be at room temperature
2 Run LATRON Primer
At the DMS
rFrom the Main Menu, select
rIf the LATRON file does not appear, press Ê File and select your current Latex file
Controls tt Control Run
QUALITY CONTROL
LATRON PRIMER AND CONTROL
4
rPress Ë
At the Diluter
rCycle LATRON Primer in the Secondary mode
tImmerse the aspirator tip completely in the Primer
tDo not place the tip against the side of the bottle
tPress and release the sample bar
tListen for the beep or verify the status line message changes from Aspirating to
Reviewing Primer Results at the DMS
rCheck the count for both DIFF and Retic when the analysis is complete
rIf both counts are
tPress È to remove the Primer Run window
tGo to step 3 and run LATRON Control
rIf either or both counts are
tRerun the LATRON Primer
tEvaluate the new count and troubleshoot if needed
Run then Ì PRIMER
Diluting before removing the bottle
≤500,
≥500,
PN 4237530BB
3 Run LATRON Control
At the DMS
rIf necessary, press È to close the Primer Run window
rPress Ë
Run then press Ë CONTROL (SECONDARY)
4-5
Page 60
QUALITY CONTROL
LATRON PRIMER AND CONTROL
At the Diluter
rGently invert the LATRON Control bottle 5 to 8 times
rCycle LATRON Control in the Secondary mode
tImmerse the aspirator tip completely in the control
tDo not place the tip against the side of the bottle
tPress and release the sample bar
tListen for the beep or verify the status line message changes from Aspirating to
Diluting before removing the bottle
Reviewing Control Results at the DMS
rReview results once the run appears on the Control Run screen
rIf a result is out-of-limits, repeat the entire process beginning with LATRON Primer
rIf the results of two control runs are unacceptable,
tFollow your laboratory’s protocol
or
tRefer to the
When LATRON is Out of Limits table in the Operator’s Guide
rCompare the results of this run with previous runs, as desired
tTo evaluate stored numerical data, select
Controls tt Review or Report
Note: Use the é and è keys to move between DIFF and Retic data.
tTo evaluate Levey-Jennings graphs, select
Controls tt Graphs
➤ ➤
4-6
PN 4237530BB
Page 61
5C CELL CONTROL
rMaterial for quality control of CBC and DIFF parameters
r5C Cell Control is assayed only for Primary mode
rThree levels of control with varying mixtures of normal and abnormal values: Normal,
Abnormal I, and Abnormal II
Storage
rControl must be stored at 2 to 8°C (35 to 46°F)
rStorage of the control product in the cap down (inverted) position is not recommended
tMight require additional mixing for complete resuspension of cellular components
Proper Handling
rDo not write an open tube date or initials in the white space between the bar code and
the tube’s cap
tBar-code reader may not be able to properly read the barcode if marks are present in
this white space
QUALITY CONTROL
5C CELL CONTROL
4
tWhen the bar-code label cannot be read, the control is stored as a sample in the
Sample Analysis data base rather than in its proper control file
Prepare 5C Cell Controls according to the Package Insert Instructions
rWarm control vials at ambient temperature for 10 to 15 minutes before mixing
rMix gently by hand using the 8 x 8 x 8 method twice
tAfter the second mixing, check the sides and bottom of the tube to verify the control
is mixed completely
tOvermixing causes hemolysis which indicates product deterioration
rDo not use mechanical mixers or rotators
rReturn to the refrigerator within 30 minutes to ensure stated open-vial stability
Stability
Open-Vial Stability
rOpen-vial stability is 13 days or 13 events (whichever comes first)
rAn event occurs anytime an operator completes the following sequence:
tRemoves a control vial from the refrigerator
tAllows the control vial to stand at room temperature 10 to 15 minutes
tMixes the control vial gently by hand using the 8 x 8 x 8 method two times
tAspirates sample
tReturns the control vial to the refrigerator within 30 minutes
PN 4237530BB
rMCV and/or RDW parameters may show trending through the product’s shelf life;
however, this is inherent to the product and should not be considered an indicator of
product instability
4-7
Page 62
QUALITY CONTROL
5C CELL CONTROL
Indications of Instability
rA slight pink color to the supernatant is normal
rInability to obtain expected values in the absence of known instrument problems or
Running 5C Cell Control
Control Run Option
rUsing the
rOnly use this option to run
gross hemolysis (darkly colored supernatant) indicates deterioration of the control
Control Run option to process 5C cell controls is not recommended
tLATRON Primer and Control in the Secondary (Open-Vial) mode
tRetic-C cell controls in the Secondary (Open-Vial) mode
tControl vials with damaged bar-code labels
Note: We recommend control vials with damaged bar-code labels be processed in
the Primary (Closed-Vial) mode using the
Bar-Code Labels
procedure in Chapter 2 of the Operator’s Guide.
Cycling Commercial Cell Controls without
tCommercial cell controls without bar-code labels
Note: We recommend these controls be processed in the Primary (Closed-Vial)
mode; see the
Cycling Commercial Cell Controls without Bar-Code Labels procedure
in Chapter 2 of the Operator’s Guide.
Recommendations for Running 5C Cell Controls
rControl tubes may be run in any order
rRun cell controls in the Primary mode in
Sample Analysis
tWill not see the control results displayed on the Run Samples screen
tContinues to display the last patient sample processed
tBar-code label on the control tube directs results to the correct file
rTo use the Secondary (Open-Vial) mode for running 5C cell control, your laboratory
must determine its own means and expected ranges for each parameter and set up a
control file using these values
4-8
PN 4237530BB
Page 63
➤ Running 5C Cell Control Summary
1 Getting the DMS Ready
QUALITY CONTROL
5C CELL CONTROL
4
If SELECT FUNCTION is displayed
rFrom the Main Menu, select Sample
Analysis
Note: Preparation for cycling in the
Primary mode with the DIFF mode
ON begins automatically.
ttRun Samples
If SELECT FUNCTION is not displayed
rFrom the Main Menu, select Sample
Analysis tt Run Samples
rPress Ë Run
rIf DIFF mode is OFF, press Ñ STOP
then press Î DIFF ON / OFF
rIs the PRIMARY: SAMPLE ANALYSIS
prompt at the top of the F3-Run
window?
tIf yes, press È
tIf no, press Ê
2 Prepare Controls according to the Package Insert
rWarm controls at room temperature 10 to 15 minutes
rMix each control tube using the 8 x 8 x 8 method twice
tDo not use a mechanical mixer!
tDo not write an open tube date or your initials in the white space between the bar
code and the tube’s cap
3 Run Controls in the same manner as Patient Samples
START PRIMARY
If using a HmX Hematology Analyzer with Autoloader
rLoad the control tubes into a cassette
tLoad in any order
tVerify the bar-code label is facing upward
rPlace the cassette in the loading bay
If using a HmX Hematology Analyzer
rScan the bar-code label of the control tube you wish to run
Note: Control levels may be processed in any order.
rAfter a successful read, insert the control tube through the entry port into the carousel
rRepeat for each control tube
PN 4237530BB
4-9
Page 64
QUALITY CONTROL
5C CELL CONTROL
4 Review Results
If using a HmX Hematology Analyzer with Autoloader
rTo verify control results are stored in the control file, select
rIf the bar code was not facing up properly or if the bar code was not read because it is
For both the HmX Hematology Analyzer with Autoloader and the HmX Hematology Analyzer
rIf a result is out-of-limits,
rCompare the results of this run with previous runs, as desired
Controls tt Review or Report
damaged, control results are stored in the patient data base
tIf the bar code was not facing up, position it correctly and run the control again
tIf the bar code is damaged, see the
Bar-Code Labels
heading in the Operator’s Guide Startup and Controls chapter
Cycling Commercial Cell Controls without
tFollow your laboratory’s protocol
or
tRefer to the
When CBC/DIFF Control is Out of Limits table in the Operator’s Guide
tTo evaluate stored numerical data, select
Controls tt Review or Report
tTo evaluate Levey-Jennings graphs for the presence of shifts or trends, select
Controls tt Graphs
rPrint results from the Control Run screen, as desired
➤ ➤
4-10
PN 4237530BB
Page 65
RETIC-C CELL CONTROL
rThree levels: Level I, Level II and Level III
rControls can only be analyzed after sample preparation is completed using the
COULTER ReticPrep Reagent Kit
rUsed for monitoring the Coulter method of reticulocyte analysis which uses VCS
technology
rCannot be used to control the manual method
Storage
rControl must be stored at 2 to 8°C (35 to 46°F)
rStore open vials upright in the rack provided
tStorage of the control product in the cap down (inverted) position might require
additional mixing for complete resuspension of cellular components
Proper Handling
Prepare according to the Package Insert Instructions
rWarm at room temperature (20 to 25°C) for 15 minutes before mixing
QUALITY CONTROL
RETIC-C CELL CONTROL
4
rMix gently by hand using the 8 x 8 x 8 method once
tAfter mixing, check the sides and bottom of the vial to verify the control is mixed
completely
tOvermixing causes hemolysis which indicates product deterioration
rDo not use mechanical mixers or rotators
rReturn to the refrigerator within 30 minutes to ensure stated open vial stability
Stability
Open-Vial Stability
rOpen-vial stability is 30 days or 30 events (whichever comes first)
rAn event occurs anytime an operator completes the following sequence:
tRemoves a control vial from the refrigerator
tAllows the control vial to stand at room temperature 15 minutes
tMixes the control vial gently by hand using the 8 x 8 x 8 method (one time)
tPipettes control sample from the vial
tReturns the control vial to the refrigerator within 30 minutes
Indications of Instability
rA pink to light red supernatant is normal
rModerate to marked hemolysis indicates deterioration of the control, usually due to
overmixing
PN 4237530BB
4-11
Page 66
QUALITY CONTROL
RETIC-C CELL CONTROL
Running Retic-C Cell Control
rUse the Control Run option for processing
rPrepare a dilution using the COULTER ReticPrep Reagent Kit before analysis
rRun in the Secondary mode
Special Reagents Needed for Retic Analysis
ReagentRecommendedFunctionCBCDIFFRetic
Open Container
Stability
ReticPrep
Reagent A
ReticPrep
Reagent B
New Methylene Blue
dye, special
formulation
Clearing reagent
(very dilute sulfuric
acid)
r Stains any reticulum
present in the RBCs
r Removes hemoglobin
from the cells without
removing the precipitated
stain-RNA complex
r Keeps the cell membrane
intact
60 Days
60 Days
4-12
PN 4237530BB
Page 67
➤ Running Retic-C Cell Control Summary
1 Prepare Controls according to the Package Insert
rWarm controls at room temperature 10 to 15 minutes
rMix each control tube using the 8 x 8 x 8 method once
rDo not use a mechanical mixer
2 Preparing the Control
rFor each control tested, label two 12 x 75 mm test tubes "A" and "B"
rPlace 4 drops of Reagent A into tube "A"
rDispense 50 µL of the well mixed control into tube "A"
rGently mix tube "A"
tLet the tube stand for a minimum of 5 minutes and a maximum of 60 minutes
3 Getting the DMS Ready
rFrom the Main Menu, select Controls tt Control Run
rPress Ê File and select the desired file
QUALITY CONTROL
RETIC-C CELL CONTROL
4
rPress Ë
Run, Ë CONTROL (SECONDARY)
4 Run Controls in the Same Manner as Patient Samples
rGently mix the "A" tube
rPipette 2 µL of the control/stain mixture from tube "A" into the bottom of tube "B"
rPlace tube "B" with the control/stain aliquot at an angle under the tip of the Reagent B
dispenser
rDispense 2 mL of Reagent B into tube "B"
rDo not mix
rWait 30 seconds then analyze the retic preparation you just made
tImmerse the aspirator tip into the retic preparation
tPress and release the sample bar
tListen for the beep or verify the status line message changes from Aspirating to
Diluting before removing the tube
5 Review Results
rIf a result is out-of-limits,
tFollow your laboratory’s protocol
or
tRefer to the
rCompare the results of this run with previous runs, as desired
tTo evaluate stored numerical data, select
When Retic Control is Out of Limits table in the Operator’s Guide
Controls tt Review or Report
PN 4237530BB
tTo evaluate Levey-Jennings graphs for the presence of shifts or trends, select
Controls tt Graphs
➤ ➤
4-13
Page 68
QUALITY CONTROL
REVIEWING CONTROL RESULTS
REVIEWING CONTROL RESULTS
Using Control Menu Options to Review 5C Cell Control or Retic-C Results
Note: MRV and IRF parameters appear on the retic screens when these parameters are
enabled.
On the Control Run Screen
rCompares control results to expected ranges
tAn
Using Î Graph . . . a Control Run Screen Option
rSemi-quantitative graphs
rDisplays results of the last 10 runs
rPress Î as many times as necessary to scroll to the graphs you want to view
Using Review or Report, CBC/DIFF (or RETIC) . . . a Control Menu Option
rComprehensive display of all runs, up to 100 runs per file
rUse the é or è key to view additional parameters
H or L is displayed when a result is outside the limits (flag does not appear in red
as it does on the Startup screen)
rAllows operator to review all parameter results
tValues for Assay and Limits are entered when a control file is set up
tParameter results are listed individually and by average (MEAN); ongoing 2SD and
CV values are also provided
tIf the CV result is greater than 20%, a +++++ code is displayed
Note: +++++ code replacing basophil parameter results is typical.
tOut-of-control results have an
rUse the
F6-Remove/Restore option as a data management tool
H or L next to the result
Using Graphs . . . another Control Menu Option
rLevey-Jennings graphs
rPlot point for each result, up to 100 runs per file
rIdeal for spotting trends or shifts
4-14
PN 4237530BB
Page 69
TABLE OF QUALITY CONTROL MATERIALS
Recommended
QC materialFunctionCBCDIFFRetic
QUALITY CONTROL
TABLE OF QUALITY CONTROL MATERIALS
4
Open Vial
Stability
LATRON PrimerCleans and prepares the pathway for
running LATRON Control
LATRON ControlChecks the functioning of the VCS
Technology
5C Cell ControlChecks recovery of CBC and Diff
parameters
Retic-C Cell Control Checks recovery of Retic parameter
S-CAL CalibratorAdjusts the accuracy of the CBC
parameters
LIN-C Cell ControlChecks the linear recovery of the CBC
parameters
IQAP (Interlaboratory Quality Assurance Program)
rThe Coulter Interlaboratory Quality Assurance Program is a free service offered to all
regular users of 4C, 4C PLUS, 5C and Retic-C cell controls as well as users of LIN-C
linearity controls
rIt complements a laboratory’s in-house quality control program by providing peer group
comparisons for each lot number of a given control product run on the same instrument
type
rThe assessment is completely independent of the assay value issued when the product
was supplied
rA separate IQAP manual (PN 4206266) is available with information on enrollment, data
submission, understanding the IQAP report, and troubleshooting IQAP problems
30 Days
30 Days
13 Days/Events
30 Days/Events
1 Hour
7 Days
PN 4237530BB
4-15
Page 70
QUALITY CONTROL
CONTROL FILE MANAGEMENT
CONTROL FILE MANAGEMENT
rOnce a control file is initially set up, the file:
tNever reverts back to a
tIs considered an existing control file
rTwo types of existing control files
tControl file that contains control run data (may be either current or old data)
tControl file with no stored control run data because the file is either newly set up or
has had all its control runs deleted from it
Existing Control Files Currently In Use
Reasons an Operator may need to Access a Control File that is Currently in Use
rTo correct a typographical error
rTo edit an assigned value (obtained from the Table of Expected Results) to a running
mean (established by the laboratory)
NOT SET UP status
4-16
PN 4237530BB
Page 71
➤ MAKING CHANGES TO A CONTROL FILE CURRENTLY IN USE
Use only if you want to make a change in a control file that is currently in use. This summary
should be used when changing individual
entering a running mean or correcting a typographical error. If you want to make a set up
change (such as changing the lot number), use the appropriate Control File Management
Summary.
1 Select the File
rFrom the Main Menu, select Special Functions tt Set Up tt Control set up
rSelect the type of file and then select the existing file you want to change
2 Answer the Questions
rRecompute Existing Runs ?
tPress ß to answer Ye s then press Û
tYes automatically adjusts the statistics and graphs to the change
tOperator must toggle to Ye s to look at the screen even if no changes are made
items in a previously set up control file, such as
QUALITY CONTROL
CONTROL FILE MANAGEMENT
4
rUse Default Limits / Expected Range ?
tTo maintain laboratory custom ranges, press Û to answer No
tTo use Expected Ranges (from the package insert), press ß to answer Ye s then
press Û
3 Make the Changes
rEnter the new information
4 Verify All Changes
rCheck all entries are correct then press Ò to save and exit the screen
➤ ➤
PN 4237530BB
4-17
Page 72
QUALITY CONTROL
CONTROL FILE MANAGEMENT
Existing Control Files No Longer In Use
Reason an Operator may need to Access an Existing File that is No Longer in Use
rTo overwrite the file with the set up information of another lot number
Note: Management of these control files differs depending on whether the laboratory
wants to use the default limits /expected ranges (from the Table of Expected Results on
the Beckman Coulter package insert) or the laboratory wants to use their own custom
ranges (expected ranges established by the laboratory).
To evaluate the difference of control file management for a laboratory using the default
limits / expected ranges from the package insert versus a laboratory using their own
custom ranges, see a comparison of procedures on the next page.
4-18
PN 4237530BB
Page 73
CONTROL FILE MANAGEMENT
Management of Existing Control Files No Longer In Use . . . a Comparison
QUALITY CONTROL
4
Using Default Limits/ Expected Range
(from the package insert)
1. Control file is initially set up using a
NOT SETUP file
2. Run controls for the month 2. Run controls for the month
3. Using the
review stored control data and use the
F6-Remove/Restore option to remove
any invalid runs
4. Print the control file4. Print the control file
5. Download the file for IQAP,
if applicable
6. Delete all the control runs using the
F8-Delete File option
Review or Report option,
Using Custom Ranges
(established by your laboratory)
1. Control file is initially set up using a
NOT SETUP file
Operator types over the default limits /
expected ranges with their laboratory’s
own established ranges
3. Using the
review stored control data and use the
F6-Remove/Restore option to remove
any invalid runs
5. Download the file for IQAP,
if applicable
6. Set up the new lot number of control
a. Answer the following questions:
Question 1:
Recompute Existing Runs ?
Review or Report option,
7. Set up a new lot in this file either now
or in the future
rPress ß to toggle to Ye s
then press Û
rYes automatically adjusts the
statistics and graphs to the
change (operator must toggle
to Yes to look at the screen even
if no changes are made)
Question 2:
Use Default Limits/ Expected Range ?
rPressing Û to answer No
maintains the laboratory’s
custom ranges
b. Set up the new lot information
7. Before
access the
delete all the control runs using the
F8-Delete File option
starting to run the new control,
Review or Report option and
PN 4237530BB
4-19
Page 74
QUALITY CONTROL
CONTROL FILE MANAGEMENT
➤ CONTROL FILE MANAGEMENT SUMMARY
(Using the Default Limits/ Expected Range from the Package Insert)
Note: To maintain customized expected ranges in a control file, use the Control File
Management Using Your Laboratory’s Custom Ranges Summary
Review the Numeric Control Data and Graphs, if applicable
1 Locate the Control File
rFrom the Main Menu, select
rPress Ê File and select the file for review then press Û
2 Review the Graphs
rUsing Î
Additional Graphs review all graphs noting any trends, shifts or plot points
inside the red area
Controls tt Graphs
.
rPress È then R to switch to the
3 Review Numeric Data
rReview %CV of all parameters (press é to see additional parameters)
rIf a CV is high, use å and examine each run for results mistakenly run into
the file (the graph review should correlate)
rDelete any mistakes from the file using Î
Print / Download the Control File Data
rFrom the Review or Report screen, press Ì to automatically print all parameters
and all runs
rIf desired, press È then G then Ì to print all graphs
rDownload IQAP data, as needed
Delete the Control File Data
1 Go to Control Review Screen
rFrom the Main Menu, select
rPress Ê File and select the file you want to delete then press Û
Review or Report option
Remove/Restore
Controls tt Review or Report
4-20
2 Delete the Control File Data
rPress Ð
Delete File
rPress ß to toggle to Ye s
rPress Û to confirm and delete the control file data
Set Up the New Lot Number of Control
Any Control File can be set up as a:
r5C cell control file (CBC/DIFF file)
rLATRON control file (Latex file)
rRetic-C cell control file (RETIC file)
PN 4237530BB
Page 75
Setting Up a 5C Cell Control File
1 Select the File
rFrom the Main Menu, select
rMove the cursor to a NOT SETUP file or an inactive file where the control run data is
deleted then press Û
2 Upload the Assay Values
rInsert the 5C cell control diskette into the A: drive of the computer
QUALITY CONTROL
CONTROL FILE MANAGEMENT
Special Functions tt Set Up tt Control set up tt CBC/DIFF file
4
rPress Í
rPress the function key for the desired level of control
3 Type Manual Entries
rSelect the Shift and manually type the Operator ID
4 Verify All Information
rMake sure the Host transmit setting correlates with your laboratory’s protocol
rCheck all entries are correct then press Ò to save and exit the screen
5 Set Up Another Level of Control, if applicable
rMove the cursor to a
deleted then press Û
rRepeat steps 2 through 5
6 Remove the Diskette from Drive A: (after all needed set ups are complete)
Upload Assay Values
NOT SETUP file or an inactive file where the control run data is
Setting Up a LATRON Control File
1 Select the File
rFrom the Main Menu, select
rMove the cursor to a NOT SETUP file or an inactive file where the control run data is
deleted then press Û
Special Functions tt Set Up tt Control set up tt Latex file
PN 4237530BB
2 Type Manual Entries
rEnter the name of the file and your Operator ID
rEnter the Lot # and expiration date from the package insert
3 Verify All Information
rMake sure the Host transmit setting correlates with your laboratory’s protocol
rCheck all entries are correct then press Ò to save and exit the screen
4-21
Page 76
QUALITY CONTROL
CONTROL FILE MANAGEMENT
Setting Up a RETIC File
1 Select the File
rFrom the Main Menu, select
rMove the cursor to a NOT SETUP file or an inactive file where the control run data is
deleted then press Û
2 Type Entries from the Table of Expected Results
rEnter the Lot #, Expiration Date, Shift, and your Operator ID
rEnter the RBC and Retic % values from the assay sheet
Note: If you enter the wrong RBC value and then correct it after running controls,
the DMS does not recalculate the incorrect RET # control results stored in the
RETIC file.
3 If the MRV and IRF Parameters Are Enabled
rEnter 99.9 for MRV and IRF assigned values and expected ranges if customized
values are not established
Special Functions tt Set Up tt Control set up tt RETIC file
4 Verify All Information
rMake sure the Host transmit setting correlates with your laboratory’s protocol
rCheck all entries are correct then press Ò to save and exit the screen
➤ ➤
4-22
PN 4237530BB
Page 77
➤ CONTROL FILE MANAGEMENT SUMMARY
(Using Your Laboratory’s Custom Ranges)
Use this summary only if you want to maintain the customized expected ranges previously
entered in an existing control file (typically a 5C control file).
Review the Numeric Control Data and Graphs, if applicable
1 Locate the Control File
rFrom the Main Menu, select
rPress Ê File and select the file for review then press Û
2 Review the Graphs
rUsing Î
points inside the red area
rPress È then R to switch to the
3 Review Numeric Data
rReview %CV of all parameters (press é to see additional parameters)
Additional Graphs review all 20 graphs noting any trends, shifts or plot
Controls tt Graphs
Review or Report option
QUALITY CONTROL
CONTROL FILE MANAGEMENT
4
rIf a CV is high, use å and examine each run for results mistakenly run into
the file (the graph review should correlate)
rDelete any mistakes from the file using Î
Print / Download the Control File Data
rFrom the Review or Report screen, press Ì to automatically print all parameters
and all runs
rIf desired, press È then G then Ì to print all 20 graphs
rDownload IQAP data, as needed
Changing the Existing File
1 Select the File
rFrom the Main Menu, select
rSelect the existing file you want to change
2 Answer the Questions
rRecompute Existing Runs ?
tPress ß to answer Ye s then press Û
tYes automatically adjusts the statistics and graphs to the change (operator must
toggle to Ye s to look at the screen even if no changes are made)
rUse Default Limits / Expected Range ?
tPress Û to answer No (maintains your laboratory’s custom ranges)
Remove/Restore
Special Functions tt Set Up tt Control set up tt CBC/DIFF file
PN 4237530BB
4-23
Page 78
QUALITY CONTROL
CONTROL FILE MANAGEMENT
3 Upload the 5C Control Assay Values
rInsert the 5C cell control diskette into the A: drive of the computer
rPress Í
rPress the function key for the desired level of control
4 Type Manual Entries
rSelect the Shift and manually type the Operator ID
5 Verify All Information
rMake sure the Host transmit setting correlates with your laboratory’s protocol
rCheck all entries are correct then press Ò to save and exit the screen
6 Set Up Another Level of Control, if applicable
rSelect another
rRepeat steps 2 through 6
Upload Assay Values
CBC/DIFF file you want to change
7 Remove the Diskette from Drive A: (after all needed set ups are complete)
IMPORTANT Risk of misleading control results. Old control runs remaining in a newly set up control file are
included in the statistical data for the new file. Delete all controls runs before starting to run the new control.
Delete the Control File Data
1 Go to Control Review Screen
rFrom the Main Menu, select
Controls tt Review or Report
rPress Ê File and select the file you want to delete then press Û
2 Delete the Control File Data
rPress Ð
Delete File
rPress ß to toggle to Ye s
rPress Û to confirm and delete the control file data
➤ ➤
4-24
PN 4237530BB
Page 79
ADDITIONAL QC CHECKS
Mode-to-Mode Comparison
rEnsures both the Primary and Secondary aspiration modes recover similar values
rMinor differences between the Primary and Secondary modes are due to differences in
the flow characteristics of the aspiration pathways
r5C cell control is assayed for Primary mode only
rRecommended procedure is located in Chapter 2 of the Operator’s Guide
rThe mode to mode limits stated in the Reference manual are established using 10 normal
bloods measured in triplicate (3 consecutive runs each); these limits should not
necessarily be used as limits for a single sample used to check differences between modes
Linearity Checks
rLinearity limits apply only to WBC, RBC, Hgb, and Plt
rUses various concentrations of blood to verify the reportable range from very low to very
high values
rLinearity can be performed using whole blood or a commercial product such as
COULTER LIN-C linearity control
rCoulter publishes performance specifications in the Reference Manual, however, a
laboratory may be required by a regulatory agency to perform an independent procedure
rWhen running very low concentrations in the Primary mode, blood detectors must be
rPerform manual differentials as a measure of good QC practice or as recommended by
your laboratory, state, and federal protocols
rManual differentials are performed to verify the automated differential
rChecks the full differential including:
tReagents
tMixing process in the mixing chamber
tSheath and sample pressures
Note: LATRON control checks the VCS technology and the integrity and alignment
of the flow cell.
rFor details, see the Differential Comparison Procedure in Appendix B of the Reference
Manual
PN 4237530BB
4-25
Page 80
QUALITY CONTROL
ADDITIONAL QC CHECKS
XB Analysis
rXB X = Mean
rPronounced x-bar-bee
rAppears on the DMS screen as XB
rXB will default to ON if instrument is reset
XB Analysis, a Quality Control Tool
rXB Analysis is a quality control method that allows a laboratory to continuously monitor
the performance of their automated hematology system and thus control the quality of
their results
rAny method used for quality control must use a material that meets two requirements
tMaterial must be stable
tMaterial must be similar in content to the patient samples that will be analyzed
rX
rNo additional cost to the laboratory because the method uses patient sample results
Analysis meets these requirements
B
tUses the patient samples themselves as a material so the similarity requirement is
tUses the red blood cell indices (MCV, MCH, and MCHC) so the stability
B = Bull (for Brian S. Bull, M.D.)
met
requirement is met
RBC Indices
rMCV, MCH, and MCHC are relatively stable parameters
tRBC indices are very tightly controlled by the body because the red cells function
best within a very narrow range of size and hemoglobin content
tOther parameters such as the white blood cells and platelets tend to have a wider
physiological range and are, therefore, not as predictable as the red cell indices
rAre typically stable for an individual patient from day to day
rAre stable for a patient population over time
tMany hospitalized patients have been investigated and it has been found that there
is no significant day-to-day or week-to-week variability in the mean of their indices
Target Values
rA constant for each RBC index (MCV, MCH, or MCHC)
rTarget values should reflect the entire patient population of the laboratory
tThe patient population of a general hospital usually includes samples from all
patient age groups, disease states and many hematologically “normal” samples
tA patient population that includes only one age group or only one disease state will
yield different target values than a patient population that includes all groups
4-26
PN 4237530BB
Page 81
QUALITY CONTROL
ADDITIONAL QC CHECKS
rAs long as the patient population remains constant, the target values of each index also
remain constant
rIf the patient population changes, the mean of an index may also change and needs to be
reevaluated
Dr. Bull's Target Values
rDefault starting values
tMCV = 89.5
tMCH = 30.5
tMCHC = 34.0
rEach laboratory should begin with these target values and then evaluate and adjust them
for their own patient population
Getting Started
Enter Target Values and Action Limits
rTarget values and action limits must be entered in the DMS:
Special Functions tt Set Up tt Sample Analysis Set Up tt Action Limits tt XB Limits
4
rUse either the suggested target values or the target values established by your laboratory
r3% action limits are suggested
Turning the XB Option ON
rX
Analysis is done automatically by the DMS only if the option is turned on
B
rOn the Run Samples screen, check the Status line
t
XB↑ indicates the X
XB↓ indicates the X
t
Analysis is turned ON
B
Analysis is turned OFF
B
rTo change the setting,
t
Sample Analysis tt Run Samples; press Í Optns then Ê to change XB ON/OFF setting
rX
Analysis is automatically turned OFF when the Run Samples ÏDisplay Only option
B
is ON
Using the XB Analysis
When XB function is turned on, the
tDMS stores the RBC results of all patient samples as they are cycled
tX
Analysis is performed on small sets of 20 patient results at regular intervals
B
(each patient set is referred to as a batch)
PN 4237530BB
rStored results (RBC, Hgb, Hct, MCV, MCH, and MCHC) are displayed on the Current XB
Batch table:
Sample Analysis tt XB tt Current XB Batch
rOf the displayed results, only the MCV, MCH, and MCHC parameters are used in the
analysis
4-27
Page 82
QUALITY CONTROL
ADDITIONAL QC CHECKS
rHaving the other RBC parameters displayed allows the operator to judge the validity of
the sample results when an out-of-limits situation occurs
rWhen the batch of 20 patient samples is completed, the DMS performs the X
and determines the batch mean of each index (MCV, MCH, and MCHC)
tMethod for obtaining the batch mean is rather complicated, but the calculation is
tThis batch mean is not a simple average value of the patients’ indices, but a type of
rDMS then compares each batch mean to its expected target value
Analysis
B
done automatically by the DMS
“weighted moving average” that statistically allows the small batch of 20 patient
samples to appear as if it were a larger set of approximately 100 samples
tResults are displayed on the XB Batch Means table:
Means
Sample Analysis tt XB tt XB Batch
tIf the batch mean is inside the expected action limits, the batch is said to be IN
tInstrument is functioning properly
t
XB↑ status message remains white to indicate the last completed batch was
inside the action limits
tIf the batch mean is outside the expected action limits, the batch is said to be OUT
tOne or more of the three indices did not come close enough to the expected
target values
tAn
t
H or L flag appears beside the out-of-limit results
XB↑ status message turns red to indicate the last completed batch was outside
the action limits
t
XB↑ status message will return to white when the next batch of XB is IN
Reviewing XB Data
Current XB Batch
rStores RBC parameter results when X
rWhen a batch of 20 samples is collected, the DMS performs the X
calculates the batch mean for MCV, MCH, and MCHC
tThe system automatically excludes samples that do not yield numeric results such
as partial aspirations
rOption primarily used for reviewing individual sample results to identify sample related
problems such as non-reportable results or non-random sampling
tOnce the batch is completed (20 patient samples are accumulated), individual
samples cannot be deleted
tDeleted sample results and partial aspiration codes are displayed and consume one
of the 20 sample spaces; however, these results are not used in the X
calculation
Analysis is turned on
B
Analysis and
B
batch mean
B
4-28
tA maximum of 5 samples may be deleted from the current batch prior to
completion
tA minimum of 15 samples are required to calculate a batch mean
PN 4237530BB
Page 83
QUALITY CONTROL
ADDITIONAL QC CHECKS
XB Batch Means Option
rUsed to view the calculated batch means for the last batch of 20 samples collected
rAlso provides the percent difference between the calculated batch mean and its
corresponding target value
rIf a percent difference is outside the action limit, the parameter is flagged with an
rMay be used to identify the affected RBC indices and to determine the direction and
amount of the change
rAfter the first 20 batches are completed, this option always displays the last 20
completed batches
tProvides a history of recent batch means which may serve as an aid during the
troubleshooting process
tAfter the 20
results are deleted to make space for the last completed batch results
XB Graphs
rThree graphs, one for each index
rUsed to view the batch means in relationship to the target value and action limit of the
corresponding index
rGraphs are good for identifying trends and shifts
rEach graph contains three areas
th
batch is completed, the table begins to rollover where the oldest batch
H or L
4
tA central white line representing the actual target value
tA black area representing the acceptable range
tAn upper and lower red line represent action limit boundaries
rAfter the first 20 batches are completed, this option always displays the plot points for
the last 20 completed batches
tProvides a history of recent batch means which may serve as an aid during the
troubleshooting process
tAfter the 20
batch results are deleted to make space for the last completed batch results
th
batch is completed, the graph begins to rollover where the oldest
What Causes a Batch to go “Out-of-Limits”?
rChange in the patient population
tOne or more types of patients added or removed from the total patient population
mix
tNon-reportable results where a batch of patients is biased by short samples,
physiologically impossible results and/or erroneous data accumulated before a
reagent or instrument problem was identified and solved
tNon-random patient sampling where a batch of patients is biased by several
abnormal patients of one particular type (oncology, neonatal and so forth)
rAn instrument problem may exist
rA calibration problem may exist
rA reagent problem may exist
PN 4237530BB
4-29
Page 84
QUALITY CONTROL
ADDITIONAL QC CHECKS
Troubleshooting when an XB Batch is “Out”
Consider a Change in the Patient Population
rIs there a change in the overall patient population?
tOne or more new patient types were added to the population
tOne or more patient types are no longer part of the population
tPossible seasonal change of the patient population (hospitals or clinics in resort
rIf a change in the overall patient population is probable,
tFuture XB batch results will also be outside the limits
tNew target values based on the current population must be established and entered
rIf a change in the overall patient population is not probable,
tIt is possible that the patient population has not really changed but just appears to
tUse the
areas)
in the computer
have changed because the batch of 20 patients does not truly represent the total
population
Current XB Batch option to check for non-reportable results and/or
non-random sampling:
Sample Analysis tt XB tt Current XB Batch
Checking for Non-Reportable Results
rAn XB batch may go outside the ±3% action limits because the batch is biased by
physiologically impossible results caused by a reagent or instrument problem, or by
physiologically abnormal specimen
Consider an Instrument Problem Exists
rConsider an instrument problem only if
tA change in the patient population is doubtful
tAll the results in the Current XB Batch are reportable (physiologically possible)
tRandom sampling is verified
rInstrument problems cause XB batch results to go “out” and stay out
tCan be a gradual change
tIf caused by a calibration drift, should go back within limits after calibration
tIf caused by a component going bad over time, should go back within limits
after the component is replaced
tCan be a sudden change
tIf caused by a component failure, should go back within limits after the
component is replaced
tIf caused by any number of instrument problems, should go back within limits
after the problem is fixed
4-30
PN 4237530BB
Page 85
QUALITY CONTROL
ADDITIONAL QC CHECKS
Assessing the Situation
rReview the XB Graphs and XB Batch Means options to identify the out-of-limits RBC
indices and note the direction of the change (increased or decreased)
tGenerally two of the three RBC indices are affected
tIt is also important to note any index that may still be inside the limit but shows a
significant change
Identifying the Parameter Causing the Problem
rThere are two ways to identify the parameter causing the problem
tKnowledge and analysis of the involved parameters
tUse of a table showing the relationship between the RBC Indices and their
associated parameters
Analysis of Involved Parameters
rMCV parameter is derived from the RBC Histogram
tAlthough derived from the histogram, the size of the individual red cells is
determined using the Coulter Principle which is a direct measurement
4
Hgb
r
r
MCH
MCHC
-----------RBC
Hgb
---------Hct
10×=
100×=where the Hct
RBCMCV×
-------------------------- -----=
10
rRBC indices are based on three directly measured parameters: MCV, RBC, and Hgb
By Use of a Parameter Relationship Table
rTable relates the RBC Indices and their associated parameters
MCV
LOW
MCVdecinc------incdec
RBC------incdecincdec
Hgb------decincdecinc
Hctdecinc------incdec
MCV
HIGH
MCH
LOW
MCH
HIGH
MCHC
LOW
MCHC
HIGH
rOnce the operator identifies the affected RBC indices and notes the direction of the
change (increased or decreased), the table reveals the suspected parameters (MCV, RBC,
or Hgb)
PN 4237530BB
4-31
Page 86
QUALITY CONTROL
ADDITIONAL QC CHECKS
NOTES
4-32
PN 4237530BB
Page 87
OBJECTIVES
When the subject is complete, you will be able to . . .
rExplain the purpose of calibration.
rTell when one needs to verify calibration.
rPerform the preliminary procedures.
rPerform the calibration procedure using the Special Procedures and Troubleshooting
(SPT) Manual.
CALIBRATION
5
5
PN 4237530BB
5-1
Page 88
CALIBRATION
OBJECTIVES
NOTES
5-2
PN 4237530BB
Page 89
CALIBRATION
rA procedure used to standardize the instrument by determining its deviation from
rNever use calibration to adjust for an instrument problem
rVerify the ambient room temperature is typical for the laboratory and within the
rIn the normal process of tracking data for an extended period of time, your laboratory
rNever adjust to a specific value based on an individual sample result
When to Calibrate
rAt installation, before you begin analyzing samples
rAfter you replace any component dealing with
rIf your Beckman Coulter Representative suggests you calibrate
calibration references and applying any necessary correction factors
instrument’s operating range (16 to 32°C; 60 to 90°F)
can make a specific decision to recalibrate a given parameter
tDilution preparation, such as the BSV
tPrimary measurement, such as an aperture
CALIBRATION
CALIBRATION
5
When to Verify Calibration
rAs dictated by your laboratory procedures, local, or national regulations
rWhen controls show evidence of unusual trends
rWhen controls exceed the manufacturer’s defined acceptable limits
rWhen average ambient room temperature changes more than 10°F (5.5°C) from the
room temperature during your last calibration
PRELIMINARY CALIBRATION PROCEDURES
rProcedures designed to ensure the instrument is operating optimally
rRecommend a thorough review of data stored in current control and X
possible bias that might require a change in a calibration factor
rThorough review of IQAP data is also recommended
files to note any
B
PN 4237530BB
5-3
Page 90
CALIBRATION
CALIBRATION PROCEDURE
Procedures to be Completed Prior to Calibration
rCheck reagent volumes to ensure sufficient volume to complete calibration procedures
rEnsure the instrument is clean
tIf the instrument is routinely shut down with COULTER CLENZ cleaning agent for
at least 30 minutes every 24 hours it is in use, initiate the
tIf the instrument is not routinely shut down with COULTER CLENZ cleaning agent
for at least 30 minutes every 24 hours the instrument is in use, the apertures must
be bleached as instructed in the SPT’s Cleaning Procedures chapter