Chemistry Analyzer
Reaction Profile Check for Cuvette
Overflow Detection
PN B06960AA
March 2012
Page 2
AU2700® and AU5400® Chemistry Analyzer
Beckman Coulter Ireland Inc.
Lismeehan
O'Callaghans Mills
Co. Clare
Ireland
Beckman Coulter do Brasil Com e Imp de Prod de Lab Ltda
Estr dos Romeiros, 220 - Galpao G3 - Km 38.5
06501-001 - Sao Paulo - SP - Brasil
CNPJ: 42.160.812/0001-44
2.1.1 SMS Test as a QC Test (Green Racks on the AU2700/AU5400, or
the Stat Table QC on the AU2700),2-2
2.1.2 SMS Test as a Routine Test (White Racks),2-2
2.2 Specific Test Parameters,2-4
2.3 QC Test Parameters (Optional),2-5
2.4 Calibration Specific Test Parameters,2-6
2.5 Reaction Profile Check (RPC) Parameters,2-7
CHAPTER 3:Troubleshooting,3-1
3.1 Confirm a Cuvette Overflow Has Occurred,3-2
3.2 Recovering from a Cuvette Overflow,3-3
3.2.1 What Causes an Overflow?,3-3
3.2.2 Recognizing an Overflow,3-3
3.2.3 Items to Check when Recovering from an Overflow,3-4
3.2.4 After the Overflow is Fixed,3-4
3.3 Troubleshooting When a Cuvette Overflow Did Not Occur,3-5
CHAPTER 4:Monitoring Results,4-1
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow,4-2
4.1.1 Searching for Samples if the SMS test is a QC Sample for
Condition A,4-3
4.1.2 Searching for Samples if the SMS test is a QC Sample for
iii
Page 4
Contents
Condition B,4-10
4.1.3 Searching for Samples if the SMS test is a Routine Sample for
Condition A,4-15
4.1.4 Searching for Samples if the SMS test is a Routine Sample for
Condition B,4-22
CHAPTER 5:As Needed Maintenance,5-1
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an
Overflow,5-2
5.1.1 Remove the Cuvette Wheel and Cuvettes,5-2
5.1.2 Clean the Cuvettes,5-6
5.1.3 Replace the Cuvette Wheel into the Incubation Bath,5-8
iv
Page 5
CHAPTER 1
Introduction
This addendum is a separate document containing changes and additions to the AU2700 and
AU5400 User's Guides. The changes and additions in this addendum pertain to the Cuvette Overflow
Modification using the Reaction Profile Check (RPC) method.
This chapter describes the principle of the RPC and how to operate effectively with the RPC method.
1.1 Reaction Profile Check (RPC) Description
1.2 Cuvette Overflow Detection Principle
1.3 Operation with Reaction Profile Check
PN B06960AA
1-1
Page 6
Introduction
1.1 Reaction Profile Check (RPC) Description
1.1 Reaction Profile Check (RPC) Description
A cuvette overflow occurs when fluid spills onto the outside of the cuvette. This may affect the
accuracy of subsequent test results.
The purpose of the RPC cuvette overflow detection method is to detect a cuvette overflow so
appropriate corrective actions can be taken immediately.
The principle of the RPC method is a new software algorithm used specifically with System
Monitoring Solution (SMS) to detect abnormal reaction kinetics caused by a cuvette overflow by
running an SMS test.
Beckman Coulter recommends manually requisitioning and running the SMS test at regular
intervals defined by Standard Operating Procedures of each laboratory. It is possible to requisition
the SMS test through the LIS.
The SMS test detects abnormal reaction kinetics which are indicative of a cuvette overflow. The
cuvette overflow check is only performed during the measurement of the SMS test. If abnormal
reaction kinetics are detected, the SMS test generates a flag (*), and an audible alarm message is
generated to indicate a possible cuvette overflow, or other analyzer error.
1.1.1 Cuvette Overflow Alarms
The alarm message "4064 REACTION PROFILE CHECK ERROR (aa, bbbb, cc, ddddd, e, ffff, ggg..gg)
hhh....hh" is generated, followed by a second alarm "4065 REACTION PROFILE CHECK CONSECUTIVE
The SMS test is run to detect abnormal reaction kinetics that can be caused by a cuvette overflow.
Beckman Coulter recommends the SMS test be performed during daily operations at a frequency
based on the Standard Operating Procedures and workflow of the laboratory.
•In Figure 1.1, the green cuvettes show where the SMS test takes place during a routine run on
an AU system.
•The (*) indicates that the analyzer has detected an abnormal reaction kinetic.
Figure 1.1 Cuvette Overflow Detection Concept
Introduction
1.1 Reaction Profile Check (RPC) Description
1
PN B06960AA
TIP If the analyzer goes to Stop mode before the requisitioned SMS test is complete, include troubleshooting
for a cuvette overflow as part of the investigation. Run the SMS test once the analyzer is functional.
1-3
Page 8
Introduction
1.2 Cuvette Overflow Detection Principle
1.2 Cuvette Overflow Detection Principle
Detection of abnormal reaction kinetics is based on flagging the elevated optical density (OD)
measured at P0 of the SMS test, which occurs during a cuvette overflow. P0 is the first available
photometric measuring point.
The OD at P0 of the SMS test is approximately 0.0000 OD when the AU is operating without a cuvette
overflow. This is due to the formulation of the SMS test without a reactive component at the
measured wavelength (800 nm).
The OD at P0 and subsequent measuring points of the SMS test becomes unstable when a cuvette
overflow occurs. This is due to the light being scattered, reflected, or refracted by the overflow.
Figure 1.2 demonstrates the typical unstable OD values throughout the reaction for overflow
reaction kinetics with the SMS test.
Figure 1.2 SMS Reaction Curve: Non Overflow and Overflow
1-4
PN B06960AA
Page 9
1.3 Operation with Reaction Profile Check
1.3.1 System Monitor Solution Assay Protocol
The RPC method to detect abnormal reaction kinetics caused by a cuvette overflow is performed
only on the SMS test. The SMS reagent part number B15601 is the only product specifically designed
to flag abnormal reaction kinetics caused by a cuvette overflow.
Beckman Coulter recommends the SMS test be performed during daily operations at a frequency
based on the Standard Operating Procedures and workflow of the laboratory, and at the beginning
and ending of each sample run. An example is to run the SMS test before patient samples are
measured, as part of routine QC, at regular sample intervals, and after the patient sample run is
complete.
It is necessary to run the SMS test on the Inner and Outer cuvette wheel.
The following sample types may be used as sample material for the SMS test:
Introduction
1.3 Operation with Reaction Profile Check
1
1. DI water
2. Quality control material
Beckman Coulter recommends performing the SMS test as QC from the green rack on the AU2700/
AU5400 or from the STAT table on the AU2700. The SMS test can also be performed from a routine
(white rack) sample.
The SMS assay requires a reagent blank with every new lot number of reagent.
PN B06960AA
1-5
Page 10
Introduction
1.3 Operation with Reaction Profile Check
1-6
PN B06960AA
Page 11
CHAPTER 2
Parameters
This chapter describes the parameters for the System Monitoring Solution (SMS) test.
The SMS Test Parameters and Reaction Profile Check Parameters are defined by Beckman Coulter.
Use the SMS Test parameter sheet to program the SMS Test parameters.
2.1 Common Test Parameters
2.2 Specific Test Parameters
2.3 QC Test Parameters (Optional)
2.4 Calibration Specific Test Parameters
2.5 Reaction Profile Check (RPC) Parameters
PN B06960AA
2-1
Page 12
Parameters
2.1 Common Test Parameters
2.1 Common Test Parameters
Refer to the AU2700 or AU5400 User's Guides for how to program test parameters. In the US and AP
markets, refer to Parameter Menu Operation in Chapter D, part 3. In the EU market, refer to
Configuring Tests in Chapter 4.
There are two options for programming and running the SMS test. Beckman Coulter recommends
performing the SMS test using QC from the green rack on the AU2700/AU5400 or stat table on the
AU2700.
1. SMS Test as a QC Test (Green Racks on the AU2700/AU5400, or the Stat Table QC on the AU2700)
2. SMS Test as a Routine Test (White Racks)
2.1.1 SMS Test as a QC Test (Green Racks on the AU2700/AU5400, or the Stat Table
QC on the AU2700)
1
Select Parameter > Common Test Parameters > Test Name.
Test Name:SMS
Long Name:System Monitor
Reagent ID:220
Use Cuv.:Both
Alarm Shots:32
For the AU5400, a bottle of SMS must be placed on each unit.
2
Select Parameter > Common Test Parameters > Round. Select SMS for all rounds and all units.
2.1.2 SMS Test as a Routine Test (White Racks)
1
Select Parameter > Common Test Parameters > Test Name.
Test Name:SMS*
Long Name:System Monitor
2-2
Reagent ID:220
Use Cuv.:**
Alarm Shots:32
* AU2700: It is necessary to program an SMS test for the Inner and Outer cuvette wheel, for
example SMSI and SMSO.
PN B06960AA
Page 13
Parameters
2.1 Common Test Parameters
* AU5400: It is necessary to program an SMS test for the Inner and Outer cuvette wheel and each
unit, for example 1SMSI, ISMSO, 2SMSI, 2SMSO, 3SMSI, and 3SMSO.
** The SMS test must be programmed on both the Inner and Outer cuvette wheel for the
AU2700, and for the Inner and Outer cuvette wheel and each unit for the AU5400.
2
Select Parameter > Common Test Parameters > Round. Select SMS for all rounds and all units.
2
PN B06960AA
2-3
Page 14
Parameters
2.2 Specific Test Parameters
2.2 Specific Test Parameters
Program the Specific Test Parameters.
1
Select Parameter > Specific Test Parameters.
2
Program the SMS test according to the SMS parameter sheet.
Enter the values according to the screen shot below.
Figure 2.1
2-4
•Test Name is user-defined depending on the option (1 or 2) selected for programming and
running the SMS test. Option 1 and 2 are described in section 2.1 Common Test Parameters.
Examples for Option 1 and 2:
Option 1 AU2700 and AU5400
Test Name:
Option 2 AU2700 Test Name:SMSI and SMSO (2 tests are programmed)
Option 2 AU5400 (2 units) Test
Name:
Option 2 AU5400 (3 units) Test
Name:
SMS (1 test is programmed)
1SMSI, ISMSO, 2SMSI, 2SMSO (4 tests are programmed)
1SMSI, ISMSO, 2SMSI, 2SMSO, 3SMSI, 3SMSO (6 tests are
programmed)
•The SMS sample type is user-defined. Select the sample type in use at the "Type" dropdown list.
PN B06960AA
Page 15
2.3 QC Test Parameters (Optional)
Programming the SMS as a QC sample or test is optional, based on selecting option 1. SMS Test as a
QC Test (Green Racks on the AU2700/AU5400, or the Stat Table QC on the AU2700) in 2.1 Common
Test Parameters. Beckman Coulter recommends performing the SMS test using QC from the green
rack on the AU2700/AU5400, or the stat table on the AU2700.
The QC sample for the SMS test may be an existing QC material, or an SMS QC Sample using DI water
can be programmed. Following is an example for programming an SMS QC Sample using DI water
for the SMS test.
1
Select Parameter > QC Control > QC Common > Control.
Control Name:SMS QC Sample
Control ID:User Defined
Expiration:User Defined
Parameters
2.3 QC Test Parameters (Optional)
2
Lot No.:No Entry: DI water
Multi Rack check box:Check if the multi-rack option is in use for other QC samples (only
available on the AU5400)
2
Select Parameter > QC Control > QC specific > Preset.
Control Name:SMS Sample
Multi/Single:Single
Mean:0.0
SD:1
Range:4
PN B06960AA
2-5
Page 16
Parameters
2.4 Calibration Specific Test Parameters
2.4 Calibration Specific Test Parameters
1
Select Parameter > Calibration > Calibration Specific. Refer to the SMS (B15601) IFU.
Test Name:SMS
Type:User Defined
Calibration Type:MB
MB Type Factor:0
Count:1
Advanced Calibration:No
Calibration Stability Period:Empty
2
Select Inner/Outer (F6) to program the MB Type Factor (0) for both the inner and outer cuvette
wheel.
3
For the AU5400, select Unit No. and program the MB Type Factor (0) for the Inner and Outer
wheel on each unit.
2-6
PN B06960AA
Page 17
2.5 Reaction Profile Check (RPC) Parameters
The Reaction Profile Check Parameters for SMS are defined by Beckman Coulter and are located in
the SMS IFU. For detailed procedures, contact Beckman Coulter Technical Services.
1
Select Parameter > Special > Data Check Parameters > Reaction Profile Check.
2
Select Set (F4).
Figure 2.2
Parameters
2.5 Reaction Profile Check (RPC) Parameters
2
PN B06960AA
3
Select SMS from the "Test Name" drop-down list.
4
Select Sample/CAL/QC from the "Sample Kind" drop-down list
5
Select the user-defined sample type from the "Type" drop-down list.
6
Enter the RPC parameters by referring to the SMS IFU.
7
Select Exit (F2).
2-7
Page 18
Parameters
2.5 Reaction Profile Check (RPC) Parameters
2-8
PN B06960AA
Page 19
CHAPTER 3
Troubleshooting
This Chapter contains new information on how to confirm a cuvette overflow, what can cause a
cuvette overflow, how to recognize an overflow, and corrective actions for how to recover from the
overflow. The possible causes and corrective actions if a cuvette overflow did not occur are
described.
3.1 Confirm a Cuvette Overflow Has Occurred
3.2 Recovering from a Cuvette Overflow
3.2.1 What Causes an Overflow?
3.2.2 Recognizing an Overflow
3.2.3 Items to Check when Recovering from an Overflow
3.2.4 After the Overflow is Fixed
3.3 Troubleshooting When a Cuvette Overflow Did Not Occur
PN B06960AA
3-1
Page 20
Troubleshooting
3.1 Confirm a Cuvette Overflow Has Occurred
3.1 Confirm a Cuvette Overflow Has Occurred
A cuvette overflow may have occurred before the SMS test that generated the "*" flag and 4064
REACTION PROFILE CHECK alarm message. Refer to the flowchart to confirm a cuvette overflow has
occurred.
To confirm a cuvette overflow has occurred, remove the cuvette wheel cover. The cuvettes should
appear frosty or white. If the cuvettes appear dark, black, or are wet when removed, a cuvette
overflow has occurred.
Figure 3.1
3-2
PN B06960AA
Page 21
3.2 Recovering from a Cuvette Overflow
The following procedure explains what can cause an overflow, how to recognize an overflow, and
corrective actions for how to recover from the overflow.
Performing scheduled maintenance correctly reduces the chances of a cuvette overflow. For more
TIP
information about maintenance for each system component, refer to the Maintenance Chapter in the
AU2700 or AU5400 User's Guide.
3.2.1 What Causes an Overflow?
A cuvette overflow may be caused by any of the following:
•A wash nozzle is clogged or partially clogged. When this happens, liquid is not aspirated from
the cuvette completely and eventually liquid spills over the side. This can occur when the wash
nozzles are not cleaned properly, or when particles such as glass are aspirated into the nozzle.
•A wash nozzle is bent or damaged.
•There are damaged or missing o-rings on the tube mounting joint manifold.
Troubleshooting
3.2 Recovering from a Cuvette Overflow
3
•The reagent probe is bent. A bent probe could be dispensing outside of the cuvette.
•The sample probe is bent. A bent probe could be dispensing outside of the cuvette.
•Cuvettes are chipped or cracked due to alignment problems with the reagent probes or wash
nozzles.
•The wash nozzle tubing is not connected to the nozzle.
3.2.2 Recognizing an Overflow
A "4064 Reaction Profile Check Error" alarm and a "4065 Reaction Profile Check Consecutive
Judgment Error" alarm are generated.
For detailed instructions, refer to section 4.1 Identifying and Re-analyzing Samples after a Cuvette
Overflow in chapter 4 of this addendum.
The cluster of error flags *, ?, @, $, D, B, and ! may indicate a cuvette overflow. The data, alarms and/
or flags will vary depending on the severity of the overflow. One or all tests could be affected by an
overflow. Items to check:
•QC error flags or alarms
•Reagent blank error flags
•Analyzer not performing as normal operation
PN B06960AA
•Numerous cuvettes fail after photocal
•Run an SMS test to confirm a cuvette overflow
3-3
Page 22
Troubleshooting
3.2 Recovering from a Cuvette Overflow
3.2.3 Items to Check when Recovering from an Overflow
WARNING
Immediate attention should be given to an overflow. If nothing is done to fix the
problem, the wheel will continue to overflow. Contact Beckman Coulter Technical
Services for assistance with performing these checks.
•The wash nozzle should be aligned over the cuvettes. Visually inspect and ensure the wash
nozzles are centered over the cuvettes and check the alignment.
•Sonicate and clean the wash nozzles with a stylus to remove any debris. Refer to the
Maintenance Chapter, Clean the Cuvette Wash Nozzle procedure in the AU2700 or AU5400 User's
Guide.
•Check the reagent and sample probes to be sure they are properly aligned. Rotate the sample
and reagent probes over the cuvette wheel.
•Check for chipped or cracked cuvettes. Replace them if necessary.
For information about replacing cuvettes, refer to section 5.1 Clean the Cuvettes, Cuvette
Wedges, and the Cuvette Wheel after an Overflow in chapter 4 of this Addendum.
•Verify the wash nozzle tubing connections are secure.
•Verify the o-rings on the manifolds are in place and not damaged.
3.2.4 After the Overflow is Fixed
After the cuvette overflow is fixed, refer to section 5.1 Clean the Cuvettes, Cuvette Wedges, and the
Cuvette Wheel after an Overflow in chapter 4 of this addendum.
After the cuvettes, cuvette wedges, and cuvette wheel are clean, perform the following:
1
Perform a Photocal. Refer to the Maintenance Chapter in the AU2700 or AU5400 User's Guide.
2
Run an SMS test to verify the overflow is fixed.
3
The results from the previous non-flagged SMS test and the SMS test that generated the "*" flag
and 4064 Reaction Profile Check alarm are invalid and must be analyzed again. Refer to 4.1
Identifying and Re-analyzing Samples after a Cuvette Overflow for the appropriate procedure.
3-4
PN B06960AA
Page 23
3.3 Troubleshooting When a Cuvette Overflow Did Not Occur
3.3 Troubleshooting When a Cuvette Overflow Did Not Occur
If the analyzer is performing abnormally, and the SMS test did not generate an "*" flag, review
analyzer data by looking at the following items:
1. Photocal and Photometer Lamp Check: Perform a photocal and photometer lamp check to
verify the integrity of the cuvettes and lamp. Refer to the AU2700 and AU5400 User's Guides,
Maintenance Chapter.
2. Reagent Blank, Calibration, and QC Data: Review reagent blank, calibration, and QC data for
accuracy and precision. Refer to the Troubleshooting and Software Chapters in the AU2700 and
AU5400 User's Guides.
3. ?, @, , $, D, B, and ! Error Flags: Review data for one or more of the listed error flags. Refer to the
Error Flags Chapter in the AU2700 or AU5400 User's Guide for a description and corrective
action for the flags.
If it is confirmed that a cuvette overflow did not occur but the SMS test generated a "*" flag and the
analyzer went to Stop mode, the possible root cause of the analyzer abnormality could be a
contaminated SMS reagent bottle, a deteriorating lamp, or a damaged mix bar.
Troubleshooting
3
1. SMS Reagent
Replace the SMS reagent and perform a reagent blank.
2. Lamp
Try to confirm how long the lamp has been in use. If the lamp has been in use over 1,000 hours,
replace the lamp.
Refer to the AU2700 or AU5400 User's Guide, Maintenance Chapter to replace the lamp.
3. Mix Bar
Refer to the AU2700 or AU5400 User's Guide, Maintenance Chapter to replace the mix bar.
PN B06960AA
3-5
Page 24
Troubleshooting
3.3 Troubleshooting When a Cuvette Overflow Did Not Occur
3-6
PN B06960AA
Page 25
CHAPTER 4
Monitoring Results
This chapter describes how to identify samples affected by a cuvette overflow. Results from samples
affected by an overflow are invalid and must be analyzed again.
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
4.1.1 Searching for Samples if the SMS test is a QC Sample for Condition A
4.1.2 Searching for Samples if the SMS test is a QC Sample for Condition B
4.1.3 Searching for Samples if the SMS test is a Routine Sample for Condition A
4.1.4 Searching for Samples if the SMS test is a Routine Sample for Condition B
PN B06960AA
4-1
Page 26
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
A cuvette overflow may have occurred before SMS test that generated the "*" flag and 4064
REACTION PROFILE CHECK alarm message.
If a cuvette overflow is confirmed, the results obtained from the previous non-flagged SMS test and
the SMS test that generated the "*" flag and 4064 REACTION PROFILE CHECK alarm are invalid and
must be analyzed again.
There are two options with two conditions to search for samples affected by the cuvette overflow
described in procedures 4.1.1, 4.1.2, 4.1.3, and 4.1.4.
NOTE
It is only necessary to follow one procedure, based on the Option (1 or 2) and Condition programmed
and in use in the laboratory.
1. Running the SMS as a QC sample. There are two conditions:
Condition A:
•Barcode analysis mode and realtime Online requisition
•Sequential analysis mode
Use the 4.1.1 Searching for Samples if the SMS test is a QC Sample for Condition A procedure.
Condition B:
•Barcode analysis mode and batch online or manual requisition
•Rack No. analysis mode
Use the 4.1.2 Searching for Samples if the SMS test is a QC Sample for Condition B procedure.
2. Running the SMS as a routine sample. There are two conditions:
Condition A:
•Barcode analysis mode and realtime Online requisition
•Sequential analysis mode
Use the 4.1.3 Searching for Samples if the SMS test is a Routine Sample for Condition A
procedure.
Condition B:
4-2
•Barcode analysis mode and batch online or manual requisition
•Rack No. analysis mode
Use the 4.1.4 Searching for Samples if the SMS test is a Routine Sample for Condition B
procedure.
PN B06960AA
Page 27
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
WARNING
If a cuvette overflow is confirmed, all samples measured after the previous
nonflagged SMS test and before the SMS test that generated the "*" flag and 4064
REACTION PROFILE CHECK alarm are invalid and must be analyzed again. If these
results have already been transferred to the Host Computer or reported, take
corrective actions according to the facility Standard Operating Procedures.
4.1.1 Searching for Samples if the SMS test is a QC Sample for Condition A
1
Select Routine >Reaction Monitor > Cal/QC.
Figure 4.1
4
PN B06960AA
2
Select SMS from the "Test Name" drop-down list.
3
For the AU5400 only, select the unit number that generated the 4064 alarm from the "Unit No."
drop-down list.
4
Select the last Index from the "Index" drop-down list.
5
Select Cuvette/Mix No. (F7). The Select Cuvette/Mix No. window appears.
4-3
Page 28
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.2
6
Select Inner or Outer that generated the 4064 alarm from the "Cuvette" drop-down list, then
OK.
select
7
Select (check the box) for QC.
8
Select Search Data (F5). The Reaction Monitor window appears.
Figure 4.3
1. Flag Display field
2. Backward arrow
3. Forward arrow
4-4
9
Select the forward arrow button to go to the last result (SMS test with the "*" flag). In this
example, go to result 8/8.
10
Select the backward arrow button to go to the previous SMS test that did not have a "*" flag.
The "*" flags displays after the Result.
PN B06960AA
Page 29
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
11
Note the date and time from the Measured Time field.
12
Select Exit (F2).
13
Select Routine>Data Report > Data List > Patient.
Figure 4.4
Monitoring Results
4
PN B06960AA
14
Select the index containing the 4064 alarm that was generated from the "Index" drop-down list.
Note the sample numbers of each sample kind (Routine, Emergency, or STAT) for Serum, Urine,
or Others measured in the index. STAT is applicable only for the AU2700.
15
Select Routine > Data Report > Data List > QC/Cal.
4-5
Page 30
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.5
16
Select the index containing the 4064 alarm that was generated from the "Index" drop-down list.
Note the sample numbers of all QC, reagent blank, and calibration samples analyzed in the
index.
17
Select Exit (F2).
18
Select Routine > Reaction Monitor.
Figure 4.6
4-6
PN B06960AA
Page 31
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
19
Select the following fields:
Test Name:(All).
Unit No.:Only for the AU5400. The unit number that the 4046 alarm generated.
Index:The index that the 4064 alarm generated.
Check box:Check one sample kind (Routine, Emergency, or STAT) for Serum, Urine, or
Others measured in the index. STAT is applicable only for the AU2700.
Sample No.: First column: "*" or approximate sample number.
Second column: "*" or approximate sample number.
Check only one sample kind at a time so it is easier to identify the approximate first Sample No.
TIP
required to search. Perform this check one sample kind at a time for all the sample kinds measured
in the index.
TIP If there is a large number of samples in the sample kind checked to search, the search takes a long
time. Enter an approximate starting (first column) and ending (second column) Sample No. to search
to reduce the search time. For example, the maximum number of photometric tests per 60 minutes
is 1600. If the average number of tests per sample is 8, the approximate number of samples
processed per 60 minutes is 200 samples. Searching the last 220 samples (200 plus 10%) should be
sufficient to find the results. If the search does not contain samples measured for 60 minutes before
the 4064 alarm was generated, it is necessary to increase the range of samples to search and
perform the search again.
4
20
If Inner or Outer displays in the Cuvette field, select Cuvette/Mix No. (F7).
Figure 4.7
•Select All from the "Cuvette" drop-down list.
•Do not change the "*" at Mix No. for R1, Sample, and R2.
•Select
21
Select Search Data (F5).
OK.
PN B06960AA
4-7
Page 32
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.8
22
Use the arrow buttons and look at the Measured Time to find the first Sample No. measured
after the time obtained in step 10.
23
Note the Sample No.
24
Repeat steps 18 through 23 for all the sample kinds (Routine, Emergency, or STAT) for Serum,
Urine, Others, QC, reagent blank, and calibration.
It is necessary to select the Cal/QC tab to search for reagent blank, calibration, and QC results.
TIP
25
Select Exit (F2) to go back to the Reaction Monitor screen.
26
Select Exit (F2).
27
Select Routine > Data Report > Data List.
4-8
PN B06960AA
Page 33
Figure 4.9
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
4
28
Select the check box for all the sample kinds (Routine, Emergency, or STAT) for Serum, Urine,
and Others measured in the index. STAT is applicable only for the AU2700.
29
Enter the first Sample No. (first column) identified in the previous steps for each sample kind.
Keep "*" as the last Sample No. (second column).
30
Select Print (F3).
The results measured after the previous SMS test before the SMS test that generated the "*" flag
and 4064 REACTION PROFILE CHECK alarm print. These results may be affected by the cuvette
overflow and are invalid. Analyze these samples again.
31
Select Routine > Data Report > QC/Cal tab and repeat steps 27 through 30.
PN B06960AA
4-9
Page 34
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
4.1.2 Searching for Samples if the SMS test is a QC Sample for Condition B
1
Select Routine>Reaction Monitor > Cal/QC.
Figure 4.10
2
Select SMS from the "Test Name" drop-down list.
3
For the AU5400 only, select the unit number that generated the 4064 alarm from the "Unit No."
drop-down list.
4
Select the last Index from the "Index" drop-down list.
5
Select Cuvette/Mix No. (F7). The Select Cuvette/Mix No. window appears.
Figure 4.11
4-10
PN B06960AA
Page 35
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
6
Select Inner or Outer that generated the 4064 alarm from the "Cuvette" drop-down list, then
select
OK.
7
Select QC.
8
Select Search Data (F5). The Reaction Monitor window appears.
Figure 4.12
4
PN B06960AA
1. Flag Display field
2. Backward arrow
3. Forward arrow
9
Select the forward arrow button to go to the last result (SMS test with the "*" flag). In this
example, go to result 8/8.
10
Select the backward arrow button to go to the previous SMS test that did not have a "*" flag.
The "*" flags displays after the Result.
11
Note the Measured Time.
12
Select Exit (F2).
13
Select Routine > Data Report > Data List > Patient.
4-11
Page 36
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.13
14
Select the index containing the 4064 alarm that was generated from the "Index" drop-down list.
Note the sample numbers of each sample kind (Routine, Emergency, or STAT) for Serum, Urine,
or Others measured in the index. STAT is applicable only for the AU2700.
15
Select Routine > Data Report > Data List > QC/Cal.
Figure 4.14
4-12
PN B06960AA
Page 37
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
16
Select the index containing the 4064 alarm that was generated from the "Index" drop-down list.
Note the sample numbers of all QC, reagent blank, and calibration samples analyzed in the
index.
17
Select Exit (F2).
18
Select Routine > Reaction Monitor.
Figure 4.15
4
PN B06960AA
19
Select the following fields:
Test Name:(All).
Unit No.:Only for the AU5400. The unit number that the 4046 alarm generated.
Index:The index that the 4064 alarm generated.
Check box:Check one sample kind (Routine, Emergency, or STAT) for Serum, Urine, or
Others measured in the index. STAT is applicable only for the AU2700.
Sample No.: First column: "*"
Second column: "*”
Check only one sample kind at a time so it is easier to identify the approximate first Sample No.
TIP
required to search. Perform this check one sample kind at a time for all the sample kinds measured
in the index.
20
If Inner or Outer displays in the Cuvette field, select Cuvette/Mix No. (F7).
4-13
Page 38
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.16
•Select All from the "Cuvette" drop-down list.
•Do not change the "*" at Mix No. for R1, Sample, and R2.
•Select
21
Select Search Data (F5).
Figure 4.17
OK.
4-14
22
Use the arrow buttons and look at the Measured Time to find the first test measured after the
time obtained from step 10. Note the sample number, test name, and sample ID.
23
Select the forward arrow button to go to the next test. Note the sample number, test name, and
sample ID.
24
Repeat step 23 up to the last result. In this example, the last result is 6400/6400.
25
Select Exit (F2).
PN B06960AA
Page 39
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
26
Select Routine > Reaction Monitor > QC/Cal and repeat steps 18 through 25.
27
All results from samples identified in this procedure are invalid and must be re-analyzed.
4.1.3 Searching for Samples if the SMS test is a Routine Sample for Condition A
1
Select Routine >Reaction Monitor > Normal.
Figure 4.18
4
PN B06960AA
2
Select the SMS test that generated the 4064 alarm from the "Test Name" drop-down list.
3
For the AU5400 only, select the unit number that generated the 4064 alarm from the "Unit No."
drop-down list.
4
Select the last Index from the "Index" drop-down list.
5
Select Cuvette/Mix No. (F7). The Select Cuvette/Mix No. window appears.
4-15
Page 40
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.19
6
Select Inner or Outer that generated the 4064 alarm from the "Cuvette" drop-down list, then
OK.
select
7
Select (check the box) for Routine for the user-defined SMS sample type in use.
8
Select Search Data (F5). The Reaction Monitor window appears.
Figure 4.20
1. Flag Display field
2. Backward arrow
3. Forward arrow
4-16
9
Select the forward arrow button to go to the last result (SMS test with the "*" flag). In this
example, go to result 8/8.
10
Select the backward arrow button to go to the previous SMS test that did not have a "*" flag.
The "*" flags displays after the Result.
PN B06960AA
Page 41
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
11
Note the date and time from the Measured Time field.
12
Select Exit (F2).
13
Select Routine>Data Report > Data List > Patient.
Figure 4.21
Monitoring Results
4
PN B06960AA
14
Select the index containing the 4064 alarm that was generated from the "Index" drop-down list.
Note the sample numbers of each sample kind (Routine, Emergency, or STAT) for Serum, Urine,
or Others measured in the index. STAT is applicable only for the AU2700.
15
Select Routine > Data Report > Data List > QC/Cal.
4-17
Page 42
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.22
16
Select the index containing the 4064 alarm that was generated from the "Index" drop-down list.
Note the sample numbers of all QC, reagent blank, and calibration samples analyzed in the
index.
17
Select Exit (F2).
18
Select Routine > Reaction Monitor.
Figure 4.23
4-18
PN B06960AA
Page 43
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
19
Select the following fields:
Test Name:(All).
Unit No.:Only for the AU5400. The unit number that the 4046 alarm generated.
Index:The index that the 4064 alarm generated.
Check box:Check one sample kind (Routine, Emergency, or STAT) for Serum, Urine, or
Others measured in the index. STAT is applicable only for the AU2700.
Sample No.: First column: "*" or approximate sample number.
Second column: "*" or approximate sample number.
TIP
Check only one sample kind at a time so it is easier to identify the approximate first Sample No.
required to search. Perform this check one sample kind at a time for all the sample kinds measured
in the index.
TIP If there is a large number of samples in the sample kind checked to search, the search takes a long
time. Enter an approximate starting (first column) and ending (second column) Sample No. to search
to reduce the search time. For example, the maximum number of photometric tests per 60 minutes
is 1600. If the average number of tests per sample is 8, the approximate number of samples
processed per 60 minutes is 200 samples. Searching the last 220 samples (200 plus 10%) should be
sufficient to find the results. If the search does not contain samples measured for 60 minutes before
the 4064 alarm was generated, it is necessary to increase the range of samples to search and
perform the search again.
4
20
If Inner or Outer displays in the Cuvette field, select Cuvette/Mix No. (F7).
Figure 4.24
•Select All from the "Cuvette" drop-down list.
•Do not change the "*" at Mix No. for R1, Sample, and R2.
•Select
21
Select Search Data (F5).
OK.
PN B06960AA
4-19
Page 44
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.25
22
Use the arrow buttons and look at the Measured Time to find the first Sample No. measured
after the time obtained in step 10.
23
Note the Sample No.
24
Repeat steps 18 through 22 for all the sample kinds (Routine, Emergency, or STAT) for Serum,
Urine, Others, QC, reagent blank, and calibration.
It is necessary to select the Cal/QC tab to search for reagent blank, calibration, and QC results.
TIP
25
Select Exit (F2) to go back to the Reaction Monitor screen.
26
Select Exit (F2).
27
Select Routine > Data Report > Data List.
4-20
PN B06960AA
Page 45
Figure 4.26
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
4
28
Select the check box for all the sample kinds (Routine, Emergency, or STAT) for Serum, Urine,
and Others measured in the index. STAT is applicable only for the AU2700.
29
Enter the first Sample No. (first column) identified in the previous steps for each sample kind.
Keep "*" as the last Sample No. (second column).
30
Select Print (F3).
The results measured after the previous SMS test before the SMS test that generated the "*" flag
and 4064 REACTION PROFILE CHECK alarm print. These results may be affected by the cuvette
overflow and are invalid. Analyze these samples again.
31
Select Routine > Data Report > QC/Cal tab and repeat steps 27 through 29.
PN B06960AA
4-21
Page 46
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
4.1.4 Searching for Samples if the SMS test is a Routine Sample for Condition B
1
Select Routine>Reaction Monitor > Normal.
Figure 4.27
2
Select the SMS test that generated the 4064 alarm from the "Test Name" drop-down list.
3
For the AU5400 only, select the unit number that generated the 4064 alarm from the "Unit No."
drop-down list.
4
Select the last Index from the "Index" drop-down list.
5
Select Cuvette/Mix No. (F7). The Select Cuvette/Mix No. window appears.
Figure 4.28
4-22
PN B06960AA
Page 47
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
6
Select Inner or Outer that generated the 4064 alarm from the "Cuvette" drop-down list, then
select
OK.
7
Select (check the box) for Routine for the user-defined SMS sample type in use.
8
Select Search Data (F5). The Reaction Monitor window appears.
Figure 4.29
4
PN B06960AA
1. Flag Display field
2. Backward arrow
3. Forward arrow
9
Select the forward arrow button to go to the last result (SMS test with the "*" flag). In this
example, go to result 8/8.
10
Select the backward arrow button to go to the previous SMS test that did not have a "*" flag.
The "*" flags displays after the Result.
11
Note the Measured Time.
12
Select Exit (F2).
13
Select Routine > Data Report > Data List > Patient.
4-23
Page 48
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.30
14
Select the index containing the 4064 alarm that was generated from the "Index" drop-down list.
Note the sample numbers of each sample kind (Routine, Emergency, or STAT) for Serum, Urine,
or Others measured in the index. STAT is applicable only for the AU2700.
15
Select Routine > Data Report > Data List > QC/Cal.
Figure 4.31
4-24
PN B06960AA
Page 49
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
16
Select the index containing the 4064 alarm that was generated from the "Index" drop-down list.
Note the sample numbers of all QC, reagent blank, and calibration samples analyzed in the
index.
17
Select Exit (F2).
18
Select Routine > Reaction Monitor.
Figure 4.32
4
PN B06960AA
19
Select the following fields:
Test Name:(All).
Unit No.:Only for the AU5400. The unit number that the 4046 alarm generated.
Index:The index that the 4064 alarm generated.
Check box:Check one sample kind (Routine, Emergency, or STAT) for Serum, Urine, or
Others measured in the index. STAT is applicable only for the AU2700.
Sample No.: First column: "*"
Second column: "*”
Check only one sample kind at a time so it is easier to identify the approximate first Sample No.
TIP
required to search. Perform this check one sample kind at a time for all the sample kinds measured
in the index.
20
If Inner or Outer displays in the Cuvette field, select Cuvette/Mix No. (F7).
4-25
Page 50
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
Figure 4.33
•Select All from the "Cuvette" drop-down list.
•Do not change the "*" at Mix No. for R1, Sample, and R2.
•Select
21
Select Search Data (F5).
Figure 4.34
OK.
4-26
22
Use the arrow buttons and look at the Measured Time to find the first test measured after the
time obtained from step 10. Note the sample number, test name, and sample ID.
23
Select the forward arrow button to go to the next test. Note the sample number, test name, and
sample ID.
24
Repeat step 22 up to the last result. In this example, the last result is 6400/6400.
25
Select Exit (F2).
PN B06960AA
Page 51
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
26
Select Routine > Reaction Monitor > QC/Cal and repeat steps 18 through 24.
27
All results from samples identified in this procedure are invalid and must be re-analyzed.
4
PN B06960AA
4-27
Page 52
Monitoring Results
4.1 Identifying and Re-analyzing Samples after a Cuvette Overflow
4-28
PN B06960AA
Page 53
CHAPTER 5
As Needed Maintenance
This chapter contains a new as needed maintenance procedure to perform after a cuvette overflow.
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
5.1.1 Remove the Cuvette Wheel and Cuvettes
5.1.2 Clean the Cuvettes
5.1.3 Replace the Cuvette Wheel into the Incubation Bath
PN B06960AA
5-1
Page 54
As Needed Maintenance
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an
Overflow
Erroneous data is generated if the cuvettes become wet on the outside (cuvette overflow). The
reaction of the sample and reagents takes place in a carefully controlled dry incubation bath. Water
on the outside of the cuvette affects the light as it passes through the cuvette. Test results are
impacted due to this change in absorbance. It is important to recognize when an overflow has
occurred, determine the cause for the overflow, and follow the appropriate procedures to recover
from an overflow. For additional information, refer to the Troubleshooting Chapter, 4.1 Recovering
from a Cuvette Overflow in this addendum.
Materials needed for this procedure:
•2% Wash Solution
•Cotton tipped applicators
•Sonicator
•Clean, dry lint-free cloth
•#2 Philips head screwdriver
•Large plastic containers to hold cuvette wedges
•Plastic containers to hold cuvettes in the sonicator
5.1.1 Remove the Cuvette Wheel and Cuvettes
1
Verify that the analyzer is in Warm up or Standby.
TIP
The AU2700 is a single unit system. The AU5400 has a two or three unit configuration.
2
Open the rear cover of the affected unit.
CAUTION
Ensure the cover is open and secured before leaning over the wash nozzle unit.
3
Loosen the knob on the wash nozzle unit. Support the wash nozzle unit when removing it from
the base so that it does not come into contact with the cuvettes, and hang it on the hook.
4
Open the upper cover of the unit.
5-2
5
Remove any mix bars positioned over the cuvette wheel. Pull the mix bars up and out. Be careful
not to scratch the mix bars.
PN B06960AA
Page 55
As Needed Maintenance
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
6
Remove the small cuvette wheel cover, and then remove the large cuvette wheel cover. It may
be necessary to angle the cover slightly to clear the mix unit.
When removing the cuvette wheel cover, be careful not to scratch or bump the sample probes,
TIP
reagent probes, or mix bars.
7
Locate the 2 large, flat screws on the metal plate in the center of the cuvette wheel.
8
Remove these two large flat screws. These screws are used as “handles” to remove the cuvette
wheel from the incubation bath.
Figure 5.1
5
PN B06960AA
1
1. Flat screws
5-3
Page 56
As Needed Maintenance
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
Figure 5.2
1. Flat screw
2. Flat screw
9
Attach the flat screws to the frame of the cuvette wheel. Place one of the screws into the
opening on the wheel in front of cuvette number 28. Place the other screw into the opening on
the wheel in front of cuvette number 111. Firmly tighten each screw into the cuvette wheel.
10
Using the “handles” (large, flat screws), carefully pull the cuvette wheel off of the two metal
positioning pins. It may be necessary to angle the frame slightly to clear the mix unit.
Loosen the screws on all 15 wedges, and then lift out each individual wedge before removing the
TIP
cuvette wheel to make the cuvette wheel lighter and easier to remove.
11
Place the cuvette wheel on a protected work surface. Loosen the screws on all 15 wedges, and
then lift out each individual wedge.
5-4
PN B06960AA
Page 57
As Needed Maintenance
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
CAUTION
Do not touch the photometric face (clear side) of the cuvette when removing the
cuvette wedge.
Figure 5.3
5
12
Place the cuvette wedge upside down on the protected surface to remove the cuvettes. Use your
finger or the reverse end of a cotton-tipped applicator stick to gently push the cuvettes out of
the wedge. Remove all 330 cuvettes.
CAUTION
The cuvettes may be scratched when they are removed from the wedge and
subsequently will not pass the photocal.
PN B06960AA
5-5
Page 58
As Needed Maintenance
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
Figure 5.4
CAUTION
Do not touch the photometric (clear) side of the cuvette. Handle by the frosted
side only.
Figure 5.5
5.1.2 Clean the Cuvettes
1
Submerge all cuvettes in a plastic container filled with 2% Wash Solution. If a sonicator is not
available, soak the cuvettes in a 5% Wash Solution overnight.
5-6
PN B06960AA
Page 59
As Needed Maintenance
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
2
Thoroughly rinse the cuvettes in deionized water, or sonicate them in deionized water for 10
minutes to remove any residual Wash Solution.
3
Allow the cuvettes to dry completely. Use one of the following methods:
•Allow cuvettes to air dry.
•Use an oven with the heat set below 50 °C (122 °F). A hair dryer works well.
•Use a lint free cloth.
CAUTION
Do not use Wash Solution or any detergent to clean the cuvette wedges or wheel.
Use only DI water. The metallic plating on the wedge or wheel may be removed if
Wash Solution or any other detergent is used.
5
4
Rinse the frame of the cuvette wheel and cuvette wedges with DI water only.
5
Dry the frame and wedges thoroughly with a lint free cloth.
6
Dry the incubation bath with a lint free cloth.
7
Replace all 330 cuvettes (with the open end facing up) back into the cuvette wedge. Ensure the
cuvette is pushed down completely into the wedge.
CAUTION
Do not touch the photometric side of the cuvette. Handle by the frosted side only.
Always replace any cuvettes removed from the cuvette wedge. If a cuvette is
missing, reagent and/or sample will cause an overflow in the cuvette wheel.
8
Replace all wedges into the cuvette wheel frame. Match the number on the wedge to the
number listed on the frame. Match the hole on the wedge with the guide pin for proper
placement.
PN B06960AA
9
Screw the wedges back into place.
TIP Replace the cuvette wheel into the incubation bath before replacing the wedges into the cuvette wheel
frame to make the cuvette wheel lighter and easier to replace.
5-7
Page 60
As Needed Maintenance
1
2
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
5.1.3 Replace the Cuvette Wheel into the Incubation Bath
1
Use the “handles” (large, flat screws) to place the cuvette wheel back into the incubation bath.
It may be necessary to angle the wheel slightly to clear the mix unit.
Figure 5.6
1. Positioning Pin
2. Alignment indicator
2
Match the white alignment indicator on the cuvette wheel with the white indicator on the
analyzer unit. When the alignment is correct, the cuvette wheel can be set onto the two metal
positioning pins.
3
Loosen the two “handles” (large, flat screws) and set them in the openings on the metal plate
in the center of the cuvette wheel.
4
Replace the large cuvette wheel cover by matching the hole in the cover with the pin on the
analyzer unit and then replace the small cover matching the green dot on the cover with the
green dot on the analyzer unit.
5-8
PN B06960AA
Page 61
As Needed Maintenance
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow
CAUTION
Be careful not to scratch the mix bars when inserting them into mix unit.
5
Remove the wash nozzle unit from the hook and carefully place it on the wash nozzle unit base
so it does not come into contact with the cuvettes. Align the holes on the wash nozzle station
over the two positioning screws on the base and tighten the knob. Verify that all tubing and
tube mounting joints are connected to the nozzles and joint manifolds. Refer to the “Replace
the Wash Nozzle Joint Tubes” in the As Needed Maintenance section in the AU2700 or AU5400 User's Guide for correct tubing placement in case any lines are disconnected.
6
Replace all mix bars that were removed. Place the correct mix bars in the designated location
(R2 = green dots). Insert the mix bars into the top of the mix unit. While inserting the mix bar,
rotate it slightly to engage the flat portion of the mix bar with the gear of the unit.
CAUTION
5
Do not force the mix bar into the unit. Damage to the gears may occur.
7
Select Maintenance > ANL Maintenance.
TIP
The AU2700 does not have an option to select a unit.
8
Select the appropriate analyzer unit number and A/Prime Washing-line.
9
Press the DIAG switch on the selected unit. The wash nozzle unit moves up and down and is
primed with DI water.
10
Repeat the previous step several times until air is removed.
11
Check for leaks around the wash nozzle unit and then close the rear and upper covers.
12
Perform a Photocal on the affected unit. Refer to the Photocal procedure in the Maintenance
chapter of the AU2700 or AU5400 User's Guide.
PN B06960AA
5-9
Page 62
As Needed Maintenance
5.1 Clean the Cuvettes, Cuvette Wedges, and the Cuvette Wheel after an Overflow