Beckman Coulter, and the Beckman Coulter symbol are
trademarks of Beckman Coulter, Inc. and are registered in
the USPTO.
All other trademarks are the property of their respective
owners.
Find us on the World Wide Web at:
www.beckmancoulter.com
Original Instructions
Revision History
These instructions apply to Power Processor systems with the latest software version listed in this
document, as well as previous software versions. When a subsequent software version changes the
information in this document, a new issue will be released.
A97244AF, June 2019
This Instructions For Use manual is for Power Processor used with PrepLink software Version 5.0.
Changes
•Safety Notice:
•CHAPTER 4,Troubleshooting
A97244AE, March 2018
This Instructions For Use manual is for Power Processor used with PrepLink software Version 5.0.
Changes
•Safety Notice:
:
Removed Moving Parts Label from the Hardware Labels section. Moved it to “Legacy Hardware
Labels” table.
Replaced the Moving Parts symbol to Crushing of hands symbol in the Symbols Glossary table
Added the California Proposition 65 symbol to the Symbols Glossary table
Removed the Country of Origin symbol from the Symbols Glossary table
Corrected the description of the Problem for Sensor AS63, in Table 4.6 3060-tube (high-speed)
Stockyard Error Codes.
Updated Figures 4.13, 4.14 and 4.16
:
Added a symbols glossary to address changes to global labeling requirements and identify the
symbols that relate to product identification, classification, cautions, and warnings.
A97244AF
A97244AD, 11/2012
Software Version 5.0
Changes:
This release of the Power Processor Stockyard IFU includes support for the 5440-tube (5K) highspeed Stockyard.
CHAPTER 1, Module Description
5440-tube (high-speed) Stockyard
Added information for the newly supported 5K Stockyard.
CHAPTER 2, Module Procedures
Control Panel on the 3060-tube and 5440-tube (high-speed) Stockyard Refrigeration Unit
5440-tube (high-speed) Stockyard added to the control panel refrigeration unit section.
iii
Revision History
CHAPTER 3, Error Recovery Procedures
Power Processor Module Function Codes
5440-tube (high-speed) Stockyard function codes added to this chapter.
CHAPTER 4, Troubleshooting
5440-tube (high-speed) Stockyard Error Codes
5440-tube (high-speed) Stockyard error codes added to this chapter.
The following sensor diagrams were added to this chapter:
•5440-tube (high-speed) Stockyard Main Body
•5440-tube (high-speed) Stockyard First Shelf
•5440-tube (high-speed) Stockyard Second Shelf
•5440-tube (high-speed) Stockyard Third Shelf
•5440-tube (high-speed) Stockyard Fourth Shelf
•5440-tube (high-speed) Stockyard Conveyor
•5440-tube (high-speed) Stockyard H-Lane
A97244AC, 12/2011
Software Version 5.0
A97244AB, 06/2011
Software Version 5.0
A97244AA (limited release), 01/2011
Software Version 5.0
iv
A97244AF
Read all product manuals and consult with Beckman Coulter-trained personnel before attempting
to operate instrument. Do not attempt to perform any procedure before carefully reading all
instructions. Always follow product labeling and manufacturer’s recommendations. If in doubt as
to how to proceed in any situation, contact your Beckman Coulter Representative.
Alerts for Warning, Caution, Important, and Note
WARNING
Warning indicates a potentially hazardous situation which, if not avoided, could
cause death or serious injury. Warning can indicate the possibility of erroneous
data that could cause an incorrect diagnosis.
CAUTION
Caution indicates a potentially hazardous situation which, if not avoided, can
cause minor or moderate injury. Caution can also alert against unsafe practices,
or indicate the possibility of erroneous data that could cause an incorrect
diagnosis.
Safety Notice
IMPORTANT
NOTE Note indicates notable information to follow.
TIP Tip indicates information to consider.
Important indicates important information to follow.
A97244AF
v
Safety Notice
General Warnings and Cautions
General Warnings and Cautions
WARNING
Do not use the equipment in a manner not specified by Beckman Coulter, Inc., as
the protection provided by the equipment can be impaired.
WARNING
Each sample tube processed by the system must have a unique sample ID and
readable bar code label. The system sends sample tubes with duplicate sample
IDs to the Error Lane. Damaged or unreadable bar code labels cause errors.
If you load multiple sample tubes with the same sample ID on the automation
system and on a connected analyzer at the same time, the system can send
duplicate results to the LIS. Contact your Beckman Coulter Representative for
suggestions to implement unique sample ID labeling.
CAUTION
To reduce risk of personal injury, operate the system only with all covers in place.
CAUTION
Do not load or view the Instructions for Use PDF files onto any computer
connected to the automation system. Failure to follow this caution can reduce
computer processing speed and system performance.
CAUTION
Use only the approved Power Processor parts and supplies as listed in the Power
Processor General System Operation IFU, Appendix B. Use only the approved
sample tubes as noted in the Power Processor General System Operation IFU,
Operational Overview.
vi
A97244AF
Electromagnetic Wave and Noise Precautions
This In Vitro diagnostic (IVD) equipment complies with the emission and immunity requirement
described in IEC 61326-2-6.
CAUTION
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment, it could cause radio interference, in which case, you might need to
take measures to mitigate the interference.
It is advised that before operation of the device, the electromagnetic environment
be evaluated. Do not use this device near sources of strong electromagnetic
radiation (for example, unshielded intentional RF sources), as they could interfere
with the correct operation.
Hardware Labels
Safety Notice
Hardware Labels
Biohazard Label
This label indicates a caution to operate only with all covers in position to decrease risk of personal
injury or biohazard.
Pneumatic Label
This label indicates a caution that the Inlet module works under a pressure of 0.7 mpa (100 PSI).
A97244AF
vii
Safety Notice
Compliance and Certification Markings
Compliance and Certification Markings
These labels and materials declaration table (the Table of Hazardous Substance's Name and
Concentration) meet People's Republic of China Electronic Industry Standard SJ/ T11364-2006
"Marking for Control of Pollution Caused by Electronic Information Products" requirements.
Recycling Label
This label is required in accordance with the Waste Electrical and Electronic Equipment (WEEE)
Directive of the European Union. The presence of this label indicates that:
1. the device was put on the European Market after August 13, 2005 and
2. the device is not to be disposed of via the municipal waste collection system of any member
state of the European Union
Customers must understand and follow all laws regarding the correct decontamination and safe
disposal of electrical equipment. For Beckman Coulter products bearing this label, contact your
dealer or local Beckman Coulter office for details on the take-back program that facilitates the
correct collection, treatment, recovery, recycling and safe disposal of these products.
For the Japan Market:
This system is considered an industrial waste, subject to special controls for infectious waste. Prior
to disposal of the system, refer to the "Waste Disposal and Public Cleaning Law" for compliance
procedures.
cNRTLus Certification Mark
This symbol indicates recognition by a Nationally Recognized Testing Laboratory (NRTL) that the
instrument has met the relevant product safety standards for the United States and Canada.
OSHA, CEC
viii
A97244AF
CE Marking
The CE marking indicates that a product has been assessed before being placed on the market, and
has been found to meet the applicable directives relating to the European Union safety, health, and
environmental protection requirements.
RoHS Caution Symbol
This symbol indicates that this electronic information product contains certain toxic or hazardous
elements, and can be used safely during its environmental protection use period. The number in the
middle of the logo indicates the environmental protection use period (in years) for the product. The
outer circle indicates that the product can be recycled. The logo also signifies that the product
should be recycled immediately after its environmental protection use period has expired. The date
on the label indicates the date of manufacture.
Safety Notice
Compliance and Certification Markings
These labels and materials declaration table (the Table of Hazardous Substance's Name and
Concentration) meet People's Republic of China Electronic Industry Standard SJ/T11364-2006
Marking for Control of Pollution Caused by Electronic Information Products requirements.
RCM Symbol
This symbol indicates compliance with the Australian Communications Media Authority (ACMA)
requirements (safety and EMC) for Australia and New Zealand.
A97244AF
ix
Safety Notice
Symbols Glossary
Symbols Glossary
Symbols Glossary
SymbolSymbol Title, Symbol Meaning, and Symbol Reference
Title of Symbol: Caution
Meaning of Symbol: Indicates the need for the user to consult the
instructions for use for important cautionary information such as
warnings and precautions that cannot, for a variety of reasons, the
presented on the medical device itself.
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 15223-1. Medical devices - Symbols to be used with medical device
labels, labelling and information to be supplied - Part 1: General
Requirements. #5.4.4
Title of Symbol: In vitro diagnostic medical device
Meaning of Symbol: Indicates a medical device that is intended to be
used as an in vitro diagnostic medical device.
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 15223-1: Medical devices. Symbols to be used with medical device
labels, labelling and information to be supplied. General requirements,
clause 5.5.1
Title of Symbol: Dangerous voltage
Meaning of Symbol: To indicate hazards arising from dangerous
voltages.
IEC 60417: Graphical symbols for use on equipment - Overview and
application, #5036
Supplemental Product-Specific Manufacturer Information
This symbol can also indicate an area of the system to not access under
any circumstances, due to possibility of high voltages and the risk of
electrical shock.
Title of Symbol: Warning; Biological hazard
Meaning of Symbol: To warn of a biological hazard.
Standard Number, Title of Standard, and Symbol Reference Number:
IEC 60878. Graphical Symbols for electrical equipment in medical
practices. #7010-W009
Supplemental Product-Specific Manufacturer Information
This label indicates a caution to operate only with all covers in position
to decrease risk of personal injury or biohazard.
This label indicates the use of biohazardous materials in the area. Use
caution when working with possible infectious samples.
Wear Personal Protective Equipment (PPE) such as gloves, eye shields,
and lab coats. Handle and dispose of biohazardous materials according
to your laboratory procedures.
x
A97244AF
Symbols Glossary (Continued)
SymbolSymbol Title, Symbol Meaning, and Symbol Reference
Safety Notice
Symbols Glossary
Title of Symbol: Warning; Crushing of hands
Meaning of Symbol: To warn of a closing motion of mechanical parts of
equipment
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 7010. Graphical Symbols for electrical equipment in medical
practices. #W024
Supplemental Product-Specific Manufacturer Information
Use caution to avoid injury to hands when close to equipment with
moving mechanical parts.
Title of Symbol: Consult instructions for use
Meaning of Symbol: Indicates the need for the user to consult the
instructions for use.
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 15223-1. Medical devices - Symbols to be used with medical device
labels, labelling and information to be supplied - Part 1: General
Requirements. #5.4.3
Title of Symbol: Date of Manufacture
Meaning of Symbol: To indicate the date when the medical device was
manufactured.
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 15223-1. Medical devices - Symbols to be used with medical device
labels, labelling and information to be supplied - Part 1: General
Requirements. #5.1.3
Title of Symbol: Authorised representative in the European
Community
Meaning of Symbol: Indicates the authorized representative in the
European community.
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 15223-1. Medical devices - Symbols to be used with medical device
labels, labelling and information to be supplied - Part 1: General
Requirements. #5.1.2
Title of Symbol: Manufacturer
Meaning of Symbol: Indicates the medical device manufacturer as
defined in EU Directives 90/385/ EEC, 93/42/EEC and 98/79/EC.
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 15223-1. Medical devices - Symbols to be used with medical device
labels, labelling and information to be supplied - Part 1: General
Requirements. #5.1.1
Supplemental Product-Specific Manufacturer Information
This symbol identifies who the legal manufacturer of the product is.
A97244AF
xi
Safety Notice
Symbols Glossary
Symbols Glossary (Continued)
SymbolSymbol Title, Symbol Meaning, and Symbol Reference
Title of Symbol: Catalogue Number
Meaning of Symbol: Indicates the manufacturer's catalogue number so
that the medical device can be identified.
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 15223-1. Medical devices - Symbols to be used with medical device
labels, labelling and information to be supplied - Part 1: General
Requirements. #5.1.4
Title of Symbol: Serial number
Meaning of Symbol: Indicates the manufacturer's serial number so that
a specific medical device can be identified.
Standard Number, Title of Standard, and Symbol Reference Number:
ISO 15223-1. Medical devices - Symbols to be used with medical device
labels, labelling and information to be supplied - Part 1: General
Requirements. #5.1.7
Title of Symbol: Stop
Meaning of Symbol: To identify the control or the indicator to stop the
active function.
Standard Number, Title of Standard, and Symbol Reference Number:
IEC 60417: Graphical symbols for use on equipment - Overview and
application, #5110A
Supplemental Product-Specific Manufacturer Information
This symbol indicates a stop button.
Title of Symbol: Cooling Unit
Meaning of Symbol: Denotes the cooling unit.
xii
Title of Symbol: Driver Box
Meaning of Symbol: Denotes the driver box.
Title of Symbol: Made in Japan
Meaning of Symbol: Indicates the country where the device hardware
was manufactured.
Title of Symbol: RxOnly Symbol
Meaning of Symbol: Caution: U.S. Federal Law restricts this device to
sale by or on the order of a licensed practitioner.
Title of Symbol: California Proposition 65
Meaning of Symbol: This product may contain chemicals known to the
State of California to cause cancer and birth defects or other
reproductive harm. For more information go to
https://www.P65Warnings.ca.gov.
A97244AF
Historical Power Processor systems may contain the following additional labels.
456161-B
TO REDUCE RISK OF PERSONAL INJURY,
OPERATE ONLY WITH ALL COVERS IN PLACE.
CAUTION
A011459L.EPS
CAUTION
SHARP OBJECTS
A16558-AA
A016351L.EPS
Legacy Hardware Labels
SymbolDescription
Safety Notice
Symbols Glossary
Please Hold This Portion
This symbol, located on the drawer in the Dynamic Inlet and
Outlet modules and the top cover of the centrifuge, indicates the
most stable position to hold the drawer or cover.
Stop Button
This symbol indicates a stop button that you can use to halt any
hazardous condition.
Power On
This symbol indicates the location of the Power On button in the
Centrifuge module.
RoHS Environmental
This symbol indicates that the product does not contain any
toxic or hazardous substances or elements. The "e" stands for
electrical, electronic, and environmental electronic information
products. This logo indicates that this electronic information
product does not contain any toxic or hazardous substances or
elements, and is green and is environmental. The outer circle
indicates that the product can be recycled. The logo also
signifies that the product can be recycled after being discarded,
and should not be casually discarded.
C-Tick Mark
The C-Tick mark is intended for use on products that comply
with the applicable Electromagnetic Compatibility (EMC)
standards in the Australian or New Zealand market.
A97244AF
Caution
To reduce the risk of electrical shock, disconnect the power
supply cord before servicing.
Caution, Biohazard Label
This caution symbol indicates a caution to operate only with all
covers in position to decrease risk of personal injury or
biohazard.
Sharp Object Label
A label reading “CAUTION SHARP OBJECTS” is found on the
Decapper device in the Decapper/Serum Level Detection (SLD)
module.
xiii
Safety Notice
CAUTION
PARTS MOVE
AUTOMATICALLY
A015081L.EPS
Symbols Glossary
Legacy Hardware Labels (Continued)
SymbolDescription
Caution, Moving Parts
This caution symbol warns the operator of moving parts that can
pinch or crush. This label is found in several locations.
Moving Parts Label
This label indicates moving parts that can pinch or crush. This
label is found in several locations.
Caution parts move automatically
While the system is in operation, do not touch or go close to any
moving parts. Close protective guards and covers during
operation. Failure to close covers correctly can cause injury or
incorrect results.
xiv
A97244AF
Revision History,iii
Safety Notice,v
General Information,xvii
CHAPTER 1:Module Description,1-1
Stockyard,1-2
Large Capacity Stockyards,1-4
Safety Shields,1-11
CHAPTER 2:Module Procedures,2-1
Splash Guard Safety Cover,2-1
Contents
How Large Capacity Stockyards are Managed,2-3
Outlet Racks,2-4
Rack Assignment with Large Capacity Stockyards,2-4
Control Panel on the 3060-tube (standard-speed) Stockyard Refrigeration
Unit,2-8
Control Panel on the 3060-tube and 5440-tube (high-speed) Stockyard
Refrigeration Unit,2-10
Retrieving a Sample from a Stockyard,2-13
How to Remove a Rack from a 1020-tube Stockyard,2-15
How to Remove a Rack from a 3060-tube (standard-speed) Stockyard,2-17
How to Remove a Rack from a 3060-tube (high-speed) Stockyard,2-21
How to Remove a Rack from a 5440-tube (high-speed) Stockyard,2-24
How to Disable the 3K or 5K Stockyard Shelves if an
Unrecoverable Error Occurs,2-28
How to Disable a Shelf in the 3K or 5K Stockyard,2-33
How to Enable the 3K or 5K Stockyard,2-35
CHAPTER 3:Error Recovery Procedures,3-1
Error Recovery General Information,3-2
System Error Recovery Procedures,3-4
xv
Contents
Stockyard Module Error Recovery Procedures,3-22
CHAPTER 4:Troubleshooting,4-1
Error Code Tables,4-1
Sensor Diagrams,4-44
Related Documents
xvi
Intended Use
The basic Power Processor is an automated sample handling system which processes sample tubes
from the pre-centrifugation, pre-sorting step to presentation of centrifuged and decapped samples
into Generic or Personality Racks for specific instruments. The Power Processor can be configured
with optional software and hardware to allow processing of sample tubes on Generic Connection
instruments.
The Power Processor performs all pre-analytical sample tube preparation, and then sorts the
sample tubes directly to Generic Connection Modules where the samples are pipetted by the
Generic Connection instrument for testing. After the samples are pipetted, the tubes can route to
other instruments for additional testing or to Outlet Racks.
General Information
Scope of this Manual
This Instructions For Use manual is for Power Processor Stockyard Modules used with PrepLink
software Version 5.0. This manual contains information and instructions that will assist you in
performing Power Processor Stockyard Module operations and troubleshooting functions.
NOTE
This document is part of the Power Processor Instructions for Use Document Set and only covers the
Power Processor Stockyard Modules. For information and instructions for other Power Processor
modules, refer to the following Power Processor Instructions for Use Document Set.
Power Processor Instructions for Use Document Set
•General System Operation IFU, PN B01683
•Inlets, Outlets, Hematology, and Bar Code Verification Modules IFU, PN B01519
•Centrifuge Module IFU, PN A97119
•Decapper and Recapper Modules IFU, PN A97252
•Aliquot Module IFU, PN A97103
•LX and DxC Connection Modules IFU, PN A97111
•AU Connection Modules IFU, PN B01540
Be sure to follow all safety cautions and warnings noted in this document.
A97244AF
•Generic/IDC and DxI Connection Modules IFU, PN A97260
•Stockyard Modules IFU, PN A97244
xvii
General Information
Scope of this Manual
xviii
A97244AF
CHAPTER 1
Module Description
This chapter provides a brief description of the Power Processor Stockyard Modules.
Stockyards provide long-time online storage of sample tubes. A Power Processor with instrument
connections can be configured with four types of stockyards. The four stockyard types are:
•200-tube Stockyard, ambient temperature (maximum of 1 can be configured)
•1020-tube (1K) Stockyard, ambient temperature (maximum of 1 can be configured)
•3060-tube (3K) Stockyard, refrigerated (maximum of 3 can be configured)
NOTE
•5440-tube (5K) Stockyard, refrigerated (maximum of 3 can be configured)
NOTE
can have a maximum of three 3060-tube Stockyards or three 5440-tube Stockyards.
The 3060-tube (3K) Stockyard is available in a standard-speed or high-speed configuration.
Configurations with multiple stockyards must all be of the same type (capacity), i.e., a configuration
A97244AF
1-1
Module Description
Stockyard
Stockyard
The stockyard is automatically configured in Outlet 2A when the PrepLink is configured to manage
the Power Processor with connected instruments. For information regarding the large capacity
stockyards, refer to Large Capacity Stockyards.
Sample tubes are routed to the stockyard for one or more of the following reasons:
•Sample analysis complete
•Reagent is not calibrated, not loaded, or test number/test volume is zero
•Reagent calibration has timed out
•Reagent level sense error
•Reagent level sense pending
•Reagent load requested
•Calibration requested
•Parameter required
•Test bypassed
•Within lot calibration pending
•Connected instrument is in MANUAL mode
•Connected instrument is Loading Paused
•There were no remaining tests for a primary tube after assigning tests to aliquots
•Remapped samples
Sample tubes in the stockyard have a status of Pending until PrepLink receives a completion
message for that sample ID from the instrument. At that time, the status changes to Outlet 2A, if all
tests are completed.
If the instrument is unable to perform any of the pending tests after a sample tube waits for the
timeout period in the stockyard, the sample tube is routed to the Pending Rack for offline handling
by the operator.
IMPORTANT
overridden if the requested test is specified as an Urgent Chemistry. When tests are configured as Urgent
Chemistries, the default stockyard timeout is 15 minutes. In all cases, the lowest timeout value for a
sample tube takes precedence.
The default stockyard timeout period for an LX/DxC sample tube is 30 minutes but can be
NOTE The timeout values listed are the defaults. All timeout values can be changed by a Beckman Coulter
Service Representative.
If all pending tests become available while a sample tube waits in the stockyard, the sample tube
does not have to wait until the timeout period has elapsed. Within minutes the sample tube is
retrieved from the stockyard and routed to the instrument. After analysis is complete, the sample
tube is routed to the stockyard for storage.
1-2
A97244AF
Module Description
Stockyard
One or more, but not all of the pending tests can be performed on a sample tube that has waited in
the stockyard. The system retrieves the sample tube from the stockyard after the completion
message is received and the instrument is ready to run the pending test. The sample tube routes to
the Pending Rack if the pending tests are not completed within the stockyard timeout period.
1
IMPORTANT
If add-on, the reflex sample or rerun sample programming arrives after the analysis of a sample
tube is complete and the tube has been sorted to the stockyard, the system considers the requested
tests as new programming. Sample tubes are retrieved from the stockyard and routed to an
instrument or to a rack in Outlet 1A or Outlet 1B. If the feature Rack ID is enabled, sample tubes in
the stockyard with a status of Outlet 2A, 2B, or 2C change to a status of Storage when the assigned
rack is removed by using the Rack Load procedure.
The stockyard timeout restarts each time the sample tube is sorted to the stockyard.
Rerun Requests or Pending Add-on Test Results
To determine the sample ID of the tube(s) with pending tests sorted to stockyard, search for
"pending results" at the DL2000 or LIS. To manually retrieve these sample tubes, use the Search by
Sample ID function on the PrepLink Locations screen. Rack and position number are shown if the
tube is still in the stockyard or if Rack Numbering is used for mapping tube location after racks are
removed from the system.
Occasionally, when sample programming includes tests with long instrument incubation times,
completed sample tubes can be routed to the Pending Rack.
IMPORTANT
the Pending Rack, and add-on sample programming is subsequently received at PrepLink, the add-on
tests cannot be performed automatically. If additional testing is needed, the operator must remove the
sample tube from the Pending Rack and process it offline. If the add-on tests are applicable to the primary
sample tube remaining in the stockyard, the add-on tests can be processed.
If the operator performs the sample retrieve process such that sample tubes are retrieved to
A97244AF
1-3
Module Description
A011826P.EPS
1
Large Capacity Stockyards
Large Capacity Stockyards
On connected systems, the outlets can be configured with a maximum of one 1020-tube stockyard,
or up to three 3060-tube or 5440-tube Stockyards for longer online sample storage. In any
configuration, the large capacity stockyards must all be of the same type.
•A maximum of one 1020-tube Stockyard, ambient temperature.
•Up to three 3060-tube (standard-speed) Stockyards, refrigerated.
•Up to three 3060-tube (high-speed) Stockyards, refrigerated.
•Up to three 5440-tube Stockyards, refrigerated.
340-tube Rack
IMPORTANT There are two types of plastic 340-tube racks. A molded rack that uses plastic standoffs
between the three rack layers, and a machined rack that uses metal standoffs. Never intermix these two
types of racks in a stockyard.
A rack in a large capacity stockyard contains positions (10 × 34) for 340 tubes as shown in Figure 1.1.
Each shelf in a 1020-tube or 3060-tube large capacity stockyard can hold three racks.
Each shelf in a 5440-tube Stockyard can hold four racks.
•The 1020-tube Stockyard has one shelf (3 racks × 340 tubes/rack = 1020 tubes).
•The 3060-tube (standard or high-speed) Stockyard has three shelves (3 shelves × 1020 tubes/
shelf = 3060 tubes).
•The 5440-tube Stockyard has four shelves with 4 racks per shelf (4 shelves × 1360 tubes/shelf =
5440 tubes).
The Transfer Load Arm loads sample tubes from left to right. The first sample tube is put in the left
rear corner (1) as shown in Figure 1.1.
Figure 1.1 340-tube Rack for a Large Capacity Stockyard
1. First Sample Tube Loaded Here
1-4
A97244AF
1020-tube Stockyard
A011833P.EPS
The 1020-tube Stockyard is shown in Figure 1.2. This stockyard has one shelf that contains three
340-tube racks. The sample tubes are loaded from the rear by a transfer-arm. Another transfer-arm
retrieves sample tubes. Removing racks is discussed in the procedures: How to Remove a Rack from
a 1020-tube Stockyard and Control Panel on the 3060-tube (standard-speed) Stockyard
Refrigeration Unit.
The 1020-tube Stockyard supports one tube size at a time, either 13 x 75 mm tubes or 13 x100 mm
tubes, depending on how the automation system is configured.
WARNING
While the transfer-arms are active, DO NOT put your hand(s) in the opening at the
rear of the 1020-tube Stockyard or in the track area. Serious injuries can occur.
Figure 1.2 1020-tube Stockyard
Module Description
Large Capacity Stockyards
1
3060-tube Stockyard
The 3060-tube Stockyard is available in two configurations: standard-speed and high-speed. The
information in this section applies to both the standard-speed and high-speed configurations.
NOTE
A 3060-tube Stockyard is shown in Figure 1.3. The 3060-tube Stockyard has three shelf units at the
bottom and a refrigeration unit at the top. The shelf units are numbered one to three, from the floor
up to the refrigeration unit.
Each shelf unit has a door and contains three 340-tube racks. The sample tubes are loaded from the
rear by a transfer-arm. The retrieve-arm retrieves sample tubes (refer to Figure 1.4). Inside the
3060-tube Stockyard, there is one Arm on each shelf.
A97244AF
The high-speed 3060-tube Stockyard is only supported with PrepLink V5.0 and higher.
1-5
Module Description
A014644P.EPS
1
Large Capacity Stockyards
The 3060-tube Stockyard supports one tube size at a time, either 13 x 75 mm tubes or 13 x100 mm
tubes, depending on how the automation system is configured.
The 3060-tube Stockyard has a door sensor that places the stockyard in PAUSE mode preventing the
Transfer or Retrieval Arms from moving when a door is opened. When the door to a shelf unit is
opened, the sensor LED displays RED. All three shelf units of the stockyard are in PAUSE mode when
any one door is opened. If the Transfer or Retrieval Arm is moving inside the stockyard when a door
is opened, the arm completes the present action, then pauses. The arm does not move to the Home
position. To resume routine operation after closing the door, press the
keypad. To prevent injury and blockage of the sample tubes, obey the warnings that follow.
WARNING
If the sensor LED is not RED when a Shelf Unit door is opened, DO NOT put your
hand(s) in the 3060-tube Stockyard before pressing the PAUSE/RUN button on the
keypad. Serious injuries can occur.
WARNING
PAUSE/RUN button on the
While the transfer-arms are active, DO NOT put your hand(s) in the opening at the
rear of the 3060-tube Stockyard or in the track area. Serious injuries can occur.
Figure 1.3 3060-tube (standard-speed) Stockyard
1. Control Panel (with cover opened)
1-6
A97244AF
Figure 1.4 Transfer-arms at the Rear of a 3060-tube Stockyard
A014646P.EPS
2
1
Module Description
Large Capacity Stockyards
1
1. Transfer-arm
2. Retrieve-arm
3060-tube (high-speed) Stockyard Features
Compared to the 3060-tube (standard-speed) Stockyard, the high-speed stockyard provides the
following additional features:
•Increased throughput, up to 600 tubes per hour; twice the speed of the standard-speed 3060tube Stockyard.
NOTE
Module throughput is dependant on factors, such as Bar Code Label quality, tube flow patterns,
test category, analyzer's performance, or other site specific/system specific factors.
CAUTION
Do not bypass the Recapper Module while tubes sort to the stockyard. Because of
the speed that sample tubes move into and out of the high-speed stockyard, the
tubes must have caps to prevent spilling.
•New rack latches to be sure the racks remain securely in place when sample tubes are being
•New temperature monitoring system with audible and visual alarms.
loaded to or retrieved from the stockyard. (Refer to Figure 1.5.)
A97244AF
NOTE
The minimum temperature and maximum temperature range, and the timer value or timer wait
time before the alarm is generated, are set by your Beckman Coulter Service Representative.
The 5440-tube (high-speed) Stockyard has four shelf units and a refrigeration unit at the top (refer
to Figure 1.6). Each shelf unit has two, side-by-side, doors and contain four 340-tube racks (1,360
tubes per shelf). The 5440-tube Stockyard supports the 13 x100 mm tubes. To configure the
stockyard for a different tube size, contact your Beckman Coulter Representative.
1-8
A97244AF
Figure 1.6 5440-tube (high-speed) Stockyard
Module Description
Large Capacity Stockyards
1
Total tube capacity of the stockyard is 5,440 tubes, with throughput of up to 600 tubes per hour.
NOTE Module throughput is dependant on factors, such as Bar Code Label quality, tube flow patterns, test
category, analyzer's performance, or other site specific/system specific factors.
CAUTION
Do not bypass the Recapper Module while tubes sort to the stockyard. Because of
the speed that sample tubes move into and out of the high-speed stockyard, the
tubes must have caps to prevent spilling.
Sample tubes are loaded at the rear of the stockyard by a transfer-arm. The retrieve-arm retrieves
sample tubes to send them to a connected instrument or an outlet rack. The transfer and retrievear ms travel up or down ver tical rails to position themselves in front of the tube access doors on each
shelf (see Figure 1.7). Inside of the 5440-tube Stockyard, there is one arm on each shelf.
A97244AF
1-9
Module Description
Large Capacity Stockyards
Figure 1.7 Transfer-arms at the rear of a 5440-tube Stockyard.
1. Transfer-arm
2. Retrieve-arm
The 5440-tube Stockyard has a door sensor that places the stockyard in PAUSE mode, preventing
the Transfer or Retrieval Arms from moving, when a door opens (see Figure 1.8). When the door to
a shelf unit opens, the sensor LED displays RED.
Figure 1.8 5440-tube Stockyard Door Sensor
1-10
1. Door Sensor
All four shelf units of the stockyard go to PAUSE mode when any door opens. If the Transfer or
Retrieval Arm is moving inside the stockyard when a door opens, the arm completes the present
action, then pauses. The arm does not move to the Home position. To resume routine operation
after closing the door, press the
PAUSE/RUN button on the keypad.
A97244AF
To prevent injury and blockage of the sample tubes, obey the warnings that follow.
WARNING
If the sensor LED is not RED when a Shelf Unit door opens, DO NOT put your
hand(s) in the 5440-tube Stockyard before pressing the PAUSE/RUN button on the
keypad. Serious injuries can occur.
WARNING
While the transfer-arms are active, DO NOT put your hand(s) in the opening at the
rear of the 5440-tube Stockyard or in the track area. Serious injuries can occur.
Safety Shields
To provide a safe operating environment for the operator, the Power Processor modules have safety
shields. These safety shields provide a protective barrier between the operator and sample
preparation activities.
Module Description
Safety Shields
1
WARNING
Removal of safety shields during operation can expose the operator to physical
and biological hazards.
A97244AF
1-11
Module Description
Safety Shields
1-12
A97244AF
This chapter contains operational procedures for Power Processor Stockyard modules.
A016605P.EPS
Splash Guard Safety Cover
The 1020-tube, 3060-tube, and 5440-tube Stockyards have splash guard covers attached to the back
side of the stockyard with a Phillips head screw.
Some error recovery procedures require the operator to remove the splash guard cover to enable
access to the module generating the error. Refer to the procedure How to Remove the Splash Guard
Cover.
WARNING
Always operate stockyards with all covers, shields, and doors in place and secured
to avoid injury.
CHAPTER 2
Module Procedures
Figure 2.1 Stockyard Splash Guard Cover (on a 3K Stockyard)
A97244AF
2-1
Module Procedures
Splash Guard Safety Cover
How to Remove the Splash Guard Cover
CAUTION
To avoid injury, be sure the module is in PAUSE mode before you remove the
splash guard cover.
CAUTION
Even when a large capacity stockyard is in PAUSE mode, the air system is still
active and applying a constant air pressure to the shuttle cups and the gripper
arm assembly. This can cause unexpected movement of the shuttle cups and the
gripper arm assembly when resolving a jammed object error, creating a possible
moving part or mechanical pinch hazard. Use caution when resolving jammed
object errors at a large capacity stockyard.
1
Press the PAUSE/RUN button on the Stockyard Module keypad to put the module in PAUSE
mode.
2
Remove the splash guard cover screw with a Phillips screwdriver.
After you remove the splash guard cover, you can address the problem that generated the
error.
3
Replace the splash guard cover and replace the screw.
IMPORTANT
Be sure to fasten the cover securely.
WARNING
Always operate stockyards with all covers, shield and doors in place to avoid
injury.
4
Press the PAUSE/RUN button on the Stockyard Module keypad to resume routine sample
processing.
2-2
A97244AF
How Large Capacity Stockyards are Managed
3060-tube Stockyards
For systems configured with one 3K Stockyard, tubes load into the stockyard by filling the bottom
shelf first. When all three racks on the bottom shelf fill, a flashing beacon activates. When the third
rack on shelf number 3 fills, an audible and visual alarm activates
Module Procedures
How Large Capacity Stockyards are Managed
2
IMPORTANT
(Refer to Figure 2.4):
The operator must remove the oldest rack by performing a Rack Load so the system can resume filling
the stockyard. The oldest rack displays on the keypad with the first number for the shelf number
and the second number for the rack position. For example, if the oldest rack is on shelf number 2,
position 2, the keypad displays “2-02.”
For systems configured with up to three stockyards, tubes load to each stockyard sequentially. A
tube loads into one stockyard, the next tube loads into the next stockyard, the next tube loads into
the last stockyard, if present, and the process repeats for each additional tube.
The system fills the bottom shelf first in each stockyard and moves upward. When a shelf of three
racks fills, the flashing beacon activates. When all three shelves in one stockyard fill, an audible and
visual alarm activates. The operator must perform a Rack Load to remove racks and resume filling
the stockyard.
Shelves in the 3K Stockyard are numbered from bottom to top and from left to right, as follows
3-013-023-03
2-012-022-03
1-011-021-03
5440-tube Stockyards
For systems configured with one 5K Stockyard, tubes load into the stockyard by filling the bottom
shelf first. When all four racks on the bottom shelf fill, a flashing beacon activates. When the fourth
rack on the shelf fills, an audible and visual alarm activates
IMPORTANT
(Refer to Figure 2.6):
The operator must remove the oldest rack by performing a Rack Load so the system can resume filling
the stockyard. The oldest rack displays on the keypad with the first number for the shelf number
A97244AF
Shelves in the 5K Stockyard are numbered from bottom to top and from left to right, as follows
4-014-024-034-04
3-013-023-033-04
2-012-022-032-04
1-011-021-031-04
2-3
Module Procedures
Outlet Racks
and the second number for the rack position. For example, if the oldest rack is on shelf number 2,
position 2, the keypad displays “2-02.”
For systems configured with up to three 5K stockyards, tubes load to each stockyard sequentially.
A tube loads into one stockyard, the next tube loads into the next stockyard, the next tube loads into
the last stockyard, if present, and the process repeats for each additional tube.
The system fills the bottom shelf first in each stockyard and moves upward. When a shelf of four
racks fills, the flashing beacon activates. When all four shelves in one stockyard fill, an audible and
visual alarm activates. The operator must perform a Rack Load to remove racks and resume filling
the stockyard.
Outlet Racks
The Rack Types for the large capacity stockyards are:
•"K1" for the 1020-tube Stockyard
•"K3" for the 3060-tube Stockyard
•"K5" for the 5440-tube Stockyard
Table 2.1 gives a summary of the V2.6 or greater rack designations. For other rack designations,
refer to Table 2.1 Rack Designations in the General System Operation IFU.
IMPORTANT
between the three rack layers, and a machined rack that uses metal standoffs. Never intermix these two
types of racks in a stockyard.
Table 2.1 Rack Designations for V2.6 or Greater
Rack
Type
K1Outlet #2, 2A, 2B
K3Outlet #2, 2A, 2B, 2C
K5Outlet #2, 2A, 2B, 2C
There are two types of plastic 340-tube racks. A molded rack that uses plastic standoffs
Rack
Description
Single Set
Triple Set
Quadruple Set
340*331Stockyard1
340*3, 340*3, 340*391Stockyard3
340*4, 340*4, 340*4,
340*4
Tube
Allocation
No.
of
Racks
161Stockyard5
Rack Assignment with Large Capacity Stockyards
No.
of Groups
(Subsections)
Rack
Section
Name(s)
2-4
At installation, the Service Representative configures the system for Outlet 2, 2A, 2B and 2C. The
correct is shown when the Racks tab is selected. The Rack screen shows a G6, K1, K3, or K5 stockyard.
A97244AF
Rack Assignment–1020-tube Stockyard
E014635S.EPS
Figure 2.2 shows a PrepLink screen for an Outlet #2 rack assignment with a 1020-tube Stockyard.
Figure 2.2 Outlet #2 Rack Assignment with a 1020-tube Stockyard
Module Procedures
Rack Assignment with Large Capacity Stockyards
2
Rack Assignment–3060-tube Stockyard
Figure 2.3 shows a PrepLink screen for an Outlet 2A rack assignment with a 3060-tube Stockyard.
The number of shelves and tubes is given under the "Outlet Overview."
Figure 2.3 Outlet 2A Rack Assignment with a 3060-tube Stockyard
A97244AF
NOTE In the event of an unrecoverable error, shelves can be disabled, to permit partial use of the stockyard.
Figure 2.4 shows the rack positions in a 3060-tube Stockyard.
2-5
Module Procedures
A007427P.EPS
3-013-023-03
2-012-022-03
1-011-021-03
Rack Assignment with Large Capacity Stockyards
Figure 2.4 3060-tube Stockyard Rack Positions
Rack Assignment–5440-tube Stockyard
Figure 2.5 shows a PrepLink screen for an Outlet 2A rack assignment with a 5440-tube Stockyard.
The number of shelves and tubes is given under the "Outlet Overview."
Figure 2.5 Outlet 2A (for PrepLink V5.0 and greater) Rack Assignment with a 5440-tube Stockyard
NOTE In the event of an unrecoverable error, shelves can be disabled to permit partial use of the stockyard.
2-6
A97244AF
Rack Assignment with Large Capacity Stockyards
Figure 2.6 shows the rack positions in a 5440-tube Stockyard.
Figure 2.6 5440-tube Stockyard Rack Positions
4-014-024-034-04
3-013-023-033-04
2-012-022-032-04
Module Procedures
2
1-011-021-031-04
A97244AF
2-7
Module Procedures
MITSUBISHI
TEMP
C
°
ON
OFF
3
EMERGENCY
STOP
MAINTENANCE
DEFROSTER
RESET
TEMP
SHIFT
MANUAL
DEFROSTER
ADJUST
TEMP
ADJUST
TIME
SET TEMP
MODE
CHANGE
CHECK
SET
1
2
A014816L.EPS
4
Control Panel on the 3060-tube (standard-speed) Stockyard Refrigeration Unit
Control Panel on the 3060-tube (standard-speed) Stockyard
Refrigeration Unit
IMPORTANT If the Control Panel is not available on your stockyard, this section does not apply.
To set the temperature on your stockyard, call your Beckman Coulter Service Representative.
Figure 2.7 shows the Control Panel for the refrigeration unit of the 3060-tube Stockyard. This panel
is at the top right side of the refrigeration unit as shown in Figure 1.3. Use this panel to:
•Set the temperature.
•Read the temperature near the refrigeration unit as described in step 6.
NOTE
This temperature does not represent the entire refrigeration unit temperature.
IMPORTANT Some buttons on the Control Panel are NOT included in the procedures in this section. Factory
or service personnel use these buttons. If you have questions, contact your Beckman Coulter Service
Representative.
Figure 2.7 Buttons and Displays on the Control Panel of the Refrigeration Unit
2-8
1. ON/OFF Button
2. Set Value Display
3. Operation Mode Display
4. SET TEMP Button
A97244AF
Module Procedures
Control Panel on the 3060-tube (standard-speed) Stockyard Refrigeration Unit
How to Set the Set Point Temperature on the 3060-tube (standard-speed)
Stockyard Refrigeration Unit
IMPORTANT If the Control Panel is not available on your stockyard, this section does not apply.
To set the temperature on your stockyard, call your Beckman Coulter Service Representative.
1
If necessary, turn the power on. Press the ON/OFF button (1) for more than one second until the
red light comes on (refer to Figure 2.7).
2
If necessary, open the cover of the Control Panel.
3
Press the SET TEMP button (4).
4
A "0" appears on the Operation Mode Display (3).
2
5
Press the ADJUST TEMP buttons to set the temperature on the Set Value Display (2).
6
Press the SET TEMP button again. The Set Point of the refrigeration unit appears. After a few
seconds, the temperature near the refrigeration unit appears.
7
Close the cover of the Control Panel.
After the Set Point temperature is established, let the 3060-tube Stockyard to equilibrate for 24
hours before monitoring and recording the temperature.
Use a calibrated thermometer to check the temperature. The temperature should be checked and
recorded daily.
A97244AF
2-9
Module Procedures
Control Panel on the 3060-tube and 5440-tube (high-speed) Stockyard Refrigeration Unit
Control Panel on the 3060-tube and 5440-tube (high-speed) Stockyard
Refrigeration Unit
The Temperature Control Panel for the 3060-tube (high-speed) or 5440-tube (high-speed) Stockyard
Refrigeration Unit, is located on the right side panel of the stockyard, as shown in Figure 1.3 in the
previous chapter. A close-up view of the control panel is shown in Figure 2.8.
The function of this control panel is to display the temperature inside of the refrigeration unit and
warn lab personnel if the temperature is out-of-range.
IMPORTANT
The minimum temperature and maximum temperature range, and timer value or timer wait time before
the alarm is generated, are set by your Beckman Coulter Service Representative.
Buttons on the Temperature Control Panel are NOT for customer use, and are not functional.
Temperature Control Panel Function
The Temperature Control Panel constantly reads the temperature inside the stockyard and
generates an alarm (audible and visual) if the temperature inside the stockyard is above or below
the acceptable range for more than the preset timer value (default is 30 minutes).
NOTE
A normal in-range operating temperature is indicated by a solid green light beacon and a green
control panel display (refer to Figure 2.8).
Figure 2.8 Temperature Control Panel Display In-Range
The normal operating temperature range is between 2 degrees and 8 degrees Centrigrade.
2-10
A97244AF
Module Procedures
Control Panel on the 3060-tube and 5440-tube (high-speed) Stockyard Refrigeration Unit
An out-of-range temperature condition is indicated by a red control panel display and one or more
of the following conditions: Keypad Display error code, Audible Alarm, Amber Light Beacon (steady
on, or blinking). Refer to Figure 2.9.
Figure 2.9 Temperature Control Panel Display Out-of-Range
2
For more information on temperature conditions and temperature error indicators, refer to the
below table.
Table 2.2 Temperature Condition and Error Indicators
Temperature
Condition
Temperature inrange (>2 to <9
degrees C)
Out-of-temperature,
but less than Timer
Value
Out-of-temperature,
and Timer Value met
Out-of-temperature,
Timer Value met,
Keypad Alarm cleared
Keypad
Display
BlankOFFOFFONGREENDisplay is without decimal point and in
BlankOFFOFFONREDTemperature greater than, or equal to 9
Error
Code
9_01
BlankOFFONOFFREDTimer restarts after Alarm has been cleared.
Audible
Alarm
ONON
Amber
Light
Beacon
(Blinking)
Green
Beacon
Light
OFFREDTemperature greater than, or equal to 9
Display
Color
Notes
degrees Centigrade. Temperature is in range
and the display turns Green.
degrees Centigrade, or Less than or equal to
2 degrees Centigrade.
degrees Centigrade, or less than or equal to
2 degrees Centigrade for the period of time
set in the Timer Value.
If temperature is still not in-range and Timer
Value is met again, the Alarm is generated
again.
A97244AF
2-11
Module Procedures
Control Panel on the 3060-tube and 5440-tube (high-speed) Stockyard Refrigeration Unit
Table 2.2 Temperature Condition and Error Indicators (Continued)
Temperature
Condition
Out-of-temperature,
Timer Value met,
other system error
Out-of-temperature,
Timer Value met,
other system error,
Keypad Alarm cleared
How to Recover from an Out-of-Range Temperature Warning
When the Temperature Control Panel senses that the internal temperature of the refrigeration unit
is outside of the preset (normal) temperature operating range, an alarm sounds, the amber warning
light beacon comes ON and begins to blink, and the keypad displays error code 9_01.
NOTE
information.
To recover from an error, do the following:
Keypad
Display
Error
Code
x_xx
BlankOFFSlow
Various conditions cause different audible and visual indicators. Refer to Table 2.2 for more
Audible
Alarm
ONONOFFREDWhile the yellow light is blinking after the
Amber
Light
Beacon
Blink
(1 per
second)
Green
Beacon
Light
OFFREDThe non-temperature system error is
Display
Color
Notes
9_01 temperature error has been
acknowledged, another type of system error
can occur, such as a stuck sample carrier on
the stockyard track.
corrected and yellow light returns to blinking
if the temperature is still out of range.
1
Press the ALARM button on the stockyard module keypad to silence the alarm. The alarm stops
and the amber warning light beacon changes from a blinking light to steady ON.
2
Confirm that all doors of the stockyard are closed, monitor the situation and, if possible, correct
the cause of the temperature problem.
3
To continue operation, press PAUSE/RUN. The amber warning light beacon changes from steady
ON to a slow blinking light.
4
Continue to observe the warning light beacon. When the temperature returns to normal, the
light goes out.
IMPORTANT If the temperature does not return to normal after 30 minutes, confirm all doors are closed. If
there are no error conditions to address and the temperature remains out of normal range, contact your
Beckman Coulter Service Representative.
2-12
A97244AF
Retrieving a Sample from a Stockyard
System-Initiated Retrieval
When the stockyard timeout expires, or it is time to test a sample that is stored in the stockyard,
the Retrieve Transfer-arm removes the sample tube from the stockyard. Then the track routes the
sample tube to a connected instrument or an outlet rack.
There are two conditions where PrepLink retrieves a sample tube and then routes the tube to a
connected instrument or an outlet rack.
•A tube is retrieved for rerun, add-on or reflex testing. The tube is retrieved from the stockyard
and routed to a connected instrument or an Outlet Rack.
•A tube is routed to the Pending Rack sort location for an operator to process it.
Module Procedures
Retrieving a Sample from a Stockyard
2
IMPORTANT
from, the stockyard.
IMPORTANT Before requesting to add on tests, confirm that the sample tube is in the stockyard. If the
sample tube has been retrieved to the Pending Rack, the sample add-on can not be processed as
expected.
DO NOT remove a rack from the stockyard when a sample tube is being loaded to, or retrieved
Operator-Initiated Retrieval to the Pending Rack
1
Select the PrepLink Locations tab.
2
Select the By option button. Select Sample ID.
3
In the For text field, type in one of the following:
•A sample ID
•Sample IDs with a comma between
•A range of Sample IDs with a hyphen between (this is the example given in Figure 2.10).
A97244AF
4
Select the Retrieve button.
5
The Retrieve Confirmation dialog box appears, refer to Figure 2.11. This screen has two lists:
•The list at the top of this screen gives the Sample IDs that will be retrieved. These samples
have Outlet #2, 2A, 2B, 2C or Pending status.
•The list at the bottom of this screen gives the Sample IDs that are not available to be
retrieved.
2-13
Module Procedures
E014633S.EPS
Retrieving a Sample from a Stockyard
6
To retrieve the samples in the list at the top of the screen, select OK.
OR
To return to the Locations screen, select Cancel.
If you typed in the primary Sample ID and you selected the Retrieve button, the primary and
associated aliquot tubes are sent to the Pending Rack sort location. If you typed in a specific aliquot
tube Sample ID and you selected the Retrieve button, the aliquot tube is sent to the Pending Rack
sort location.
Example:
•Primary sample ID 132436 and aliquot sample ID 132436.1 are in the stockyard.
•Type sample ID 132436 and select
route to the Pending Rack sort location.
•Type sample ID 132436.1 and select
Rack sort location, while the primary tube remains in the stockyard.
Retrieve. The primary tube 132436 and aliquot tube 132436.1
Retrieve. The aliquot tube 132436.1 routes to the Pending
Figure 2.10 Retrieve Sample by Range of Sample IDs
2-14
A97244AF
Figure 2.11 Retrieve Confirmation Screen
E014634S.EPS
Module Procedures
How to Remove a Rack from a 1020-tube Stockyard
2
How to Remove a Rack from a 1020-tube Stockyard
Safety
Be careful when removing a rack from a 1020-tube Stockyard.
WARNING
To prevent injury, be sure a large capacity stockyard is in PAUSE mode before you
open a door or put your hands inside.
CAUTION
Use two hands to remove a rack from a large capacity stockyard. Support the rack
and keep it horizontal to prevent tubes from falling.
When replacing the rack, push the rack to the back of the rack position.
IMPORTANT
to, or retrieved from, the large capacity stockyard.
DO NOT remove a rack from a large capacity stockyard when a sample tube is being loaded
A97244AF
2-15
Module Procedures
How to Remove a Rack from a 1020-tube Stockyard
System-Initiated Rack Load
When a 1020-tube Stockyard is full the flashing beacon, the audible alarm and the RACK LOAD light
on the 1020-tube Stockyard keypad come on. The transfer-arm completes the current operation
and then moves to the Home position.
1
Press the ALARM button on the 1020-tube Stockyard keypad to silence the alarm and stop the
flashing beacon.
2
Read the 1020-tube Stockyard keypad display to determine which rack positions to unload.
Rack positions are numbered 01 through 03.
3
Remove the racks indicated by the keypad display. Hold each rack by one of the hand-holes in
the front of the rack. When pulling out the rack, hold the rack with both hands and keep it
horizontal.
4
Load empty racks into the vacant positions. The PAUSE/RUN light flashes after a rack is loaded.
If the PAUSE/RUN light does not flash and the rack to be removed is still displayed on the LED,
verify that the Line Control Computer is on the Main Screen. If the Line Control Computer is on
the Database Query/Sample Search screen, return to the Line Control Computer Main Screen.
5
Press the PAUSE/RUN button on the keypad to resume routine operation.
IMPORTANT
PAUSE/RUN button is pressed to resume normal operation, the sample tubes do not sort to the stockyard
as expected.
The sample tubes stop at the stockyard with no error indication on the LED display. After the rack is fully
inserted, the sample tubes resume processing when the PAUSE/RUN button is pressed.
If a rack in the 1020-tube Stockyard is not fully inserted after a rack Removal, when the
Operator-Initiated Rack Load
Racks in the 1020-tube Stockyard can be unloaded before they fill using the Rack Load procedure.
1
Press the RACK LOAD button on the 1020-tube Stockyard keypad. The alarm and beacon
activate.
2-16
2
Press the ALARM button to silence the alarm and stop the flashing beacon.
3
Remove the racks you want to unload. Hold each rack by one of the hand-holes in the front.
When pulling out the rack, hold the rack with both hands and keep it horizontal.
A97244AF
Module Procedures
How to Remove a Rack from a 3060-tube (standard-speed) Stockyard
4
Load racks in the empty positions. The PAUSE/RUN light flashes after a rack is loaded.
If the PAUSE/RUN light does not flash and the rack to be removed is still displayed on the LED,
verify that the Line Control Computer is on the Main Screen. If the Line Control Computer is on
the Database Query/Sample Search screen, return to the Line Control Computer Main Screen.
5
Press the PAUSE/RUN button to resume routine operation.
2
IMPORTANT
PAUSE/RUN button is pressed to resume normal operation, the sample tubes will not sort to the
stockyard as expected.
The sample tubes stop at the stockyard with no error indication on the LED display. After the rack is fully
inserted, the sample tubes resume processing when the PAUSE/RUN button is pressed.
If a rack in the 1020-tube Stockyard is not fully inserted after a rack Removal, when the
How to Remove a Rack from a 3060-tube (standard-speed) Stockyard
Safety
Be careful when removing a rack from a 3060-tube Stockyard. Support the rack and keep it
horizontal to prevent tubes from falling. Figure 2.12 shows a rack not fully removed from the shelf
unit of a 3060-tube Stockyard.
WARNING
To prevent injury, be sure the large capacity stockyard is in PAUSE mode before
you open a door or put your hands inside.
A97244AF
CAUTION
Use two hands to remove the rack from a large capacity stockyard. Support the
rack and keep it horizontal to prevent tubes from falling.
When replacing the rack, push the rack to the back of the rack position.
IMPORTANT
to, or retrieved from, the large capacity stockyard.
DO NOT remove a rack from the large capacity stockyard when a sample tube is being loaded
2-17
Module Procedures
A014645P.EPS
How to Remove a Rack from a 3060-tube (standard-speed) Stockyard
Figure 2.12 A Rack Not Fully Removed from a Shelf Unit of a 3060-tube Stockyard
System-Initiated Rack Load
When a 3060-tube Stockyard is full, the flashing beacon, the audible alarm and the RACK LOAD light
on the 3060-tube Stockyard keypad come on. The transfer-arm completes the current operation
and then moves to the home position.
Select the image below to start a video on how to perform a System-Initiated Rack Load.
CAUTION
Before viewing this video, confirm that the IFU PDF files are not installed on, or
viewed from, a computer connected to the Power Processor system.
2-18
A97244AF
How to Remove a Rack from a 3060-tube (standard-speed) Stockyard
NOTE This video is provided as an aid to help perform the following procedure.
Module Procedures
2
1
Press the ALARM button on the 3060-tube Stockyard keypad to silence the alarm and stop the
flashing beacon.
2
Read the 3060-tube Stockyard keypad display to determine which rack positions to unload. The
keypad displays the racks that need be removed. Refer to Figure 2.4.
IMPORTANT
3
Open the door of the related shelf unit of the 3060-tube Stockyard.
4
Remove the racks indicated by the keypad display. Hold each rack by one of the hand-holes in
the front of the rack. When pulling out the rack, hold the rack with both hands and keep it
horizontal.
Shelves are numbered from bottom to top and from left to right as follows:
Shelf Three3-013-023-03
Shelf Two2-012-022-03
Shelf One1-011-021-03
A97244AF
5
Load empty racks into the vacant positions. The PAUSE/RUN light flashes after a rack is loaded.
Close the door.
If the PAUSE/RUN light does not flash and the rack to be removed is still displayed on the LED,
verify that the Line Control Computer is on the Main Screen. If the Line Control Computer is on
the Database Query/Sample Search screen, return to the Line Control Computer Main Screen.
2-19
Module Procedures
How to Remove a Rack from a 3060-tube (standard-speed) Stockyard
6
Press the PAUSE/RUN button on the keypad to resume routine operation.
IMPORTANT
PAUSE/RUN button is pressed to resume normal operation, the sample tubes will not sort to the
stockyard as expected.
The sample tubes stop at the stockyard with no error indication on the LED display. After the rack is fully
inserted, the sample tubes resume processing when the PAUSE/RUN button is pressed.
If a rack in the 3060-tube Stockyard is not fully inserted after a rack Removal, when the
Operator-Initiated Rack Load
Racks in the 3060-tube Stockyard can be unloaded before they fill using the Rack Load procedure.
Refer to Figure 2.4.
Select the image below to start a video on how to perform an Operator-Initiated Rack Load.
CAUTION
Before viewing this video, confirm that the IFU PDF files are not installed on, or
viewed from, a computer connected to the Power Processor system.
2-20
NOTE This video is provided as an aid to help perform the following procedure.
1
Press the RACK LOAD button on the 3060-tube Stockyard keypad. The alarm and beacon
activate.
2
Press the ALARM button to silence the alarm and stop the flashing beacon.
A97244AF
Module Procedures
How to Remove a Rack from a 3060-tube (high-speed) Stockyard
3
Open the door of the related shelf unit of the 3060-tube Stockyard.
4
Remove the racks you want to unload. Hold each rack by one of the hand-holes in the front of
the rack. When pulling out the rack, hold the rack with both hands and keep it horizontal.
5
Load racks into the empty positions. The PAUSE/RUN light flashes after a rack is loaded. Close
the door.
If the PAUSE/RUN light does not flash and the rack to be removed is still displayed on the LED,
verify that the Line Control Computer is on the Main Screen. If the Line Control Computer is on
the Database Query/Sample Search screen, return to the Line Control Computer Main Screen.
6
Press the PAUSE/RUN button to resume routine operation.
2
IMPORTANT
PAUSE/RUN button is pressed to resume normal operation, the sample tubes will not sort to the
stockyard as expected.
The sample tubes stop at the stockyard with no error indication on the LED display. After the rack is fully
inserted, the sample tubes resume processing when the PAUSE/RUN button is pressed.
If a rack in the 3060-tube Stockyard is not fully inserted after a Rack Removal, when the
How to Remove a Rack from a 3060-tube (high-speed) Stockyard
Safety
Be careful when removing a rack from a 3060-tube Stockyard. Support the rack and keep it
horizontal to prevent tubes from falling. Figure 2.12 shows a rack not fully removed from the shelf
unit of a 3060-tube Stockyard.
WARNING
To prevent injury, be sure the large capacity stockyard is in PAUSE mode before
you open a door or put your hands inside.
A97244AF
CAUTION
Use two hands to remove the rack from a large capacity stockyard. Support the
rack and keep it horizontal to prevent tubes from falling.
When replacing the rack, push the rack to the back of the rack position.
IMPORTANT
to, or retrieved from, the large capacity stockyard.
DO NOT remove a rack from the large capacity stockyard when a sample tube is being loaded
2-21
Module Procedures
1
How to Remove a Rack from a 3060-tube (high-speed) Stockyard
Removing a Rack from the 3060-tube (high-speed) Stockyard
The 3060-tube (high-speed) Stockyard has a rack latching mechanism to confirm that each rack on
a shelf is securely held in place. Refer to Figure 2.13.
In front of each rack is a spring-loaded latch that is depressed by the rack as it slides over it while
being inserted into the shelf unit. When the rack is fully inserted, the latch rises-up and snaps into
position to secure the rack.
To remove the rack, press down on the latch to move it below the front edge of the rack, and pull
the rack toward you. Refer to Figure 2.14.
2-22
A97244AF
How to Remove a Rack from a 3060-tube (high-speed) Stockyard
1
Figure 2.14 Removing a Rack from the 3060-tube (high-speed) Stockyard
Module Procedures
2
1. Rack Latching Mechanism
System-Initiated or Operator-Initiated Rack Load
Performing a System-Initiated or Operator-Initiated rack load on a 3060-tube (high-speed)
Stockyard, is the same as performing it on a 3060-tube (standard-speed) Stockyard.
For instructions, or to view a video, on these procedures, select one of the following:
System-Initiated Rack Load
Operator-Initiated Rack Load
A97244AF
2-23
Module Procedures
How to Remove a Rack from a 5440-tube (high-speed) Stockyard
How to Remove a Rack from a 5440-tube (high-speed) Stockyard
Safety
Be careful when removing a rack from a 5440-tube Stockyard. Support the rack and keep it
horizontal to prevent tubes from falling. Figure 2.12 shows a rack not fully removed from the shelf
unit of a 3060-tube Stockyard.
WARNING
To prevent injury, be sure the large capacity stockyard is in PAUSE mode before
you open a door or put your hands inside.
CAUTION
Use two hands to remove the rack from a large capacity stockyard. Support the
rack and keep it horizontal to prevent tubes from falling.
When replacing the rack, push the rack to the back of the rack position.
IMPORTANT
to, or retrieved from, the large capacity stockyard.
DO NOT remove a rack from the large capacity stockyard when a sample tube is being loaded
Removing a Rack from the 5440-tube (high-speed) Stockyard
The 5440-tube (high-speed) Stockyard has a rack latching mechanism to confirm that each rack on
a shelf is securely held in place. Refer to Figure 2.15.
How to Remove a Rack from a 5440-tube (high-speed) Stockyard
In front of each rack is a spring-loaded latch that is depressed by the rack as it slides over it while
being inserted into the shelf unit. When the rack is fully inserted, the latch rises-up and snaps into
position to secure the rack.
To remove the rack, press down on the latch to move it below the front edge of the rack, and pull
the rack toward you. Refer to Figure 2.16.
Figure 2.16 Removing a Rack from the 5440-tube (high-speed) Stockyard
2
1. Rack Latching Mechanism
System-Initiated Rack Load
When a 5440-tube Stockyard is full, the flashing beacon, the audible alarm, and the RACK LOAD light
on the stockyard keypad come on. The transfer-arm completes the current operation and then
moves to the home position.
1
Press the ALARM button on the 5440-tube Stockyard keypad to silence the alarm and stop the
flashing beacon.
2
Read the 5440-tube Stockyard keypad display to determine which rack positions to unload. The
keypad displays the racks to be removed. Refer to Figure 2.6.
IMPORTANT
Shelves are numbered from bottom to top and from left to right as follows:
Shelf Four4-014-024-034-04
1
A97244AF
Shelf Three3-013-023-033-04
Shelf Two2-012-022-032-04
Shelf One1-011-021-031-04
2-25
Module Procedures
How to Remove a Rack from a 5440-tube (high-speed) Stockyard
3
Open the door of the related shelf unit of the 5440-tube Stockyard.
4
Remove the racks indicated by the keypad display. Hold each rack by one of the hand-holes in
the front of the rack. When pulling out the rack, hold the rack with both hands and keep it
horizontal.
5
Load empty racks into the vacant positions. The PAUSE/RUN light flashes after a rack is loaded.
Close the door.
If the PAUSE/RUN light does not flash and the rack to be removed is still displayed on the LED,
verify that the Line Control Computer is on the Main Screen. If the Line Control Computer is on
the Database Query/Sample Search screen, return to the Line Control Computer Main Screen.
6
Press the PAUSE/RUN button on the keypad to resume routine operation.
IMPORTANT
PAUSE/RUN button is pressed to resume normal operation, the sample tubes will not sort to the
stockyard as expected.
The sample tubes stop at the stockyard with no error indication on the LED display. After the rack is fully
inserted, the sample tubes resume processing when the PAUSE/RUN button is pressed.
If a rack in the 5440-tube Stockyard is not fully inserted after a rack Removal, when the
2-26
A97244AF
Operator-Initiated Rack Load
Racks in the 5440-tube Stockyard can be unloaded before they fill using the Rack Load procedure.
Refer to Figure 2.6.
1
Press the RACK LOAD button on the 5440-tube Stockyard keypad. The alarm and beacon
activate.
2
Press the ALARM button to silence the alarm and stop the flashing beacon.
3
Open the door of the related shelf unit of the 5440-tube Stockyard.
4
Remove the racks you want to unload. Hold each rack by one of the hand-holes in the front of
the rack. When pulling out the rack, hold the rack with both hands and keep it horizontal.
Module Procedures
How to Remove a Rack from a 5440-tube (high-speed) Stockyard
2
5
Load racks into the empty positions. The PAUSE/RUN light flashes after a rack is loaded. Close
the door.
If the PAUSE/RUN light does not flash and the rack to be removed is still displayed on the LED,
verify that the Line Control Computer is on the Main Screen. If the Line Control Computer is on
the Database Query/Sample Search screen, return to the Line Control Computer Main Screen.
6
Press the PAUSE/RUN button to resume routine operation.
IMPORTANT
PAUSE/RUN button is pressed to resume normal operation, the sample tubes will not sort to the
stockyard as expected.
The sample tubes stop at the stockyard with no error indication on the LED display. After the rack is fully
inserted, the sample tubes resume processing when the PAUSE/RUN button is pressed.
If a rack in the 5440-tube Stockyard is not fully inserted after a rack Removal, when the
A97244AF
2-27
Module Procedures
How to Disable the 3K or 5K Stockyard Shelves if an Unrecoverable Error Occurs
How to Disable the 3K or 5K Stockyard Shelves if an
Unrecoverable Error Occurs
In the event of an unrecoverable error, individual shelves can be disabled to permit partial use of
the stockyard.
CAUTION
Do not disable all stockyards or all shelves at the same time. A minimum of one
shelf of one stockyard is needed for continued sample processing on the system.
If all stockyards or all shelves are disabled at the same time, samples destined for
the stockyard will have no place to sort and will circulate around the track until
one shelf is enabled.
CAUTION
Do not bypass all of the stockyard modules at the same time. If all stockyards are
bypassed at the same time, samples destined for the stockyard will have no place
to sort and will circulate around the track until one stockyard is put into AUTO
mode.
Select the image below to start the video for enabling/disabling the stockyard.
CAUTION
Before viewing this video, confirm that the IFU PDF files are not installed on, or
viewed from, a computer connected to the Power Processor system.
2-28
NOTE
This video is provided as an aid to help perform the following procedure.
A97244AF
Module Procedures
How to Disable the 3K or 5K Stockyard Shelves if an Unrecoverable Error Occurs
1
Silence the alarm and follow the error recovery for the error code at the keypad display.
2
NOTE
2
At the Line Controller main screen, select System Setup from the menu bar.
3
Select the System Setup option. The System Setup dialog box appears.
4
In the System Setup dialog box, select Unit Enable/Disable.
The Unit Enable/Disable dialog box appears. The status (enabled or disabled) of the stockyard
shelf and other connections is shown.
5
Select the status of the correct shelf to Disable, then select OK. The Verify Stockyard Inactive
window displays (refer to Figure 2.17).
Figure 2.17 Verify Stockyard Inactive
If the problem is a single shelf and you cannot recover from this problem, then continue with this
procedure and disable the shelf.
A97244AF
6
Do the following:
a. Verify the stockyard is inactive.
b. Check that the
c.Select
7
If all shelves in all stockyards are disabled, the following pop-up window will display (PP
version 5.0 or later) indicating that no stockyard is available to store samples (refer to
Figure 2.18).
a. Select
b. If this pop-up did not display, continue to step 8.
NOTE
OK from the pop-up window.
Cancel to stop this action.
If all stockyards or all shelves are disabled at the same time, samples destined for the stockyard
will have no place to sort and will circulate around the track until one shelf is enabled.
PAUSE/RUN on the keypad display is lit.
2-29
Module Procedures
How to Disable the 3K or 5K Stockyard Shelves if an Unrecoverable Error Occurs
Figure 2.18 No Stockyard(s) Available
8
Select Exit to close the System Setup menu.
9
At the stockyard, verify the Control Panel is in Manual mode. Press PAUSE/RUN to clear the
Error at the keypad.
10
Remove the splash guard cover at the rear of the stockyard as follows:
•Press the
mode.
PAUSE/RUN button on the stockyard module keypad to put the module in PAUSE
CAUTION
Before you continue with this procedure, be sure the module is in PAUSE mode.
CAUTION
Even when a large capacity stockyard is in PAUSE mode, the air system is still
active and applying a constant air pressure to the shuttle cups and the gripper
arm assembly. This can cause unexpected movement of the shuttle cups and the
gripper arm assembly when resolving a jammed object error creating a possible
moving part or mechanical pinch hazard. Use caution when resolving jammed
object errors at a large capacity stockyard.
•Remove the splash guard cover screw with a Phillips screwdriver.
2-30
A97244AF
How to Disable the 3K or 5K Stockyard Shelves if an Unrecoverable Error Occurs
A016605P.EPS
Figure 2.19 Splash Guard (on a 3K Stockyard)
Module Procedures
2
NOTE Only remove the rear stockyard cover if access to the 3rd shelf is needed.
11
If tubes are in front of the bar code reader at the load or unload position, remove the tubes.
12
Remove any tubes from either the transfer-arm or retrieve-arm.
13
At the rear of the stockyard, verify there are no tubes or other obstructions in the path of the
load and unload shuttle cups for each level.
14
Remove any tubes in the load and unload shuttle cups as follows:
•Press the
•Press the
position to remove tubes.
•Remove any tubes from the shuttle cups. Press
FUNCTION +/- buttons to set the display at “31.”
ENTER button on the keypad to move the six shuttle cups to the Load/Unload
ENTER to clear FUNCTION “31.”
A97244AF
2-31
Module Procedures
How to Disable the 3K or 5K Stockyard Shelves if an Unrecoverable Error Occurs
15
At the front of the stockyard, remove any tubes inside the stockyard as follows:
•Press the
•Press the
PAUSE/RUN button to exit PAUSE mode.
FUNCTION +/- button to set the display at “30.”
CAUTION
After Function “30” has been activated, wait 30 seconds before opening any unit
doors. Serious injuries can occur if the user places their hands inside the unit while
the internal gripper arms are moving.
•Press the
front of the stockyard. Wait 30 seconds before opening any door.
16
Follow the steps below one shelf at a time.
•Remove the middle rack below the gripper.
•Check for the presence of a tube in the internal gripper, if a tube is present remove the tube.
•Replace the removed rack.
•Proceed to next shelf.
17
Remove all racks and tubes on the disabled shelf. Those samples can be remapped for future
use.
ENTER button on the keypad to move the internal grippers on each shelf to the
CAUTION
Failure to remove all samples from the disabled shelf or shelves can cause system
errors, hardware damage, sample tube breakage, and operator exposure to
biohazard conditions.
18
Replace empty racks in the stockyard shelf.
19
Reinstall all covers.
20
Press ENTER to clear FUNCTION “30.”
21
On the stockyard control panel, press AUTO button twice and the stockyard will initialize.
When initialization is complete the stockyard will continue normal operation.
2-32
A97244AF
How to Disable a Shelf in the 3K or 5K Stockyard
22
Verify the status in PrepLink, or the LIS, for sample tubes recovered from the BCR load and
unload, transfer-arm, retrieve-arm, internal grippers, and shuttle cups to determine if the
samples need further processing.
a. If needed, manually process any recovered sample tube(s).
b. If processing is complete, the sample(s) can be remapped to the stockyard for future use.
23
Contact your Beckman Coulter Service Representative if needed.
How to Disable a Shelf in the 3K or 5K Stockyard
If you want to disable a shelf of the stockyard for maintenance, for example, and you Do Not have
an unrecoverable error, perform the following procedure.
Module Procedures
2
CAUTION
Do not disable all stockyards or all shelves at the same time. A minimum of one
shelf of one stockyard is needed for continued sample processing on the system.
If all stockyards or all shelves are disabled at the same time, samples destined for
the stockyard will have no place to sort and will circulate around the track until
one shelf is enabled.
1
At the Line Controller main screen, select the System Setup from the menu bar.
2
Select the System Setup option. The System Setup dialog box appears.
3
In the System Setup dialog box, select Unit Enable/Disable.
The Unit Enable/Disable dialog box appears. The status (enabled or disabled) of the stockyard
shelf and other connections are shown.
A97244AF
2-33
Module Procedures
How to Disable a Shelf in the 3K or 5K Stockyard
4
Select the status of the correct shelf to Disable, then select OK. The Verify Stockyard Inactive
window displays (refer to Figure 2.20).
Figure 2.20 Verify Stockyard Inactive
5
Do the following:
a. Verify the stockyard is inactive.
b. Check that the
c.Select
6
If all shelves in all stockyards are disabled, the following pop-up window will display (PP
version 5.0 or later) indicating that no stockyard is available to store samples. Refer to
Figure 2.21.
a. Select
b. If this pop-up did not display, continue to step 7.
NOTE
Figure 2.21 No Stockyard(s) Available
OK from the pop-up window.
Cancel to stop this action.
If all stockyards or all shelves are disabled at the same time, samples destined for the stockyard
will have no place to sort and will circulate around the track until one shelf is enabled.
PAUSE/RUN on the keypad display is lit.
2-34
7
Select Exit to close the System Setup menu.
8
At the stockyard, verify that the Control Panel is in Manual mode.
9
Remove all the racks and tubes on the disabled shelf.
Those samples can be remapped for future use.
A97244AF
10
When ready to resume operation, press the AUTO/MANUAL button on the stockyard control
panel twice to initialize the stockyard.
When initialization is complete, the stockyard continues normal operation.
How to Enable the 3K or 5K Stockyard
1
Confirm that the sample tubes are not in the process of being transferred into, or out of, the
stockyard.
2
At the Line Controller main screen, select the System Setup from the menu bar.
Module Procedures
How to Enable the 3K or 5K Stockyard
2
3
Select the System Setup option. The System Setup dialog box appears.
4
In the System Setup dialog box, select Unit Enable/Disable.
5
The Unit Enable/Disable dialog box appears. The status (enabled or disabled) of the stockyard
shelf and other connections is shown.
6
Select the status box of the correct shelf to switch from Disabled to Enabled.
7
Select OK.
8
Select Exit to close the System Setup menu.
9
At the stockyard, press AUTO/MANUAL button on the keypad twice to change from MANUAL
mode to AUTO mode. When the initialization is complete, the stockyard will continue normal
operation.
A97244AF
2-35
Module Procedures
How to Enable the 3K or 5K Stockyard
2-36
A97244AF
CHAPTER 3
Error Recovery Procedures
This chapter provides information to identify and resolve errors related to the stockyard modules,
and system operational errors that are not specific to any Power Processor module.
NOTE
For error recovery information for other Power Processor Modules, refer to the correct Power
Processor Module Instructions for Use documentation.
WARNING
Perform Power Processor maintenance activities with caution.
Wear Personal Protective Equipment (PPE), such as gloves, eye shields, lab coats,
etc.
Wash hands thoroughly after contact with sample media and after all
maintenance activities.
Observe correct laboratory policies and laboratory procedures related to the
handling of biohazardous materials.
Refer to safety material sources (Material Safety Data Sheets, etc.) for specific
hazard information.
CAUTION
Before performing an error recovery procedure, confirm that the PAUSE button is
lit on that module.
A97244AF
CAUTION
Even when a large capacity stockyard is in PAUSE mode, the air system is still
active and applying a constant air pressure to the shuttle cups and the gripper
arm assembly. This can cause unexpected movement of the shuttle cups and the
gripper arm assembly when resolving a jammed object error creating a possible
moving part or mechanical pinch hazard. Use caution when resolving jammed
object errors at a large capacity stockyard.
Contact a Beckman Coulter Representative for assistance resolving operational issues not discussed
in this chapter.
3-1
Error Recovery Procedures
Error Recovery General Information
Error Recovery General Information
When an error condition occurs, the system generates an error message, sounds an alarm and
flashes a warning beacon. Error messages are shown on the keypad of the module where the error
has occurred, and in the Line Control Computer System Event Log, except for the Inlet, Decapper,
Recapper and the Secondary Decapper. The System Event Log automatically stores the 200 most
recent error messages in the order they occurred.
Power Processor Module Function Codes
The procedure to recover from an error condition can include entering one or more Function Codes
into one of the keypads on the Power Processor or a connected instrument.
The following sections offer specific procedures for recovering from identified errors. Table 3.1 lists
the Power Processor Module Function Codes by associated stockyard, with descriptions of the
action that was triggered by the function code. All Function Codes must be entered in MANUAL
mode.
Table 3.1 Power Processor Module Function Code Definitions
UnitFunction
Code
Outlet01Lane through
54Stop the arm while the gripper fingers are opened
5440-tube (high-speed)
Stockyard
3060-tube (high-speed)
Stockyard
3060-tube (standard-speed)
Stockyard
1020-tube Stockyard01Lane through
01Lane through
30Move Inside Arms on all shelves to the door side with grippers closed
31Move shuttle cups on all shelves to the outside of unit
01Lane through
30Move Inside Arms on all shelves to the door side with grippers closed
31Move shuttle cups on all shelves to the outside of unit
01Lane through
30Internal gripper arm moves to front, closed gripper
31Move cups out (back side of Stockyard)
Definition
3-2
A97244AF
General Error Recovery Procedure
When an error occurs at a hardware module, a flashing beacon and audible alarm activate. The
keypad display at the affected module will show which error code the error generated. This error
code is a four-digit numeral that the keypad display shows in 2, two-digit segments.
The procedure below should be performed any time an error occurs on the system.
1
Read the two-digit code from the keypad display on the affected hardware module.
Error Recovery Procedures
Error Recovery General Information
3
NOTE
2
Press the ALARM button on the keypad of the affected hardware module to silence the alarm.
3
Press one or both of the FUNCTION +/- buttons on the keypad to view the second two-digit code
from the keypad display.
4
Refer to the Error Code tables for the correct hardware module. For example, if the error occurs
at the 3K Stockyard, refer to Table 4.5, 3060-tube (standard-speed) Stockyard Error Codes in
CHAPTER 4.
5
Look up the code in the "Keypad Display" column of the error code table.
6
Read the information in the "Sensor," "Problem" and "Solution" columns for that error code.
Do not attempt to perform steps suggested in the Solution column yet.
7
Locate the sensor on the correct Sensor Diagram.
The leading zero is not displayed.
A97244AF
8
Spend a moment looking at the hardware. Try to find the cause of the error.
9
After investigating the cause of the error, perform the steps suggested in the "Solution"
column.
10
Press PAUSE/RUN to resume routine operation. Pay attention to the area around the sensor that
generated the error. Confirm that the error has been resolved.
11
If the error persists, repeat this procedure. Contact your local Beckman Coulter Representative
for any unrecoverable errors.
3-3
Error Recovery Procedures
System Error Recovery Procedures
System Error Recovery Procedures
This section provides information to identify and resolve errors related to system operational
problems. There are no specific error recovery procedures for Stockyard modules.
System Error Recovery Procedures include:
•Gripper Finger Error Recovery
•Jammed Sample Tube Carrier Recovery
•Bypassing a Module
•Flashing AUTO/MANUAL Button on Keypad
•Stop Button Recovery without Instrument Connections
•Stop Button Recovery with Instrument Connections
Gripper Finger Error Recovery
Error Recovery
The operator must evaluate gripper finger/transfer-arm errors to understand where the error
occurred in the loading or unloading process. Study the sample tube positioning and the Error
Tables to resolve gripper finger errors.
CAUTION
Even when a module is in PAUSE mode, gripper fingers can open or close
unexpectedly, creating a possible pinch hazard. Use caution when resolving the
gripper finger error conditions.
Jammed Sample Tube Carrier Recovery
The Power Processor system uses a series of sensors and stoppers to control the movement of
sample tube carriers through the system. The pneumatic stoppers engage and disengage to halt
sample tube carriers at specific locations. If a sensor is out of adjustment, the stopper can engage as
a sample tube carrier passes by, pinching the sample tube carrier and causing a jam. When a sample
tube carrier becomes jammed, the system generates an error message, sounds an alarm and flashes
a warning beacon.
3-4
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
CAUTION
Before attempting to release a jammed carrier, verify that the carrier is really
jammed by gently pushing it in the opposite direction that the conveyor belt is
moving. After doing this, if the carrier moves up the track, it is NOT jammed and
this procedure must NOT be used. Refer to the following NOTE for corrective
action.
3
NOTE
If the carrier moves up the track, check the sensor number nearest the stopped sample carrier and
refer to the correct error code table to resolve the error. Contact your Beckman Coulter Representative
for assistance.
Error Recovery
To recover from a jammed sample tube carrier error, follow this procedure (refer to Figure 3.1).
1
Press the ALARM button on the keypad to silence the alarm.
2
Gently push the sample tube carrier in the direction that the belt is moving until it "clicks" free.
IMPORTANT
pressure can damage the system.
3
Press the PAUSE/RUN button to resume routine operation.
Figure 3.1 Module Keypad (can vary slightly)
A jammed tube carrier should move with only a moderate amount of force. Excessive
A97244AF
A011079P.EPS
3-5
Error Recovery Procedures
System Error Recovery Procedures
Bypassing a Module
When a module is bypassed, transfer belts move sample tube carriers through the system. Sample
tube carriers can be bypassed at all modules by following the same procedure.
How to Bypass a Module
Any module of the Power Processor system can be bypassed using the procedure below. When
bypassing the Centrifuge, this must be done at the Centrifuge Track keypad.
CAUTION
Do not bypass all of the stockyard modules at the same time. If all stockyards are
bypassed at the same time, samples destined for the stockyard will have no place
to sort and will circulate around the track until one stockyard is put into AUTO
mode.
1
Press the AUTO/MANUAL button twice to switch to MANUAL mode. The AUTO/MANUAL button
indicator light stays lit to indicate that the module is in MANUAL mode.
2
Press the FUNCTION +/- buttons to set the readout to "01."
3
Press the ENTER button on each module keypad.
4
Wait for all sample tube carriers to pass through the module.
5
Press the AUTO/MANUAL button twice to return the module to AUTO mode.
Flashing AUTO/MANUAL Button on Keypad
Error Recovery
To recover from a condition where the AUTO/MANUAL button flashes continuously on the module
keypad and the module cannot be recovered, follow the steps below.
3-6
IMPORTANT
IMPORTANT The following procedure applies to Inlet, Outlet, Centrifuge, Connection Module and Aliquot
keypads.
If an error occurs after a module returns to AUTO mode, press PAUSE/RUN to clear the error.
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
Error Recovery Procedure
1
Press and hold the flashing AUTO/MANUAL button for a minimum of 10 seconds on the keypad.
The keypad changes to MANUAL mode.
If the condition occurs at...Then...
3
the Bar Code Verification Module in front
of the bar code reader,
the Hematology Outlet and a sample tube
is being placed in a personality or generic
rack,
the LX/DxC Connection Module and
sample tubes are being placed in the
Unload Shuttle,
OR
the sample tubes are in the Load Shuttle
or being placed in the Load Shuttle,
OR
there is a sample tube in front of the bar
code reader in the LX/DxC Connection
Module,
the Generic (CLSI)/IDC Connection
Module,
the 1K, 3K, or 5K Stockyard and a sample
tube is being loaded into the stockyard,
OR
the sample tube is being unloaded from
the stockyard,
remove the sample tube from the sample tube
carrier, and process the sample tube offline.
remove the sample tube from the gripper
fingers, and place it in the correct rack.
remove the sample tubes from the Unload
Shuttle, and place the tubes in the waiting
sample tube carriers.
remove the sample tubes, and process the
sample tubes offline.
remove the sample tube from the sample tube
carrier, and process the sample tube offline.
clear the sample tubes in the queue (from
BCR03 to BCR02) by entering Function 84 at
the Generic Connection module keypad.
These sample tubes must be processed
offline.
remove the sample tube from the sample tube
carrier.
remove the sample tube from the retrievearm and place in the waiting sample tube
carrier.
A97244AF
2
Press the AUTO/MANUAL button twice to return to AUTO mode at the module keypad. The
module Homes and sample processing continues.
IMPORTANT
error.
If an error occurs after a module returns to AUTO mode, press PAUSE/RUN to clear the
3-7
Error Recovery Procedures
System Error Recovery Procedures
Stop Button Recovery without Instrument Connections
Pressing any STOP button (except the Centrifuge STOP Button) will disconnect power to the Power
Processor system at any time during operation. The Line Control Computer and the PrepLink will
remain ON. The air system is not shut off when the STOP buttons are pressed.
NOTE
If the Centrifuge STOP Button is pressed, refer to the procedure, Stop Button Recovery for the
Centrifuge in the Centrifuge IFU.
IMPORTANT The STOP button can be pressed to halt any hazardous condition.
IMPORTANT The location of the last sample tube that was sorted by the system, cannot be accurately
represented at the Sample Locations screen in the Line Control Computer.
CAUTION
For system configurations with the Aliquot Module, when a STOP button is
pressed, aliquot tips containing a serum sample will drip slowly. The Serum Drip
Trays might need to be cleaned. For information on how to clean the serum drip
trays, refer to CHAPTER 5, “Daily Maintenance Procedures” in the General System
Operation IFU.
Stop Button Recovery
IMPORTANT Only perform the STOP Button Recovery steps that apply to the laboratory’s specific Power
Processor configuration.
1
Disengage the STOP button by pressing the button(s) again.
3-8
Figure 3.2 STOP Button
1
A011814P.EPS
1. STOP Button
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
2
For system configurations with the Aliquot Module, at the Aliquot Unit, lift and raise the aliquot
tip transfer-arms.
3
At the Outlet Modules, if any sample tubes are in the gripper fingers or in front of the bar code
reader, manually remove the tubes and process offline.
4
Press the green ON button at the Inlet Module to restore power to the system.
3
IMPORTANT
track. Correct any errors on each module and place modules back in AUTO mode, recovering
backward to the Inlet.
5
At the Outlet Module(s), press the AUTO/MANUAL button twice to set to AUTO mode. The gripper
arm(s) will initialize and return to Home position.
6
For system configurations with an Aliquot Module and a sample tube in front of bar code reader
#1, place a blockage between the sample at bar code reader #1 and the samples that are
preceding bar code reader #1 in the Aliquot Unit.
7
At the Aliquot Unit, press the FUNCTION +/- buttons to set the readout to "84" and press ENTER.
The aliquot sample tubes will sort to the Aliquot Error Rack.
8
At the Labeler Unit, remove any dropped aliquot tubes from the top of the aliquot tube tray, in
the roller area, and in the aliquot tube chute.
9
Set the AUTO/MANUAL button on the Labeler Unit to AUTO mode. Aliquot sample tubes that
were being processed will be discarded into the defective aliquot tube disposal container.
Perform the system recoveries beginning with the Outlet or stockyard at the end of the
A97244AF
10
Set the AUTO/MANUAL button on the Aliquot Unit to AUTO mode. The aliquot tip transfer-arms
will move forward and discard the aliquot tips into the biohazard container. The primary tubes
will sort to the Aliquot Error Rack.
11
At the bar code reader #1 in the Aliquot Unit, remove the blockage that was set there in step 6.
12
Set the AUTO/MANUAL button on the Decapper Module to AUTO mode. Sample tubes that are
waiting will be decapped.
3-9
Error Recovery Procedures
System Error Recovery Procedures
13
For system configurations with the Aliquot Module, if there are no sample tubes present in the
Serum Level Detector Unit, set the
14
If tubes are in the Serum Level Detector gripper fingers, or in the optical wells, remove the
tubes and place them in sample tube carriers inside the Serum Level Detector Unit.
15
Press the FUNCTION +/- buttons to set the readout to "84" and press ENTER. This will move the
sample tube carriers from the Serum Level Detector Unit.
16
Set the AUTO/MANUAL button on the Serum Level Detector Unit to AUTO mode.
17
If there are no samples at the Centrifuge, set the AUTO/MANUAL button on the Centrifuge to
AUTO mode, and proceed to step 36.
AUTO/MANUAL button to AUTO mode and proceed to step 16.
18
If a sample tube is in front of the Centrifuge Track bar code reader, it can proceed down the
track and bypass the Centrifuge when the Centrifuge is set to AUTO mode in step 35. This
situation could be a problem if that tube required centrifugation.
a. Check the sample programming to determine if the tube requires centrifugation.
b. If centrifugation is needed, set the sample tube aside and record the sample ID
(identification of this tube will be needed in step 31b.
c.If centrifugation is not needed, do not remove the tube.
19
Turn the Centrifuge power OFF and remove the Centrifuge safety shield.
20
Remove any sample tubes or balance tubes from the grippers. Set them aside and determine if
these tubes were being loaded or unloaded from the Centrifuge.
21
Confirm there are no jammed sample tube carriers in the Centrifuge Track queue (D-Lane).
22
Remove any sample tubes from sample tube carriers that have entered the Centrifuge Track DLane. Set them aside and determine if these tubes were being loaded or unloaded from the
Centrifuge.
3-10
23
Push the Centrifuge transfer-arm to the Home position over the loading area of the Centrifuge
Track D-Lane.
24
Turn the Centrifuge power ON. Confirm the Centrifuge is in MANUAL mode.
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
25
Turn the Door Release Lock switch to the right to unlock the Centrifuge. If the Centrifuge does
not unlock, confirm that the Centrifuge transfer-arm is completely in the Home position.
26
Open the Centrifuge lid.
a. If there are tubes in the Centrifuge, follow your laboratory procedures to determine if the
tubes have been fully spun or need to be re-spun. If the tubes need to be re-spun, proceed
to step 27.
b. If you determine that the tubes do not need to be re-spun, proceed to step 31c.
c.If there are no tubes in the Centrifuge, proceed to step 27.
27
Manually load the Centrifuge with the tubes set aside in steps 18b, 20, and 22 that needed to be
spun. Confirm the load is balanced.
3
28
Close the Centrifuge lid and confirm both latches are secure.
29
Confirm that the Centrifuge is in MANUAL mode.
30
On the Centrifuge, press the FUNCTION +/- buttons to set the readout to "80" and press ENTER.
Press the
31
After centrifugation is complete,
a. Turn the release door switch to the Right to open the Centrifuge lid.
b. Manually unload the sample tube from step 18b and return it to the empty carrier in front
c.Manually unload the sample tubes into the sample tube carriers in the Centrifuge Track D-
d. Close the Centrifuge lid and confirm both latches are secure.
e. Replace the Centrifuge safety shield.
32
If there were any tubes set aside in step 20 and 22 that did not need to be re-spun, manually load
those tubes into the sample tube carriers in the Centrifuge Track D-Lane.
PAUSE/RUN button to start centrifugation.
of the Centrifuge Track bar code reader.
Lane.
A97244AF
33
On the Centrifuge Track keypad, confirm the Centrifuge is in MANUAL mode. Press the
FUNCTION +/- buttons to set the readout to "84" and press ENTER.
34
On the Centrifuge keypad, press the PAUSE/RUN button to move the sample tube carriers from
Centrifuge Track D-Lane area.
3-11
Error Recovery Procedures
System Error Recovery Procedures
35
After all the sample tubes have cleared the last Centrifuge Track D-Lane, set the AUTO/MANUAL
button on the Centrifuge to AUTO mode.
36
If the system is configured with a Hematology Outlet,
•remove the tube from the gripper fingers, and process it offline. Set the
button to AUTO mode.
If the system is configured with a Bar Code Verification Module,
•set the
37
At the Inlet Module...
AUTO/MANUAL button to AUTO mode.
•If any samples are still in the Inlet gripper finger(s), they must be removed and placed into
the partially unloaded rack.
•If a pa rti all y un loa ded rac k (a ny f our Inl et r ack s) n eed s to be p roc ess ed, pla ce a sam ple tub e
in location #1 of the rack and move all tubes forward so there are no empty tube positions.
IMPORTANT
Rack, as an empty tube position will cause the Priority Rack to be ignored.
For a Dynamic Inlet, this step is important if the rack that is being processed is a Priority
•Finally, set the AUTO/MANUAL button to AUTO mode to continue sample tube processing.
Stop Button Recovery with Instrument Connections
If the STOP button at any Power Processor module (except the Centrifuge STOP Button) is pressed
when samples are on the instrument(s), follow the procedure below to recover.
AUTO/MANUAL
3-12
NOTE
If the Centrifuge STOP Button is pressed, refer to the procedure, Stop Button Recovery for the
Centrifuge in the Centrifuge IFU.
This procedure includes all module options and supported instrument types. Only perform steps for
modules and instruments that are specific to the laboratory’s system configuration.
IMPORTANT
The STOP button can be pressed to halt any hazardous condition.
IMPORTANT If a sample tube breaks in the Load or Unload Shuttle on the instrument Connection Module
and all debris can be completely removed from the shuttle, proceed with established laboratory cleanup
procedures. If all debris CANNOT be removed from the shuttle, contact your local Beckman Coulter
Representative.
IMPORTANT The location of the last sample tube that was sorted by the system can not be accurately
represented at the Sample Locations screen.
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
CAUTION
For system configurations with the Aliquot Module, when a STOP button is
pressed, aliquot tips containing a serum sample will drip slowly. The Serum Drip
Trays might need to be cleaned. For information on how to clean the serum drip
trays, refer to CHAPTER 5, “Daily Maintenance Procedures” in the General System
Operation IFU.
Stop Button Recovery
IMPORTANT Only perform the STOP Button Recovery steps that apply to the laboratory’s specific Power
Processor configuration.
Perform the system recoveries beginning with the LX/DxC connected instruments, Aliquot Module, and
Outlet or stockyard at the end of the track, and work your way towards the Inlet Module. Correct any
errors on each module and place modules back in AUTO mode, recovering backward to the Inlet.
1
Let the instrument(s) continue to run and continue to process samples.
3
2
If sample tubes are in the quadruple-gripper transfer-arm of the LX/DxC connection unit, do
the following to remove them:
a. Open the access door above the LX/DxC auto-loader to gain access to the quadruple-
gripper transfer-arm.
b. Reach through the access door and manually pull the transfer-arm toward the door so the
grippers can be accessed from the door opening.
c.Locate solenoid SL17 on the transfer-arm and note the button on top of the solenoid. This
button can be either blue or orange. (Refer to Figure 3.3.)
NOTE
This button will open the quadruple-grippers. Do Not press this button until you have
inserted a LX/DxC loading rack under the tubes to hold them in place when the grippers open.
A97244AF
3-13
Error Recovery Procedures
System Error Recovery Procedures
Figure 3.3 Gripper release solenoid SL17
d. Place a LX/DxC rack under the tubes, raise it so the tubes are inside the rack and hold it in
place.
e. While holding the LX/DxC rack with tubes securely in the rack, use the end of a lab swab or
other implement to push the button on top of solenoid SL17. The grippers will open and the
tubes can be removed. (Refer to Figure 3.4.)
NOTE
Figure 3.4 Opening the quadruple-grippers
Hold down the button securely to keep the grippers open while you remove the tubes.
3-14
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
f.Lower the rack and tubes, and then release the solenoid button.
g. Remove the rack and tubes and close the access door.
3
Check the sample tube status at the LX/DxC instrument console to see if they are completed.
If…Then…
samples are completed,transfer these tubes to the sample tube
carriers that are waiting to be unloaded at
the Connection Module after the Connection
Module is put back in AUTO mode.
samples are not completed,transfer these tubes to an offline instrument
rack in the output tray on the right side of
the instrument.
4
Remove sample tubes from the racks in the LX/DxC instrument load tray. Transfer these tubes
to another rack(s) and put this rack(s) in the output tray on the right side of the instrument(s).
3
5
Remove sample tubes from the instrument load shuttle. Place these tubes into another rack and
place this rack in the output tray on the right side of the instrument.
6
Check the instrument console to confirm that the instrument is functional. If it is not, refer to
the instrument IFU for instructions on restoring the instrument to operation.
7
Process the racks in the right side output tray using PRIORITY LOAD. When these samples are
completed, transfer these tubes to sample tube carriers that are waiting for unloading at the
Connection Module after the Connection Module is placed into AUTO mode. Process a
maximum of 2 racks at a time to permit the automation system to continue loading racks onto
the LX/DxC instrument after the track is re-started.
8
Disengage the STOP button by pressing the button(s) again.
9
For system configurations with the Aliquot Module, lift and raise the aliquot tip transfer-arms.
10
Remove the stockyard splash guard cover.
A97244AF
3-15
Error Recovery Procedures
System Error Recovery Procedures
11
Remove the sample tube from the stockyard grippers. The Home position for the input arm
opens the grippers and the sample tube falls after the stockyard is put in AUTO mode.
If…Then…
the STOP button was pressed
during a sample tube
retrieval from the stockyard,
a tube was stuck on or
between the third shelf,
12
Replace the stockyard splash guard cover.
13
Open each large capacity stockyard door and check each rack for tubes that fell inside or behind
the stockyard during the stop procedure. Remove all fallen tubes.
14
At the PrepLink Locations tab, search by Sample ID and record the location for each tube
removed from the stockyard. Put all samples shown on the
rack locations. Process any samples that are not shown on the
15
At Outlet Modules 1A and 1B, if any sample tubes are in the gripper fingers, manually remove
the tubes and process offline.
remove the sample tube(s) in the stockyard
from all gripper fingers (inside and back) and
loading or unloading shuttles.
remove the protective rear cover to retrieve
the sample.
Locations tab back in their identified
Locations tab offline.
16
Press the green ON button on the Inlet Module to restore power to the system.
17
At Outlet Modules 1A and 1B, set the AUTO/MANUAL button to AUTO mode.
18
Set the AUTO/MANUAL button on the Secondary Decapper to AUTO mode.
19
If there are no large capacity stockyard errors, proceed to step 21.
If…Then…
the STOP button was pressed during a
sample tube load at the stockyard,
with a tube in the grippers,
20
At the large capacity stockyard keypad(s), press the PAUSE/RUN button.
after powering ON, wait for the arm to move
to the second shelf input area.
3-16
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
21
At the large capacity stockyard keypad(s), set the AUTO/MANUAL button to AUTO mode.
22
Set the AUTO/MANUAL button on the Recapper to AUTO mode.
23
If there are no sample tubes at the Recapper, proceed to step 30.
24
Check for caps in the Recapper chute.
a. At the Recapper, open the front door and rear door.
b. Look for caps in the chute and lateral cap pusher (1).
c.If there are caps, proceed to step 25. If there are no caps, proceed to step 26.
25
To remove caps from the chute:
3
a. From the rear of the Recapper, lift and remove the curved (clear plastic) cap deflector (3)
from over the chute loop.
b. Remove all caps in the chute.
c.Reinstall the cap deflector.
26
From the rear of the Recapper, remove caps from the lateral cap pusher:
a. If there are sample tubes in the capping area, push them back away from the capping area.
b. Remove all caps at the lateral cap pusher. If there is a cap in the metal cylinder, use a flat
head screwdriver to push the cap down and out of the cylinder. Confirm there are no caps
on the track.
IMPORTANT
27
In the Recapper:
there is a sample tube
with a cap,
If one or more caps stay in the lateral cap pusher, there might be a 2_10 error.
If…Then…
remove the sample tube from the sample tube
carrier.
IMPORTANT
carrier, there will be a 1_06 error.
If the capped tube stays in the sample tube
A97244AF
there is a sample tube
without a cap,
do NOT remove the sample tube from the sample
tube carrier.
3-17
Error Recovery Procedures
System Error Recovery Procedures
28
At the Recapper, set the AUTO/MANUAL button to AUTO mode.
If…Then…
there is a sample tube
without a cap,
a sample tube with a cap was
removed in step 27.
29
Follow the instructions in the Error Code section for any errors that occur at the Recapper.
Refer to the Decapper and Recapper Modules IFU.
30
Confirm there are 8 empty automation racks in the LX/DxC instrument load tray. (Refer to the
“Returning LX/DxC Load Tray to 8 Racks” procedure in the LX and DxC Connection Modules IFU.)
Set the
31
Remove any remaining tubes from the instrument(s) unload tray and transfer them to the
sample tube carriers that are waiting for unloading at the Connection Module.
IMPORTANT
AUTO/MANUAL button on the Connection Module to AUTO mode.
Pause the Connection Module while placing tubes into carriers. Press PAUSE/RUN to
resume processing.
the Recapper puts a cap on the sample tube.
the empty sample tube carrier will route to the
bar code reader at the large capacity stockyard.
Put the capped tube in the sample tube carrier.
32
At the Generic Connection Modules:
If…Then…
the Generic Connection is an
AU2700/5400 connection,
a sample tube is at the aspiration
location,
the sample was aspirated before
the STOP button was pressed,
the sample was NOT aspirated
before the STOP button was
pressed,
33
Sample tubes in the queue (from BCR03 to BCR02), must be processed offline. At the processing
keypad, press
FUNCTION +/- buttons to set the readout to "84" and press ENTER.
contact your local Beckman Coulter
Representative for error recovery
instructions for this module.
check the Sample ID status at the Generic or
IDC Connection instrument(s). If the sample
is in process, it will go to sample completion.
sample processing will continue at the
instrument(s).
the sample must be processed offline.
3-18
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
34
After the sample queue is cleared, set the AUTO/MANUAL button at each Generic and IDC
Connection Module keypad to AUTO mode.
35
Set the AUTO/MANUAL button on the H-Lane to AUTO mode.
36
For system configurations with an Aliquot Module and a sample tube in front of bar code reader
#1, place a blockage between the sample at bar code reader #1 and the samples that are
preceding bar code reader #1 in the Aliquot Unit.
37
At the Aliquot Unit, press the FUNCTION +/- buttons to set the readout to "84" and press ENTER.
Any sample tubes that were not aliquoted, will route to Aliquot Error Rack.
38
At the Labeler Unit, remove any dropped aliquot tubes from the top of the aliquot tube tray, in
the roller area, and in the aliquot tube chute.
3
39
Set the AUTO/MANUAL button on the Labeler Unit to AUTO mode. Aliquot sample tubes that
were being processed will be discarded into the defective aliquot tube disposal container.
40
Set the AUTO/MANUAL button on the Aliquot Unit to AUTO mode. The aliquot tip transfer-arms
will move forward and discard the aliquot tips into the biohazard container. The primary tubes
will sort to the Aliquot Error Rack.
41
At the bar code reader #1 in the Aliquot Unit, remove the blockage that was set there in step 36.
42
Set the AUTO/MANUAL button on the Decapper to AUTO mode. Waiting sample tubes will be
decapped.
43
If there are no sample tubes present at the Serum Level Detector Unit, set the AUTO/MANUAL
button to AUTO mode. Proceed to step 47.
44
If tubes are in the Serum Level Detector gripper fingers, or in the optical wells, remove the
tubes and place them in sample tube carriers inside the Serum Level Detector Unit.
A97244AF
45
At the Serum Level Detector keypad, press the FUNCTION +/- buttons to set the readout to "84"
and press
ENTER. This will move the sample tube carriers from the Serum Level Detector Unit.
3-19
Error Recovery Procedures
System Error Recovery Procedures
46
Set the AUTO/MANUAL button on the Serum Level Detector Unit to AUTO mode.
47
If there are no samples at the Centrifuge, set the AUTO/MANUAL button to AUTO mode and
proceed to step 66.
48
If a sample tube is in front of the Centrifuge Track bar code reader, it might proceed down the
track and bypass the Centrifuge when the Centrifuge is set to AUTO mode in step 65. This
situation could be a problem if that tube required centrifugation.
a. Check the sample programming to determine if the tube requires centrifugation.
b. If centrifugation is needed, set the sample tube aside and record the sample ID
(identification of this tube will be needed in step 61b).
c.If centrifugation is not needed, do not remove the tube.
49
Turn the Centrifuge power OFF and remove the Centrifuge safety shield.
50
Remove any sample tubes or balance tubes from the grippers. Set them aside and determine if
these tubes were being loaded or unloaded from the Centrifuge.
51
Confirm there are no jammed sample tube carriers in the Centrifuge Track queue (D-Lane).
52
Remove any sample tubes from sample tube carriers that have entered the Centrifuge Track DLane. Set them aside and determine if these tubes were being loaded or unloaded from the
Centrifuge.
53
Push the Centrifuge transfer-arm to the Home position over the loading area of the Centrifuge
Track D-Lane.
54
Turn the Centrifuge power ON. Confirm the Centrifuge is in MANUAL mode.
55
Turn the Door Release Lock switch to the right to unlock the Centrifuge. If the Centrifuge does
not unlock, confirm that the Centrifuge transfer-arm is completely in the Home position.
3-20
56
Open the Centrifuge lid.
a. If there are tubes in the Centrifuge, follow your laboratory procedures to determine if the
tubes have been fully spun or need to be re-spun. If the tubes need to be re-spun, proceed
to step 57.
b. If you determine that the tubes do not need to be re-spun, proceed to step 61c.
A97244AF
Error Recovery Procedures
System Error Recovery Procedures
c.If there are no tubes in the Centrifuge, proceed to step 57.
57
Manually load the Centrifuge with the tubes set aside in steps 48b, 50, and 52 that needed to be
spun. Confirm the load is balanced.
58
Close the Centrifuge lid and confirm both latches are secure.
59
Confirm that the Centrifuge is in MANUAL mode.
60
On the Centrifuge, press the FUNCTION +/- buttons to set the readout to "80" and press ENTER.
Press the
61
After centrifugation is complete,
PAUSE/RUN button to start centrifugation.
3
a. Turn the release door switch to the Right to open the Centrifuge lid.
b. Manually unload the sample tube from step 48b and return it to the empty carrier in front
of the Centrifuge Track bar code reader.
c.Manually unload the sample tubes into the sample tube carriers in the Centrifuge Track D-
Lane.
d. Close the Centrifuge lid and confirm both latches are secure.
e. Replace the Centrifuge safety shield.
62
If there were any tubes set aside in step 50 and 52 that did not need to be re-spun, manually load
those tubes into the sample tube carriers in the Centrifuge Track D-Lane.
63
On the Centrifuge Track keypad, confirm the Centrifuge is in MANUAL mode. Press the
FUNCTION +/- buttons to set the readout to "84" and press ENTER.
64
On the Centrifuge keypad, press the PAUSE/RUN button to move the sample tube carriers from
Centrifuge Track D-Lane area.
65
After all the sample tubes have cleared the last Centrifuge Track D-Lane, set the AUTO/MANUAL
button on the Centrifuge to AUTO mode.
A97244AF
66
If the system is configured with a Hematology Outlet,
•remove the tube from the gripper fingers and then process it offline. Set the
button to AUTO mode.
AUTO/MANUAL
3-21
Error Recovery Procedures
Stockyard Module Error Recovery Procedures
If the system is configured with a Bar Code Verification Module,
•set the
67
At the Inlet Module...
•If any samples are still in the Inlet gripper finger(s), they must be removed and placed into
the partially unloaded rack.
•If a pa rti all y un loa ded rac k (a ny f our Inl et r ack s) n eed s to be p roc ess ed, pla ce a sam ple tub e
in location #1 of the rack and move all tubes forward so there are no empty tube positions.
AUTO/MANUAL button to AUTO mode.
IMPORTANT
Rack, as an empty tube position will cause the Priority Rack to be ignored.
•Finally, set the AUTO/MANUAL button to AUTO mode to continue sample tube processing.
For a Dynamic Inlet, this step is important if the rack that is being processed is a Priority
Stockyard Module Error Recovery Procedures
This section provides information to identify and resolve errors related to the Stockyard modules.
Stockyard error recovery procedures include:
•Outlet Module Error Recovery
•Fallen Sample Tube Inside the 1020-tube Stockyard
•Fallen Sample Tube Inside the 3060-tube or 5440-tube Stockyard
Outlet Module Error Recovery
1
Press the ALARM button on the Outlet Module keypad to silence the alarm.
3-22
2
Read the error code from the keypad display.
3
Press the FUNCTION +/- buttons on the keypad to show the complete error code.
4
Locate the error code in the Outlet Module Error Codes section of the Power Processor manual.
5
Resolve the condition causing the error message.
6
Press the PAUSE/RUN button on the keypad to resume routine operation.
A97244AF
Fallen Sample Tube Inside the 1020-tube Stockyard
1
Press the ALARM button on the module keypad to silence the alarm.
Error Recovery Procedures
Stockyard Module Error Recovery Procedures
3
IMPORTANT
2
If required, move the Transfer Load or Transfer Retrieval Arm before reaching inside the
stockyard to remove the dropped sample tube.
Hold the transfer-arm while retrieving the dropped sample tube.
3
After removing the dropped sample tube, release the transfer-arm. The arm moves to the Home
position.
4
Press PAUSE/RUN to resume routine operation and then process the sample tube offline.
Confirm that the module is in PAUSE mode before proceeding.
Fallen Sample Tube Inside the 3060-tube or 5440-tube Stockyard
NOTE During this procedure, if you see that a cap is missing from a sample tube in the stockyard, double-
check that the module is in PAUSE mode and then place a cap on the tube. Also remove any fallen caps
from within the stockyard. Contact your Beckman Coulter Service Representative to make any needed
adjustments to the Recapper.
1
Press the ALARM button on the module keypad to silence the alarm.
IMPORTANT
2
Determine where the error occurred. Open the shelf doors and then use a flashlight to search
inside the module.
3
If required, move the transfer load or transfer retrieval-arm before reaching inside the
stockyard to remove the dropped sample tube.
Hold the transfer-arm while retrieving the dropped sample tube.
4
If the sample tube has dropped on top of a rack, pull out the rack and retrieve the tube.
5
If the sample tube has dropped near the shuttle cup area, remove the tube as follows:
•Press the
Confirm that the module is in PAUSE mode before proceeding.
FUNCTION +/- buttons to set the display at “31.”
A97244AF
3-23
Error Recovery Procedures
Stockyard Module Error Recovery Procedures
•Press the ENTER button on the keypad to move the six shuttle cups to the Load/Unload
position and remove the dropped tube from the shuttle cup area.
•Press
6
If the sample tube dropped to the bottom level of the stockyard, open the lowest level doors.
Use a flexible claw-like grabber tool to reach inside the stockyard to retrieve the sample tube.
7
If you are unable to remove the dropped sample tube, contact your Beckman Coulter Service
Representative.
8
After removing the dropped sample tube, release the transfer-arm. The arm moves to the Home
position.
9
Verify the status in PrepLink, or the LIS, for the sample tube recovered from the stockyard to
determine if the sample needs further processing.
a. If needed, manually process the recovered sample tube.
b. If processing is complete, remap the sample to the stockyard for future use.
ENTER to clear FUNCTION “31.”
10
Press PAUSE/RUN to resume routine operation.
3-24
A97244AF
This chapter provides important troubleshooting information on how to locate and resolve
problems with the Stockyard module. It is divided into 2 sections:
•Error Code Tables
•Sensor Diagrams
Error Code Tables
The Error Code tables describe error messages that are generated by the Power Processor system.
The tables provide system error codes, a brief explanation of the possible problem, and possible
solutions.
CHAPTER 4
Troubleshooting
Error Code Explanation
The "Sensor" column contains the alphanumeric codes as they appear in the Line Control Computer
System Error Log. These codes, (SN09 PAS, for example) refer to specific sensors on the track.
The "Keypad Display" column contains the error codes shown on the keypads located at each
module of the system.
The "Problem" column provides a description of the cause of the error.
The "Solution" column provides a way to resolve an error condition.
A97244AF
4-1
Troubleshooting
Error Code Tables
Nomenclature
Table 4.1 contains definitions of the alphabetic portion of error codes found in the "Sensor" column
of the Error Code tables in this chapter.
Use this table and the color diagrams in the Sensor Diagrams section, to locate errors on the system.
Table 4.1 Error Code Definitions
CodeDefinition
AMAC synchronous motor
ASMagnetic auto switch
BRBar code reader
BZAudible alarm (buzzer)
DMDC motor
LPLamp (keypad and warning light)
LSMechanical limit switch
PMPulse/stepper motor
SLPneumatic solenoid
SNSensor
SWKeypad switch
Unit Error Code Categories
•The Error Code is a three-digit number. The keypad on the error unit shows the 1_digit segment.
The first digit represents the error category. (Refer to Table 4.2.)
•Press the
the component (for example, solenoid or sensor) causing the error.
Example:
If the Error Code is 1_03, then this represents Sensor Error 03(SN03). The sensor for the
Stockyard indicates that Rack #3 is not set correctly. Refer to Table 4.4, 1020-tube Stockyard
Error Codes, and Figure 4.2, 1020-tube Stockyard Main Body Side in this chapter.
•Press the
sets of digits.
FUNCTION – button on the keypad and the last two digits appear. These digits refer to
FUNCTION + and FUNCTION – button alternately to toggle back and forth to see both
4-2
A97244AF
Error Code Tables
Table 4.2 Unit Error Code Categories
CategoryError DescriptionProblemSolution
0_xxCould not read bar code.Invalid bar code. • Attach valid bar code.
• Check that the bar code
label is correct and clean.
Sample information was not
received from LIS.
Sample is not at the correct
position.
• Valid communication
needed.
Troubleshooting
4
1_xxCannot pass carrier to sensor
SNxx.
Can also refer to ASxx.
At Startup sample tube
carrier was located in the
sensor.
2_xxCannot move cylinder. • Check the solenoid/cylinder
3_xxBar code reader not working
or label not facing bar code
reader.
4_xxPulse Motor/AC Motor Error. • Check motors and drive
5_xxAnalyzer is not ready, or the
tube has stopped
somewhere.
7_xxAuto sensor error (ASxx). • Confirm that the sensor
9_xxAC motors on arm. • Confirm that the AM1 and
• Check the sensor position
and then look for possible
jamming.
• Check that carrier is
separated from the sensor
at Startup.
position and look for
possible jamming.
• Check that the bar code
reader light comes on and
that the carrier rotates.
belts.
• Confirm that the Analyzer is
Ready or whether the
cassette is jammed
between Analyzer and
Connection Unit.
position and the cylinder
work correctly.
AM2 work correctly.
A97244AF
Temperature errorStockyard internal
temperature out-of-range.
• Check the condition of
refrigerator. To continue
operation, press the
PAUSE/RUN button.
4-3
Troubleshooting
Error Code Tables
200-tube Stockyard Error Codes
Table 4.3 represents Error Codes as they appear in the Line Control Computer Systems Error Log
and Error Signals as they appear on the 200-tube Stockyard keypad display. A brief description of
the problem and a possible solution are also found in the table.
Table 4.3 200-tube Stockyard Error Codes
SensorKeypad
Display
SN22
PAS
SN23
PAS
SN33
PAS
SL202_20The first stopper on the sample tube
SL21
ON
SL21
OFF
1_22A sample tube carrier was passed
through from the sorting point on the
through lane, but the sample tube
carrier was not detected.
1_23A sample tube carrier was diverted to
the empty sample tube carrier waiting
lane, but the sample tube carrier was
not detected.
1_33The stopper retracted at the
intersection of the empty sample tube
carrier waiting lane, but the sample tube
carrier was not detected.
carrier sorting lane did not engage
completely.
2_21The stopper on the sorting point on the
sample tube carrier sorting lane did not
retract completely.
2_21Error at the sensor before the sorting
point of the sorting lane.
ProblemSolution
1. Check that a sample tube carrier is not
jammed or stuck. Refer to CHAPTER 3,
Jammed Sample Tube Carrier Recovery.
2. Place a metallic object in front of SN22 to turn
it ON.
1. Check that a sample tube carrier is not
jammed or stuck. Refer to CHAPTER 3,
Jammed Sample Tube Carrier Recovery.
2. Place a metallic object in front of SN23 to turn
it ON.
1. Check that a sample tube carrier is not
jammed or stuck. Refer to CHAPTER 3,
Jammed Sample Tube Carrier Recovery.
2. Place a metallic object in front of SN33 to turn
it ON.
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Tube Carrier Recovery.
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Tube Carrier Recovery.
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Tube Carrier Recovery.
SL222_22Diverter malfunction at the intersection
of the through lane and the empty
sample tube carrier waiting lane.
SL31
ON
SL31
OFF
SL32
ON
2_31The stopper at the empty sample tube
carrier waiting lane did not engage
completely.
2_31The stopper at the empty sample tube
carrier waiting lane did not retract
completely.
2_32Sensor error in the empty sample tube
carrier waiting lane.
4-4
Confirm that a sample tube carrier is not jammed
and that nothing is jammed on the diverter.
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Tube Carrier Recovery.
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Tube Carrier Recovery.
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Tube Carrier Recovery.
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Tube Carrier Recovery.
Confirm that a sample tube carrier is not jammed
or stuck. Refer to CHAPTER 3, Jammed Sample
Tube Carrier Recovery.
Confirm that a sample tube carrier is not jammed
and that the diverter is open.
Confirm that a sample tube carrier is not jammed
and that the diverter is closed.
A97244AF
4-5
Troubleshooting
Error Code Tables
1020-tube Stockyard Error Codes
Table 4.4 represents Error Codes as they appear in the Line Control Computer Systems Error Log
and Error Signals as they appear on the 1020-tube Stockyard keypad display. A brief description of
the problem and a possible solution are also found in the table.
Table 4.4 1020-tube Stockyard Error Codes
SensorKeypad
Display
BUZZ
CMD
BUZZ
CMD
BUZZ
CMD
BUZZ
CMD
BUZZ
CMD
BUZZ
CMD
SN01
OFF
SN02
OFF
SN03
OFF
0_11Bar code reader #1: Error at bar code
0_12Bar code reader #1: Positioning error at
0_13Bar code reader #1: Invalid route code
0_21Bar code reader #2: Error at bar code
0_22Bar code reader #2: Positioning error at
0_23Bar code reader #2: Invalid route code
1_01Rack #1 is not set correctly.Set the rack correctly.
1_02Rack #2 is not set correctly.Set the rack correctly.
1_03Rack #3 is not set correctly.Set the rack correctly.
ProblemSolution
reader.
bar code reader.
at bar code reader. Additional sample
programming caused a sorting conflict.
reader.
bar code reader.
at bar code reader. Additional sample
programming caused a sorting conflict.
1. Inspect the sample to see if the bar code label
is readable.
2. Press PAUSE/RUN.
1. Inspect the sample to see if the bar code label
is readable.
2. Remove the sample and press PAUSE/RUN.
1. Remove the sample tube and press ENTER.
2. Remove the sample tube and press PAUSE/
RUN.
1. Inspect the sample to see if the bar code label
is readable.
2. Press PAUSE/RUN.
1. Inspect the sample to see if the bar code label
is readable.
2. Remove the sample and press PAUSE/RUN.
1. Remove the sample tube and press ENTER.
2. Remove the sample tube and press PAUSE/
RUN.
SN04
ON
SN05
ON
SN401_40Error at proximity sensor SN40 in the
1_04A sample was not loaded into the
sample tube carrier.
1_05A sample was not retrieved from the
sample tube carrier.
inner lane.
4-6
Confirm that loading/retrieving of sample tubes
works correctly. If necessary, adjust the fiber
sensor SN04.
Confirm that loading/retrieving of sample tubes
works correctly. If necessary, adjust the fiber
sensor SN05.
1. Confirm that the sensor works correctly. The
Red LED should be on when a carrier is within
range of SN40.