BEAMEX FB150, FB350, FB660 User Manual

0 (0)
FBXXX___tgeng0000
Field Temperature Block series
Beamex® FB150, FB350, FB660
User Guide
ii
iii
Table of Contents
1 Before You Start .......................................................................1
1.1 Introduction ............................................................................................... 1
1.2 Unpacking ................................................................................................ 2
1.3 Symbols Used ........................................................................................... 3
1.4 Safety Information ..................................................................................... 4
1.4.1 Warnings .........................................................................................................4
1.4.2 Cautions ..........................................................................................................6
1.5 CE Comments ........................................................................................... 7
1.5.1 EMC Directive .................................................................................................7
1.5.2 Immunity Testing .............................................................................................8
1.5.3 Emission Testing ..............................................................................................8
1.5.4 Low Voltage Directive (Safety) ........................................................................8
1.6 Authorized Service Centers ...................................................................... 9
2 Specications and Environmental Conditions .................... 11
2.1 Specifications ......................................................................................... 11
2.2 Environmental Conditions ....................................................................... 12
3 Quick Start .............................................................................. 13
3.1 Setup ....................................................................................................... 13
3.2 Parts and Controls .................................................................................. 14
3.2.1 Display Panel ................................................................................................15
3.2.2 Display ..........................................................................................................16
3.2.3 Power Panel ..................................................................................................17
3.2.4 -R Option Panel (-R models only) ..................................................................20
3.3 Languages .............................................................................................. 21
3.3.1 Language Selection ......................................................................................22
3.3.2 Reset to English Language ...........................................................................22
4 Menu Structure .......................................................................23
4.1 Temp Setup Menu ................................................................................... 23
4.2 Prog Menu .............................................................................................. 24
4.3 System Menu .......................................................................................... 25
4.4 Input Setup (-R only) ............................................................................... 26
5 Controller operation...............................................................27
iv
5.1 Main Screen ............................................................................................ 27
5.2 Main Menu .............................................................................................. 28
5.2.1 Temp Setup ................................................................................................... 28
5.2.2 Prog Menu .....................................................................................................30
5.2.3 System Menu .................................................................................................32
5.2.4 INPUT SETUP (-R model only) ......................................................................38
6 Digital communication interface...........................................43
6.1 Wiring ...................................................................................................... 43
6.1.1 Setup .............................................................................................................43
6.1.2 Serial Operation ............................................................................................43
6.2 Command Syntax ................................................................................... 44
6.3 Commands by Function or Group .......................................................... 45
6.4 Serial Commands - Alphabetic Listing ................................................... 48
6.5 Non-SCPI Process Commands ............................................................... 66
6.6 Non-SCPI Commands by Function or Group ......................................... 66
7 Troubleshooting .....................................................................69
8 Maintenance ...........................................................................71
8.1 Field Temperature Block Performance Analysis ..................................... 71
v
Tables
Table 1 Symbols used ........................................................................................ 3
Table 2 Base Unit Specifications ..................................................................... 11
Table 3 -R Option Specifications .................................................................... 12
Table 4 Matching Certificate Values to ITS-90 Coefficients ............................. 39
Table 5 Setting Coefficients Rtpw, a8, b8, and b4 ........................................... 39
Table 6 Commands by Function or Group ....................................................... 45
Table 7 PROG:SEQ:PAR parameters ............................................................... 55
Table 8 Troubleshooting, problems, causes and solutions .............................. 69
vi
Figures
Figure 1 Clamp-on ferrite installation ................................................................ 8
Figure 2 FBXXX Field Temperature Block ........................................................ 14
Figure 3 Display panel and keys ..................................................................... 16
Figure 4 FBXXX display ................................................................................... 17
Figure 5 FB150 power panel ........................................................................... 19
Figure 6 FB350 and FB660 power panel ......................................................... 19
Figure 7 -R option panel ................................................................................. 20
Figure 8 Probe connector wiring ..................................................................... 21
Figure 9 Steps to language selection .............................................................. 22
Figure 10 Main Menu - Temp SetUp ................................................................ 23
Figure 11 Main Menu - Prog Menu .................................................................. 24
Figure 12 Main Menu - System Menu .............................................................. 25
Figure 13 Main Menu - Input Setup ................................................................ 26
Figure 14 RS-232 wiring .................................................................................. 44
1
Before You Start
Introduction
1 Before You Start
1.1 Introduction
Field Temperature Blocks (FB150, FB350, and FB660) are designed to be reliable, stable heat sources that can be used in the eld or laboratory. They offer accuracy, portability, and speed for nearly every eld calibration application. The instruments have been designed with the eld user in mind and are easy to use while maintaining stability, uniformity, and accuracy comparable to some laboratory instruments.
Special built-in features make Field Temperature Blocks extremely adaptable. The exclusive Voltage Compensation allows the technician to plug into mains power with voltage from 90 V ac to 250 V ac without degradation to the instrument. The Ambient Temperature Compensation provides the largest operating range in the industry (0°C to 50°C) with the largest guaranteed temperature range (13°C to 33°C). The Gradi­ent Temperature Compensation keeps the axial gradient within specication over the entire temperature range of the instrument and over the specied guaranteed operat­ing temperature range. These combined features along with the rugged design, light weight, and small size make this line of instruments ideal for eld applications.
Unique Patent Pending safety features make these the safest eld heat sources avail­able. The unique Air Flow Design keeps the probe handle cool protecting delicate instruments and the user. The Block Temperature Indicator shows the user when the well temperature is above 50°C letting the user know when it is safe to remove the insert or move the instrument. The indicator light illuminates when the instrument is energized and the well is above 50°C. If the instrument is removed from mains power, the indicator light ashes until the well has cooled to less than 50°C.
The optional “-R” version (“FBXXX-R”) combines the heat source with a built-in reference.
The Field Temperature Blocks’ controller uses a PRT sensor and thermoelectric mod­ules or heaters to achieve stable, uniform temperatures throughout the block.
The LCD display continuously shows many useful operating parameters including the block temperature, the current set-point, block stability, and heating and cooling status. For the -R version, the reference temperature readings are displayed. The display can be set to show the information in one of eight different languages; English, Japanese, Chinese, German, Spanish, French, Russian, and Italian.
The instrument’s rugged design and special features make them ideal for the eld or the laboratory. With proper use, the instrument provides continued accurate calibration of temperature sensors and devices. Before use, the user should be familiar with the warnings, cautions, and operating procedures of the block as described in the User’s Guide.
FBXXX Field Temperature Blocks
Unpacking
2
1.2 Unpacking
Unpack the instrument carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
FB150
FB150 Field Temperature Block
Insert: FB150-MH2, FB150-MH1, or FB150-B
Power Cord
RS-232 Cable
User Guide
Calibration Certicate and calibration label
LEMO Connector (-R model only)
Well Insulator
Clamp-on ferrites (3) [-R model only]
Tongs (insert removal tool)
FB350
FB350 Field Temperature Block
Insert: FB350-MH2, FB350-MH1, or FB350-B
Power Cord
RS-232 Cable
User Guide
Calibration Certicate and calibration label
LEMO Connector (-R model only)
Clamp-on ferrites (3) (-R model only)
Tongs (insert removal tool)
FB660
FB660 Field Temperature Block
Insert: FB660-MH2, FB660-MH1, or FB660-B
Power Cord
RS-232 Cable
User Guide
Calibration Certicate and calibration label
LEMO Connector (-R model only)
Clamp-on ferrites (3) [-R model only]
Tongs (insert removal tool)
3
Before You Start
Symbols Used
If all items are not present, contact an Authorized Service Center (see Section 1.6Au­thorized Service Centers on page 9).
1.3 Symbols Used
Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this guide.
Table 1 Symbols used
Symbol Description
AC (Alternating Current)
AC-DC
Battery
Complies with European Union directives
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
FBXXX Field Temperature Blocks
Safety Information
4
Symbol Description
Read the User’s Guide (Important Information)
Off
On
Canadian Standards Association
C-TICK Australian EMC mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/ EC) mark.
1.4 Safety Information
Field Temperature Blocks are designed in accordance with IEC 61010-1, IEC 61010­2-010 and CAN/CSA 22.2 No 61010.1-04. Use this instrument only as specied in this manual. Otherwise, the protection provided by the instrument may be impaired. Refer to the safety information in the Warnings and Cautions sections below.
The following denitions apply to the terms “Warning” and “Caution”.
“Warning” identies conditions and actions that may pose hazards to the user.
“Caution” identies conditions and actions that may damage the instrument being used.
1.4.1 Warnings
To avoid personal injury, follow these guidelines.
GENERAL
DO NOT use this instrument in environments other than those listed in the User’s
Guide.
Inspect the instrument for damage before each use. Inspect the case. Look for cracks or missing plastic. DO NOT use the instrument if it appears damaged or operates abnormally.
5
Before You Start
Safety Information
Follow all safety guidelines listed in the User’s Guide.
Calibration equipment should only be used by trained personnel.
If this equipment is used in a manner not specied by the manufacturer, the protection provided by the equipment may be impaired.
Before initial use, or after transport, or after storage in humid or semi-humid environ­ments, or anytime the instrument has not been energized for more than 10 days, the instrument needs to be energized for a “dry-out” period of 2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-2. If the product is wet or has been in a wet environment, take necessary measures to remove moisture prior to applying power such as storage in a low humidity temperature chamber operating at 50°C for 4 hours or more.
DO NOT use this instrument for any application other than calibration work. The instrument was designed for temperature calibration. Any other use of the instrument may cause unknown hazards to the user.
DO NOT place the instrument under a cabinet or other structure. Overhead clearance is required. Always leave enough clearance to allow for safe and easy insertion and removal of probes.
Use of this instrument at HIGH TEMPERATURES for extended periods of time requires caution.
Completely unattended high temperature operation is not recommended due to safety hazards that can arise.
This instrument is intended for indoor use only.
Follow all safety procedures for the test and calibration equipment you use.
Do not use the instrument if it operates abnormally. Protection may be impaired. When in doubt, have the instrument serviced.
DO NOT operate the Field Temperature Block around explosive gas, vapor, or dust.
DO NOT operate instrument at orientations other than upright. Tilting the instrument
or laying it down on its side during use could create a re hazard.
BURN HAZARD
The instrument is equipped with a Block Temperature Indicator (front panel LED HOT indicator) even when the instrument is unplugged. When the indicator is ash­ing, the instrument is disconnected from mains power and the temperature of the block is above 50°C. When the indicator is illuminated, always on, the instrument is pow­ered and the block temperature is above 50°C.
DO NOT turn the instrument upside down with the inserts in place; the inserts will fall out.
DO NOT operate near ammable materials.
FBXXX Field Temperature Blocks
Safety Information
6
Use of this instrument at HIGH TEMPERATURES for extended periods of time requires caution.
DO NOT touch the well access surface of the instrument.
The block vent may be very hot due to the fan blowing across the heater block of the instrument.
The temperature of the well access is the same as the actual display temperature, e.g. if the instrument is set to 600°C and the display reads 600°C, the well is at 600°C.
Probes and inserts may be hot and should only be inserted and removed from the instrument when the instrument indicates temperatures less than 50°C.
DO NOT turn off the instrument at temperatures higher than 100°C. This could create a hazardous situation. Select a set-point less than 100°C and allow the instrument to cool before turning it off.
The high temperatures present in Field Temperature Blocks designed for operation at 300°C and higher may result in res and severe burns if safety precautions are not observed.
ELECTRICAL HAZARD
These guidelines must be followed to ensure that the safety mechanisms in this instru­ment operate properly. This instrument must be plugged into an AC only electric outlet according to Table 2, Specications . The power cord of the instrument is equipped with a grounding plug for your protection against electrical shock hazards. It must be plugged directly into a properly grounded receptacle. The receptacle must be installed in accordance with local codes and ordinances. Consult a qualied electrician. DO NOT use an extension cord or adapter plug.
If supplied with user accessible fuses, always replace the fuse with one of the same rating, voltage, and type.
Always replace the power cord with an approved cord of the correct rating and type.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE INJURY or DEATH may result if personnel fail to observe safety precautions. Before working inside the equipment, turn power off and disconnect power cord.
1.4.2 Cautions
To avoid possible damage to the instrument, follow these guidelines:
DO NOT leave the inserts in the instrument for prolonged periods. Due to the high operating temperatures of the instrument, the inserts should be removed after each use and buffed with a Scotch-Brite® pad or emery cloth (see Section 8Maintenance on page 71).
Always operate this instrument at room temperature between 5°C and 50°C (41°F to 122°F). Allow sufcient air circulation by leaving at least 15 cm (6 in) of clearance
7
Before You Start
CE Comments
around the instrument. Overhead clearance of 1 meter (3 ft) is required. DO NOT place instrument under any structure.
Component lifetime can be shortened by continuous high temperature operation.
DO NOT use uids to clean out the well. Fluids could leak into electronics and dam­age the instrument.
Never introduce any foreign material into the probe hole of the insert. Fluids, etc. can leak into the instrument causing damage.
Unless recalibrating the instrument DO NOT change the values of the calibration con­stants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the block.
DO NOT allow the probe sheath or inserts to drop into the well. This type of action can cause a shock to the sensor and affect the calibration.
The instrument and any thermometer probes used with it are sensitive instruments that can be easily damaged. Always handle these devices with care. DO NOT allow them to be dropped, struck, stressed, or overheated.
DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty environ­ment. Always keep the well and inserts clean and clear of foreign material.
The Field Temperature Block is a precision instrument. Although it has been designed for optimum durability and trouble free operation, it must be handled with care. Al-
ways carry the instrument in an upright position to prevent the inserts from dropping out. The convenient handle allows for hand carrying the instrument.
If a mains supply power uctuation occurs, immediately turn off the instrument. Power bumps from brown-outs could damage the instrument. Wait until the power has stabilized before re-energizing the instrument.
The probe and the block may expand at different rates. Allow for probe expansion inside the well as the block heats. Otherwise, the probe may become stuck in the well.
Most probes have handle temperature limits. If the probe handle limits are exceeded, the probe may be permanently damaged. Due to a unique Air Flow Design, Field Tem-
perature Blocks protect the probe handle temperature and provide a safer temperature handle for the user.
1.5 CE Comments
1.5.1 EMC Directive
Beamex equipment has been tested to meet the European Electromagnetic Compatibil­ity Directive (EMC Directive, 89/336/EEC). The Declaration of Conformity for your instrument lists the specic standards to which the instrument was tested.
FBXXX Field Temperature Blocks
CE Comments
8
The instrument was designed specically as a test and measuring device. Compliance to the EMC directive is through IEC 61326-1 Electrical equipment for measurement, control and laboratory use.
As noted in the IEC 61326-1, the instrument can have varying congurations. The instrument was tested in a typical conguration with shielded RS-232 cables.
1.5.2 Immunity Testing
Using Clamp-On Ferrites
For the –R model only, clamp-on ferrites are provided for use in improving its electro­magnetic (EM) immunity in environments of excessive EM interference. During EMC testing we found that ferrites clamped around probe cables for the Reference PRT in­put reduced the risk the EM interference affects measurements. Therefore, we recom­mend that the clamp-on ferrites provided be used on the cables of probes attached to the readout, especially if the product is used near sources of EM interference such as heavy industrial equipment.
To attach a ferrite to a probe cable, make a loop in the cable near the connector and clamp the ferrite around half of the loop as shown in the diagram. The ferrite can be easily snapped open and moved to a new probe when needed.
Figure 1
clamp-on ferrite
connector
probe
Clamp-on ferrite installation
1.5.3 Emission Testing
The instrument fullls the limit requirements for Class A equipment. The instrument was not designed to be used in domestic establishments.
1.5.4 Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (2006/95/EC), Bea­mex equipment has been designed to meet the EN 61010-1 and the EN 61010-2-010 standards.
9
Before You Start
Authorized Service Centers
1.6 Authorized Service Centers
Please contact the following Authorized Service Center to coordinate service on your Beamex product:
Beamex Oy Ab
Ristisuonraitti 10 FI-68600 Pietarsaari Finland
When contacting a Service Centers for support, please have the following information available:
Model Number
Serial Number
Voltage
Complete description of the problem
11
Specications and Environmental Conditions
Specications
2 Specications and Environmental Conditions
2.1 Specications
Table 2 Base Unit Specifications
Base Unit Specifications
FB150 FB350 FB660
Temperature Range at 23°C
–25°C to 150°C
(-13°F to 302°F)
33°C to 350°C
(91°F to 662°F)
50°C to 660°C
(122°F to 1220°F)
Display Accuracy ±0.2°C Full Range ±0.2°C Full Range ±0.35°C at 50°C
±0.35°C at 420°C
±0.5°C at 660°C
Stability ±0.01°C Full Range ±0.02 °C at 33°C
±0.02 °C at 200°C ±0.03°C at 350°C
±0.03°C at 50°C ±0.05°C at 420°C ±0.05°C at 660°C
Axial Uniformity at 40 mm (1.6 in)
±0.05°C Full Range ±0.04°C at 33°C
±0.1°C at 200°C ±0.2°C at 350°C
±0.05°C at 50°C ±0.35°C at 420°C
±0.5°C at 660°C
Axial Uniformity at 60 mm (2.4 in)
±0.07°C Full Range ±0.04°C at 33°C
±0.2°C at 200°C
±0.25°C at 350°C
±0.1°C at 50°C ±0.6°C at 420°C ±0.8°C at 660°C
Radial Uniformity ±0.01°C Full Range ±0.01°C at 33°C
±0.015°C at 200°C
±0.02°C at 350°C
±0.02°C at 50°C
±0.05°C at 420°C
±0.1°C at 660°C
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C Full Range ±0.015°C Full Range ±0.015°C at 50°C
±0.025°C at 420°C ±0.035°C at 660°C
Loading Effect (versus display with 6.35 mm probes)
±0.08°C Full Range ±0.2°C Full Range ±0.1°C at 50°C
±0.2°C at 420°C ±0.2°C at 660°C
Hysteresis 0.025°C 0.06°C 0.2°C
Operating Conditions 0°C to 50°C, 0% to 90% RH (non-condensing)
Environmental conditions for all specications except temperature range
13°C to 33°C
Immersion (Well) Depth 150 mm (5.9 in)
Insert OD 30 mm (1.18 in) 25.3 mm (1.00 in) 24.4 mm (0.96 in)
Heating Time 16 min: 23°C to 140°C
23 min: 23°C to 150°C
25 min: –25°C to 150°C
5 min: 33°C to 350°C 15 min: 50°C to 660°C
Cooling Time 15 min: 23°C to –25°C
25 min: 150°C to –23°C
32 min: 350°C to 33°C
14 min: 350°C to 100°C
35 min: 660°C to 50°C
25 min: 660°C to 100°C
Resolution 0.01°
Display LCD,°C or°F user-selectable
Key Pad Arrows, Menu, Enter, Exit, 4 soft keys
Size (H x W x D) 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)
Environmental Conditions
12
Base Unit Specifications
FB150 FB350 FB660
Weight 8.16 kg (18 lbs) 7.3 kg (16 lbs) 7.7 kg (17 lbs)
Power Requirements 100 V to 115 V (±10%)
50/60 Hz, 575 W
230 V (±10%) 50/60 Hz,
575 W
100 V to 115 V (±10%), 50/60 Hz, 1400 W
230 V (±10%), 50/60 Hz, 1800 W
System Fuse Ratings 115 V: 6.3 A T 250 V
230 V: 3.15 A T 250 V
15 A, 250 V Thermal Circuit Breakers
Computer Interface RS-232
Safety EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04
Table 3 -R Option Specifications
-R Specifications
Built-in Reference Thermometer Readout Accuracy
(4-Wire Reference Probe)
± 0.013 °C at -25 °C ± 0.015 °C at 0 °C ± 0.020 °C at 50 °C ± 0.025 °C at 150 °C ± 0.030 °C at 200 °C ± 0.040 °C at 350 °C ± 0.050 °C at 420 °C ± 0.070 °C at 660 °C
Reference Resistance Range 0 ohms to 400 ohms
Reference Resistance Accuracy‡ 0 ohms to 42 ohms: ±0.0025 ohms
42 ohms to 400 ohms: ±60 ppm of reading
Reference Characterizations ITS-90, CVD, IEC-60751, Resistance
Reference Measurement Capability 4-wire
The temperature range may be limited by the reference probe connected to the readout. The Built­In Reference Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors.
Measurement accuracy specications apply within the operating range and assume 4-wires for PRTs.
2.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance section. The instrument operates safely under the fol­lowing environmental conditions:
ambient temperature range: 0-50°C (32-122°F)
ambient relative humidity: 0 % to 90 % (non-condensing)
mains voltage: within ±10% of nominal
vibrations in the calibration environment should be minimized
altitude: less than 2,000 meters
indoor use only
13
Quick Start
Setup
3 Quick Start
3.1 Setup
Note: The instrument will not heat, cool, or control until the “SET PT.” parameter is “Enabled”.
Place the block on a at surface with at least 15 cm (6 in) of free space around the in­strument. Overhead clearance is required. DO NOT place under a cabinet or structure.
Plug the instrument power cord into a mains outlet of the proper voltage, frequency, and current capability (see Section 2.1, Specications, on page 11 for power details). Observe that the nominal voltage corresponds to that indicated on the front of the block.
Carefully place the insert into the well. Inserts should be of the smallest hole diam­eter possible still allowing the probe to slide in and out easily. Various insert sizes are available. Contact an Authorized Service Center for assistance (see Section 1.6, Authorized Service Centers, on page 9). The well must be clear of any foreign objects, dirt and grit before an insert is installed. The insert is installed with the two small tong holes positioned upward.
Turn on the power to the block by toggling the switch on the power entry module. After a brief self-test, the controller should begin normal operation. The main screen appears within 30 seconds. If the instrument fails to operate, please check the power connection. The display shows the well temperature, and waits for user input before further operation.
Press “SET PT.” and use the arrow keys to set the desired set-point temperature. Press “ENTER” to save the desired set-point and enable the instrument. After ve (5) sec­onds, the instrument should start to operate normally and heat or cool to the designated set-point.
Parts and Controls
14
Figure 2 FBXXX Field Temperature Block
3.2 Parts and Controls
This section describes the exterior features of the Field Temperature Block. All inter­face and power connections are found on the front of the instrument (see Figure 2).
15
Quick Start
Parts and Controls
3.2.1 Display Panel
Figure 3 on next page shows the layout of the display panel.
Display (1)
The display is a 240 x 160 pixel monochrome graphics LCD device with a bright LED backlight. The display is used to show current control temperature, measurements, status information, operating parameters, and soft key functions.
udlr Arrow Keys (2)
The Arrow Keys allow you to move the cursor on the display, change the display lay­out, and adjust the contrast of the display. The contrast can only be adjusted using the u and d arrow keys while viewing the main display window.
Enter Key (3)
The Enter Key allows you to select menus and accept new values.
SET PT. (4)
The Set Pt. Key allows you to enable the instrument to heat or cool to a desired set­point. Until this key is enabled, the instrument will not heat or cool. It is in a “sleep” state for safety of the operator and instrument.
°C/°F Key (5)
The °C/°F Key allows you to change the displayed temperature units from °C to °F and vice versa.
Menu Key (6)
The Menu Key allows the user to access all parameter and settings menus. From the main menu, the user can use the soft keys to access submenus and functions.
Exit Key (7)
The Exit Key allows you to exit menus and cancel newly entered values.
Soft Keys (8)
The Soft Keys are the four buttons immediately below the display (labeled F1 to F4). The functions of the soft keys are indicated on the display above the buttons. The function of the keys may change depending on the menu or function that is selected.
Parts and Controls
16
Block Temperature Indicator (9)
The Block Temperature Indicator lamp allows users to know when the block tempera­ture is safe (50°C to 60°C) to remove inserts or move the Field Temperature Block. The indicator light is lit continuously once the block has exceeded approximately 50°C (varies 50°C to 60°C). The indicator light stays lit until the block cools to less than approximately 50°C. If the instrument is disconnected from mains power, the indicator light ashes until the block temperature is less than approximately 50°C.
Figure 3 Display panel and keys
3.2.2 Display
The front panel display is shown in detail in Figure 4 on opposite page.
Heat Source Temperature (1)
The most recent block temperature measurement is shown in large digits in the box at the top of the screen.
Set-point Temperature (2)
The current set-point temperature is displayed just below the Process Temperature.
Reference Thermometer Temperature (3) [-R models only]
When installed, the most recent reference thermometer measurement is shown on the screen.
17
Quick Start
Parts and Controls
Stability Status (4)
On the right hand side of the screen, you will nd a graph displaying the current status of the stability of the Field Temperature Block.
Heating/Cooling Status (5)
Just below the stability graph there is a bar graph that will indicate HEATING, COOL­ING, or CUTOUT. This status graph indicates the current level of heating or cooling if the instrument is not in cutout mode.
Soft Key Functions (6)
The four texts at the bottom of the display (not shown) indicate the functions of the soft keys (F1–F4). These functions change with each menu.
Editing Windows
While setting up and operating the instrument, you are often required to enter or select parameters. Editing windows appear on the screen when necessary to show the values of parameters and allow edits.
Figure 4 FBXXX display
3.2.3 Power Panel
The following are found on the lower front panel of the instrument (see Figures 5 and Figure 6 on page 19).
Parts and Controls
18
Power Cord Plug (1)
The power supply cord attaches to the lower front power panel. Plug the cord into an AC mains supply appropriate for the voltage range as specied in the specications tables.
Power Switch (2)
For the FB150, the power switch is located on the power entry module of the unit at the lower center of the power panel.
For the FB350 and FB660, the power switch is located between the RS-232 and the circuit breakers.
Serial Connector (3)
On the FB150, the serial connector is a 9-pin subminiature D type located on the power panel above the power entry module. On the FB350 and FB660, the serial connector is a 9-pin subminiature D type located on the power panel to the left of the power switch. The serial (RS-232) interface can be used to transmit measurements and control the operation of the instrument.
Fuses
For the FB150, the fuses are located inside the power entry module of the unit (Figure 5 on opposite page).
If necessary, fuses must be replaced according to Specications (see Section 2.1, Specications, on page 11.
Thermal Circuit Breakers (5)
For the FB350 and FB660, the thermal circuit breakers are separate from the power connector (Figure 6 on opposite page). Circuit breakers can be reset by pressing the button in the center of each circuit breaker.
Loading...
+ 54 hidden pages