BD Vacutainer Pronto Quick Reference Card

Page 1
BD Vacutainer
®
Pronto
™
Quick Release Needle Holder
QUICK REFERENCE CARD
• Ensure BD Pronto™ Holder
mechanism is ready to accept new collection device
• If not, press white reset
button as shown in step 6
1 2
• While pushing into BD
Pronto™ Holder, secure collection device to holder by turning hands in opposite directions 1/4 turn until needle is fully sealed onto holder
• DO NOT OVERTORQUE
• As indicated for particular
collection device, remove coloured (patient-end) needle cap
• Perform venipuncture
according to your facility's established procedure. As indicated for particular collection device, engage safety mechanism
4
3
RESETREADY
• As indicated for particular
collection device, remove white (non-patient-end) cap
• BD Pronto™ Holder is
compatible with all BD Vacutainer® blood collection devices
Page 2
CAUTION:
• Handle all biologic samples and blood collection sharps
(lancets, needles, luer adapters and blood collection sets) according to the policies and procedures of your facility
• Obtain appropriate medical attention in the event of any
exposure to biologic samples (for example, through a puncture injury) since they may transmit viral hepatitis, HIV (AIDS) or other infectious diseases
• Utilize any safety-engineered feature if the blood collection
device provides one
• Discard all blood collection "sharps" in biohazard containers
approved for their disposal.
• As verified by testing, the possible number of uses per BD
Pronto™ Quick Release Needle Holder is 100. As per CLSI guidelines, the holder may be cleaned periodically using a 10% bleach solution (NCCLS M29-A2, Vol. 21, No. 23). Autoclaving or the use of 100% bleach solution will impact performance of these products.
• Dispose of collection device
into nearest approved sharps container:
- Hold device in a downward position over container opening, push green button on holder to release collection device into container
- Ensure white reset button is not obstructed by other fingers
• To reuse BD Pronto™
Holder, push white reset button
BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. ©2007 BD. Franklin Lakes, NJ 07417 1.800.268.5430 www.bd.com/ca Printed in Canada 09/07 BDPAS0077-CE