Use this quick reference guide once you are familiar with the procedures in
the BD FACSAria II User’s Guide.
Starting Up the Cytometer
Follow these steps to start up your BD FACSAria™ II cytometer. For more
details, refer to Chapter 4 in the user’s guide.
1 Start up the computer.
2 Turn on the cytometer main power and the lasers. Wait 30 minutes for
the lasers to warm up.
main power
;Tip If you are using temperature
control, start heating or cooling the
sample injection chamber or the
recirculating water bath while the
lasers are warming up.
3 Start BD FACSDiva™ software and log in with your user
name and password.
4 Verify fluidics levels in the Cytometer window. Replenish
fluids or empty the waste, if needed.
sheath wastebleachDIethanol
5 Select Cytometer > Fluidics Startup.
6 Follow the instructions in the Fluidics Startup wizard to complete the
process.
Verify that the air and fluid lines are connected to the sheath tank, then
click Done.
1
Page 2
Caution: Do not run fluidics startup with the air and fluid lines
connected to the ethanol shutdown tank. This can cause damage to
the system. Always verify the lines are connected to the sheath tank.
7 Start the stream.
a Click the Sorting button () on the Workspace toolbar to display
the Breakoff and Side Stream windows.
b Click the Stream button () in the Breakoff window to turn on the
stream.
8 Check the stream position in the waste aspirator and inspect the stream
image in the Breakoff window.
If you see any dripping or spraying or the stream appears abnormal,
turn off the stream and refer to Troubleshooting the Stream in
Chapter 7 in the user’s guide.
9 Select a new Sort Setup mode, if necessary.
Sort SetupNozzle (micron)Default pressure (psi)
70 micron7070
85 micron8545
100 micron10020
130 micron13010
10 Adjust the Amplitude to achieve fast-merging satellites within 6 drops.
For troubleshooting assistance, see Troubleshooting the Breakoff on
page 5 or refer to Sorting Troubleshooting in Chapter 7 in the user’s
guide.
11 Enter the generated value as the Drop 1 target.
12 Turn on the Sweet Spot when the drop pattern is stable.
2
Page 3
Sorting Setup
Do the following to set up for sorting. For more details, refer to Chapter 5
in the user’s guide.
1 Check the laser delay and area scaling factors for your sheath pressure.
Refer to Checking Performance with Cytometer Setup and Tracking in
Chapter 4 in the user’s guide or perform quality control as described in
Appendix E.
2 Optimize cytometer settings for the sample to be sorted.
Refer to Data Collection in Chapter 4 in the user’s guide.
3 Install the required collection device and set up the side streams.
4 Use BD FACS™ Accudrop to calculate the drop delay.
5 Run the sample to be sorted. Use gating tools and subsetting methods
to define the population(s) of interest.
NOTICE Snap-to gates cannot be used for sort gates.
6 Define a sort layout for the tube containing the defined sort
populations.
Sorting
1 Install the collection tubes, plate, or slide.
2 Install the sample tube onto the loading port and click Load.
3 Verify that the current tube pointer is set to the appropriate tube in the
Browser and click Sort.
4 Click OK if you are prompted to open the aspirator drawer or turn on
the deflection plates.
3
Page 4
Sorting continues until the required number of cells has been sorted. If
the number of Target Events is set to Continuous, sorting continues
until you manually stop sorting by clicking the Stop Acquiring button
or the Sort button.
; Tip Click Record Data to save data for the tube. Acquisition and
sorting continue when the required number of events has been recorded.
Shutting Down the Cytometer
1 Turn off the stream.
2 Remove the nozzle and install the closed-loop nozzle.