BC Biomedical ULT-2000 User Manual

ULTRASOUND TRANSDUCER
LEAKAGE TESTER
ULT-2000 SERIES
USER MANUAL
BC BIOMEDICAL
ULT-2000 SERIES
TABLE OF CONTENTS
WARNINGS ................................................................................................................ 2
DESCRIPTION ............................................................................................................ 6
OVERVIEW.................................................................................................................. 9
KEYS............................................................................................................................ 10
SCREENS ................................................................................................................... 12
SETUP ......................................................................................................................... 17
OPERATIONS ............................................................................................................. 21
MANUAL REVISIONS.................................................................................................. 26
WARRANTY................................................................................................................. 26
SPECIFICATIONS ....................................................................................................... 27
NOTES......................................................................................................................... 28
1
WARNING - USERS
The ULT-2000 Testers are for use by
skilled technical personnel only.
WARNING - USE
The ULT-2000 Testers are intended for testing
only and they should never be used in
diagnostics, treatment or any other capacity
where they would come in contact with a patient.
WARNING - MODIFICATIONS
The ULT-2000 Testers are intended for use within
the published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Tester. A serious hazard may occur
if the patient is connected when
testing with the Tester.
Do not connect any leads from the patient
directly to the Tester or DUT.
WARNING – POWER ADAPTOR
Unplug the Power Adaptor before
cleaning the surface of the Tester.
WARNING - LIQUIDS
Do not submerge or spill liquids on the Tester.
Do not operate the Tester if internal components
not intended for use with fluids may have been
exposed to fluid, as the internal leakage may
have caused corrosion and be a potential hazard.
2
WARNING
High Voltages are generated when running
electrical leakage tests. Do not touch any surface that is in contact with the Tester, including the ultrasound transducer
and the adapter.
Do not touch the liquid in the test basin, as it is
at an electrical potential that is equal to the
“challenge” voltage being used to perform the
electrical safety test.
CAUTION - SERVICE
The ULT-2000 Testers are intended to be
serviced only by authorized service personnel.
Troubleshooting and service procedures
should only be performed by
qualified technical personnel.
CAUTION - ENVIRONMENT
The ULT-2000 Testers are intended to function
between 15 and 30 °C.
Exposure to temperatures outside this range can
adversely affect the performance of the Tester.
CAUTION - CLEANING
Do not immerse. The Tester should be cleaned
by wiping gently with a damp, lint-free cloth.
A mild detergent can be used if desired.
CAUTION - INSPECTION
The ULT-2000 Testers should be
inspected before each use for wear and
the Tester should be serviced
if any parts are in question.
3
NOTICE – SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)
Electrical Caution
(Consult Manual for Further Information)
Center Negative
Direct Current
NOTICE – ABBREVIATIONS
Amps Ampheres
c centi- (10
C Celsius
CF Crest Factor
° degree
DFA Digital Fast Acquisition
DUT Device Under Test
hrs hours
Hz hertz k kilo- (10
kg kilograms kHz kilohertz
lbs pounds
L1, L2, L3 Location 1, 2, 3
M Mega- (10
MHz Megahertz
µ micro- (10
µA microampere
m milli- (10
mA milliampere
mHz millihertz
mm millimeter
ms millisecond
mV millivolts
ohm
PC Personal Computer
Pk peak
RF Radio Frequency
RMS Root Mean Square
US United States
V volt
3
)
-3
-2
)
6
)
-6
)
)
4
USER ASSUMES FULL RESPONISIBILITY FOR UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS. SUCH MODIFICATIONS OR APPLICATIONS MAY RESULT IN EQUIPMENT DAMAGE OR PERSONAL INJURY.
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
NOTICE – CONTACT INFORMATION
NOTICE – DISCLAIMER
NOTICE – DISCLAIMER
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
Manual ULT-2000 Series Copyright © 2007
www.bcgroupintl.com Made in the USA
11/07 Rev 04
5
BC BIOMEDICAL
ULT-2000 SERIES
ULTRASOUND TRANSDUCER LEAKAGE TESTER
The Model ULT-2000 Series is a family of Microprocessor based, Ultrasound Transducer Leakage Testers. The ULT-2010 measures both the conductivity of the cleaning medium and the leakage current of the ultrasound transducer. The ULT-2020 offers a Meter mode and a Datalog.
The following are highlights of some of the main features:
ULT-2010 (BASIC FEATURES):
LARGE GRAPHICS DISPLAY WITH CURSOR SELECTION OF OPTIONS AND SETUP OF PARAMETERS
USER-SELECTABLE CHALLENGE (TEST) VOLTAGE (90 TO 275 VAC)
USER SELECTABLE CHALLENGE (TEST) FREQUENCY (50 OR 60 HZ)
USER-SELECTABLE TEST LIMITS BY ULTRASOUND TRANSDUCER MANUFACTURER AND MODEL
1% FS MEASUREMENT ACCURACY
AUTO RANGING WITH 10, 250, OR 500 µA FS
AUTOMATIC INTERNAL SELF TEST
SINGLE BUTTON PRESS FOR FULL SYSTEM TEST
SIMPLE GO-NOGO MODE – FOR NON-TECHNICAL USERS
ANALYTICAL MODE – FOR TECHNICAL USERS
DIGITAL CALIBRATION – NO POTS TO TURN
USER-SELECTABLE DISPLAY OPTIONS
BATTERY LIFE DISPLAY (0 to 100%)
PROGRAMMABLE BACKLIGHT TIMER
CONTRAST IS SOFTWARE ADJUSTABLE
FLASH UPGRADEABLE FIRMWARE
RS232 & PC SOFTWARE
SERIAL PRINTER OUTPUT WITH PROGRAMMABLE USER HEADER
REAL TIME CLOCK
COMPATIBLE WITH DALE
BIOMEDICAL ULT-800® ULTRASOUND TRANSDUCER ADAPTERS AND DUAL CONDUCTIVITY PROBES
®
TECHNOLOGY DALE800® AND FLUKE®
6
ULT-2020 (METER, DATALOG)
HAS ALL THE BASIC MODEL FEATURES PLUS:
METER MODE FOR EXTENDED MEASUREMENT PERIODS
PROGRAMMABLE METER CHALLENGE VOLTAGE AND FREQUENCY
PROGRAMMABLE METER TIMER
DATALOG WITH STORAGE OF 99 TEST RECORDS
OPTIONAL ACCESSORIES:
BC20 - 21103 BATTERY ELIMINATOR (US Version) BC20 - 21101 BATTERY ELIMINATOR (Euro Version) BC20 - 41337 RS232 COMMUNICATIONS CABLE (7 Pin Mini-Din to DB 9 F)
BC20 - 41339 COMMUNICATION CABLE ADAPTER (USB to DB9 M) (For use with BC20-41337) BC20 - 30106 BC BIOMEDICAL SMALL SOFT SIDED CARRYING CASE
ULT-PC-10 DUAL CONDUCTIVITY PROBE (Short)
ULT-PC-20 DUAL CONDUCTIVITY PROBE (Long)
BC20 - 42200 CIDEX® COMPATIBILITY TRAY
BC20 - 42300 EXTERNAL PRINTER
BC20 - 42310 PACKAGE OF 5 ROLLS OF PAPER FOR BC20 – 42300
7