OPERATION MANUAL
Humidity- temperature probe
Humidity- temperature probe
with USB-Interface
with USB-Interface
DescriptionDescription
Characteristics features
• Combined Temperature and Humidity measurement
• Sensor in Stainless steel casing with Sinterlter
• Resolution 0.01 % RH, 0.01°C
• Accuracy 2 % RH, 0.3°C
• Inclusive of Windows-Software “Recorder”
• Calibration possibility with salt reference cells
Areas of application
• Monitoring of stock rooms, in quality assurance or air conditioning
systems
Technical data
Humidity measurement
Humidity measuring range 0 ... 100 % RH
Humidity resolution 0.01 % RH
Typical accuracy ±2 % RH (at 23 °C)
Temperature measurement
Temp. measuring range –40 ... 80 °C
Temperature resolution 0.01 °C
Accuracy ±0.3 K between 0 and 40 °C
Module
Power supply Over the USB port
Operating current Approx. 30 mA
Interface USB-interface, 1.1 und 2.0 compatible
Probe dimensions 150 x Ø12 mm
CE-conformance 2004/108/EG
EMV-noise emission: EN 61000-6-3:2011
EMV-noise immunity: EN 61000-6-2:2011
Scope of supply Humidity probe in stainless steel
casing, interface adapter with USBconnection cable, Windows-Software
and extensive documentation
Article number HYTELOG-USB
Windows-Software
• Calculation and display of dew point, abso-lute humidity, vapour
pressure, saturated vapour pressure and enthalpy
• Tabular representation of measured values
• Storing of data on hard disk
Description
The product offers an efcient measurement and display system for
temperature and relative humidity. The compact measuring probe with
overall dimensions of Ø12 x 150 mm is housed in a stainless steel
casing. The scope of supply includes a port converter, which enables
direct operation through the USB port of a PC.
The front area of the sensor is provided with a polyethylene-sinter lter
for protection against sprinkling water and mechanical damages. An
accurate NTC has been used as temperature sensor. The humidity
measurement operates with a long term stable, capacitive polymer
sensor.The micro controller of the port converter compensates the
linearity error and temperature drift of sensor elements. The high
quality polymer sensor guarantee outstanding measuring accuracy
and long-term stability, also under extreme operating conditions.
The current measured values are transferred to the connected PC
through the USB interface. The display and graphical representation
of the meas-ured values appear on the PC. An easy to use Windows
software for display of measured values and data representation is
included in the scope of supply.
The USB driver software emulates a serial COM-port. The ASCII-protocol for data communication is documented and enables integration
with user’s own developed programs. The probe is supplied in calibrated condition. With the help of reference cells available as accessories, the accuracy of the probe can be checked by the user himself.
Further calibration and adjusting the scale is also possible.
Technical changes reserved
0141 0316-196 27.08.2015
B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
1 / 4
OPERATION MANUAL
Humidity- temperature probe
Humidity- temperature probe
with USB-Interface
with USB-Interface
WINDOWS-Software RECORDER
With the help of this program, which is covered in the scope of supply,
the measured values can be received through the USB-interface and
displayed on the PC. The displayed le is compatible with any desired
spreadsheet program, with which it is possible to further process,
statistically evaluate or visualise the measurement data.
In addition, the PC-Software also calculates dew point, absolute
humidity, enthalpy and vapour pressure from the measured values of
relative humidity and temperature. The calculated gures can also be
stored.
First time operation
Connect the humidity meas-uring system to the USB-interface of a
PC. After rst time run of the software, go to menu option „settings“
and select device type as „HYTELOG 4800Bd“ and also select the
type of interface to be used under „connections“ (Note: For USB-ver-
sion, mention the virtual COM-port specied during driver installation).
The remaining settings (Data rate, Parity, Start and Stop bit) are
automatically selected and need not be changed. If the connection is
established, the data communication appears on the terminal window.
Then select „Close“. The current settings will be stored.
If you are not able to establish data link between PC and the
measuring device or the measuring adaptor, then rst please check
the power supply and also cable connection to the PC. Further information on debugging is available under FAQ´s on the CD or at our
Homepage under SUPPORT.
Data recording
First activate all the hooked up measurement channels that are to be
recorded. In ‚Text 1‘ and ‚Text 2‘, you can enter a description, which
has to appear as heading on the top of data le. The data is recorded
in a le, which is de-clared as path in the ‚Start‘ button. The recording
begins with the ‚Start‘ button.
View of the Software “RECORDER“
System requirements
Windows 98, 2000 or XP, RS232 or USB-interface. Generally, older
PCs are also suitable.
Important hint: First connect the USB-Version to the PC after
installing the software. This simplies driver installation and enables
“Plug&Play“ feature.
Installation
A detailed installation instruction is provided on the CD, which
automatically gets started on inserting the CD (prerequisites: InternetExplorer 5.0 or higher). Follow these instructions for installation:
Manual Installation
Insert the enclosed CD into your drive and select „Run“ in the start-
menu and then browse to select the le ‚setup.exe‘ under the path
LW:\software\RECORDER\TEMPLOG\disk1. Then follow the instructions of the installation program.
EXCEL™
The created le is compatible with CSV-Format. In order to display
the measured data, you can use graphic tools, for example, the
diagram-assistant. However, other programs can also be used to
graphically represent or evaluate the measured data.
Calibration
The humidity probe is supplied in calibrated
condition. The accuracy at 23°C is of the order
of ±0.3° K and ±2 % RH. Under normal opera
ting conditions, it is not necessary to again
calibrate the probe. The cross checking of
measuring accuracy of the humidity measu
rement part can be done by end user with the
salt reference cells available as special ac
cessories.
The cross checking must be done in temperature stable environment.
In case, the measured value does not match with the reference value
of the cell, the device should be sent to our calibration laboratory for
check-up.