B + B Thermo-Technik DS 1820-PUR-2M, DS 1820-PUR-5M, DS1820-PUR-10M User guide

OPERATION MANUAL
Temperature probe DS18S20 with PUR-cable and RJ12 connector
DescriptionDescription
Characteristic features
• Ready made, plug-in type temperature probe
with DALLAS semi-conductor sensor DS18S20
• High quality PUR-connection cable
• Continuous operating temperature –40 ... +125 °C
• Sensor in stainless steel protective sleeve
• Splash waterproof
• RJ12 plug connector with breakage protection
• Resolution 0.06 °C
• Accuracy ±0.5 °C nominal (from -10...+85 °C),
as per data sheet of manufacturer
Scratchpad memory for probe identication
Areas of application
• Monitoring of frozen goods as per cold storage regulations
• Building instrumentation
• Air conditioning systems
• Quality assurance
• Science, research and laboratories
• Industrial temperature logging
Technical data
Temperature probe DS18S20 with PUR-cable and RJ12 connector
Accuracy ±0,5 °C (-10...+85 °C)
Sensor DS18S20
Probe Protection sleeve, stainless steel
1.4571; Ø6xL40 mm
Cable TPE/Cu/TPE, 3x0,14 mm²
2000 mm
Connection RJ12 connector
CE-conformance 2004/108/EG
EMV-noise emission EN 61000-6-3:2011
EMV-noise withstanding EN 61000-6-1:2007
Article Art.-No.
Temperature probe DS18S20 with cable 2 m DS1820-PUR-2M
Temperature probe DS18S20 with cable 5 m DS1820-PUR-5M
Temperature probe DS18S20 with cable 10 m DS1820-PUR-10M
Technical changes reserved
0141 0316-186 21.08.2015
B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
Available probes
The reasonably priced probes with PUR-connection cable are meant for measurement in open atmosphere, on surfaces or in non-aggressi­ve gas medium. The probes are sealed and can bear water contact for some time. However, long term immersion in liquid is not recommen­ded.The models with stainless steel probe tube are chemically stable and the tube portion of probe can be submerged in liquid, the stainless steel (1.4571) does not get affected.
Temperature range
The Dallas temperature sensors are semiconductor sensors. The un-housed sensors are suitable for temperature measurement in the
range of –55 … +125 °C. These temperature values are the nal limits
and operating above these values is not at all recommended, otherwi­se the component can get damaged. In addition, the allowable application temperature also depends on the connection cable and type of protection sleeve used. The operation range of the cable is -40...+130 °C, so the probe should not be used above these values.
1 / 2
OPERATION MANUAL
Temperature probe DS18S20 with PUR-cable and RJ12 connector
Installing and conguration
The Dallas temperature sensor of type 1820 has an internal iden-
tication (serial number) and can be operated in parallel together
with several other components on a three-wire bus. After wiring all
the sensors, the PC adapter must be individually congured for the connected sensors. Operation is not possible without prior congu­ration of the system. Since the conguration is stored in the internal
EEPROM of adapter, this process is to be done only once. Only if an additional sensor is to be used on the existing network, the congura­tion needs to be repeated. The sorting of probes, found on the network, is done on the basis of binary serial number.
Pin conguration of RJ12-plug connector
The Western-plug connector is congured as follows (View on the
cable, i.e. contact surfaces of the plug!):
1 Shielding or unoccupied 2 Ground 3 Dallas Data or unoccupied 4 Dallas Data 5 +5 V 6 +5 V or unoccupied
Thermal heat transfer of sensor-ambient temperature
During surface measurements, the measuring arrangement should be thermally insulated from the surroundings, for example, with some foam material or mineral wood.
Thermal heat transfer of sensor-connecting wires
This measuring error can be minimised by itself, for example, if the connecting lead used is as thin as possible and the connecting mate­rial is a bad thermal conductor or if the connecting wire is tempered with the measuring object. In principle, of course, the highest measuring accuracy is achieved through immersion in liquids or in a mounting tube. However, an addi­tional measuring error should be included while taking measurements on surfaces.
Accessories
Accessories Articleno.
Distribution box for temperature measuring system, 10 sockets RJ12
Temperature measuring system TLOG with RS232-Interface
Temperature measuring system TLOG with USB-Interface
Humidity/temperature measu­ring system with USB-interface Hytelog Multisensor USB
VERT-GEH
0567 0002
0567 0004
0567 0001
Measuring accuracy
The sensors are calibrated during manufacture and have a typical measuring accuracy of ±0.5 °K at 23 °C application temperature. At the upper and lower limit of measuring range, the accuracy is some­what on the lower side. Further information is available in the data sheet of component at the website of manufacturer. During all temperature measurements, the physical conditions are also to be taken care of in order to avoid measuring error, which mainly decides the precision of measuring arrangement.
Thermal transition resistance of measuring object­sensor
This is the main measuring error which occurs during surface mea­surements. This can be eliminated by providing good thermal contact through mounting in a tube, applying thermal conducting paste or thermal conducting adhesive.
Technical changes reserved