PERCENTAGE DISPLAY F2.1 = 2
Display specification
Display [
Pr 20.
], representing 20.5 %.
Pr = current actual weight / range ×100%.
Press
to display switch in percentage and weight.
CHECK WEIGHT AND SELECTION SCALE FUNCTION F2.1 = 3
Function description
Set parameters such as F2.2 = A, F2.3 = B, F2.4 = C and F2.5 = D.
When display weight is X.
If X ≤ A, do not conduct check weight and selection.
If X < (B – D), it lacks of weight, and the display flickers.
If (B – D) ≤ X ≤ (B – C), it is qualified and the display has normal display
If X > (B – C), it is overweight and the display flickers.
Acquisition of target value
Press
long until the display shows [
TAR ET
], and then press
to show current target value and flicker.
If F2.6 = 0, press
, the instrument will take the weight on current scale as the new target value and quit the setting interface.
If F2.6 = 1, the display shows [
000000
], to request manual change of target value. After change, press
to save setting data and quit
setting interface.
FUNCTION OF ACCUMULATION SCALE F2.1 = 4
Operation method
In normal weighing state, when the scale is in , add weight to the scale and press
on operation panel, if the display shows [
Add--] progress bar,
it indicates that the current display weight is included accumulated value, and then it returns to normal weighing state. If the
display shows [
--no--
] for a
second, and returns to the normal weighing state, it indicates operation is invalid. Reason: 1. Between two
accumulation operations, the scale
must have back-to- process, otherwise, accumulation is refused. 2. Accumulation operation is effective
only when display
weight≥F2.2
is set. 3. The
scale is in
dynamic
state.
Adjustment, clearing and printing of accumulated value
In normal weighing state, press
on operation panel for more than 2 seconds, the display will show [
Zero
] for a second, and then the
display shows current total accumulated value [
A 9. 00
] and flickers. To clear accumulated value, press [
Pr 20.
], to make flicker weight be 0.
Press
to print accumulated data. Press
to quit the interface.
Attention: set whether to be over detailed data or accumulated data in F4.6.
FUNCTION OF COUNTING SCALE F2.1 = 5
Instrument display
[
c 128
], showing current quantity.
Sampling method
1.
Check whether the scale is in , if not, press
for setting.
2.
Place materials counted on the scale.
3.
Press
long until the display shows [
AMPLE
], and then press
. If F2.6 = 0, the display shows [
PC 00
]. Input the quantity
counted just now, and press for confirmation. The instrument saves sampling data and quits the sampling interface. If F2.6 = 1, the