Prior to servicing these pumps, read this manual and the pump’s Operator’s
Manual carefully to fully understand the pump’s functionality and to ensure
safe and proper servicing.
Warning
There are risks associated with using anything other than the recommended sets with this device. Sets
designated for use with this device are identified in Chapter 3 of the Operator’s Manual. Baxter’s warranty on
this device will be null and void and Baxter will assume no responsibility for incidents which may occur if the
product is not used in accordance with product labeling. See Chapter 1 for a complete list of warnings and
cautions.
Duplication or distribution of this manual and any information contained within is strictly prohibited without the
express written permission of Baxter. This manual and any information contained within, may not be
reproduced, distributed, or transmitted in any form, or by any means, for any purpose, without the express
written permission of Baxter. To order additional copies of this manual, or other related manuals, contact your
local Baxter Service Center.
Disclaimer
The information in this document has been carefully examined and is believed to be entirely reliable. However,
no responsibility is assumed for inaccuracies. Furthermore, Baxter reserves the right to make changes to any
products herein to improve readability, function, or design. Baxter does not assume any liability arising out of
the application or use of any product or circuit described herein; neither does it cover any license under its patent
rights nor the rights of others.
Trademark Information
Baxter, the Baxter wordmark, Buretrol, Colleague, Continu-Flo, Personality, and the shape of the On/Off clamp
are registered trademarks of Baxter International Inc. Astec is a trademark of Emerson Network Power (division
of Emerson Electric, Inc.). Condor is a trademark of Condor D.C. Power Supplies, Inc. LOCTITE and PRISM
are trademarks of Loctite Corporation. Lubriko is a trademark of Master Lubricants Company. MicroMo is a
trademark of MicroMo Electronics, Inc. Namiki is a trademark of Namiki Precision Jewel Co., Ltd. Red Glpt is
a trademark of G.C. Electronics. Whisk is a trademark of Allegiance Corporation. Other product names or
trademarks appearing in this manual are the property of their respective owners.
Year 2000 Compliance
This product is Year 2000 Compliant and will function as designed through 2035 without interruption or failure
by the occurrence of dates prior to, during or after year 2000.
Patent Information
This device is protected under one or more of the following U.S. and Foreign Patents: United States: 5,151,019;
5,764,034; 5,782,805; 5,799,207; 5,842,841; 5,843,035; 6,013,057; 6,068,612; 6,129,517; D390,654. Foreign:
Australia: 130,693; 706,742; 710,286; 712,859; 713,132; 721,076; 723,365; 727,479; 730,132; 736,366; 739,019;
740,655; 742,059; 744,398; 750,280; Benelux: 27657; Canada: Des. 80218 (BII); 2,224,239; Denmark: 165/97;
Europe: 0837708; Germany: M9608875.3; GB: 2,059,861; 2,224,239; 2,312,049; 2,312,055; 2,319,079; 2,338,753;
2,338,756; 2,338,757; 2,338,758; 2,338,992; 2,342,188; Hong Kong: HK1002291; Japan: 1002447;
Korea: 10-344380; 207012; 376076; New Zealand: 329316; 329317; 329318; 329320; 333087; 333088; 333089;
333090; 333091; 333092; 333093; Des. 28022; R.O.C.: 098653; 096216; 090525; 092501; 101,347; 102,585;
Des. 058282; Singapore: 47257; 54666; 75193; 79298; 83175; South Africa: 2000/5266; Sweden: 61479; other U.S.
and Foreign patents pending.
Table of Contents
Table of Contents
List of Illustrations.................................................................................................x
List of Tables ........................................................................................................xiv
07-19-H2-646COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL xv
List of Tables
xviCOLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07-19-H2-646
Overview
Chapter 1
Chapter 1 Introduction
This manual provides service and maintenance information for the Colleague and
Colleague 3 Volumetric Infusion Pump (hereinafter referred to as “the pump”).
The manual is intended for trained, qualified personnel and qualified Baxter
employees. Refer to the appropriate Operator’s Manual for detailed operating
instructions and applicable precautionary statements. Table 1-1 lists all Colleague
Pump product codes covered by this service manual.
This chapter contains the following information:
Product Features
■ “Product Features,” 1-1
■ “Manual Conventions,” 1-2
■ “Labeling,” 1-4
■ “Safety Summary,” 1-6
■ “Authorized Service Centers,” 1-13
The Colleague and Colleague 3 Pumps are electromechanical devices that deliver
fluid at user-selected rates through clinically acceptable routes of administration,
such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation
of fluid spaces applications. Colleague pumps are approved for use in hospital and
alternate care facilities, Colleague 3 pumps are approved for hospital use only.
07-19-H2-646COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL 1-1
1Manual Conventions
The Colleague Pump has one pump module and the Colleague 3 pump has three
pump modules, enabling it to deliver up to three independently-programmed
infusions simultaneously.
■ Fluid delivery applications include parenteral fluids, drugs, electrolytes,
total parenteral nutrition (TPN) solutions, whole blood, and blood
products.
■ Uses Baxter Standard Administration Sets equipped with Keyed On/Off
Clamps
■ Automatically loads administration set tubing
■ Can be used with a variety of source containers
■ Micro and Macro rate range
■ Piggybacking Secondary Medications with Callback Option
■ Dose programming modes
■ Configurable air sensor settings
■ User selectable downstream occlusion pressure settings (psig and mmHg)
Note:For information on conformance to the CE Mark, refer to the appropriate Opera-
Manual Conventions
■ Panel Lockout with optional Auto Lock function minimizes potential for
tampering and administration set removal
■ Configurable Personality feature sets
■ The Colleague 3 Pump has three independent pump channels that allow
up to three infusions simultaneously.
■ Label library
■ Battery charge level indicator
■ Nurse call
■ Communications port available for computer monitoring of pump
■ Built-in diagnostics
■ Optional Colleague Guardian feature allows limits to be programmed for
drugs in the label library, including custom labels.
tor’s Manual.
Pump Items
In this manual, all keys and switches are shown using the exact name shown on
the pump itself. Examples are provided below. See Figure 2-1 or Figure 2-2 for
the location of front panel keys and displays.
1-2COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL07-19-H2-646
IntroductionManual Conventions
Pump keys or switches with a single function are shown like this: START
The function of the pump’s four soft keys changes depending on the pump’s
current activity. The current function of each soft key is shown on the Main
Display above the key. Soft keys are shown in text like this:
Main display messages (other than the soft keys) are shown like this:
Main Display Text
Pump module messages are shown like this: PUMPING
Warnings, Cautions, and Notes
Certain items in this manual are highlighted by special messages. The definitions
are provided below.
Warnings
Warning messages indicate a possible hazard which, if not avoided, could result in
personal injury or death.
Done
Cautions
Caution messages indicate a problem or unsafe practice which, if not avoided,
could result in minor or moderate personal injury, product or property damage.
Notes
Notes provide supplemental information to the accompanying text. Notes precede
the text to which they apply.
07-19-H2-646COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL1-3
1Labeling
Labeling
See Figure 11-6 for a diagram identifying all labels affixed to the pump.
Labeling Symbol Definitions
IPX1Drip-proof equipment; enclosed equipment protected against dripping
fluids.
~
U
U
L
L
CLASSIFIED
MEDICAL ELECTRICAL
EQUIPMENT, PROFESSIONAL
IN ACCORDANCE WITH UL2601-1 AND
CAN/CSA C22.2 NO. 601.1.
5R78
Alternating current or equipment intended to be connected to an
alternating current source.
Fuse.
Attention, consult accompanying documents.
Type CF applied part.
Marked item is recyclable.
This product is classified by Underwriters Laboratories Inc. with
respect to electric shock, fire, and mechanical hazards only in
accordance with UL-2601-1, Second Edition, and CAN/CSA C22.2
No. 601-1.
The symbol of conformity to the Council directive 93/42/EEC.
EU Authorized Representative: Baxter S.A. B-7860 Lessines, Belgium
Labeling Abbreviations
COMM. PORTCommunications Port
VOL.Vo l u m e
CONT.Contrast
1-4COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL07-19-H2-646
IntroductionLabeling
Serial Number Format
This manual is applicable to all the Baxter product codes listed in Table 1-1. The
table lists the serial number suffix applicable to each Colleague pump product
code covered by this manual, and the language of the pump’s exterior labeling and
software user interface. The date of manufacture can be determined from the
serial number as shown in the following example.
Year of Manufacture
(12 = 2002, 9 = 1999)
Month of
Manufacture
12040362CS
Sequential NumberSee Table 1-1
Examples:
8060517CSU.S. Colleague Pump manufactured June 1998
10025678TGGerman Colleague 3 Pump manufactured Feb. 2000
11081234TESpanish Colleague 3 Pump manufactured Aug. 2001
Table 1-1Product Code to Serial Number Suffix Cross-Reference List
Serial Number
Product CodeProduct
2M8151U.S. English Colleague PumpCS
2M8151KU.K. English Colleague PumpCK
2M8153U.S. English Colleague 3 PumpCT
2M8153K U.K. English Colleague 3 PumpTK
BRM8151Brazilian Portuguese Colleague PumpCB
BRM8153Brazilian Portuguese Colleague 3 PumpTB
Suffix
CNM8151Swedish Colleague PumpCW
CNM8153Swedish Colleague 3 PumpTW
DNM8151French Colleague PumpCF
DNM8153French Colleague 3 PumpTF
GNM8151Danish Colleague PumpCN
GNM8153Danish Colleague 3 PumpTN
HNM8151German Colleague PumpCG
HNM8153German Colleague 3 PumpTG
PNM8151Spanish Colleague PumpCE
PNM8153Spanish Colleague 3 PumpTE
TRM8151Turkish Colleague PumpCY
TRM8153Turkish Colleague 3 PumpTY
WNM8151Dutch Colleague PumpCD
WNM8153Dutch Colleague 3 PumpTD
07-19-H2-646COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL1-5
1Safety Summary
Safety Summary
General precautions to observe while using the pump are shown below. Warnings
and cautions related to personal safety that appear throughout the manual are also
summarized in this section.
General
■ BEFORE INITIALLY POWERING ON THE PUMP, CHARGE
THE BATTERIES FOR AT LEAST 12 HOURS. A COMPLETE
CHARGE MAY TAKE LONGER THAN 12 HOURS.
■ In accordance with the International Standard, IEC 60601-1 (1988-12)
Medical Electrical Equipment — Part 1: General Requirements for
Safety, Colleague and Colleague 3 Pumps are classified as:
•Class 1
•Type CF
•Drip-proof
•Not suitable for use with flammable anesthetic mixtures with air,
oxygen or nitrous oxide
•Continuous operation
■ This manual has been developed with consideration to the requirements in
the International Standard, IEC 60601-2-24 (1998-02) Medical Electrical
Equipment — Part 2-24: Particular Requirements for Safety of Infusion
Pumps and Controllers. Data presented in the Technical Specifications
reflect specific test conditions defined in this standard. Other external
factors such as varying back pressure, temperature, head height, set usage,
fluid restrictions, solution viscosity, or combinations of these factors, may
result in deviations from the performance data enclosed.
■ This product is classified by Underwriters Laboratories Inc. with respect
to electric shock, fire and mechanical hazards only in accordance with
UL2601-1, and CAN/CSA C22. 2 No. 601-1-M90. In accordance with
these documents, this equipment is classified as:
•Class 1
•Type CF
•IPX1 (Drip-proof)
•Not suitable for use with flammable anesthetic mixtures with air,
oxygen or nitrous oxide.
•Continuous operation
■ Prior to operating the pump, carefully read its Operator’s Manual to fully
understand the pump’s functionality and to ensure safe and proper
operation. An Operator’s Manual in the appropriate language is shipped
with each pump.
■ Although the pump has been designed and manufactured to exacting
specifications, it is not intended to replace trained, qualified personnel in
the supervision of infusion therapy.
1-6COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL07-19-H2-646
IntroductionSafety Summary
■ Read and understand this manual before attempting to perform service or
maintenance on the pump.
■ To ensure that the pump continues to perform within specifications,
perform the routine maintenance procedures described in Chapter 6 of
this manual at least once every 12 months.
Warnings
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
Before using the pump, charge the batteries for at least 12 hours. A complete charge may
take longer than 12 hours.
Pump modules from Colleague and Colleague 3 pumps are not interchangeable. Do not
replace Colleague pump modules with Colleague 3 pump modules or vice-versa.
U.S. AND CANADA: Use only Baxter standard administration sets equipped with keyed
On/Off Clamps that have an “s” after the code number, or are labeled as pump
compatible. If you have questions about administration set compatibility, contact the
Baxter Product Information Center at the number shown on the administration set
labeling. There are risks associated with using anything other than the recommended
administration sets with this device.
LATIN AMERICA: Use only the following administration sets with the pump:
•Baxter “Green Label” C96XX administration sets equipped with keyed On/Off
Clamps, or
•Baxter standard administration sets equipped with keyed On/Off Clamps that
have an “s” after the code number, or are labeled as pump compatible. If you
have questions about administration set compatibility, contact the Baxter
Product Information Center at the number shown on the administration set
labeling. There are risks associated with using anything other than the
recommended administration sets with this device.
EEC Countries: Use only Baxter “Green Label” C96XX administration sets equipped with
keyed On/Off Clamps. There are risks associated with using anything other than the
recommended administration sets with this device. See “Recommended Administration
Sets” in Chapter 3 of the appropriate Operator’s Manual for more information.
In EEC Countries, use only Continu-Flo “Green Label” C96XX administration sets as the
primary fluid line when administering a secondary medication. Carefully follow the
directions on the primary and secondary administration set labels.
Always read and follow the instructions which accompany the source container and the
administration sets you are using. Carefully follow any label copy instructions for
loading, removing, and reloading the set, as well as the recommended set change
interval. For optimal pump performance, set use should not exceed the change interval
shown on the set's label copy or 72 hours, whichever is less.
For infection control purposes, consider the set change interval recommended by the
United States Centers for Disease Control and Prevention (CDC), your institution’s
guidelines, and the instructions provided with the administration set, using whichever is
most appropriate.
Baxter will assume no responsibility for incidents which may occur if the product is not
used in accordance with product labeling.
! WARNING !
07-19-H2-646COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL1-7
DO NOT CONNECT THE ADMINISTRATION SET TO THE PATIENT WHEN PRIMING.
1Safety Summary
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
DO NOT USE FUNCTIONAL TEST MODE ON PATIENTS.
While the pump automatically closes the keyed On/Off clamp, always close the regulating
clamp on the administration set before removing the tubing from the pump module.
The administration sets should be disposed of in an appropriate manner, considering the
nature of the residual fluid that may be contained within, in accordance with the hospital
disposal practices.
If the device has been dropped or appears to be damaged, it should be taken out of
service and inspected by qualified service personnel.
Prior to starting an infusion, verify that no drops are falling in the drip chamber and the
programmed information is correct.
This device should be used only with Baxter accessories specified for this device. There
are risks associated with using anything other than the recommended accessories with
this device. Accessories designated for use with this device are listed in Chapter 4.
To avoid the possibility of electric shock, use caution when the power supply is
connected to the AC input during calibration.
This device should be repaired only by qualified Baxter employees or Baxter-trained,
qualified personnel, using only Baxter recommended parts. There are risks associated
with using anything other than Baxter recommended parts. Baxter will assume no
responsibility for incidents which may occur if the product was not repaired in
accordance with procedures authorized by Baxter.
Epidural administration of drugs other than those indicated for epidural use could result
in serious injury to the patient.
! WARNING !
! WARNING !
! WARNING !
! WARNING !
! WARNING !
•Epidural administration of anesthetics is limited to short term infusion (not to
exceed 96 hours) with indwelling catheters specifically indicated for short term
anesthetic epidural drug delivery.
•Epidural administration of analgesics is limited to use with indwelling catheters
specifically indicated for either short term or long term analgesic epidural drug
delivery.
•To prevent infusion of drugs not indicated for epidural use, do not use
administration sets incorporating injection sites during epidural delivery.
•Clearly distinguish pumps used for epidural drug delivery from pumps used for
other routes of administration.
DO NOT CHANGE THE USER INTERFACE LANGUAGE TO OTHER THAN THAT OF THE
PUMP’S EXTERIOR LABELS.
Ensure that the pump is powered off and unplugged from the AC receptacle before
performing maintenance procedures.
To avoid the possibility of electric shock, use caution when the power supply is
connected to the AC input during calibration. Power supply line input is energized
whenever the pump is connected to an AC outlet.
Before connecting the pump to an AC outlet, make certain that the plastic insulator shield
is in place over the power supply assembly.
During the power-up test, do not wear a grounding wrist strap. Put the wrist strap back on
only when the pump is disconnected from the AC outlet.
1-8COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL07-19-H2-646
IntroductionSafety Summary
Cautions
Caution
Caution
Caution
Caution
Caution
Use of this device is restricted by Federal Law (USA) to sale or use by, on the order of, or
under the supervision of a physician or other licensed health care professional.
The power cord must be connected to a 100-120 VAC/220-240 VAC, 50/60 Hz, properly
grounded 3-wire receptacle designated as “Hospital Grade.”
To avoid possible damage to the pump, make Nurse Call connections only as shown in
Figure 3-1.
Use only accessory equipment complying with the device’s safety requirements; failure to
do so may lead to reduced safety levels of the resulting system. Consideration relating to
accessory choice shall also include:
•use of the accessory in the patient vicinity
•evidence of the safety certification of the accessory has been performed in
accordance with the appropriate UL2601-1 or IEC 601-1 and/or IEC 601-1-1
harmonized national standard.
As with all medical electronic equipment, care must be exercised to avoid exposing this
device to powerful sources of electromagnetic interference. This device design has been
tested to current U.S. and European standards and guidelines for medical devices. The
device was not found to be adversely affected by these susceptibility tests and will
perform safely. The device’s emissions were also found to be acceptable. Using the pump
near operating equipment which radiate high energy radio frequencies (such as
electrosurgical/cauterizing equipment, two-way radios, or cellular telephones) may cause
false alarm condition. If this happens, reposition the pump away from the source of
interference; or turn off the pump and, if clinically necessary, manually regulate the flow
with the regulating clamp according to your institution’s guidelines.
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
When infusing through a central line catheter, Baxter recommends using sets with a Luer
lock adapter.
Follow the cleaning schedule and methods described in Chapter 6 to ensure proper
maintenance of the device.
When mounting the Colleague pump to an IV pole or other mounting location, ensure it
has been securely clamped. The Colleague 3 Pump, due to its weight, cannot be mounted
on a headboard or other location except an IV pole.
To avoid personal injury, ensure that the IV pole is stable and secure. Ensure that the pole
is able to support the pump, along with any other devices, without tipping or falling. The
pole diameter should be between 0.95 and 3.81 cm (0.37 and 1.5 in.).
Never place the pump on the bed with the patient.
If mounting a Colleague pump on a headboard or footboard, ensure that the headboard or
footboard is stable, secure, and has a depth greater than 1.91 cm (0.75 in.) and height of at
least 5.08 cm (2 in.).
Ensure pump is mounted where it is easily accessible and the administration set can be
installed without stretching or kinking the tubing.
This device may interact with some electrocardiograph (ECG) equipment during infusions.
Proper setup of ECG should eliminate any difficulties. Consult the ECG instruction manual
for more information.
07-19-H2-646COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL1-9
1Safety Summary
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Do not clean, disinfect, or sterilize any part of the device by autoclaving, or with ethylene
oxide gas. Doing so may damage the device and void the warranty. Only external parts of
the device should be disinfected.
Do not use hard instruments for cleaning. Do not spray cleaners directly into the tubing
channel, the bezel openings adjacent to the Volume and Contrast controls, or the area
where the power cord enters the housing.
Do not use the following chemicals on the pump, as they will damage the front panel:
acetoaldehyde, acetone, ammonia, benzene, hydroxytoluene, methylene chloride, and
ozone. Do not use cleaners containing n-alkyl dimethyl ethyl benzyl ammonium chloride.
Wear a grounding wrist strap when assembling and disassembling the pump.
Do not lay the pump face down on components or tools which could scratch or damage
the keypad.
Ensure that wires are not pinched or overstressed.
Grounding reliability can only be achieved when this pump is connected to an
earth-grounded receptacle marked “Hospital Grade.” (When grounding receptacle
reliability is in doubt, the equipment should be powered by its batteries.)
When troubleshooting the pump, do not inject or apply signals unless the procedures in
this chapter specifically instruct you to do so. Damage to the pump or its subassemblies
could result.
If any evidence of damage is found, repair as necessary or contact your authorized
service provider.
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Do not use anything other than distilled water and a foam swab to clean the prisms. Cloth
or cotton swabs may scratch the prisms’ surfaces and anything other than distilled water
can fog the prisms.
Do not use hot air or attempt to dry the area using a cloth.
Do not store the pump unplugged and powered on. The batteries may discharge
completely, permanently damaging them.
Failure to plug the pump into an AC power supply when a Battery Depleted alert occurs
may cause permanent damage to the batteries.
Unlock locking connectors before attempting to disconnect them.
Ensure that the locking flange on connector J5 is unlocked before removing the flat
ribbon cable.
Ensure that connectors are mated and observe the correct orientation.
LED/icon PCB harness connector P15 is not keyed. Make certain that the connector is
correctly oriented before connecting. The connector is correctly oriented when the black,
yellow, and green leads face toward the handle (top) of the front bezel housing.
Unlock the flat ribbon connector flange for CN1 before removing the flat ribbon cable.
P109 connector harness is not keyed. Ensure that the connector is oriented so that the
green, yellow, and black leads face away from the PCB surface when it is connected.
1-10COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL07-19-H2-646
IntroductionSafety Summary
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Make certain that the contrast control wiring harness is routed under the EMI shield when
routing wiring harness under adjacent UIM PCB mounting tab.
Unlock the volume control header connector before disconnecting the connector.
Overtightening the fuse caps may cause the fuse holder to break.
Where connectors are identified as “locking,” unlock the connectors before removing
cables.
Ensure that the battery bracket lower mounting tabs are correctly engaged in the housing.
Ensure that the wiring harnesses beneath the batteries are not pinched, and that all four
harnesses protrude away from the batteries.
Ensure that the DC output harness connector is oriented on the power supply connector
such that the lock on the harness connector engages the lock on the PCB connector.
Ensure that fan leads are correctly oriented within the connector SK2 harness.
If the backup beeper is secured in place with push nuts and they are loose, or a standoff
breaks, you must replace the rear housing. Ensure that no loose parts remain inside the
pump. A broken standoff or loose backup beeper could cause the pump to malfunction.
Keep the electrical connections on the beeper away from the clip’s center tab even though
the connections are insulated.
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
When removing the 12 VDC connector nut, use care to ensure that the 12 VDC connector
wiring is not caught in the flare nut socket.
To avoid cracking the PANEL LOCKOUT button, do not squeeze it with pliers or pry it
sideways off of the shaft.
Wiring must be routed clear of standoffs and not pulled around standoffs.
Improper positioning of ferrite rings can result in short circuits.
To ensure correct reassembly, note the routing and connecting points of AC fuse holder
wiring before removing AC fuse holder(s).
Ensure that the AC leads that connect to the fuse holders are not touching the fan blades.
Note the routing of the wires before removing them.
Note the routing and connecting points of the AC Connector wiring before removing the
wiring.
This procedure involves removing the screws that secure the pump modules to the user
interface module. The pump can suddenly separate and be damaged if it is not adequately
supported when the securing screws are removed. To support the pump while it is
upside-down, use the top half of the shipping carton packing insert as a holding fixture
(see Figure 8-9).
Caution
07-19-H2-646COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL1-11
Do not short across any other solder joints other than those shown.
1Safety Summary
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Do not disconnect the lithium backup battery at any time during the calibration procedure.
The pump’s Battery and Pump History will be erased if you disconnect the backup battery.
The potentiometers are held in adjustment with Red Glpt varnish. If adjustments are
necessary, use a stylus to carefully remove the varnish without damaging the
potentiometer. Rotate the pot fully clockwise and then counterclockwise once before
making the final adjustment.
Correct placement of the wiring and fuse holders is critical; incorrect placement can
damage the elevated IC heat sink on the UIM PCB.
Do not damage or change the setting of VR1 while removing the varnish.
Motor magnets may attract metal debris to motors or PCBs. To prevent debris from
entering the pump mechanism, always maintain a clean work area when performing
procedures involving the pump module.
The clamp VB locking tab must be installed with the beveled surface facing away from the
“V” portion.
Ensure that the wiring harnesses are clear of the module mounting surfaces before mating
the module to the mounting surface.
Unlock the flat ribbon cable connector flange on J7 before removing the keypad flat
ribbon cable.
Forcing the housing off of the pump mechanism can damage the mechanism. If you have
difficulty removing the housing, recheck the jaw opening, and repeat the procedure.
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Forcing the mechanism into place can damage it.
The accessory harness connector is not keyed. The connector is correctly oriented when
the violet and blue leads on the harness face toward the accessory connector on the
housing.
Unlock flat ribbon cable connector flange before removing cable.
Maintain a soldering pencil tip temperature of 536° to 662° F (280° to 350° C) to avoid
damaging the insulation.
Tip the pump module on its side to avoid solder splashes.
Do not damage the flex ribbon by pulling it too hard.
Check the bezel for broken alignment tabs. You may remove damaged tabs by cutting or
breaking them off. After inspecting the interior of the pump to verify that no broken pieces
remain, mark the broken tab(s) with indelible ink to indicate that the broken pieces have
been accounted for.
Ensure that wiring is not pinched and that the ring terminal is parallel to the case edge.
Ensure that connectors are mated correctly. Failure to correctly orient connector P8 can
cause a blown fuse.
1-12COLLEAGUE VOLUMETRIC INFUSION PUMP SERVICE MANUAL07-19-H2-646
Loading...
+ 392 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.