There are risks associated with using anything other than the recommended IV extension sets with this device. Sets
designated for use with this device are identified in the section “Simplified Instructions”. Baxter’s warranty on this
device will be null and void and Baxter will assume no responsibility for incidents which may occur if the product is
not used in accordance with product labeling. See the section “Precautions” for a complete list of warnings and
cautions.
For use only by Baxter Healthcare Corporation. The software contains proprietary information belonging to Baxter
Healthcare Corporation. The software must not be reproduced or disclosed to others without prior written approval.
Any unauthorized use of this information may subject the user to substantial liability.
Documentation Copyrights
Duplication or distribution of this manual and any information contained within is strictly prohibited without the
express written permission of Baxter. This manual and any information contained within, may not be reproduced,
distributed, or transmitted in any form, or by any means, for any purpose, without the express written permission of
Baxter. To order additional copies of this manual, or other related manuals, contact your local Baxter Service Center.
Disclaimer
The information in this document has been carefully examined and is believed to be entirely reliable. However, no
responsibility is assumed for inaccuracies. Furthermore, Baxter reserves the right to make changes to any products
herein to improve readability, function, or design. Baxter does not assume any liability arising out of the application
or use of any product or circuit described herein; neither does it cover any license under its patent rights nor the rights
of others.
Trademark Information
Baxter, the Baxter wordmark, Micro-Volume, and Auto Syringe are trademarks of Baxter International Inc. in the
U.S. Other trademark applications pending. Terumo is a trademark of Terumo Medical Corporation. ASCEND,
and Hi-Tor Plus are trademarks of Huntington Professional Products. LpH SE and Septisol are trademarks of
Steris Corporation. Cidex 7 is a trademark of Johnson & Johnson.
The selected options can easily be reviewed and the chosen configuration can be
changed to meet new or different requirements.
The pump operates on an internal rechargeable battery pack and can also be operated
while attached to a “900-Series” battery charger (model number CAS40-01-900) or to a
Multi-Charger (product code 1M8563).
The pump is supplied with a pole clamp and a built-in IV pole loop. The pump can also be
used on a tabletop.
Information on writing custom software for the pump is available in the Auto Syringe AS50
Infusion Pump Programmer’s Manual. Information on troubleshooting and repair of the
pump is available in the Auto Syringe AS50 Infusion Pump Service Manual.
Federal (USA) law restricts this device to sale by or on the order of a physician.
The charger port is to be used only with “900 Series” chargers or other accessories that are
labeled specifically for use with Auto Syringe AS40A or AS50 Series Infusion Pumps, including
!
the 1M8563 Multi-Charger.
CAUTION: Refer all service, repair, and calibration to qualified technical
!
!
●Though the factory-supplied configuration settings are suitable for most therapies, the
personnel.
CAUTION: Do not operate the Auto Syringe AS50 Infusion Pump in the
presence of flammable anesthetics, oxygen-enriched, or explosive
atmospheres.
operator and hospital professionals should verify that the pump’s settings are
appropriate for the clinical application.
●The Protocol Library feature is a configuration option developed to facilitate infusion
programming. Before using a protocol for any specific drug, refer to the full
prescribing information supplied by the drug manufacturer.
●As with all medical electronic equipment, care must be exercised to avoid exposing
this device to powerful sources of electromagnetic interference.
●Do not expose the pump to X-rays, gamma rays, or other ionizing radiation, or to
strong electric or magnetic fields.
●Do not autoclave, steam sterilize, EtO sterilize, or subject the pump or charger to
temperatures in excess of 104° F (40° C).
●Do not use the following chemicals on the device, as they will damage the front panel:
acetone, acetoaldehyde ammonia, benzene, hydroxytoluene, methylene chloride,
and ozone, Do not use cleaners containing n-alkyl dimethyl ethylbenzyl ammonium
chloride unless they appear in the list of recommended cleaners in the Pump
Maintenance section of this manual.
●If a LINE OCCLUDED alarm occurs, RELIEVE THE RESIDUAL PRESSURE IN THE
SYRINGE BY RELEASING THE PLUNGER DRIVER. If the pressure is not relieved
prior to clearing the occlusion, an unintentional bolus delivery may occur when the
blockage is cleared. Check for kinked tubing, clogged catheter, closed stopcocks, etc.
Check for foreign material preventing movement of the pump mechanism.
●Be sure to PURGE THE SYSTEM OF ALL AIR BEFORE ADMINISTERING ANY
MEDICATION. Failure to follow this normal infusion procedure could precipitate
serious consequences.
●Remember that the volume of fluid contained in the connecting tubing is a residual
amount and cannot be infused. Allow for this extra volume of fluid when initially filling
the syringe.
●In MANUAL SCHEDULE mode, START must be pressed to deliver each dose and
continue the infusion regimen. Pressing any other key silences the audio portion of
DOSE DUE alarm and leaves the pump in Standby state, without initiating a dose
the
delivery.
●Although extreme care has been taken in assembly, component selection, and quality
control during manufacture of the Auto Syringe AS50 Infusion Pump, routine checks
and maintenance procedures must be performed so that patients receiving infusions
can be assured of trouble-free operation.
●Caution must be exercised in the selection of drugs intended to be delivered via any
infusion pump. If the drug contained in the syringe will be exposed to extreme
environmental conditions for prolonged time periods, IT IS IMPORTANT TO SELECT
DRUGS THAT WILL NOT CHANGE PHARMACOLOGICALLY UPON SUCH
EXPOSURE.
●As with all automatic infusion devices, whenever a TOXIC OR DANGEROUS LEVEL
of drug is stored in the reservoir, CONSTANT MONITORING of the infusion is
required.
●CAUTION must be employed to assure that the Auto Syringe AS50 Infusion Pump is
in good working order before putting it into use. If the pump is being operated on
battery power alone, ensure that the battery has been charged as described in this
manual.
●When the pump is first turned on, verify that the Lamp Test is accurate (see “Display,
Battery, and Syringe Sensor Check” on page 75).
●Verify all program data before pressing START.
●Wipe off spills immediately. Do not allow fluid or residues to remain on the pump.
●Repeated incorrect identification of syringe size by Syringe Recognition may signify a
pump fault condition, or that a syringe manufacturer has made a dimensional change.
The pump should be removed from service as soon as possible, so that the problem
can be investigated.
The status display has eight information areas, called
fields. Except for the text field at the bottom, each field
displays three types of information:
INF
B
D
N
L LIMIT
k
nit
m
mgh
m
mLmc
● The markings on the left and right side of each field are
called annunciators. The left annunciators identify the
type of data contained in the field, while the right
annunciators identify the units of measure.
● The central portion of each field displays variable
k
numeric data, either decimal numbers or time. This data
generally represents either infusion program information
or an active display such as a running total of the
amount of drug infused.
● The bottom field of the status display is a two-line text
display, used to prompt the programming sequence or to
display a message about the infusion.
After the infusion mode has been selected, the pump
automatically displays the necessary fields to be
programmed. When the pump needs information, it
provides a text prompt and flashes the current field.
Note: The pump’s backlight is on whenever the charger is connected and the pump is
switched on. The pump can also be configured to turn the backlight on for a specified time (0 to 60 minutes) after each keypress when the charger is not connected. See the Auto Syringe AS50 Infusion Pump Technical Manual for details.
Field #2 displays the volume to be infused
(mL), or the infusion rate (mL/hr or mL/min).
General Information
OD
g
C
s
g
OSE
g
g
TOT
s
g
O
TOT
O
Field #3 displays dose duration (HH:MM), or
patient body weight (kg).
Field #4 displays dose interval (
HH:MM), drug
concentration (mg/mL, or U/mL), or bolus size
(Units or mL).
Field #5 displays time remaining until the next
dose is due (hh:mm), dose size (mcg/min,
mcg/hr, mg/min, mg/hr, mcg/kg/min, mcg/kg/hr,
mg/kg/min, mg/kg/hr, mU/min, mU/hr, U/min, or
U/hr) or bolus size (mg, mcg/kg, mg/kg, or
Units).
Field #6 displays total delivered (mL, mcg, mg,
or grams), volume limit (mL), bolus size (mL),
or final volume (mL).
B
N
D
V
k
nit
m
m
mcgk
ram
mLmc
Field #7 displays total delivered (mL), or drug
volume to be mixed with infusion base solution
(mL).
Arrow KeysThe up arrow key (▲) increases a number,
The keypad includes a standard set of number keys and
decimal point (referred to in this document as digit keys),
arrow keys, and various function keys.
PUR
BOLU
L
NFIR
selects a different field, or displays the
next item in a list.
The down arrow key (▼) decreases a
number, selects a different field, or
displays the previous item in a list.
Digit KeysThe digit keys are used to enter numeric
values directly.
Function KeysThese keys activate specific functions.
Action KeysThese keys start and stop the pump motor.
START
STOP
Confirm KeyThis key is used to complete a
programming step, to respond to a
prompt, or to display additional infusion
program information.
When the pump is in Standby state, the infusion is stopped and the STANDBY light
flashes. If the pump is left in Standby state for more than two minutes, a
alert or alarm occurs.
Run State
There are two Run states: Run Delivering state and Run Counting state.
In Run Delivering state, the infusion program is running and the pump is delivering an
infusion. The three
In Run Counting state, the infusion program is counting time until the next dose is due,
and the pump is not delivering an infusion. The top
PUMP IS IDLE
RUN lights flash in a “falling drop” pattern.
RUN light flashes.
Note: Bolus and Purge operations are special infusion states. The
unison when bolusing or purging in both Standby and Run states.
Data Display and Entry
Numeric DataNumbers are displayed in a right-justified format. A leading
Time Data FieldsIn time fields, values are displayed right justified. Leading
Automatic DecimalIf a number is entered without a decimal point, the pump
Flashing DataThe data within a field flashes to show that the field has
Flashing AnnunciatorsThe annunciators for the current field flash. The current
RUN lights flash in
zero appears before decimal values when the field size
permits. For example, the value “.12” is displayed as
0.12.
zero(es) are automatically entered if needed. For example,
1 minute is entered by pressing
displayed as
0:01.
1, CONFIRM and is
automatically enters a decimal point to the right of the
number when
CONFIRM is pressed.
been opened and is ready to be programmed or edited.
field will be opened for editing when
EDIT is pressed.
Data TestAll entered data is tested when
CONFIRM is pressed. If the
entered value is too large or too small, the pump issues a
PUMP LIMIT alert and substitutes the nearest acceptable
Arrow KeysArrow keys can be used instead of individual digits for
Key SoundsAn audible click provides confirmation of each keypress.
DashesDashes are displayed when a field is clear and open for
Protocol Library Feature
The Protocol Library feature is a configurable option that allows authorized personnel to
store named sets or protocols of frequently used programming data. Protocols may also
be grouped into libraries.
General Information
selecting predefined responses, incrementing or
decrementing fields. In general, arrow keys cannot be used
to initialize fields that have been cleared to dashes, except
for syringe size and manufacturer fields.
The pump indicates an invalid keypress by sounding a short
beep. Simultaneous keypresses are ignored.
programming. In general, digit keys must be used to enter a
value in a dashed field. If an arrow key is pressed while
dashes are displayed, the pump beeps and displays the
message
USE DIGITS.
Using a protocol can simplify the programming process by providing default selections at
some of the data entry steps. For instructions on selecting a protocol, see “Using
Protocols” on page 32.
The Protocol Library feature is enabled or disabled through the pump configuration mode.
This procedure can be done by authorized personnel on site, and is described in the Auto
Syringe AS50 Infusion Pump Service Manual.
A protocol may specify any of the following:
Default Infusion ModeIf a protocol specifies a default infusion mode, the infusion
mode programming step is skipped. The pump briefly
displays the infusion mode name before proceeding directly
to the syringe manufacturer programming step.
Default ConcentrationIf a protocol specifies a default concentration, the default
value appears automatically at the start of the
concentration programming step. This value may be
changed, or it may be accepted by pressing
CONFIRM.
Default Dose or RateIf a protocol specifies a default dose or rate, the default
value appears automatically at the start of the
ENTER RATE programming step. This value may be
or
changed, or it may be accepted by pressing
ENTER DOSE
CONFIRM.
Bolus PreventionIf a protocol specifies bolus prevention, the pump will not
allow a bolus to be programmed or delivered. If
pressed, the pump beeps and displays the message
Default Bolus SizeIf a protocol specifies a default bolus size, the default value
appears automatically when the Bolus Size field is first
opened for editing. This value may be changed, or it may be
accepted by pressing
Bolus Maximum RateThe bolus delivery rate is normally limited to the smaller of:
the limit specified by the RATE RANGE configuration
option, or the maximum rate for the programmed syringe. If
a protocol specifies a bolus maximum rate, the pump will
further limit the bolus delivery rate so it does not exceed the
specified bolus maximum rate, which is defined in terms of
dosage (such as mcg/kg/min).
Bypassing the Protocol Library Feature
Note: The NONE selection may not be available, depending upon configuration.
CONFIRM.
Selecting
Protocol Library feature and allows infusion mode selection and programming to proceed
in the normal way.
NONE at the SELECT LIBRARY or SELECT PROTOCOL step bypasses the
Serial Communication Feature
The pump provides remote control and monitoring capabilities. A standard 9-pin
connector (DB9) allows the connection of an RS-232 cable between the pump and a
computer. The cable allows an operator at a remote computer to control pump operations
and to query the pump for data such as infusion totals, program values, and operational
status. The remote control feature is a configurable option. See the Auto Syringe AS50
Infusion Pump Programmer’s Manual for details.
The pump provides 18 infusion modes, which fall into two major categories: continuous
and timed infusions. Continuous infusions deliver at a steady, programmed rate. Timed
infusions deliver one or more equal doses according to a programmed schedule. The
infusion modes are briefly described below. For more information see “Detailed
Instructions” on page 27.
Continuous Infusion Modes
mL/hr and mL/min: The rate of infusion is programmed in mL/hr or
Units/hr, Units/min, mUnits/hr and mUnits/min: The drug concentration is entered in
mg/hr, Mg/min, mcg/hr and mcg/Min: The drug concentration is entered in mg/mL, and
General Information
mL/min, depending on the mode selected. A volume limit
can be programmed for these modes only.
units/mL, and the dose is entered in Units/hr, Units/min,
mUnits/hr or mUnits/min, depending on the mode selected.
The pump automatically displays the effective rate in mL/hr.
the dose is entered in mg/hr, mg/min, mcg/hr or
mcg/min, depending on the mode selected. The pump
automatically displays the effective rate in mL/hr.
mg/kg/hr, mg/kg/min, mcg/kg/hr and mcg/kg/min: The patient's body weight is
entered in kg, the drug concentration is entered in mg/mL,
and the dose is entered in mg/kg/hr, mg/kg/min, mcg/kg/hr
or
mcg/kg/min, depending on the mode selected. The pump
automatically displays the effective rate in mL/hr.
CUSTOM DILUTION: In this mode the pump is programmed for a continuous
infusion and the drug concentration is adjusted for a target
infusion rate and dose. The target rate is entered in mL/hr,
the patient’s body weight is entered in kg, the initial drug
concentration is entered in mg/mL, the target dose is
entered in mcg/kg/min, and the final volume is entered in
mL. The pump calculates and displays the volume of drug
(in mL) to be added to the infusion base solution to produce
a final concentration that achieves the target rate and target
dose. The final concentration is displayed when the infusion
is initiated.
SINGLE DOSE: In this mode, the pump is programmed to infuse a single
MANUAL SCHEDULE:In this mode, the pump is programmed to infuse a dose
AUTO SCHEDULE:In this mode, the pump is programmed to infuse a dose
dose over a specified period of time. The dose is entered in
mL, and the time period is entered in hours and minutes.
over a specified period of time, and to repeat the dose at
specified intervals. The dose is entered in mL, and the time
period and interval are entered in hours and minutes.
The pump alerts the operator when a dose is due to start.
The operator must then press
START to initiate each dose.
over a specified period of time, and to repeat the infusion at
specified intervals. The dose is entered in mL, and the time
period and interval are entered in hours and minutes.
Whereas
to press
MANUAL SCHEDULE mode requires the operator
START to initiate each dose, AUTO SCHEDULE
mode automatically starts each dose when it is due.