Baxter Arena Schematic Diagrams

CC
BB
74m
Filter
“B”
24.84 mL/min
AAMI Standard
Water
461.58 mL/min461.58 mL/min
DWG No.DWG No.
DESCRIPTIONDESCRIPTIONREVREV DATEDATEECO #ECO # APPROVEDAPPROVED
Dialyze Mode
Numeric values shown are using Drake Style Concentrate @ 500 mL/min Dialysate Flow
Light Blue: Water Purple: Fresh Dialysate
Red: A Concentrate Green: Ultrafiltrate
Dark Blue: B Concentrate
Pink: Water & Acid
461.58 mL/min461.58 mL/min
“B”
Concentrate
Pump
“B”
Concentrate
Regulator (Optional)
74m
Filter
13.58 mL/min
Approved
Backflow
Preventer
as Required by
Local Code
510 mL/min510 mL/min
100 step
1/3 mL/stroke
“A”
Drain Line
“A”
Concentrate
Pump
“A”
Concentrate
Regulator (Optional)
Safety Thermistor
Water Pressure Regulator
13.58 mL/min
100 step
1/3 mL/stroke
-500
-500
mmHg
mmHg
Supply
Manifold
Removal
Water Temp Control
Thermistor
Heater
Adjust to 15 PSI
Adjust to 15 PSI
During Flow
During Flow
500 mL/min500 mL/min
Air Gap
Chamber
Mixer
Air Removal
Sprayer
Air
Pump
Water
On /Off
Valve
Orange: Spent Dialysate
Black Dots: No Flow this Mode
Vent
to
Atmosphere
Hall-effect
Level
Sensor
(open/close
the water valve)
Magnet
Air
Trap
Drain
“A” Conductivity
Thermistor/Probe Set
1500 mL/min1500 mL/min
(Adjust vacuum by RPM)(Adjust vacuum by RPM)
Heat
Exchanger
X
Heat
Clean
Recirculation
Valve
2 .8
m / L4 4 m in
36.83X
Adjust to 16 PSIAdjust to 16 PSI
Supply Regulator
X’s flow
“B” Conductivity
Probe/Thermistor Set
“B” Mix
Chamber
500 mL/min500 mL/min
NOTES: All 4 flow path thermistors values are:
o
25 C = approx. 5000 ohms
o
37 C = approx. 3000 ohms
150m
Particle
Filter
Supply
Pump
(lower
middle)
Rinse Ports
With Proximity Sensors
Disinfect
Loop
Variable
Variable
mL/min
mL/min
(Runs higher than
(Runs higher than
500mL/min)
500mL/min)
Dialysate Monitoring Manifold (Also called “Final Manifold”, “Dialysate Manifold” or “Primary Manifold”)
Dialysate Conductivity
Probe/Thermistor Set
Pressure Equalizer
Variable Positive PSI
Flow
Variable Positive PSI
Not Adjustable
Not Adjustable
Restriction
Blood Leak Detector
2
X
8
Drain
C4C4
Flow
Equalizer
(valves will switch
every 13.68 seconds)
114
114
mL
mL
when
when
full
full
3
Bracket
500 mL/min500 mL/min
X X
2 t o 3 P S I
2 t o 3 P S I
filling,
filling,
16 PSI
16 PSI
@EOS
@EOS
Flow
Restriction
Output
End-of-Stroke Sensors
1
X
66
77
C2C2C3C3
711
114
114
mL
mL
when
when
full
full
2
Input Pressure Equalizer
C1C1
Connector
Bypass Valve (cycled during disinfect)
Dialysate Pressure Transducer
Flow Sensor
500 mL/min500 mL/min
Dialysate
Dialysate
Pressure
Pressure
Starts Here
Starts Here
Dialyzer
VenousVenous
5
Connector
1144
Adjust
Adjust
to
to
200 to
200 to
300
300
mmHg
mmHg
UF Flow
Meter
True
Check Valve
510 mL/min510 mL/min
ArterialArterial
Dialysate Pressure
(left side)
Speed Varies
with Flow
UF Removal
Regulator
22
1.5
1.5
mL
mL
when
when
11
UF Drain Line
Connector
Particle
Pump
full
full
150m
Filter
44
33
Rinse Valve
(Dialysate Pressure Relief
DDDD
CC
Valve)
BB
Air Gap
{
AA
Drain
223344
CONFIDENTIALCONFIDENTIAL
INFORMATIONINFORMATION
UNLESS OTHERWISE SPECIFIEDUNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN INCHES.DIMENSIONS ARE IN INCHES.
TOLERANCES ARE:TOLERANCES ARE:
.XX ± .01 .XXX ± .005
SURFACE FINISH:SURFACE FINISH:
DO NOT SCALE DRAWINGDO NOT SCALE DRAWING
BREAK ALL SHARP CORNERSBREAK ALL SHARP CORNERS
.005/.010 CHAMFER/RADIUS.005/.010 CHAMFER/RADIUS
ANGLESANGLESDECIMALSDECIMALS
± .5^
6363
Baxter
DESCRIPTIONDESCRIPTION
Arena Standard Hydraulics Flow Diagram
MATERIAL / FINISH / SPECIFICATIONSMATERIAL / FINISH / SPECIFICATIONS
157-1278-587
BB
L. Pittman
SCALESCALE
None
11
DATEDATEDRAFTERDRAFTER
1-8-04
SHEETSHEET OFOF
1 1
CorelDRAW FILECorelDRAW FILE
AA
REVREVDWG No.DWG No.
A