1. CPU :Intel Pentium 4 1.5Ghz / AMD XP 1500 ; Memory: 2GB (winXP)/3GB (win7
32bit)/8GB (win7 64bit)
2. Operating system (OS): windows XP(service pack 3)、windows 7 (service pack 1)
3-2. Software & Diver Installation
Open the VSpec2.1Sr software folder and click setup.exe as shown in Fig.3-2-1. Follow the
instructions of the setup program.
<Fig. 3-2-1>
First click Next in the window shown in Fig. 3-2-2.
<Fig. 3-2-2>
SEC2000 Spectra System
www.als-japan.com
11
Page 12
SEC2000 Operating Manual
When you have finished reading the License Agreement, select I accept the terms in the
license agreement and click Next ( Fig. 3-2-3).
<Fig. 3-2-3>
Fill in the User Name and Organization, you can use the option to make this program
available to everyone or personal. Click Next to move on.
<Fig. 3-2-4>
Please select the Complete setup type, and click Next. If you are satisfied with your
installation, select Install. The icon of Visual Spectra 2.1 is created on your computer screen.
SEC2000 Spectra System
www.als-japan.com
12
Page 13
SEC2000 Operating Manual
<Fig. 3-2-5>
Next, you should install the diver of the hardware. Open the Divers folder from the
VSpec2.1Sr software source folder.
<Fig. 3-2-6>
SEC2000 Spectra System
www.als-japan.com
13
Page 14
SEC2000 Operating Manual
Select the suitable diver and click the program, follow the InstallShield Wizard to install the diver.
<Fig. 3-2-7>
<Fig. 3-2-8>
Clicking Finish to complete installation of the diver as shown in Fig 3-2-9.
<Fig. 3-2-9>
SEC2000 Spectra System
www.als-japan.com
14
Page 15
SEC2000 Operating Manual
3-3. Connect SEC2000 to the PC
Using the enclosed USB cable, connect SEC2000 to the PC as shown in Fig.3-3-1.
<Fig. 3-3-1>
3-3-1. Windows XP
The computer automatically recognizes the USB interface. In the Found New
Hardware Wizard screen, check Install the software automatically
(Recommended) and click the Next.
<Fig. 3-3-2>
SEC2000 Spectra System
www.als-japan.com
15
Page 16
SEC2000 Operating Manual
When the Hardware installation is completed, click the Finish.
<Fig. 3-3-3>
3-3-2. Windows 7
The PC will recognize SEC2000 and you can start to operate the spectrometer.
<Fig. 3-3-4>
SEC2000 Spectra System
www.als-japan.com
16
Page 17
SEC2000 Operating Manual
4. Operation
4-1. Connection
SEC2000 consists of a light source, cuvette holder, detector and other accessories. The
pre-set mode is for measuring absorbance or transmittance. It looks like a straight bar in
the order of light source, cuvette holder, and detector, as shown in Fig. (a). To measure
a reflectance spectrum, reconfigure it, as shown in Fig. (b). To measure an irradiance
spectrum, remove light source, as shown in Fig. (c) of Fig. 4-1-1.
Note: When changing the combination of the device, the position of the light source and
detector are also changed slightly. The spectral shape of the light source in Scope mode
may be changed, due to changes of the intensity of the UV lamp and the VIS lamp. This
will not affect the measurement.
(a) (b) (c)
<Fig. 4-1-1>
Connect the Spectra System detector to a USB interface of the computer using the enclosed
USB cable and plug in the enclosed electrical cord to Light Source. Then turn on the light.
<Fig. 4-1-2>
SEC2000 Spectra System
www.als-japan.com
17
Page 18
SEC2000 Operating Manual
4-2. Program Start
Start the Visual Spectra 2.1 program and click the GO icon. The trace line will appear, when
the Spectrometer (SEC2000) has been connected properly with the computer.
<Fig. 4-2-1>
If the Spectra System driver is not installed properly or the USB connection fails between
Spectra System and your computer, Failed in loading SEC2000 device message will
appears. In this case, check the USB connection or reinstall the USB driver.
<Fig. 4-2-2>
SEC2000 Spectra System
www.als-japan.com
18
Page 19
SEC2000 Operating Manual
on
4-3. Software Operation
4-3-1. Menu Descriptions
From the Main Menu bar, the following selections will produce the following drop-down
options.
① File
- Open
Dark (Ctrl+D) A Dark spectrum (*. dark) is read and displayed on Overlay1.
Sample Dark
Reference (Ctrl+R)
Sample (Ctrl+O)
Processed
Settings
- Save
Dark Save the current dark spectrum as a file (*.dark).
Sample Dark Save the current sample dark spectrum as a file (*.spldark)
Reference Save the current reference spectrum as a file (*. refer).
Sample Save the Master Trace’s data as a file (*.sample).
A sample dark spectrum (*.spldark) is read and displayed on
Overlay1.
A reference spectrum (*.refer) is read and displayed on Overlay
2.
A sample spectrum (*.sample) is read and displayed
Master Trace.
Read the processed (absorbance/ transmittance/ irradiance)
spectrum to Master Trace.
It is activated when the dark and reference spectra are opened.
Set for a measuring spectrum and read the graphical view
setting file (*.ini).
Save the Master Trace’s data to file (processed spectrum,
Processed
Establish
environment
SEC2000 Spectra System
www.als-japan.com
saved as *.absorb/ *.trans/ *.irrad.).
It is activated when the dark and reference spectra are saved.
Establish environment for a measuring spectrum and save the
graphical view setting file (*.ini).
19
Page 20
SEC2000 Operating Manual
- Auto Increment
Enable
Base Name
Starting Index
Generates a filename automatically for saved spectra.
Set the base name for auto incremented files.
Ex. of file name: ALS1204,date→20110112, etc.
Set the starting index for auto incremented files.
Ex. of file name: ALS1204-1, ALS1204-2, etc.
- Print Setup Select and configure a printer for graphical spectrum printing
<Fig. 4-3-1>
SEC2000 Spectra System
www.als-japan.com
20
Page 21
SEC2000 Operating Manual
- Page Setup Write Title, Comment and Footer for printing graphical spectrum.
<Fig. 4-3-2>
- Print Preview Preview the graphical spectra printing.
<Fig. 4-3-3>
- Print (Ctrl+P) Print current graphical spectra.
- Exit End program.
SEC2000 Spectra System
www.als-japan.com
21
Page 22
SEC2000 Operating Manual
② Edit
- Copy Graph (Ctrl+C) Copy current graphical spectrum to a temporary site.
- Copy DataCopy current data to a temporary site.
- Add Marker (Ctrl+A) Make a marker for current graphical spectrum.
<Fig. 4-3-4>
- Erase Marker (Ctrl+E) Erase a maker that is located near a current cursor.
- Erase All Markers Erase all markers.
③ View
- Rename Window Rename a graphical spectrum window.
<Fig. 4-3-5>
SEC2000 Spectra System
www.als-japan.com
22
Page 23
SEC2000 Operating Manual
- Mouse Tracking
Cursor
Zoom
Pan
- Spectrum Scale
Auto Scale
Change Scale
Change mouse status to cursor.
Change mouse status to Expand/Reduce.
Change mouse status to drawing.
Adjusts the scale of the spectrum in the current spectral
window so that the peaks fill the display vertically.
Adjusts the scale of the spectrum optionally.
Full Scale
- Cursor Display Enable or disable the display of cursor for the spectral window.
- Grid DisplayEnable or disable the display of grid for the spectral window.
- Marker Display Enable or disable the display of maker for the spectral window.
- Color
Trace
Background
Frame
Cursor
Grid
SEC2000 Spectra System
www.als-japan.com
Adjusts the scale for all data to be displayed.
Master: Choose a master color for the spectral windows.
Overlay1~10: Choose each overlay for the spectral windows.
Choose a background color for the spectral window.
Choose a frame color for the spectral window.
Choose a cursor color for the spectral window.
Choose a grid color for the spectral window.
23
Page 24
SEC2000 Operating Manual
Text
Tick
Marker
Default
- Main Status bar Toggle the status bar display in the main visual spectral
- Spectrum Tool bar Toggle the tool bar display in the main visual spectral
Choose a text color for the spectral window.
Choose a tick color for the spectral window.
Choose a marker color for the spectral window.
Return all colors to the original.
window.
window.
<Fig. 4-3-6>
If the“Quantitative Analysis” is checked, the window will be displayed on the right
side of the main window as below figure.
<Fig. 4-3-7>
SEC2000 Spectra System
www.als-japan.com
24
Page 25
SEC2000 Operating Manual
④ Overlay
- Spectrum
- Time Series
- Add Overlay 1 ~ 10 Display the data for each spectral window (1 ~ 10).
- Clear Clear selected overlay data from the spectral window.
- Add Overlays (Multiple Selection) Display several data at the same time.
- Clear All Clear all overlay data from the spectral window.
⑤ Spectrum
- Config. Spectrometer Confirm the parameters of spectrometers for
wavelength calibration.
<Fig. 4-3-8>
- Reset Spectrometer This function cannot be used for user.
- Calib. Spectrometer This function cannot be used for user.
- Store Dark Save the spectrum measurement result as dark spectrum to
memory and display on overlay 1.
- Store Reference Save the spectrum measurement result as reference
spectrum to memory and display on overlay 2.
- Snapshot Halt data acquisition and take a snapshot of activity in the
spectral window.
- Scope Mode Present the data in scope mode.
- Scope Mode Minus Dark Switch the current spectral window into scope mode
and subtract the stored dark spectrum.
- Absorbance ModeDisplay data in absorbance mode. Please save the Dark
and Reference spectra before using this function.
- Transmittance Mode Display data in transmittance mode. Please save the
SEC2000 Spectra System
www.als-japan.com
25
Page 26
SEC2000 Operating Manual
Dark and Reference spectra before using this function.
- Reflectance Mode Display data in reflectance mode. Please save the Dark
and Reference spectra before using this function.
- Relative Irradiance Mode Display data in relative irradiance mode. Please
save the Dark and Reference spectra before using this
function.
- Reference Color Temp. Input the color temperature (K) of your blackbody
reference lamp used for relative irradiance measurements.
<Fig. 4-3-9>
- Quantitative Analysis It is effective only on absorbance mode.
- Peak Search Search peaks in the specified range. Click Save, the peak data
will be saved as *.csv file.
<Fig. 4-3-10>
SEC2000 Spectra System
www.als-japan.com
26
Page 27
SEC2000 Operating Manual
⑥Time Time course measurements of the specific wavelength or full spectrum
change is performed.
- Configure Time Acquisition Configure and establish the parameters for time
acquisition process.
<Fig. 4-3-11>
- Configure Time Acq. Channel Set up the channels for time acquisition process.
Six specific wavelengths can be set up from channel 1 to channel 6.
<Fig. 4-3-12>
SEC2000 Spectra System
www.als-japan.com
27
Page 28
SEC2000 Operating Manual
Combination 1 and 2 can perform operation among channel 1 ~ channel 6.
- Restore Parameters
<Fig. 4-3-13>
Select and open a file with the complete set of time acquisition parameters,
including the configuration settings for Time Acq. Channel.
- Save Parameters
Save a complete set of time acquisition parameters, including the configuration
settings for Time Acq. Channel.
- Activate Activate the time acquisition mode.
⑦ Window
- Cascade Arrange all spectral windows in a horizontally overlapping design.
- Tile Arrange all spectral windows side by side.
⑧ Help
- Help Topics Present the user manual for the system.
- About Visual Spectra 2.1 Present the version information of Visual Spectra 2.1.
SEC2000 Spectra System
www.als-japan.com
28
Page 29
SEC2000 Operating Manual
4-3-2. Functions
① Icons
Minus
Dark
Select this icon if the baseline of the resulting graph is unstable
based on the present Y coordinate, or to adjust the Y coordinate
based on the dark level. If this icon is clicked, the baseline will
be forced to be Y (Intensity) = 0 which will be reflected in the
graph. Though this function is used for adjusting an unstable
distribution of a spectrum, you must pay attention to the fact that
it is based on the original data when saved as a file.
Auto
Scale
Full
Scale
Scale
Enable
For a detailed view, select this button to enlarge the Ycoordinate into full screen. The X-coordinates may be enlarged
beyond the edge of the screen in this mode.
This icon displays the X-axis and Y-axis in a full screen.
For a detailed view, this mode allows the user to examine the graph
with the X-axis and Y-axis scaled to the user’s preference. Simply
input the upper range of each axis accordingly.
This function provides the coordinate of a particular peak or
point. After choosing “ENABLE” icon, select the target point.
Then the result will be recorded on the point that the mouse
indicates.
Cursor
With this function selected, clicking on a point will provide the
coordinate information (Wavelength, Intensity) shown on the
status bar at the bottom of the screen. In accordance with the
movement of the mouse, new coordinate information will be
shown.
SEC2000 Spectra System
www.als-japan.com
29
Page 30
SEC2000 Operating Manual
Zoom
Pan
Snap
Shot
② Integration Time, Average, & Boxcar
Integration
Time
This shows an enlarged image of the desired region. You can
observe the graph by zooming in and out of spectrum with the
method mentioned in the “Full scale”. After choosing the
“Zoom” button, click and drag the cursor over the desired
region with the left button of the mouse pressed, and then you
can see the detailed graph in the scoped area.
To see a broader spectrum, choose the “PAN” button. Click and
drag the graph, then X-axis and Y-axis also change following the
graph movement.
Snap Shot makes real-time values halt. When the “SNAP.S”
button is clicked, the window will be stopped and all other
icons wil l bec om e inactiv at ed. I f yo u c l ic k “SNAP.S ”
button again, the icons become active.
Function of spectral signal amplifying. The larger the number, the larger
the signal (increase in detection sensitivity).
If the signal intensity exceeds 16,000 counts, the signal will be saturated
and not be detected. Therefore, the signal should be adjusted below
16,000 counts. In case that the concentrations between samples have
large difference, make the integration time as low as possible.
<Fig. 4-3-14>
Average Function of spectral signal stabilizing. The higher the figure, the more the
signal is stabilized. When the figure is too high, it is hard to get real-time
data. Decide the least figure so the spectral signal of sample does not
move up and down too much. If the real time values of the sample
concentration continue to change very frequently, lower this function to
meet your experiment.
Boxcar Function of spectral signal smoothing. The higher the figure, the more the
signal smoothes. If the figure is too high, the limit of resolution lowers.
Decide the lowest figure so the spectral signal does not vary much. When
you want to separate intensities of more than two nearby wavelengths,
lower this function to meet your experiment.
SEC2000 Spectra System
www.als-japan.com
30
Page 31
SEC2000 Operating Manual
5. Measurement
5-1. Calibration
All spectrometers are delivered to the user after an initial calibration but some can drift
slightly over a period of time and depending on the environment (If the mirror and
diffraction grating are not aligned, please contact us.).
It can be confirmed by the following way whether wavelength calibration is need. Open
the software and turn on the UV lamp. Confirm the following peaks in scope mode. As
shown in Fig. 5-1-1, sharp peaks at 656 nm labeled Dα, are the feature peaks of the
deuterium arc lamp. The wavelength calibration is required when the feature peaks is
gap.
<Fig. 5-1-1> Emission spectrum of the ultraviolet deuterium arc lamp showing characteristic
hydrogen Balmer lines.
5-2. Dark and Reference Measurements
When Dark and Reference are measured, continue with the experiment applying the
same Integration Time value. To block the light towards the detector SEC2000, close the
shutter of the light source.
① The power supply is switched on and turned ON the UV lamp and halogen lamp.
Wait for 20 to 30 minutes until UV lamp is stabilized.
② Put the reference sample into the cuvette holder, and open the light source shutter
SEC2000 Spectra System
www.als-japan.com
31
Page 32
SEC2000 Operating Manual
fully.
③ Adjust the Integration Time to control the maximum value of the Y axis of the graph
(less than 16000).
④ By clicking the “Reference” icon in the program, the spectrum is obtained at the
maximum light intensity. Save the reference spectrum by
“File”/”Save”/”Reference”.
⑤ Close the light source shutter, and click the “Dark” icon in the program to obtain a
spectrum under the minimum conditions without light. Save the dark spectrum by
“File”/”Save”/”Dark”.
⑥ Scope mode is the basic measurement mode. After measuring the spectrum in
Dark and Reference, the “Absorbance”, “Transmittance”, “Reflectance” and
“Irradiance” measurement mode icons become active.
5-3. Sample Measurement
Apply the same condition with which the Reference was measured (the shutter open &
the same Integration Time). Put the sample into the cuvette holder, and close the cover.
Save the sample spectrum by “File”/ ”Save”/ ”Sample”.
5-4. Absorbance/Transmittance Measurement
Absorbance/Transmittance spectra are used to measure how much light is absorbed by
a sample. For most samples, absorbance is linearly related to concentration of the
substance. The software calculates absorbance (A) using Beer’s law and shows the
relationship between absorbance and concentration as below equation:
A=bc
Where A is the absorbance of the solution, is the molar absorptivitymol-1cm-1, b
is the pathlength of radiation through the absorbing medium (cm) and c is the
concentration (mol/L) of the solution.
The absorbance is measured with the following procedure.
SEC2000 Spectra System
www.als-japan.com
32
Page 33
SEC2000 Operating Manual
<Fig 5-4-1>
① First, the light source and detector of SEC2000 are set as shown in Fig. 5-4-1.
② The power supply is switched on and turned ON the UV lamp and halogen lamp.
Wait for 20 to 30 minutes until the lamps are stabilized.
③ Open the software and click “Go” icon, the spectrum signal will appear. On Scope
Mode, put the reference sample into the cuvette holder, and open the light source
shutter fully.
④ Adjust the Integration Time to control the maximum value of the Y axis of the graph
(less than 16,000).
⑤ By clicking the “Reference” icon in the program, the spectrum is obtained at the
maximum light intensity. Save the reference spectrum by “File”/”Save”/
“Reference”.
⑥ Close the light source shutter (Do not close the power switch), and click the “Dark”
icon in the program to obtain a spectrum under the minimum conditions without
light. Save the dark spectrum by “File”/”Save”/”Dark”.
⑦ Put the sample into the cuvette holder, and open the light source shutter fully. Click
the “Absorbance Mode”/”Transmittance Mode” icon to change the Mode. Store
the spectrum by “File”/”Save”/”Processed” or “Sample” to save an absorbance or
transmittance spectrum.
Notes
When absorbance/transmittance is saved by “File”/”Save”/”Processed”, the data of
absorbance/transmittance is displayed only absorbance/transmittance mode. However,
if absorbance/transmittance is saved “File”/”Save”/“Sample”, the saved data is
convertible for other mode.
SEC2000 Spectra System
www.als-japan.com
33
Page 34
SEC2000 Operating Manual
5-5. Reflectance Measurement
Reflection is the change in direction of a wave front at an interface between two
dissimilar media so that the wave front returns into the medium from which it originated.
The reflection of light may be Specular (that is, mirror-like) or Diffuse (that is, not
retaining the image, only the energy) depending on the nature of the interface. The
glossier the surface is, the closer to specular reflection it is.
Reflection is expressed as a percentage (%r) of reflective light quantity from a standard
reference substance:
%r = {(S-D)/(R-D)}100 (%)
Where r is the reflectance of the sample, S is the reflect light amount which means
applied light to the sample surface; R is the reflect light amount which applied light to the
standard reflective material surface; D is the amounts of dark light when light intercepts.
5-5-1. Reflectance Measurement 1
In this measurement, the shape of the sample should be board like and it can be set into
the cuvette holder at 45 degrees.
The reflectance is measured with the following procedure.
<Fig. 5-5-1> Configuration example of reflectance measurement.
① First, the light source and detector of SEC2000 are set as shown in Fig. 5-5-1.
② The power supply is switched on and turned ON the UV lamp and halogen lamp.
SEC2000 Spectra System
www.als-japan.com
34
Page 35
SEC2000 Operating Manual
Wait for 20 to 30 minutes until the lamps are stabilized.
③ Open the software and click “Go” icon, the spectrum signal will appeared. On
Scope Mode, put the standard reflective material into the cuvette holder, and open
the light source shutter fully.
④ Adjust the Integration Time to control the maximum value of the Y axis of the graph
(less than 16,000).
⑤ By clicking the “Reference” icon in the program, the spectrum is obtained at the
maximum light intensity. Save the reference spectrum by “File”/”Save”/
“Reference”.
⑥ Close the light source shutter (Do not close the power switch), and click the “Dark”
icon in the program to obtain a spectrum under the minimum conditions without
light. Save the dark spectrum by “File”/”Save”/”Dark”.
⑦ Put the sample into the cuvette holder, and open the light source shutter fully. Click
the “Reflectance Mode” icon to change the Mode. Save the spectrum by
“File”/”Save”/”Processed” or “Sample”.
5-5-2. Reflectance Measurement 2
In this measurement, the Reflection probe is used.
The reflectance is measured with the following procedure.
Reflectance Standard
<Fig. 5-5-2> Configuration example of reflectance measurement using reflection probe.
Computer
Sample
SEC2000-UV/VIS SEC2000-D
Reflection Probe
Probe holder
SEC2000 Spectra System
www.als-japan.com
35
Page 36
SEC2000 Operating Manual
① First, the light source and detector of SEC2000 are set as shown in Fig. 5-5-2.
② The power supply is switched on and turned ON the UV lamp and halogen lamp.
Wait for 20 to 30 minutes until the lamps are stabilized.
③ Open the software and click “Go” icon, the spectrum signal will appeared. On
Scope Mode, open the light source shutter. Put the reflection probe to the
standard reflective material. Move the probe tip to adjust the reflection light
intensity to maximum. Fix the probe tip to holder by screw, and then the distance
from the tip to measurement surface is fixed.
④ Adjust the Integration Time to control the maximum value of the Y axis of the
graph (less than 16,000).
⑤ By clicking the “Reference” icon in the program, the spectrum is obtained at the
maximum light intensity. Save the reference spectrum by “File”/”Save”/
”Reference”.
⑥ Close the light source shutter (Do not close the power switch), and click the
“Dark” icon in the program to obtain a spectrum under the minimum conditions
without light. Save the dark spectrum by “File”/”Save”/”Dark”.
Put the reflection probe to sample surface, and open the light source shutter fully.
Click the “Reflectance Mode” icon to change the Mode. Save the spectrum by
“File”/”Save”/”Processed” or “Sample”.
5-6. Irradiance Measurement
Have you ever seen the colors of light emitted by the sun, fluorescent lights, white
lights and lamps? These colors look white to our eyes. Red, yellow and green traffic
signal colors are radiation of light. The material emitting light is called the light source.
The emission of light by the light source is called Luminescence.
How is the color of the light source generated? Shall we perform experiments for the
Luminescence Spectrum of the light?
5-6-1. Relative irradiance measurement
① First, connect the detector of SEC2000 and integrating sphere by optical fiber. Set
the standard light source to integrating sphere as shown in Fig. 5-6-1.
② Turn on the power of the standard light source. Wait for 20 to 30 minutes until the
SEC2000 Spectra System
www.als-japan.com
36
Page 37
SEC2000 Operating Manual
lamps are stabilized.
③ Open the software and click “Go” icon, the spectrum signal will appear.
<Fig. 5-6-1> Configuration example of relative irradiance measurement.
④ “Spectrum”/” Reference Color Temp.” Input the color temperature (K) of the
standard light source used for relative irradiance measurements.
⑤ On Scope Mode, Adjust the Integration Time to control the maximum value of the
Y axis of the graph (less than 16,000).
⑥ By clicking the “Reference” icon in the program, the spectrum is obtained at the
maximum light intensity. Save the reference spectrum by “File”/ ”Save”/
“Reference”.
⑦ Close the light source shutter (Do not close the power switch), and click the “Dark”
icon in the program to obtain a spectrum under the minimum conditions without
light. Save the dark spectrum by “File”/”Save”/”Dark”.
⑧ Connect a luminescence sample to integrating sphere (set a LED bulb to
integrating sphere), Click the “Irradiance Mode” icon to change the Mode.
<Fig. 5-6-2> Relative irradiance measurement of LED bulb.
SEC2000 Spectra System
www.als-japan.com
37
Page 38
SEC2000 Operating Manual
⑨ Save the spectrum by “File”/”Save”/”Processed” or “Sample”.