BARNSTEAD|THERMOLYNE FB1300, FB1400 Operation And Repair Manual And Parts List

Page 1
INSTRUCTION MANUAL
Furnace
Type: FB1400
Model: FB1410 M
Brand:
Page 2
BARNSTEAD|THERMOLYNE CORPORATION
Type FB1300 & FB1400
Furnaces
OPERATION AND REPAIR MANUAL
AND PARTS LIST
SERIES 1049 & 1050
Model Numbers
FB1310M, FB1310M-26, FB1314M, FB1315M, FB1318M,FB1310M-33
FB1410M, FB1410M-26, FB1414M, FB1415M, FB1418M, FB1410M-33
Page 3
2
Table of Contents
Safety Information ..............................................................................................................................................3
Important Information....................................................................................................................................3
Warnings ......................................................................................................................................................4
Introduction..........................................................................................................................................................5
Intended Use ................................................................................................................................................5
General Usage..............................................................................................................................................5
Principles of Operation ................................................................................................................................5
General Specifications ........................................................................................................................................6
Declaration of Conformity ............................................................................................................................7
Installation ..........................................................................................................................................................9
Unpacking ....................................................................................................................................................9
Site Selection ................................................................................................................................................9
Electrical Connections..................................................................................................................................9
Operation ..........................................................................................................................................................10
Power Switch..............................................................................................................................................10
Cycle Light..................................................................................................................................................10
Door Safety Switches ................................................................................................................................10
Digital Readout ..........................................................................................................................................10
Push To Set Temperature Button ................................................................................................................11
Temperature Adjustment Knob....................................................................................................................11
Digital Control..............................................................................................................................................11
Sensor Break Protection ............................................................................................................................12
Furnace Loading ..............................................................................................................................................13
Preventive Maintenance....................................................................................................................................14
Troubleshooting ................................................................................................................................................15
Maintenance and Servicing ..............................................................................................................................16
To Replace Heating Element......................................................................................................................16
To Replace Thermocouple..........................................................................................................................19
To Replace Insulation ................................................................................................................................21
To Replace Door Switches ........................................................................................................................22
To Replace Solid State Relay ....................................................................................................................23
To Replace PC Board (Controller) ..............................................................................................................24
Wiring Diagram..................................................................................................................................................25
Exploded View ................................................................................................................................................26
Replacement Parts List ....................................................................................................................................27
Ordering Procedures ........................................................................................................................................28
One Year Limited Warranty ..............................................................................................................................29
Page 4
3
Important Information
This manual contains important operating and safety information. You must carefully read and understand the contents of this manual prior to the use of this equipment.
Your Thermolyne FB1300 Model or FB1400 Model Furnace has been designed with function, reliability, and safety in mind. It is your responsi­bility to install it in conformance with local elec­trical codes. For safe operation, please pay attention to the alert signals throughout the man­ual.
Warnings
To avoid electrical shock, this furnace must:
1. Use a properly grounded electrical outlet of correct voltage and current handling capacity.
2. Be disconnected from the power supply before servicing.
3. Have the door switch operating properly.
To avoid burns:
1. “Caution: Hot Surface. Avoid Contact.” Do not touch the exterior or interior sur­faces of the furnace during use or for a period of time after use.
To avoid personal injury:
1. Do not use in the presence of flammable or combustible materials — fire or explo­sion may result. This device contains components which may ignite such material.
2. Refer servicing to qualified personnel.
Safety Information
Warning
Warnings alert you to a possibility of personal injury.
Caution
Cautions alert you to a possibility of damage to the equipment.
Note
Notes alert you to pertinent facts and conditions.
Hot Surface
Hot surfaces alert you to a possibility of personal injury if you come in contact with a surface during use or for a period of time after use.
Alert Signals
Page 5
4
Please note the following WARNINGS:
WARNING
This warning is presented for compliance with California Proposition 65 and other regulato­ry agencies and only applies to the insulation in this product. This product contains refracto­ry ceramic, refractory ceramic fiber or fiberglass insulation, which can produce respirable dust or fibers during disassembly. Dust or fibers can cause irritation and can aggravate preexisting respiratory diseases. Refractory ceramic and refractory ceramic fibers (after reaching 1000°C) contain crystalline silica, which can cause lung damage (silicosis). The International Agency for Research on Cancer (IARC) has classified refractory ceramic fiber and fiberglass as possibly carcinogenic (Group 2B), and crystalline silica as carcinogenic to humans (Group 1).
The insulating materials can be located in the door, the hearth collar, in the chamber of the product or under the hot plate top. Tests performed by the manufacturer indicate that there is no risk of exposure to dust or respirable fibers resulting from operation of this product under normal conditions. However, there may be a risk of exposure to respirable dust or fibers when repairing or maintaining the insulating materials, or when otherwise disturbing them in a manner which causes release of dust or fibers. By using proper handling proce­dures and protective equipment you can work safely with these insulating materials and minimize any exposure. Refer to the appropriate Material Safety Data Sheets (MSDS) for information regarding proper handling and recommended protective equipment. For addi­tional MSDS copies, or additional information concerning the handling of refractory ceramic products, please contact the Customer Service Department at Barnstead|Thermolyne Corporation at 1-800-553-0039.
SAFETY INFORMATION
Page 6
5
Intended Use
The FB1300 Model and FB1400 Model fur­naces are general purpose laboratory and heat treating furnaces. For optimum element life, observe the following temperature ranges: 100°C (212°F) to 982°C (1800°F) for continuous use, or from 982°C (1800°F) to 1100°C (2012°F) for intermittent use. Continuous use is operating the furnace for more than three straight hours, and intermittent use is operating the furnace for less than three hours.
The unit consists of a heating chamber and a digital controller. See Figure 1 for the overall shape and general features of the unit.
General Usage
Do not use this product for anything other than its intended usage.
Principles of Operation
The furnace chamber is heated by a single three section resistant heater which is embed­ded in a refractory material. The chamber is insulated with a ceramic fiber insulation. The temperature is controlled by an electronic con­trol which provides one temperature setting. The temperature is measured by a thermocou­ple and is registered on a digital display. For safety, door switches are incorporated to remove power from the heating elements when the door is opened. The furnace is supported by the control unit which also houses the electrical connections.
Introduction
Page 7
66
General Specifications
FB1300 Models
Dimensions: (handle not included) Chamber: 4” W x 3.75” H x 4.5” D (10 x 9.5 x 11 cm) Overall : 7.9” W x 13.2” H x 8.5” D (20 x 33.5 x 22 cm)
Weight: 15.4 lb. (7 kg) Electrical Ratings:
Model #
Volts Amps Watts Phase Frequency FB1310M 220-240 4.4 1050 1 50/60 FB1310M-26 220-240 4.4 1050 1 50/60 FB1310M-33 220-240 4.4 1050 1 50/60 FB1314M 100 10.5 1050 1 50/60 FB1315M 120 8.8 1050 1 50/60 FB1318M 208 5.0 1050 1 50/60
Temperature: Operating Range (continuous): 982°C; (intermittent): 1100°C. Environmental Conditions:
Operating: 17°C - 27°C; 20% - 80% relative humidity, non-condensing. Installation Category II (over-voltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664. Altitude limit: 2,000 meters. Storage: -25°C - 65°C; 20% - 80% relative humidity.
FB1400 Models
Dimensions: (handle not included) Chamber: 5” W x 4.25” H x 6” D (13 x 11 x 15 cm) Overall: 9.7” W x 15.4” H x 11.1” D (25 x 39 x 28 cm)
Weight: 23 lb. (10.4 kg) Electrical Ratings:
Model #
Volts Amps Watts Phase Frequency FB1410M 220-240 6.3 1500 1 50/60 FB1410M-26 220-240 6.3 1500 1 50/60 FB1410M-33 220-240 6.3 1500 1 50/60 FB1414M 100 14.4 1500 1 50/60 FB1415M 120 12.5 1500 1 50/60 FB1418M 208 7.2 1500 1 50/60
Temperature: Operating Range (continuous): 982°C; (intermittent): 1100°C.
Environmental Conditions:
Operating: 17°C - 27°C; 20% - 80% relative humidity, non-condensing. Installation Category II (over-voltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664. Altitude limit: 2,000 meters. Storage: -25°C - 65°C; 20% - 80% relative humidity.
Page 8
7
Declaration of Conformity
Barnstead|Thermolyne hereby declares under its sole responsibility that this product conforms with the techni­cal requirements of the following standards (-33 models only):
EMC: EN 50081-1 Generic Emission Standard;
EN 50082-1 Generic Immunity Standard.
Safety: EN 1010-1-92 Safety requirements for
electrical equipment for measurement,control and laboratory use; Part I: General Requirements
EN 1010-2-010 Part II: Particular requirements
for laboratory equipment for the heating of materials
per the provisions of the Electromagnetic Compatibility Directive 89/336/EEC, as amended by 92/31/EEC and 93/68/EEC, and per the provisions of the Low Voltage Directive 73/23/EEC, as amended by 93/68/EEC.
The authorized representative located within the European Community is:
Electrothermal Engineering, Ltd.
419 Sutton Road
Southend On Sea
Essex SS2 5PH United Kingdom
Copies of the Declaration of Conformity are available upon request.
GENERAL SPECIFICATIONS
Page 9
8
Figure 1: Front and Back Views
Thermocouple
Spec. Plate
Furnace Chamber
Power Switch
Push to Set Temperature Button
Setpoint Temperature Adjustment
Cycle Light
Digital Readout
Page 10
9
Unpacking
Unpack furnace from the box and remove packing material from inside furnace cham­ber. The furnace is supplied with a three wire cord and plug.
Site Selection
Install furnace on a sturdy surface and allow space for ventilation. The furnace should be installed at least six inches away from any combustible material.
Electrical Connections
The electrical specifications are located on the specification plate on the back of the fur­nace. Consult Barnstead/Thermolyne if your electrical service is different than those listed on the specification plate. Prior to connecting your furnace to your electrical supply, be sure the ON/OFF switch is in the OFF position.
Installation
Warning
Do not mount on a surface of flammable material. The furnace should be installed at least six inches away from any com­bustible material.
Warning
Use a properly grounded electri­cal outlet of correct voltage and current handling capacity.
Page 11
10
Figure 2: Digital Control
Power Switch
When power is switched ON, the On/OFF power switch will illuminate and the digital dis­play will illuminate.
Cycle Light
The amber cycle light will illuminate whenever the power is being applied to the heating ele­ments. The cycle light will flicker on/off as fur­nace reaches setpoint.
Door Safety Switches
The door safety switches remove power from the heating elements when the door is opened. Open and close the door a few times; note that the amber CYCLE light will be out when door is open. This check must be done when furnace is heating or when cycle light is illuminated. If this condition is not true, consult the Troubleshooting section before proceeding.
Digital Readout
The digital readout continuously displays cham­ber temperature unless the Push To Set Temperature button is depressed. Then setpoint temperature is displayed.
Operation
Warning
Do not use in the presence of flammable or combustible mate­rials — fire or explosion may result. This device contains components which may ignite such material.
To avoid electrical shock, this furnace must have the door switch operating properly.
Hot Surface
Caution: Avoid Contact. Do not touch the exterior or interior sur­faces of the furnace during use or for a period of time after use.
Page 12
11
Push To Set Temperature Button
When this button is depressed, the digital dis­play will indicate setpoint temperature. When the button is released, the actual chamber tem­perature is displayed.
Temperature Adjustment Knob
Turning this knob clockwise will increase set­point temperature. Turning the knob counter­clockwise will decrease temperature.
Digital Control
This digital controller provides accurate control at one single temperature setting. To set tem­perature, simply:
1. Turn Power switch ON.
2. While depressing the Push To Set Temperature button, turn the tempera­ture knob until you reach the desired setpoint temperature as indicated on digital display.
3. Release the Push To Set Temperature button.
The digital display will now indicate the actual chamber temperature. The furnace will heat to the new setpoint temperature. The CYCLE light will remain on until the furnace temperature is within 1°C of the setpoint temperature; then the CYCLE light will flicker on and off as the con­troller maintains the setpoint temperature.
OPERATION
Note
If at any time the Temperature Adjustment knob is turned in either direction, the setpoint will change, even if the Push To Set Temperature button is not depressed. To view the current setpoint temperature, depress the Push To Set Temperature button.
Page 13
12
Sensor Break Protection
This controller provides sensor break protec­tion in the event the thermocouple opens. If an open thermocouple condition occurs, the digital display will indicate 5 degrees or less and the power to the heating element will be shut off (CYCLE light will extinguish).
O
PERATION
Page 14
13
For best results of furnace loading, use less than two-thirds of any dimension of the chamber. Maintain a 3/4" clearance between the load and the sides of the chamber.
If you are heating a number of small parts, spread them throughout the mid­dle two thirds of the chamber.
Keep objects away from thermocouple.
Raise your load up off the furnace floor with small pieces of ceramic or a hearth plate to promote even heating.
Use insulated tongs and mittens when loading and unloading furnace.
Always wear safety glasses.
Furnace Loading
Caution
Do not overload your furnace chamber or allow the load to touch the thermocouple. If the load is to be heated uniformly, it should not occupy more than two-thirds of any dimension of the chamber. Failure to observe these cautions could result in damage to furnace components and/or load.
Page 15
14
Contamination is a major cause of element failure, therefore, when possible, remove the fume forming material before heating (e.g., cleaning cutting oil from tool steel).
The resistance wire is high-grade nickel-chromium. Some chemicals, notably sulphur, halogens, and cyanides, attack this wire at high temperatures, so avoid spilling these chemicals in the furnace or heating them any hotter than necessary. The refractory cement helps to protect the wire, but will not completely immunize it from damage.
All heating elements must be considered expend­able, and replacement is expected; however, rea­sonable care in their use will greatly extend the ser­vice they will give. As the manufacturer has no con­trol over the use or care of the elements, no specif­ic service guarantee can be made.
Housekeeping is vital to your electric furnace— KEEP IT CLEAN! Run your furnace up to 871°C (1600°F) empty occasionally to burn off the conta­mination that may exist on the insulation and ele­ments. Run for approximately two hours with the door slightly open.
Element life is reduced somewhat by repeated heating and cooling. If the furnace is to be used again within a few hours, it is best to keep it at the operating temperature or at a reduced level such as 260°C (500°F).
During normal use, the thermocouple in your fur­nace can become oxidized and cause inaccurate readings; therefore, we suggest that if you regularly use your furnace you should change your thermo­couple once every six months to assure the accura­cy of your controller readings.
Clean by wiping the outside case of the unit with a damp cloth and mild soap solution.
Preventive Maintenance
Warning
Before using any cleaning or decontamination method except those recommended by Barnstead|Thermolyne, users should check with Barnstead|Thermolyne that the proposed method will not damage the equipment.
Warning
Disconnect the furnace from power supply before clean­ing.
Page 16
Problem Possible Causes Corrective Action
The furnace does not heat No power. Check power source and fuses or (CYCLE light does not illuminate). breakers.
Defective electrical hookup. Repair electrical hookup. Thermocouple has oxidized and Replace thermocouple.
opened the circuit. (Open thermocouple is indicated on the display as a temperature of 0-5 degrees).
Controller malfunction. Replace controller. Door switches malfunction. Re-align or replace door safety switches. Defective solid state relay. Replace output relay.
Door switches do not cut Door switches are not functioning. Re-align or replace door switches. power to heating elements.
Slow heatup. Low line voltage. Install line of sufficient size and proper
voltage. (Isolate furnace from other electrical loads.)
Heavy load in chamber. Lighten load in chamber to allow
heat to circulate.
Wrong heating element. Install proper element.
Repeated element burnout. Overheating furnace. Keep furnace under maximum
temperature.
Heating harmful materials. Enclose material in container. Clean up
spills in and on chamber. Ventilate chamber by opening door slightly when heating known harmful reagents.
Contamination present from Replace insulation material. pervious burnout.
Inaccurate temperature Oxidized or contaminated Replace thermocouple. readout. thermocouple.
Poor thermocouple connection. Tighten connections. Improper loading procedures. Use proper loading procedures. Poor ventilation of base. Clear area around furnace base. Thermocouple connections Reconnect thermocouple correctly.
reversed. (Indicated by downscaling or temperature on display.)
Controller malfunction. Replace controller.
15
The Troubleshooting section is intended to aid in defining and correcting possible furnace problems. When using the chart, select the problem category that resembles the malfunc­tion. Then proceed to the possible causes category and take necessary corrective action.
Troubleshooting
Page 17
16
Figure 3: Replacement of a Heating Element
Maintenance and Servicing
To Replace Heating Element
1. Set the furnace on its top. (See Figure
3). Remove thermocouple cover.
2. Remove screw and clamp holding ther­mocouple, then grasp the thermocouple at the bend where it enters the furnace chamber and pull straight back. Retain porcelain feed-through insulator.
3. When the thermocouple tip is clear of the furnace back, bend it out of the way.
4. Remove the screws holding the steel back plate to the case.
5. Remove the steel back plate.
Thermocouple Cover
Warning
Disconnect the furnace from the power supply before servicing. Refer servicing to qualified personnel.
Note
Perform only maintenance described in this manual. Contact an authorized dealer or our factory for parts and assistance.
Note
It is seldom necessary to disconnect the thermocouple from the controller if the thermocouple is in good condition.
Page 18
17
6. Remove the back insulation block by opening the door and gently pushing it out. Support this insulation block while removing it, as it is quite soft and easily crumbled at the edges.
7. Remove bottom cover to obtain access to terminals.
These steps will expose the heating element leads and insulating bushings in the bottom plate of the furnace.
8. Cut the element leads between the ele­ment and the terminal block. (There are two leads.) The element and hearth plate unit may now be removed by pushing it straight back out of the fur­nace. Use care not to damage the chamber insulation when removing the hearth plate and element as it can be reused if it has not been contaminated.
9. Remove the old element lead wire and power wires from the terminal block, and save the glass sleeving for re­installation on the new element leads. Sleeving must be replaced if cracked or brittle!
10. New elements are shipped flat to pro­tect them from damage in shipment, and to save space in storage. They must be formed before installation.
11. Place the element on a flat surface with the smooth (exposed element) side up. Place a board or other straight edge along one row of notches.
12. Gently bend the element along the straight edge. The refractory cement will break along the row of notches.
MAINTENANCE AND SERVICING
Note
The hearth plate contains no heating coil, and may be saved for re-installation if it is in good condition. The hearth plate is replaceable independent of the heating element.
Page 19
18
Make the bend 90°, avoiding exces­sive bending. (The element wires will be exposed at the corner thus formed. This will not affect its life or perfor­mance.)
13. Bend the other side of the element.
14. Place the hearth plate across the open end of the ‘’U” shaped element.
15. Slide the element and hearth plate unit into the chamber, pushing it firmly against the hearth collar. Use care not to damage the soft insulation. Remove any crumbs of insulation that may get between the unit and the hearth collar.
16. Thread the element leads through the ceramic bushings. Bend the leads so they lie close to the refractory plate and the bottom insulation block. (The easiest and safest way to do this is to press the wire flat with a stick or blunt pusher. Do not use a sharp object or nick the wire.)
17. Replace the glass sleeving and bend the lead 3/4 turn around the terminal screw. Cut off the excess wire. Replace power wires on top of ele­ment lead wires and tighten screw. Do not cross the wire over itself around the terminal; this makes it difficult to keep the connection tight and prevents good electrical contact. If you have excess wire, cut it off. Make sure ele­ment lead wires are not touching any other wires.
18. Replace bottom cover of control unit.
M
AINTENANCE AND SERVICING
Note
Nicking or damaging the ele­ment leads will cause prema­ture element failure.
Page 20
19
19. Replace the back insulation block and back plate.
20. Examine the thermocouple, and, if it is good, reinsert it into the chamber. It should extend about 1-1/2" into the chamber. Make sure porcelain insula­tor is in place for the thermocouple to pass through on the steel back plate. Replace clamp and screw. (Excessive scaling, pitting, or cracks are some indications that the thermocouple may need to be replaced.)
21. Replace thermocouple cover.
22. If replacement of back insulation is necessary, carefully redrill hole for thermocouple, using back cover as guide.
23. Reconnect furnace to power supply.
24. Test operation of furnace.
To Replace Thermocouple
1. Set furnace on its top.
2. Remove thermocouple cover.
3. Remove screw and clamp holding the thermocouple, then grasp the thermo­couple at the bend where it enters the chamber and pull it straight back from the furnace. Retain porcelain feed­through insulator.
4. Remove bottom cover.
M
AINTENANCE AND SERVICING
Warning
Disconnect the furnace from the power supply before servicing.
Page 21
20
5) Disconnect the thermocouple from the terminal block by removing the screws on the terminals. Pull the thermocouple through the hole in the furnace base and discard.
6) Insert the new thermocouple into the back of the furnace until the tip extends 1-1/2" into the chamber. Make sure the porcelain insulator extends through the steel back plate to prevent the lead wires from touching metal.
7) Bend the thermocouple sharply so that it lies flat along the back of the furnace.
8) Thread the thermocouple through the hole in the base which has a nylon insulator, replace clamp and screw.
9) Bend the thermocouple sharply toward terminal block.
10) Secure the two yellow wires marked “+” together on the terminal block. Secure the two red wires “-” together on the adjacent terminal. Make sure connec­tions are secure to terminal block.
A polarity test of the thermocouple and lead wire is easily made with the use of a magnet. On chromel alumel thermocouples and lead wire, the non-magnetic wire is positive ( + ) and the magnetic wire is negative ( - ).
11) Replace bottom plate.
12) Replace thermocouple cover.
13) Reconnect furnace to power supply.
14) Test operation of furnace.
M
AINTENANCE AND SERVICING
Note
If the thermocouple touches metal, this could short out the signal, causing the control to display room temperature. This could cause the furnace temper­ature to run away, possibly dam­aging furnace components.
Note
If the furnace temperature moves downward, the thermo­couple leads are reversed.
Page 22
21
To Replace Insulation
1. Remove thermocouple cover.
2. Set furnace on its top and remove
screw and clamp securing thermocou­ple, then grasp thermocouple and remove by pulling it straight back. Retain porcelain feed-through insulator.
3. Remove back plate.
4. Remove bottom plate.
5. Disconnect the element leads and white
insulated wires from terminal block in control section.
6. Disconnect thermocouple leads from
terminal block.
7. Remove four screws holding control
section to furnace chamber.
8. Remove screws from door switch
bracket.
9. Remove the control section from the
furnace chamber. Remove plates. Be sure to note how plates are secured together for reassembly.
10. Remove back piece of insulation by
opening door and pushing it out gently.
11. Remove bottom piece of insulation by
lifting it out.
12. Remove element and hearth plate by
pulling it straight back out of the furnace chamber. (Be careful not to damage elements.)
13. Remove side insulating pieces.
M
AINTENANCE AND SERVICING
Warning
Disconnect furnace from power supply before servicing.
Note
Identify or mark wires discon­nected to ensure proper place­ment and connection when rein­stalling.
Page 23
22
14. To remove top insulating piece and hearth collar, position the furnace on its side. Remove both objects from fur­nace.
15. Reposition furnace on its top. Reinsert new hearth collar and the new top piece of insulation. Insert the new side pieces of insulation last.
16. Reinsert element and hearth plate unit into the chamber, pushing it firmly against the hearth collar. (Be careful not to damage insulation.)
17. Reinsert new bottom piece of insula­tion over hearth plate. (Element leads and ceramic bushings should be exposed above insulation bottom piece.)
18. Thread the element leads and ceramic bushings through the bottom plate. Bend the leads so they lie close to the refractory plate and the bottom insula­tion block. (The easiest and safest way to do this is to press the wire flat with a stick or blunt pusher. Do not use a sharp object or nick the wire.) Secure plate to furnace chamber.
19. Reverse steps 1-9 to reassemble fur­nace.
To Replace Door Switches
1. Turn furnace upside down and remove bottom cover.
2. Disconnect wires from door switches.
M
AINTENANCE AND SERVICING
Page 24
23
Identify or mark wires disconnected from door switches to ensure proper place­ment and connection when reinstalling.
3. Remove two screws from door switches and slide door switches out.
4. Install new door switches to bracket. Place furnace in an upright position. Adjust door switches until a click is heard from the switches, when furnace door is approximately 2" from being completely closed. Secure door switches to bracket.
5. Place furnace upside down. Reconnect the wires to new door switches.
6. Replace bottom cover and turn furnace upright.
7. Reconnect furnace to power supply.
8. Test operation of door switches as described in step d.
To Replace Solid State Relay
1. Turn furnace upside down and remove the bottom cover.
2. Disconnect the wires from the solid state relay. Identify or mark the wires discon­nected to ensure proper placement and connection when re-installing.
3. Remove solid state relay from bottom cover. Note placement of solid state relay.
M
AINTENANCE AND SERVICING
Warning
Disconnect furnace from power supply before servicing.
Warning
Disconnect furnace from power supply before servicing.
Page 25
24
4. Install new solid state relay and recon­nect wires.
5. Replace bottom cover and turn furnace upright.
6. Reconnect furnace to power supply.
To Replace PC Board (Controller)
1. Turn furnace upside down and remove bottom cover.
2. Disconnect wires from PC Board. Identify or mark wires disconnected to ensure proper placement and connec­tion when re-installing.
3. Remove PC Board from bottom cover.
4. Install new PC Board and secure.
5. Reconnect wires identified in step 3 to new PC board.
6. Replace bottom cover and turn furnace upright.
7. Reconnect furnace to power supply.
M
AINTENANCE AND SERVICING
Warning
Disconnect furnace from power supply before servicing.
Page 26
25
Wiring Diagram
Ref. No.
Description
CN1 Control DS1 Pilot Light F1 Fuse F2 Fuse FL1 Capacitor H1 Heating Element R1 Resistor, Pot RY1 Relay, Solid State S1 Switch, Power S2 Switch, Door S3 Switch, Door S4 Switch, Setpoint TB1 Terminal Block TB2 Terminal Block, TC TC1 Thermocouple
J6
S4
R1
CN1
GND
J3 J1
HIAC
J2 J5
LOWAC
J8 J9
-+
J7
SSR
YEL.
+
J4
-
RED
TB2
YELLOW
RED
TC1
S1
1 2
- 4 3 +
RY1
DS1
H1
TB1
S2
L2
G
L1
FL1
FB1310M-33 ONLY
S3
F1
F2
Page 27
26
Exploded View
Page 28
27
Quantity
Key# Description Required Part Number
1 Heating element (100V - FB1314M) 1 EL140X2 1 Heating element (120V - FB1315M) 1 EL44X1 1 Heating element (208V - FB1318M) 1 EL44X3 1 Heating element (240V - FB1310M, -33) 1 EL44X2 1 Heating element (100V - FB1414M) 1 EL186X2 1 Heating element (120V - FB1415M) 1 EL48X1 1 Heating element (208V - FB1418M) 1 EL48X3 1 Heating element (208V - FB1410M, -33) 1 EL48X2 2 Thermocouple (FB1300 all models) 1 TC745X1A 2 Thermocouple (FB1400 all models) 1 TC746X1A 3 Solid State Relay (FB1300, FB1400 all models) 1 RYX54 4 Door Switches (all models) 2 SWX163 5 PC Board (FB1315M, FB1415M) 1 PC745X1A 5 PC Board (FB1310M, -26, FB140M, -26) 1 PC745X2A 5 PC Board (FB1314M, FB1414M) 1 PC745X3A 5 PC Board (FB1318M, FB1418M) 1 PC745X4A 6 Insulation Btm (all FB1300 models) 1 JC44X1 6 Insulation Btm (all FB1400 models) 1 JC48X2 7 Insulation Back (all FB1300 models) 1 JC44X4 7 Insulation Back (all FB1400 models) 1 JC48X1 8 Insulation Top (all FB1300) 1 JC44X3 8 Insulation Top (all FB1400) 1 JC48X3 9 Insulation Hearth Collar (FB1300) 1 HC44X1
9 Insulation Hearth Collar (FB1400) 1 HC48X1 10 Insulation Sides (FB1300) 2 JC460X2 10 Insulation Sides (FB1400) 2 JC48X4
11 Cycle Light (amber) (all 100V & 120V) 1 PLX103 11 Cycle Light (amber) (all 208V & 240V) 1 PLX104
12 Push To Set button 1 SW745X1A 13 Temperature setpoint adjustment (all models) 1 RS745X1A 14 Power Switch (FB1300 100 & 120V models) 1 SWX137 14 Power Switch (FB100 208 & 240V models) 1 SWX138 14 Power Switch (FB1400 100 & 120V models) 1 SWX143 14 Power Switch (FB1400 208 & 240V models) 1 SWX144 15 Door Assembly (FB1300 all models) 1 DR347X1A 15 Door Assembly (FB1400 all models) 1 DR348X1A 16 Element Sleeving (all models) 1 SL745X1 17 Element Sleeving (all FB1300 models) 1 SL59X2 17 Element Sleeving (all FB1400 models) 1 SL745X1 18 Hearth Plate (all FB1300 models) 1 PH44X1 18 Hearth Plate (all FB1400 models) 1 PH48X1 19 Filter (all -33 models) 1 CAX98 20 Fuse Holder (all -33 models) 1 FZX26 21 Fuse 10 amp .25 + 1.25 - (Buss™ Type ABC, 2 FZX30
all -33 models)
Replacement Parts List
Page 29
28
Please refer to the Specification Plate for the complete model number, serial number, and series number when requesting service, replacement parts or in any correspondence con­cerning this unit.
All parts listed herein may be ordered from the Barnstead|Thermolyne dealer from whom you purchased this unit or can be obtained promptly from the factory. When service or replacement parts are needed we ask that you check first with your dealer. If the dealer cannot handle your request, then contact our Customer Service Department at 319-556­2241 or 800-553-0039.
Prior to returning any materials to Barnstead|Thermolyne Corp., please contact our Customer Service Department for a “Return Goods Authorization” number (RGA). Material returned without a RGA number will be refused.
Ordering Procedures
Page 30
29
Barnstead|Thermolyne Corporation warrants that if a product manufactured by Barnstead|Thermolyne and sold by it within the continental United States or Canada
proves to be defective in material or construction, it will provide you, without charge, for a period of ninety (90) days, the labor, and a period of one (1) year, the parts, necessary to remedy any such defect. Outside the continental United States and Canada, the warranty provides, for one (1) year, the parts necessary to remedy any such defect. The warranty period shall commence either six (6) months following the date the product is sold by Barnstead|Thermolyne or on the date it is purchased by the original retail consumer, whichever date occurs first.
All warranty inspections and repairs must be performed by and parts obtained from an authorized Barnstead|Thermolyne dealer or Barnstead|Thermolyne (at its own discre­tion). Heating elements, however, because of their susceptibility to overheating and conta­mination, must be returned to our factory, and if, upon inspection, it is concluded that fail­ure is not due to excessive high temperature or contamination, warranty replacement will be provided by Barnstead|Thermolyne. The name of the authorized Barnstead|Thermolyne dealer nearest you may be obtained by calling 1-800-446-6060 (319-556-2241) or writing to:
Barnstead|Thermolyne
P.O. Box 797
2555 Kerper Boulevard
Dubuque, IA 52004-0797
USA
FAX: (319) 589-0516
E-MAIL ADDRESS: mkt@barnstead.com
Barnstead|Thermolyne’s sole obligation with respect to its product shall be to repair or (at its own discretion) replace the product. Under no circumstances shall it be liable for incidental or consequential damage.
THE WARRANTY STATED HEREIN IS THE SOLE WARRANTY APPLICABLE TO Barnstead|Thermolyne PRODUCTS. Barnstead|Thermolyne EXPRESSLY DISCLAIMS ANY AND ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR USE.
One Year Limited Warranty
Page 31
30
Page 32
31
Page 33
Barnstead|Thermolyne
2555 Kerper Blvd. P.O. Box 797 Dubuque, IA 52004-2241 Phone: 319-556-2241 800-553-0039 Fax: 319-589-0516 E-Mail Address: mkt@barnstead.com
IS O
9001
REGISTERED
Loading...