1. PMP REPLACEMENT KIT
1.1 Introduction
Kit description
The PMP kit contains the 6000 series PMP
R763864
InfoT 386
This installation manual gives som e background information about the R6000 software/PMP compatibility The InfoT 386 handles this issue in detail.
PMP and Software compatibility
The change from V1.xx to V2.02 improved the uniformity alignment from a 1-level to a 3-level correction. Upgrade to V3.xx is necessary for the NEW PMP board.
Soft V1.xx : 1-level uniformity corre ction for ALL color temperatures and Projector White.
Soft V2.02 onwards : 3-level uniformity corrections as follows : 6500°K: 2 levels = grey + full white Projector White : 1 level = grey
1. PMP replacement Kit
Software version Characterisitcs
R6300
R6400
R6400i
R6500
V1.12 V3.01
V1.12 V3.01
V1.12 V2.02 V3.01
V1.12 V2.02 V3.01
Std version
- for new PMP
-3levelunif Std version
- for new PMP
-3levelunif Std version
-3levelunif
- for new PMP
-3levelunif Std version
-3levelunif
- for new PMP
-3levelunif
V1.xx to V2.02 Upgrade
In the new soft V2.02, because of the 3 levels, we ha ve now two tables for corrections: one for 6500K and another one for PW. Both tables MUST be filled in, they cannot remain "empty". An upgrade from V1.xx to V2.02 is the
no values available for these two tables. There are two options when an upgrade is required :
refore not possible because these new tables will never be filled in by the upgrade. There are
R5976881 R6000 SERIES PMP 15/03/2005
1
Page 4
1. PMP replacement Kit
1. The projector is returned to the factory for a complete new 3-level uniformity alignment and software upgrade. When the pro­jector returns, a full re-alignment based on 3 leve ls uniformity is done, beside the upgrade to V2.02. This return can be an opportunity to replace the consumables like polars, analyzers and/or panels. In this case a full TCR re-alignment will be done as well
1. One does the upgrade himself a nd uses the special program CGU625PW.exe to make the new soft compatible with the old 1-level uniformity correction. With this program the 1-level grey uniformity corrections are copied to the new empty tables and the full white corrections are set to a "no correction" status. This all means that the uniformity is the same as before, not improved by the upgrade…
New PMP board
The (first) PMP version R763067 was replaced by the R763864 due to a different mapping of new FPGA’s. This hardware change has implied a change of the software version to V3.xx.
This means that the new PMP board R763864 requires software versions from V3.xx onwards.
V3.xx is not downwards compatible i.e. works only with the new PMP R763864 and does not work with the old PMP! Whenever you replace the "old" R763067 by the newer version R763864 you MUST update the soft to minimum V3.01. The procedure in case you replace the old PMP by the new version will depend on the existing software in the projector. If the software in the projector is lower than V2.02, you have to upgrade to V3.01 and this will necessitate all 4 steps (
hereafter) In case you have to change from a 1-level version to a 3-level uniformity version you can decide to return the projector to the factory
for an upgrade including the 3-level uniformity adjustment. If however you decidenot to return the projector, then,youcan upgrade yourself provided that you
(see later for more details). Without this additional program the upgrade is impossible. When the soft in the projector is V2.02 (thus already with the 3-level uniformity) you just have to upgrade to V3.01. No need to run
the special program.
run the program CGU652PW.exe
as described
1.2 The upgrades
Software files
When you unzip the V3.01 Reality 6500 software you get two files:
Br65.3.01_def.bin: default upgrade
Br65.301.bin: soft upgrade
How to upgrade from V1.xx to V2.02 with OLD PMP (R763067)
1. Unzipping the R6500 V3.01 soft you get two bin. files (see hereafter).
2. When you upgrade to V2.02 you just upgrade the soft with the default switch in the OFF position, not in the ON position as
described in step B hereafter.
3. Run the CGU652PW.exe program
How to upgrade from V1.xx to V3.01 with NEW PMP (R763864)
1. Upgrade of the default program
2. Reboot the CPU.
3. Upgrade the full software from default
4. Run the program
How to upgrade from V2.02 to V3.01 with NEW PMP (R763067)
1. Upgrade of the default program
2. Reboot the CPU.
3. Upgrade the full software from default
2
R5976881 R6000 SERIES PMP 15/03/2005
Page 5
1.3 Blockschematic overview
Overview
1. PMP replacement Kit
There is no real need for an
upgrade from V1.xx to
V2.02 except if the projector
is re- aligned (taking
advantage of the 3 level
uniformity correction)
Send projector to
Barco for upgrade
and re-alignment
3 level uniformity corrections : all active 2 grey levels (for t = 6500° and PW) and Full white at t =6500°)
Result :
NO
Need to Upgrade
from V1.xx to V2.02
in the field ?
YES
SW = OFF
PMP = R763067 ?
YES
PMP needs to be
Replaced by
R763864 ?
YES
Replace PMP by
R763864
in the
field
Actual Soft = V1.xx
?
YES
SW = ON
Upgrade Default
program
Br65.3.01_def.bin
PMP = R763864 ? END
NO
YES
Projector up to date
YES
Send projector to
Barco for PMP
replacement and
upgrade
NO
Actual Soft = V2.02 ?
YES
SW = ONSTEP 1
NO
3 level uniformity corrections : all active 2 grey levels (for t = 6500° and PW) and Full white at t =6500°)
Result :
NO Contact Barco
STEP 2
2 identical ( grey levels for 6500 and PW))
Reboot projector
(Power switch)
Upgrade standard
STEP 3
STEP 4
3 level uniformity corrections :
software
Br65.3.01.bin
Run
CGU652PW.exe
Result :
1 inactive ( full white
Upgrade Default
program
Reboot projector
(Power switch)
Upgrade standard
software
Image 1-1
1.4 Extra info
The CGU652PW program
The program does two things:
1. Initialization of the full white settings in 6500K
2. Re-use the 1-level grey uniformity corrections for 6500K and copy to the Projector White- grey table.
This program does not improve the uniformity !
R5976881 R6000 SERIES PMP 15/03/2005 3
Page 6
1. PMP replacement Kit
CAUTION: VERY IMPORTANTREMARK! BE CAREFUL! This program do es not check the software version in
the p rojector before you start. There is a big risk when you run this program in a projector having already the 3-level uniformity correction.
The final result is that you overwrite the factory set 3 - level corrections. Thefinal result is a worse uniformity !
Summary
If case of doubts, please contact Barco before starting the upgrade. Check the soft of the projector before you start the CGU625PW.exe program. When the projector was upgraded with the above special uniformity program to e.g. V2.02, it is not possible to differentiate it from a