The MDRC-1219 (TS) is a 19” one megapixel color LCD display intended for review of medical images with
a resolution of SXGA (1280*1024) and an aspect ratio of 5:4. The display is not intended for diagnosis.
Thereare2versions:
•MDRC-1219: standard version
•MDRC-1219 TS: version equipped with projective capacitive touchscreen
Warnings, cautions, notes and tips
There are four levels of precautionary or advisory statements that may be used in this user guide. In
descending order of importance, they are:
WARNING: Describes hazards or dangers that might result in personal injury or death.
CAUTION: Describes hazards that could damage the product.
Gives additional information about the described subject.
Gives extra advice about the described subject.
1.1What’s in the box
Overview
•MDRC-1219 (TS) display
•1xUSBcable(2m)
•1x DisplayPort video cable (2 m)
•a set of AC power cords (depending on the region of operation)
•1xuserguide
•1x CD containing MediCal QAWeb Agent and translated user guides
The user guide is available in other languages on www.barco.com/support
WARNING: Read all the important safety information before installing and operating
your monitor. Please refer to the dedicated chapter in this user guide.
WARNING: Sufficient expertise is required to install this equipment. All devices and
complete setup must be tested before taking into operation.
CAUTION: When the display is assembled in the medical system, take care of the fixa-
tion of all cables, to avoid unwanted detachment.
CAUTION: The monitor is not intended to be sterilized.
2. Installation
2.1Cable connections
To connect the cables
1. Connect one or more video source(s) to the corresponding video inputs. Use the appropriate video
cable(s) to do this.
The input source to be displayed can be selected in the OSD menus (see "Input source selection", page
13).
Image 2-1
2. Connect the USB upstream connector to a PC USB host to make use of QAWeb or any of the display
USB downstream connectors (e.g. to connect a keyboard, mouse or other peripheral).
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
7
2. Installation
Image 2-2
3. Connect the mains power input to a grounded power outlet.
Image 2-3
4. Route all cables trough the provided cable routing clip at the back of the display stand.
2.2Display position adju stment
To adjust the display position
You can safely tilt, pivot, raise and lower the display as desired.
8
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
Image 2-4
2. Installation
WARNING: The display must be in its highest position before it c an be properly pivoted.
The standard orientation of the video input is landscape. If you use the display in portrait
mode, make sure you change the orientation of the video input via the screen settings
of the computer.
2.3VESA-mount installation
To mount the display on a VESA arm
The display panel, standard attached to the stand, is compatible with the VESA 100 mm standard.
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
9
2. Installation
1. Unscrew the four fixation screws to detach the panel from the stand.
Image 2-5
2. Use 4 M4 screws to attach the panel to a VESA approved arm. Please respect the following rule to
Enabling the Display Power Management System (DPMS) of your display will optimize its lifetime by automatically switching off the backlight when the display is not used for a specified period of time. By default,
DPMS is enabled on your display, but it also needs to be activated on your workstation. To do this, go to
“Power Options Properties” in the “Control Panel”.
Barco recommends setting DPMS activation after 20 minutes of non-usage.
Use a screen saver to avoid image retention
Prolonged operation of an LCD with the same content on the same screen area may result in a form of
image retention.
You can avoid or significantly reduce the occurrence of this phenomenon by using a screen saver. You
can activate a screen saver in the “Display properties” window of your workstation.
Barco recommends setting screen saver activation after 5 minutes of non-usage. A
good screen saver displays moving content.
In case you are working with the same image or an application with static image elements for several hours
continuously (so that the screen saver is not activated), change the image content regularly to avoid image
retention of the static elements.
Understand pixel technology
LCD displays use technology based on pixels. As a normal tolerance in the manufacturing of the LCD, a
limited number of these pixels may remain either dark or permanently lit, without affecting the performance
of the product. To ensure optimal product quality, Barco applies strict selection criteria for its LCD panels.
To learn more about LCD technology and missing pixels, consult the dedicated white
papers available at w
ww.barco.com/healthcare.
Maximize quality assurance
QAWeb guarantees optimum and stabilized image quality in every private practice.
The front sensor on the MDRC-1219 (TS) works seamlessly with QAWeb to ensure a consistent image
over time. It automatically stabilizes the image from the moment you switch on the display. What’s more,
QAWeb provides you with instant feedback on the status of the display.
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
11
3. Operation
3.2Standby switching
About
•Push the control wheel long (5 sec) to put your display in standby mode
•Push the control wheel short (1 sec) to exit standby mode and activate your display
3.3OSD menu use
To open the OSD menu
Shortly push the control wheel during normal operation to open the OSD menu. If the control wheel is
locked, first unlock it as described in "Control wheel locking/unlocking", page 12.
The OSD main menu comes up in the left top of the screen. If no further actions are taken within the
following 20 seconds, the OSD menu will disappear again (and the keyboard will lock if enabled).
To navigate the OSD menu
•Turn the control wheel (counter) clockwise to scroll through the different menu pages, to change values
or to make selections.
•Push the control wheel to go into a submenu or confirm adjustments and selections.
3.4Power status LED
Overview
The power status of the display is indicated by a LED at the front of the display. Below is an overview of
the different LED color modes:
•Off:Notpowered
•Steady green*: Operational
•Blinking green*: Entering standby mode
•Steady orange: In standby mode
* This default behavior can be changed so that the power status LED is also off when the display is operational or when entering standby mode.
To change the behavior of the power status LED
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Settings menu.
3. Enter the Power Status LED submenu.
4. Change the behavior of the power status LED as desired and confirm.
3.5Control wh eel locking/unlocking
About
To avoid unwanted or accidental activation of the control wheel, a lock mechanism can be enabled. This
mechanism will lock the keyboard automatically, except while using the OSD menus.
12
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
3. Operation
To lock the control wheel
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Settings > Keyboard lock menu.
3. Switch the keyboard lock on or off.
4. Exit the OSD menu to activate the selected option.
To unlock the co ntrol w heel
During normal operation, turn and hold the control wheel counter clockwise for 5 seconds, until the
OSD unlocked message appears.
3.6Input source selectio n
About input source selection
The MDRC-1219 (TS) can have multiple video inputs connected. Switching between the different inputs
can be done easily in the OSD menu.
To select the input source
1. Bring up the OSD main menu.
2. Navigate to the Input selection menu.
3. Select one of the available input sources and confi rm.
3.7Luminance adjustment
To adjust the luminance
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Luminance menu.
3. Set a luminance value as desired and confirm.
3.8sRGB color space
About sRGB color space
The sRGB color space combines a display function and white point selection and is designed to match
typical home and office viewing conditions. It is widely used in most computer applications.
When selecting sRGB,theDisplay function and White point selection options in the Adjustments menu will be disabled.
To select sRGB color space
1. Bring up the OSD main menu.
2. Navigate to the Adjustments menu.
3. Select sRGB and confirm.
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
13
3. Operation
3.9QAWeb presets
About QAWeb presets
Display function, white point selection and ambient light conditions for your display can be applied from
within the MediCal QAWeb application.
When selecting QAWeb,theDisplay function and White point selection options in the
Adjustments menu will be disabled.
To select QAWeb presets
1. Bring up the OSD main menu.
2. Navigate to the Adjustments menu.
3. Select QAWeb and confirm.
The display USB upstream port must be connected to a PC with QAWeb installed on it
before the QAWeb presets can be applied.
3.10Display functions
Display function selection is disabled when sRGB or QAWeb are selected in the Aments menu.
About display functions
Native, uncorrected panels will display all grayscale/color
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
display functions have been defined. These functions emphasize on these parts containing crucial diagnostic information by correcting the native panel behavior.
The available display functions for your MDRC-1219 (TS) are:
•Native: If you select Native, the native panel behavior will not be corrected.
•Gamma 1.8 or 2.2: Select one of these display functions in case the display is to replace a CRT display
with a gamma of 1.8 or 2.2 respectively.
•DICOM: DICOM (Digital Imaging and Communications in Medicine) is an international standard that
was developed to improve the quality and communicat
DICOM display function results in more visible grayscales in the images. Barco recommends selecting
the DICOM display function for most medical viewing applications.
The DICOM display function applies ambient light compensation (ALC) taking the ambient light conditions of your reading room into account. The available reading room options are:
- Office: Selects DICOM function optimized for office conditions (60-180 Lux)
- Operation Room: Selects DICOM function optimized for operating room conditions (300-400 Lux)
levels with luminance increments that are not
ion of digital images in radiology. In short, the
djust-
14
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
The settings of the display must be adapted t o suit the requirements of the visualization
software. In case of doubt, please contact the vendor of the visualization software.
To select a display function
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Display function menu.
3. Select one of the available display functions and confirm.
3.11White point selectio n
White point selection is disabled when sRGB or QAWeb are selected in the Adjustments
menu.
About white point selection
3. Operation
This setting allows you to modify the display white point, used as reference for all other colors to be displayed.
The available white point settings for your display are:
•Native: The native, unmodified color temperature of the LCD panel.
•Bluebase: Simulation of the bluebase film color temperature.
•Clearbase: Simulation of the clearbase film color temperature.
•Programmable: When selecting this setting, you will be able to manually adjust the video gain for the
red, green and blue channel in separate submenus.
To select the white point
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > White point menu.
3. Select one of the available white point presets.
3.12Analog video settings
The following settings are only available when an analog video input source (DVI-A) is
selected.
About analog video settings
When the analog video input source is active, a number of analog video settings becomes available:
•Auto Adjust: The analog video setting will automatically be adjusted
•Geometry: Allows to manually adjust the geometry settings of the analog video (clock frequency, clock
phase, horizontal position, vertical position)
•Level: Allows to manually adjust the contrast and brightness levels of the analog video
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
15
3. Operation
To adjust the analog video settings
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Analog menu.
3. Adjust one of the available analog video settings as desired.
3.13Power save mode
About power save mode
Enabling power save mode on your MDRC-1219 (TS) will optimize the display lifetime by automatically
switching off the backlight when no video signal is detected after approximately 10 seconds.
To enable/disable power save mode
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Settings menu.
3. Enter the Power save submenu.
4. Select On or Off as desired and confirm.
3.14OSD menu language
About the OSD menu language
By default, the OSD menu comes up in English. However, there’s a wide range of other languages available for the OSD menu of your Eonis display.
To select the language of the OSD menu:
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Settings menu.
3. Enter the Language submenu.
4. Select one of the available languages.
3.15OSD menu orientation
About the OSD menu orientation
The orientation of the OSD menu can be changed depending on the orientation of your display (landscape
or portrait).
To change the orientation of the OSD menu
1. Bring up the OSD main menu.
2. Navigate to the Adjustments > Settings menu.
3. Enter the OSD orientation submenu.
4. Select Landscape or Portrait as desired and confirm.
16
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
3.16Factory reset
About factory reset
A factory reset allows you to fully restore the display to its original factory setting.
The MDRC-1219 (TS) does not require any scheduled maintenance or calibration activities. We recommend to use QAWeb with the Barco default tests and frequencies to calibrate and maintain the display,
or to return the display to a Barco approved maintenance organization. In any case of doubts, please
contact Barco Healthcare.
4.2Cleanin g
WARNING: Unplug the power cable from the mains power input before cleaning the
display.
CAUTION: Take care not to damage or scratch the front glass or LCD. Be careful with
rings or other jewelry and do not apply excessive pressure on the f
ront glass or LCD.
CAUTION: Do not apply or spray liquid directly to the display as excess liquid may cause
damage to internal electronics. Instead, apply the liquid to a cleaning cloth.
To clean the display
Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized cleaning product for medical equipment. Read and follow all label instructions on the cleaning product. In case
of doubt about a certain cleaning product, use plain water.
Possible cleaning solutions:
•70% isopropyl alcohol
•1.6% aqueous ammonia
•Cidex® (2.4% glutaraldehyde solution)
•Sodium hypochlorite (bleach) 10%
•“Green soap” (USP)
•0.5% Chlorehexidine in 70% isopropyl alcohol
•Like Cleansafe® optical cleaning liquid
Do not use following products:
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
19
4. Maintenance
•Alcohol/solvents at higher concentration > 70%
•Strong alkalis lye, strong solvents
•Acid
•Detergents with fluoride
•Detergents with ammonia at higher concentration > 1.6%
•Detergents with abrasives
•Steel wool
•Sponge with abrasives
•Steel blades
•Cloth with steel thread
20
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
5. IMPORTANT INFORMATION
5.1Safety information
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or fire hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualified personnel.
Do not expose this apparatus to rain or moisture.
5. Important information
Modifications to the unit:
Do not modify this equipment without authorization of the manufacturer.
Type of protection (electrical):
Monitor with internal power supply: Class I equipment.
Degree of safety (flammable anesthetic m ixture):
Equipment not suitable for use in the presence of a flammable anesthetic mixture
or nitrous oxide.
The equipment shall not be operable when the air oxygen content is above 25%.
with air or with oxygen
Non-patient care equipment
•Equipment primarily for use in a health care facility that is intended for use outside the patient area.
•The equipment may not be used with life support equipment.
•The user should not touch the equipment, nor its signal input ports (SIP)/signal output ports (SOP)
and the patient at the same time.
Mission critical applications
We strongly recommend there is a replacement monitor immediately available in mission critical applications.
Use of electrical surgical knives
Provide as much distance as possible between the electrosurgical generator and other electronic equipment (such as monitors). An activated electrosurgical generator may cause interference with them and
can disrupt the functionality of the display.
Power connection – Equipment with internal power supply
•This equipment must be earthed.
•Power requirements: The equipment must be powered by the AC mains voltage.
•The equipment is intended for continuous operation.
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
21
5. Important information
Transient over-voltage
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd. Ed. or Table BBB.201ofIEC 60601-1-1 for the medical electrical systems.
To maintain compliance with EMC Regulation, use only shielded interface cables for the connection to
peripheral devices.
Power cords:
•Europe: H05VV-F or H05VVH2-F PVC cord with appropriate EU plug.
US and Canada: “hospital grade” cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent receptacle marked hospital only or hospital grade. These instructions need to be marked either on the equip.
or on a tag on the power cord
•Do not overload wall outlets and extension cords as this may result in fire or electric shock.
•Mains lead protection: Power cords should be routed so that they are not likely to be walked upon
or pinched by items placed upon or against them, paying particular attention to cords at plugs and
receptacles.
•The power supply cord should be replaced by the designated operator only at all time.
•Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Grounding reliability
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle.
Liquids and moisture
Never expose the monitor to liquids or moisture.
Never use the monitor near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry
tuborinawetbasement.
The equipment is IP20 compliant.
Moisture condensation
Do not use monitor under rapid temperature and humidity change condition or avoid cold air from airconditioning outlet directly.
Moisture may condense on the surface or inside of the unit, or create a mist residue inside the protection
plate, this is not a malfunction of the product itself, although it may cause damage to the monitor.
If condensation happens, let the monitor stand unplugged until
there is no condensation.
Ventilation
Do not cover or block any ventilation openings in the cover of the set. When installing the device in a
cupboard or another enclosed location, heed the necessary space between the set and the sides of the
cupboard.
Installation
Place the device on a flat, solid and stable surface that can support the weight of at least 3 devices. If you
use an unstable cart or stand, the device may fall, causing serious injury to a child or adult, and serious
damage to the device.
The display has been designed to be used in landscape and portrait position with a tilt of -5° to 22°.
22
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
5. Important information
This apparatus conforms to:
•CE (MDD 93/42/EEC class I product), CE-2014/30/EU
•EN 60601-1:2006 +A11:2011 +A1:2013 +A12:2014
•IEC 60601-1(ed.3), IEC 60601-1(ed.3);am1
•ANSI/AAMI ES60601-1: 2005/(R)2012
•CSA CAN/CSA-C22.2 NO. 60601-1:14
•IEC 60529:1991 + A1: 2000 (Degrees of protection IP code)
•IEC 62471: 2006; EN 62471: 2008 (Photobiological Safety Of Lamps And Lamp System)
Finland: "Laite on liitettävä suojamaadoituskoskettimilla varustettuun pistorasiaan"
Norway: "Apparatet må tilkoples jordet stikkontakt"
Sweden: "Apparaten skall anslutas till jordat uttag"
CAUTION: The enclosure has to be checked upon collision damage, refer to qualified
service personnel.
5.2Environmenta l information
Disposal Information
Waste Electrical and Electronic Equipment
This symbol on the product indicates that, under the European Directive 2012/19/EU governing
waste from electrical and electronic equipment, this product must not be disposed of with other municipal
waste. Please dispose of your waste equipment by handing it over to a designated collection point for the
recycling of waste electrical and electronic equipment. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate these items from other types of waste
and recycle them responsibly to promote the sustainable reuse of materia
For more information about recycling of this product, please contact your local city office or your municipal
waste disposal service.
l resources.
For details, please visit the Barco website at: h
ttp://www.barco.com/en/AboutBarco/weee
Turkey RoHS compliance
Türkiye Cumhuriyeti: AEEE Yönetmeliğine Uygundur.
[Republic of Turkey: In conformity with the WEEE Regulation]
According to the “Management Methods for the Restriction of the Use of Hazardous Substances in Electrical and Electronic Products ” (Also called RoHS of Chinese Mainland), the table below lists the names and
contents of toxic and/or hazardous substances that Barco’s product may contain. The RoHS of Chinese
Mainland is included in the MCV standard of the Ministry of Information Industry of China, in the section
“Limit Requirements of toxic substances in Electronic Information Products”.
零件项目(名称)
Component name
印制电路配件
Printed Circuit Assemblies
液晶面板
LCD panel
外接电(线)缆
External Cables
內部 线路
Internal wiring
金属外壳
Metal enclosure
塑胶外壳
Plastic enclosure
散热片(器)
Heatsinks
电源供应器
有毒有害物质或元素
Hazardous substances and elements
铅
Pb
xooo oo
xooo oo
xooo oo
oo o o oo
oo o o oo
oo o o oo
oo o o oo
xooo oo
汞
Hg
镉
Cd
六价铬
Cr6+
多溴联苯
PBB
多溴二苯
醚
PBDE
Power Supply Unit
风扇
Fan
文件说明书
Paper Manuals
光盘说明书
CD manual
本表格依据SJ/T 11364的规定编制
This table is prepared in accordance with the provisions of SJ/T 11364.
o: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for
this part is below the limit requirement in GB/T 26572.
x: Indicates that this toxic or hazardous substance contained in at least one of the homogeneous
materials used for this part is above the limit requirement in GB/T 26572.
All Electronic Information Products (EIP) that are sold within Chinese Mainland must comply with the
“Marking for the restriction of the use of hazardous substances in electrical and electronic product” of Chinese Mainland, marked with the Environmental Friendly Use Period (EFUP) logo. The number inside the
EFUP logo that Barco uses (please refer to the photo) is based on the “General guidelines of environment-friendly use period of electronic information products” of Chinese Mainland.
10
5.3Biological hazard and returns
Overview
The structure and the specifications of this device as well as the materials used for manufacturing makes
it easy to wipe and clean and therefore suitable to be used for various applications in hospitals and other
medical environments, where procedures for frequent cleaning are specified.
However, normal use shall exclude biological contaminated en
tions.
Therefore use of this device in such environments is at the exclusive risk of Customer. In case this device
is used where potential biological contamination cannot be excluded.
Customer shall implement the decontamination process as defined in the latest edition of the ANSI/AAMI
ST35 standard on each single failed Product that is returned f
investigation to Seller (or to the Authorized Service Provider). At least one adhesive yellow label shall be
attached on the top site of the package of returned Product and accompanied by a declaration statement
proving the Product has been successfully decontaminated.
Returned Products that are not provided with such external decontamination label, and/or whenever such
declaration is missing, can be rejected by Seller (or by the Authorized Service Provider) and shipped back
at Customer expenses.
vironments, to prevent spreading of infec-
or servicing, repair, reworking or failure
5.4Regulatory information
Indications for use
The MDRC-1219 (TS) is intended to be used as
digital mammography) for review and analysis by trained medical practitioners. Caution (USA): Federal
law restricts this device to sale by or on the order of a physician. (Details & exemptions are in the Code
of Federal Regulations Title 21, 801 Part D).
a tool in displaying and viewing digital images (excluding
FCC class B
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
This device has been tested and found to comply with the limits for a Class B digital device, pursuant to
Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This device generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause harmful interference
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
25
5. Important information
to radio communications. However, there is no guarantee that interference will not occur in a particular
installation. If this device does cause harmful interference to radio or television reception, which can be
determined by turning the device off and on, the user is encouraged to try to correct the interference by
one or more of the following measures:
•Reorient or relocate the receiving antenna.
•Increase the separation between the device and receiver.
•Connect the device into an outlet on a circuit different from that to which the receiver is connected.
•Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by the party responsible for compliance could void the
user’s authority to operate the equipment.
Canadian notice
CAN ICES-1/NMB-1
5.5EMC notice
Electromagnetic emissions
The MDRC-1219 (TS) is intended for use in the electromagnetic environment (IEC
specified below. The customer or the user of the MDRC-1219 (TS) should assure that it is used in such
an environment.
60601-1-2 4
th
edition)
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
This MDRC-1219 (TS) complies with appropriate medical EMC standards on emissions to, and interference from surrounding equipment. Operation is subject to the following two conditions: (1) this device
may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
Compliance
Group 1The MDRC-1219 (TS) uses
Class B
Not applicable: consumption less
than 75W
Complies
Electromagnetic environment –
Guidance
RF energy only for its internal
function. Therefore, its RF
emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
The MDRC-1219 (TS) is suitable
for use in all establishments,
including domestic establishments
and those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
•Reorient or relocate the receiving antenna or equipment.
•Increase the separation between the equipment and receiver.
•Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•Consult the dealer or an experienced technician for help.
26
K5902137 (451920612601) MDRC-1219 (TS) 09/03/2017
5. Important information
Electromagnetic immunity
The MDRC-1219 (TS) is intended for use in the electromagnetic environment (IEC 60601-1-2 4thedition)
specified below. The customer or the user of the MDRC-1219 (TS) should assure that it is used in such
an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601-1-2 4
edition (2014)
Test levels
± 8kV contact
±15kVair
th
Compliance level
± 8kV contact
± 15kV air
Electromagnetic
environment –
guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5%U
U
T
40% U
1
(> 95% dip in
T
) for 0.5 cycle
(60%dipinUT)
T
for 5 cycles
70% U
(30%dipinUT)
T
for 25 cycles
<5%U
U
) for 5s
T
(>95% dip in
T
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 1 kV line(s) to line(s)
±2kVline(s)toearth
<5%U
U
) for 0.5 cycle
T
40% U
(> 95% dip in
T
(60% dip in UT)
T
for 5 cycles
70% U
(30% dip in UT)
T
for 25 cycles
<5%U
U
)for5s
T
(>95% dip in
T
Mains power quality
should be that of a typical
commercial or hospital
environment
Mains power quality
should be that of a typical
commercial or hospital
environment
Mains power quality
should by that of a typical
commercial or hospital
environment. If the
user of the MDRC-1219
(TS) requires continued
operation during power
mains interruptions, it
is recommended that
the MDRC-1219 (TS)
be powered from an
uninterruptible power
supply or a battery.
30 A/m30 A/mPower frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
3V/m@150kHzto80
MHz
9to28V/min
communication service
bandsupto6GHz
3V/m @ 150 kHz to
80MHz
9to28V/min
communication service
bands up to 6 GHz
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the MDRC-1219
(TS), including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d=1.2√P
1. is the a.c. mains voltage prior to application of the test level.
should be less
than the compliance
level in each frequency
3
range.
At 80 MHz and 800 MHz, the higher frequency range applies.
Interference may occur in
the vicinity of equipment
marked with symbol:
2. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the MDRC-1219 (TS) is used exceeds the applicable RF compliance level above, the MDRC-1219 (TS) should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the MDRC-1219 (TS).
3. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than3V/m.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Recommended separation distance
The MDRC-1219 (TS) is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer of the user of the MDRC-1219 (TS) can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the MDRC-1219 (TS) as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter
W
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection form structures, object and people.
Separation distance according to frequency of transmitter
4
150kHz to 80MHz
d=1.2√P
80MHz to 800MHz
d=1.2√P
800MHz to 2.5GHz
d=2.3√P
5.6Explanation of symbols
Symbols on the device
On the device or power supply, you may find the following symbols (nonrestrictive list):
Indicates the device meets the requirements of the applicable EC directives.
Indicates compliance with Part 15 of the FCC rules (Class A or Class B)
Indicates the device is approved according to the UL Recognition regulations
Indicates the device is approved according to the UL Demko regulations
4. For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter. Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Indicates this device must not be thrown in the trash but must be recycled,
according to the European WEEE (Waste Electrical and Electronic Equipment)
directive
Indicates Direct Current (DC)
Indicates Alternating Current (AC)
Stand-by
Equipotentiality
Protective earth (ground)
or
Symbols on the box
On the box of the device, you may find the following symbols (nonrestrictive list):
Indicates a device that can be broken or damaged if not handled carefully when
being stored.
Indicates a device that needs to be protected from moisture when being stored.
Indicates the storage direction of the box. The box must be transported, handled
and stored in such a way that the arrows always point upwards.
15
n
xx-yy Kg
-20 °C
Indicates the maximum number of ident
n
other, where “n” is the limiting number.
or
Indicates the weight of the box and that it should be carried with two persons.
xx-yy
xx-yy
or
Indicates that the box should not be cut with a knife, a cutter or any other sharp
object.
Indicates the temperature limits6to which the device can be safely exposed when
Indicates the range6of humidity to which the device can be safely exposed when
being stored.
x
5 %
yyy
106 kPa
Indicates the range6of atmospheric pressure to which the device can be safely
exposed when being stored.
xx
50 kPa
5.7Legal disclaimer
Disclaimer notice
Although every attempt has been made to achieve technical accuracy in this document, we assume no
responsibility for errors that may be found. Our goal is to provide you with the most accurate and usable
documentation possible; if you discover errors, please let us know.
Barco software products are the property of Barco. They are distributed under copyright by Barco NV or
Barco Inc., for use only under the specific terms of a software licen
Barco Inc. and the licensee. No other use, duplication, or disclosure of a Barco software product, in any
form, is authorized.
Trademarks
se agreement between Barco NV or
All trademarks and registered trademarks are property of their respective owners.
Copyright notice
This document is copyrighted. All rights are reserved. Neither this document, nor any part of it, may
be reproduced or copied in any form or by any means - graphical, electronic, or mechanical including
photocopying, taping or information storage and retrieval systems - without written permission of Barco.