BaoLai Medical P9-2, P9-3 Instruction Manual

P9-2/P9-3
成型尺寸:140*210MM 80g双胶纸 双色印刷 粘胶成本
This manual is suitable for P9-2&P9-3 series ultrasonic scaler
P9-2/P9-3
1 Symbol instruction
3 Product structure, scope of application and contraindication
4 Components
5 Technical specifications
6 Usage
7 Sterilization
8 Precautions
9 Maintenance
10 Storage and transportation
11 After-service
12 Environmental protections
13 Declaration
Contents
1
2
7
7
8
9
10
12
13
15
14
15
15
1
IPX0
IPX1
1 Symbol instruction
Trademark
Caution
Class equipment
Use indoor only
Type applied part
Alternating current
Foot switch
Water control switch
Follow instructions for use
Ordinary equipment
Drop-proof
Date of manufacture
Appliance compliance with WEEE directive
Autoclavable
SN
Serial number
Manufacture
2
Power knob
Gear indicator
Water control switch
Handpiece holder
Power switch
Picture1
Cable
Foot switch
Bottle
Cover of the bottle
Power plug
Water filter
2.2 Accessories
2 Acquaintance of the product and its parts
2.1 The top and bottom map of main unit
Scaling tips
Endo chuck
Pictur e2-1:
Applied parts
Endo file
Pictur e2-2:
Applied parts
D2 handpiece
LED
or
D3 handpiece
light guide
Torque wrench
Endo wrench
"O" circle
3
Pictur e2-5
Pictur e2-4
Pictur e2-3
2.3 Connection sketch map
Sketch map for how to connect handpiece Sketch map for how to install the bottle
D2
D3
Picture3
Picture4
4.Assemble the water bottle to the machine
1.Turn the water control switch to the maximum.
2.Disassemble the handpiece
3. Make sure the power supply is connected correctly and the power switch is turned on, and then step on the foot switch.
4
Picture5
Sketch map for how to replace water filter
Picture6
Sketch map for how to install tip
Tighten
Note: The tip is tightened when the wrench slips for 2 circles.
Sketch map for how to install endo file
Loosen
Tighten
Tighten
Picture7
5
Picture9
Align
"O" ring
light guide
LED
lower shell
D3 handpiece
2.4 Sketch map of assembling D3
Picture8
2.5 Sketch map of assembling bottle
upper shell
MAX
Power switch
2
6
Picture10
Picture11
Power knob
Picture12
Water control switch
Picture13
7
3 Product structure, scope of application and contraindication
3.1 Product performance and structure Ultrasonic scaler is composed of electrocircuit, waterway and ultrasonic transducer.
3.2 Intended use
A powered device utilizes a vibrating ultrasonic tip to remove calculus and other accretions from tooth surfaces during dental cleaning, periodontal therapy and root canal treatment.
3.3 Contraindication
3.3.1 The hemophilia disease patients or patients with thrombocytopenia purpura are forbidden to use this equipment.
3.3.2 The patients or doctors with heart pacemaker are forbidden to use this equipment.
3.3.3 The heart disease patients, pregnant women and children should be cautious to use the equipment.
4 Components
Number Description
Main unit
Type
Foot switch
Handpiece
Scaling tip
Scaling tip
Scaling tip
Scaling tip
Scaling tip
Endo file
Endo chuck
01
02
03
04
P9-2/P9-3
F1
S1
D2 or D3
S2
S3
S4
S5
ES1
Torque wrench
Endo wrench
Bottle
Cover of bottle
Instruction manual
Qualified certificate
Warranty card
Packing list
"O" circle
Water filter
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
Φ120mm×19mm
Φ120mm×70mm
Φ16mm×3mm
W1
E
P9-2/P9-3
P9-2/P9-3
P9-2/P9-3
P9-2/P9-3
8
5 Technical specifications
5.1 Performance technical specifications
5.2 The conditions of working environment:
5.2.1 Environmental temperature:+10 to +40
5.2.2 Environmental humidity: 0 to 80%
5.2.3 Atmospheric pressure: 700hPa to 1060hPa
Output power
Weight of main unit
Operating mode
Electric shock protection type
Degree of protection against electric shock
Degree of protection against harmful ingress of water
Degree of safety of application in the presence of a Flammable Anesthetic Mixture with air or with Oxygen or Nitrous Oxide
Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
Continuous operation
3W to 20W
2.5kg
2N
B F
Ordinary equipment (IPX0), Foot switch(IPX1)
Output primary tip vibration excursion
Output tip vibration frequency
Output half-excursion force
Class
Classification 93/42/EEC
Class
a
0~50ml/min
Water flow rate
230V~ 50/60Hz 150mA
Input:
Main unit fuse
F1(250VT 0.3AL) F2(250VT 0.3AL) F3(250VT 1.6AL)
110V~ 50/60H z 300 mA
F1(250VT 0.5AL) F2(250VT 0.5AL) F3(250VT 1.6AL)
9
6 Usage
6.1 Open the packing, make sure that all the parts and accessories are
complete according to the packing list. Take the main unit out of the box
and put it on a stable plane.
6.2 Turn the water control switch to the max based on symbol as shown
in 9.2.1[Note 1]
6.3 Insert the plug of the foot switch to its socket.
6.4 Install the bottle:
Clean the connection interface of the bottle.Install the bottle to the main unit (Picture 3).Put proper amount of pure water or special liquid medicine, such as
Hydrogen Peroxide, Chlorhexidine, Sodium Hypochlorite into the bottle.
Note: The liquid medicine can not be lower than the lower limit to insure
flowing smoothly, and can not be higher than the upper limit to insure safety. Install the cover (Picture 9).
6.5 Connect the handpiece with the cable (Picture 4). Hold the handpiece
in the same gesture as a pen in hand.
6.6 Select a suitable scaling tip as you need and screw it on the
handpiece ( Picture 6).
6.7 Connect the power and press the power switch. The power is
connected when it shows blue color( Picture 10).
6.8 Adjust the power knob as you need, step on the foot switch and
start the machine .(Picture 11)
6.9 Vibration intensity: Adjust the vibration intensity according to the
scaling or endo requirement at any time in the clinical treatment.(Picture 11)
6.10 Water volume adjustment: Step on the foot switch to let the tip vibrate,
and then adjust the water control switch to choose suitable water flowing
out and form spray to cool down the handpiece and clean the teeth
.(Picture 13)
10
7 Sterilization
7.1 Precautions
a) Please pay attention to the handpiece during the process of operation
and sterilization to see whether it is externally broken. No protective oil is
allowed to be painted on the handpiece. b) There are two "O" circles in each handpiece which need to be
sterilized, plugged and unplugged repeatedly.For the purpose of
extending life time, dental lubricant should be used.
6.11 Function selection. Adjust the power knob to select the needed function.
6.11.1 Scaling function
Adjust the power knob until the blue indicator lights.Select a suitable scaling tip according to your need and screw it on the
handpiece tightly by the torque wrench (Picture 6). During clinical scaling, please do not vertically touch the teeth with the
point of tip so as to avoid damage of both teeth and tip. The working frequency is extremely high during operation. A light touch
and a certain to and fro motion will be able to eliminate the tartar without
heating. Overexertion and long time lingering are forbidden.
6.11.2 Endo function (Picture 10)
Adjust the power knob until the green indicator lights.Screw the endo chuck on the handpiece with endo wrench.Loosen the nut of endo chuck.Insert the endo file into the front hole of endo chuck and use endo wrench
to tighten the nut so as to clamp the file. Put the endo file into the patient's root canal slowly, step on the foot switch
and conduct the ultrasonic root canal therapy. Power and water can be
adjusted according to actual needs during clinical treatment.
Note: Don't step on the foot switch before putting the endo file into the root canal.
6.12 Disassemble the scaling tip and unplug the handpiece. Sterilize the
scaling tip, handpiece, endo chuck, endo file and torque wrench.
11
The circles should be replaced once they are broken or over abrasive.
12
8 Precautions
8.1 Performing endo function under the scaling statusthe blue area is forbidden.
8.2 Please use Baolai's S series scaling tip, as the other brand's scaling tip may not be compatible with the device's handpiece, and forcibly using it may result in damaging handpiece beyond repair.
8.3 Please check the water filter routinely ,replace if necessary(Picture 7).
8.4 Please keep the device clean.
8.5 The handpiece, scaling tip, endo chuck, endo file, endo wrench and torque wrench must be sterilized before each treatment.
8.6 Don't screw or unscrew handpiece and scaling tip when stepping on the foot switch.
8.7 The scaling tip much be fastened by torque wrench and there must be fine spray coming from the tip when operating.
8.8 Replace a new scaling tip when it is worn by 2mm.
8.9 Don't twist the scaling tip o rub it.
8.10 Keep the connection of cable and handpiece dry.
8.11 Choosing the liquid according to clinic treatment.
8.12 Don't pull the handpiece forcibly during operation in case damage the handpiece or the cable.
8.13 Don't knock or rub the handpiece.
8.14Turn off the switch and cut off the power source after stopping operation.
8.15Put the liquid into the bottle and close it with its cover after the bottle is properly installed on the main unit.
8.16 Only use the device after the bottle is closed with its cover.
8.17 After stopping operation, let the device continuously work under condition of water getting through for 20 to 30 seconds to clean the water circuit, and then clean the connection of bottle.
8.18 If don't use the device for a long time, please install the dry and clean bottle on the main unit and close its cover, in case the dust fly into the connection of bottle.
8.19 Only Baolai or authorized distributors have the right to maintain, repair or amend the device. Baolai takes the safety responsibility only if the replaced spare parts belong to Baolai and operator properly operate according to the instruction manual.
9.1 Troubleshooting and notes
9 Maintenance
Fault
Solutions
Possible cause
The scaling tip doesn't vibrate and there is no water flowing out when stepping on the switch
The foot switch is in loose contact
Insert the foot switch to its socket tightly
The fuse in the main unit is broken
Contact our dealers or us
The scaling tip doesn't vibrate but there is water flowing out when stepping on the switch
The tip is in loose contact
Screw the tip on the handpiece tightly (Picture 6 )
The connect plug between the handpiece and the circuit board is in loose contact
Contact our dealers or us
Contact our dealers or us
Problem of handpiece
The scaling tip vibrates but there is no spray when stepping on the switch
The water control switch is not on
Turn on the water control switch Note 1
The handpiece generates heat
The water control switch is in a low grade
Turn the water control switch to a higher gradeNote 1
The vibration of the tip becomes weak
The joint of the handpiece and cable has water leakage
The endo file doesn't vibrate
There is noise coming from the endo chuck
The tip hasn't been screwed onto the handpiece tightly
Screw the tip on the handpiece tightly(as showed in Picture 6)
The tip is loose because of vibration
"O" is broken
The screw is loosen
The screw is loosen
The endo chuck is damage
Screw on the tip tightly (as showed in Picture 6)
The tip is damagedNote 2
Change a new one
Change a new one
Change a new one
Tighten it(as showed in Picture 4)
Tighten it(as showed in Picture 4)
If the problem still can't be solved, please contact with local dealer or manufacturer.
13
There is air in the water pipe
Step on the foot switch, disassemble the water tank and assemble it. (Picture 3)
14
9.2 Notes
9.2.1 [Note 1] Adjust the water control switch according to the direction
shown control the water volume
9.2.2 [Note 2] If the scaling tip has been screwed on tightly and there is
fine spray too, the following phenomena show that the scaling tip is damaged: The vibrating intensity and the water atomization degree become weak
obviously. ② During treatment, it produces the sound like "buzz" from the scaling tip.
9.3 If not use for a long time, please make the machine get through the
electricity and water once per month for five to ten minutes.
10 Storage and transportation
10.1 Environmental conditions of storage and transportation:
a) Relative humidity: 0 to 80%
b) Atmospheric pressure: 50kPa to 106kPa c) Environmental temperature: -10 to +50
10.2 Prevent excessive shock and vibration in transportation, be sure to
handle with care and avoid inversion.
10.3 Don't mix with dangerous goods during transportation.
10.4 Avoid the sun, rain or snow during transportation.
10.5 The equipment should be handled carefully and lightly. Be sure that
it's far from the vibration, and install or store in a cool, dry and ventilated place.
10.6 Don't store the machine with the articles that are combustible,
poisonous, caustic and explosive.
15
13 Declaration
The device has been tested and homologated in accordance with
EN 60601-1-2 for EMC.This does not guarantee in any way that this
device will not be effected by electromagnetic interference.
Avoid using the device in high electromagnetic environment.
PA-P9-M-E-1
Rev.A6