Hold Function
PHscan40 pocket pH tester contains two data hold modes. When the Auto-Hold function is enabled, the tester will automatically sense a stable endpoint
reading and freeze it, “HOLD” indicator appears on the display. If the Auto-Hold function is disabled, press HOLD key, the tester will immediately freeze
currently displayed value. Press the key again to resume measuring.
Electrode Care and Maintenance
Since pH electrode is susceptible to dirt and contamination, clean as necessary depending on the extent and condition of use.
AFTER MEASURING:
Rinse the pH electrode in distilled water, store the electrode into the electrode storage solution.
CLEANING THE ELECTRODE:
● Salt deposits: Soak the electrode in warm tap water to dissolve deposits, then thoroughly rinse with distilled water.
● Oil or Grease film: Wash the electrode bulb gently in some detergents and water. If necessary, using the alcohol to clean the electrode bulb, then
rinse with distilled water. Place the sensor in the electrode storage solution for 30 minutes.
● Clogged reference junction: Heat a diluted KCl solution to 60 oC to 80oC. Place the sensing part of the electrode into the heated solution for about 10
minutes. Allow the electrode to cool in some unheated KCl solution.
● Protein deposits: Prepare a 1% pepsin solution in 0.1M of HCL. Place the electrode in the solution for 10 minutes. Rinse the sensor with distilled
water.
REACTIVATING THE ELECTRODE:
If stored and cleaned properly, the electrode should be ready for immediate use. However, a dehydrated bulb may cause sluggish response. To rehydrate the
bulb, immerse the electrode in a pH4.01 buffer solution for 10 to 30 minutes. If this fails, the electrode requires activation.
● Soak the pH electrode in 0.1M HCl for 5 minutes.
● Remove and rinse with deionized water, then place in 0.1M NaOH for 5 minutes.
● Remove and rinse again, and soak in electrode storage solution for 30 minutes.
HOW TO PREPARE THE ELECTRODE STORAGE SOLUTION:
Dissolve 22.365 grams of potassium chloride reagent (KCL) in the 100mL distilled water. Stir the solution until the solution has thoroughly mixed.