Baltech TL-0208C, TL-0212C, TL-0215C Operation Manual

www.baltech.biz
OPERATION MANUAL
infrared thermometer
CONTENT
2 2 2 3 4 4 5 5 6 6 6 6 7 7 7 7 9
9 10 12 13 15 15 16 17 18
1. INTRODUCTION
2. DEVICE DESCRIPTION
2.1 Application
2.2 Technical data
2.3 Delivery Set
2.4 Design and principle of operation
2.5 Marking
2.6 Packing
3. DEVICE APPLICATION
3.1 Operating limitations
3.2 Prestarting procedures
3.2.1 Safety measures
3.2.2 External examination
3.2.3 Preparation
3.2.4 Selection of measurement units and laser pointer
3.2.5 Battery changing
3.3 Work with the device
3.3.1 LCD display and control panel
3.3.2 Operation modes of the device
3.3.3 Field of View (FOV)
3.3.4 Emissivity of a controlled object
4. MAINTENANCE
5. STORAGE AND TRANSPORTATION
6. ACCEPTANCE AND PACKING CERTIFICATE
7. CALIBRATION CERTIFICATE
8. WARRANTY CERTIFICATE
www.baltech.biz
OPERATION MANUAL
infrared thermometer
INTRODUCTION
Tha nk you fo r purch asi ng the BALTE CH TL inf rare d thermometer. The device allows you to perform non-contact one-touch temperature measurement. The built-in laser pointer improves targeting, while backlit LCD display and handy buttons make the device more ergonomic. The non­contact infrared thermometers can be used for temperature measurement of surfaces, that can't be measured with traditional (contact) thermometers (e.g. moving or hard-to­get-to objects, surfaces of pressurized objects).
Before you start using the device read this operation manual carefully.
DEVICE DESCRIPTION
The infrared thermometer BALTECH TL is designed for non­contact temperature measurements of industrial and household objects. The device can be used for temperature monitoring of electrical equipment, conditioning and ventilation systems, industrial and residential buildings, in scientific experiments.
2.1 Application
No.
2.2.1
2.2.3
2.2.6
2.2.7
2.2.8
2.2.12
2.2.14
2.2.15
2.2.16
2.2.17
2.2.18
BALTECH
TL-0215C
-50+1500°C
(-58...+2732°F)
BALTECH TL-0212C
-50+1000°C
(-58...+1832°F)
BALTECH TL-0208C
-50+800°C
(-58...+1472°F)
Characteristics
Value
Temperature measurement range, °C (°F)
Response time, sec
Field of view
Emissivity setting
Power supply, V
Display illumination
Alarm when reaching defined temperature
Defined and current temperature display
Dimensions, mm
Weight, g
0.5 0.5 0.5
±1.0 ±1.0 ±1.0
50:1 50:1 50:1
0.1-1.0 0.1-1.0 0.1-1.0
9 9 9
Max/Min/Avr/ T
Yes Yes Yes
Yes Yes Yes
Yes Yes Yes
Yes Yes Yes
Yes, 10 points
Yes, 10 points
200х127х47 200х127х47 200х127х47
~330 ~330 ~330
2.2 Technical data
Operating temperature range relative humidity, %
2.2.9
0~50°C (32°F~122°F)
10 ~90% RH
2.2.2
±3°C (±5°F) From -50~-20°C (-58°F~-4°F) ±2°C (±3°F) From -20~-100°C (-4°F~212°F)
±2% From
100~-1500°C (212°F~2732°F)
Accuracy
2.2.4
Repeatability, °C (°F)
±1°C (±2°F)
2.2.5
Resolution, °C (°F)
0.1°C (0.1°F)
2.2.11
Laser pointer
Yes, switchable
2.2.13
Memory
Yes, 10 points
Yes ,~6sec
2.2.10
Auto shutdown
REFLECTED RADIATION
TRANSMITTED RADIATION
SELF-RADIATION
Th e In f ra re d th e rm o m et e r pe rf o rms no n - co n ta c t temperature measurement of object surfaces through measurement of thermal radiation. Radiated, reflected and transmitted energies (figure 2.1) are collected and focused on the infrared detector. A built-in digital processor converts the information into temperature data in accordance with the sele cted op erati on mode, an d displa ys it on th e multifunction LCD.
2.3 Delivery set
Figure 2.1
2.4 Design and principle of operation
1 Infrared thermometer 2 Strap 3 Battery (9V, type 006P) 4 Operation manual 5 Carrying case 6 Packing
1 1 1 1 1 1
2.5 Marking
2.6 Packing
laser pointer
LCD display
lens
measurement start button
battery and cover
The device is marked with a nameplate on the side face of the measurement unit. The nameplate contains the device name, measurement range, field of view and manufacturer name.
The device is packed in a plastic carrying case with polyurethane foam inserts.
Device appearance and functions
of main elements are shown on the figure 2.2.
Figure 3.1
DEVICE APPLICATION
3.1 Operating limitations
The device can be used at the environmental temperature from 0 to 50°C and relative humidity from 10 to 90%. A controlled object must be bigger than a size of a spot, which is calculated according to the diagram of FOV (paragraph 3.3.3. of the this operation manual)
The device can't perform measurements through transparent surfaces, for example glass or plastic material. It will measure the temperature of these materials.
Steam, dust and fume or other substances can lead to inaccurate measurement results due to the creation of obstacles for the device's optics.
When operating with the laser pointer, do not direct the laser beam towards eyes or mirror surfaces (figure 3.1).
3.2 Prestarting procedures
3.2.1 Safety measures
It is necessary to observe the following safety measures:
- to protect the device from exposure of electro-magnetic stirri ng, emitted during arc we lding and when us ing induction heaters.
- t o pr otec t the device fr om heat shock (a t a high temperature or sharp temperature change, it is necessary to wait for 30 minutes before you start using the device).
- do not leave the device on or near objects with a high temperature.
- to protect the device from static electricity.
3.2.2 External examination
Before you start using the device, it is necessary to carry out an external examination of the device concerning damages and contamination of the lens and display. If there is contamination, it is necessary to perform maintenance of the device in accordance with paragraph 4 of this operation manual. If there are mechanical damages, make sure, that they will not influence readings, or send the device for repair.
Preparation consists of evaluation of emissivity of controlled objects and to enter these values into the device. The em i ss iv i ty is de te r mi ne d by me an s o f t he ta b le or experimentally (paragraph 3.3.4 of this operation manual). If necessary, one can perform modification of a controlled surface for the purpose of emissivity increasing. It is also necessary to make sure, that the controlled spot size is smaller, than the size of a controlled object at a defined distance to the object.
To select measurement units (°C or °F) and switch on (off) the laser pointer, open the cover in the device handle (figure
3.2) and set the corr espond ing switche s to a pro per position.
If the device screen displays « », open the cover in the device handle (figure 3.2) and change the battery. You can use both alkaline battery, type 6F22 and nickel-cadmium battery or nickel metal hydride battery of the same size.
3.2.3 Preparation
3.2.4 Selection of measurement units and laser pointer
3.2.5 Battery changing
DEVICE APPLICATION
LASER ON/OFF, °C / °F
Figure 3.2
The measured values and the operation mode are displayed on the LCD. The device management is carried out with a push-button control panel. The functions of the display and control buttons are shown on the figure 3.3.
3.3 Work with the device
3.3.1 LCD display and control panel
Laser pointer on/off
Battery Low
Data storage
∆T/Emissivity
Alarm t min. or t max.
Memory
Down button
Scan/ hold/auto
Measurement units
°C or °F
Main temperature
display
Max/min/avg
Defined/recorded
value
Lock button
Up button
Mode button
Figure 3.3
To select an operation mode and set up parameters of the device, use the button «MODE». The selection of the required mode is performed by multiple pressing of a button and through control of the required mode on the device display. The device operation modes and sequence of their selection are shown on the figure 3.4. After the mode is selected, press the measurement start button and start measuring. Release the button to switch off the device. The device will switch off in about 6 seconds.
3.3.2 Operation modes of the device
Operation mode
Parameter Setup
Figure 3.4
1. Scanning and displaying temperature during measurement
2. Maximum temperature during measurement
3. Minimum temperature during measurement
4. Average temperature during measurement
5. The difference between the maximum and the minimum temperature during measurement
6. Recalling the stored measurement results by using or buttons.
7. Setting of a controlled surface emissivity
8. Setting of the highest alarm temperature
9. Setting of the lowest alarm temperature
(The parameters are set with or buttons)
DEVICE APPLICATION
REMARKS
1. MEMORY: To save the values, press the button
«M». To delete all the records, select «DATAO», using the  or  button, and press the button «M».
2. FIXATION: After the operation mode is selected, press the measurement start button to start measuring (figure 2.2). Push the lock button for continuous measurements without having to press the start measurement button. The device will operate continuously in the selected mode without automatic shutdown.
3. In scan mode, the LCD displays both the current temperature in Celsius or in Fahrenheit. The unit will hold the last reading for 6 seconds after the measurement start button is released. When the battery is low, the battery icon will be displayed, but the device will continue to operate.
4. When DATA# flashes, the temperature value on the main display can be recorded in the # memory unit by pressing the button «M».
The field of view is the angle of vision of the device, which is determined by the optics of the device. The field of view is the ratio of the distance from the device lens to the measurement spot* to the measurement spot diameter. The FOV diagram is shown on the figure 3.5.
3.3.3 Field of View (FOV)
The field of view for this device is determined at the distance of 1 meter from the lens. At smaller distances the spot diameter is almost constant and is ~17-19 mm, and at bigger distances the FOV starts decreasing.
*Measurement spot is an area, from which more than 90% of emission is fallen to the detector.
Most of organic matters and painted or oxidized surfaces have the emissivity 0,95 (predefined in the device). But when measuring the temperature of shiny or polished metal surfaces, you can have inaccurate measurement results. To remove this error, set up the emissivity of a controlled object in the device (see the table below and setting description in the paragraph 3.3.2) , or stick a dark scotch tape (friction tape) on the measured surface or coat it with black mat paint. Carry out measurement of the modified surface, when this material has the same temperature, as the controlled object has (in several minutes). If the object surface allows no painting or black tape, you can measure the temperature, using a contact thermometer and pyrometer, after which calculate emissivity as the ratio of pyrometer readings to contact thermometer readings. The emissivity of the most common materials is specified in the table.
3.3.4 Emissivity of a controlled object
DEVICE APPLICATION
Material
Emissivity
Aluminum
Asbestos
Asphalt
Basalt
Brass
Brick
Coal
Ceramics
Concrete
Copper
Dirt
Frozen food
Hot food
Glass (plate)
Ice
Lead
Limestone
Oil
Paint
Paper
Plastic
Rubber
Sand
Leather
Snow
Steel
Textiles
Wood
Water
0,30
0,95
0,95
0,70
0,50
0,90
0,85
0,95
0,95
0,95
0,94
0,90
0,90
0,85
0,98
0,50
0,98
0,94
0,93
0,95
0,95
0,95
0,90
0,98
0,90
0,80
0,94
0,94
0,93
MAINTENANCE
STORAGE AND TRANSPORTATION
Repair and maintenance, not described in this operation manual, must be performed only by qualified trained staff. Clean the device housing with a dry cloth periodically. Do not use abrasives and solvents for the device cleaning. The lens should be cleaned with compressed air jet or soft brush. It is recommended to clean the device housing, using soft jaw or cloth, moistened with soap solution.
The device must be transported and stored in the carrying case. The packed device must be stored in dry heated space at air temperature from 0 to 40 °C and relative humidity to 80%, in the absence of acid and alkali fumes and other aggressive environments. After transportation at negative temperature, wait for 2 hours before starting using the device.
ACCEPTANCE AND PACKING CERTIFICATE
The infrared thermometer BALTECH TL-____№_____ is manufactured according to the specifications and is ready for operation.
Issue date «___»__________201_
Signature ______________ Job position and signature of the manufacturer's representative
The infrared thermometer BALTECH TL-____№_____ is manufactured according to the specifications and is ready for operation.
Issue date «___»__________201_
Signature ______________ Job position and signature of the manufacturer's representative
CALIBRATION CERTIFICATE
WARRANTY CERTIFICATE
Signature/print full name Commissioning date (date, month, year) Responsible person (job position, full name) Department E-mail Fax (country code, area code, №) Phone number (country code, area code, №) Company full name Postal address , index Country Device model (serial number)
No.
Item
Details
8
The warranty period begins from the date the warranty certificate was filled in, but not later than 1 month after the delivery of the equipment. The warranty is not valid, if the customer registration form is not filled in. The guarantee gives the right for free training of 2 specialists in the company BALTECH GmbH (Spengl erstrasse 39-4 1, 23 556 Lübe ck Ge rmany, Tel: +49 (0) 451-370-87-700, e-mail: info@baltech.biz, www.baltech.biz) within coordinated time periods, not later than 6 months after the delivery date.
BALTECH GmbH
www.baltech.biz
Kronsforder Allee 45,
23560 Lübeck Germany
Tel: +49 (0) 451-370-87-700
E-Mail: info@baltech.biz