Badger Meter 4020 User Manual

The COX Series 4000 Flow Computer is an integral part of a complete system from Cox which provides the most accurate flow metering capabilities in the industry. Its state-of-the-art electronics automatically compensate for nonlinearity caused by high fluid viscosity, and nonlinearity inherent with LO FLO meters, thus maintaining the calibrated accuracy of the turbine meters over their entire flow ranges.
• Linearization Program up to 40 calibration points via front panel keypad, RS-485 port, or COX QuickLoad Software
• Compensates for inherent nonlinearity caused by:
- Higher viscosity fluids such as hydraulic oils, diesel fuel, automotive oils
- Lo Flo Series characteristic nonlinearity
• Flow totalization, local and remote
• RS-485 communications port Useful for changing calibration data in the field and interfacing with a central PC
• Direct indication in standard engineering units: GPM, PPM, LPM, etc.
Options for both Models 4010 and 4020:
NEMA 4 housing for wall mount applications
Explosion-proof housings
- Class I, Division I, Groups C and D
- Class I, Division I, Groups E, F, and G
RS-232 serial communications port in separate enclosure
QuickLoad Software (c/w RS485
RS232 Converter)
In addition to all of the key features listed, the Model 4020 Flow Computer also provides:
Temperature compensation Using an RTD temperature input and viscosity vs. temperature data fluctuations in viscosity and specific gravity caused by fluid temperature variations can be automatically corrected
Data acquisition system interface Optically isolated 4-20 mA DC analog output, linearly proportion to flow rate, for connection to chart recorders or other instrumentation
15555 North 79th Place Scottsdale, AZ 85260 tel: (480) 922-7446 fax (480) 948-3610 www.cox-instruments.com
COX 4010 and 4020 Flow Computers
Precision Bulletin 4010/20-FC
©2007- Cox Instruments 830-XX 901TC2
COX 4010 and 4020 Flow Computers
Unit Specifications
Overall Accuracy
±
0.05%
Operating Temperature: 32˚F - 131˚F (0 - 55˚C)
Environmental
Storage Temperature: -40˚F - 158˚F (-40˚ - 70˚C)
Power
Nominal Line Voltage: 100, 110, 220 or 240 VAC (50/60 Hz)
Power Consumption 5 Watts max
Temperature Inputs
Compatible RTD type: 100Ω Platinum
(4020) Only
Lead Wire Compensation: 4 Wire Configuration 2, 3 or 4 Wire
Range: 0-24 VDC, sink only.
Flow Inputs
Thresholds: High = 2.8 - 24 VDC
Low = 0 - 1.3 VDC
Impedance: 5.8 k Six Digit Display
Flow Rate Display 3 Character units of measure identifier
(GPM, LPM, PPH, CCM, etc.) Type: Dry Contact, Form C
Relay Outputs Contact Rating:
10A@115/230VAC/28VDC Type: NPN
Transistor Outputs
150 mA Maximum 30 VDC
Auxiliary Power
(AC powered units only)
Outputs
Voltage: 24 VDC regulated and filtered Current: 0 to 100mA
Analog Output
4-20 mA, 12-27 VDC
ACC: ± .1% F.S. update = 2Hz Frequency Range 0-7.5 kHz Communications RS-485 Multidrop
FLOW COMPUTER APPLICATION
Turbine flowmeter sends digital signal to Flow Computer
100W RTD direct hook-up to Flow Computer (4020 Only)
Flow Computer Linearizes, Temperature Compensate (4020 Only) and calculates corrected Flow Rate & Total
Scaled pulse out to remote totalizer in supervisory area
4-20 mA out, corrected flow rate, to strip chart recorder tracks trends
Alarm relay and transition outputs to activate bell and/or light as needed
For more information, contact COX Instruments